Anda di halaman 1dari 9

Talanta 172 (2017) 221–229

Contents lists available at ScienceDirect

Talanta
journal homepage: www.elsevier.com/locate/talanta

Some practical considerations for linearity assessment of calibration curves MARK


as function of concentration levels according to the fitness-for-purpose
approach
⁎ ⁎
J.M. Juradoa, , A. Alcázara, R. Muñiz-Valenciab, S.G. Ceballos-Magañac, F. Raposod,
a
Departament of Analytical Chemistry, Faculty of Chemistry, University of Seville, c/ Profesor García González 1, 41012 Seville, Spain
b
Facultad de Ciencias Químicas, Universidad de Colima, Carretera Colima-Coquimatlán km 9, 28400 Coquimatlán, Colima, México
c
Facultad de Ciencias, Universidad de Colima, c/ Bernal Díaz del Castillo 340, 28045 Colima, México
d
Instituto de la Grasa, Consejo Superior de Investigaciones Científicas (IG-CSIC), Campus Universitario Pablo de Olavide, Carretera de Utrera km 1, Edificio
46, 41013 Seville, Spain

A R T I C L E I N F O A BS T RAC T

Keywords: Since linear calibration is mostly preferred for analytical determinations, linearity in the calibration range is an
Calibration curves important performance characteristic of any instrumental analytical method. Linearity can be proved by
Fitness-for-purpose approach applying several graphical and numerical approaches. The principal graphical criteria are visual inspection of
Linearity the calibration plot, the residuals plot, and the response factors plot, also called sensitivity or linearity plot. All
Response factor
of them must include confidence limits in order to visualize linearity deviations. In this work, the graphical
Relative error
representation of percent relative errors of back-calculated concentrations against the concentration of the
calibration standards is proposed as linearity criterion. This graph considers a confidence interval based on the
expected recovery related to the concentration level according to AOAC approach. To illustrate it, four
calibration examples covering different analytical techniques and calibration situations have been studied. The
proposed %RE graph was useful in all examples, helping to highlight problems related to non-linear behavior
such as points with high leverage and deviations from linearity at the extremes of the calibration range. By this
way, a numerical decision limit which takes into account the concentration of calibration standards can be easily
included as linearity criterion in the form of %RETh=2·C−0.11. Accordingly, this %RE parameter is accurate for
the decision-making related to linearity assessment according to the fitness-for-purpose approach.

1. Introduction tion level, are very useful to indicate the range in which sensitivity
remains practically constant within a confidence limit of 5% according
Linearity is one of the most important performance characteristics to Dorschel et al. [6] and Huber [7]. A number of statistical tests, such
of an instrumental analytical method when linear response is assumed; as experimental Fratio (IUPAC) [8], ANOVA lack-of-fit (LOF) [9,10],
thus it is an interesting parameter to be evaluated in validation studies Mandel's test [11,12] and significance of the quadratic term [5] have
[1]. After adjusting a linear function to a set of calibration data, points been proposed to check linearity. Literature on linearity also consider
are expected to be randomly distributed around the obtained line. But several numerical parameters, such as correlation coefficient (not
reality goes its own way and points can show trends differing from advisable alone), standard deviation of regression, standard deviation
linear behavior, mainly at low or high concentration levels. The of the slope [13], quality coefficients [14,15] and back-calculated
accuracy of an analytical method is highly linked to linearity, which concentrations expressed as percentage of relative error (%RE)
must be assured over the whole calibration range [2]. There are [3,16–19]. The main problem is that some of these procedures are
numerous approaches for linearity testing, including both graphical not equivalent and contradictory results can be obtained. For this
and numerical methods [3]. Visual inspection of calibration and reason, common sense must be used to select adequate tools in order to
residuals plot are very useful, but they must be accompanied by assure linearity according to the fitness for purpose principle [20].
significance tests or numerical decision parameters [4,5]. Linearity There is no magical recipe for linearity testing, but two important
plots, obtained by representing the response factors versus concentra- aspects should be always considered, the easy visualization of data and


Corresponding authors.
E-mail addresses: jmjurado@us.es (J.M. Jurado), fraposo@cica.es (F. Raposo).

http://dx.doi.org/10.1016/j.talanta.2017.05.049
Received 3 March 2017; Received in revised form 10 May 2017; Accepted 15 May 2017
Available online 18 May 2017
0039-9140/ © 2017 Elsevier B.V. All rights reserved.
J.M. Jurado et al. Talanta 172 (2017) 221–229

Table 1
Calibration examples used in this work.

