SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

INNOVATECIS CIA LTDA.


General Jose Maria Guerrero
N69-170 y Alfonso del Hierro
Quito, Ecuador 170103
Diego Almeida Phone: 593 260 40607

CALIBRATION

Valid To: July 31, 2020 Certificate Number: 4038.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 6:

I. Dimensional

Parameter/Equipment Range CMC2 () Comments

Calipers3 (1.0 to 350) mm 9.5 µm Master gauge blocks


(0.0625 to 1.0) in 11 µin
(>1 to 2) in 16 µin
(>2 to 4) in 19 µin
(>4 to 6) in 23 µin

Depth Gauges3 (1.0 to 350) mm 9.5 µm Master gauge blocks


(0.0625 to 1.0) in 11 µin
(>1 to 2) in 16 µin
(>2 to 4) in 19 µin
(>4 to 6) in 23 µin

Depth Micrometers3 (1.0 to 350) mm 0.58 µm Master gauge blocks


(0.0625 to 1.0) in 11 µin
(>1 to 2) in 16 µin
(>2 to 4) in 19 µin
(>4 to 6) in 23 µin

Height Gauges3 (1.0 to 350) mm 9.5 µm Master gauge blocks


(0.0625 to 1.0) in 11 µin
(>1 to 2) in 16 µin
(>2 to 4) in 19 µin
(>4 to 6) in 23 µin

(A2LA Cert. No. 4038.01) 11/14/2018 Page 1 of 10


Parameter/Equipment Range CMC2 () Comments

Indicators3 Up to 25 mm 0.59 µm Master gauge blocks


(>25 to 50) mm 0.61 µm
(>50 to 100) mm 0.89 µm
(0.0625 to 3.8) in 23 µin

Inside Micrometers3 (1.0 to 350) mm 0.58 µm Master gauge blocks


(0.0625 to 0.25) in 7.0 µin
(>0.25 to 1.0) in 11 µin
(>1.0 to 2.0) in 15 µin
(>2.0 to 4.0) in 22 µin
(>4.0 to 6.0) in 27 µin

Outside Micrometers3 (1.0 to 350) mm 0.58 µm Master gauge blocks


(0.0625 to 0.25) in 7.0 µin
(>0.25 to 1.0) in 11 µin
(>1.0 to 2.0) in 15 µin
(>2.0 to 4.0) in 22 µin
(>4.0 to 6.0) in 27 µin

Thickness Gauges3 23 µm to 50 mm 1.0 µm Master gauge blocks


(0.000 92 to 0.038) in 59 µin
(0.0625 to 1.0) in 11 µin
(>1.0 to 2.0) in 16 µin
(>2.0 to 4.0) in 23 µin

II. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 5 () Comments

DC Voltage – Measure (0 to 100) mV 1.0 µV Transmille 8081


(0.1 to 1) V 9.4 µV
(1 to 10) V 96 µV
(10 to 100) V 1.2 mV
(0.1 to 1) kV 13 mV

DC Voltage3 – Generate (0 to 100) mV 12 µV Transmille 3041A


(0.1 to 1) V 82 µV
(1 to 10) V 0.68 mV
(10 to 100) V 8.2 mV
(0.1to 1) kV 82 mV

(A2LA Cert. No. 4038.01) 11/14/2018 Page 2 of 10


Parameter/Equipment Range CMC2, 5 () Comments

Resistance3 – Generate (0 to 10) Ω 7.3 mΩ Transmille 3041A


(10 to 100) Ω 13 mΩ
(0.1 to 1) kΩ 0.12 Ω
(1 to 10) kΩ 1.0 Ω
(10 to 100) kΩ 10 Ω
(0.1 to 1) MΩ 0.20 kΩ
(1 to 10) MΩ 5.4 kΩ

Resistance (2 Wire) – (10 to 200) Ω 98 mΩ Precision multimeter


Measure (0.2 to 2) kΩ 0.54 Ω
(2 to 20) kΩ 4.2 Ω
(20 to 200) kΩ 40 Ω
(0.2 to 2) MΩ 0.8 kΩ
(2 to 20) MΩ 46 kΩ
(20 to 100) MΩ 0.23 MΩ

