Anda di halaman 1dari 8

Gastrointestinal Imaging • Original Research

Hara et al.
CT Iterative Reconstruction Technique

Gastrointestinal Imaging
Original Research

Iterative Reconstruction Technique


for Reducing Body Radiation Dose
at CT: Feasibility Study
Amy K. Hara1 OBJECTIVE. The purpose of this study was to evaluate the image noise, low-contrast
Robert G. Paden resolution, image quality, and spatial resolution of adaptive statistical iterative reconstruction
Alvin C. Silva in low-dose body CT.
Jennifer L. Kujak MATERIALS AND METHODS. Adaptive statistical iterative reconstruction was used
Holly J. Lawder to scan the American College of Radiology phantom at the American College of Radiology
reference value and at one-half that value (12.5 mGy). Test objects in low- and high-contrast
William Pavlicek
and uniformity modules were evaluated. Low-dose CT with adaptive statistical iterative re-
Hara AK, Paden RG, Silva AC, Kujak JL, construction was then tested on 12 patients (seven men, five women; average age, 67.5 years)
Lawder HJ, Pavlicek W who had previously undergone routine-dose CT. Two radiologists blinded to scanning tech-
nique evaluated images of the same patients obtained with routine-dose CT and low-dose CT
with and without adaptive statistical iterative reconstruction. Image noise, low-contrast reso-
lution, image quality, and spatial resolution were graded on a scale of 1 (best) to 4 (worst).
Quantitative noise measurements were made on clinical images.
RESULTS. In the phantom, low- and high-contrast and uniformity assessments showed
no significant difference between routine-dose imaging and low-dose CT with adaptive sta-
tistical iterative reconstruction. In patients, low-dose CT with adaptive statistical iterative re-
construction was associated with CT dose index reductions of 32–65% compared with routine
imaging and had the least noise both quantitatively and qualitatively (p < 0.05). Low-dose CT
with adaptive statistical iterative reconstruction and routine-dose CT had identical results for
low-contrast resolution and nearly identical results for overall image quality (grade 2.1–2.2).
Spatial resolution was better with routine-dose CT (p = 0.004).
CONCLUSION. These preliminary results support body CT dose index reductions of
32–65% when adaptive statistical iterative reconstruction is used. Studies with larger statisti-
cal samples are needed to confirm these findings.

T
he explosive growth of CT can Dose reduction with CT has been limited
be attributed to its wide avail- because the current CT reconstruction algo-
ability, speed, and diagnostic rithm (filtered back projection [FBP]) does
benefits. In a 2007 report [1] it not produce consistently diagnostic images if
was estimated that more than 68.7 million tube current is substantially reduced. Iterative
CT examinations are performed each year in reconstruction is a reconstruction algorithm
Keywords: body CT, low-dose CT, radiation dose
the United States, a dramatic upsurge com- whereby image data are corrected with an as-
DOI:10.2214/AJR.09.2397 pared with the 3 million performed in 1980. sortment of models. Although new to CT, it-
Although it accounts for only 11–13% of ra- erative reconstruction is widely used in PET
Received January 15, 2009; accepted after revision diologic examinations performed overall in and was used when CT was introduced [5]. A
February 24, 2009. the United States, CT is responsible for more current limitation of iterative reconstruction,
1 than two thirds of the total radiation dose as- however, is the long computing time. There-
All authors: Department of Diagnostic Radiology, Mayo
Clinic Arizona, 13400 E Shea Blvd., Scottsdale, AZ 85259. sociated with medical imaging [2, 3]. Public fore, a modified and computationally faster
Address correspondence to A. K. Hara. concern with radiation exposure escalated iterative reconstruction technique, adaptive
when a widely publicized article [4] claimed statistical iterative reconstruction, was devel-
AJR 2009; 193:764–771
that the estimated cancer risk in the United oped in which only one corrective model is
0361–803X/09/1933–764 States attributable to CT radiation has grown used to address image noise. This technique
from 0.4% to 1.5–2.0% owing to the substan- is used to solve one of the primary problems
© American Roentgen Ray Society tial increase in use of CT. of dose reduction for CT with FBP: increased

