1.0 Introduction..................................................................................................................... 1
2.0 Mechanical Interface ...................................................................................................... 1
3.0 Electrical Interface .......................................................................................................... 2
4.0 Data Interface ................................................................................................................. 2
5.0 Communication Protocol ................................................................................................ 3
6.0 Identification (ID) Segment ............................................................................................. 5
7.0 Results Segment ............................................................................................................ 8
6.0.1 For counting purposes, the STX mentioned in Section 5.4.3 is taken as byte 1.
6.1 The Identification segment of each message identifies the type of message and the
specimen the message represents.
6.2 Message Type—Field 1
The Message Type field identifies the message. There are four message types:
WBC Histogram Message— Type "WBC"
RBC Histogram Message— Type "RBC"
PLT Histogram Message— Type "PLT"
Count Data Message— Type "__ _" (3 blanks)
6.14 Dr Name—Field 13
The Dr Name field is an alphanumeric field of twenty two (22) characters enclosed in
double quotation marks.
* Clinical significance has not been established for these parameters; therefore
they are not reportable in the U.S.
There are four sets of measurement units. A problem arises under the following conditions:
The numeric results data, starting in field 17, are always sent as Standard USA units even
though that data may be displayed in a different unit of measure on the instrument. The
displayed unit of measure, sent in field 74, is transmitted for informational purposes only and
may or may not correspond to the Standard USA format of the transmitted numeric results data.
NOTE: A comma (,) follows the last field which immediately precedes the CRC characters.
Note: If a field contains 3 characters plus double quotation marks and a comma, the
total number of bytes used is 6. See the Histogram Message Example.
02 22 57 42 43 22 2C 22 20 43 44 31 37 30 30 22 . "WBC", "CD1800"
2C 22 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 22 2C , "– – – – – – – – – – – –",
34 32 38 38 2C 30 30 2C 22 2D 2D 2D 22 2C 22 31 4288,00, "– – –","1
30 2F 31 39 2F 39 34 22 2C 22 31 36 3A 34 34 22 0/19/94", "16:44"
2C 22 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D 2D , "– – – – – – – – – – – – – – – –" , “
2D 2D 22 2C 22 20 20
20 20 20 20 20 20 20 20 20 20 20 20 20 31 22 2C 1",
22 4D 22 2C 22 20 20 2F 20 20 2F 20 20 22 2C 22 "M" , " / / " , "
20 20 20 20 20 20 20 20 20 20 20 20 20 20 20 20
Note: Each line contains 16 characters starting with 1, not zero. The “02” and “.” bolded on the
first line refer to the STX character. The “03” and “65” bolded on the last line refer to the
ETX and CRC characters respectively.
Note: If a field contains 3 characters plus double quotation marks and a comma, the total
number of bytes used is 6. See the Count Data Message Example.
E7 [checksum]
NOTE: Sample checksum value may
not match this record’s data.
– – – – –, [spare field]
– – – – –, [spare field]
– – – – –, [spare field]
– – – – –, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
–, [spare field]
1, [limits set]
(continued)
0, [sample error/incomplete
aspiration] Field 73
9B [checksum]
NOTE: Sample checksum value