MOBILE C ARM UNIT

Brand TECHNIX
Type TCA 6S‐ 9 With RTP 1000
Origin Italy
General  Mobile C Arm with high frequncy monobloc and image intensifier for radiography and fluoroscopy procedure
Application Able to applied both for radiography and fluoroscopy imaging on surgery procedure (urology, digestive and orthopaedi), guidance 
for interventional proccess and brachytherapy.
1 Functioning Features
User Interface : ‐ Membrane keyboard with alphanumeric touchscreen 5,7 "LCD display
    User interface control could be rotate 600
  ‐ Alphanumeric keyboard and trackball for the data input
  ‐ Service program for faults finding
  ‐ Microprocessor management, friendly user interface
I.I field selection :   Electronic zoom selection according to the number of the I.I fields with automatic limitation
    of the radiated field by the iris collimator
Thermal unit : ‐ Check and visualization in real time of thermal units on the display according to the applied load
  ‐ From 100 % to 5 % XR enabled
  ‐ In radiography it is not possible to control the x ray emission any more when the 5 % of 
    the available HUT is reached
Flouroscopy control :  Multi function footswitch
   1. Pedal for continuous, pulsed, and single shot flouroscopy
   2. Storage image on memory by footswitch without interrupting flouroscopy
On Stand : ‐ General magneto thermic switch
  ‐ Key safety switch
  ‐ Emergency pushbutton
Monitor Position : Flat Screen monitor 19", high resolution 625/50, monitor trolley with automatic ambient light sensor 
Safeties : ‐ Filament current
  ‐ Monobloc temperature
  ‐ Overload
  ‐ kV min/max or fault HV
  ‐ Store data check
DICOM : Image Sharing and Verify
2 Operational Modes
Continuous flouroscopy
     Dose control : Automatic or manual
     Small focus : 0,5 mm ( IAE Tube )
     kV variation range : 40 ‐ 110 kV
     mA variation range : 0,5 ‐ 8 mA
     kV ‐ mA correlation : 40 kV ‐ 0,5 mA , 80 kV ‐ 7,6 mA, 110 kV ‐ 8 mA
     Safety timer : ‐ Audible alarm restorable after 5' X Ray
  ‐ X Ray stop after 10' continuous radiation according to IEC 60601‐2‐7
Snapshot digital radiography
     mA variation range : 1 ‐ 12 mA
     XR flash time : < 1s
Anatomic APR selection in flouroscopy
     APR 1 mA low ( pediatric, from 0,7 to 6,6 mA ) : 40 kV ‐ 0,7 mA, 80 kV ‐ 6,4 mA, 110 kV ‐ 6,6 mA
     APR 2 mA lung ( Lungs, from 0,5 to 7 mA ) : 40 kV ‐ 0,5 mA, 80 kV ‐ 7 mA, 110 kV ‐ 5,5 mA
Radiography
     Working technique : 2 points with kV and mAs setting
     Large focus : 1,5 mm ( IAE Tube )
     kV variation range : 40 ‐110 kV
     mA variation range : @ 230 VAC ( 25 mA fixed from 40 kV to 100 kV ), ( 20 mA at 110 kV )
     mAs variation range : @ 230 VAC ( 1‐125 mAs from 40 to 100 kV ), ( 1‐100 mAs from 101 to 110 kV )
     Times range : 0,04 ‐ 5 sec
3 Radiological Data
Generator power in DC current : 3,5 kW ( 100 kV, 35 mA, 91 ms ), mAs < 2,2 at nominal voltage 230 VAC
High frequency inverter : 40 kHz
Max high voltage ( radiography and flouroscopy ) : 110 kVp
Max current in continouos flouroscopy : 8 mA
Max current in flouroscopy "SNAPSHOT" : 10 mA
Max current in radiography : 25 mA @230 VAC, 18 mA @ 115 VAC
Max mAs in radiography : 125 mAs @230 VAC, 90 mAs@ 115 VAC
Exposure time in radiography : 0,03 ‐ 5 s
Max flouroscopy time : HU safety after 21' of flouroscopy @ 110 kV, 5 mA ( 550 W )
Cooling at 30 0 : 300 minutes
4 Monobloc
Anode : Stationary
Anode inclination : 120
Focal spot size : 0,5 ( small ), 1,5 ( large )  
