Anda di halaman 1dari 14

NI~~D .

NIPRO INFUSION PUMP

FP-960
OPERATION MANUAL

NIPRO CORPORATION
INTRODUCTION

This manual is written for the operator(s) of the Nipro FP-960


infusion pump.The procedures recommended in this book have
been developed to provide reliable, safe and efficient operation
of FP-960. The operator should read and thouroughly understand
the contents of this manual before attempting to operate FP-960.
This manual should always be kept at your hand.

CONTENT
1. Precausions in use of pump ·1

2. Components of pump and functions 3

3. Components on operation panel and functions 5

4. Specifications and package ·8

5. Directions for use ·9

6. Alarms and notice 13

7. Monitors and special functions 19

8. Maintenance and storage 20

9. Ni-cd battery 20

10.Procedure for battery replacement and cautions 20

11.Guarantee and service 21

12.Precausions in use of electric medical device


for safe and non-dangerous operation 22
. .
1.PRECAUTIONS IN USE OF PUMP • Don't close radiatorgrid placed at rear side of the pump.
To prevent your or any other personal injury and property • When error is alerted with buzzing, take proper actions
damage, the following precautions are specified for safety immediately to correct the error cause in accordance with the
operation. procedure specified in P.13-P.16.
• During operation, check correct function of the pump by
& WARNING indicates that : frequent monitoring of gradual reduction of solution volume in
If the pump is used incorrectly without vial.
considering the WARNING, death or • Don't operate the high frequency generating devices such as
serious injury may be caused. electric knife, wireless portable telephone or the like at a
& CAUTION indicates that : place near the pump because malfunction of the pump may
be caused by jamming.
If the pump is used incorrectly without
considering the CAUTION, injury or • Don't use the pump for blood transfusion because it is not
property damage may be caused. designed so.
• Don't infuse quick drying or extremely viscous solution.
• Ground the pump. If not grounded, electric shock and/or
&WARNING malfunction of the pump may be caused.
• Perform inspection and function test of the pump before use.
• Use infusion set exclusively specified for the pump.
• Don't reuse infusion set. Store a suitable number of infusion
.If
& CAUTION
the pump is contaminated with sol ution, remove it
sets for allowing ready use in emergency.
• Before opening door of the pump, close tube of infusion set immediately by wiping with gauze or clean cloth.
by roller clamp. .If the pump is contaminated with securely adhered dirt,
• To load infusion set in the pump, set tube of infusion set to remove it by wiping with lukewarm water wetted gauze or
tube clamp in accordance with the procedure specified in P.9. cloth after removing the excess water by strong squeezing.
• Don't wet the pump with solution because the pump is not • Take care not to wet the pump, in particular, power inlet,
water proof designo power SW and fuses because the pump is not water proof
• Don't apply strong force or impact to cam blades. designo
• Don't apply strong force or impact to bubble sensor (emitter • Don't use organic solvente such as thinner to clean the pump
and receiver). Clean the sensor regularly by wiping with casing to avoid damage of it.
lukewarm water wetted gauze or the like. If not cleaned, • Store the pump at an ambient temperature between -10 and
malfunction of the pump may be caused. 40°C and at a relative humidity between 30 and 80% in a
• Don't apply strong force or impact to pressure sensor. Clean place without fear of dew condensation.
the sensor regularly by wiping with lukewarm water wetted • Don't discard used Ni-Cd battery. Deliver it to appropriate
gauze or the like. If not cleaned, malfunction of the pump may organization responsible for recycling it.
be caused.

2
2.COMPONENTS OF PUMP ANO FUNCTIONS

I FRONT VIEW I
Handle I REAR VIEW I Name plate

Radiator grid Buzzer volume


Hinge (don't cover to
Operation panel
allow smooth
(see P.5-P.?)
ventilation from Knobbed bolt
Knobbed bolt pump inside)

WARNING for use

Power switch
Name plate
I.V. pole clamp Fuse (O.8A)

Door Power inlet


Battery cover (AC220Vor AC110V)
(Ni-Cd battery
Rubber leg Is encased.) Serial No.