764 AJR:193, September 2009


CT Iterative Reconstruction Technique

noise with decreased radiation dose. The pur- original FBP and the full adaptive statistical itera- abdomen had been imaged, and in six, only the ab-
pose of this study was to determine the fea- tive reconstruction images can result in a blended domen and pelvis. Ten of the 12 comparisons had
sibility of adaptive statistical iterative recon- image with markedly decreased noise that retains been performed with IV contrast enhancement
struction for low-dose body CT through an a more typical CT appearance. This blended im- (nine venous phase, one arterial phase). In the
evaluation of image noise, low-contrast res- age can be adjusted from 1% to 100% in adaptive other two examinations, CT was unenhanced. For
olution, image quality, and spatial resolution statistical iterative reconstruction. A mathematic routine-dose CT, the peak kilovoltage had been
both in a phantom and in patients. description of adaptive statistical iterative recon- 140 kVp in eight examinations and 120 kVp in four.
struction is shown in Appendix 1. Seven of the CT examinations had been performed
Materials and Methods with dose modulation software with variable tube
Study Design Phantom Study current at a slice thickness of 5 mm in two exami-
All examinations were performed on a 64-MDCT The American College of Radiology (ACR) CT nations, 3.75 mm in five, and 3 mm in five exam-
scanner (CT750 HD, GE Healthcare). This retro- phantom (Gammex 464, Gammex) was scanned inations. In most cases, 64-MDCT scanners had
spective HIPAA-compliant study was approved by twice, once with the ACR reference values (www. been used (three, VCT, GE Healthcare; six, Sen-
the institutional review board of our institution. acr.org/accreditation/computed/ct_reqs.aspx) and sation 64, Siemens Healthcare). The other three
then at one-half this value (12.5 mGy). Helical examinations were performed with a 16-MDCT
Adaptive Statistical Iterative Reconstruction scanning was performed in the manner required scanner (Sensation 16, Siemens Healthcare).
In computation with iterative reconstruction, for submission of images for scanner accreditation.
the image has an initial condition of values, which Low-contrast resolution, high-contrast resolution, Technique of Low-Dose CT With Adaptive
are iteratively optimized according to the rules of and uniformity modules were imaged, and these Statistical Iterative Reconstruction
the model. The FBP image is used for the initial test objects were evaluated by CT physicists not Low-dose CT was performed with the follow-
condition in adaptive statistical iterative recon- blinded to scanning technique. Radiation dose and ing parameters: fixed noise index, 30.9; collima-
struction (the initial value of each pixel) for the noise estimates were made in accordance with ACR tion, 0.625 mm; reconstruction slice thickness,
following reasons: it is presumably close to the fi- protocol. Images were reconstructed with both FBP 3.75 mm; tube potential, 120 kV; variable tube cur-
nal optimized solution (lessening the need for it- and multiple values of adaptive statistical iterative rent determined by x, y, z-axis dose modulation;
erations); it is a valid indicator of specific-slice reconstruction ranging from 10% to 100%. gantry rotation time, 0.5 second. The CT dose was
image noise; and it can be quickly obtained. For reduced through an increase in accepted noise in-
modeling and use of iterative reconstruction, min- Patient Study dex for the study from 22.1 to 30.9. With the dose
imum convergence is achievable with the adaptive The study cohort consisted of 12 patients (seven modulation software of the scanner, the tube cur-
statistical iterative reconstruction model (Fig. 1). men, five women; average age, 67.5 years; range, rent was automatically reduced to match the ac-
A fully converged 100% adaptive statistical itera- 53–86 years) who consecutively underwent low- ceptable noise index. The dose varied with patient
tive reconstruction image, however, tends to have dose CT and who had undergone routine-dose CT size; that is, larger patients needed a higher tube
a noise-free appearance with unusually homoge- of the same region (abdomen or abdomen and pel- current for maintenance of the desired noise index
neous attenuation. Because some noise is inherent vis) within an average of 10.1 months (range, 3 than did thinner patients. All 12 low-dose CT ex-
in CT, use of 100% adaptive statistical iterative re- days–5 years) before low-dose CT. The compari- aminations were reconstructed twice, once with
construction may not be immediately appealing to sons were matched for IV contrast enhancement FBP and once with 40% adaptive statistical itera-
most radiologists. However, a linear mixture of the and imaging phase. In six comparisons, only the tive reconstruction. The 40% level was chosen on

A B C
Fig. 1—Production of adaptive statistical iterative reconstruction image.
A, Filtered back projection image obtained at 120 kVp and 300 mA at 12.5 mGy (half dose).
B, Image from 100% adaptive statistical iterative reconstruction generated through multiple iterations in accordance with rules of noise reduction model.
C, Linear combination of A and B produces blended image (50% adaptive statistical iterative reconstruction), which has less noise than filtered back projection image but
without artifactual smoothing of 100% adaptive statistical iterative reconstruction image.