Example variable Data

1 [Cu], µg L−1 2, 2, 2, 4, 4, 4, 6, 6, 6, 8, 8, 8, 10, 10, 10


Absorbance, a.u. 0.023, 0.022, 0.023, 0.046, 0.047, 0.045, 0.066, 0.066, 0.065, 0.089, 0.090, 0.088, 0.113, 0.112, 0.113

2 [Quinine], mg L−1 1, 1, 1, 2, 2, 2, 3, 3, 3, 4, 4, 4, 5, 5, 5, 6, 6, 6, 7, 7, 7, 8, 8, 8, 9, 9, 9, 9, 10, 10, 10


Intensity, RFU 138.455, 137.885, 137.535, 248.2, 247.956, 247.466, 354.044, 354.117, 353.668, 455.47, 454.756, 454.312, 558.784, 558.449, 558.006,
650.048, 650.142, 649.991, 723.440, 723.672, 723.068, 815.512, 815.315, 815.493, 895.258, 895.022, 894.987, 972.986, 971.926,
970.385

3 [Zn], µg L−1 10, 10, 10, 20, 20, 20, 50, 50, 50, 75, 75, 75, 100, 100, 100, 500, 500, 500
Intensity, counts 161, 148, 155, 233, 224, 217, 455, 459, 448, 629, 639, 621, 795, 814, 799, 4043, 4031, 3970

4 [Isoamil acetate], mg L−1 0.5, 1, 2, 3, 4, 5, 10, 15, 20


Area ratio, dimensionless 0.013, 0.028, 0.058, 0.087, 0.131, 0.178, 0.379, 0.599, 0.804

Fig. 1. Linearity study for the calibration curve of copper by ET-AAS in the calibration range 2–10 µg L−1 (Example 1). A, Absorbance; a.u., absorbance units; RF, response factor.

a clear acceptance criterion allowing the decision-making. Finally, a plot of %RE of back-calculated concentrations against
In this work, a combination of graphical and numerical criteria for concentration level offers additional information. By this way, a
linearity testing is proposed by considering the adequacy to the numerical decision limit which takes into account the concentration
purposes of the method in a specified concentration range. Four of calibration standards can be easily included.
calibration datasets covering usual situations that can occur in
chemical analysis laboratories were selected. Calibration and residuals
2. Theory
plots are mandatory to inspect the fitting quality [21]. They must
include confidence limits, allowing us to visualize the goodness of
Consider a set of n calibration points (xi, yi), with apparent linear
linear fitting and detect outliers according to statistical criteria.
relationship, where each xi-value corresponds to a known concentra-
Linearity plot, based on the variation of the response factors, is also
tion of an analyte and each yi-value correspond to an instrumental
considered because it can highlight the needed of non-linear fitting.
signal. Data can be fitted by a linear calibration model obtained by

222
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 2. Linearity study for the calibration curve of quinine by fluorescence spectroscopy in the calibration range 1–10 mg L−1 (Example 2a). I, intensity; RFU, relative fluorescence units;
RF, response factor.

ordinary least squares (OLS) method by minimizing the sum of squares mental data are inside the confidence limits of the fitted model. The
of residuals: confidence limits must consider the leverage effect of each calibration
n n point on the regression line. The leverage score, hi, varying from 0 to 1,
S= ∑ ei2 = ∑ (yi − yˆ)i 2 measures the amount by which the predicted value would change if the
i =1 i =1 (1) observation was shifted one unit in the y direction. Leverage score of
where ŷi is the predicted dependent variable obtained from each each point in the case of XY linear regression is obtained by the
independent value xi as expression [4].

