CENTRAL GAS SUPPLY SYSTEMS

COMPONENTS AND ACCESSORIES


3

TABLE OF CONTENTS

INTRODUCTION P.004

PRODUCT RANGE OVERVIEW P.008

TECHNOLOGY OVERVIEW P.010


-- Supply boards P.010
-- Switchover Boards P.010
-- Pressure Regulators P.012
-- Quality Standards P.013
SPECIAL SERVICES P.014
-- Ceodeux meditec services P.014
-- Electronic solutions …gas monitoring system P.015

HOW TO CHOOSE A SUPPLY SYSTEM P.016

PRODUCTS P.018
-- Switch over board P.019
-- Supply boards - emergency modules P.024
-- High pressure regulators P.028
-- Line pressure regulators P.032
-- Section valves P.038
-- Non return valves P.042
-- Terminal units P.044
-- Connexions P.048
-- Accessories P.054

ISO
13485 All Rotarex regulators are produced in Europe in accordance
with international standards (ISO; CGA....) and are guaran-
QUALITY
0029 ASSURANCE
teed to provide safe and reliable performance in operation.
All locations are ISO 13485.
4 INTRODUCTION

CEODEUX MEDITEC COMPLETE PIPELINE SYSTEM

Patient therapy and safety is our main focus in the design, installation,
commissioning and operation of medical gas pipeline systems

QUANTITY OF SUPPLY

Pipeline systems are designed in such a way sufficient medical gas capac-
ity can be ensured in order to provide the required flows of gases for the
intended number of patients to be treated at any time.

IDENTITY OF SUPPLY

All products included in our medical gas supply system are designed, pro-
duced and controlled for a specific medical gas. All points to which the user or
technical staff can connect are gas specified. There is no risk of misconnection.

CONTINUITY OF SUPPLY

This is achieved by installing, as a minimum, duplex components and provid-


ing additional means of supply provision in the event of failure of the primary
and secondary supply systems.
Means for connection of ultimate external and transportable sources are also
considered at different stages..

QUALITY OF SUPPLY

Our pipeline distribution systems maintain quality of supplied medical gases


thanks to appropriate selection of material, components cleanliness and
internal filtration equipment.
Products designs ensure safe supply by minimizing ignition risks, by avoid-
ing patient poisoning risks and by guaranteeing very good performances for
accurate therapies.

QUALITY FIRST...
CEODEUX MEDITEC products are designed, validated, produced and con-
trolled during their whole lifetime according to a very stringent quality
system which exceeds requirements of ISO 13485 standard.

DESIGN

All our products comply with local and international standards. Some
reference standards:
Medical gas pipeline systems ISO 7396-1:2007 Medical gas pipeline systems Part 1: Pipelines for compressed medical gases and vacuum
ISO 10524-1:2006 Pressure regulators for use with medical gases - Part 1: Pressure regulators and pressure
regulators with flow-metering devices
Products design and validation ISO 10524-2:2006 Pressure regulators for use with medical gases - Part 2: Manifold and line regulators
ISO 10524-4:2008 Pressure regulators for use with medical gases - Part 4: Low pressure regulators
Materials selection and cleanliness ISO 15001:2011 Anesthetic and respiratory equipment. - Compatibility with oxygen

Product design is done through a risk assessment process and is combining PRODUCTION
proven technologies and innovation to reach the best performance levels.
All products are produced in a controlled environment in order to maintain
cleanliness
VALIDATION An individual traceability is ensured with individual serial number
Products are individually controlled and data are recorded for a perfect
Product validation process guarantees that products are developed and product management.
produced to withstand the complete lifecycle (installation, commission-
ing, normal service and maintenance)
ALL RIGHTS OF CHANGE RESERVED
INTRODUCTION 5

THE MAIN MEDICAL GASES IN PIPELINE SYSTEMS…

Symbol Name Application Source Type


Oxygen is one of the most extensively used gases for respiratory -- Compressed gas cylinder
O2 Oxygen therapy and life-support and is additionally used in anesthetic -- Liquid tank
procedures. -- Concentrator
Medical air is mainly used in respiratory therapy as a power source -- Compressor
Air Air for patient ventilators, and for blending with oxygen. It is also used -- Compressed gas cylinder
as the driving gas for nebulized drugs and chemotherapy agents.
N2O Nitrous Oxide Nitrous oxide is used for anesthetic and analgesic purposes, being -- Liquefied gas cylinder
mixed with air, oxygen, and nebulized agents.
CO2 Carbon Dioxide Carbon dioxide is used less commonly now as a respiratory -- Liquefied gas cylinder
stimulant, and for insufflation during surgery.
Surgical air (of medical air quality) is also used, at a higher pressure, -- Compressor
Air Surgical Air to power a variety of surgical tools and other devices such as -- Compressed gas cylinder
tourniquets.
N2 Nitrogen As an alternative, nitrogen can be used for the same purpose as -- Compressed gas cylinder
surgical air.

SOME DEFINITIONS OF OUR CORE BUSINESS

3 main categories of pressure regulators have to be considered into a medi-


cal gas supply systems.

1. MANIFOLD PRESSURE REGULATORS

Manifold pressure regulators are used to reduce high cylinder pressure to a


lower pressure within a source of supply of a medical gas pipeline system.
They are mainly integrated into supply and switchover boards.

2. LINE PRESSURE REGULATORS

Line pressure regulators are used to reduce the pressure supplied by


manifold pressure regulators or by cryogenic vessels to the lower pressure
required at the terminal units of medical gas pipeline systems. They are
intended to be used for inlet pressure up to 30bar.

3. CYLINDER PRESSURE REGULATORS

A cylinder pressure regulator is used to reduce high cylinder pressure to a


lower pressure suitable for use with medical equipment or for delivery of
gas directly to a patient.

It is often coupled to a device which controls the flow, such as a flow control
valve or a fixed orifice.

Cylinder pressure regulators are also used in pipeline systems as transport-


able source for maintenance purpose or ultimate source supply.

ALL RIGHTS OF CHANGE RESERVED


6 INTRODUCTION

ALPI-LINE
MEDICAL GAS SUPPLY SYSTEMS
PURGE GASES COLLECTION LINE
VENTED OUTSIDE BUILDINGS

PRIMARY SOURCE LINE

ALPI-MULTI
Cylinders Extensions ALPI-LINK
Pigtail Extension - Extension

ALPI-SWITCH
Switch over Board

ALL RIGHTS OF CHANGE RESERVED


INTRODUCTION 7

I.E: According to ISO 7396-1


Double stage pipeline system

DISTRIBUTION LINE
-- Toward ALPI-REG Line Pressure
Regulator Double-stage distribution
systems
-- Toward ALPI-DIST Area Shut-Off Valve
Single-stage distribution systems

EMERGENCY SOURCE LINE


ALPI-NRV
LP Non-return Valve

ALPI-SAFE
Emergency Module
ALPI-NRV AF 10
HP Non-return Valve ALPI-MSV
Main Section Valve

ALPI-LINK
Pigtail Manifold - Extension

Flexible Hose
Connection Cylinder - Extension
Cylinder Holder
Cylinder fixation

ALL RIGHTS OF CHANGE RESERVED


8 PRODUCT RANGE OVERVIEW

SWITCH OVER BOARD


AUTOMATIC VERSION

ALPI-SWITCH M826  P. 019 ALPI-SWITCH M825  P. 019 ALPI-SWITCH M820  P. 020 ALPI-CEN P. 022
Type Double stage line regulator Single stage line regulator Basic version Single stage
Inlet Pressure 200 bar / 2900 psi 200 bar / 2900 psi 200 bar / 2900 psi 200 bar / 2900 psi
Outlet Pressure 10 / 5 bar 10 / 5 bar 10 / 5 bar 10 bar
Flow Rate 80 / 50 Nm3/h (N2) 80 / 50 Nm3/h (N2) 80 / 50 Nm3/h (N2) 50 Nm3/h (N2)
Raw Brass Raw Brass Raw Brass
Material Raw Brass
Plastic Protection Cover Plastic Protection Cover Plastic Protection Cover

MANUAL VERSION

ALPI-SWITCH MAN  P. 020 ALPI-CEN MAN P. 022


Type Basic version Single stage
Inlet Pressure 200 bar / 2900 psi 200 bar / 2900 psi
Outlet Pressure 7 bar 10 bar
Flow Rate 50 Nm3/h (N2) 50 Nm3/h (N2)
Raw Brass
Material Raw Brass
Plastic Protection Cover

SUPPLY BOARDS - EMERGENCY MODULES

ALPI-SAFE M830  P. 024 ALPI-SAFE M831 P. 024 ALPI-MOD P. 026


Type Single stage Single stage Single stage
Inlet Pressure 200 bar / 2900 psi 200 bar / 2900 psi 200 bar / 2900 psi
Outlet Pressure 7 bar 7 bar 10 bar
Flow Rate 50 Nm3/h (N2) 50 Nm3/h (N2) 50 Nm3/h (N2)
Raw Brass Raw Brass
Material Raw Brass
Plastic Protection Cover Plastic Protection Cover

HIGH PRESSURE REGULATORS

BSC 290/390 P. 028 DM200 GD P. 030 DM200 TGD P. 031


Type Diaphragm - Cartridge Diaphragm - Single stage Diaphragm - Single stage
200/300 bar
Inlet Pressure 200 bar / 2900 psi 200 bar / 2900 psi
2900/4350 psi
4 /8 bar
Outlet Pressure 10 bar 20 bar
58 / 116 psi
Flow Rate 7 / 25 Nm3/h (N2) 100 Nm3/h (N2) 250 Nm3/h (N2)
Material Chrome Plated Brass Raw Brass Raw Brass

ALL RIGHTS OF CHANGE RESERVED


PRODUCT RANGE OVERVIEW 9

LINE PRESSURE REGULATORS

ALPI-REG 1 P. 032 ALPI-REG 2 P. 034 DC50 MED P. 036 DM20 / DM20 GD P. 037
Type Single second stage Dual second stage Diaphragm Diaphragm
Inlet Pressure 10 bar / 145 psi 10 bar / 145 psi 50 bar / 725 psi 20 bar / 290 psi
8/15 bar 10 bar / 145 psi
Outlet Pressure 5 bar / 72 psi 5 bar / 72 psi
116/217 psi GD : 12 bar / 290 psi
98 Nm3/h (N2)
Flow Rate 50 Nm3/h (N2) 50 Nm3/h (N2) 150/300 Nm3/h (N2)
GD : 124 Nm3/h (N2)
Material Chrome Plated Brass Chrome Plated Brass Raw Brass Raw Brass

SECTION VALVES

ALPI-MSV P. 038 ALPI-DIST P. 040


5 / 10 bar 5 / 10 bar
Working Pressure
72 / 145 psi 72 / 145 psi
Cv 0.35 0.35
Ext: Chrome Plated Brass Ext: Chrome Plated Brass
Material
Int: Raw Brass Int: Raw Brass
Type 3x Ball Valves 1x Ball Valve
Handwheel 1/4 turn 1/4 turn

NON RETURN VALVES TERMINAL UNITS

ALPI-NRV M855 P. 042 ALPI-NRV AF10 P. 043


Working Pressure 25 bar / 362 psi max. 200 bar / 2900 psi max.
Material Raw Brass Raw Brass PLUG-IN SYSTEM P. 044

CONNEXIONS

ALPI-MULTI P. 048 ALPI-LINK P. 050 CYLINDER PIGTAIL P. 052 CYLINDER P. 053


CYLINDER EXTENSION INTERMEDIATE PIGTAIL FLEXIBLE HOSE

ACCESSORIES

PRESSURE P.54 CONTACT P. 056 BA12 P. 058 GAS CYLINDER P. 059


SENSOR GAUGE ALARM BOX HOLDER

PROTECTION P. 059 OMEGA PLATE P. 060 MAINTENANCE P. 061


COVER KIT FIXING ELEMENT TOOL CASE

ALL RIGHTS OF CHANGE RESERVED


10 TECHNOLOGY OVERVIEW

TECHNOLOGY OVERVIEW

SUPPLY BOARDS

A supply board is used in a central gas supply system in order to reduce the
cylinder pressure to an appointed secondary pressure. The supply board will
then supply a stable pressure to line regulators or points of use.

