ADR-1000A Maintenance Manual


REV 2.1

TABLE OF CONTENTS

3. BOARDS AND SIGNALS DESCRIPTION ....................................................................3-1

3.1 DEEB........................................................................................................................................3-1
3.2 RTEE ........................................................................................................................................3-2
3.3 RC.............................................................................................................................................3-3
3.3.1 General ................................................................................................................................3-3
3.3.2 NPRCB.................................................................................................................................3-3
3.3.3 RC-IO...................................................................................................................................3-3
3.4 BFSB ........................................................................................................................................3-6
3.5 INTERFACE SIGNALS ..................................................................................................................3-7
3.5.1 I/O signals properties ..........................................................................................................3-7
3.5.2 I/O types...............................................................................................................................3-9
3.5.3 Signals Description - Inputs ..............................................................................................3-11
3.5.4 Signals Descriptions - Outputs ..........................................................................................3-13
3.6 MODES OF OPERATION............................................................................................................3-14
3.6.1 General ..............................................................................................................................3-14
3.6.2 Stand By.............................................................................................................................3-14
3.6.3 Fluoro ................................................................................................................................3-14
3.6.4 Photospot (DR, DA, DSA, BOLUS DA, BOLUS DSA) ......................................................3-15
3.6.5 Film Radiography (Spot) ...................................................................................................3-15
3.6.6 Backup Fluoro (Emergency Bypass)..................................................................................3-16
3.6.7 Data flowcharts.................................................................................................................. 3-16
CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3. BOARDS AND SIGNALS DESCRIPTION

3.1 DEEB
DEEB (Display and Edge Enhancement board) is a PCI board located in the MPU. The
board receives the pixels data from PRCB via the video bus and performs ‘realtime’ image
processing, display images on monitor in 1kx1k resolution, and send the images to the host
via PCI bus.
An optional RTEE daughter board that can be attached to the DEEB enables performing
real-time edge-enhancement operation (see next section).
• Switch module is responsible for video streams routing inside the DEEB, function
minify and zoom of the images
• Local bus Arbiter module is responsible for PLX9054 chip control, histogram function
and control of the service memory
• PCI frame buffer allows transferring images between DEEB and the host memory.
This module also performs collimating function
• Display module is responsible for display video images with 1kx1k resolution and text
data with 1280x1024 resolution
• RTEE board is responsible for calculating convolution with 11x11 kernel for real-time
Edge Enhancement function and display images to a reference second monitor

P R C B V id e o B u s P R C B B iD ir V id e o
Control

Control

Control
ROI
Data 16b

Data 16b

C o n tro l S ig n a ls
D a ta in 1 6 b
C o n tro lS ig n a ls O p tio n a l
B o a rd
C o n tro lS ig n a ls
D isp la y V id e o D a ta 1 6 b S W IT C H & Z O O M & M IN IF Y (R E F D isp la y ,
EDGE
M o d u le D a ta o u t 1 6 b
E n h a n c em e n t)
C a m C lo c k
S y s C lo c k
Control

Control
Data 16b

Data 16b
Data 16b

Data 16b

Data 16b

P C I B u ffe r
RamDacBus 10b

H is to g ra m
M o d u le
Local Bus

D a ta 1 0 b M AU M O D U LE
Local Bus

D o u b le
R O I 1b
B u ffe r(3 2 M x 2 )
Local Bus

Local Bus

LO CAL BUS
Local Bus

Local Bus

C o n tro l
DEEB
Local B us C o n tro l PLX 9054
S e rv ic e D a ta 1 0 b A rb itr & B o a rd
M e m o ry
H isto g ra m
CONTROL

DATA 32b

P C I IN T E R F A C E 3 2 b it

Figure 3- 1: DEEB Block Diagram

Boards And Signals Description 3-1


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.2 RTEE
RTEE board is a daughter board connected to the DEEB via 120-pin connector. The board
can perform realtime image Edge Enhancement as well as send images to the reference
monitor.
The RTEE is comprised of two main modules:
• Main Controller module - responsible for video streams routing inside RTEE and
outside interface, performs convolution and minify functions.
• Display module - receives 1024x1024 pixels images and 1280x1024 overlay
graphics pixels and converts the data to an analog composite video to be displayed
on the reference monitor.