AJR:193, September 2009 765


Hara et al.

30.00
10% increments, the result was a linear de-
Noise (standard deviation)
Full Dose
Half Dose
crease in noise as measured with SD (Fig. 2).
25.00 For full dose scanning, at 0% adaptive statis-
20.00 tical iterative reconstruction, the SD (noise)
was 20. At 100% adaptive statistical iterative
15.00
reconstruction, the noise was minimum (SD
10.00 4), an approximately 75% reduction of noise
5.00 from the original data. At approximately
50% adaptive statistical iterative reconstruc-
0.00
0 10 20 30 40 50 60 70 80 90 100 tion, the noise was approximately one-half
% Adaptive Statistical Iterative Reconstruction
that of a full-dose FBP image.
When the phantom was scanned at 50%
Fig. 2—Noise reduction in images reconstructed with adaptive statistical iterative reconstruction in phantom. lower dose, the noise as measured with SD
Graph shows linear decrease in image noise (SD) as percentage adaptive statistical iterative reconstruction was 1.4 times greater (28.6 vs 20.4) with 0%
increases. Images acquired with 50% dose reduction (half dose) have 1.4 times SD value (28.57 compared with adaptive statistical iterative reconstruction. At
20.39) without adaptive statistical iterative reconstruction. Reconstructing images with 30% adaptive statistical
iterative reconstruction for half-dose acquisitions produces images with noise nearly equivalent to that of full- 30% adaptive statistical iterative reconstruc-
dose images without adaptive statistical iterative reconstruction (double arrow) (SD 20.52 compared with 20.39). tion, the noise was equivalent to that of a full-
dose FBP image, and further reductions in
the basis of results of the phantom analysis, which Image noise, image quality, low-contrast resolu- noise occurred as percentage adaptive statisti-
indicated that 40% adaptive statistical iterative re- tion, and spatial resolution were graded on a scale cal iterative reconstruction was increased.
construction should produce a diagnostically ac- from 1 (best) to 4 (worst). A score of 1 meant that Comparison of low-contrast images
ceptable image with less noise than a full-dose the image was better than expected at routine-dose showed comparable appearance of the ACR-
FBP image. CT, 2 meant the image was equivalent to that ex- required 6-mm objects at both routine-dose
pected at routine-dose CT, 3 meant the image was CT with FBP and low-dose CT with adaptive
Dose Comparison worse than expected at routine-dose CT, and 4 statistical iterative reconstruction (Fig. 3).
Volume CT dose index (CTDI) and dose–length meant the image was nondiagnostic. The readers Comparison of the high-contrast object
product (DLP) were compared for low-dose (n = independently assessed image noise, image qual- (12 line pairs/cm) showed that adaptive sta-
12) and routine-dose (n = 12) CT examinations. ity, and low-contrast resolution. Readers were in- tistical iterative reconstruction was compa-
For comparison of radiation doses, the patients structed to assess low-contrast resolution by eval- rable with FBP in terms of spatial resolution
were divided into three groups based on body uating the conspicuity of the hepatic veins within and easily exceeded the spatial resolution re-
mass index (BMI) (weight in kilograms divided by the liver or solid organ cysts. Spatial resolution was quirement of 6 line pairs/cm required for ac-
height squared in meters): greater than 25 (n = 3), assessed through consensus evaluation by grading creditation (Fig. 4).
20–24.9 (n = 6), and less than 20 (n = 3). of the sharpness of the hepatic or renal edges. Uniformity was maintained at low-dose CT
with adaptive statistical iterative reconstruc-
Quantitative Analysis Results tion (the maximum deviation between the cen-
Two abdominal imaging fellows not involved Phantom Study tral ROI and peripheral ROIs was less than
in qualitative data analysis made quantitative When adaptive statistical iterative recon- 5 HU) and was within ACR specifications
noise measurements on a total of 36 data sets: 12 struction was applied to the FBP image in [6]. Uniformity measurements on routine-
low-dose CT without adaptive statistical iterative
reconstruction, 12 low-dose CT with adaptive sta-
tistical iterative reconstruction, and 12 routine-
dose comparison CT. Noise measurements were
made by recording the SD in an identically sized
2,500-mm 2 region of interest (ROI) placed 5 mm
outside the anterior abdominal wall at the level of
the umbilicus.

