Faculty of Engineering
An Undergraduate Study
presented to the
Department of Chemical Engineering
Faculty of Engineering
University of Santo Tomas
In Partial Fulfillment
of the Requirements for the Degree of
Bachelor of Science in Chemical Engineering
by
April 2019
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING i
APPROVAL SHEET
__________________________________
Prof. Edna C. Quinto, Ph.D, ASEAN Eng
Thesis Adviser
_______________________ _______________________
Engr. Cristina E. Tiangco, Ph.D Engr. Rose Ann Z. Tamolang
Department of Chemical Engineering Department of Chemical Engineering
Panelist Panelist
_______________________
Engr. Mildred A. Ebuengan
Entrepreneur, Owner EM6 Flight Enterprise
Panelist
_________________________
Engr. Mark Emile H. Punzalan
Chair, Department of Chemical Engineering
University of Santo Tomas
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING ii
ACKNOWLEDGEMENT
To our thesis adviser, Prof. Edna C. Quinto, PhD., ASEAN Eng., the
accomplishment of this study would have not been possible without your
guidance, undying support, understanding, and precious time. We express our
sincerest gratitude for sharing us your expertise and for assisting our group in
every experiment that we had to do. We have been privileged with such an
adviser who is extremely patient especially during the times we had to stay until
late night in order to finish our experiments, and for never doubting us to do this
study.
To the Research Center for the Natural and Applied Sciences (RCNAS), for
aiding us during the experimentation phase of our research study and for giving us
permission in using the necessary equipment that we needed for our study.
To Mr. Florencio De los Reyes, M.Sc., for allowing us to use the UV-Vis
Spectrophotometer available in Lab 11 and work in the balance room especially in
urgent times. Also, for always reminding us the proper rules when working inside
the laboratory.
Lastly, to the Almighty God, for surrounding us all these good people who
contributed to the completion of our study. We express our deepest gratitude to
You for always giving us the grace to continue this research despite all the
problems we have encountered. Thank you for guiding us all the way and thus, all
praises and the achievement of this study is given unto You.
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING iv
ABSTRACT
The study shows the response of standards of Iron (III) to excess ammonium
the wavelength of 480nm. The method shows an RSD of 6.56% when a 20ppm
iron (III) standard was read continuously for 5 minutes. The signal is repeatable
with an RSD of 8.12% when the 20ppm standard is compared to a blank solution.
The calculated LOD is 3.38M/ΔAbs for the dynamic range of 10ppm to 35ppm. The
close to the target samples for which the detection method may work. The figures
of merit of the method show promising application for iron (III) detection.
Preliminary steps were done to test the method on milk samples. However the
opacity of the milk or casein shows the reverse response. Probably other
components of milk are interfering with the reaction with the thiocyanate
reaction. Further steps are therefore recommended for the use of the method.
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING v
TABLE OF CONTENTS
LIST OF TABLES
LIST OF FIGURES
desktop computer
Figure 4.3.1 Calibration Curve for ∆Abs Range: 0.0106 to 0.1235 (n=3)
Figure 4.3.2 Calibration Curve for ∆Abs Range: 0.0106 to 0.1235 with error bars
CHAPTER 1
THE PROBLEM
1.1 Introduction
Iron is a vital nutrient that is present in almost all aspects of our lives. Its main
role is in the production of hemoglobin in red-blood cells and for oxygen transport
mechanism in the blood for all vertebrate and some invertebrate animals [1].
However, iron in excess can cause significant effects ranging from liver damage to
methods in detecting different mineral, these current techniques require the use
methods [2].
Due to the importance of heavy metals at trace level in the human health and
metals in the environmental samples have been continuously carried out on the
analytical and environmental laboratories around the world. [3-5]. Among heavy
metals, iron in small amounts is an essential element for most life on Earth,
including humans and animals. It is well known that an iron deficiency is the most
common cause of anemia. On the other hand, too much iron can cause several
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 2
health problems. High levels of iron are associated with an increased risk for
cancer, heart disease and other illnesses such as endocrine problems, arthritis,
coupled plasma mass spectrometry (ICP-MS) have been proposed for the
natural water and other real samples of environmental interest in which they are
The present work describes the attempt for the determination of iron ions
samples.
