Anda di halaman 1dari 4

ProKnow Page 1 of 4

UW TG244 Case 05 Head and Neck


Your Submission 

Instructions (/planning/studies/591463b21740cdf604f500897ce5beee/instructions)

Download (/planning/studies/591463b21740cdf604f500897ce5beee/download)

Upload, Review and Submit (/planning/studies/591463b21740cdf604f500897ce5beee/candidate)

Your Submission (/planning/studies/591463b21740cdf604f500897ce5beee/submission)

Population Results (/planning/studies/591463b21740cdf604f500897ce5beee/population)

Learn (/planning/studies/591463b21740cdf604f500897ce5beee/learn)

Submission
Download Report (/api/psp/studies/591463b21740cdf604f500897ce5beee/solutions/submission/result/5dc4877d2640c6aae7e56f056de8f4e6/report)

Remove Submission

Powered by PlanIQ™
130.63 / (http://www.sunnuclear.com/solu
Score:
How are scores calculated? (http://support.proknowsystems.com/hc/en-us/articles/360022094014) 150.00

Slides (32)

https://proknowsystems.com/planning/studies/591463b21740cdf604f500897ce5beee/subm... 11/8/2019
ProKnow Page 2 of 4

Info

Participant Name Stephanie Hufnal

Participant Email hufnal3058@uwlax.edu

Participant Role student

Participant Institution Univ. of WI Lacrosse

Creation Date 2019-11-07T21:07:00.021795+00:00

Scoring Version 2.2.0

TPS Manufacturer Varian Medical Systems

TPS Model Eclipse

Radiation Type(s) Photon

Delivery Type(s) VMAT

Number of Treatment Beams 3

Beam Energies (MV) 6

Total Monitor Units 650.4476

Machine Info Varian Medical Systems XMLC (5.0, 10.0 mm)

Dose Grid Max Dimension (mm) 2.5

Dose Grid Voxel Volume (mm^3) 15.625

Study Name UW TG244 Case 05 Head and Neck

Results

METRIC RESULT GOAL 1 (Min GOAL 2 POINTS POINTS


Requirement) (Ideal) (MAX)

[01] Volume (%) of the 95.80 PASS [Goal = 90] PASS 15.00 15.00
PTV70 covered by 70 (Gy) [Goal = 95]

[02] Volume (%) of the 0.75 PASS [Goal = 10] FAIL [Goal 6.47 7.00
PTV70 covered by 73.5 = 0]
(Gy)

[03] Dose (Gy) covering 74.40 PASS [Goal = 77] FAIL [Goal 3.71 7.00
0.03 (cc) of the PTV70 = 71.3]

[04] Volume (%) of the 99.92 PASS [Goal = 95] PASS 7.00 7.00
CTV70 covered by 70 (Gy) [Goal = 99]

[05] Volume (%) of the 93.18 PASS [Goal = 90] FAIL [Goal 13.18 15.00
PTV63-PTV70 covered by = 95]
63 (Gy)

[06] Volume (%) of the 23.60 PASS [Goal = 60] FAIL [Goal 6.37 7.00
PTV63-PTV70 covered by = 20]
66.15 (Gy)

[07] Volume (%) of the 99.25 PASS [Goal = 95] PASS 7.00 7.00
CTV63-CTV70 covered by [Goal = 99]
63 (Gy)

https://proknowsystems.com/planning/studies/591463b21740cdf604f500897ce5beee/subm... 11/8/2019
ProKnow Page 3 of 4

[08] Volume (%) of the 96.08 PASS [Goal = 90] PASS 15.00 15.00 
PTV56-PTV63 covered by [Goal = 95]
56 (Gy)

[09] Volume (%) of the 17.66 PASS [Goal = 60] PASS 7.00 7.00
PTV56-PTV63 covered by [Goal = 30]
58.8 (Gy)

[10] Volume (%) of the 97.95 PASS [Goal = 95] FAIL [Goal 5.95 7.00
CTV56-CTV63 covered by = 99]
56 (Gy)

[11] Conformation Number 0.75 PASS [Goal = FAIL [Goal 8.47 12.00
[53.2 (Gy), PTV56] 0.5] = 1]

[12] Dose (Gy) covering 46.80 PASS [Goal = 48] PASS 0.00 0.00
0.03 (cc) of the [Goal =
SPINAL_CORD 47.999]

[13] Dose (Gy) covering 50.78 PASS [Goal = 52] PASS 0.00 0.00
0.03 (cc) of the [Goal =
BRAINSTEM 51.999]

[14] Dose (Gy) covering 32.91 PASS [Goal = 45] PASS 5.00 5.00
0.03 (cc) of the RT [Goal = 35]
COCHLEA

[15] Dose (Gy) covering 35.65 PASS [Goal = 45] FAIL [Goal 4.68 5.00
0.03 (cc) of the LT = 35]
COCHLEA

[16] Volume (%) of the 28.78 PASS [Goal = 40] FAIL [Goal 4.12 5.00
LIPS covered by 30 (Gy) = 20]

[17] Mean dose (Gy) to the 26.67 PASS [Goal = 30] FAIL [Goal 4.16 7.00
RT_PAROTID = 24]

[18] Volume (%) of the 12.66 PASS [Goal = 25] FAIL [Goal 4.34 5.00
MANDIBLE covered by 70 = 10]
(Gy)

[19] Mean dose (Gy) to the 41.44 PASS [Goal = 50] FAIL [Goal 6.42 7.00
LARYNX = 40]

[20] Volume (%) of the 16.21 PASS [Goal = 30] FAIL [Goal 3.76 5.00
POST NECK covered by = 10]
35 (Gy)

[21] Structure(s) containing BODY, CTV56, PASS [Goal = FAIL [Goal 3.00 5.00
the global max dose point CTV63, CTV63- Elsewhere] = CTV70]
CTV70, PTV56,
PTV63, PTV70

[22] Number of treatment 3.00


beams

[23] Number of unique 1.00


isocenters

[24] Number of unique 1.00


couch angles

[25] Cumulative meterset 650.45


over all treatment beams

[26] Estimated 'beam-on' 3.00 


time, all beams (minutes)
 

https://proknowsystems.com/planning/studies/591463b21740cdf604f500897ce5beee/subm... 11/8/2019
ProKnow Page 4 of 4

[27] Max dose voxel 2.50


dimension (mm)

[28] Structure(s) not fully NONE


covered by dose grid

TOTALS 21 total metrics 21 8 130.63 150.00

Comments
----------- DICOM DETECTIVE ALERTS ----------- Inconsistency in Patient ID: 002445RS [RT Plan] 002445 [RT
Structure Set] 002445RS [RT Dose] 002445 [CT Imageset] Inconsistency in Plan <-> Structure Set: The current RT
Plan does not reference the SOP Instance UID of the current RT Structure Set.

https://proknowsystems.com/planning/studies/591463b21740cdf604f500897ce5beee/subm... 11/8/2019

Anda mungkin juga menyukai