XK3100-B2 INDICATOR

(NEW MOLD)

MANUAL WIM GPS


E-105

XK3100 SERIES

INDICATOR

Thank you for using our company’s XK3100


series indicator. Please read the manual before using
and operate according to this manual. If you have any
problems, please contact with us. We will offer you
the satisfied service.
E-105

Catalog

CHAPTER 1 Summary --------------------------------------------------------------------------- (4)

CHAPTER 2 Main Parameter ------------------------------------------------------------------ (4)

CHAPTER 3 Installation ------------------------------------------------------------------------- (5)

1. The front of indicator ------------------------------------------------------------------- (5)

2. The back of indicator ------------------------------------------------------------------- (5)

3. Load cell connected with indicator ------------------------------------------------- (5)

CHAPTER 4 Calibration ------------------------------------------------------------------------- (6)

1. Weight Calibration ---------------------------------------------------------------------- (6)

(1) Division setting ------------------------------------------------------------------ (6)

(2) Point setting ----------------------------------------------------------------------- (6)

(3) Zero range setting --------------------------------------------------------------- (6)

(4) Max capacity setting ------------------------------------------------------------ (6)

(5) Weight calibration ---------------------------------------------------------------- (6)

2. Calibration parameter modification --------------------------------------------------- (7)

(1) Point modification -------------------------------------------------------------- (7)

(2) Zero range modification --------------------------------------------------------- (7)


CHAPTER 5: RS232 transmission confer (optional) --------------------------------------- (7)
1. RS232 SETTING ---------------------------------------------------------------------------- (7)

2. Data format ----------------------------------------------------------------------------------- (8)

CHAPTER 6 Operation ------------------------------------------------------------------------- (8)

1. Keyboard instruction --------------------------------------------------------------------- (8)

2. Operation instruction --------------------------------------------------------------------- (9)

(1) Tare ------------------------------------------------------------------------------------ (9)

(2) Zero ------------------------------------------------------------------------------------ (9)

(3) Weight mode ----------------------------------------------------------------------- (9)

(4) Counting mode --------------------------------------------------------------------- (9)

(5) Precentum mode -------------------------------------------------------------------- (10)

3. Warning and Error message ----------------------------------------------------------- (10)

CHAPTER 7. Using Ambient and Condition ------------------------------------------------- (10)


E-105

CHAPTER 1 Summary

The XK3100-B2 Indicator is a new product of our company. It is a kind of general weighing
indicator. It adopted the AD IC of famous semiconductor factory in USA. It has high precision,
high stability, high credibility, zero code adjust and widely of the input signal, upper and lower
limit setting and alarm, peak test (selectable). The indicator can be made many kinds of
electronic scale connected with load cells easily such as platform scale, ground scale, vehicle
scale, oil scale and animal scale. It is a perfect indicator which can be used in business
industry and etc.
Main Function of XK3100-B2:
1. With AC/DC power supply, 6V 4Ah no-maintenance battery and high efficiency recharging circuit
2. Low power cost, can be used for 200 hours after recharging(only with one load cell and
not open the backlight)
3. Auto adjust of zero code (tare automatically), input signal range: - 6 mv ----- +21 mv
4. Calibrate with digital keyboard (locked by hardware and software)
5. 6 digits LCD display, 24 mm high of symbol, backlight can be selected Zero auto track
6. Several divisions and point selecting, calibrate with any point
7. Low voltage check, calibration self-check, overload warning and error information

CHAPTER 2. Main parameter

1 Style ------------------------------ XK3100-B2 indicator


2 Precision ------------------------ OIML
3 Display --------------------------- 24 mm high of symbol (Backlight is selectable)
4 Keyboard ------------------------ key-press keyboard with 8 function keys
5 Division -------------------------- Several divisions selecting
6 Tare range ---------------------- 99999
7 Load cell ------------------------- Drive 12 of 700Ω or drive 6 of 350Ω
8 Voltage ---------------------------- AC: 220V ( 10 15 ) 50 Hz 10 W
110V.60Hz
DC: Rechargeable battery (6V, 4Ah)
9 Storage-humidity -------------- ≤70 RH Non-Dew
10 Working-humidity ------------- ≤90 RH Non-Dew
11 Storage-temperature -------- -10 50
12 Working-temperature ------- 0 40
13 Fuse ------------------------------- 1 A
14 Size -------------------------------- 290×170×140 (mm)
15 Weight ---------------------------- 1.75 kg
E-105

CHAPTER 3. Installation
1. The front of indicator

Keyboard
Display

Figure 3-1
2. The back of indicator

Power plug

Load cell interface


Power switch

Figure 3-2
3. Load cell connected with indicator

1) It is 5-pin plug. Figure 3-3 indicate different pins define.