A supply board can be considered like a simplified switch over board (without
the continuous gas supply from several high-pressure sources).

Most of our supply boards have 3 common inlets available. This avoids
installation of extensions and increases safety of the installation.

PRODUCT
FINDER
Alpi-Safe P. 020
Alpi-Mod P. 022

SWITCH OVER BOARDS

SAFETY FIRST EASE OF HANDLING AND INSTALLATION

-- DUOBLOC: 6 cylinders (2x3) connections possible w/o -- Easy access of purging and isolation valves
extension – to improve the global system tightness and reduce -- Easy installation with all components pre‐mounted on an Omega plate
leakage points. Also, with DUOBLOC concept you can purge -- Easy access to all functions
independently each side. The purge can be collected. -- Easy access to all components submitted to maintenance
-- Fitted with a safety relief valve in case of over
pressure which can be collected
-- Shut‐off valves design approved for 300 bar for a higher safety factor LOWER OPERATING COSTS
-- Automatic switch for a continuous gas supply
-- Dedicated HP and LP pressure gauges for a permanent indication. -- A continuous gas supply for a better cylinders change productivity
Contact gauges can also be used to connect to an alarm box -- Easy and quick maintenance operations
-- With a gas monitoring system, you can easily check gas -- Product design for a long service lifetime
consumption and sources status from your desk -- Multiple large cylinders connection to reduce cylinders quantity
-- Protected with an inviolable plastic cover and associated rental, transportation and handling costs
-- A complete and compact system for a surface
saving in hospital technical areas
HIGH PERFORMANCE

-- Outlet pressure and flow stability thanks to


its Balances‐Valve Technology
-- Internal mechanical forces reduced for a longer
lifetime and a lower ownership cost
-- 2 levels of pressure regulation for a higher
flow and a much better stability
-- Smart choice of material in conformance with EN ISO 15001

ALL RIGHTS OF CHANGE RESERVED


TECHNOLOGY OVERVIEW 11

TECHNOLOGY OVERVIEW (continued)

AUTOMATIC SWITCH OVER BOARDS WITH MANUAL RESET

An automatic switch over board with manual reset is a system which


provides a continuous gas supply to the piping system. One source of gas is
used as the primary source, while a second source is held in reserve.

When the primary source reaches a predetermined pressure, the reserve


supply automatically begins to supply gas to the system.

Once the switch over occurs and primary source is replaced, the switch over
board is reset, such that the reserve supply supplying gas is now desig-
nated as primary source and the secondary source is now held in reserve.
The empty cylinder can be replaced without interrupting the gas flow.

PRODUCT
FINDER
Alpi-Switch P. 016
Alpi-Cen P. 018

SPECIAL FEATURES

MAINTENANCE SYSTEM PURGING AND SETTING DEVICE

For serviceable components, our maintenance systems can be selected for Special device integrated on second stage regulators for effective mainte-
your maintenance operations. nance operations and setting purpose.

--Enable easy, quick and effective --Enable a safe pressure regulator


maintenance operations on pressure gauge, dismantling by an easy and appropriate
pressure sensor and safety relief valve venting when isolated
--Manual operation without any tool --After your maintenance operation
--O-ring sealing for an easily manageable effectively set the regulator as it is done
tightness in our production line
--Enable maintenance operations under --Enable settings in static and dynamic
pressure without interrupting medical conditions in order to properly set pressure
gas supply into the hospital and flow according to hospital alarm levels
--No gas leaks during maintenance -Specific
- tooling available
operations for as safe replacement without
being exposed to pressurized gas flow

When selected for second stage regula- PRODUCT


tors maintenance system gives installer FINDER
much more flexibility and enables easy Alpi-MSV P. 034
PRODUCT
product reversibility on site Alpi-Dist P. 036 FINDER
Alpi-Reg 1 P. 028 Alpi-Reg 1 P. 028
Alpi-Reg 2 P. 030 Alpi-Reg 2 P. 030

ALL RIGHTS OF CHANGE RESERVED


12 TECHNOLOGY OVERVIEW

TECHNOLOGY OVERVIEW (continued)

PRESSURE REGULATOR TECHNOLOGIES

Rotarex Supply Panels and Switch over Panels use 3 main pressure regulator DIAPHRAGM
technologies to achieve a stable and reliable pressure regulation: -- Our most-used technology (cylinder regulation, line, supply panel...)
-- Compact design
BALANCED VALVE -- Good precision
-- Best-in-class pressure stability
-- Minimizes the effect of inlet pressure fluctuations on outlet pressure CARTRIDGE
-- Increases regulator lifetime and reduces cost of ownership by -- High outlet pressure stability
reducing seat effort -- Compact design
-- Diaphragm technology only -- Long product life = very resistant technology

BALANCED VALVE TECHNOLOGY

Our Balanced-Valve Technology regulator gives you dual stage regulator


performance in a single stage design. Due to its proprietary design, it is able PRODUCT
to balance the internal forces within the regulator and compensates for the FINDER
Alpi-Switch P. 016
pressure fluctuation on the inlet. It provides a constant outlet pressure like a
Alpi-Cen P. 018
dual stage regulator but with a lower total ownership cost. Alpi-Safe P. 020
Alpi-Mod P. 022

Ergonomic hand wheel


for pressure adjustment

Spring applies
BV Technology force on diaphragm

Single stage
P2

Inlet (P1) Outlet (P2)


Balanced-Valve
and Dual stage

P1

ALL RIGHTS OF CHANGE RESERVED


TECHNOLOGY OVERVIEW 13

TECHNOLOGY OVERVIEW (continued)

CARTRIDGE REGULATOR

Superior technical performance with cartridge technology: PRODUCT


-- Better outlet pressure stability due to the cartridge design. Outlet FINDER
pressure remains stable despite any fluctuation of inlet pressure. SCL 290 MED P. 024
-- Longer product life due to less impingement on the diaphragm. Alpi-Reg 1 P. 028
-- Compact design with reduction of dead volume Alpi-Reg 2 P. 030
(minimal purge requirements)
-- Sintered inlet filter provides better filtration without restricting flow.

Fine threads to Ergonomic hand wheel


easily set up for pressure adjustment
the outlet pressure

Spring applies
force on diaphragm

Cartridge Diaphragm

Sintered filter

Outlet (P2) Inlet (P1)

DIAPHRAGM REGULATOR

Ergonomic hand wheel


for pressure adjustment PRODUCT
FINDER
DM200 GD P. 026
DM200 TGD P. 027
DM20
Spring applies DM20 GD P. 033
force on diaphragm DC50 P. 032

Poppet
Diaphragm

Inlet (P1) Outlet (P2)

ALL RIGHTS OF CHANGE RESERVED


14 SPECIAL SERVICES

CEODEUX MEDITEC SERVICES


SYSTEMS PRE - ENGINEERING

We pre-engineer total solution from source to point of use regarding the


standard, the performances and the type of source you are considering

INSTALLATION / SERVICE SUPPORT

We can provide you installation support in order to easily and effectively


implement our equipments in hospitals or clinics.
We can offer you service kits, instructions and tooling in order to make our
equipments easily serviceable

PRODUCTS TRAININGS

We suggest trainings for installers and hospital technical staff and


documentation support in order to teach how to manipulate and service
our equipments

TECHNICAL SUPPORT

With our experience, expertise, analysis and testing equipments, we are


able to give you strong technical assistance and help you to solve problems

ALL RIGHTS OF CHANGE RESERVED


SPECIAL SERVICES 15

ELECTRONIC SOLUTIONS …GAS MONITORING SYSTEM


KEY FEATURES

Switch over boards with integrated Contact gauge Pressure sensor Pressure switch Gas Source Monitoring
electronic panel

Pneumatic Control Contact pressure gauges for The pressure sensor The pressure switch is a device Complete Gas Source Monitoring
with Electronic Monitoring ROTAREX regulators, points precisely measures pressure which indicates when a certain And possible reports in different
of use, supply boards or of gas in use. It acts as a set pressure has been reached requested locations
switch over boards transducer; it generates a on its input. The switch may be
signal as a function of the designed to make contact either
pressure imposed. on pressure rise or on pressure
fall.

DISPLAY DETAILS

1 HP CYLINDER PRESSURE 6 SINGLE SOURCE


Primary Liquid Source or
2 LINE PRESSURE IN BAR Emergency Gas Source
2 5
6
3 LINE PRESSSURE 7 SETTING AND CONTROL
BARGRAPH INTERFACE
Color coding (Green, Yellow, - Values
Red) based on defined toler- - Units
ances - Tolerances
- Alarm levels
4 SOURCE STATUS
In Operation or in Reserve
(Reset Option)
1 3
4 7 5 SOURCE STATUS
AND INFORMATION
- Empty to be replaced
- Set as next source in service
- Failure Mode alarms

ALL RIGHTS OF CHANGE RESERVED


16 HOW TO CHOOSE A SUPPLY SYSTEM

SELECTING THE RIGHT SUPPLY SYSTEM

To choose the right supply solution for your application and get the best results, you should identify the following technical parameters:

TECHNICAL PARAMETER EXAMPLE


Product Supply board, switchover board, manifold regulator, line regulator, cylinder regulator, valve,...
Gas type O2 , Air, N2O, CO2
Type of source Compressed cylinder, liquid tank,...
Nominal inlet pressure Bar or psi
Nominal outlet pressure Bar or psi
Outlet pressure limits lower and upper alarm levels
Nominal flow (N2) Nm3/h, Nlpm, Slpm or SCFM
Material Raw brass, chrome plated brass, stainless steel
Inlet connection Ports quantity and type
Outlet connection Ports quantity and type
Gauges Pressure range and type (bourdon tube, inductive contact, sliding contact)
Pressure sensor yes/no, pressure range and location/ output signal : 4-20mA
Pressure switch yes/no
Temperature range -20°C to +60°C / -4° F to +140°F
Outdoor or indoor use Environment
Product applicable standards ISO 10524-1 for cylinder pressure regulators
ISO 10524-2 for manifold and line pressure regulator
ISO 10524-4 for low pressure regulator
Medical gas supply system standard ISO 7396-1, HTM 02-01, NFPA 99, ABNT NBR 12188,...
Certification / Approval / Registration CE marked - 93/42/EEC European Medical Device Directive, FDA 510(k) – USA, TGA (Therapeutic Goods
Administration) - Australia, Health Canada, HSA (Health Sciences Authority) – Singapore, ANVISA – Brasil

Each product page is designed to provide you the essential technical information at a glance:

16 SUPPLY BOARDS SUPPLY BOARDS 17

M820 / M825 ALPI-SWITCH | SWITCH OVER BOARD SPECIFICATIONS

Inlet Ports G3⁄8” Male Diaphragm AISI 304 Inlet pressure 200 bar / 2900 PSI
KEY FEATURES REGULATORY
- M820 Automatic Supply Outlet Port G½” Female Temperature -20°C to +60°C Outlet pressure 10 bar / 145 PSI
System with Manual Reset - 2 x duobloc with shut‐off valves - Medical certified according to MDD 93/42/EEC
range
is made for hospitals - 2 x 3 inlets amended 2007/47/EC Seat seal Polyamide Gauges HP : 0-315 bar Nominal Flow M820 : 50 Nm3/h (N2)
and medical clinics - 1 x outlet - Complies with EN ISO 7396‐1 LP : 0-16 bar M825 : 80 Nm3/h (N2)
- M820 enables a continuous - 1 x safety relief valve O-ring EPDM Weight ± 13.8 kg Leak rate 1.10-4 mbar.l/s He
gas supply. One source is - 2 x purge valve
in service while the second - 2 x high pressure gauges
one is held in reserve - 1 x low pressure gauge
- Suitable for Oxygen - 2 x pressure regulators
and all medical gases
- Automatic switch from
the source in service to
the one in reserve
- Manual reset when 1 PURGE PORT 7 OUTLET PORT
6 7 - G3⁄8 (standard) - G1/2 (standard)
replacing empty cylinder
- Empty cylinder - Collected (option) - Compression tube fitting
replacement without (option)
interrupting gas flow 2 HP PRESSURE SENSOR
- Good looking design with - Plug (standard) 8 PURGE VALVE
5
possible alarm integration 8 - HP pressure sensor (option) FIXATION SUPPORT
4
3 HP PRESSURE GAUGE 9 SAFETY RELIEF VALVE
OPTIONS 9 - HP Pressure gauge (standard)
- HP contact gauge (option) 10 RESET LEVER
★ M825 serie with integrated
dual stage regulator 3 4 LP PRESSURE GAUGE
10 11 DUOBLOC
- LP Pressure gauge (standard)
- LP contact gauge (option)
11 12 ISOLATION VALVE
2
12 5 LP PRESSURE SENSOR
- Plug (standard) 13 PURGE VALVE
1 13 - LP pressure sensor (option)
M820 VERSION M825 VERSION M820 / M825 14 INLET PORT(S)
14 6 OUTLET - w/o Extension Block - 1 inlet
- Integrated second stage port (standard)
regulator - w/ Extension Block 2 or 3 inlet
ports (option)

