The guard ring, shown in view A of figure 3 -15, shunts any leakage currents
to the negative side of the generator. This prevents such current from
flowing through coil a and affecting the meter reading.
If the test leads are open, no current will flow in coil a. However,
current will flow internally through coil and deflect the pointer to
infinity. This reading indicates a resistance too large to measure. When a
resistance, such as R x , is connected between the test leads, current also
flows in coil a; the pointer tends to move clockwise. At the same time,
coil still tends to move the pointer counterclockwise.
Therefore, the moving element, composed of both coils and the pointer,
comes to rest at a position in which the two forces are balanced. This
position depends upon the value of R x , which controls the amount of the
current in coil a. Because changes in voltage affect both coils in the same
proportion, the position of the moving element is independent of the
voltage. If you short the test leads together, the pointer will come to
rest at zero because the current in coil a is relatively large. Since R3
limits the current, the instrument will not be damaged under these
circumstances. The external appearance of one type of megger is shown in
view B of figure 3 -15.
Most meggers you will use are rated at 500 volts; however, there are other
types. Meggers are usually equipped with friction clutches, which are
designed to slip if the generator is cranked faster than its rated speed.
This prevents the generator speed and output voltage from exceeding rated
values. A 1,000 -volt generator is available for extended ranges. When an
extremely high resistance, such as 10,000 megohms or more, is to be
measured, a high voltage is needed to cause enough current flow to actuate
the meter movement.
0
c
!"
c
"
c D
#
"
c 0
As shown in figure 3 -16, the fix ed coils are connected in series and
positioned coaxially (in line) with a space between them. The two movable
coils are also positioned coaxially and are connected in series. The two
pairs of coils (fixed pair and movable pair) are also connected in serie s
with each other. The movable coil is pivot -mounted between the fixed coils.
The main shaft on which the movable coils are mounted is restrained by
spiral springs that restore the pointer to zero when no current is flowing
through the coil. These springs also act as conductors for delivering
current to the movable coils. Since these conducting springs are very
small, the meter cannot carry a high value of current.
Figure 3-16. - Internal construction of an electrodynamometer.