LUT 18x16 LUT 18x16

Address -17b data-16b Address -17b data-16b

Optional module
A
ControlSignals Main Controller module (Cyclone-C12)
Display
VideoData
Module Kernel Calculation
Minify functions
Switch functions
(Stratix - 1S10) Optional module
B
(Virtex II - 250)
Sys Clock
Local Bus

Data 16b from Switch

Data 16b to Switch

Cofiguration
Sys Clock

Local Bus
Control

RTEE
Board
DEEB CONNECTOR

Figure 3- 2: RTEE Block Diagram

Boards And Signals Description 3-2


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.3 RC

3.3.1 General
The Radiography Control Unit (RC) is divided in two printed circuit boards: the NPRCB
and the RC-I/O. These boards communicate through a dual fiber optic cable. During
acquisition image frames data in the camera are dumped to the NPRCB and sent forward
to the DEEB while room control and x-ray related signals are interfaced through the
RC-IO to the NPRCB and then to the software.

3.3.2 NPRCB
The NPRCB board is located in a PCI slot of the MPU. Images composed of 1024x1024
pixels/frame by 12bits/pixel are dumped from the camera to the NPRCB at a rate up to
30f/sec. The incoming pixels are analyzed by a real-time histogram and then corrected for
equalization, blemishes, circular blank and other proprietary functions. The corrected
images are then sent to the DEEB via two dedicated video bus.
Main features of the NPRCB are:
• Sustained high-rate acquisition of 1024 x 1024 x 8-bit @ 30 fps or 1024 x 1024 x
10bits @ 15 fps, for 10 seconds (with High-Rate option)
• 10-bit real time histogram calculation
• Offset correction
• Gain correction
• Blemish correction
• Automatic cine review directly from on-board memory

3.3.3 RC-IO
The RC-I/O board is located in the Power Distribution Unit (Rack). The RC_IO interfaces
to the X-ray room host system (x-ray generator, table, I.I., Injector, etc.).
The RC-I/O includes a daughter board named IOB that controls the camera IRIS as well
as interfaces the host system analog signals.
Additional daughter board named BFSE (optional) is connected to the opposite side of the
RC_IO. This board relays the power of the camera and is used for automatic camera reset
in case of the MPU failure (Emergency Fluoro).

The following block diagram describes the main modules and interconnections of the
NPRCB, RC-IO and IOB.

Boards And Signals Description 3-3


ADR-1000A Maintenance Manual

3-4
Camera Camera PDB
control PDB
Data Mon lock
RC I/O
AT&T
UART ABC/ABS
12 to10 Interface
LUT
PCI
Acquisition CDI
Interface

Figure 3- 3: RC Block diagram


interrupt Unit
B
U Address R
F
F
O
Fiber Optic Connection
E O
Integrate I/O's
R
- M
S Controls Main
P Interrupt
SCLK SV
Buffers,
control Opto's
C UART
module C
I &
Connectors
- O
B L.Lock N
U Timing Unit T
S R
CMT Medical Technologies Ltd.

Boards And Signals Description


O
SCLOCK (33Mhz)
L
B
U
F
Data
F
E
R I/O Interrupts
S IRIS
CONTROL
REV 2.1

IOB and Analog


RC BASE Signals interface
ADR-1000A Maintenance Manual

3-5
REF
Analog input 6 Analog 8 8
ADC
from room filters
Iris
position
Analog
Multiplexer
Input
IRIS
select
Bypass

Figure 3- 4: IOB Block Diagram


select
3
8 8
IRIS Opto isolator
RC_IO isolated 8
In position
outputs
RC_IO
nonisolated
8
input
REF Motor
Current test Opto isolator
Fault
Over Current
Iris DAC +
CMT Medical Technologies Ltd.