Qualitative Analysis
Qualitative image analysis was performed by
two board-certified and fellowship-trained ab-
dominal radiologists with 8 and 10 years of CT
experience. The 36 data sets were randomized and
deidentified so the readers were unaware of the
postprocessing algorithm and dose. Only the axial
images were displayed on a PACS (Centricity ver- A B
sion 2.1, GE Healthcare). All data sets were dis-
Fig. 3—Low-contrast objects of comparable quality.
played at soft-tissue settings (window, ~ 400 HU; A, Filtered back projection image obtained with 25-mGy routine body image protocol.
level, ~ 40). B, Adaptive statistical iterative reconstruction image obtained at 50% reduced dose (12.5 mGy).

766 AJR:193, September 2009


CT Iterative Reconstruction Technique

dose CT with adaptive statistical iterative re-


construction had the least visually assessed
image noise, and the average of the two read-
ers was significantly better compared with
those for low-dose CT without adaptive sta-
tistical iterative reconstruction (p < 0.0001)
and routine-dose CT (p = 0.01) (Table 3). Av-
eraged scores from both readers in the com-
parison of routine-dose CT and low-dose CT
with adaptive statistical iterative reconstruc-
tion were equivalent or nearly equivalent for
image quality and low-contrast resolution.
Low-dose CT with adaptive statistical iter-
ative reconstruction was significantly better
A B than low-dose CT without it in low-contrast
resolution (p = 0.01) and image quality (p =
Fig. 4—High-contrast objects. 0.0002). The consensus score for spatial res-
A, Filtered back projection image obtained at routine dose of 25 mGy shows 12 line pairs/cm.
B, CT scan obtained with adaptive statistical iterative reconstruction at low dose of 12.5 mGy shows 12 line olution was significantly better for routine-
pairs/cm despite more image degradation than in A. dose CT compared with low-dose CT with-
out (p = 0.002) and with (p = 0.004) adaptive
dose FBP images and low-dose CT images CTDI and DLP of 65% compared with only statistical iterative reconstruction.
with adaptive statistical iterative reconstruc- 29–35% for patients with a BMI was greater Three comparisons had identical scanning
tion were nearly identical (Fig. 5). than 25. parameters (peak kilovoltage, slice thickness,
Quantitative comparisons—Quantitative scanner model), and the results were nearly
Patient Study ROI noise measurements were highest for identical to those in the overall group anal-
Dose comparison—Overall, use of a low low-dose CT without adaptive statistical it- ysis. The only difference was that the aver-
dose reduced the CTDI 47% and the DLP erative reconstruction (average, 35; range, aged image quality grade was slightly higher
44% (Table 1). For low-dose CT, the average 11–52). The noise indexes were nearly iden- for low-dose CT with adaptive statistical it-
CTDI was 12 compared with 22 for routine- tical for low-dose CT with adaptive statis- erative reconstruction (2.0) than for routine-
dose CT. The average DLP was 470 mGy·cm tical iterative reconstruction (average, 31; dose CT (2.2). In the overall analysis, rou-
for low-dose CT compared with 894 mGy·cm range, 7–51) and routine-dose CT (average, tine-dose CT (2.1) had slightly better overall
for routine-dose CT. Not surprisingly, when 32; range, 10–45). image quality (low-dose CT with adaptive
the 12 examinations were compared on the Qualitative comparisons—Low-dose CT statistical iterative reconstruction, 2.2).
basis of BMI (Table 2), the percentage re- without adaptive statistical iterative recon-
ductions in CTDI and DLP increased as BMI struction had the worst scores for visually Discussion
decreased. Therefore, patients with a BMI assessed image noise, image quality, spatial Several approaches have been used in the
less than 20 had percentage reductions in resolution, and low-contrast resolution. Low- effort to minimize radiation dose at CT [7].
These include use of automated tube current
modulation [8–12], noise reduction filters [13,
14], and low-dose protocols for specific clini-
cal indications. With automated tube modula-
tion techniques, CTDI reductions of 40–60%
have been achieved without compromise of
image quality and are now routinely used on
most scanners [7, 12, 15]. Further dose re-
ductions have been achieved mainly by im-
plementation of specific low-dose protocols
for indications such as renal stones and co-
lonic polyps [16–19]. These protocols do not
include IV contrast administration, and us-
ing them can reduce dose more than 50% if
image quality outside the area of interest is
sacrificed. The renal stone and CT colonog-
A B
raphy protocols are not widely implemented
Fig. 5—Uniformity module in phantom. for routine body CT, however, because most
A and B, Filtered back projection image obtained at routine dose of 25 mGy (A) has uniformity nearly identical routine body CT examinations are performed
to that of low-dose (12.5 mGy) CT scan obtained with adaptive statistical iterative reconstruction (B). Central
region of interest (ROI) has attenuation between +3 and –3 HU. Maximum deviation between central ROI and with IV contrast material. In addition, radi-
peripheral ROIs must be less than 5 HU. ologists perform many of these examinations

