SD 70

Con
GB
Insert the batteries

1 2 3

+ +
- -
-
+ -
+

4 5 6

MEASUREMENT ∞
Press for Press and
Remove protective cap from the probe 3 seconds hold
Press

Automatically turns off after 8 minutes of


inactivity.

off on off

Lock/Unlock Data!

!! Possible slight delay after pressing


before freeze of the data (!).

unlock lock unlock

C. Backlight ON/OFF

F. Store Data

off on off
D. Store Data
1. In lock mode: 2s
After 3 sec. the file no. temporarily appears
on the display, e.g. 01.
Once the storage has reached max. capacity,
2. In measurement mode: 2s (or the data
2s will overwrite the oldest
) files.
3. When the storage is full, it will overwrite the oldest data.
CON CALIBRATION
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

G. System Setup
Characters in RED represent „Flashing symbols“.
Menu selection:
• Re-Calibrate according to individual recommendations! Rinse the electrode
with destilled water after each measurement.
• If temperature is out of range, „ERR“ will appear. Meter automatically returns to Measurement Mode.
2s • If the default reading is more
(latest than 30% out of standard buffer range, „ERR“ will appear.
memory#)
E1. Auto Mode
Remove protective cap from the probe.
Power ON (see A)
(7.00 or 6.86) (dd-mm-yy/24hr or mm-dd-yy/12hr) (°C or °F)
Auto-recognition for : 1413 µS and : 12.88 mS

(31-12 24 or 12-31 12ampm) .....


Immerse the probe
into calibration
To enter: solution (1413 µS). wait

____________________________________________________________________________________________ 3
Note: Icons/Characters in RED meaning "BLINK"
: Hit once / : Press and hold / : Release
Immerse the probe into
calibration solution
(12.88 mS). wait

E2. Manual Mode


Remove protective cap from the probe.
Power ON (see A)

: Limited to 200 ~ 2000 µS; : limited to 2.00 ~ 20.00 mS

Immerse the probe


into calibration
solution.

Adjustment To increase (+): + To decrease (-):

to store wait

Immerse the probe into


calibration solution.


Adjustment: + to increase (+); To decrease (-):

to store wait
SETUP
F. System Setup
Characters in RED represent „Flashing symbols“.
Unfreeze data first (see frontside B) before Menu selection.
3 sec
Recall Data Select Buffer System Set Date and Time
go to l. 1x go to ll. 2x go to lll.

Set Temperature Set Data/Time display To enter.


3x go to lV. 4x type go to V.

I. Recall Data
3 sec
Press for
further files. Exit

Delete stored data.


3 sec 3 sec
Automatic Exit

II. Select Calibration Mode


3 sec

1x

Exit

III. Set Date and Time


Select V. `Set Data/ Time display type‘ first.
3 sec

2x
24 hour clock

24 h Minutes Hours Year Month Day


Exit
am/pm 12 hour clock

AM Minutes Hours Year Day Month


PM
Exit

Adjustment: to increase (+); to decrease (-)

IV. Set Temperature system (°C/°F)


3 sec

3 sec
°C °C °F
3x Exit

V. Set Data/Time display type


3 sec
select select
4x
or 24 hour Exit
am/pm
12 hour clock clock

SD70 Conductivity Specification/Features


Content
1 Con Meter SD 70 in a robust
plastic case with strap
2 Batteries
1 Instruction Manual
Order Code: 19 48 02

Accessories
195 00 17 Battery CR 2032 (2 Batteries required)
19 48 22 Replacement electrode
38 48 01 Beaker, 100 ml

VI. Electrode Replacement


1 2 3

4 5 6

new electrode

Instruction manuals available in further languages, visit www.lovibond.com