Control IRIS DRIVE

Boards And Signals Description


IRIS
position
Potentiometer -
RC_IO
Motor Power Power IRIS
nonisolated Opto isolator
enable switch Motor
output
REV 2.1
CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.4 BFSB
The Backup Fluoroscopy Switch Board connects as a daughter board to the RC_IO. A 15V
power supply connects as power-in to the BFSB and then continues through an
optocoupled relay as power-out to the Camera CCU. In case of emergency, like a serious
MPU failure, the RC_IO main control module will be notified via the watchdog unit. As a
result, the relay on the BFSB will be turned off and on again, resulting in Camera
automatic reset. This power reset will enter the Camera into internal independent
continuous interlaced operation (master mode) enabling visualization of continuous
Fluoroscopy mode (this is provided also by controlling the Fluoroscopy request signal in
emergency mode and the video switcher auto switching the CCU video signal directly to
the In Room Monitor). In case of need it is possible to bypass the BFSB board by
disconnecting the power cables and connecting them directly to each other (shunt
connection).

Shunt Connection

Shunt SW

15 V
Power
BSFB Camera
Supply

RCIO

Figure 3- 5: BFSB interconnection

Figure 3- 6: ADR 1000A RC-IO Connection Panel

Boards And Signals Description 3-6


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.5 Interface Signals

3.5.1 I/O signals properties


Input: A signal incoming to ADR.
The input has 4 parameters to be defined:
Wet/Dry defines weather the input signal is power driven (Wet/Dry = 0) or
contact-close input (Wet/Dry = 1), (see related figure).
Polarity defines the input active polarity (depends on Wet/Dry polarity).

HOST V iso HOST RC I/O


RC I/O

Input signal, "wet/dry" = 1 Input signal, "wet/dry" = 0


If polarity = 1, then "active" when If polarity =0, then "active" when
switch is closed switch is closed

Figure 3- 7: Input signals isolation configuration

Half Debounce/Debounce: both signals define the type of debounce applied to


the relevant input signal. Full debounce: upon assertion and negation. Half
debounce: debounce upon assertion only.

Debounce Half debounce Action


0 0 No debounce
0 1 No debounce
1 0 Full debounce
1 1 Half debounce
Table 3- 1: Debounec properties

Interrupt: defines whether ADR responds to the specific input:


Responds interrupt = 1
Neglects interrupt = 0

Boards And Signals Description 3-7


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

Output:
A signal outgoing from ADR
Output can be defined using up to 3 parameters (depending on type - refer to
next section).
Wet/Dry defines whether output from ADR is power driven Wet/Dry = 1
Or contact close Wet/Dry = 0.
SSR type, always dry contact style (unaffected by Wet/Dry - refer to next
section).
Polarity defines whether the output signal is inverted on its way out or not:
Not inverted: Polarity =1
Inverted: Polarity =0

RC V iso
RC
I/O I/O

Output signal, "wet/dry" = 0 Output signal, "wet/dry" = 1

Figure 3- 8: Output signals isolation configuration

Outputs 0..3 are special outputs that differ from all other outputs. They are not
effected by polarity, active output is always 1.
Mux value defines one output of 4 possibilities to be driven to outputs 0..3

MUX Value Actual Output


0 The output itself
1 Pulse fluoroscopy
2 EXPACK
3 CINE
Table 3- 2: Outputs 1..3 multiplexing options

Boards And Signals Description 3-8


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.5.2 I/O types

3.5.2.1 FET Based solid state relay output (OCF):


• Description:

4.7Ω

Figure 3- 9: "OCF" Output

• Specifications:
Maximal voltage withstand......................... 30V.
“ON” resistance .......................................... ~25 Ω.
“OFF” resistance ....................................... 1010Ω.
Isolation to RC_I/O “iso” power.................3500V.
Current limitation ..................................... 200 mA.
Mispolarization withstand ......................... Yes 30V cont.
Max. operating frequency............................0.5 kHz.