I RIGHT SIDE VIEW I


Front casing I FULL OPEN VIEW I

Infusion Rear casing Tube holder Tube clamp


tube inlet
Tube clamp lever
Operating (depress tube
Door procedure clamp lever to
open tube clamp)
I~~-t---+-Cam blades

Bubble sensor
I
I Battery cover (Emitter)
1
I
Door clamp I~; __ --'IIt~-WARNING for
tube setting

Pressure sensor
Bubble sensor Harness Infusion
(Receiver) tube outlet
3 4
3.COMPONENTS ON OPERATION PANEL ANO (J)AC POWER

FUNCTIONS The LEO is turned on during connection of the pump with external
power supply.
(?lCHARGE
The LEO is turned on during charge of encased Ni-Cd battery.
About 6 hours are required to full charge the battery When full
charged, the LEO is turned off. The battery is charged as far as the

®®®® pump is connected to external power supply irrespective of ON/OFF of


power SW of the pump.
( llnEAOY monitor
, 11 LEO is turned on when the pump is ready for use after successful
c.ompletion of preparation process. Flashing of the LEQ indicates that
I rror is caused or that preparation process is not yet completed.

® ~ OCCLUSION alarm
1 h LEO is turned on when occlusion in infusion tube is detected.
Alnrrn range is between 68.65 and 1l768kPa (0.7 and 1.2kgUcm') and the
.uarrn limit is directly proportional to infusion rateo
1, I OW BATTERY alarm
111 \ LEO is turned on when voltage of the battery is reduced to preset
1I v 1 and further operation
is disabled under normal condition.
"AIR-IN-LlNE alarm
111 ' LEO is turned on when a bubble of a size to be alerted is detected
()f infusion tube is not set to the pump. Flashing of the LEO indicates
dI l ction of a dangerous bubble a little smaller than the above siz e.
IllOOR OPEN alarm
111 \ LEO is turned on when door is fully or partially opened.
Il KVO (Keep-Vein-Open)

111 LEO is turned on when the pump is in KVO condition after the
lnluslon is complete.
11 HATE
1110 LEO is turned on when infusion rate is displayed on data screen.
IDI) LlMIT
20 lile LEO is turned on when delivery limit is displayed on data screen.
11'1 l, TOTAL
111 LEO is turned on when delivery total is displayed on data screen.
I? 1111
111 LEO is turned on when data displayed on screen is mL unit.
13 II1I/h
J

111 LEO is turned on when data displayed on screen is mLlh unit.

5 6
@OATA SCREEN 4.SPECIFICATIONS ANO PACKAGE
RATE, OELlVERY LlMIT, OELlVERY TOTAL or fast infusion rate during
its setting or the pump operation ; or error message is displayed on +SPECIFICATIONS
data screen by 7 segment LEO. e Product name infusion pump
@SELECT SW e Model FP-960
By each pressing of the SW, cycle of "RATE •• O.LlMIT •• O.TOTAL" i e Power supply AC220V±20V, 50/60Hz
repeated. If O.LlMIT or O.TOTAL is left to be displayed for 10 seconds Ni-Cd battery 9.6V, 700mA/h
after the pressing, it is changed to RATE automatically. (usable for 2-3 hours after charge for about
@RESET SW 6 hours)
The SW is used to clear O.TOTAL. To reset the volume, press the S *AC110V± 1OV, 50/60Hz type FP-960 is also
for 2 seconds or more after displaying O.TOTAL. available upon request.
e Maximum power consumption 25VA
@UPSW
eRATE setting range 1-1000mL/h(1mLlh steps)
The SW is used to increase RATE or O.LlMIT displayed on data scree
el .LlMIT setting range 1-9999m Uh(1 mLlh steps)
by pressing. When the SW is pressed one second or more, th
en,TOTAL display range 1-9999mL/h(1 mUh steps)
displayed rate or volume is increased continuously.
e, ast infusion rate About 1000mLlh
@OOWN SW elllfusion pressure 68.65 - 11768kPa (0.7 -1.2kgf /cm )
The SW is used to decrease RATE or O.LlMIT displayed on dat elllfusion accuracy + 10% with Nipro FPC-96 infusion set.
screen by pressing. When the SW is pressed one second or more, th el\pplicable infusion set FPC-96 series
displayed rate or volume is decreased continuously.
@STOP/MUTE SW el\lnrms and notice Occlusion, low battery, Air-In-Line, door open,
The SW is used to stop the pump operation or to mute buzzing. KVO and drive error
e Monitors
@)STOP LEO
The LEO is turned on when the pump operation is stopped.
e, IJ es
External power connection, charge and ready
0.8A/250VX2
*1.5A/125V X2 for AC110V type FP-960.
@START/FAST SW e 01' ration environment Ambient temperature 10~35'C
The SW is used to start infusion or fast infusion. To start infusion, pres
atmospheric pressure 700~ 1060hPa
the SW momentarily. To start fast infusion, press the SW for 1-
Relative humidity 30~85%(No dew condensation)
seconds or more. By the pressing, fast infusion rate is displayed o ')Ior ge environment Ambient temperature -1O~40'C
data screen.
Atmospheric pressure 700~1 060hPa
@START LEO Relative humidity 30~85%(No dew condensation)
The LEO is turned on when the pump operation is started. I It ctric shock protective class and type Class 1, type CF
Flashing of the LEO indicates that the pump is being operated. Highe e p ration mode • Continuous operation
(Iower) infusion rate is reflected by higher (Iower) flashing frequency. Ollt r dimension (W)86mmX(L)130mmX(H)155mm
(excluding protruded part)
Wüight about 1.5kg