yˆi = b1⋅xi + b0 (2) 1 (x − x ) 2


hi = + n i
n ∑i =1 (xi − x )2 (3)
being b0 the estimate of intercept and b1 the estimate of slope for the
considered linear model. This leverage score is used to obtain the standard error (sŷ) of each
Following, different graphical criteria for linearity testing are estimate ŷi at position xi:
explained.
syˆ = sy / x⋅ hi (4)
2.1. Calibration plot and coefficients of correlation & determination
where sy/x is the standard error of the estimates, also known as residual
standard deviation.
The inspection of calibration plot must be the first action to be
And the 95% confidence interval (CI) for each estimate can then be
considered in the evaluation of linearity. This practice can be useful to
obtained from:
find out possible outliers, points of influence and non-linear data
trends. Relating to the coefficient of correlation (r), although widely CI = yˆ ± tcrit ·sŷ (5)
used, it is important to note that must not be used to check regression
data [5]. With respect to the coefficient of determination (R2), it must where tcrit is the t-Student value for α=0.05 and n-2 degrees of
not be considered as a measure of linearity taking into account that R2 freedom.
is the ratio of the variability explained by the model with regard to the A representation of CI against the concentration allows us to obtain
total (observed) variability. Sometimes, R2 values of 0.999 can be the confidence region for the calibration line. If we are interested in the
obtained for linear regression but the distribution of data along the confidence limits for the points used to obtain the regression line, the
calibration line indicates non-linear behavior. On the contrary, lower prediction interval (PI) must be computed as [8]:
R2 can be assumed for linear relationship depending on the fitness for
1
purpose of the method [5]. There are lots of well-established scientists PI = yˆ ± tcrit ·sy / x · + hi
m (6)
recommending not using this parameter as synonymous of linearity
[3]. where m is the number of replicates used to obtain the mean signal for
A better approach to evaluate linearity is verifying that experi- each calibration point. Experimental data must lie inside this region as

223
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 3. Linearity study for the calibration curve of quinine by fluorescence spectroscopy in the calibration range 1–6 mg L−1 (Example 2b). I, intensity; RFU, relative fluorescence units;
RF, response factor.

a first test to assure that the model really fit the data. Accordingly, standards [6,7]. The median of the computed response factors, defined
outliers and deviations from linearity are highlighted when calibration as Rc, is traced as a central line along the calibration range, and two
points appear out of this region. For simplicity, only PI will be plotted 5%-confidence lines obtained as 0.95·Rc and 1.05·Rc, are also plotted.
in the graphics presented in this work. If the computed response factors lie within the two limits, it clearly
indicates that linearity can be assured over the calibration range.
2.2. Residuals plot Positive and negative deviations from linearity are usually found at low
and high concentration levels, respectively.
Residuals plots are strongly recommended in several manuals on
chemometrics and quality control [5,21–23]. Plotting residuals can 2.4. Percent relative error of back-calculated concentrations (%RE)
help us to detect outliers, influence points, lack of fit or uneven
variance. If the selected function fits well the data, residuals must When calibration data are fitted by a model and the calibration
appear normally distributed and centered around zero. The use of function is obtained, it is possible to calculate a predicted concentration
scaled residuals, i.e., residuals (ei) divided by sy/x is strongly recom- value for each calibration standard by substituting their corresponding
mended. They are very useful because a sample from the standard measured signal into the function. The difference between this so called
normal distribution N (0, 1) is obtained and the computed residuals back-calculated concentration and the nominal concentration is the
can be easily compared with the corresponding z-value. But this error associate to the concentration of each standard when it is
standardization does not consider the leverage effect of each point on predicted by the model. This difference can be used to calculate a RE
the regression line. The standard error of each residual (sei ) can be dividing its value by the nominal content. This parameter was used
obtained by considering the leverage score and then compute the so previously in the analytical literature to try obtaining some measure of
called Studentized residual as: the error when using the proposed regression equation [16–19].
ei ei According to some documents, to verify linear range these relative
ti = = errors must lie between ± 15% for all the calibration range, and
sei sy / x⋅ 1 − hi (7)
between ± 20% when limit of quantification (LOQ) is reached [3,24].
Studentized residuals must lie between ± 1.96 limits (for conve- Although these literature limits are accepted, however the dependence
nience rounded to ± 2) over the whole calibration range. On the on the concentration of calibration standards is not considered. The
contrary, the selected model is not acceptable. Guidelines for Standard Method Performance Requirements published
in the AOAC Official Methods of Analysis proposes an expected
2.3. Response factors plot recovery interval as a function of analyte concentration [25] (Table
S1 in the Supplementary electronic material). The differences between
Linearity plots, which are very used in chromatographic method the limits of this interval and 100% could be considered as the expected
development, are obtained plotting the signal-to-concentration ratio relative error depending on the concentration level. For instance, at
(the so called response factor) against the concentration of calibration analyte concentrations of 0.1%, the AOAC recovery lies between 95%

