preoperative checkout
In this section
1. Every day before your first patient
2. Before every patient
i
o Set the bag/ vent switch to vent.
o Fill the bellows using 02 flush.
o Check that mechanical ventilation starts. Check that the bellows inflate
and deflate. Check that the display shows the correct ventilator date.
Check that there are no inappropriate alarms.
- Turn the system circuit breaker off and check that mechanical ventilation
continues while the system is running on battery power. After completing the
check, turn the sytem circuit breaker on. The mains indicator is lit when AC
power is connected.
- Set the appropriate control and alarm limits for the case.
ii
o Set the ventilator to VCV mode and the settings for TV to 400ml, RR to
12, I;E ti 1:2, T pause to off, and P max to 40cm h2o.
o Set the gas flow to the minimum settings.
o Start a case.
o Set the bag/ vent switch to vent.
o Fill the bellows using oksigen flush.
o Check that mechanical ventilation starts. Check that the bellows inflate
and deflate. Check that the display shows the correct ventilator date.
Check that there are no inappropriate alarms.
- Set the appropriate controls and alarm limits for the case.