UC 71
Thermal Analysis Application
Application published in METTLER TOLEDO Thermal Analysis UserCom 7
DSC cell
The FRS5 must be well ‘burned-in’ (e.g.
Thermal Analysis Application
METTLER TOLEDO
TA Application No. UC 71 2
Sample quantity
The size of the measurement signal is
propotional to the sample amount. This
means, in principle, with cp measure-
ments (except ADSC): the larger the sam-
Thermal Analysis Application
METTLER TOLEDO
TA Application No. UC 71 3
1.3 The ‘steady state’ ADSC
In this method the sample is subjected
to a periodic heating and cooling pro-
gram (sawtooth). The heating and cool-
ing rates lie between 2 K/min and
Thermal Analysis Application
METTLER TOLEDO
TA Application No. UC 71 4
is entered as the sample weight), heat capacity is however quite small The longer measurement times (3 to 10
3. Sample (in the crucible with lid with (corresponds to the mass of the lid). It is times), compared with the saphire meth-
similar mass (difference < 20 mg) as for better to use smaller samples with ADSC od, are only justified if the specific heat
the calibration). The evaluation routine in order to keep the signal damping low. is to be measured (also isothermally)
Math/ADSC Cp determines the amplitude For the same reasons high frequencies during chemical reactions or certain
Thermal Analysis Application
using Fourier analysis. In addition to the are not used. Normally the cycle time physical transitions. Please note: the Al
formula shown in figure 4, the signal is not less than 1 minute. The average crucibles are only suitable for certain
damping is compensated mathematically heating rate lies between -5 K/min and chemical reactions. Series of samples
by the phase change of the sample mea- +5 K/min (the maximum heating and with good conductivity gave errors in the
surement. cooling rates of the method must be range -2.5 % and +3 %.
lower than the maximum heating and
Possible errors: cooling rates of the modules).
The error possibilites are similar to those An amplitude of 1 K is normally used.
of the saphire method. The calibration
1) With the new Erel temperature function (in version 5.12) of the DSC sensor FRS5, 2.5 % to 3 % can be attained.
Publishing Note:
This application has been published in
the METTLER TOLEDO Thermal Analysis
UserCom No. 7.
See www.mt.com/ta-usercoms