PRODUCT CONFIGURATOR

Inlet Version Outlet Outlet Pressure Sensor/


Pressure Type Gas Type Pressure Inlet Connection Inlet Ports Connections Gauges Switch Configuration

ALPISW 200 V1 02 10 G38 2P G12 1 B A


200 bar 200 M820 V1 O2 O2 10 bar 10 G3⁄8” Male G38 2 x 1 port 1P G½” - G12 with gauges - 1 without A Standard A
2900 145 PSI with O-ring Female standard pressure sensor configuration
PSI
M825 V2 Air AIR 2 x 2 port 2P Tube T12 with HP 2 with HP B with collected CL
Fitting inductive pressure sensors purge and
12mm contact gauge safety *
N2O N2O 2 x 3 port 3P Tube T16 with LP 3 with LP pressure C * Special on demand

Fitting inductive sensor


16mm contact gauge
CO2 CO2 with HP & 4 with HP & LP D
LP inductive pressure sensors
contact gauge
N2 N2 with LP Pressure E
sensor
with HP sensor F
& LP switch

ALL RIGHTS OF CHANGE RESERVED ALL RIGHTS OF CHANGE RESERVED

ALL RIGHTS OF CHANGE RESERVED


17

NOTES

ALL RIGHTS OF CHANGE RESERVED


18 SWITCH OVER BOARDS

M820 / M825 /M826 ALPI-SWITCH | SWITCH OVER BOARD

KEY FEATURES REGULATORY


-- M820 Automatic Supply
System with Manual Reset -- 2 x duobloc with shut‐off valves -- Medical certified according to MDD 93/42/EEC
is made for hospitals -- 2 x 3 inlets amended 2007/47/EC
and medical clinics -- 1 x outlet -- Complies with EN ISO 7396‐1
-- M820 enables a continuous -- 1 x safety relief valve
gas supply. One source is -- 2 x purge valve
in service while the second -- 2 x high pressure gauges
one is held in reserve -- 1 x low pressure gauge
-- Suitable for Oxygen -- 2 x pressure regulators
and all medical gases
-- Automatic switch from
the source in service to
the one in reserve
-- Manual reset when
replacing empty cylinder
-- Empty cylinder
replacement without
interrupting gas flow
-- Good looking design with
possible alarm integration

Electronic display available


for these three versions of
Alpi-switch (see page 15 for
more informations)

ALL RIGHTS OF CHANGE RESERVED


SWITCH OVER BOARDS 19

M826 VERSION

Dual stage line regulator

M825 VERSION

Single stage line regulator

ALL RIGHTS OF CHANGE RESERVED


20 SWITCH OVER BOARDS

M820 VERSION

No line regulator

M815 VERSION

Single stage line regulator

ALL RIGHTS OF CHANGE RESERVED


SWITCH OVER BOARDS 21

SPECIFICATIONS

Inlet Ports G3/8” Male Diaphragm AISI 304 Inlet pressure 200 bar / 2900 PSI

Outlet Port G1/2” Female Temperature -20°C to +60°C Outlet pressure 10 bar / 145 PSI
range
Seat seal Polyamide Gauges HP : 0-315 bar Nominal Flow M820 : 50 Nm3/h (N2)
LP : 0-16 bar M825 : 80 Nm3/h (N2)
O-ring EPDM Weight ± 13.8 kg Leak rate 1.10-4 mbar.l/s He

1 PURGE PORT
6 7 7 OUTLET PORT
- G3/8 (standard)
- Collected (option) - G1/2 (standard)
- Compression tube fitting
2 HP PRESSURE SENSOR
(option)
- Plug (standard)
5 8 PURGE VALVE
8 - HP pressure sensor (option)
4 FIXATION SUPPORT
3 HP PRESSURE GAUGE
- HP Pressure gauge (standard) 9 SAFETY RELIEF VALVE
9
- HP contact gauge (option)
10 RESET LEVER
3 4 LP PRESSURE GAUGE
10 - LP Pressure gauge (standard) 11 DUOBLOC
- LP contact gauge (option)
11
2 12 ISOLATION VALVE
12 5 LP PRESSURE SENSOR
- Plug (standard) 13 PURGE VALVE
1 13 - LP pressure sensor (option)
14 INLET PORT(S)
14 6 OUTLET - w/o Extension Block - 1 inlet
- Integrated dual stage port (standard)
regulator, or single stage, or - w/ Extension Block 2 or 3 inlet
without line regulator ports (option)

PRODUCT CONFIGURATOR

Inlet Version Outlet Outlet Pressure Sensor/


Gas Type Inlet Connection Inlet Ports Gauges Configuration
Pressure Type Pressure Connections Switch
ALPISW 200 V1 02 10 G38 1P G12 1 A A
200 bar 200 M820 V1 O2 O2 10 bar 10 G3/8” Male G38 2 x 1 port 1P G1/2” - G12 with gauges - 1 without A Standard A
2900 145 PSI with O-ring Female standard pressure sensor configuration
PSI
M825 V2 Air AIR 5 bar 5 2 x 2 port 2P Tube T12 with HP 2 with HP B with collected CL
72,5 PSI Fitting inductive pressure sensors purge and
12mm contact gauge safety *
M826 V3 N2O N2O 2 x 3 port 3P Tube T16 with LP 3 with LP pressure C * Special on demand

Fitting inductive sensor


16mm contact gauge
M815 V4 CO2 CO2 with HP & 4 with HP & LP D
LP inductive pressure sensors
contact gauge
N2 N2 with LP Pressure E
switch
with HP sensor F
& LP switch

ALL RIGHTS OF CHANGE RESERVED


22 SWITCH OVER BOARDS

ALPI-CEN | SWITCH OVER BOARD

KEY FEATURES REGULATORY


- Automatic Switch Over
Board with Manual - 2 x duobloc with shut‐off valves - Medical certified according to MDD 93/42/EEC
Reset for hospitals - 2 x 2 inlets (one with non‐whipping filter) amended 2007/47/EC
and medical clinics - 1 x outlet - Complies with EN ISO 7396‐1
- Enables a continuous gas - 1 x safety relief valve
supply. One source is in - 2 x purge valve
service while the second - 2 x high pressure gauges
one is held in reserve - 1 x low pressure gauge
- Automatically switches
from the source in service
to the one in reserve
- Manual reset when CEN Automatic version
replacing empty cylinder
- Allows replacement of
empty cylinders without
interrupting gas flow

CEN Manual version

KEY ADVANTAGES

SAFETY FIRST
- DUOBLOC: 6 cylinders (2x3) connections possible w/o EASE OF HANDLING AND INSTALLATION
extension – to improve global system tightness and reduce - Easy access of purging and isolation valves
leakage points. Also, with DUOBLOC concept you can purge - Easy installation with all components pre‐mounted on an Omega plate
independently each side. The purge can be collected. - Easy access to all components submitted to maintenance
- Fitted with a safety relief valve in case of over
pressure which can be collected LOWER OPERATING COSTS
- Automatic switch for a continuous gas supply - A continuous gas supply for better cylinder change-over productivity
- Dedicated HP and LP pressure gauges for a permanent indication. - Reduces cylinder quantity and associated rental,
Contact gauges can also be used to connect to an alarm box transportation and handling costs
- With a gas monitoring system, you can easily check gas - A complete and compact system for better space
consumption and source status from your desk. efficiency in hospital technical areas
HIGH PERFORMANCE
- Exceptionally stable outlet pressure and flow
thanks to Balanced‐Valve Technology
- Reduces internal mechanical forces for a longer
lifetime and a lower ownership cost

ALL RIGHTS OF CHANGE RESERVED


SWITCH OVER BOARDS 23

SPECIFICATIONS

Female ports G 3/8” (inlet) Leak rate 1.10-4 mbar ℓ/s He Inlet pressure 200 bar / 2900 psi
G 1/2” (outlet)
Seat seal Polyamide Temperature -20°C to + 60°C Outlet pressure 10 bar / 145 psi
range
O-ring EPDM Gauges High and low pressure Nominal Flow 50 Nm3/h (N2)
(M10 x 1 or G 1/4”)
Diaphragm AISI 304 Weight ± 13.8 kg Setting 8 bar / 10 bar

CEN Automatic version

PRODUCT CONFIGURATOR

Body End O-ring


Inlet Pressure Version type Outlet Pressure Gauges Outlet Valve Configurations
Material Connections Material
CEN 200 SEMI 10 LB G EPDM 1 NV A
200 bar 200 Automatic SEMI 10 bar 10 Raw LB In: G 3/8 G EPDM - with gauges - 1 without outlet NV Standard A
2900 psi 145 psi Brass Out: G 1/2 - standard standard shut-off valve configuration
Female (standard)
Manual MAN with HP inductive 2 with CL
contact gauge connected
purge
with HP sliding 3
contact gauge
with LP inductive 4
contact gauge
with LP sliding 5
contact gauge
with HP & LP sliding 6
contact gauges
Others versions and possibilities available upon request

ALL RIGHTS OF CHANGE RESERVED


24 SUPPLY BOARDS- EMERGENCY MODULES

M830 /M831 ALPI-SAFE | SAFETY SUPPLY BOARD

KEY FEATURES REGULATORY


- Safety Supply System
for hospitals and - 1 x duobloc with shut‐off valves - Medical certified according to MDD 93/42/EEC
medical clinics - 3 x inlet connections amended 2007/47/EC
- Enables gas supply - 1 x outlet connection - Complies with EN ISO 7396‐1
continuity if both primary - 1 x safety relief valve
and secondary sources - 1 x purge valve
are unable to supply - 1 x high pressure gauges
the gas pipelines or for - 1 x low pressure gauge
maintenance purpose - 1 x Pressure regulators
- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)
- Permanently connected
with an automatic M830: no line regulator M831: line regulator (with optional reset lever)
activation
- Attractive design thanks to
its protective plastic cover

ALL RIGHTS OF CHANGE RESERVED


SUPPLY BOARDS- EMERGENCY MODULES 25

SPECIFICATIONS

Inlet Ports G 3/8” male Diaphragm AISI 304 Inlet pressure 200 bar / 2900 PSI

Outlet Port G1/2” Female Temperature -20°C to +60°C Outlet pressure 7 bar / 101.5 PSI
range
Seat seal Polyamide Gauges HP : 0-315 bar Nominal Flow 50 Nm3/h (N2)
LP : 0-16 bar
O-ring EPDM Weight ± 5.0 kg Leak rate 1.10-4 mbar.l/s He

1 INLET PORT(S) 9 LP PRESSURE GAUGE


-- w/o Extension Block - 1 inlet -- Bourdon tube 0-16 bar
port (standard) (standard)
-- w/ Extension Block 2 or 3 inlet -- Inductive contact gauge
port(s) (option) (option)

2 PURGE VALVE 10 LP PRESSURE SENSOR


-- Plug (standard)
7 3 DUOBLOC -- 0-16 bar / 4-20 mA
(option)
6 8
4 ISOLATION VALVE
5 9 11 PURGE PORT
5 SAFETY RELIEF VALVE -- G3/8 (standard)
10 -- Collected (option)
6 HP PRESSURE GAUGE
-- Bourdon tube 0-315 bar 12 FIXATION SUPPORT
4 14 (standard)
3 -- Inductive contact gauge 13 HP PRESSURE SENSOR
(option) -- Plug (standard)
2 11
-- 0-250 bar / 4-20 mA
7 OUTLET PORT (option)
1
-- G1/2 (standard)
12 -- Compression tube fitting 14 RESET LEVER (OPTIONAL)
(option) -- Plug (standard)
-- 0-250 bar / 4-20 mA
8 OUTLET (option)
13
-- Outlet block (standard)
-- Integrated second stage
regulator (for M831)