AJR:193, September 2009 767


Hara et al.

TABLE 1:  Comparison of Low-Dose CT and Routine-Dose CT of Same Patients


Dose–Length Product (mGy⋅cm) CT Dose Index
Body
Patient IV Contrast Mass Percentage Percentage
No. Type of Examination Enhancement Index Routine Dose Low Dose Reduction Routine Dose Low Dose Reduction
1 Abdomen Yes 34 707 441 38 27 17 37
2 Abdomen No 30 1,008 773 23 31 20 37
3 Abdomen, pelvis Yes 28 1,209 886 27 26 18 32
4 Abdomen Yes 25 376 305 19 14 9 35
5 Abdomen Yes 25 848 502 41 26 15 44
6 Abdomen, pelvis No 22 921 549 40 20 11 43
7 Abdomen, pelvis Yes 22 860 548 36 18 11 39
8 Abdomen Yes 20 396 197 50 13 6 52
9 Abdomen, pelvis Yes 20 1,128 451 60 21 8 62
10 Abdomen Yes 19 353 114 68 14 5 65
11 Abdomen, pelvis Yes 19 1,198 442 63 26 9 65
12 Abdomen, pelvis Yes 18 1,073 430 60 22 8 62
Average 24 840 470 44 22 11 48

without a clear idea of the diagnosis and fear tive reconstruction for low-dose body CT of tion and routine-dose CT. Not surprisingly,
missing an important finding owing to poor both a phantom and patients. An ACR phan- when CTDI was reduced without application
image quality. A method of reducing dose at tom was used to validate the use of adaptive of adaptive statistical iterative reconstruc-
routine abdominal CT has been lacking. statistical iterative reconstruction. We found tion, both readers found the images noisier
In this preliminary study, we evaluated an that use of a dose reduced by 50% (12.5 than images from full-dose routine examina-
alternative approach to reducing CT radiation mGy) and adaptive statistical iterative recon- tions, and quantitative noise measurements
dose for routine abdominal CT that has not struction yielded low- and high-contrast res- also were higher. These results support the
been previously available for clinical use, to olution and image uniformity within ACR ability of adaptive statistical iterative re-
our knowledge. This approach, adaptive sta- specifications. Images obtained with the construction to allow substantial reduc-
tistical iterative reconstruction, is a unique technique easily exceeded the spatial reso- tions in radiation dose without the compro-
CT reconstruction algorithm compared with lution requirement for ACR accreditation. mise in image quality due to noise that once
the only one previously available (FBP). Un- We are applying for ACR accreditation using was so troublesome. Furthermore, reader as-
like with FBP, with adaptive statistical iter- this technique. sessments of overall image quality and low-
ative reconstruction, it is not assumed that After the validation, low-dose CT with contrast resolution were nearly identical for
noise is evenly distributed across the entire adaptive statistical iterative reconstruction low-dose CT scans with adaptive statistical
image. Instead, matrix algebra is used to se- was tested in a group of 12 patients who iterative reconstruction and routine-dose im-
lectively identify and then subtract noise from had undergone routine-dose imaging of the ages, further supporting the potential of this
the image with a mathematic model. The re- abdomen or abdomen and pelvis. This pre- technique for routine imaging.
sult is a less noisy image. The ability to selec- liminary evaluation showed use of adaptive Spatial resolution was the only parameter
tively reduce image noise allows generation statistical iterative reconstruction reduces that was worse with adaptive statistical iter-
of a higher-quality image at a lower radiation readers’ perception of image noise in spite ative reconstruction than with routine-dose
dose with adaptive statistical iterative recon- of a decrease in CTDI that was as high as imaging in this pilot study (Fig. 7), although
struction than with FBP techniques. 65% (Fig. 6). Actual measurements of image the actual score (2.5) meant the image had
For this study, we sought to determine the noise were nearly identical for low-dose CT only slightly lower spatial resolution than was
feasibility of using adaptive statistical itera- with adaptive statistical iterative reconstruc- expected for routine-dose CT. In the review
of adaptive statistical iterative reconstruc-
tion images with low scores, the main differ-
TABLE 2:  Subset Comparison According to Body Mass Index ences were slightly decreased sharpness of
Average Dose–Length Product cyst edges and a mildly irregular or jagged
(mGy⋅cm) Average Volume CT Dose Index margin of solid organs. These imperfections
Body
Mass No. of Routine Percentage Routine Percentage did not appear to affect the diagnostic value
Index Patients Dose Low Dose Reduction Dose Low Dose Reduction of the image itself. This factor was evaluat-
> 25 3 975 700 29 28 18 35 ed only on axial images. Coronal and sagit-
20–24.9 6 755 425 41 19 10 46
tal multiplanar reformations were not avail-
able or included in the evaluation. This issue
< 20 3 875 328 64 21 7 64
will be evaluated in a future study involving