Nominal anode power : 800 W ( small ), 4000 W ( large )
Anode heat storage capacity : 40 kHU/min ( 54 kJ/min )
Monobloc thermal capacity : 590 kJ ( 790 kHU )
Max continuous thermal dissipation : 80 W, 109 HU/sec, 6528 HU/min
Total filtration : > 2,8 mmAl
5 Image Intensifier Tube
Field number : 3
Input nominal diameter : 230 mm
Resolution ( central ) : 48/56/64 lp/cm
Conversion factor : 240/120/60 ( cdm‐2/mRs‐1 )
Contrast ratio : 23:1/25:1/30:1
Integral distortion : 4%/2%/1% 
Antiscatter fixed grid : Ratio 8:1, 36 shutters/cm, focus 80 cm
6 Collimator
Type :   Iris collimator and parallel shutter with cassette holder for spot photo
Iris : ‐ Control by console with adjustable continuous opening until the max
  ‐ Iris automatic limitation on the I.I field selected
Orientable shutters : Control by console for the opening /closing and clockwise/anticlockwise rotation
MOBILE C ARM UNIT

Brand TECHNIX
Type TCA 6S‐ 9 With RTP 1000
7 CCD Camera 1K x 1K
Technology : CCD interline progressive scanning
Active pixels : 1024x1024
Acquisition : Matrix 1024x1024, 10 bits, 25 fps
Contrast resolution : 1024 grey level
Sensitivity : 0,2 lux
Signal‐Noise Ratio : 60 dB
8 Mechanical Characteristics
Vertical motion : 500 mm motorized in 60 sec
Horizontal motion : 210 mm
Panning motion : ± 270˚ manual motion
Orbital motion : 135˚ ( 92˚ to ‐ 43˚ )
Pivotal motion 0 0
: ± 12,5  / ± 9
Free space : 770 mm
Depth : 690 mm
S.I.D : 988 mm
Focus skin distance : 218 mm
Minimum distance from the floor : 172 mm
Distance floor focus : 339 mm
Distance floor skin : 557 mm
Movement : Manual, steering rear wheels with manual control by the operator, swiveling front wheel, stationary brake by hand
Stand wheels diameter : Rear ( double wheels 125x40 mm ), front double wheels 80x30 mm
Trolley wheels diameter : Base trolley ( 5x80x35 mm ) with brake, High configuration trolley ( 4x125x30 mm ) with brake
9 Weight and dimension
Weight Stand : 360 kg
Width : 859 mm
Depth in transport position : 2000 mm
Height in transport position : 1780 mm
10 Digital Image System With 1K Memory
Characteristics
Image acquisition : from digital camera CCD 1024x1024x12 bit
Image storage : 1024x1024x12
D/A converter : SXGA 1024x1024 60Hz
Video input : Digital
Acquisition
Number of images : 80.000 on hard disk
Recursive filter : 2,4,8,16/smart filter/motion detector
Horizontal, vertical image inversion : YES
Contrast and brightness : YES
Grey scale inversion : YES
Post Processing
Spatial filter : YES
Cineloop : YES
Electronic shutter : YES, continuous 1 of 3
Shifting pixels : YES
Electronic zoom : YES
Land marking : YES
Overview : YES, 4/9/16 images
Print on windows printer USB 2 : YES
11 Monitor ( GEV955IMI / EM19TFTI/M‐II )
Panel size/type : 19" TFT active matrix LCD, antiglare, and hard coated, medical  /  19" TFT B/W antiglare, medical
Brightness and contrast control : OSD menu / Front pushbuttons
Resolution : 1280x1024  /  1280x1024
Pixel pitch : 0,294x0,294 mm  / 0,294x0,294 mm
Active area( HxV ) : 376 mm x 301 mm  /  376 mm x 301 mm
Viewing area ( H/V ) 0 0 0 0
: 170 /170   ,   170 /170
Max luminance 2   2
: 1500 cd/m /  1000 cd/m
Contrast ratio : 1000:1  /  900:1
Gray scale : 12 bit LUT  / 10 bitLUT
Calibration : Dicom curve / Dicom curve
Respons time : 20 ms / 18 ms
Video input : VGA,BNC,Svideo  /   DVI,VGA
12 Electrical Data
Voltage : ‐ 230 VAC ± 10 % monophase standart
Frequency : 50 Hz
Standby current : 1 A @ 230 VAC
Line compensation : Automatic
Isolation Class : Class I with applied part type B