PRODUCT CONFIGURATOR

Outlet Outlet
Inlet Pressure Version type Gas Type Inlet Connection Inlet Port Gauges Pressure Sensor Configuration
Pressure Connections
ALPISAFE 200 O2 7 G38 2 G12 4 1 A
200 bar 200 M830 V1 O2 O2 7 bar 7 G3/8” Male G38 1 port 1P G1/2” - G12 with gauges - 1 without A Standard A
2900 PSI 101.5 PSI with O-ring Female standard pressure configuration
sensor
M831 V2 Air AIR 2 port 2P Tube Fitting T12 with HP 2 with HP B with collected CL
12mm inductive pressure purge and
contact gauge sensors safety *
M831 N2O N2O 3 port 3P Tube Fitting T16 with LP 3 with LP C * Special on demand

with reset 16mm inductive pressure


lever contact gauge sensor
CO2 CO2 with HP & 4 with HP & D
LP inductive LP pressure
contact gauge sensors
N2 N2

ALL RIGHTS OF CHANGE RESERVED


26 SUPPLY BOARDS- EMERGENCY MODULES

ALPI-MOD | SUPPLY BOARD

APPLICATIONS
-- Safety Supply System
for hospitals and -- Used in combination with a switch over board for
medical clinics the regulation of the emergency source during
-- Enables gas supply maintenance on the principal source. This avoids
continuity if both primary installing extensions and reduces the amount of
and secondary sources leaking points.
are unable to supply -- Suitable for the high flow supply of industrial gases
the gas pipelines or for except toxic and corrosive gases.
maintenance purpose
-- Suitable for Oxygen and all
medical gases (Breathing
Air, CO2, Ar, N2O)
-- Permanently connected
with an automatic
activation
-- Attractive design thanks to
its protective plastic cover

KEY FEATURES

-- Possible to connect 3 gas cylinders and a gas for purging operation (up
to 3 cylinders without any extension - without using the purge line).
-- Ready to install: all components are pre-mounted on a board.
-- Best-of-class pressure stability with Balanced-Valve Technology: the
effect of inlet pressure fluctuations on outlet pressure are minimized.
Balanced-Valve Technology enables the delivery of a very stable outlet
pressure and flow even with high flow line regulators.
-- Can be connected to an alarm box using contact gauges.

ALL RIGHTS OF CHANGE RESERVED


SUPPLY BOARDS- EMERGENCY MODULES 27

SPECIFICATIONS

Female ports In: G 3/8 - Out: G 1/2 Leak rate 1.10-4 mbar ℓ/s He Inlet pressure 200 bar
2900 psi
Seat seal Polyamide
Temperature -20°C to + 60°C Outlet pressure 10
range 145psi
O-ring EPDM
Gauges High and low pressure Nominal Flow 50 Nm3/h (N2)
Diaphragm AISI 304 (M10 x 1 or G 1/4)
(regulator)
Weight ± 6,0 kg
± 13.0 lbs

PRODUCT CONFIGURATOR

Body O-ring Fix or adjustable


Inlet pressure Outlet End Connections Gauges Oulet valve Configuration
Material Material Outlet Pressure
ALPIMOD 200 16 L G EPDM 1 FX N A
200 bar 200 10 bar 10 Raw LB In: G 3/8 G EPDM - With gauges - 1 With fixed FX Without NV Standard A
2900 psi 145 psi brass Out: G 1/2 Standard standard P2 outlet shut- configuration
Female off valve
16 bar 16 With HP 2 “Mirror” version R
232 psi inductive - duobloc on
contact gauge right side
With LP 4 With connected CL
inductive purge
contact gauge
With HP & 6 “Mirror” with RCL
LP inductive connected
contact gauges purge

ALL RIGHTS OF CHANGE RESERVED


28 HIGH PRESSURE REGULATOR

SERIES BSC 290/390 | HP CYLINDER PRESSURE REGULATOR

INCLUDED FUNCTIONS REGULATORY


-- Ultimate and/or
Maintenance source -- Inlet port -- Medical certified according to MDD 93/42/EEC
of supply for hospitals -- Outlet port amended 2007/47/EC
and medical clinics -- Pressure regulator -- Product complies with EN ISO 10524-1
-- It enables gas supply -- Safety relief valve
continuity if primary, -- High pressure measurement
secondary and emergency -- Low pressure measurement
sources are unable to
supply the gas pipelines or
for maintenance purpose
-- Much higher safety
level for the complete
gas supply system
-- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)
-- To be connected on
medical cylinder valve

KEY ADVANTAGES

SAFETY FIRST Right inlet Left inlet


-- Fitted with a safety relief valve in case of over
pressure which can be collected
-- Dedicated HP and LP pressure gauges for a permanent indication
-- Compatible materials with Oxygen and medical applications
-- No gas contamination due to a metal-metal sealing
Plugged Plugged
HIGH PERFORMANCE
-- Proven Cartridge Seat Technology
-- Better outlet pressure and flow stability
-- Smart choice of material in conformance with EN ISO 15001 Pressure gauges
Inlet (HP)
LOWER OPERATING COSTS
-- Compact and lightweight design for an easy handling and storage Outlet (LP)
-- Longer product life thanks to less impingement on the diaphragm
Safety valve
-- Easy maintenance operations thanks to Cartridge Technology

ALL RIGHTS OF CHANGE RESERVED


HIGH PRESSURE REGULATOR 29

SPECIFICATIONS

Body Chrome plated brass Inlet Connection Cylinder connection Inlet pressure 200 / 300 bar
2900 / 4350 psi
Seat seal Polyamide Outlet Connection Hose Barb or Medical connector Outlet pressure 4 / 8 bar
58 / 116 psi
O-ring EPDM Leak rate 10- mbar l/s He
8 Nominal Flow 7 / 25 Nm3/h (N2)

Diaphragm Hastelloy® Weight ± 1.4 kg Temperature -40°C to +60°C


range

PRODUCT CONFIGURATOR

Material Inlet Pressure Port Configuration Outlet pressure Inlet Connection Outlet connection Pressure sensor Gas Type
BSC L 290 R 4 NF NIST B O2
Chrome plated L 200 bar 290 Right inlet R 4 bar 4 G1/4” Female O2 Hose Barb HB without pressure A O2 O2
brass 2900 psi 58 psi 6 mm sensor
300 bar 390 Left inlet L 8 bar 8 NF 29-650 NF AFNOR Medical NF with HP pressure B Air AIR
4350 psi 116 psi Connector sensors
10 bar 10 DIN 477-1 DIN DIN Medical DIN with LP pressure C N2O N2O
145 psi Connector sensor
BS 341-3 BS NIST Medical NIST with HP & LP D CO2 CO2
Connector pressure sensors
NEN 3268 NEN N2 N2

UNI 11144 UNI

CGA V1 CGA

ALL RIGHTS OF CHANGE RESERVED


30 HIGH PRESSURE REGULATOR

DM200 GD | MANIFOLD REGULATOR

APPLICATIONS KEY FEATURES


-- Diaphragm single stage
-- Purity up to 5.5 -- Designed for cylinder regulator applications. -- Exceptionally durable
-- Inlet pressure: -- Ideally used in industrial applications requiring high
200 bar (2900 psi) -- Medical version available (see Meditec catalogue).
flow - particulary to supply gas to welding machines.
-- Outlet pressure:
10 bar (145psi)

TO ORDER

★ HP pressure gauge:


290043000000
★ HP & LP pressure gauge:
290044000000

SPECIFICATIONS

Ports inlet: 16 x 1.336 (Female) Weight ± 4,6 kg Inlet pressure 200 bar (2900 psi)
outlet: M20 x 1,5 (Male) ± 10.1 lbs
Seat seal PA 6.6 Leak rate 10-3 mbar ℓ/s He Outlet pressure 10 bar / 145 psi

Diaphragm Butyl Temperature range - 20°C to + 60°C Nominal Flow 100 Nm3/h (N2)
- 4°F to + 140°F
Body Material Raw brass Gauges High and low pressure
(M10 x 1)

ALL RIGHTS OF CHANGE RESERVED


HIGH PRESSURE REGULATOR 31

DM200 TGD ALPI-LINE | LINE REGULATOR

APPLICATIONS TO ORDER
-- Diaphragm single stage
-- Purity up to 5.5 -- Ideally suited for distribution of medical gases in -- Option 1:
-- Inlet pressure: hospital applications requiring very high flow with HP pressure gauge: 295004000000
200 bar (2900 psi) -- Option 2:
-- Outlet pressure: with HP & LP pressure gauge: 295006000000
20 bar (290psi)

★ High flow regulator


★ 1 inlet / 1 outlet
★ O2 application compatible
★ Inlet/outlet pressure gauges

Special requirements on request

SPECIFICATIONS

Ports inlet: 16 x 1.336 (Female) Weight ± 4,6 kg Inlet pressure 200 bar
outlet: M20 x 1,5 (Male) ± 10.1 lbs 2900 psi
Seat seal PA 6.6 Leak rate 10-3 mbar ℓ/s He Outlet pressure 20 bar
290 psi
Diaphragm Butyl Temperature range - 20°C to + 60°C Nominal Flow 250 Nm3/h (N2)
- 4°F to + 140°F
Body Material Raw brass Gauges High and low pressure
(M10 x 1)

ALL RIGHTS OF CHANGE RESERVED


32 LINE PRESSURE REGULATORS

M845 ALPI-REG1 | SINGLE LINE PRESSURE REGULATOR

KEY FEATURES REGULATORY


-- Single Line Pressure
Regulator for hospitals -- 1 x pressure regulator -- Medical certified according to MDD 93/42/EEC
and medical clinics -- 1 x inlet connections amended 2007/47/EC
-- Mainly used in double‐stage -- 1 x outlet connection -- Complies with EN ISO 7396‐1
pipeline systems where -- 1 x first stage pressure gauge
pressure reduction from the -- 1 x second stage pressure gauge
supply system is required -- 2 x quick connector / pressure sensor
in order to provide the
nominal distribution pressure
to the terminal units
-- Enables supply to several
hospital or clinic floors and
can be easily integrated in
interior design thanks to
its good looking design
-- Maintain gas supply during
maintenance and/or
emergencies thanks to two
quick connectors that bypass
the pressure regulator
-- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)

Left Inlet Side Configuration

Reversed Configuration

Turn it at 180° and exchange pressure gauges


and quick connectors and you will easily
get a right inlet side pressure regulator

ALL RIGHTS OF CHANGE RESERVED


LINE PRESSURE REGULATORS 33

SPECIFICATIONS

Inlet Ports G1/2” female Leak rate 10-3 mbar l/s He Inlet pressure 10 bar / 145 psi

Outlet Port G1/2” female Temperature +10°C to +40°C Outlet pressure 4 bar / 58 psi
range

Other Ports Plug G3/8” Weight 4.0 Kg Nominal Flow 50 Nm3/h (N2)

1 INLET ISOLATION VALVE 7 O UTLET PRESSURE


5 REGULATOR
2 SERVICE CONNECTION
-- Plug G3/8” (standard) 8 P RESSURE GAUGE
-- Medocal connector (option) CONNECTION
-- G1/4” (standard)
6 3 INLET CONNECTION -- w/maintenance system
4
7 -- G1/2” female(standard) (option)
-- Compression tube fitting
(option) 9 OUTLET CONNECTION
8 -- G1/2” female(standard)
3 9 4 INLET PRESSURE GAUGE (MP) -- Compression tube fitting
-- Bourdon tube 0-16 bar (option)
2 10 (standard)
-- Inductive contact gauge 10 SERVICE CONNECTION
(option) -- G1/2” female(standard)
-- Compression tube fitting
5 P URGING & SETTING (option)
DEVICES
1 11 11 OUTLET ISOLATION VALVE
6 C ARTRIDGE PRESSURE
REGULATOR