768 AJR:193, September 2009


CT Iterative Reconstruction Technique

TABLE 3:  Visual Assessments by Two Readers


Image Noise Low-Contrast Resolution Overall Image Quality Spatial Resolution
Low Dose Low Dose Low Dose Low Dose
Non- Routine Non- Routine Non- Routine Non- Routine
Reader ASIR ASIR Dose ASIR ASIR Dose ASIR ASIR Dose ASIR ASIR Dose
A 2.8 1.6 2.2 2.3 2.2 2.2 2.5 2.3 2.3 NA
B 2.8 1.6 2.3 2.5 2.0 2.1 3.0 2.2 1.9 NA
Averagea 2.8 1.6 2.2 2.4 2.1 2.1 2.8 2.2 2.1 2.6 2.5 1.9
Note—Values are qualitative grading scale: 1, better than routine-dose CT; 2, similar to routine-dose CT; 3, worse than routine-dose CT; 4, nondiagnostic. ASIR = low-dose
CT with adaptive statistical iterative reconstruction, NA = not applicable.
aLow-dose CT with adaptive statistical iterative reconstruction was significantly better than routine-dose CT for image noise (p = 0.01). Low-dose CT with ASIR was

significantly better than low-dose CT without ASIR for image noise, low-contrast resolution, and overall image quality (p < 0.01). Routine-dose CT was significantly better
than low-dose CT with or without ASIR for spatial resolution (p ≤ 0.004).

A B C
Fig. 6—57-year-old woman with body mass index of 18.
A–C, Low-dose CT scan obtained at 120 kVp, 3.75-mm slice thickness, and CT dose index (CTDI) of 8 without adaptive statistical iterative reconstruction (A) has more
image noise in liver than low-dose CT scan with adaptive statistical iterative reconstruction (B) and routine-dose CT scan (140 kVp; 3-mm slice thickness; CTDI, 22) (C). B
and C have nearly identical image quality.