3.5.2.2 High power low resist. Solid State Relay Output (SSR):
• Description:

4.7Ω

Figure 3- 10: "SSR" Output

• Specifications:
Maximal voltage withstand........................ 60V.
“ON” resistance ....................................... ~0.7 Ω.
“OFF” resistance ....................................... >1MΩ.
Isolation to RC_I/O “iso” power.................2500V.
Current limitation ..................................... 750mA.
Max. operating frequency.......................... ~500 Hz.

Boards And Signals Description 3-9


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.5.2.3 Transistor based input (OCI):


• Description:

Figure 3- 11: - “OCI” Input

• Specifications:
Breakdown voltage .................................. 60V.
Isolation to RC_I/O “iso” power...............10MΩ.
Input voltage range................................... 0-30 V
Input “1”................................................... 0-0.4 V
Input “0”................................................... 5-30V
Operating current ..................................... 1-3 mA.
Max. operating frequency......................... 2 kHz.

3.5.2.4 Analog Input:


• Specification:
Breakdown voltage ............................................. 40V
Input resistance ...................................................<3KΩ
Input voltage dynamic range.............................. -10..+10 V

Boards And Signals Description 3-10


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.5.3 Signals Description - Inputs


Name Meaning Description
FLREQ Fluoro Request Generator response to Fluoro foot switch, active when foot switch
is pressed while the rest of generator conditions are fulfilled.
PF_MODE Fluoro Mode Selects continuous Fluoro (0) or Pulsed Fluoro (1).
PF_RATE Pulse Fluoro Rate When Pulse Fluoro selected, determines the pulse trigger rate, see
Table 3- 4 below.
PREPREQ Prepare request Generator responses to first radiography hand switch, active when
first hand switch is pressed while the rest of generator conditions
are fulfilled.
EXPREQ Exposure request Generator response to second radiography hand switch, active
when second hand switch is pressed and the rest of generator
conditions are fulfilled.
EXON (RX) Exposure on Generator indicates actual X_RAY radiation.
LIH Last Image Hold Signal informing whether to show LIH or not in Fluoro mode and
(FL LIH) Pulsed Fluoro mode.
SPOT VIEW Signal used in FR mode used to determine FSS or SVS. (see
(SP LIH) paragraph 3.5)
READY Signal indicating generator is ready in Digital Radiography (DR)
(GENRDY) and Film Radiography (FR) modes.
MOOP (Mode Input analog signal indicating system’s mode of operation.
Of Operation) (Alternative to MAG signals).
Indicator The mode of operation is defined according to the “station” and
I.I magnification as selected by the operator at the Console, and
by the depressed foot/hand switch. Table 3- 5 below shows the
dependency of the MOOP to the signal voltage.
VHRef X-ray Collimator These signals indicate Shutters position @ specific SID.
Vref ANALOG signal to IOB.
Href
V-rev, H-rev CCD Camera Inputs to Controls camera Vertical and Horizontal reverse (see
Reverse Scan Table 3- 6 below).
F_CODE -N.A. Format mode Encoded film format to ADR-1000A.
DR mode Selects between DR or Film mode.
SYS STANDBY Defines X-Ray generator in Standby mode.
DOSE signals Signals encode dose level from generator in Fluoro and Pulsed
Fluoro modes.

Table 3- 3: Signals description – Inputs

PF-Rate 1 PF-Rate 0 Rate


0 0 15 fr/sec
0 1 30 fr/sec
1 0 7.5 fr/sec
Table 3- 4: Pulsed Fluoro Rates
Note: When using the serial communication to the generator the PF rate is determined via
the serial communication, allowing additional rates.