PRODUCT CONFIGURATOR

Outlet
Inlet Pressure Gas Type End Connections Service Connection Gauges Maintenance System Configuration
Pressure
ALPIR1 10 O2 4 G12 PG 1 SS L
10 bar 10 O2 O2 4 bar 4 G1/2” - Female G12 Plug G3/8” PG with gauges - 1 without maintenance SS Left inlet L
145 PSI 58 PSI standard system (standard)
Air AIR 8 bar 8 Tube Fitting T12 AFNOR Medical NF with MP inductive 2 with maintenance MS Right inlet R
116 PSI 12mm Connector contact gauge system
N2O N2O Tube Fitting T16 DIN Medical DIN
16mm Connector
CO2 CO2

N2 N2

ALL RIGHTS OF CHANGE RESERVED


34 LINE PRESSURE REGULATORS

M840 ALPI-REG2 | DOUBLE LINE PRESSURE REGULATOR

KEY FEATURES REGULATORY


-- Double Line Pressure
Regulator for hospitals -- 2 x pressure regulator -- Medical certified according to MDD 93/42/EEC
and medical clinics -- 1 x inlet connections amended 2007/47/EC
-- Mainly used in double‐stage -- 1 x outlet connection -- Complies with EN ISO 7396‐1
pipeline systems where -- 1 x first stage pressure gauge
pressure reduction from the -- 1 x second stage pressure gauge
supply system is required -- 2 x quick connector
in order to provide the -- 2 x pressure sensor
nominal distribution pressure
to the terminal units
-- Enables to supply several
hospital or clinic floors and
can be easily integrated in
interior architecture thanks
to its good looking design
-- Gas supply continuity
during maintenance and/
or emergency thanks to
its triple safety level
-- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)

Left Inlet Side Configuration

Reversed Configuration

Turn it 180° and exchange pressure gauges


and quick connectors/nuts/sensors and you
will easily get a right inlet side product

ALL RIGHTS OF CHANGE RESERVED


LINE PRESSURE REGULATORS 35

SPECIFICATIONS

Inlet Ports - Welded for tube Leak rate 10-3 mbar l/s He Inlet pressure 8 to 10 bar max.
Ø12/16/18/21mm
- RDB tube fitting Ø12mm
Outlet Port - Welded for tube Temperature +10°C to +40°C Outlet pressure 4 bar 0/+1 bar
Ø12/16/18/21mm range
- RDB tube fitting Ø12mm
Other Ports - G3/8” Weight 8.0 Kg Nominal Flow 50 Nm3/h (N2)
2x first stage - Quick connector according to
+2x second stage
national standards

5 1 INLET ISOLATION VALVE 7 O UTLET PRESSURE


GAUGE
2 SERVICE CONNECTION
-- Plug G3/8” (standard) 8 P RESSURE GAUGE
6 -- Medocal connector (option) CONNECTION
4 -- G1/4” (standard)
7 3 INLET CONNECTION -- w/maintenance system
-- G1/2” female(standard) (option)
-- Compression tube fitting
8 (option) 9 OUTLET CONNECTION
3 -- G1/2” female(standard)
9
4 INLET PRESSURE GAUGE (MP) -- Compression tube fitting
2 -- Bourdon tube 0-16 bar (option)
(standard)
-- Inductive contact gauge 10 P RESSURE SENSOR
10 (option) CONNECTION
-- G1/2” female(standard)
5 P URGING & SETTING -- Compression tube fitting
DEVICES (option)

6 C ARTRIDGE PRESSURE 11 OUTLET ISOLATION VALVE

1 11
REGULATOR

PRODUCT CONFIGURATOR

Inlet Outlet Maintenance


Gas Type End Connections Service connection Gauges Pressure Sensor Configuration
Pressure Pressure System
ALPIR2 10 N2O 4 G12 NF 1 A MS L
10 bar 10 O2 O2 4 bar 4 G1/2” - Female G12 G3/8” Female G38 with gauges - 1 without pressure A without SS Left inlet L
145 PSI 58 PSI standard sensor maintenance
system
Air AIR 8 bar 8 Tube Fitting T12 AFNOR Medical NF with MP 2 with MP B with MS Right inlet R
116 PSI 12mm Connector inductive pressure sensor maintenance
contact gauge system
N2O N2O Tube Fitting T16 DIN Medical DIN with LP pressure C
16mm Connector sensor
CO2 CO2 with LP & MP D
pressure sensors
N2 N2

ALL RIGHTS OF CHANGE RESERVED


36 LINE PRESSURE REGULATORS

M847 ALPI-REG | LINE PRESSURE REGULATOR

-- Diaphragm single stage APPLICATIONS KEY FEATURES


-- Balanced-Valve Technology
-- Inlet pressure: -- For all applications requiring a low pressure with high -- Low pressure regulator with high flow, without
50 bar (725 psi) flow. vibration.
-- Outlet pressure: -- Ideally suited as line regulator in combination either -- Best-in-class pressure stability with Balanced-Valve
8/15 bar with supply board or switch over board. Technology: the effect of inlet pressure fluctuations
116/217 psi on outlet pressure is minimized. BV-technology
enables the delivery of a very stable outlet pressure
and flow even with high flow line regulators.
★ 1 inlet / 1 outlet -- reduced strain on the seat increases regulator life and
★ Rear thread for panel reduces the ownership cost.
mounting
★ O2 application compatible
★ High flow

Special requirements on request

SPECIFICATIONS

Female ports G 1/2 or 1/2 NPT Weight ± 1,4 kg Inlet pressure 50 bar (725 psi)
(inlet/outlet) ± 3.1 lbs
Seat seal EPDM Leak rate 10-3 mbar ℓ/s He Outlet pressure 8/15 bar
116/217 psi
O-ring EPDM Temperature range - 20°C to + 60°C Nominal Flow 150/300 Nm3/h (N2)
- 4°F to + 140°F
Diaphragm EPDM Gauges Low pressure Oxygen use OK

PRODUCT CONFIGURATOR

Outlet Pressure End Connections O-ring Material Body Material Gauges


ALPI-REG 4 G EPDM CHR 1
4 bar 4 G 1/2 - G 1/2 G EPDM - Standard Chrome plated CHR Without 0

8 bar 8 With 1

ALL RIGHTS OF CHANGE RESERVED


LINE PRESSURE REGULATORS 37

DM20 / DM20GD

APPLICATIONS KEY FEATURES


-- Medical line regulator
-- Low pressure gauge -- This regulator is the ultimate solution for the -- Outlet pressure can be setted from 0 to 10bar and
-- Inlet and outlet -- transportation of your medical gases. from 0 to 12bar.
connections -- Simple, robust, safe and durable design.
-- Made in France.
PART NUMBER
★ DM20 type:
293520000000
★ DM20 GD:
295304000000
Special requirements on request

SPECIFICATIONS

Body Raw brass Inlet Connection G3/8”F Inlet pressure 20 bar / 290 psi
GD : G1”F

Seat seal Polyamide Outlet Connection Hose barb Ø8-10mm Max. Outlet 10 bar / 145psi
(Hose barb) pressure GD : 12bar / 174psi
GD : HB Ø13mm
O-ring EPDM Leak rate 10-8 mbar.l/s He Nominal Flow 98 Nm3/h (N2)
GD : 124 Nm3/h (N2)

Diaphragm NBR Weight ± 1.6 kg Temperature -20°C to +60°C


GD : ± 2.3 kg range

ALL RIGHTS OF CHANGE RESERVED


38 SECTION VALVES

M850 ALPI-MSV | SECTION VALVE

KEY FEATURES REGULATORY


-- Main Section Valve for
hospitals and medical -- 1 x brass bloc body -- Medical certified according to MDD 93/42/EEC
clinics gas supply system -- 3 x ball valves amended 2007/47/EC
-- Connects ALPI‐SWITCH, -- 1 x safety relief valve -- Complies with EN ISO 7396‐1
ALPI‐SAFE and Ultimate -- 3 x inlet connections
/ Maintenance sources -- 1 x outlet connection
together in order to get -- 1 x pressure gauge
one single outlet whatever -- 1 x pressure sensor (optional)
the source in service
-- Enables isolation
and/or selection of
a specific source
-- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)

ALL RIGHTS OF CHANGE RESERVED


SECTION VALVES 39

SPECIFICATIONS

Inlet Ports 1,2,3 G1/2” Female Gauge 0-16 bar Office size Ø10mm

Weight 4.3 Kg
Outlet Port G1/2” Female Temperature -20°C to +60°C
(Top right) Working Pressure 10 bar max. range

1 SAFETY RELIEF VALVE 6 PRESSURE GAUGE


2 4 -- Bourdon tube 0-10 bar (standard)
2 ISOLATION VALVE -- Inductive contact gauge (option)
5
3 I NLET CONNECTION 7 P RESSURE GAUGE
7 PORT 1, 2 & 3 CONNECTION
a. Port 1 /b. Port 2/ c. Port 3 -- G1/4” (standard)
3a -- w/ Maintenance system (option)
8 -- G1/2” Female (standard)
-- CTF (option)
3b -- Medical connector (option) 8 OUTLET CONNECTION
9 -- G1/2” (standard)
2 2 4 P RESSURE SENSOR -- Compression tube fitting (option)
CONNECITON
-- G1/4” (standard) 9 INLET PORT 3
-- w/ Maintenance system (option) -- G1/2” Female (standard)
1
-- CTF (option)
5 PRESSURE SENSOR -- Medical connector (option)
-- Plug G3/8” (standard)
6 3c -- 0-10 bar 7 4-20 mA

KEY ADVANTAGES

SAFETY FIRST EASE OF HANDLING AND INSTALLATION


-- Made in one single brass bloc, ALPI‐MSV improves global -- Easy access of isolation valves
system tightness and reduces potential leakage points. -- Easy installation with all components mounted on an Omega plate
-- Fitted with a safety relief valve in case of over-pressure -- Flexible product. Select the most appropriate
-- Dedicated pressure gauge and sensor for a permanent inlet port regarding your configuration
visual indication and signal control.
-- Available with a transparent plastic cover LOWER OPERATING COSTS
-- Easy and quick maintenance operations
HIGH PERFORMANCE -- Designed for a long service lifetime
-- Innovative and proven ball valve technology with compensation -- A complete and compact system for a better space
-- Smart choice of material conforming to EN ISO 15001 efficiency in hospital technical areas

PRODUCT CONFIGURATOR

Working Maintenance
Gas Type Inlet port 1 Inlet port 2 Inlet port 3 Outlet Port Gauges Pressure Sensor
Pressure System
ALPIMSV 10 N2O G12 G12 G12 G12 1 A SS
10 bar 10 O2 O2 G1/2” - Female G12 G1/2” - Female G12 G1/2” - Female G12 G1/2” - Female G12 with gauge - 1 without A without SS
145 PSI standard pressure maintenance
sensor sytem
Air AIR Tube Fitting T12 Tube Fitting T12 Tube Fitting T12 Tube Fitting T12 with inductive 2 with pressure B with MS
12mm 12mm 12mm 12mm contact gauge sensor maintenance
system
N2O N2O Tube Fitting T16 Tube Fitting T16 Tube Fitting T16 Tube Fitting T16
16mm 16mm 16mm 16mm
CO2 CO2 AFNOR NF AFNOR NF AFNOR NF
Medical Medical Medical
Connector Connector Connector
N2 N2 DIN Medical DIN DIN Medical DIN DIN Medical DIN
Connector Connector Connector

ALL RIGHTS OF CHANGE RESERVED


40 SECTION VALVES

ALPI-DIST | SECTION VALVE

KEY FEATURES REGULATORY


-- Distribution Manifold for
hospitals and medical -- 1 x brass bloc body -- Medical certified according to MDD 93/42/EEC
clinics gas supply system -- 1 x ball valves amended 2007/47/EC
-- The manifold is made to -- 1 x inlet connections -- Complies with EN ISO 7396‐1
isolate a low pressure line -- 1 x outlet connection
into the hospital facility in -- 1 x pressure gauge
case of emergency and/or -- 1 x pressure sensor (optional)
maintenance operations
-- With the emergency/
maintenance inlet port,
possibility of supplying gas
only in the isolate line
-- Suitable for Oxygen
and all medical gases
(Breathing Air, CO2, N2O)

KEY ADVANTAGES

SAFETY FIRST EASE OF HANDLING AND INSTALLATION


-- Made in one single brass bloc, ALPI‐MSV improves global -- Easy access of isolation valves
system tightness and minimizes potential leakage points. -- Easy installation with all components mounted on an Omega plate
-- Dedicated pressure gauge and sensor for a permanent
visual indication and signal control. LOWER OPERATING COSTS
-- Easy and quick maintenance operations
HIGH PERFORMANCE -- Designed for a long service lifetime
-- Innovative and proven ball valve technology with compensation -- A complete and compact system for better space
-- Smart choice of material in conformance with EN ISO 15001 efficiency in hospital technical areas

ALL RIGHTS OF CHANGE RESERVED


SECTION VALVES 41

SPECIFICATIONS

Inlet Ports G 1/2” Female Leak rate 10-3 mbar l/s He Working Pressure 10 bar max.

