MINIDOP ES-100VX
POCKET DOPPLER
CONTENTS
* Features . . . . . . . . . . . . . . . . . . . . . . . . . . 1
* Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . 2
* Clinical applications . . . . . . . . . . . . . . . . . 4
* Operating controls . . . . . . . . . . . . . . . . . . 5
* How to turn the unit ON and OFF . . . . . . . 7
* External outputs . . . . . . . . . . . . . . . . . . . . 8
* Operation . . . . . . . . . . . . . . . . . . . . . . . . . 8
* Probe selection and applications . . . . . . 10
* Principles . . . . . . . . . . . . . . . . . . . . . . . . 11
* Block diagram . . . . . . . . . . . . . . . . . . . . . 12
* Specifications . . . . . . . . . . . . . . . . . . . . . 13
* Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 14
* Warranty . . . . . . . . . . . . . . . . . . . . . . . . . 14
The POCKET DOPPLER MINIDOP ES-100VX can be used with five inter-
changeable probes having different frequencies of 2, 4, 5, 8 and 10 MHz, to
detect fetal heartbeats as well as arterial and venous blood flows in the
extremities.
Please read this manual carefully.
FEATURES
ES-100VX is designed for the use to detect fetal heartbeats and arterial &
venous blood flows in the extremities, and has the following features.
* GREAT SENSITIVITY
1
CAUTIONS
1. General
Non-conformability with expendables/accessories/other device
Please use designated battery, ultrasonic gel, probe and headset. Do not
use those accessories for other devices.
1-1. PROBE
(1) The standard probe is for transcutaneous use only.
(2) The probe transducer is a very thin and delicate part. Please take
care not to drop or hit the probe tip. After use, protect the transducer
with probe cover.
The probe cover also prevents against switching ON when storing the
unit after use.
1-3. BATTERY
(1) When battery is low, the power indicator becomes dark. Also there will
be no speaker sounds. Replace the battery.
(2) Use a 9 volt alkaline square type battery. A non-alkaline may cause a
shortage of power.
(3) When not using the unit for a long time, remove the battery.
2
2. Before using the unit
(1) Do not place near water or other liquid.
(2) Do not place where atmospheric pressure, ventilation, sunlight, dust,
salt, sulfur and so forth will not affect the unit adversely.
(3) Use only on conditions of temperature between 10 and 40 degrees
Centigrade and humidity less than 75 %.
(4) Take care of the stability conditions such as inclination, vibration, and
shock during transportation.
(5) Do not place where chemicals are stored, or where gas may be gen-
erated.
(6) Keep the unit and the patient from a computer more than 1.5 m apart.
(7) When using the unit after a long time, do not forget to check whether
the unit operates normally and safely.
(8) Make sure that there are no damages on the unit and probe.
8. Destruction
(1) In case of destruction of the unit, follow the instructions for disposition
of the destruction appointed by each country or local government.
CLINICAL APPLICATIONS
4
OPERATING CONTROLS
Front panel
5
Rear panel
6
HOW TO TURN THE UNIT ON and OFF
5. REPLACING BATTERY
Replace the battery with new one if the power indicator becomes
dark. Use a 9 volt ALKALINE square type battery. A non-alkaline may
cause a shortage of power.
7
EXTERNAL OUTPUTS
1. HEADSET
Connect the headset when necessary. The headset cuts off the
speaker.
OPERATION
8
(4) Put the probe on the measurement area
and move it slowly to locate the point
where the maximum Doppler sounds
are heard. An ideal probe angle to the
vessel is approximately 60 degrees.
9
PROBES and APPLICATIONS
* CLINICAL APPLICATIONS
10
PRINCIPLES
The voltage is amplified. The electronics amplifies the signal, filters out
noise and improves the S/N ratio. The Doppler shift signals are amplified
further and converted to audible sound via a speaker.
11
BLOCK DIAGRAM
12
SPECIFICATIONS
Manufacturing date : The first 2 digits and following 2 digits of the serial
number represent the year and month of manufac-
turing, respectively.
The serial number is located inside of the battery
compartment and it consists of 4 to 8 digits and
may start with "Serial number" or "S/N".
Examples:
03020001: Feb/2003
0401: Jan/2004
13
Accessories: Carrying case
(Optional) Ultrasonic gel
Battery
Headset
Camera strap
CLEANING
1. PROBE
Remove the Doppler gel from the probe head after use.
Clean the probe using damp cloth and then wipe with a soft dry cloth, but
take great care that any water may not penetrate into the probe.
If using disinfectant, please consult in advance with the manufacturer.
2. MAIN UNIT
To clean the main unit, use a damp cloth and then wipe with a soft dry
cloth, but take great care that any water may not penetrate into the unit.
WARRANTY
This equipment is guaranteed for a period of one year after the date of
purchase when used under normal conditions.
In the event of any trouble during the warranty period, please contact the
dealer from who you purchased the unit.
In case the warranty period is over, please consult the dealer for a charged
service.
14
Hadeco, Inc.
7-11 Arima 2-chome, Miyamae-ku, Kawasaki, 216-0003 Japan
100VX OM E-26
August 25, 2004
Printed in Japan