ABX Pentra HbA1c WB Cal

ABX Pentra
Ref.: A11A01703
Volume R1: 1 x 8 ml HbA1c WB Cal
Volume R2-6: 5 x 2 ml

Calibrator for the measurement of haemoglobin A1c 2008/06/23


A93A00782A US

(HbA1c) and total haemoglobin.


A11A01703
Intended use 1 x 8 ml
ABX Pentra HbA1c WB Cal is used to calibrate ABX Pentra HbA1c WB,
Ref. A11A01702. 5 x 2 ml

Characteristics
• ABX Pentra HbA1c WB Cal is a liquid multicalibrator.
Different constituents are added in order to obtain specific
concentrations:
CAL 1 (1 x 8 ml): contains 0.4% NaOH, surfactant and chlorohemin.
CAL 2-6 (1 x 2 ml): contains Buffer, Hapten. HORIBA ABX
BP 7290
34184 Montpellier - cedex 4 - France
• ABX Pentra HbA1c WB Cal is ready-to-use. The kit is composed of 6
vials:
CAL 1: 1 vial of 8 ml
CAL 2-6: 5 vials of 2 ml
On ABX Pentra 400:
ABX Pentra HbA1c WB Cal should be used according to this calibrator
notice and as specified in the respective instructions for use of the Levels HbA1c (g/dl) HbA1c (μmol/l)
reagent. HORIBA ABX cannot guarantee its performance if used CAL 1 0.00 0.00
otherwise. CAL 2 0.42 1.60
CAL 3 0.76 2.90
Handling CAL 4 1.11 4.20
1. Break the vial cap. CAL 5 1.85 7.00
2. Invert the calibrator at least five times in order to obtain a CAL 6 2.85 10.80
homogenous mixture.
3. Transfer the calibrators to different cups and place them in the Level 1 contains the equivalent of 18.0 g/dl (68.29 μmol/l) of total
appropriate positions on the ABX Pentra 400. haemoglobin.
4. Unused products should be refrigerated. Mix the content
thoroughly before reuse. Storage and Stability
Calibrators, unopened or after opening, are stable up to the expiry date
Materials required but not provided on the label if stored at 2-8°C and contamination is avoided.
• HORIBA ABX reagents and clinical chemistry analyser: ABX Pentra 400.
• Standard laboratory equipment. Waste Management
• NaCl solution, 9 g/l. Please refer to local legal requirements.

Assigned values General Precautions


ABX Pentra HbA1c WB Cal is standardised in relation to a HPLC 1. ABX Pentra HbA1c WB Cal should be used only for the
method (HbA1c) and the Drabkin method (total haemoglobin). determination of the calibration curve.
2. Calibrator, for professional in-vitro diagnostic use only.
The total haemoglobin is calibrated with the kit's CAL 1.
3. Caution. Irritant. Avoid contact with the eyes, skin and clothing.
CAL 1 is used only with Total hemoglobin.
Wash hands carefully after handling the product. Do not ingest.
For the level 1 of HbA1c, use NaCl (9 g/l).
4. The calibrator vials should be discarded after use.
5. Please refer to the MSDS associated with the calibrator.

Warning
It is the user's responsibility to verify that this document is applicable
Form-0849 Rev. 2

to the calibrator used.

S.A.S. au capital de 41.700.000 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com
ABX Pentra
HbA1c WB Cal

S.A.S. au capital de 41.700.000 € - RCS Montpellier 328 031 042 - SIRET 328 031 042 000 42 - APE 332 B Latest version documents on www.horiba-abx.com

Anda mungkin juga menyukai