Resistance (4 Wire) – (0 to 200) Ω 98 mΩ Precision multimeter


Measure (0.2 to 2) kΩ 0.54 Ω
(2 to 20) kΩ 4.2 Ω
(20 to 200) kΩ 39.7 Ω
(0.2 to 2) MΩ 0.8 kΩ
(2 to 20) MΩ 45 kΩ
(20 to 100) MΩ 0.23 MΩ

Capacitance3 – Generate, (10, 20, 50, 100) nF 0.42 nF Transmille 3041A


at 1 kHz 1 µF 6.5 nF
10 µF 0.1 µF

DC Current – Measure (0 to 10) nA 0.16 nA Transmille 8081


(10 to 100) nA 0.32 nA
(0.1 to 1) µA 0.37 nA
(1 to 10) µA 0.62 nA
(10 to 100) µA 1.6 nA
(0.1 to 1) mA 61 nA
(1 to 10) mA 0.62 µA
(10 to 100) mA 5.6 µA
(0.1 to 1) A 0.26 mA
(1 to 10) A 6.2 mA
(10 to 30) A 29 mA

(A2LA Cert. No. 4038.01) 11/14/2018 Page 3 of 10


Parameter/Equipment Frequency CMC2, 5 () Comments

DC Current3 – Generate (0 to 100) µA 0.061 µA Transmille 3041A


(0.1 to 1) mA 0.18 µA
(1 to 10) mA 1.1 µA
(10 to 100) mA 13 µA
(0.1 to 1) A 0.19 mA
(1 to 10) A 5.4 mA
(10 to 30) A 16 mA

DC Current3 – Clamp-On (1 to 19) A 0.7 A Transmille 3041A and


Meters (>19 to 100) A 1.5 A Transmille EA002

(>100 to 500) A 5.3 A Transmille 1000A and


(>500 to 1500) A 15 A Transmille EA002

Parameter/Range Frequency CMC2, 5 () Comments

AC Voltage – Measure

(0 to 100) mV 40 Hz to 20 kHz 0.10 mV Transmille 8081


(0.1 to 1) V 0.86 mV
(1 to 10) V 11 mV
(10 to 100) V 0.10 V
(0.1 to 1) kV 0.96 V

AC Voltage3 – Generate

(0 to 100) mV 45 Hz to 1 kHz 0.18 mV Transmille 3041A


(0.1 to 1) V 1.3 mV
(1 to 10) V 12 mV
(10 to 100) V 0.24 V
(0.1 to 1) kV 2.1 V

AC Current – Measure

(0 to 100) µA 40 Hz to 1 kHz 0.10 µA Transmille 8081


(0.1 to 1) mA 0.66 µA
(1 to 10) mA 15 µA
(10 to 100) mA 0.15 mA
(0.1 to 1) A 1.7 mA
(1 to 10) A 18 mA
(10 to 30) A 46 mA

(A2LA Cert. No. 4038.01) 11/14/2018 Page 4 of 10


Parameter/Range Frequency CMC2, 4, 5 () Comments

AC Current3 – Clamp-On
Meters

(1 to 19) A 50 Hz 0.7 A Transmille 3041A +


(>19 to 100) A 1.5 A Transmille EA002
(>100 to 500) A 5.3 A Transmille 1000A +
(>500 to 1500) A 15 A Transmille EA002

AC Power3 – Generate,
PF = 1

(0 to 100) mA (50 to 60) Hz 1.1 % Transmille 3041A


(0.1 to 2) A (0 to 1) kV 0.56 %
(2 to 3) A (0 to 1) kV 0.57 %
(3 to 30) A (0 to 1) kV 0.84 %