a group of patients with lesions to determine struction capability, we have used these results may become diagnostic, improving detection
whether diagnostic confidence is affected. to investigate ways to reduce our standard- and characterization of lesions.
Future releases of adaptive statistical itera- dose CT protocols, particularly in imaging of This initial evaluation had limitations.
tive reconstruction software also may help to smaller patients. Studies in neurologic, mus- First, because of the retrospective nature of
resolve this issue. culoskeletal, chest, and cardiac CT are ongo- the study, the low-dose and routine-dose CT
The degree of dose reduction was great- ing to determine whether low-dose protocols examinations did not have identical scan-
est for patients with a lower BMI. In patients can be used in these areas. ning parameters. Changes in peak kilovolt-
with a BMI less than 20, the average CTDI Even more aggressive reductions in radi- age and slice thickness can affect noise and
dose reduction was 64% compared with 35% ation dose may be possible in the future. In image quality. Results of prospective studies
for patients with a BMI greater than 25. That effect, scanning may be performed at doses with similar imaging parameters will be help-
adaptive statistical iterative reconstruction low enough to render images nearly nondiag- ful for confirming the initial results. Second,
allows dose reductions for smaller patients nostic but with advanced iterative reconstruc- the small sample size had limited power, and
may help with pediatric imaging, which was tion techniques to return image quality to an prospective studies with larger samples are
not evaluated in this study. The idea of dose acceptable level. Currently, the use of itera- needed. In addition, for the purposes of this
reduction in CT of even larger patients by use tive reconstruction at CT is limited by long study, we chose an adaptive statistical itera-
of adaptive statistical iterative reconstruction reconstruction times. As hardware and soft- tive reconstruction level of 40% because it ap-
is encouraging because the number of obese ware improve, more complex iterative recon- proximated the levels in the phantom study. It
patients in the United States continues to in- struction algorithms may be used clinically, is possible that higher levels of adaptive statis-
crease [20]. resulting in even greater improvements in im- tical iterative reconstruction may improve re-
These preliminary results may help to en- age quality. Iterative reconstruction also may sults. Finally, we did not assess lesions spe-
courage more widespread use of low-dose CT allow routine image reconstruction at thin- cifically. It is possible that adaptive statistical
protocols. In our practice, we have instituted ner slices. Currently, increased noise limits iterative reconstruction may affect lesion con-
use of low-dose CT with adaptive statistical the evaluation of thin reconstructed images spicuity and detection, and this factor has to
iterative reconstruction for all body CT per- (< 2.5 mm) in abdominal imaging. With it- be assessed with future studies.
formed on scanners with this reconstruction erative reconstruction and adaptive statistical We conclude that low-dose body CT with
algorithm available. For our scanners that do iterative reconstruction, whether or not radia- adaptive statistical iterative reconstruction
not have adaptive statistical iterative recon- tion dose is reduced, thinner reconstructions has quantitative and qualitative image noise

AJR:193, September 2009 769


Hara et al.

A B C
Fig. 7—75-year-old man with body mass index of 22.
A–C, Low-dose CT scans without (A) and with (B) adaptive statistical iterative reconstruction (120 kVp; 3.75-mm slice thickness; CT dose index [CTDI], 11) and routine-
dose CT scan (C) (140 kVp; 3-mm slice thickness; CTDI, 20) all show hepatic cysts, but sharpness of cyst edges is best in C. B has least image noise.