Boards And Signals Description 3-11


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

Operating Mode I.I Magnification Signal voltage


(volts ± 0.1V)
N 2
Fluoro M1 2.5
M2 3
M3 3.5
Digital N 4
Radiography M1 4.5
(DR) M2 5
M3 5.5
Film N 6
Radiography M1 6.5
(FR) M2 7
M3 7.5
N 8
Spot Filming M1 8.5
M2 9
M3 9.5
Table 3- 5: I.I. magnification I/F signal

TV Scan Mode VREV HREV


0 0
0 1
1 0
1 1
Table 3- 6: TV Scan Mode

Boards And Signals Description 3-12


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.5.4 Signals Descriptions - Outputs

Name Meaning Description


FLGNT Fluoro Grant ADR response, to FLREQ, indicating all conditions are
met. Directly controls Fluoroscopy Radiation.
PF Trigger Pulse Fluoro Signal that triggers the generator to “perform a Fluoro
Pulse Trigger Pulse”.
This signal is present only when Pulse Fluoro mode is
selected.
PREPACK Prepare System’s response to PREPREQ while all system’s
Acknowledge conditions are met
EXPACK Exposure Signal that triggers the generator to “Perform an X_RAY
Acknowledge pulse”. Usually implemented as an envelope to the actual
X_RAY pulse. This signal indicates that all system’s
conditions are met for Radio Pulse operation.
INJ Injector Output signal used to activate injector at specific time
EXPOSE Expose LI Output signal from system to Laser Imager keypad to
activate image transfer from ADR to Laser Imager.
F-POS Fluoro Select mirror position for Fluoroscopy.
Position
R-POS Radio Select mirror position for Radiography.
position
VCR_REC VCR Record Activate VCR Recording. In stand-by the VCR should be
ON with REC and PLAY activated and PAUSED.
When one of the Foot/Hand SW is depressed, the system
will generate a pulse or change the state of this signal
(SSPM configurable) to Play-Record the VCR.
ADR Active ADR Ready MPU is on and the application has loaded correctly.
Table 3- 7: Signals description - output

Boards And Signals Description 3-13


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.6 Modes Of Operation

3.6.1 General
The ADR-1000A supports 5 modes of operation :
1. STAND BY
2. FLUORO
i) FL (Continuous Fluoro)
ii) PFL (Pulsed Fluoro)
3. PHOTOSPOT
i) DR (Digital Radiography)
ii) DA (Digital Angiography)
iii) DSA (Digital Subtracted Angiography)
iv) High Rate acquisition, (15 or 30 PPS).
v) BOLUS DA
vi) BOLUS DSA
4. SPOT (Film Radiography)
i) (FSS) Film Spot Shot
ii) (SVS) Spot View Shot
5. BACKUP FLUORO (EMERGENCY BYPASS).
3.6.2 Stand By
Default mode. This mode applies when none of the foot/hand switches are
activated.
3.6.3 Fluoro
3.6.3.1 Continuous Fluoroscopy (FL)
The ADR-1000A enters this mode when the Fluoro foot switch is asserted PF
mode = 0 and ends this mode when Fluoro foot switch is released.
Fluoro foot switch assertion generates FLREQ “active” to ADR-1000A.
RC_IO then sends interrupt FLREQ to software.
When all conditions are met, ADR system returns acknowledge FLGNT to
host system and start to integrate from camera.
Fluoro foot switch released, generates FLREQ “inactive” to ADR-1000A.
RC_IO generates stop interrupt upon Debounced FLREQ inactive.
Software, upon stop interrupt requests, disables FLGRT and INTEGRATE
outputs.
If the Radiography Hand Switch is depressed during Fluoro foot switch,
Fluoroscopy is aborted and the system enters the selected radio mode.
3.6.3.2 Pulse Fluoro (PFL)
Before entering this mode the user should select Pulsed Fluoro (PF) button and
the desired PF Rate.
The ADR 1000A enters this mode when the Fluoro foot switch is asserted and
ends this mode when Fluoro foot switch is released.