Parameter/Equipment Range CMC2, 4, 5 () Comments

DC Power – Generate

(0 to 100) mA (0 to 1) kV 0.11 % Transmille 3041A


(0.1 to 2) A 0.041 %
(2 to 3) A 0.051 %
(3 to 30) A 0.063 %

Phase – Generate

(20 to 220) Vrms (0 to 20) A, 50 Hz 0.28° Transmille 3041A

(A2LA Cert. No. 4038.01) 11/14/2018 Page 5 of 10


Parameter/Equipment Range CMC2 () Comments

Electrical Simulation of RTD (-100 to 800) °C 0.13 °C Transmille 3041A


Temperature Indicators3

Electrical Simulation of
Thermocouple Temperature
Indicators3 –

Type B (600 to 1000) °C 1.3 °C Transmille 3041A with


(>1000 to 1820) °C 0.94 °C Transmille EA001A

Type C (10 to 650) °C 0.56 °C


(>650 to 2316) °C 1.1 °C

Type E (-240 to -25) °C 0.86 °C


(>-25 to 1000) °C 0.26 °C

Type J (-210 to 0) °C 0.42 °C


(>0 to 1200) °C 0.33 °C

Type K (-190 to -100) °C 0.48 °C


(>-100 to 1370) °C 0.45 °C

Type L (-200 to 900) °C 0.56 °C

Type N (-190 to -25) °C 0.74 °C


(>-25 to 120) °C 0.30 °C
(>120 to 1300) °C 0.41 °C

Type R & S (0 to 400) °C 1.4 °C


(>400 to 1760) °C 0.87 °C

Type T (-240 to 0) °C 1.0 °C


(>0 to 400) °C 0.20 °C

Type U (-200 to 0) °C 0.70 °C


(>0 to 600) °C 0.50 °C

(A2LA Cert. No. 4038.01) 11/14/2018 Page 6 of 10


III. Fluid Quantities

Parameter/Equipment Range CMC2 () Comments

Volumetric Containers >100 µl to 5 ml 1.8 µl Gravimetric method


(> 5 to 10) ml 1.8 µl using analytical
(>10 to 30) ml 2.0 µl balance
(>30 to 50) ml 2.3 µl
(>50 to 100) ml 3.5 µl
(>100 to 200) ml 5.0 µl
(>200 to 2000) ml 6.4 µl
(>2000 to 3 000) ml 0.31 ml
(>3000 to 10 000) ml 0.65 ml
(>10 000 to 25 000) ml 0.87 ml

IV. Mechanical

Parameter/Equipment Range CMC2, 4, 5 () Comments

Gauge Pressure3 – (-12 to 150) psig 0.023 psig (0.17 kPa) Crystal XP2i, Fluke
Pneumatic, Hydraulic (>150 to 325) psig 0.14 psig (0.96 kPa) 700G, Ametek Jofra
(>325 to 500) psig 0.15 psig (1.0 kPa) IPI MK. II, or
(>500 to 1300) psig 0.19 psig (1.3 kPa) Martel BGPIR-
(>1300 to 1900) psig 0.42 psig (2.9 kPa) PRO-10K, Martel
(>1900 to 2500) psig 0.53 psig (3.7 kPa) T-140
(>2500 to 3000) psig 0.87 psig (6.0 kPa)
(>3300 to 5000) psig 1.0 psig (7.2 kPa)
(>5000 to 6000) psig 1.5 psig (11 kPa)
(>6000 to 7000) psig 1.7 psig (13 kPa)
(>7000 to 10 000) psig 1.9 psig (14 kPa)

Torque – Measure Up to 10 N∙m 0.70 % Stahlwille 7728-1,


(10 to 100) N∙m 0.40 % Stahlwille 7728-1S,
(100 to 1000) N∙m 0.22 % Stahlwille 7728-10,
Stahlwille 7728-100

Torque Transducers (1 to 1000) N∙m 0.053 % Arm and M1 class


weights

(A2LA Cert. No. 4038.01) 11/14/2018 Page 7 of 10


Parameter/Equipment Range CMC2 () Comments

Scales and Balances3 1 mg to 250 g 0.60 mg OIML F1 weights

(250 to 600) g 1.5 mg OIML Class E2 weights

(600 to 3500) g 60 mg OIML Class E2 and F1


(3.5 to 50) kg 1.3 g weights

(50 to 300) kg 0.0060 kg OIML Class M1 weights


(300 to 600) kg 0.029 kg

V. Thermodynamic

Parameter/Equipment Range CMC2 () Comments

Baths3 (-25 to 0) °C 0.041 °C PRT probe and


(0 to 140) °C 0.033 °C temperature indicator
(140 to 200) °C 0.066 °C
(200 to 250) °C 0.089 °C
(250 to 660) °C 0.091 °C