and image quality similar to or better than system. Br J Radiol 1973; 46:1016–1022 14. Kalra MK, Maher MM, Sahani DV, et al. Low-
those of routine-dose CT. Compared with 6. McCollough CH, Bruesewitz MR, McNitt-Gray dose CT of the abdomen: evaluation of image im-
those of conventional imaging, with adaptive MF, et al. The phantom portion of the American provement with use of noise reduction filters pilot
statistical iterative reconstruction, CTDI and College of Radiology (ACR) computed tomogra- study. Radiology 2003; 228:251–256
DLP both were reduced an average of nearly phy (CT) accreditation program: practical tips, 15. Kalra MK, Maher MM, Toth TL, Kamath RS,
50% and up to 65% in some patients using artifact examples, and pitfalls to avoid. Med Phys Halpern EF, Saini S. Radiation from “extra” im-
adaptive statistical iterative reconstruction 2004; 31:2423–2442 ages acquired with abdominal and/or pelvic CT:
when compared with routine dose imaging. 7. Valentin J; International Commission on Radia- effect of automatic tube current modulation. Radi-
The low-dose technique with adaptive sta- tion Protection. Managing patient dose in multi- ology 2004; 232:409–414
tistical iterative reconstruction met ACR ac- detector computed tomography (MDCT): ICRP 16. Cohnen M, Vogt C, Beck A, et al. Feasibility of
creditation standards in a phantom analysis. Publication 102. Ann ICRP 2007; 37:1–79 MDCT colonography in ultra-low-dose technique
On the basis of our results, we have imple- 8. Mulkens TH, Bellinck P, Baeyaert M, et al. Use of in the detection of colorectal lesions: comparison
mented this low-dose technique for routine an automatic exposure control mechanism for with high-resolution video colonoscopy. AJR
body CT in our practice. Future studies are dose optimization in multi-detector row CT ex- 2004; 183:1355–1359
needed to confirm these preliminary results aminations: clinical evaluation. Radiology 2005; 17. Niemann T, Kollmann T, Bongartz G. Diagnostic
and determine the effect of adaptive statisti- 237:213–223 performance of low-dose CT for the detection of
cal iterative reconstruction on lesion detec- 9. Smith AB, Dillon WP, Lau BC, et al. Radiation urolithiasis: a meta-analysis. AJR 2008; 191:396–
tion and conspicuity. dose reduction strategy for CT protocols: success- 401
ful implementation in neuroradiology section. 18. Poletti PA, Platon A, Rutschmann OT, Schmidlin
Acknowledgment Radiology 2008; 247:499–506 FR, Iselin CE, Becker CD. Low-dose versus stan-
We thank Qing Wu for statistical help. 10. Graser A, Wintersperger BJ, Suess C, Reiser MF, dard-dose CT protocol in patients with clinically
Becker CR. Dose reduction and image quality in suspected renal colic. AJR 2007; 188:927–933
References MDCT colonography using tube current modula- 19. Luz O, Buchgeister M, Klabunde M, et al. Evalu-
1. 2007 CT market summary report. Des Plaines, tion. AJR 2006; 187:695–701 ation of dose exposure in 64-slice CT colonogra-
IL: IMV Medical Information Division, 2007 11. Lee CH, Goo JM, Ye HJ, et al. Radiation dose phy. Eur Radiol 2007; 17:2616–2621
2. Kalra MK, Maher MM, Toth TL, et al. Strategies modulation techniques in the multidetector CT 20. Ford ES, Mokdad AH. Epidemiology of obesity in
for CT radiation dose optimization. Radiology era: from basics to practice. RadioGraphics 2008; the western hemisphere. J Clin Endocrinol Metab
2004; 230:619–628 28:1451–1459 2008; 93:S1–S8
3. Linton OW, Mettler FA Jr. National conference on 12. McCollough CH, Bruesewitz MR, Kofler JM Jr. 21. Thibault JB, Sauer KD, Bouman CA, Hsieh J. A
dose reduction in CT, with an emphasis on pediat- CT dose reduction and dose management tools: three-dimensional statistical approach to im-
ric patients. AJR 2003; 181:321–329 overview of available options. RadioGraphics proved image quality for multislice helical CT.
4. Brenner DJ, Hall EJ. Computed tomography: an 2006; 26:503–512 Med Phys 2007; 34:4526–4544
increasing source of radiation exposure. N Engl J 13. Kalra MK, Maher MM, Blake MA, et al. Detection 22. Zhang Y, Fessier J, Hsieh J. Fast variance image
Med 2007; 357:2277–2284 and characterization of lesions on low-radiation-dose predictions for quadratically regularized statisti-
5. Hounsfield GN. Computerized transverse axial abdominal CT images postprocessed with noise cal image reconstruction in fan-beam tomogra-
scanning (tomography). Part 1. Description of reduction filters. Radiology 2004; 232: 791–797 phy. IEEE Nuclear Science Symp Conf 2005; 4:4

(Appendix appears on next page)

770 AJR:193, September 2009


CT Iterative Reconstruction Technique

APPENDIX 1: Mathematical Description of Adaptive Statistical Iterative Reconstruction

Mathematically, the adaptive statistical iterative reconstruction model [21] is based on matrix algebra whereby the measured value of each
pixel (y) is transformed to a new estimate of the pixel (y'). This pixel value is compared with the ideal answer predicted with the noise model
(A). If needed, another iteration ensues.
xˆ = argmin{L(Ax,
x
y) + αG(x)}

A final pixel value for the adaptive statistical iterative reconstruction image (x̂) results when repeated y' values ultimately converge. G is the
regularization term that enforces smoothing and edge preservation of the data [22].

F O R YO U R I N F O R M AT I O N
PQI Connect is the latest addition to the ARRS Website and serves as a source for information on meeting the
growing demand for quality review programs in today’s radiology practices and facilities. The interactive and
easy-to-navigate site focuses on five critical topics that guide you through news items, relevant articles, and links
to important information on each topic.

AJR:193, September 2009 771

Anda mungkin juga menyukai