224
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 4. Linearity study for the calibration curve of zinc by ICP-AES in the calibration range 10–500 µg L−1 (Example 3a). I, emission intensity; RF, response factor.

and 105%, and the expected relative error at this level could be The estimated %RE-value at C=10−7 is 11.9%. Although this value is
established as ± 5%. We propose as novelty the use of these concen- higher than 10%, we consider our approximation adequate because it is
tration-dependent relative errors as a limit for linearity testing over the lower than the 15% limit proposed by previous literature [3,24]. From
calibration range. The main problem is that the intervals proposed by 10−6 to 10−3, our proposed limit always is more restrictive than the
AOAC are not symmetric. For instance, at 100 ppm the recovery expected according to AOAC recovery model. By this way, Eq. (9) is
interval is 90–107%. When lower concentration levels are considered proposed to estimate the theoretical %RE-values from any mass
this asymmetry grows. With the aim of establishing a single criterion, fraction over the calibration range.
we propose the use of the upper level of the AOAC recovery intervals, as
%RETh = 2·C −0.11 (9)
it is the most restrictive situation, to obtain the %RE for each
concentration level. The computed relative errors are included in By using this expression, the calculated confidence %RE-limits can
Table S1. Our final objective is to obtain a plot of relative errors of be easily included on the %RE-plot for the graphical evaluation of
back calculated concentrations versus concentration levels, henceforth linearity.
called the %RE-plot. In order to facilitate the drawing of confidence
limits on the %RE-plot, the computed relative errors has been related
3. Examples of analytical calibration
with the analyte ratio expressed in mass fraction (C) using a function in
the form:
Analytical calibration is an overall procedure to express the
%RE = a·C −b (8) relationship between concentration of analytes and detector response
that can be applied to different analytical techniques (atomic absorp-
As can be seen, this function is very similar to the Horwitz function
tion spectroscopy-AAS, fluorescence spectroscopy-FS, inductively
relating reproducibility and analyte content [25,26]. The fitting proce-
coupled plasma-ICP, gas chromatography-GC, and high performance
dure was carried out in an Excel Spreadsheet by using Solver comple-
liquid chromatography-HPLC). On this manuscript, the examples have
ment to obtain parameters a and b. In order to ensure the adequate
been selected mainly to cover different situations for calibration curves,
value of %RE at analyte ratio of 1, the value of a was restricted to ≤2
independently of the technique selected to carry out the measurements.
(that is the expected %RE at this concentration level). The errors of the
The data sets are shown in Table 1.
fitted parameters were evaluated by means of numerical differentiation
[27]. The results obtained were a=2.0 ± 0.4 and b=0.11 ± 0.01; giving a Example 1. External calibration data used for the direct
R2 value of 0.95. The supplementary electronic material includes the % determination of Cu in aniseed spirits by electrothermal atomic
RE-plot (Fig. S1), calculated from AOAC, against analyte ratio in absorption spectroscopy (ET-AAS) [28]. Each standard was measured
logarithmic scale. This plot also includes the representation of the fitted in triplicate from genuine replicates.
function. As can be seen, the function fits better at the extremes of the
analyte ratio that at the central points. It is logical because from 10−7 to Example 2. Calibration curve for quinine measured by FS.
10−5 the %RE-value expected according AOAC is always the same, 10%. Fluorescence measurements were carried out by using a Shimadzu

225
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 5. Linearity study for the calibration curve of zinc by ICP-AES in the calibration range 10–100 µg L−1 (Example 3b). I, emission intensity; RF, response factor.