+ ACKAGE
l 'ump · · 1
f'owor cord ····· ·· ·1
Op ration manual ·1
''1)1cifications and design may be changed without giving prior notice.

7 8
To unlock tube clamp, press tube clamp lever marked with ~shown in
5.DIRECTIONS FOR USE Ihe lower right Fig, of P.9.
1.0peration
( Y closing door, the tube clamp is released.
G)Set the pump to I.V. pole or on table.
~Connect the pump to external power supply .
• Connect power cord to power (11) ot infusion rate (1 mUh steps).
inlet ~ placed at rear side of -:. Ouring infusion, infusion rate
the pump and connect plug change is disabled.
placed at opposite side of the After stopping the pump operation,
cord to AC220V * receptacle provid change the rate by pressing
-ed with ground connection. "UP" SW and/or "OOWN" SW
*For AC110Vtype FP-960, connect m rked with ••.
plug to AC110V receptacle .
• After the connection completion,
LEO's of" . r "and"CHARGE"
on operation panel are turned on
if battery is not full charged. '1o ot O.LlMIT (mL), display the volume on data screen by pressing
"',1 LECT" SW followed by pressing "UP" SW and/or "OOWN" SW
Q)Turn on power SW Q) placed at
111ukod with ••.
rear side of the pump.
• f)uring infusion, the volume change is disabled.
At the moment of the turning on,
LEO's on operation panel are turned on from upper left end laterally on "11 et to OmL, KVO mode can not be executed.
by one toward lower end. Check the stepwise turning on. .·LJ ually infusion rate (mLlh) is displayed on data screen. By each
pt ssing of "SELECT" SW, cycle of "RATE" •• "O.LlMIT" ••
®Fill infusion set (FPC-96 series) with solution. it i "D.TOTAL" is repeated. If "O.LlMIT" or "O.TOTAL" is left to be
Remove bubbles from infusion tube completely. If n~t removed, di played for 10 seconds, the display is changed to "RATE"
detected by BUBBLE sensor, disabling the pump function. .uuornatically,
@Set infusion tube firmly as : "tube clamp" •• "tube guide" ••. "ea 1 1111
opon roller clamp of infusion set.
blades" •• "BUBBLE sensor (emitter)". Close door and check turrunq o
'1I'fl s "START/FAST" SW for 1-2 seconds or more to start fast infusion.
of "OOOR OPEN" alarm LEO.
1'11H "STOP 1M UTE" SW to stop the infusion after checking few drops
,,1 ',olution discharge from needle tip and no bubbles in infusion tube.
Ilocause the fast infusion volume is integrated, reset by pressing
"HESET" SW for 2 seconds or more after displaying "O.TOTAL" by
prossing "SELECT" SW.
1il',Ioltllnfusion by pressing "START" SW after venipuncture.
11111ch ck O.LlMIT or O.TOTAL during infusion, press "SELECT" SW to
di play it.

r ,M thod of changing RATE or D.LlMIT


ss "STOP" SW to discontinue the pump operation.
ss "UP" SW and/or "OOWN" SW to change the RATE or O.LlMIT.
ss "START IFAST" SW momentarily to start infusion.

