TotalCare® Bed System Service Manual (MAN112 REV 7) Correction

Since the release of the TotalCare® Bed System Service Manual (MAN112 REV 7), changes
have been made that require updating. Listed below are the page numbers and changes that
require updating. Make these changes in your manual. These changes will be included in
the next revision of the manual. If you have any questions, refer to the service manual’s back
cover for contact information.

Page Number Change Required


Cover Replace the cover with the cover in this document.
6-39 Replace the Bed Scale procedure with this:

CAUTION:
Local Country laws or regulations may require mandatory re-
verification of OIML scales that are certified according to the
NAWI Directive.

For OIML scale beds, do as follows:

Fully raise the bed.


Make sure the bed is flat.
Tare the scale.
Use calibrated weights, and do as follows:
1. Put 100 kg on the bed spread between the seat and chest section.
2. Take a weight reading. The reading must be between 99.5 kg and
100.5 kg.
3. If the weight reading is not between 99.5 kg and 100.5 kg, inspect
the weigh frame for interference. Repair as needed and repeat the
step 2. If no interference is found, calibrate the scale, refer to “Ver-
ify Operation of the NAWI Class IIII Scale (NAWI Class IIII
Scale—European Version Only)” on page 6-10.

If any repairs or adjustments were made or if mandatory re-verifica-


tion after inspection is required by local law, do as follows:
Have a qualified and approved person re-verify the scale. Refer to
Verify Operation of the NAWI Class IIII Scale (NAWI Class IIII
Scale—European Version Only)” on page 6-10.

September 30, 2008 154426 REV 1 Page 1 of 2


6-39 continued For non-OIML scale beds, do as follows:
Fully raise the bed.
Make sure the bed is flat.
Zero the scale.
Use calibrated weights, and do as follows:
• Put 200 lb (90.7 kg) on the bed spread between the seat and chest
section.
Take a weight reading. If the reading is between 198 lb (89.8 kg)
and 202 lb (91.6 kg), no further action is required.
• If the weight reading is not between 198 lb (89.8 kg) and 202 lb
(91.6 kg), inspect the weigh frame for interference. Repair as
needed and repeat the weight check. If no interference is found,
calibrate the scale, refer to “Patient Weigh System Calibration
(Non-NAWI Class IIII Scale Only)” on page 4-17.

Test for proper operation of the scale functions and adjustments.

September 30, 2008 154426 REV 1 Page 2 of 2


DRAFT 9/30/08

SERVICE MANUAL
TotalCare®
Bed System
From Hill-Rom

Product No. P1900 (L Model and Earlier)

MAN112 REV 7
DRAFT 9/30/08
Global Headquarters US
Hill-Rom, Inc.
1069 State Route 46 E
Batesville, IN 47006-9167
Tel: 800-445-3720
www.hill-rom.com

US Rental Therapy St. Paul, MN International


Hill-Rom, Inc. Hill-Rom, Inc. Hill-Rom, Inc.
Tel: 800-638-2546 Tel: 651-490-1468 or 800-426-4224 Tel: +1 (0)812 934 8173
www.thevest.com Fax: +1 (0)812 934 7191
www.hill-rom.com
international@hill-rom.com

Australia Belgique/België Canada


Hill-Rom Australia Pty. Ltd. Hill-Rom Medical Services BV Hill-Rom Canada
Tel: +61 (0)2 8814 3000 Tel: +31 (0)347 / 32 35 32 Tel: 800-267-2337
Fax: +61 (0)2 8814 3030 Fax: +31 (0)347 / 32 35 00

Deutschland España
Hill-Rom Shanghai Hill-Rom GmbH Hill-Rom Iberia S.L.
Tel: +86 (0)21 5396 6933 Tel: +49 (0)211 16450 0 Tel: +34 (0)93 685 6009
Fax: +86 (0)21 5383 3136 Fax: +49 (0)211 16450 182 Fax: +34 (0)93 666 5570

France Hong Kong Ireland


Hill-Rom SAS Hill-Rom Asia Ltd. Hill-Rom Ltd.
Tel: +33 (0)2 97 50 92 12 Tel: +852 (0)2297-2395 Tel: +353 (0)1 413 6005
Service: +33 (0)820 01 23 45 Fax: +852 (0)2297-0090 Fax: +353 (0)1 413 6030
Fax: +33 (0)2 97 50 92 00 dublin.sales@hill-rom.com

Italia
Hill-Rom S.p.A. Hill-Rom Japan c/o Hill-Rom Japan
Tel: +39 (0)02 / 950541 Tel: +81 (0)3 5715 3420 Tel: +81 (0)3 5715 3420
Fax: +39 (0)02 / 95328578 Fax: +81 (0)3 5715 3425 Fax: +81 (0)3 5715 3425

Nederland New Zealand Nordic Region:


Hill-Rom Medical Services BV Hill-Rom Australia Pty. Ltd. Sverige, Denmark, Norge
Tel: +31 (0)347 / 32 35 32 Tel: +61 (0)2 8814 3000 Hill-Rom AB
Fax: +31 (0)347 / 32 35 00 Fax: +61 (0)2 8814 3030 Tel: +46 (0)8 564 353 60
Fax: +46 (0)8 564 353 61
se.marketing@hill-rom.com

Österreich Portugal South East Asia


Hill-Rom Austria GmbH Hill-Rom Iberia S.L. Hill-Rom Singapore
Tel: +43 (0)2243 / 28550 Tel: +34 (0)93 685 6009 Tel: +65 (0)6391 1322
Fax: +43 (0)2243 / 28550-19 Fax: +34 (0)93 666 5570 Fax: +65 (0)6391 1324
austria@hill-rom.com

Suisse/Schweiz United Kingdom


Hill-Rom SA Hill-Rom Ltd.
Tel: +41 (0)21 / 706 21 30 Tel: +44 (0)1530 411000
Fax: +41 (0)21 / 706 21 33 Fax: +44 (0)1530 411555
hrch.info@hill-rom

Anda mungkin juga menyukai