1.2 Objectives
blood for all vertebrate and few invertebrate animals [10]. There are several
to find an alternative that gives the result without the required protocol of
analytical methods. Thus, this study will be a significant endeavor in modifying the
determine the changes in absorbance in the liquid samples. The principle behind is
visible (390–780 nm) radiation by chemical species in solution” [11]. Some organic
and inorganic species inherently absorb radiation in the UV-vis range, and the
species. [11]
1.5 Hypothesis
reagent of an optical sensor for iron detection in liquid samples. Being the active
reagent of the optical method, ammonium thiocyanate will react and show visible
samples. Preliminary steps would be done to test the method on milk samples.
Although, other components of milk that might interfere and obstruct with the
reaction with the thiocyanate reaction, we would like to state that this aspect of
the study would be beyond the scope and limitation of the study. Furthermore, a
non-selective type of method would be develop, thus the iron present in a sample
will have to be isolated if possible. The actual data gathering of the study using a
would be conducted at the Research Center for the Natural Sciences of Thomas
February 2019.
Iron. The most common metallic element in the universe. Pure iron is a soft,
lustrous, dark silvery-gray metal. Iron is a strongly reactive metal, being very
reactive with acids, and forms oxides, commonly known as rust, with air and
water.
soluble in water. The odorless inorganic compound has many industrial and
detector.
CHAPTER 2
REVIEW OF RELATED LITERATURE
This chapter contains the main principles about optical sensing, its
The oxidation of ferrous iron, the use of hydrogen peroxide, effects of iron
in the human body and in water on the aquatic life, as well as other sources of
iron and its effect to the environment are also discussed in this chapter. The
different methods presented in this chapter for determining the presence of iron
chemistry is a major source of oxidative stress in media. The reaction involves the
ferrous iron catalyzed conversion of hydrogen peroxide into a hydroxide ion and a
hydroxyl free radical with the concurrent oxidation of ferrous iron to ferric iron.
creates the most damaging agent found in the cell culture system-the hydroxyl
free radical. This radical will react with virtually any molecule in the medium or
the cell. It is so reactive that it typically reacts very close to its site of formation. A
major action of hydroxyl free radicals in cell culture involves the initiation of lipid
peroxidation. Ferrous iron mediated lipid peroxidation can occur with iron in a
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 8
oxygen species (ROS), which include hydrogen peroxide (H2O2) and hydroxyl
radicals (·OH) occur in rain and surface waters [14, 15]. The generation of ·OH, via
presence of ferrous iron, H2O2 forms ·OH via the Fenton reaction:
·OH is far more reactive than H2O2 and its reaction with aqueous species is
Iron is vital for almost all living organisms by participating in a wide variety
transport. Disorders of iron metabolism are among the most common diseases of
common micronutrient insufficiency. One of the main reasons for the limited
the optimal epidemiologic approach for identifying it and for measuring its
severity. In a 1985 World Health Organization (WHO) report, it was estimated that
15% to 20% of the world's population had iron deficiency anemia. Current
highlighted the need for reliable epidemiologic methods to assess iron status.
[18].
reduced ferrous iron (Fe2+) and insoluble or oxidized ferric iron (Fe3+). In
oxygenated waters, ferrous iron (Fe2+) oxidizes to ferric iron (Fe3+) which is capable
parts per billion in ocean. It might likewise be discharged to water from natural
deposits, industrial wastes, and refining of ores. This discharge makes the level of
iron much higher and can cause harm in the marine ecosystem.
green plants. The oxygen bound with iron travels in the blood transporting the
carbon dioxide out. Green plant uses iron for nitrogen binding while
water restricts the amount of phytoplankton that can survive. However, when the
iron concentration is above 0.1 parts per million, this will potentially damage the
The toxicity of iron is dependent on the specie and the size of the fish.