9 10
* During infusion, change of RATE or D.LlMIT is disabled.
.Fast infused volume is added to D.TOTAL.
.To check D.TOTAL during fast infusion, press "SELECT" SW.Relevant
RATE change O.LlMIT change LED is turned on and D.TOTAL is displayed. O.TOTAL
Fast infusion rate display display during fast infusion

Press "SELECT" SW under


3.Display and reset of D.TOTAL pressing "START/FAST" e«.
• D.TOTAL is displayed on data screen by pressing "SELECT" SW.
• D.TOTAL is displayed for 10 seconds followed by changing to flo IKVO
rate display. • Ihe LED comes on when the pump is in KVO condition after the
• To reset memorized D.TOTAL, stop the pump operation by pressin Infusion is complete. (Infusion rate at 1mLlh)
"STOP/MUTE" SW. Display D.TOTAL, by pressing "SELECT" S .I)uring KVO mode, buzzing is caused at 1-2 second intervals.
followed by pressing "RESET" SW for 2 seconds or more.
* During infusion, reset is disabled.
• To return to normal infusion mode (mUmin), press "STOP/MUTE" SW

* When "POWER" SW is turned off, all settings and memorize


D.TOTAL are cleared.
lo discontinue the pump operation for changing D.LI M IT or resetting
I .TOTAL.

.ln1uslon completion.
• Press "STOP/MUTE" SW.
4.Fast infusion • Remove infusion set from the pump.
• During fast infusion, fast infusion rate is displayed on data screen. • Turn off "POWER" SW.
• About 2 seconds after releasing press of "START/FAST" SW, t
display is changed to infusion rate and infusion is continued.

11 12
6.ALARMS ANO NOTICE
1.Alarms

BUZZING CA ACTION NOTE


ALARM DISPLAY PATTERN
e Closed roll I'111fl roller clamp. elf infusion rate is set to 100mLlh or less,
Pi,Pi,Pi ...
OCCLUSION on infusion H, plnce infusion set. OCCLUSION LEO may be flashed. Thls
(intermittent)
e Occluded 11"lold the tube or replace infusion is caused by the response to pressure
tip of infusi I1 fluctuation due to use of cam blades of
eFolded infus 11, IIIH • t supplier to checklrepair. waveform infusion system. When pressure
eOamaged is increased to alarm limit, pump opera
-tion is stopped with the LEO ON and
buzzing.

eEven though LOW BATTERY is alerted,


LOW BATTERY infusion is maintained for 1-15 minutes in
proportion to infusion rateo By connection
of pump with AC supply, the alarm is
released and normal infusion is continued.
eSee P.20 to know method of battery
replacement and cautions.

e No solutio 1,,1' lnluslon or start after refilling e Flashing of AIR-IN-L1NE LEO shows
Pi,Pi,Pi ...
AIR-IN-L1NE (intermittent) e Bubble in •11111 ItII!. small bubble inclusion in infusion tube.
tube. 110''''IVI bubble or replace infusion When bubble of alarm limit size is
e Unloaded ,1 detected, pump operation is stopped
tube or tub 1" 111111 tube or set it to sensor with the LEO ON and buzzing.
-ed from 1111'I Ily. eOuring DOOR OPEN, the LEO is ON
e oa.maged 11,11111 ' t supplier to check/repair. irrespective of presence/absence of infusion
tube because sensor (receiver) is placed
at door inside.

eFully/par I 1,,', uoor completely.


Pi,Pi,Pi. ..
OOOR OPEN opened d ", 11"' ! t supplier to check/repair.
(intermittent)
e Oamaged

eAt the moment that KVO LEO is turned


KVO on, infusion rate is reduced to 1mLlh.

14
13
BUZZING NOTE
ALARM DISPLAY PATTERN CAU ACTION

E--1

1111" 011 POWER SW to find tne


1 11 I ourco. After the source removal,
""11111111 SWagain.

E--2 Pii. .. .Internal elect


(continuous) circuit error.

• Affected by
frequency nois
E--3

. 4 I ms is not displaved .
• Durinq infuslon. E-- a ar
E--4 • Damaged mot uppuer to check/repair. The alarm is displayed only w~en .POW.ER
• Damaged SW is turned on. If alerted dunng mfusl.on
rotation sens without error display, stop pump operatlon
• Internal elect and press the SW to see error codeo
circuit error.

16
15
ALARM DISPLAY BUZZING

E-S [O---..Q)(9 CIWlGE -l~


PATTERN
Pii···
CAUSE 1 ACTION NOTE
e Damag~d Bubble sensor circu pp lier to check I repai r.
~rnrJ~~ SATJ UNE OP~N KV (Coutinuous)
e Disconnect or degradation of
~r= F =
= = Bubble sensor circuit.
111
E-6 lO-------G)[oc"'"'"_)~
~------, '-= Pii···
e Damaged pressure sensor circ ~ lier to checkl repair, eln following cases, Alarm is desplayed.
~Cill[2JG:;]@! (Coutinuous)
@!DL= Sffi¡/§D
e~~ = =
e Disconnect or Transformation €
pressure sensor circuit.
• Using other tubes.
= ~ • Setting condition is not good.
~
E-7 [O-------GlLoc_-l~
Pii··· e Less moter rotation than settin ~ pller to check I repair.
rrIDlllí:J~1ill: (Coutinuous)
8::=
'N BAIT klNE OPEN

e
o=]§D
eDamaged moter,Degradation
IQ§J
~
AlE /
= = vEJ
Qml stuck of drive mechanism
h