RF-1501 spectrofluorophotometer. Quinine standards were prepared be seen, all calibration points are found within the prediction interval
in 0.05 M sulfuric acid and fluorescence intensity was recorded. (Fig. 1A) and R2 present a value of 0.9991. Studentized residual plot
Excitation and emission wavelengths were set at 350 nm and (Fig. 1B) also shows all the computed residuals within the ± 2 limits.
450 nm, respectively. Each standard was measured in triplicate from The same can be observed in the case of response factors (Fig. 1C),
genuine replicates. whose values are included between ± 5%. In Fig. 1D, back-calculated
relative errors (%RE) are represented against concentration of stan-
Example 3. External calibration data used for the determination of Zn
dard solutions. This plot includes the ± 15% limits (dotted lines) and
in acid-digested aniseed spirits by inductively coupled plasma atomic
the limits, proposed by authors, related to the concentration levels
emission spectroscopy (ICP-AES) [29]. Each standard was measured in
(dashed lines). Note that the proposed limits are slightly greater than
triplicate from genuine replicates.
± 15% at the lowest levels. This fact can soften the decision-making
Example 4. Internal calibration data used for the determination of when calibration is performed at very low concentrations. In this case,
isoamil acetate in wines by head-space solid-phase microextraction and all %RE values lie within both limits (theoretical and calculated) and
GC with flame ionization detector (HS-SPME-GC-FID) [30]. Ethyl then linearity can be assured on the whole calibration range.
heptanoate (10 mg L−1) was used as internal standard. As the HS-
SPME extraction time was 60 min for each standard, the calibration 4.2. Example 2. A curvilinear calibration with evenly spaced data
points were measured once.
Calibration plots, standardized residuals plots, response factors Molecular fluorescence is a highly sensitive technique. At low
plots and %RE-plots have been obtained in order to comparatively concentration levels, the dependence between signal and analyte
study the linearity for each of the four selected examples. All these content can be considered linear. However, it is usual to find non-
methods have been applied by plotting their corresponding variables linear relationships between signal and concentration at high levels
and drawing the confidence limits. For each figure four graphs have and, accordingly, non-linear calibration models can be fitted [31,32]. In
been plotted: A) Calibration plot including prediction interval; B) Plot this example, the calibration range was intentionally selected from 1 to
of Studentized residuals including ± z limits (z=2); C) Response factor 10 mg L−1 to highlight this kind of situations. As can be seen in Fig. 2A
linearity plot including ± 5% limits (α=0.05); D) %RE-plot including linear model does not fit well to data and the residual distribution,
± 15% limits (dotted lines) and the proposed limits according to the although all calibration points appeared within the limits of the
concentrations of calibration standards (dashed lines). prediction interval. The distribution of the residuals (Fig. 2B) also
shows a non-linear behavior. According to the profile that can be seen
4. Results and discussion on the response factor plot (Fig. 2C), sensitivity decreases when
concentration of calibration standards is raised. Note that in the %
4.1. Example 1. A linear calibration with evenly spaced data RE-plot, showed in Fig. 2D, the proposed limits are lower than 15%,
due to the concentration of the standards. In this graph, a high
Fig. 1 shows the plots obtained in the case of the Example 1. As can deviation of the first calibration point is observed, although the

226
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 6. Linearity study for the calibration curve of isoamyl acetate by HS-SPME-GC-FID in the calibration range 0.5–20 mg L−1 (Example 4a). AR, area ratio (dimensionless); RRF,
relative response factor, obtained as AR divided by C.