Outlet Port G 1/2” Female Weight ± 2.8 kg Orifice size Ø10mm

Service port Plug G3/8” Gauge 0-16 bar Temperature -20°C to +60°C
range

4 5 6 1 P RESSURE SENSOR 5 PRESSURE GAUGE


CONNECITON -- G1/4” (standard)
-- G1/4” (standard) -- w/ Maintenance system
-- w/ Maintenance system (option)
(option)
6 OUTLET CONNECTION
2 ISOLATION VALVE -- G1/2” (standard)
-- Compression tube fitting
3 3 INLET CONNECTION (option)
-- G1/2” (standard)
-- Compression tube fitting
(option)

4 S ERVICE CONNECTION
2 1 -- Plug G3/8” (standard)
-- Medical connector (option)

PRODUCT CONFIGURATOR

Working Pressure Gas Type End Connections Service Connection Gauges Pressure Sensor Maintenance System
ALPIDIST 5 02 G12 NF 1 A SS
5 bar 5 O2 O2 G1/2” - Female G12 AFNOR Medical NF with gauge - 1 without pressure A without SS
72.5 PSI Connector standard sensor maintenance sytem
10 bar 10 Air AIR Tube Fitting T12 DIN Medical DIN with inductive 2 with pressure B with maintenance MS
145 PSI 12mm Connector contact gauge sensor system
N2O N2O Tube Fitting T16 NIST Medical NIST Tube Fitting T16
16mm Connector 16mm
CO2 CO2

N2 N2

ALL RIGHTS OF CHANGE RESERVED


42 NON RETURN VALVES

M855 ALPI – NRV LP | LOW PRESSURE NON RETURN VALVE

APPLICATIONS KEY FEATURES


Low pressure non-return
valve for medical gas -- Used to prevent back feeding from one source to -- Medical certified according to MDD 93/42/EEC
distribution systems another one amended 2007/47/EC
-- For all medical gases -- Avoid reverse flow to disturb pressure level -- Product complies with EN ISO 7396-1
-- Soft sealing monitoring on sources (different pressure settings)
-- Spring loaded -- Used for automatic activation of the emergency
-- No vibration source

KEY ADVANTAGES

Safety First
-- Smart choice of materials in conformance
with ISO 15001 and the most restrictive
additional requirements
-- Low pressure drop for a very limited impact
on your installation performances
-- Suitable for all types of medical gases
-- Soft sealing for a proper tightness level

Installation and Operating Costs


-- In-line design for a perfect integration
into the medical pipeline
-- Spring loaded design – No orientation impact
-- Flexible – Identical design for all medical gases
-- Product design for a long service
lifetime without maintenance

OPTIONS

Tube fitting
Double ring tube
fittings Ø12 mm and Ø16mm

SPECIFICATIONS

End Connection G3/4” Female Leak rate < 1.7 X 10-3 mbar.l/s He Max. working 25 bar / 362.5 PSI
(Inlet/Outlet) Pressure
Seat EPDM Temperature -20°C to +65°C Opening Pressure 1 bar / 14.5 PSI
range
O-ring EPDM Dimensions Ø35 x 68 mm Orifice size 9.1 mm

Spring CuSn 6 Weight ± 0.350 kg Cv 0.95

PRODUCT CONFIGURATOR

Max. Inlet Pressure End Connections


ALPINRV-LP 25 G34
25 bar 25 G3/4” - Female G34
362.5 PSI
Tube Fitting T12
12mm
Tube Fitting T16
Others versions and possibilities
available upon request
16mm

ALL RIGHTS OF CHANGE RESERVED


NON RETURN VALVES 43

ALPI-NRV HP | HIGH PRESSURE NON-RETURN VALVE

KEY FEATURES KEY ADVANTAGES


High pressure non-return
valve for medical gas -- Used to prevent any high pressure exposure in case of Safety First
distribution systems flexible hose or pigtail failure -- Smart choice of materials in conformance
-- For all medical gases -- Flexible product able to be connected on switchover with ISO 15001 and the most
-- Metal-Metal Sealing board directly or cylinder extension -- restrictive additional requirements
-- Integrated filter -- Resistant to Oxygen adiabatic compression -- Suitable for all types of medical gases
-- No vibration -- Integrated filter for a proper protection against -- Resistant to Oxygen adiabatic compression
contamination -- Integrated filter for a proper protection
against contamination
REGULATORY
Installation and Operating Costs
-- Medical certified according to MDD 93/42/EEC -- Ready-to-install elements
amended 2007/47/EC -- O-ring sealing connection, Only a simple
-- Product complies with EN ISO 7396-1 wrench requested to install
-- Flexible thanks to its universal design
whatever the configuration is
Regional version
standard version

HIGH PRESSURE
NON-RETURN VALVE

SPECIFICATIONS

Material Brass External leak rate < 10-4 mbar.l/s He Outlet connections G 3/8” M
G 3/8” F
Filter Bronze Temperature range -20°C to +60°C CV 0.74 (Ø6mm)
(acc. versions) 0.43 (Ø4mm)
Internal leak rate < 2 Lpm Inlet connection G 3/8” Male or regional Weight ± 0.3kg
(closed) standard

PRODUCT CONFIGURATOR

Max. Inlet Pressure Inlet Connections Outlet Connection


ALPINRV-HP 200 M16 G38F
200 bar 200 M16 x 125 - O2 M16 G3/8” Female G38F
2900 PSI
M24 x 250 (Air) M24 G3/8” Male G38M

M27 x 150 (N2O) M27

M20 x 250 (CO2) M20

G3/8” Male G38

ALL RIGHTS OF CHANGE RESERVED


44 TERMINAL UNITS

ALPI-CONNECT | MEDICAL OUTLET SYSTEM

KEY FIGURES APPROVAL


-- ALPI-CONNECT is the
terminal unit at the end -- Base equipped with a valve for carrying out any -- approved in accordance with European Directive
of the medical gas supply maintenance operations on the wall outlet without 93/42/EEC
system disconnecting the supply line -- Meets the requirements of EN ISO 9170-1
-- Enables link from patient to -- Plug head equipped with a double polarizing system -- Meets the requirements of EN ISO 7396-1
flexible pipes, respiratory eliminating the risks of confusion or delivery of
machinery, flow meter .. for the the wrong fluid according to the standards of each
administration of gas therapy country
-- Available in 8-bar version -- Various available accessories and configurations to
for supplying surgical tools cover all types of fixation:
·· Bed head unit
-- Available in vacuum ·· Integrated
version (under pressure) for ·· On panel mounted
medical suction
-- Compatible with all types
of medical fluids

ADVANTAGES

SAFETY TOTAL COST OF OWNERSHIP


-- Base and connection head polarized for each gas to eliminate the risk of -- Reduced frequency and time of maintenance
mix-ups during use and during maintenance operations -- Minimum parts to be replaced
-- Smart choice of material in accordance with EN ISO 15001 -- No need to interrupt the gas supply and treatment of patients with an
·· No aluminium alternative solution.
·· No plastic materials containing fluorine and ·· Sealing exclusively metal/metal at the base block
releasing toxic products in event of ignition
-- Longer lifetime due to reliable construction

HOW TO COMPOSE YOUR SOLUTION

According to available configurations

+ + =

Base block Socket Accessories kits Complete solution

ALL RIGHTS OF CHANGE RESERVED


TERMINAL UNITS 45

ALPI-CONNECT | 1. BASE BLOCK

TECHNICAL SPECIFICATIONS

Body Brass Gas O2 / Air 4 bar / Air 8 bar / Inlet connection - Ø 8/10 mm
N2O / CO2 / N2 / Vacuum - Tube Cu Ø 8/10 x2 – L150 mm
- Connection hose Ø 6 mm
Ball Stainless steel Temperature range -20°C to +60°C Outlet connection Coding by gas type
– attachment by 2 CHC screws

Spring Stainless steel Max. 12 bar Dimensions Ø 45 mm x L18,7 mm


working pressure

VERSIONS / DIMENSIONS

Piping connection Flexible connection

Rear inlet

A = 23,7 0B = 8/10 A = 18,7 0B = 6

Vertical inlet

A = 23,7 0B = 6
A = 18,7 0B = 8/10

Oblique inlet 45°

A = 18,7 0B = 8/10

ALL RIGHTS OF CHANGE RESERVED


46 TERMINAL UNITS

ALPI-CONNECT | 2. SOCKET

TECHNICAL SPECIFICATIONS

Body Chrome-plated brass Gas O2 / Air 4 bar / Air 8 bar / Inlet connection Coding specific to gas type
N2O / CO2 / N2 / Vacuum – attachment by 2 CHC screws
Valve Seal EPDM Temperature range -20°C to +60°C Outlet connection In accordance with the
NFS90-116 and FDS90-119
standards
Filter 60 μm Max. 12 bar Dimensions Ø 47 mm x L 38 mm
working pressure

AVAILABLE VERSIONS ACCORDING TO STANDARDS AND GAS TYPES

O2 Vacuum Air Dimensions

DIN 13260-2

BS 5682 : 1998

AFNOR NF S90-116
IFDS 90-119

SS 875 24 30

JIST 7101

ALPI-CONNECT | 3. ACCESSORIES

NUT

AFNOR version for ceiling pendent


(+indentification ring)

A = 18,7 B = 36,6

ALL RIGHTS OF CHANGE RESERVED


TERMINAL UNITS 47

ALPI-CONNECT | 3. ACCESSORIES

HOUSING AND COVER ASSEMBLY


3 AVAILABLE CONFIGURATIONS: INTEGRATED / ON PANEL MOUNTED / BED HEAD UNITS 1 KIT 1: NUT ONLY
Metal locking screw included
2 KIT 2: COVER+NUT

Cover
3 KIT 3: COVER+NUT+BOX

4 BOX KIT

Nut Box Box kit

ALPI-CONNECT | PRODUCT SELECTION TOOL


PRODUCT CONFIGURATOR

Type Standard Gas Inlet connection Inlet orientation Configuration


ALPI-CONNECT NF 02 T10 AR BH1
NFS 90-116 NF 02 02 Free 08 D8 Vertical VE Bed head unit with BH1
cover
DIN 13260-2 DIN Air 4 bar A4 Free 010 D10 Oblique 45° 45 Bed head unit without BH2
cover
SS 875 2430 SS Air 8 bar A8 Pre-tubed 08 T8 Rear AR Ceiling pendent CP

BS 5682:1998 BS C02 CO2 Pre-tubed 010 T10 Integrated INT

JIS T7101 JIS N2 N2 Connecting hose HB6 Wall mounted WM


06
Vacuum VAC

TYPE OF MOUNTING

Bed head unit Bed head unit


Projection Wall mounted Ceiling pendent*
with cover without cover
Base inlet VE – Tube Cu x x x x /
Base inlet 45° - Tube Cu / x (or) x (or) x (or) /
Base inlet AR – Tube Cu / x (or) x (or) / /
Base inlet VE – Connecting hose / / / / x
Base inlet AR – Connecting hose / / / / x (or)
Connector head x x x x x
Box kit 4 / / x x(others) /
Kit 1 1 / x / / /
Kit 2 2 x / x / /
Kit 3 3 / / / x(AFNOR) /
* See nut accessories for AFNOR model only