=~,
Encoder trouble. 11

E-8
Pii···
e More moter rotation than settín 1- pller to check I repai r.
~rnrJ~ (Coutinuous)
-
OCClU lOW "'IR.IN

,r= E
(Q@J
@Q@
=
e Electronic noise, Encoder troub
e Electronic circuit error.
ER SW to find the noise
lhe source removal, turn

I1

2. Notice

NOTICE BUZZING PATTERN


Remirder of
CAUSE ACTI-=O~N +- N_OT_E I
Pi(single):
infusion start elnfusion is not started. I IART/FAST SW or torn off eTo mute buzzing, press STOP/MUTE Sw.
3 min after READY LEO
START/FAST SW is not accepts W
ON followed by 30 sec lppll r to check I repair,
due to insufficient SW operation
intervals e SW contact error.

17
18
7.MONITORS ANO SPECIAL FUNCTIONS
1.REAOY LEO M INTENANCE ANO STORAGE
-1:0:. READY) Flashing indicates : I~complete. infusion preparatio
11111 pump is contaminated with solution, remove it immediately
Wlpll1(J wllh gauze, cloth or the like.
by

disable d pump operation I I r nr not to wet the pump with any fluid because it is not water
@-READY] Lighting indicates : Successfully completed infusi
I f J "el, Ign, in particular power inlet, POWER SW and fuses.
11'1I1 ( organic solvents such as thinner to clean the pump casing
preparation, ready for infusio 111, 11Is attcked.
start.
1" 111,f ump at an ambient temperature range from -10 to 40'C and
lo READY J Extinguishment indicates : f II\lVI humidity range from 30 to 85% in a place without fear of dew.

Pump operation under norm fl le 11 ulon.


condition
2.CHARGE LEO BATTERY
111 1> utory is limited. Ouring the life, it is possible to repeat
§"CHARGEJ Lighting indicates : ~uring battery charge by conne , e I di charge cycle about 700 times which is equivalent to use for
. tion to external power supply 111J vunr If one cycle of charge/discharge is performed daily.
[O CHARGE] Extinguishment indicates:
1111111
111 1 hour
l3ATTERY LEO is turned off (initially, the LEO is turned off
after charge start) is gradually reduced. The reduced
I /f 111el d by gradually reduced operable time of the pump.
Oisconnection from external pow
supply or full charged battery , 'ClIllITl nded to replace battery with new one when the operable
Note I 11' luc d to about one hour.
Charge can be controlled always irrespectiv 111xl section to know procedure for battery replacement and
of ON/OFF of POWER SW as far as pump i 111111119
the replacement.
connected to external power supply.
3. r ~~ JLEO DURE FOR BATTERY REPLACEMENT
~J Lighting indicates : Pump connected to external powe
AUTIONS
supply '"11 ,1 pow r cord from power inlet of the pump and AC power
Note' The LEO' . 1 I I h,IIClWIel by turning off POWER SW.
. IS turned on Irrespective of ON/OF 111 f '1 he Il marked wtth •• in Fig. of the next page to remove
of POWER SW f .
as ar as pump IS connected t \ Iy! IIV' 1
external power supply.
IVIII el 1) ttery to ¡eplace with new one followed by engaging
un ,1111 wlth lhe new one.
4.Special functions 11\ I1 I IIH lnl tool such as forceps to engage them for avoiding
1 11 ''1 el '"' 19 causad by short-circuit.
~he. followin.g special functions are enabled' I I 'CIVil 1 Y screwing the bolts.
Flne adJustment of OCCLUSION . ti, 1 !lnl r ctory charged. Charge by connecting the pump to AC
sensitivities. or AIR-IN-UNE detecting
W 1 IIpply. About 6 hours are required to full charge.
• Fine adjustment of infusion t ( . WI 11(IIAI! ¡: LEO is turned off, disconnect the pump from AC power
rate +9%) ra e adJustable range: displayed infusion
11I 1'1 ',y \ op rate the pump to check whether the battery is full
• ~~~~~ ~:Vi~~to~~~eT:~ ~TART/FAST
please contact your near~st
cy.
z: SW, STOP/MUTE SW and
the above functions is required,
11 lId
1111II IIH f urnp to AC power supply again to charge battery for the
I 11 1 (1\ lIa pump.
W 1.1111 Iy Ior FP-960 is available from your nearest agency.