problem becomes from high concentration levels. It is known that this 0.9997, however the calibration must not be considered adequate.
deviation is due to a bad selection of the calibration levels and, The calibration points lie inside the prediction interval lines (Fig. 4A),
accordingly, some reduction in the calibration range must be necessary. but the residuals plot (Fig. 4B) shows a not random distribution of the
A new shorter calibration line was obtained in the range 1– five residuals corresponding to lower levels. Linearity plot (Fig. 4C)
6 mg L−1. As can be seen in Fig. 3A, all calibration points lie within shows a high deviation of the two first response factors. Finally, paying
the prediction interval of the regression curve, but the distribution of attention to Fig. 4D, the first point is out of any acceptance limit of %
residuals still shows non-linear trends. In fact, a second grade RE.
polynomial would fit also the data leading to a R2 value of 0.9999. In Considering these results, a new shorter calibration range must be
addition, linearity plot based on response factors also indicates a non- selected to obtain reliable results. Obviously, the problematic concen-
linear response. From our experience with this kind of plots, in some tration level is 500 µg L−1 and, accordingly, the calibration line has
situations it is usual to find strong variations in the response factors, been optimized by using the range from 10 to 100 µg L−1. The results
but adequate linearity for the fitness for purpose of the determination can be seen in Fig. 5. In this case, the R2 value is 0.9996. Fig. 5B shows
method is achieved. According to the profile of the response factor that the Studentized residuals were not randomly distributed. In
showed in Fig. 3C, sensitivity still decreases when concentration of addition, four of the response factors (Fig. 5C) appear out of the ±
calibration standards is raised. This is a typical deviation that occurs in 5% limits. Actually, response factor behaves as if the model was not
a number of real situations, but analysts can accept it if interpolation linear. This is the same situation found in Example 2, although
errors are kept within adequate limits. When Fig. 3D is checked, it can polynomial model could fit better, the linear model is selected due to
be seen that the %RE of all back-calculated concentrations are easy interpretation and calculations because the fitness for purposes is
distributed within the calculated acceptance limits and, in fact, they achieved. In fact, %RE of back-calculated contents are lower than 3%
are lower than 6%. In view of such results, linearity can be assured in and lie within the accepted calculated limits, as can be seen in Fig. 5D.
this calibration range.
4.4. Example 4. A linear calibration with deviations at lowest
4.3. Example 3. A linear calibration using uneven data distribution concentration levels

Uneven data distribution is one of the principal causes of linearity Deviations from linearity are also frequent at low concentration,
failing due to the influence of points with high leverage. Ideally, the when the measurement is done close to the quantification limit. At that
concentrations should be evenly spaced across the calibration range to concentration levels the signal usually tends at the background signal.
avoid this problem. As can be seen in Fig. 4, in the case of Example 3, In this example, positive deviation is observed in Fig. 6A, although all
not equidistant calibration points were selected and the distance calibration points appear inside the prediction interval, and no random
between concentration levels grow along the x-axis. In Fig. 4A, it can distribution is observed for the four first residuals (Fig. 6B). As
be seen how the last calibration point presents a high influence on the expected, the relative response factors (RRFs) increases with the
calibration line. The determination coefficient presents a value of concentration of the calibration standards (Fig. 6C). The two first %

227
J.M. Jurado et al. Talanta 172 (2017) 221–229

Fig. 7. Linearity study for the calibration curve of isoamyl acetate by HS-SPME-GC-FID in the calibration range 3–20 mg L−1 (Example 4b). AR, area ratio (dimensionless); RRF,
relative response factor, obtained as AR divided by C.