Small amount of ferric iron particles can cause irritation of the gill which leads to
gill damage and secondary bacterial and fungal infections of fishes. Although the
extent of the damaged that can be caused by the physical presence of iron has
elements are elements present at low at low concentrations (mg kg−1 or less) in
agroecosystems. They are also called micronutrients and are crucial for plant
growth. Trace elements in an agroecosystem are either inherited from soil parent
materials or inputs through human activities. Heavy metals when repeatedly used
There are many iron sources in freshwater lakes. These may include: the
precipitation, and hydrology [22]. Due to its low solubility, iron concentration in
water is quite low despite being the second most abundant metal and fourth most
environmental problems occur. Thus, the structure and function of the different
bodies of water ecosystems are altered as a result of the variations of its iron
content [23].
2.6 Sensors
may define it as a device that receives a stimulus and responds with a beam of
light. [24]
modulated [24].
have a functional and good sensor. Properties like accuracy, resolution and
2.6.2.1 Accuracy
a sensor from the ideal or true value from its input. Accuracy simply
means the closeness of the value given by the sensor to the true
value. The higher its accuracy, the better the data acquired using
2.6.2.2 Resolution
2.6.2.3 Reliability
the liquid sample. It works by passing a beam of light through the sample inside
the cuvette and estimating the intensity of light reaching a detector. Correlation
2.8.1 Stability
time. The term stability has been defined as the stability of the samples
2.8.2 Repeatability
identical conditions. It is the ability of the sensor to give the same values in
the conditions is not changed. If the sensor’s output changes even if the
sample and other parts and properties of the setup was not changed, then
the said sensor has low repeatability. A sensor should ideally give the same
2.8.3 Sensitivity
voltammetry or aerometry have been offered for the determination of iron on the
detector. While the element-specific AAS detector determines the total iron
to Fe (III) [26].
Another method for the detection of iron, is with the use of 1, 10-
Phenanthroline. A solution containing both iron (II) and iron (III) will turned a
reddish-orange iron (II) complex and yellow iron (III) complex when 1, 10-
Phenanthroline is added. Both iron (II) and iron (III) complex had the identical
absorbance coefficients of 396 mµ. A study conducted by Harvey, Smart and Amis,
enables the determination of iron (II) and total iron in the same solution by
CHAPTER 3
METHODOLOGY
This chapter describes and defines each research method that was
followed through the course of the study. From the preparation of reagents,
instrumentation system that was used for data collection, the determination of
suitable wavelength, the stability, repeatability and sensitivity tests, as well as the
preparation and preliminary test for the real sample all were clearly discussed in
this chapter. Lastly, the researchers also discussed the methods that were used to
The type of research that was conducted was both qualitative and
quantitative since the researchers detected the amount of Iron by the optical
and the addition of thiocyanate for the solution to show varying coloration.
Concentrations of both the iron solutions and thiocyanate solution are known. The
optical density of the solutions was then identified using the said instrument by
allowing both the iron and thiocyanate solution to turn red. Thus, generating a
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 18
calibration curve that determined if the liquid samples contain any iron as
claimed.
water. The standard iron solutions used were in the form of ferric ammonium
concentrations of iron, 35, 30, 25, 20, 15 and 10 were derived from the stock
ppm solution. The indicator was prepared according to its solubility to water.
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 19
Spectrophotometer. The equipment used was shown in Figure 3.1. The sample
was placed inside a cuvette and was then introduced into UV-Vis
computer for processing. Data acquisition and off-line analysis were performed
using Microsoft Excel (Microsoft Corporation, USA). Spectra readings were taken
from the chosen range of wavelengths which was from 385 nm to 750 nm. All
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 20
done in two different conditions. In the first condition, the sample that was tested
was the 20 ppm iron solution. For the first condition, a cuvette was filled with 2mL
iron solution. Spectra readings were taken from 385 nm to 750 nm which is the
range for the visible region. The other condition was the addition of 150 μL of
obtained in the two conditions were plotted against the wavelength range
considered (385 to 750 nm). The two resulting spectra were compared to plot
their difference. The wavelength that gave the largest difference between the two
absorbance readings was the working wavelength used in the succeeding sensor
tests. The working wavelength was determined graphically and was represented
by the abscissa with the largest vertical difference occurred. All of the succeeding
wavelength.