ALL RIGHTS OF CHANGE RESERVED


48 CONNEXIONS

M859 ALPI | MULTI CYLINDER EXTENSIONS

KEY FEATURES REGULATORY


Left or right, 2 or 3
cylinders extension for -- High pressure header to connect cylinder batteries -- Medical certified according to MDD 93/42/EEC
medical gas and switch available for various gases amended 2007/47/EC
over boards and -- Resistant to Oxygen adiabatic compression -- Product complies with EN ISO 21969
emergency modules -- Standard G3/8” F connections inlet/outlet
-- 2 or 3 cylinder version
EXTENSIONS -- Delivered with plate support

★ 2 or 3 cylinders version
★ Non-return valve installed
at each manifold inlet
connection
Special requirements on request

CYLINDER EXTENSION

KEY ADVANTAGES

SAFETY AND PERFORMANCE INSTALLATION AND OPERATING COSTS


-- Smart choice of materials in conformance with ISO 15001 and the most -- Ready-to-install elements
restrictive additional requirements -- Only a simple wrench requested to install
-- Suitable for all types of medical gases -- Flexible thanks to its universal design whatever the configuration is
-- Resistant to Oxygen adiabatic compression -- 2 and 3 cylinders versions can be easily combined together to use
-- O-ring sealing for a proper tightness level required number of cylinders
-- Integration of hole at each inlet for safety cable fixation

ALL RIGHTS OF CHANGE RESERVED


CONNEXIONS 49

SPECIFICATIONS

Material Brass/copper Temperature range -20°C to +60°C End connections G 3/8” Male

O-ring EPDM Internal diameter Ø4 mm Plate Stainless steel

Leak rate < 10-7 mbar.l/s He Connections 2 or 3 cylinders Weight ± 1.6 kg (2C)
± 2.8 kg (3C)

PRODUCT CONFIGURATOR

Material Type End Connections


ALPIMULTI B 2C G38
Brass B 2 Cylinders 2C G3/8” Male G38

3 Cylinders 3C

ALL RIGHTS OF CHANGE RESERVED


50 CONNEXIONS

M858 ALPI-LINK CONNECTION PIGTAILS

APPLICATIONS KEY FEATURES


Straight or elbow pigtail
ideally suited to connect - Pre-defined dimensions for a perfect positioning - Medical certified according to MDD 93/42/EEC
ALPI-SWITCH / ALPI-SAFE - Resistant to Oxygen adiabatic compression amended 2007/47/EC
to extensions or - Standard G3/8” F connection - Product complies with EN ISO 21969
extensions together in
order to connect more gas
cylinders

PIGTAIL MANIFOLD-EXTENSION

PIGTAIL EXTENSION-EXTENSION

KEY ADVANTAGES

SAFETY AND PERFORMANCE INSTALLATION AND OPERATING COSTS


- Smart choice of materials in conformance with ISO 15001 and the most - Ready-to-install elements
restrictive additional requirements - Only a simple wrench requested to install
- Suitable for all types of medical gases - Flexible thanks to its universal design whatever the configuration is
- Resistant to Oxygen adiabatic compression

ALL RIGHTS OF CHANGE RESERVED


CONNEXIONS 51

SPECIFICATIONS

End Connection G3/8” Female Temperature -20°C to +60°C Tube size Diameter 10x2,5
(Inlet/Outlet) range
Leak rate < 10-7 mbar.l/s He Max. working 200 bar
Pressure 2900 PSI
Material Copper brass Weight ± 0.165kg (EE)
± 0.295 kg (ME)

PRODUCT CONFIGURATOR

Material Type End Connections


ALPILINK SS EE G38
Cooper CU Extension- EE G3/8” Female G38
Extension
Manifold-Extension ME

Extension corner EC

ALL RIGHTS OF CHANGE RESERVED


52 CONNEXIONS

CYLINDER PIGTAILS

KEY FEATURES OPTIONS


Straight or elbow pigtail
ideally suited to connect -- Cylinder connector according the following standard: -- Different outlet connection
supply boards or switch over -- AFNOR, DIN, NEN, UNI,… -- Shut off valve
boards to gas cylinders -- Other connections: on demand
-- Outlet connections: G 3/8 - Female
-- Material: stainless steel, electro polished
PIGTAILS
★ high pressure
★ straight or elbow
★ stainless steel, electro
polished
Special requirements on request

STRAIGHT VERSION ELBOW VERSION

PRODUCT CONFIGURATOR

Gas Type Cylinder Connection Outlet connection Verion


MED PIGTAIL O2 NF G6 S
O2 O2 NF 29-650 NF G3/8” - Femalew/ G6 Straight version S
rotating nut
Air AIR DIN 477-1 DIN acc. to French NF Elbow version E
recommendation
N2O N2O BS 341-3 BS

CO2 CO2 NEN 3268 NEN

N2 N2 UNI 11144 UNI

CGA V1 CGA

ALL RIGHTS OF CHANGE RESERVED


CONNEXIONS 53

ALPI-FLEX | CYLINDER FLEXIBLE HOSES

KEY FEATURES OPTIONS


Flexible hoses for various
pressures used for connecting -- Stainless steel hose -- Without safety cable version
supply boards, switch over -- Compatible with neutral and corrosive gases -- Elbow version
boards and other equipment according to the hose type.
at the source of gas supply -- The hose is composed of a stainless steel double REGULATORY
braid, a stainless steel or PTFE inside, and end ISO 21969
FLEXIBLE HOSES connections.
-- The hose is standardly equipped with a stainless steel
★ high pressure
safety cable as a safety best practice.
★ stainless steel

Special requirements on request

10

2 1
4 5 6 7 8 9

PRODUCT CONFIGURATOR

Inner Diameter Length Gas Type Cylinder Connection Outlet connection Options
AFLEX DN6 1500 O2 NF G6 C
6 mm DN6 1500 mm 1500 O2 O2 NF 29-650 NF G3/8” - Femalew/ G6 Safety cable C
rotating nut
10 mm DN10 3000 mm 3000 Air AIR DIN 477-1 DIN acc. to French NF Elbow on cylinder B
recommendation side
N2O N2O BS 341-3 BS

CO2 CO2 NEN 3268 NEN

N2 N2 UNI 11144 UNI

CGA V1 CGA

ALL RIGHTS OF CHANGE RESERVED


54 ACCESSORIES

PRESSURE SENSOR

KEY FEATURES
-- Thin film pressure
transducer -- Total error band includes linearity, repeatability, -- Temperature range limited to -15°C to +85°C by use
hysteresis and temperature errors in temperature of mating connector
range -30°C to 100°C. -- M12 - 4 pole
-- Response time: 5 ms max to 63% of full scale pressure -- Sensor meets ROHS requirements.
with step change on input -- CE-conformity according to EMC directive 2004/108/
-- Vibration: iec 60068-2-6 fc vibrating (sinusoidal) 20g eg emission and immunity
-- Operating life: 1 million full pressure cycles min
-- Media: all class II fluids and gases compatable with
stainless steel OPTIONS

-- Pressure Input pressure range: 10 / 16 / 250 barG

1 2 3 4

M12 CONNECTOR BUILT IN ELE

with the indications of the portions of this data which are subject to such limitations shall be included on any reproduction hereof which includes any part of the portions subject to such limitations.
government (B) used in whole or in part by the government for manufacture, or, in the case of computer software documentation, for preparing the same or similar computer software, or (C)used

disclosure, or (2)release to a foreign government, as the interest of United States may repair or overhaul work by or for such government under the conditions of (1)above. This legend, together
available to enable timely performance of the work, provided that the release or disclosure hereof outside the government shall be made subject to a prohibition against further use, release or
Those portions of this technical data indicated as limited rights shall not, without the written permission of Kavlico, be either (A)used, released or disclosed in whole or in part outside of the

by a party other than the government, except for (1)emergency repair or overhaul work only, by or for the government, where the item or process concerned is not otherwise reasonably
MODEL SHOWN A
DIN 175301-803 A (18mm)

35 max.
LABLE LA

23 max.
27 -0.3 HEX 27 -0.3 H

(2) (3)

(4) (1)
C

18 mm PIN CALL OUTS


OUTPUT PIN 1 PIN 2 PIN 3 PIN 4
4-20 mA Vsup Iout N/C N/C
0.5-4.5 VDC ratiometric Vsup Vout N/C GND 0.5-4.5
0-5 VDC
Vsup Vout N/C GND
0-10 VDC

D
BUILT-IN ELECTRICAL PRESSURE CONNECTION
CONNECTOR PORT PRESSURE CONNECTION PORT OPTIONS
5 6 7 8

G1/4A DIN 3852-E 7/16-20 UNF-2B (FEMALE) 7/16-20


CTRICAL CONNECTOR OPTIONS SAE J1926/1 (MODIFIED) SAE J19
(PRESSURE PORT G1/4A SHOWN) WITH 45° CONE AND SCHRADER WIT
A
S 207 INDUSTRIAL STANDARD (9.4 mm) M12 - 4 POLE
5.60 ±0.05

E
2
2
34 max.
32 max.

LABLE
21.5 max.

E
24 max.

12.70
11.30
7.65
12 ±0.20

9.60 ±0.10

27 -0.3 HEX 3
8.90 ±0.10

X
12 ±0.20

90
3 °
B
G1/4A 0 7/16-20 UNF
2 - 0.10 0
0
18.90 - 0.20 7/16-20 UNF 18.90 - 0.20
(2) F 15.88
(2) (3) 18.90
(3)

1 2 3 4

(1)

(4)
(4) (1)
C

9.4 mm PIN CALL OUTS M12 PIN CALL OUTS


UTPUT PIN 1 PIN 2 PIN 3 PIN 4 OUTPUT PIN 1 PIN 2 PIN 3 PIN 4
20 mA Vsup Iout N/C N/C 4-20 mA Vsup N/C Iout N/C
DC ratiometric Vsup Vout N/C GND 0.5-4.5 VDC ratiometric Vsup N/C Vout GND
5 VDC 0-5 VDC
Vsup Vout N/C GND Vsup N/C Vout GND
10 VDC 0-10 VDC

ALL RIGHTS OF CHANGE RESERVED

2:1
LABLE EXAMPLE:
ACCESSORIES 55

SPECIFICATIONS

Inlet pressure max. 10 / 16 / 250 bar Calibrated output 4-20mA Inlet connection G1/4A
signal (Linear) vs applied
pressure(0...100%)
Operating -30°C to 100°C Operating supply 8-30 VDC Protection class M12: IP67
temperature range

Storage temperature -30°C to 100°C External O-Ring Flourocarbon fkm (viton®)


range

FIGURE 1: TOTAL ERROR BAND


TOTAL ERROR BAND IN % OF FULL SPAN

Temp °C

PRODUCT CONFIGURATOR

Pressure Built in electrical


Pressure Range Output External O-Ring connection port connector*
PTE 5000 010 1 B 1 A
0-10 barG 010 4-20 mA 1 Flourocarbon fkm B G1/4 A 1 DIN 175301-803 A A
(viton®) DIN 3852-E (18 mm)
0-16 barG 016

0-250 barG 250


*Mating connector to be ordered separatly

ALL RIGHTS OF CHANGE RESERVED


56 ACCESSORIES

CONTACT PRESSURE GAUGES

KEY FEATURES Sliding contact gauge


Spare part contact pressure -- Normally Open (NO)
gauges for ROTAREX -- Vertical (6 o’clock) or rear mounting -- Accuracy class: 2,5
regulators, points of use, -- Connections: -- Adjustment of switching point with a key
supply boards or switch over M10 x 1 - Male, 1/4 NPT - Male or G 1/4 - Male -- Contact load ≤10 Watt / 10 VA.
boards -- Many pressure ranges available -- Switching current min. 20 mA, max. 500 mA.
-- Material: cuprous alloy or stainless steel -- Cable length 2 m, cable outlet left-hand
-- Nominal diameter: Ø50 mm -- Cannot be used with explosive or combustive gases

Inductive contact gauge


-- Normally Open (NO)
-- Accuracy class: 2,5
-- Adjustment by twisting of contact hood
-- Contact-free “contact release” without wear
-- Cable length 2 m, cable outlet right-hand
-- Compatible with explosive or combustive gases

Available with vertical or rear


mounting connections (normally open)