Dry Blocks3 (-25 to 0) °C 0.062 °C PRT probe and


(0 to 140) °C 0.057 °C temperature indicator
(140 to 350) °C 0.074 °C
(350 to 660) °C 0.091 °C

Climatic Chambers3 – (-25 to 0) °C 0.9 °C PRT probe and


(Including Oven, (0 to 60) °C 1.1 °C temperature indicator
Incubator, Refrigerator, (>60 to 140) °C 1.3 °C
Freezer, Autoclave,
Sterilizers, Muffles) (>140 to 300) °C 1.6 °C Thermocouple and
(>300 to 660) °C 3.6 °C readout
(>660 to 1050) °C 4.5 °C

Humidity Chambers (25 to 80) % RH 3.1 % RH Control Company 4085


data loggers

Bi-Metal Thermometers3 (0 to 350) °C 0.12 °C PRT probe and


(350 to 660) °C 0.58 °C temperature indicator

(A2LA Cert. No. 4038.01) 11/14/2018 Page 8 of 10


Parameter/Equipment Range CMC2 () Comments

Digital Thermometers (-25 to 140) °C 0.06 °C ISOTECH PRT T100-


(TC or RTD Probes and (>140 to 400) °C 0.080 °C 250 probe with
Indicators)3 (>400 to 660) °C 0.091 °C Transmille 8081

Liquid-In-Glass (-25 to 140) °C 0.30 °C PRT probe &


Thermometers3 temperature indicator

Relative Humidity – (25 to 80) % RH 1.7 % RH Control Company 4085


Thermo-Hygrometer

VI. Time & Frequency

Parameter/Equipment Range CMC2, 5 () Comments

Frequency – Measuring (0 to 100) Hz 5.8 mHz Transmille 3041A


Equipment 100 Hz to 1 kHz 5.8 Hz
(1 to 10) kHz 5.8 Hz
(10 to 100) kHz 5.8 Hz
100 kHz to 1 MHz 0.6 kHz
(1 to 10) MHz 5.8 kHz

Frequency – Measure 1 Hz to 1 kHz 0.1 Hz Transmille 8081


1 kHz to 10 kHz 0.012 kHz
10 kHz to 100 kHz 0.013 kHz
100 kHz to 1 MHz 0.022 kHz

Stopwatches and Timers3 10 s to 1.0 h 0.075 s Reference stopwatch


(>1.0 to 2.5) h 0.095 s
(>2 .5 to 5.0) h 0.14 s
(>5.0 to 10.0) h 0.26 s
(>10.0 to 24.0) h 0.60 s

Tachometers (60 to 60 000) RPM 0.58 RPM Transmille 3041A and


(>60 000 to 100 000) RPM 0.82 RPM LED artifact
___________________________________________
1
This laboratory offers commercial and field calibration service.

(A2LA Cert. No. 4038.01) 11/14/2018 Page 9 of 10


2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.
3
Field calibration service is available for this calibration and this laboratory meets A2LA R104 – General
Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer's site can normally be
expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such
as the environment at the place of calibration and for other possible adverse effects such as those caused
by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced
by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result
in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4
In the statement of CMC, percentages are to be read as percent of reading, unless otherwise noted.
5
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter.
6
This scope meets A2LA’s P112 Flexible Scope Policy.

(A2LA Cert. No. 4038.01) 11/14/2018 Page 10 of 10


Accredited Laboratory
A2LA has accredited

INNOVATECIS CIA. LTDA.


Quito, ECUADOR
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005
General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific
Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a
defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 14th day of November 2018.

_______________________
President and CEO
For the Accreditation Council
Certificate Number 4038.01
Valid to July 31, 2020