Boards And Signals Description 3-14


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

In addition to the continuous Fluoro handshake signals, the ADR 1000A


outputs a PF Trigger Pulse that triggers the generator at the requested rate
status. The system will respond to changes in the PF rate status at all times.
3.6.4 Photospot (DR, DA, DSA, BOLUS DA, BOLUS DSA)
The ADR-1000A System enters this mode when PREPREQ is activated
(usually when first step of double step Hand switch is asserted) and ends when
PREPREQ is negated.
RC_IO generates PREP interrupt upon PREPREQ.
SW responds with PREPACK (if valid) sets mirror position (F_POS &
R_POS- if valid) and camera’s gain & IRIS position.
EXPREQ activation (usually second step of Hand switch assertion) generates
EXPREQ interrupt from RC_IO to software. Software responds with
EXPACK triggering the generator for an X-ray pulse. Upon receiving the
generator feedback EXON, indicating the actual X-ray pulse, ADR system
will initiate light integration to the Camera CCD till negation of EXON. Upon
negation of EXON, Camera transfers the frame from CCD and sends the data
to the NEW PRCB, then it is sent to DEEB and then to the display monitors
and to the disk. If series acquisition was selected additional EXPACK will be
sent to trigger the generator for pulses at the rate that was selected by the user.
Inactivation of EXPREQ (first step Hand switch release), holds the system in
Photospot mode. If then PREPREQ is deactivated, the system will exit
Photospot mode.

3.6.5 Film Radiography (Spot)


3.6.5.1 Film Spot Shot (FSS)
The system enters into this mode when FR mode is selected, Spot-view signal
is not active and Hand switch is depressed (first step of hand switch).
System ends this mode when Hand and switch is released.
Signal flow:

• Hand switch depression activates READY signal (before PREPREQ


signal) which is detected by ADR-1000A.
• ADR-1000A switches to FSS mode when it detects READY active and
PREPREQ, EXPREQ and SPOT VIEW not active.
• ADR-1000A sets FSS mode - Iris & gain.
• During EXON, ADR-1000A Camera integrates light from I.I.
• ADR-1000A displays a flash of light.
• Upon HSW release READY signal is negated (as well as PREPREQ and
EXPREQ).
• ADR-1000A exits Film Spot Shot mode.

Boards And Signals Description 3-15


CMT Medical Technologies Ltd. ADR-1000A Maintenance Manual
REV 2.1

3.6.5.2 Spot View Shot (SVS)


System enters this mode when FR mode is selected, Spot-view signal is active
and Hand switch is depressed (first step of hand switch).
System ends this mode when Hand and switch is released.
Signal flow:
• Hand switch depression activates READY signal (before PREPREQ
signal) which is detected by ADR-1000A.
• ADR-1000A switches to SVS mode when it detects READY and SPOT
VIEW active, and PREPREQ and EXPREQ not active.
• ADR-1000A sets SVS mode - Iris & gain.
• During EXON, ADR-1000A Camera integrates light from I.I. till EXON is
off.
• ADR-1000A acquires and displays the image on the monitors (without
automatic saving of the image to the disk).
• Hand switch release generates READY not active to ADR-1000A.
• ADR-1000A exits Film Spot Shot mode.

3.6.6 Backup Fluoro (Emergency Bypass)


After less than 200ms, from powering up the ADR-1000A system, RC_IO and
the Camera CCU loads from their NOVRAM to their main programmable
chips, the Bypass configuration and the system enters Backup Fluoro mode.
This configuration includes setting the Camera to continuous interlace mode,
setting the Iris opening to a default position and bypassing the “FLREQ”
signal to the “FLGNT” signal. This procedure enables the operation of the
system under Fluoroscopy radiation without using the MPU (PC) unit. When
the MPU logs-on the application software, it will override the Camera and
RC_IO Bypass setting by loading them their normal operating programs and
will enter to normal operation mode. In case of emergency (MPU lockup
during procedure), the operator is required to make the following procedure:
1. Shut down the system main power.
2. Switch off the MPU (PC power switch).
3. Power up the system (without the MPU).