19
20
L I
CAUTIONS IN USE OF ELECTRIC MEOICAL
© VICE FOR SAFE ANO NON-OANGEROUS
ALARMOFF !',m ON
PERATION
1;fQ
! ARNING
I 1lile, medical device shall be operated
ffiCAUTION
by authorized expert only.
I 1111,.medical device shall be set to a place of:
(1 )I~II contact with water
)r 11 idverse effects caused by exposure to abnormal atmospheric
1'11 ure, temperature or humidity; direct sunlight; or air containing
"" 1, ! alt and/or sulfur
)NII 111 itlnation, vibration or impact (also during transport)
11.GUARANTEE ANO SERVICE )tlll ,11 micals or no harmful gas generation
)1'1 1 nce of power supply of correct frequency/voltage and
The pump is guaranteed by Nissho for a . 111 ptable current level (or power consumption)
date ot your purchase If the umo : .penod of one year after the
)1 11 IlIqh frequency generating device such as electric knife or
due to our faulty work~anshi ~ d P IS f~,led oro damaged by defects
n WII( It s portable telephone
be repaired by us free of ch:rge p~atenals dunng the period, it shall
If failure or damage is caused b :ase contact your nearest agency. 11" following actions prior to use:
guarantee for one year is invalid Yd ny ?f the following reasons the 111t.k SW contacts, data setting SW's data screen and the like to
1. Misuse or unreasonable alterati~~ rep.alr cost shall be paid by y~u: IlIt correct functions of electric medical device.
person n reparr carned out by unauthorized )1 111ck correct and safe connection of all cords.
2. Drop or roug h transfer /trans t )1 111ck correct combination with other device(s) to eliminate potential
3. Fire, expOsure to salt or P~~r~~~~ the date of your purchase ""'11110 Is miss or danger
gas
abnormal voltage application o t f G , earthquake, storm, flood, )1 III ok Infusion line to be connected to patient to assure its safety.
r ac o od of any nature )1111I k correct ratings of power supply.
)11111'111 frequency generating device such as electric knife or wireless
1'"11(1)1 telephone is present near infusion pump, turn off the device.
11 lIu following actions during use
~1)1111ck whether limit time or volume required to diagnose or treat
l' ,lit nt ls not exceeded.
~ ) 11111110r patient and electric medical device frequently to assure safe
lid normal operation of the device.
~ )11 uouble of patient or electric medical device is detected, take
111111' lo cope with the trouble; for example, stop the device operation
111r.uustderation of the patient safety.
~ \)1 111ck no placement of high frequency generating device such as
I 1It le, knife or wireless portable telephone near infusion pump.

21
22
5.Take the following actions after use:
(1)Return operation SW and the like to original states followed
turning off POWER SW in accordance with prescribed procedure.
(2)Store electric medical device at a place of:
1 No contact with water
TI No adverse effects caused by exposure to abnormal atmospheri
pressure, temperature or humidity; direct sun-light; or air containin
dust, salt and/or sulfur
III No inclination, vibration or impact
NNo chemicals or no harmful gas generation
(3)Store electric medical device after cleaning for the next normal use.
(4)lf electric medical device is damaged, indicate the fault condition
by suitable means. The device shall be repaird by authorized exper
only.
7.Don't alter electric medical device.
8.Maintenance and check
(1)Check and maintain electric medical device and its components
regularly.
(2)lf electric medical device is used after some time of resting, check
correct and safe functions of it prior to use.
9.0ther
(1)Use infusion set exclusively specified for infusion pump.
(2)Take care not to wet infusion pump with solution or the like.
(3)During infusion, monitor reduction of solution level in vial and infusion
pump operation frequently for assuring safety.
(4)When safety guard is operated, take corrective actions immediately
after finding the error cause.

10. Maintenance control


(1)Keep electric medical device clean. If contaminated with splashed
solution or dirt, remove by wiping with alcohol or lukewarm water
watted gauze or cloth.
(2)Store electric medical device under avoiding exposure to direc
sunlight, high ternperature/humidity and excess amount of dust.
(3)Oiling is not required.
(4)lf electric medical device is not used for a long time, check and
maintain regularly.

23

Anda mungkin juga menyukai