2) It must use shielded cable and the shielded cable must be connected with GND (can
through the GND of AC power plug), otherwise the indicator reads will not be stable.
3) If using the shielded 4-pin cable, you must connect +V, +S and –S, –V. Otherwise it will
affect the indicator.
4) You could turn on the indicator after connecting load cell, Otherwise it will display
abnormally.
+V Shield

-V -IN
+IN

Figure 3-3
Note: +V, -V (1, 2): Load cell’s excitation
+S, -S (1, 2): Load cell’s feedback signal
+IN, -IN (3,4): Load cell’s output signal
Shield (5): the load cell’s shield cable (connected with GND of AC plug)
E-105

CHAPTER 4. Calibration
When you calibrate the indicator, the load cell’s shield cable must connected with GND,
otherwise the indicator reads will have errors. (Figure 3-3)
Calibration should be operated by measure technician is as follows:
1. Weight calibration
STEP OPERATION DISPLAY NOTICE
Power on, press “calibration” and “ Enter” together when self-checking, loosen when hear the
1
buzzer.

pass It will display pass after Self checking and ask to


2
input calibration password

Input calibration ------1 Press ↑ to input data press Enter to confirm;


password “010201” press Tare to modify last bit of password or exit
3
Press Enter d= xx calibration. Password right to enter division setting,
otherwise exit calibration
d= xx Division setting
4 press ↑ d= yy (1, 2, 5, 10, 20, 50, 100, 200); Press Enter to confirm
press Enter dot= x Confirm the division and enter point setting
dot= x Point setting
dot= y Press ↑ to modify decimal digits, at best 5 digits
5 press ↑
ZEroxx confirm the decimal digits and enter zero range
press Enter
setting
ZEroxx Zero range setting
press ↑ Zeroyy Press ↑ to modify the zero range, the most zero
6
tracing range will be 3d ;Press Enter to confirm
press Enter Full Confirm the setting and enter capacity setting
Full Full capacity setting(DOS prompt and original setting
display in turn)
7 Press ↑ xxxx Press ↑ to modify data, press Enter to confirm
Press Enter LoAd present data and enter next bit, after finish 6 bit will
enter weight calibration
LoAd Weight calibration
Add poise
LoAd Set the poise on the platform
Press ↑
8 Press ↑ and Enter to enter poise weight, if the
Press Enter
present data, press Enter , calibration will be finished
Press Enter
xxxx when the Tare signal is on if off, finish data entering
E-105

Note: 1. When calibrate, tare flag will be acted as stablization.

2. Press Tare If you want to quit without save settings. It keeps the old.

3. Power off the indicator after calibration, check the data have been saved or not. If not

saved, please check the shoot ring is connected credibility or not.

2. Calibration parameter

STEP OPERATION Y NOTICE

1 Press CAL in normal status till it shows PASS


PASS Display“PASS”to ask input the password to modify
2
calibration parameter
Input calibration ------1 Press ↑ to enter data, press Enter to confirm ;
password press tare to modify the last bit of password or
3 “010201” exit;
Press Enter dot= x Password right to modify the parameter, otherwise
exit the setting
dot= x Point setting
Press ↑ dot= y press ↑ to modify decimal digits, at best 5 digits
4
confirm the decimal digits and enter zero range
Press Enter ZEroxx setting
ZEroxx Zero range setting
Press ↑ ZEroyy Press ↑ to modify the zero range, the most zero
5
tracing range will be 3d
Press Enter Confirm the setting and exit modification

CHAPTER 5. RS232 transmission confer


(Optional)
1.RS232 SETTING

STEP OPERATION DISPLAY INSTRUCTION


Switch on, press “Set” and “Enter” when the indicator is self-checking. Stop press the
1
keys when hearing the buzzer rings.
CLOSE Display the COM condition after self-checking:“CLOSE”
“ –Ser- ”
2 “CLOSE”means COM is closed, no data transmitting;
“ -SEr- ”means RS232 is in transmission condition;
“ -Stb- ”means RS232 will in transmission condition
E-105

after weighing data stable;


“ -Etb- ”Press“Enter“ to transmit data
Press ↑ -SEr- Press ↑ to choose the RS232 working condition.
3 Press Enter Press Enter to confirm the current working condition.
Entering the baud rate setting.
Press ↑ 2400 Display the current Baud rate
4800 Press ↑ to choose the baud rate, the option baud
4 Press Enter rate is (1200,2400.4800,9600)
Press Enter to confirm the set baud rate, exit
setting mode, enter the normal work mode.

2. The indicator use currency RS232 UART signal every 10BIT is a data frames, the format of
frames as follows

Bit Bit Bit Bit Bit Bit Bit Bit Bit Bit

1 2 3 4 5 6 7 8 9 10
BIT1 start bit
BIT2~BIT9 data bit
BIT10 stop bit

Data format Every message include 6 bytes, they are following:


No. 1: D0~D7 —— 0FFH Message Flag
No. 2: D0~D2 —— Decimal point 0-5
D3~D4 —— Current mode: 00 - weighting; 01- counting; 10 - percentum;
D5 —— 1 means weight is negative, otherwise is positive
D6 —— 1 means weight is stable, otherwise is unstable
D7 —— 1 means weight is overflow, otherwise is normal
No. 3: D0~D7 —— BCD1 LSB
No. 3: D0~D7 —— BCD2
No. 5: D0~D7 —— BCD3 MSB
No. 6: D0~D7 —— Unit for weight: 1 - lb; 0 - Kg;