RE values are higher than 25%, being both data out of the proposed points and deviations at low concentration levels. In these cases, the
limits. results provided by the different graphical plots could be controversial.
The calibration line was accordingly built by reducing the calibra- To a better understanding of linearity subject, the main objective of this
tion range from the lowest concentration level (Fig. 7). These points paper is to propose the different graphical plots in the form of a flow
were sequentially removed and finally, the calibration plot was chart to definitively making a decision about the goodness-of-fit for
obtained using 6 points in the range of 3–20 mg L−1. The R2 value calibration curves. First of all, the quality of the adjustment is observed
increased from 0.9988 to 0.9998. In addition, the residuals profile is in the calibration and residual plots. Following, the response factors
randomly distributed, as can be seen in Fig. 7B. The response factor plot allow us to know any deviation on sensitivity in the considered
still increases with concentration and most of the points appear out of calibration range. The evolution of the response factors could indicate
± 5% limits. However, the %RE-plot presents all points within the some deviations from linearity and the need of other calibration
calculated confidence interval, with values lower than 2.5%. Taking function. Finally, it is important to note the goodness-of-fit provided
into account these results, linear response can be assumed within this by the %RE-plot in all these examples. By this way, after selecting the
calibration range according to the fitness for purpose approach. appropriate calibration range, the linearity of response was assured by
means of this graph. Therefore, we propose here that for linearity
assessment the fitness for purpose of the method should prevail over
5. Conclusions
any other criterion. For that reason, the final decision about linearity
should be made accurately considering either numerically or graphi-
The evaluation of linearity of calibration functions is a very
cally representing the plot of %RE of back-calculated concentrations. In
important subject relating to the selection of a model for unknown
any case, the limits should be established according to the analyte
samples quantitation. Traditionally, three different plots were consid-
concentration level.
ered for the assessment of linearity. Firstly, the widely used calibration
plot by means of the lack of reliability R2 criterion. Secondly, the
Appendix A. Supplementary material
valuable residuals plot or the more sophisticated Studentized version.
Thirdly, the very useful response factors plot. A fourth graphical
Supplementary data associated with this article can be found in the
criteria based on the relative error of back-calculated concentrations
online version at doi:10.1016/j.talanta.2017.05.049.
(%RE-plot) has been also considered. For the first time, decision limits
are included in the plot depending on the concentration levels of
References
calibration standards. These limits are derived from the AOAC
expected recovery at each analyte level.
[1] E. Hsieh, J.P. Liu, On statistical evaluation of the linearity in assay validation, J.
For typical linear calibration curves, the four graphical plots show Biopharm. Stat. 18 (2008) 677–690.
agreement between theory and experimental data. However, for day-to [2] International Conference on Harmonisation (ICH) of technical requirements for
day analytical calibration procedures the trends including non-linear registration of pharmaceuticals for human use. Harmonized Tripartite Guideline.
Validation of Analytical Procedures: text and Methodology Q2 (R1), 2005.
behavior are very frequent, mainly due to the influence of high leverage

