Water connection
1930
1284
Drain connection
710
Drain connection
180
104
46T-5 = 677 692 46T-5 = 677
174 Max 265 46T-4 = 1906,5
850 Anti-tip bracket, to be fastened to the wall for single door model. 46T-5 = 2046,5
Steam connection
Condensate connection female G1/2"
Female G1/2"
6
The measurements refer to both single and double doors
7
1. Dist. Water, 1280 mm above floor, 2 m flexible hose female G1/2" incl. In delivery.
THIS DRAWING CONTAINS CONFIDENTAL INFORMATION AND REMAINS THE PROPERTY OF GETINGE DISINFECTION AB WITHOUT WHOSE
2. Warm water, 1280 mm above floor, 2 m flexible hose female G1/2" incl. In delivery.
3. Cold water, 1280 mm above floor, 2 m flexible hose female G1/2" incl. In delivery.
4. Electrical conn. 500 mm above floor, 5 m incl. In delivery.
5. Drain, Ø75mm 180mm above floor
PERMISSION MAY NOT BE COPIED, SHOWN OR HANDED TO A THIRD PART OR USED AT THE BASIS OF ALTERNATE DESIGNS
Clean side
140
110
75
Drain connection 5
185
460
4
300
200 / 250
500
140
Recommended service area,
0
450
500
Solid side Drain cooling Connection in wall Max height 46T-4 = 1299 ISO 2768-mK JK 2010-01-15 Layout Drawing
Reviewed
Yes 75mm 200mm 46T-5 = 1369 46-T
MAN 2011-10-31
No 75mm 250mm Approved Replacing Replaced by
Yes 50mm 200mm
No 50mm 250mm Min.200 (When the drain is wall connected) MAN 2011-10-31
Status Dwg.no. Sheet Rev
Approved Production
Ljungadalsgatan 11
PO Box 1505 SE 351 15 Växjö Sweden File no
503277600.idw
5032776 1 /1 A