CHAPTER 6. Operation
1. Keyboard instruction:
(1) Cali. To modify some calibration parameter.
(2) Mode To select function: Weighing, Counting & percentum.
(3) Unit To select unit: kg or lb.
(4) Set To select up and low limits range in weighing model.
To sample the weight in counting and percentum model.
(5) Zero To make weight zero when nothing on the platform but the display is not zero.
Press this key 2-4 seconds, the backlight will be switched on or off.
(6) Tare To delete the tare.
(7) ↑ To input number.
E-105

(8) Enter To confirm

2. Operation instruction:
When power on,the indicator will be in weighing model. The “Zero” flag and “kg” flag
disappear. You can press Mode to select model: weighing, counting, percentum.

Tare:
Press Tare to make the weight window display “Zero” after the weight is stable. The “Tare”
flag

disappears. The weight is Net. Remove the poise, the displays is negative. Press Tare again,
the displays is zero. The “Tare” flag disappear.
Note: (1). Tare function can set and set continuously only the weight is over zero and stable.
(2). Tare range: 99999 (if the display data is over 99999, not include the decimal, tare
function is ineffective)

Zero:
Press Zero to make display zero.(Zero range: <=4%FS, it is invalid under tare state.)

Weighing mode
Select Unit: Press Unit to select the unit between “kg” and “lb”. The corresponding flag
disappears.

Setting Alarm Function:


1. Up limit setting--Press Set will display the up limit and the signal “-HH-” circularly. Press
Enter to confirm and setting the low limit. Press ↑ to set the number. In the process of
inputting the number, you can press ↑ to increase the recent digit. Press Enter to confirm
recent digit and set the next digit. If the digit is the last one you may press Enter to set the
low limit.
2. Low limit setting--It will display the set weight low limit and the signal “-LL-” circularly when
you begin to set it. The operation is the same as setting up limit. It will go into the alarm
method setting after the low limit setting finished.
3. Alarm method setting—It means in the range alarm when the scale displays “-IN-”. It means
out of the range alarm when the scale displays “-OUT-”. It means no alarm when the scale
displays “-NO-”. You may press ↑ to switch. Press Enter to confirm.
Press Set to exit from the setting when in the process of setting. The set value is invalid.

Counting mode
Press Set to sample weight in the counting model. It displays “SAP X” (X is the sample
number).
1. Put enough samples on platform and press ↑ to select samples number: 10, 20, 50,100 &
200,500,1000. Press Enter confirm.
2. The window displays “LOAD-C” after the sample number is selected. Press Enter to finish
sample.
3. There are two instance of unit weight lacking:
a. When the weight window displays “-SLAC-”, it means the unit weight of sample is less
than 80% of division. You still can count but the counting may be inaccurate. The signal
will disappear after about 3 seconds.
b. When the weight window displays “-CSL-”, it means the unit weight of sample is not
enough at all. Please press Set to re-sample. Press Enter to exit and return weighting.
Press Set again it will exit from the sample state.

Percentum mode
1. Press Set in the percentum state to display “LOAD-P”. There are two ways to sample:
a. Put enough samples on the platform and then press Enter to finish sample.
b. Press Unit then the scale displays “000000”. Press ↑ and Enter to input the weight
value.
E-105

2. When the sample weighed less than 0.1% FS., the window displays “-CSL-”. It means the
sample is not enough. You should re-sample. Press Set to re-sample. Press Enter to exit
return weighing.
Press Set again it will exit from the sample state

Caution: The up and low limit range and alarm ways will be saved, different unit VS different
alarm range and ways.

3. Warning and Error message:


(1) The weight window displays “-----” with continuously alarm sound when the weight over
100%FS+9d, if the ADC overflows, the weight window displays “—Adc—”with continuously
alarm sound. For these instance, the weight should be taken away.
(2) If the voltage of the indicator is low, the window displays “—Lo—” while the window
displays zero The display will recover to normal when it is loaded . Under this
condition, you can use the indicator in a short time, but you should recharge the
indicator as soon as possible.
(3) The load cell signal over±6mV when power on it will display “HHHHH” “LLLLL”
(4) The weight show“UNSTA” if the scale is unstable when power on please check the
load cell connection either credibility or in free state.
(5) Display“—SYS—” when power on, the calibration data is wrong please calibrate again
or repair
(6) Please extend the zero range if can not enter sleep state or return back once enter
sleep state

CHAPTER 7. Using Ambient and Condition


1. The indicator is not used for the surroundings as follows:
(1). At libation and rock places.
(2). At sunshine places.
(3). At high temperature, much dust and moist places.
2. GND well, ensure safety to use.
3. Be care to charge the battery in time when use DC power supply and set off the power when
do not use and do not charge
4. Washing with cauterant is forbidden.
5. Indicator without water and mixture.
6. Weighting indicator is precision measuring instrument. Nation calibration measuring
department will responsible for the adjustment of theindicator.