228
J.M. Jurado et al. Talanta 172 (2017) 221–229

[3] F. Raposo, Evaluation of analytical calibration based on least squares linear [18] P. Hubert, P. Chiap, J. Crommen, B. Boulanger, E. Chapuzet, N. Mercier,
regression for instrumental techniques: a tutorial review, TRAC-Trend Anal. Chem. S. Bervoas-Martin, P. Chevalier, D. Grandjean, P. Lagorce, M. Lallier, M.C. Laparra,
77 (2016) 167–185. M. Laurentie, J.C. Nivet, The SFSTP guide on the validation of chromatographic
[4] S.V.C. De Souza, R.G. Junqueira, A procedure to assess linearity by ordinary least methods for drug bioanalysis: from the Washington Conference to the laboratory,
squares method, Anal. Chim. Acta 552 (2005) 23–35. Anal. Chim. Acta 391 (1998) 135–148.
[5] S.L.R. Ellison, V.J. Barwick, T.J. Duguid, Practical Statistics for the Analytical [19] J.W. Dolan, Calibration curves, part 3: a different view, LC-GC Eur. 22 (2009)
Scientist, The Royal Society of Chemistry, Cambridge, 2009. 304–308.
[6] C.A. Dorschel, J.L. Ekmanis, J.E. Oberholtzer, F.V. Warren Jr., B.A. Bidlingmeyer, [20] Eurachem Guide. The fitness for purpose of analytical methods. A Laboratory
LC detectors: evaluation and practical implications of linearity, Anal. Chem. 61 Guide to Method Validation and Related Topics, Second ed., 2014.
(1989) 951A–968A. [21] M. Thompsom, P.J. Lowthian, Notes on Statistics and Data Quality for Analytical
[7] L. Huber, Validation of analytical methods: review and strategy, LC-GC Eur. 11 Chemist, Imperial College Press, London, 2011.
(1998) 96–105. [22] E. Mullins, Statistics for the Quality Control Chemistry Laboratory, The Royal
[8] K. Danzer, L.A. Currie, Guidelines for calibration in analytical chemistry. Part 1: Society of Chemistry, Cambridge, 2003.
fundamentals and single component calibration, Pure Appl. Chem. 70 (1998) [23] J.M. Miller, J.C. Miller, Statistics and Chemometrics for Analytical Chemistry, 6th
993–1014. ed., Pearson Education Ltd, Harlow, 2010.
[9] Analytical Methods Committee, Is my calibration linear? Analyst 119, 1994, 2363– [24] US EPA. Method 8000C-Determinative Chromatographic Separations, Revision 3,
2366. March, 2003, p. 66.
[10] M.G. Akritas, N. Papadatos, Heteroscedastic one-way ANOVA and lack-of-fit tests, [25] AOAC, Appendix F: guidelines for standard methods performance requirements,
J. Am. Stat. Assoc. 466 (2004) 368–382. in: AOAC Official Methods of Analysis, 2012.
[11] J. Mandel, The statistical Analysis of Experimental Data, Dover Publications, New [26] W. Horwitz, L.R. Kamps, K.W. Boyer, Quality assurance in the analysis of foods and
York, 1964. trace constituents, J. Assoc. Off. Anal. Chem. 63 (1980) 1344–1354.
[12] J.M. Andrade, M.P. Gómez-Carracedo, Notes on the use of Mandel´s test to check [27] E.J. Billo, Excel for Scientist and Engineers: numerical methods, John Wiley &
for nonlinearity in laboratory calibrations, Analyst 5 (2013) 1145–1149. Sons, Hoboken, EEUU, 2007.
[13] A.M. Cuadros-Rodríguez, C. García Campaña, Jimenez Linares, M. Roman Ceba, [28] J.M. Jurado, M.J. Martín, F. Pablos, A. Moreda, P. Bermejo-Barrera, Direct
estimation of performance characteristics of an analytical method using the data set determination of copper, lead and cadmium in aniseed spirits by electrothermal
of the calibration experiment, Anal. Lett. 26 (1993) 1243–1258. atomic absorption spectroscopy, Food Chem. 101 (2007) 1296–1304.
[14] P. Vankeerberghen, J. Smeyers-Verbeke, The quality coefficient as a tool in [29] J.M. Jurado, A. Alcázar, M.J. Martín, F. Pablos, A.G. González, Determination of
decisions about the quality of calibration in graphite furnace atomic absorption Zn, B, Fe, Mg, Ca, Na and Si in anisette samples by inductively coupled plasma
spectrometry, Chemom. Intell. Lab. Syst. 15 (1992) 195–202. atomic emission spectrometry, Talanta 63 (2004) 297–302.
[15] J.O. de Beer, C. Naert, E. Deconinck, The quality coefficients as performance [30] J.M. Jurado, O. Ballesteros, A. Alcázar, F. Pablos, M.J. Martín, J.L. Vilchez,
assessment parameter of straight line calibration curves in relationship with the A. Navalón, Differentiation of certified brands of origins of Spanish white wines by
number of calibration points, Accredit. Qual. Assur. 17 (2012) 265–274. HS-SPME-GC and chemometrics, Anal. Bioanal. Chem. 390 (2008) 961–970.
[16] E.L. Jhonson, D.L. Reynolds, D.S. Wright, L.A. Pachla, Biological sample pre- [31] D.A. Skoog, F.J. Holler, S.R. Crouch, Principles of Instrumental Analysis, 6th ed.,
paration and data reduction concepts in pharmaceutical analysis, J. Chromatogr. Thomson Brook/Cole, Belmont, USA, 2007.
Sci. 26 (1988) 372–379. [32] J.M. Jurado, R. Muñiz-Valencia, A. Alcázar, S.G. Ceballos-Magaña, J. González,
[17] M. Mulholland, D.B. Hibbert, Linearity and the limitations of least squares Fitting chemical data with Excel: a practical tutorial (in Spanish), Educ. Quím. 27
calibration, J. Chromatogr. A 762 (1997) 73–82. (2016) 21–29.

229

Anda mungkin juga menyukai