6. Calibration
To calibrate the sensor, proceed as follows:
Main
Display
Main
Calibrate
Calibrate
Display
Calibrate
Cal Sensor
Inhibiting 4 - 20
Please Wait
75
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
(7) Wait and the display will show:
Processing Data
Please wait
Zero Calibrated
(10) Either:
Press Enter To
Select Gas Conc.
76
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
This means set the gas concentration (ie. 50% LEL:
for 2.2% methane if working in accordance with
EN61779).
Processing Data
Please wait . . . .
Span Calibrated
77
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
(17) Wait and the display will show:
Press Enter To
Release 4 - 20
4 - 20 released
Calibrate
Cal Sensor
Main
Display
78
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
(23) The Hand-Held Interrogator may be switched off by
pressing esc and ↵ together.
79
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
(2) Connect the brown lead of the SHC Protection Device
to the orange lead of the Optima communications link.
To Optima brown
Communications To Hand-Held
Link Connections Interrogator
in Junction Box blue SHC-1
SHC Protection Device
Z - SHC 1 Optima
Interrogator 4v0
Z - SHC 1 Optima +
Interrogator 4v0
62
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
Z - SHC 1 Excel
Interrogator 4v0
Main
Calibrate
Main
Diagnose
Main
Gas Tables
63
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
(9) Press and the display will show:
Main
Configure
Main
Change Mode
Main
Power Off
64
Z- SHC1 Optima +
Interrogator 4v0
MAN0551
2108M0501
65
Change Mode
Force 4-20 Self Test Select Gas
Calibrate Diagnose
6. Calibration
44
4. ELECTRICAL CONNECTIONS
Issue 08 - 01/09
Standard Configuration for Retrofit Application (With SHC Protection Device and Hand-Held Interrogator Option.)
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
8.1 Troubleshooting
The table below describes some possible faults and problems that can
be encountered and their most likely causes and remedies. The user
should be aware that there are often significant differences between what
appears to be the problem and the actual problem when fully diagnosed.
Experience has shown that a methodical, whole-system orientated
approach is the most effective for solving problems with gas detection
systems and installations.
Fault / Problem Cause / Remedy
Output is 0mA Cause: Unit is in FAULT condition
97
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
98
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
99
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
100
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
101
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
102
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
8.2 DIAGNOSIS OF WARNING AND FAULT MESSAGES
The table below describes the WARNING and FAULT messages which can
be reported by a unit when the DIAGNOSTIC menu options are used on
the Handheld Interrogator. Where possible, appropriate remedial actions
to clear these WARNINGS and FAULTS are identified. Note: If the remedial
action is successful, the warning or fault will automatically clear.
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Inst Guide WARNING: (1) Relocate the unit
away from any intense
Temp Lim Specified operating sources of heat or cold.
Exceded temperature limits nearly
exceeded. (High or low.) (2) Shield the unit from
intense sources of heat
or cold.
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Inst WARNING: (1) Check the 0V and
Guide 24V connections and
Bad 24V Supply The 24V supply voltage cabling to the unit.
is just outside of the
specified limits. (18V (2) Verify that the round-
to 32V) trip resistance of the
supply cable is less
than 25.5 ohms.
104
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Inst (3) Check the 4-20mA
Guide output of the unit as
described in Section
Bad 4-20mA 5.2
FAULT:
Loop
The unit is unable to In addition to the above,
transmit a current check that the unit’s
which represents a gas 4-20mA output is correctly
concentration in excess configured for the
of the units internal controller input.
alarm threshold
W-See Ops WARNING: (1) Clean the optics. If
Guide One of the lamps’ this WARNING does
Lamp Output output has been not clear, the unit’s
diagnosed as being diagnosis of low lamp
close to the minimum output is probably
operating limit. correct.
105
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Ops FAULT: Replace the Optima Plus
Guide The lamp output unit with a spare and return
Lamp Output has fallen below the it to Honeywell Analytics
minimum operating limit. for service or repair.
106
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Ops (2) Zero offsets can
Guide also be caused by
Neg Gas exposing the unit to
Reading extreme temperatures.
Check log for Temp
Lim Exceded entries.
(3) Provided that optics
are clean and dry,
re-zero unit. (See
Section 7.2)
107
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Ops (3) Check that the unit is
Guide not being exposed to
Signal Quality excessive vibration or
temperatures.
108
MAN0551 Issue 08 - 01/09
2108M0501
8. FAULT FINDING
SHC-1
Description Remedial Action(s)
Handheld Text
W-See Ops FAULT (ONLY): (1) Attempt to restart
Guide An internal hardware the unit by switching
Internal Failure fault has been the power off and on
diagnosed. again or selecting
the DIAGNOSTIC >
SOFTWARE RESET
option on the SHC-1.
(2) If the problem
persists, return the
unit to Honeywell
Analytics for repair /
replacement.
109
MAN0551 Issue 08 - 01/09
2108M0501
110
MAN0551 Issue 08 - 01/09
2108M0501
6. Calibration
6.2 SHC-1 HANDHELD INTERROGATOR
The SHC-1 Handheld Interrogator is a portable, Intrinsically
Safe (IS) device which can be used to calibrate, test and
commission all of Honeywell Analytics Searchpoint and
Searchline optical gas detection products. The SHC-1
interrogator communicates with detectors via a digital RS485
link. Electrically, this link can be made either by using the IS
socket on the DVC100 / DX100 or by connecting the flying
leads on the SHC Protection Device (04230-A-1025) to the
appropriate junction box terminals.
WARNING
The IS certification of the SHC-1 Handheld Interrogator
only remains valid if it is used in conjunction with the
Termination Unit or the SHC Protection Device. Do not
attempt to connect the SHC-1 interrogator directly to
units, always use either the Termination Unit or the SHC
Protection Device.
60