Anda di halaman 1dari 12

Operation Manual

开发手册

Products Name(产品名称): SpO2 Module

Model No.(模块型号): SNR-S-5


Revision(版本): 1.3
Issued Date(日期): 2011-6-6

1
Chapter 1. Characteristic ................................................................................................................. 3
Chapter 2. Range and accuracy ....................................................................................................... 4
2.1. Range................................................................................................................................. 4
2.2. Resolution.......................................................................................................................... 4
2.3. Accuracy............................................................................................................................ 4
Chapter 3. Environmental Requirements .................................................................................... 6
3.1. Temperature .................................................................................................................... 6
3.2. Humidity ........................................................................................................................... 6
3.3. Altitude .............................................................................................................................. 6
Chapter 4. Power Specifications .................................................................................................. 7
4.1. Power Specification ........................................................................................................ 7
Chapter 5. Communication Protocol............................................................................................... 8
5.1. Communication Protocol ................................................................................................... 8
Chapter 6. Connections ................................................................................................................. 10
6.1. Connections ..................................................................................................................... 10
6.2. Probe connector J3 .......................................................................................................... 10
6.3. Host Connector J4 ........................................................................................................... 10
6.4. Probe connector J1 .......................................................................................................... 11
6.5. Host Connector J2 ........................................................................................................... 11
6.6. Probe ............................................................................................................................... 11
Chapter 7. Mechanical .................................................................................................................. 12
7.1. dimension Size ............................................................................................................... 12

SpO2 Module 2
Chapter 1. Characteristic

 Strong anti-electromagnetic interference

 Dimension compatible with Nellcor & BCI oximeter module

 Fast & Accurate to measure SpO2 and Pulse Rate

 CE standards

SpO2 Module 3
Chapter 2. Range and accuracy

2.1. Range

Pulse rate: 25 ~ 250 bpm

SpO2: 35 ~ 100%

2.2. Resolution
Pulse rate: 1 bpm

SpO2: 1%

2.3. Accuracy

Pulse rate:(25~250BPM)

2bpm Normal

3bpm Motion or low perfusion

SpO2:(70~100%)

2% Normal

3% Motion or low perfusion

Below 70% Not specified

2.4. Low Perfusion


SpO2: 0.1%
PR: 0.1%

SpO2 Module 4
SpO2 Module 5
Chapter 3. Environmental Requirements

3.1. Temperature
Operating temperature: 5~40 °C

Storage temperature: -20~55°C

3.2. Humidity

Operating Humidity:15%~85%, non-condensing

Storage Humidity:10~ 95 %, non-condensing

3.3. Altitude
Operating Altitude:70~106kPa

Storage Altitude: 50~106kPa

SpO2 Module 6
Chapter 4. Power Specifications

4.1. Power Specification


Par. Min. Typ. Max.

5V +4.75V +5V +5.25V

-5V -4.75V -5V +5.25V

Ripple --- --- 100mVp-p

Current --- 50mA ---

SpO2 Module 7
Chapter 5. Communication Protocol

5.1. Communication Protocol


SNR-S-5 Communication Protocol compatible with BCI. It communicates
with the host via a asynchronous serial port at TTL levels, the data format is
4800baud, 8 data bits, odd parity, 1stop bit(4800,8-Odd-1). We use
5-byte-format to send data packets, from oximeter module to host computer.
We send 60 packets every second, and unpack the packets according to sync
bit(bit7). Define as following:

Byte Bit Description Byte Bit Description

0 Signal strength 0 0 Pulse rate 0

1 Signal strength 1 1 Pulse rate 1

2 Signal strength 2 2 Pulse rate 2

3 Signal strength 3 3 Pulse rate 3


1 4
4 1 = Searching too long 4 Pulse rate 4

5 1 = Sensor unplugged 5 Pulse rate 5

6 1 = Pulse beep 6 Pulse rate 6

7 1 (Syn. bit) 7 0 (Syn. bit)

0 Plethysmogram 0 0 SpO2 0

1 Plethysmogram 1 1 SpO2 1

2 Plethysmogram 2 2 SpO2 2

2 3 Plethysmogram 3 5 3 SpO2 3

4 Plethysmogram 4 4 SpO2 4

5 Plethysmogram 5 5 SpO2 5

6 Plethysmogram 6 6 SpO2 6

SpO2 Module 8
7 0 (Syn. bit) 7 0 (Syn. bit)

0 Bargraph 0

1 Bargraph 1

2 Bargraph 2

3 Bargraph 3
3
4 1 = Sensor error

5 1 = Searching for pulse

6 Pulse rate 7

7 0 (Syn. bit)

SpO2 Module 9
Chapter 6. Connections

6.1. Connections

As the picture shows, there are J3 and J5 connectors, define as following:

6.2. Probe connector J5


P1 Pin Description Probe Connections
1 PRB_DET
2 SHIELD
3 RED- According to the
4 RED+ sample connection
5 SHIELD
6 PD-
7 PD+

6.3. Host Connector J3


P3 Pin Description
1 TX Output
2 RX INPUT
3 GND
4 +5V

J1 and J2 Connectors on the back, compatible with Nellcor Pin

SpO2 Module 10
6.4. Probe connector J1
Pin Signal description
1 NC
2 Reset input
3 TX Transmit data output
4 RX Receive data input
5 Ground
6 NC
7 NC
8 +5.0V Power input
9 NC
10 Ground

6.5. Host Connector J2


Pin Description
1 PD-
2 PD+
3 SHIELD (Ground)
4 SHIELD (Ground)
5 NC
6 NC
7 RED-
8 RED+

6.6. Probe
This Module connected to the oximeter sensor with a analogous PD. and the LED wave
lenth is 660nm Red light or 905nm infrared light.

SpO2 Module 11
Chapter 7. Mechanical

7.1. Dimension Size


Length Unit:mm

SpO2 Module 12