NORMALLY OPEN CONTACT PRESSURE GAUGE, Ø50 M10 X 1 MALE VERTICAL CONNECTION
Diameter Scale Material Connection Male thread Contact KIT part number
Ø50 0 + 16 bar Cuprous alloy Vertical M10 x 1 Inductive On demand
Ø50 0 + 300 + 400 bar Cuprous alloy Vertical M10 x 1 Inductive 360021990001
Ø50 0 + 400 bar Cuprous alloy Vertical M10 x 1 Sliding 390000990013

NORMALLY OPEN CONTACT PRESSURE GAUGE, Ø50 M10 X 1 MALE REAR CONNECTION
Diameter Scale Material Connection Male thread Contact KIT part number
Ø50 0 + 16 bar Cuprous alloy Rear M10 x 1 Inductive 390001990004
Ø50 0 + 40 bar Cuprous alloy Rear M10 x 1 Inductive 299178990028
Ø50 0 + 100 bar Cuprous alloy Rear M10 x 1 Inductive On demand
Ø50 0 + 100 bar Cuprous alloy Rear M10 x 1 Sliding On demand
Ø50 0 + 250 bar Cuprous alloy Rear M10 x 1 Sliding 390000990011
Ø50 0 + 250 bar Cuprous alloy Rear M10 x 1 Inductive 390000990012
Ø50 0 + 400 bar Cuprous alloy Rear M10 x 1 Inductive 390003990002
Ø50 0 + 400 bar Cuprous alloy Rear M10 x 1 Sliding On demand
Ø50 0 + 250 bar Cuprous alloy Rear M10 x 1 Sliding On demand

ALL RIGHTS OF CHANGE RESERVED


ACCESSORIES 57

BA 12 | ALARM BOX

KEY FEATURES OPTIONS


-- Signal sent automatically
for notifying gas shortage. -- Detecting a drop in pressure when the gas bottle is -- Intrinsically safe barrier for Ex environment (Isolating
The message is visual and empty switching amplifier)
acoustic -- Messages are displayed visually by LEDs and audibly -- 230V AC or 115V AC power supply
-- Optional EX protection buzzer
(installation outside -- Remote message with potential free contacts possible
Ex-area) KEY ADVANTAGES
-- Inputs for magnetic spring contact and inductive
-- Devices available in three contact pressure gauge are suitable. Only NC contacts
versions: For 2, 6 and 10 -- Product flexibility : three versions available according
for safety! to your process( 2, 6 or 10 contacts )
pressure gauge -- Plastic case with IP65 seal for wall and panel -- Potential-free output as change-over contact
mounting -- Group message and New value message
ALARM BOXES -- Cage clamp connection and pluggable -- Integrated LED allow visual information
-- Easy to configure when the device is closed -- Integrated Buzzer for acoustic alarm
★ 2/6/10 contacts
★ Ex Version

Special requirements on request

BA 12-02 BA 12-10 BA 12-06

SPECIFICATIONS

Voltage 230 VAC/50 Hz Type 1 Potential free relay contact Connection 2-storey cage clamps
115 VAC/60 Hz

Power < 3VA Rating 8A/230 VAC Terminal voltage 10 VDC/10mA (unstabilized)
w/ resistive load 0.9

Function Group Message Material ABS

Type 2 For external horn or lamp Protection IP 65

Rating 8A/230 VAC Dimension 200 x 120 x 75 (mm)


w/ resistive load (W x H x D)

Temperature range 0°C to 55°C


32°F to 131°F

PRODUCT CONFIGURATOR

Contacts Voltage Ex protection


BA12 02 230 0
2 contacts 02 230 VAC 230 Without 0

6 contacts 06 115 VAC 115 With EX

10 contacts 10
Others versions and possibilities available upon request

ALL RIGHTS OF CHANGE RESERVED


58 ACCESSORIES

GAS CYLINDER HOLDERS

Designed for the storage of


one or large number of gas
cylinders in an appropriate
area

★ Can be fixed permanently


to the wall
★ Securely holds cylinder in
place
★ Allows permanent
designation of appropriate
cylinder storage area
★ Delivered with a fixing belt
★ Many cylinder holders can
be used together, side by
side
★ Part number: 202500000007

Special requirements on request

Rear view

ALL RIGHTS OF CHANGE RESERVED


ACCESSORIES 59

PROTECTION COVER KIT

KEY FEATURES
Universal protection cover
kit for : -- PETG transparent plastic cover
-- ALPI-REG 1 -- Stainless steel wall fixation
-- ALPI-REG 2 -- Rubber elements to fit wall ruggedness
-- ALPI- MSV

★ Securely hold equipment in


place with 4 fixation point on
a robust plate
★ Avoids access for non-
allowed people
★ Guarantees clear and quick
visibility of the product status
(on/off)
★ Protection cover assembling/
dismantling without any tool
★ Part number : 059900454

ALL RIGHTS OF CHANGE RESERVED


60 ACCESSORIES

OMEGA PLATE

-- Wall Fixation support for :


• ALPI-REG 1
• ALPI-REG 2
• ALPI- MSV
• ALPI-DIST
-- Supplied with rubber
elements to fit wall
ruggedness
-- Securely hold equipment
in place
-- 2 existing dimensions
• 100 mm height
• 160 mm height
-- Made of stainless steel

OMEGA Plate 100 mm OMEGA Plate 160 mm


Ref. 059900476 Ref. 059900477

COMPATIBILITY TABLE

ALPI-REG 1 ALPI-REG 2 ALPI-MSV ALPI-DIST


OMEGA Plat 100 mm x x
OMEGA Plat 160 mm x x

ALL RIGHTS OF CHANGE RESERVED


ACCESSORIES 61

MAINTENANCE TOOL CASE

KEY FEATURES
-- Maintenance tool case for
the complete ALPI-LINE -- Tool case
product range maintenance -- Preformed integration foam
operations -- Pressure gauge 16 bar for M825
-- Special keys
★ Easily transportable -- Medical oxygen compatible leak detector
★ All components properly -- Purging and setting device for ALPI-REG
positioned and integrated in
foam and case and properly
identified
★ Quick and more effective
maintenance operations
★ Part number : 294500000000

ALL RIGHTS OF CHANGE RESERVED


A FULL LINE OF GAS
CONTROL SOLUTIONS

COMPLETE SOLUTIONS FROM SOURCE TO PROCESS.


ROTAREX is helping engineers worldwide to get better gas results: from ultra high purity production
and medical care facilities to industrial and LPG applications, as well as alternative energy vehicles,
fire suppression, diving, aerospace, cryogenics, laboratory, petro-chemical and welding. ROTAREX
applies over 90 years of know-how and experience to custom design, develop and manufacture the high
performance valves, regulators and fittings to suit your needs, all in one hand. Discover the difference
ROTAREX can make in your world.

CYLINDER VALVES EQUIPMENT FIRETEC AUTOMOTIVE LPG/SRG MEDITEC


ULTRA HIGH PURITY VALVES MEDICAL VALVES & EQUIPMENT INDUSTRIAL CYLINDER VALVES REFRIGERANT CYLINDER VALVES

PRESSURE REGULATORS SUPPLY & SWITCH OVER BOARDS LINE VALVES FITTINGS & ADAPTORS

FIXED INSTALLATION OBJECT FIRE SUPPRESSION AUTOMOTIVE VALVES


FIRE SYSTEMS SYSTEMS & REGULATORS WATER CARBONATION

LPG TANK VALVES LPG CYLINDER VALVES DIGITAL MEASUREMENT


& REGULATORS & REGULATORS SYSTEMS PLASTIC INJECTION MOULDING

AUSTRALIA / BRASIL / CHINA / FRANCE / INDIA / ITALY / JAPAN / LUXEMBOURG / MIDDLE EAST / RUSSIA / SINGAPORE / SPAIN / THAILAND / USA
AUSTRALIA / BRASIL / CHINA / FRANCE / INDIA / ITALY / JAPAN / LUXEMBOURG / MIDDLE EAST / RUSSIA / SINGAPORE / SPAIN / THAILAND / USA
www.rotarex.com

WORLDWIDE HEADQUARTERS

Ceodeux Meditec Customer Service


24 rue de Diekirch, B.P. 19
7505 Lintgen, Luxembourg
Tel.: +352 32 78 32 1
Fax: +352 32 78 32 854
E-mail: info@meditec.rotarex.com

REGIONAL / COUNTRY HEAD QUARTERS

NORTH AMERICA EUROPE ASIA OCEANIA


USA FRANCE, BENELUX, GERMANY MIDDLE EAST INDIA AUSTRALIA
Rotarex North America 24, rue de Diekirch, BP 19 Rotarex Middle East Rotarex ENGG. PVT. LTD. Rotarex Australia/New Zealand
221 Westec Drive L-7505 Lintgen, Luxembourg Jebel Ali Free Zone - Lob 17, Office 415 C-105, Mahindra Apartments Hornsby Heights,
Westmoreland Technology Park I Tel.: +352 32 78 32-1 P.O. Box 261749 Vikaspuri, New Delhi 110018 Sydney, NSW, 2077
Mt. Pleasant, Pennsylvania 15666 USA Fax: +352 32 78 32 317 Dubai, U.A.E. mob.: +91-9873467666 Tel.: +61 477 477 481
Tel.: +1 724-696-43 40 E-mail: salesequipment@rotarex.com Tel.: +971 (0) 4 887 6701 E-mail: atul.nigam@rotarex.in E-mail: info@australia.rotarex.com
Fax: +1 724-696-43 64 Fax: +971 (0) 4 887 6702
E-mail: info@northamerica.rotarex.com ITALY E-mail: info@middle-east.rotarex.com THAILAND
Rotarex Italia S.r.l. Rotarex (Thailand) Co Ltd.
SOUTH AMERICA 46 Via Giacomo Matteotti CHINA 297, Wanglee Tower
I-25080, Ciliverghe di Mazzano (BS) Italy Rotarex Star 3rd floor, Unit C, Surawong Road
BRASIL Tel.: +39 030 212 05 50 60 Yuan Zhong Road Suriyawong, Bangrak,
Rotarex Brazil Ltda Fax: +39 030 212 23 62 Shanghai Nanhui Industrial Zone Bangkok 10500 Thailand
Cond. Ind. Portal da Anhanguera Estr. Municipal E-mail: info@italia.rotarex.com 201300, Shanghai China Tel.: +66 2635 71 79
Gov. Mário Covas, S/N Tel.: +86-21 5800 4000 E-mail: info@thailand.rotarex.com
13279-411 Bairro Macuco -Valinhos SPAIN Fax: +86-21 5800 3226
São Paulo Brazil Rotarex Spain E-mail: info@china.rotarex.com SOUTH KOREA
Tel.: +55 19 3518 0800 C/Alcala, 4° Oficina 404 Rotarex Luz Tech Co.
Fax: +55 19 3869-1503 28014 Madrid SINGAPORE 305-500, 547 Yongsan-dong
E-mail: info@brazil.rotarex.com Tel.: +34 (91) 570 89 05 Rotarex Fareast Pte Ltd Yuseong-gu
Mobile: + 34 (650) 908 856 10 Ubi Crescent Taejon, Korea
E-mail: info@spain.rotarex.com Ubi Techpark, Lobby C, #06-55 Tel.: +82 42 825 8911
408564 Singapore Fax: +82 42 825 8913
RUSSIA Tel.: +65 64 72 37 27 E-mail: info@rotarex.co.kr
Rotarex Rus Fax: +65 64 72 45 28
Tverskaya street, 20/1 bldg. 1 of 506 E-mail: info@singapore.rotarex.com
125009, Moscow Russia
Tel.: +7 985 125 7776 JAPAN
Skype: rotarexrus Rotarex Japan Ltd
E-mail: info-russia@rotarex.com 6F, Yawaragi Bldg., 1-27-11 Taito
Taito-ku, Tokyo
POLAND 110-0016 JAPAN
Rotarex Polska Tel.: +81 3 5817 8108
13 Gröbli Mob.: +81 80 3755 2539
PL 49-300 Brzeg, Poland Fax: +81 3 5817 8168
Tel.: +48 77 416 40 16 E-mail: info@japan.rotarex.com © 2012 - ROTAREX S.A.
Fax: +48 77 416 20 99 CEODEUX MEDITEC - CEODEUX S.A.
E-mail: info@poland.rotarex.com MGSS-EN-1115

Anda mungkin juga menyukai