Two dipswitches on the IOB board define 1 of 4 available openings of the Iris.
Two dipswitches (from an array of 8) that are located on the RC_IO, define the
Mirror position and the “ADR_ACTIVE” signal for bypass (refer to
Installation chapter for dip-switches settings for BACKUP FLUORO).

3.6.7 Data flowcharts


The ADR-1000A is a multitasking system.
This means that several operations can be performed simultaneously.
In order to get a better view of the system internal operations, different types of
operations are separately described in the following data flowcharts:

Boards And Signals Description 3-16


General configuration
IN ROOM MONITOR REFERENCE MONITOR
REV 2.1

Optical comunication to
RC_I/O SYSTEM IMAGE 2'nd IMAGE
DVD/CD DISK DISKS DISKS
(PDU - Interface Rack)

TV CAMERA
ROOM IMAGE Edge + Ref. IDE RAID
CONTROL

Boards And Signals Description


PROCESSING Display
CMT Medical Technologies Ltd.

Controller Controller
(NPRCB) (DEEB) (RTEE)

VIDEO BUS SB

PCI BUS

AGP BUS
ETHERNET NETWORK
DRAM PCI
CPU GRAPHIC INTERFACE
MEMORY CHIPSET
CONTROLLER (OPTION)
(SVGA)

DICOM
INPUT DEVICE
INTERFACE

Figure 3- 12: General Data processing block diagram


W/S MONITOR

MOUSE
KEYBOARD
KEYPAD Legend:
High Rate systems
A. Angio systems

3-17
ADR-1000A Maintenance Manual
Fluoroscopy Acquisition
IN ROOM MONITOR REFERENCE MONITOR
REV 2.1

Optical comunication to
RC_I/O SYSTEM IMAGE 2'nd IMAGE
DVD/CD DISK DISKS DISKS
(PDU - Interface Rack)

TV CAMERA
ROOM IMAGE Edge + Ref. IDE RAID
CONTROL

Boards And Signals Description


PROCESSING Display
CMT Medical Technologies Ltd.

Controller Controller
(NPRCB) (DEEB) (RTEE)

SB
VIDEO BUS

PCI BUS

AGP BUS
ETHERNET NETWORK
DRAM PCI
CPU GRAPHIC INTERFACE
MEMORY CHIPSET
CONTROLLER (OPTION)
(SVGA)

DICOM
INPUT DEVICE
INTERFACE

Figure 3- 13: Dataflow during Fluoroscopy acquisition


W/S MONITOR

MOUSE
KEYBOARD
KEYPAD Legend:
High Rate systems
A. Angio systems

3-18
ADR-1000A Maintenance Manual
Radio (Photospot) Acquisition
IN ROOM MONITOR REFERENCE MONITOR
REV 2.1

Optical comunication to
RC_I/O SYSTEM IMAGE 2'nd IMAGE
DVD/CD DISK DISKS DISKS
(PDU - Interface Rack)

TV CAMERA
ROOM IMAGE Edge + Ref. IDE RAID
CONTROL

Boards And Signals Description


PROCESSING Display
CMT Medical Technologies Ltd.

Controller Controller
(NPRCB) (DEEB) (RTEE)

SB
VIDEO BUS

PCI BUS

AGP BUS
ETHERNET NETWORK
DRAM PCI
CPU GRAPHIC INTERFACE
MEMORY CHIPSET
CONTROLLER (OPTION)
(SVGA)

DICOM
INPUT DEVICE
INTERFACE

Figure 3- 14: Dataflow during Radio acquisition


W/S MONITOR

MOUSE
KEYBOARD
KEYPAD Legend:
High Rate systems
A. Angio systems

3-19
ADR-1000A Maintenance Manual