Stability was determined by plotting the absorbance with time. The stability
of the signal at the chosen wavelength was tested by measuring the absorbance of
the ferric ammonium sulfate solution using the middle concentration of the
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 21
thiocyanate was added to 2000 μL of the iron solution. This was tested for 120
The stability was checked with the use of the relative standard deviation of
the absorbance readings during the 120 minute run of the sample. The relative
literatures.
The repeatability test is a measure of how consistently the method will show
the same thread The 20 ppm concentration of the diluted iron solution of was
was read for about 300 seconds. The absorbance of this sample was also read for
a response time of 300 seconds each run. This step was repeated 6 times and the
data collected from the UV-Win Lab Data Processor would be plotted using
Microsoft Excel.
run, making sure that what was done in the previous sample will also be done in
the next samples to be tested. The calculated RSD must be ≤ 10% for the proposed
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 22
calibration curve with delta absorbance (y-axis) vs. the standard iron solution
concentration (x-axis).
present in the standard solution was tested by getting the absorbance of the iron
solution before and after adding ammonium thiocyanate. The test was conducted
2000 μL of the standard iron solution was placed in the cuvette and
concentration of 3M was added to the iron solution and was read for another 5
minutes.
The same procedure was done to the other remaining concentrations of the
standard iron solution. Cuvettes were washed with ultra-pure water and with the
next solution to be used to make sure residues were removed from the surfaces of
the cuvettes.
After the absorbance measurement, the data were collected from the UV-
Winlab Data Processor and Viewer Software; it was then plotted using Microsoft
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 23
(3 ⋅ 𝑆𝑡𝑑𝑒𝑣 𝑜𝑓 𝑠𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦)
𝐿𝑂𝐷 =
(𝑠𝑙𝑜𝑝𝑒 𝑜𝑓 𝑠𝑒𝑛𝑠𝑖𝑡𝑖𝑣𝑖𝑡𝑦)
The commercial milk samples that were used in this research are Nestlé
Nestogen Two, Mead Johnson Enfamil Two A+, and Wyeth Bonamil. Table 1 shows
Nestogen 4.9610 30
Enfamil 4.4658 30
Bonamil 10.3501 60
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 24
Each milk sample was prepared according to the suggested amount of milk
times and was placed in the centrifuge for 10 minutes to improve its opacity.
2000 µL of the iron solution was placed in a cuvette and was placed inside
the UV-Vis for 5 minutes to get the absorbance reading. This solution was oxidized
the sample, 150 µL of ammonium thiocyanate was added to the solution and its
absorbance was read for another 5 minutes. This procedure was repeated 3 times
After the absorbance measurement, the data were collected from the UV-
Winlab Data Processor and Viewer Software; it was then plotted using Microsoft
Excel.
Chapter 3. Methodology
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 25
CHAPTER 4
RESULTS AND DISCUSSION
This chapter describes the result of the tests done in chapter 3. It starts
with the determination of suitable wavelength to work on during the tests. This
stability test is done based on the chosen concentration of iron and indicator.
test on its repeatability and finally its sensitivity and limit of detection.
formed. During the experiment, it was observed that the higher the concentration
of iron that will react with ammonium thiocyanate, the more intense the observed
color.
tests for the proposed method in determining iron was done with milk samples in
The results from the stability test were plotted as absorbance vs.
wavelength. Based on the graph shown in the Figure 4.0, the optimum difference
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 26
was found at the wavelength of 480 nm. This wavelength shows that if readings
were done here, there is a wider scale to be able to read the absorbance with
better accuracy. Therefore, the working wavelength that was used in the
1 Iron
0.8 Iron + SCN
0.6 SCN
0.4
Water
0.2
0
385 485 585 685
Wavelength
Stability test was done for 20 ppm because it is the value close to the
amount of iron in milk, our target sample. This test was done to prove the ability
of the sensor to give a constant absorption reading. This was also done to identify
The concentrations above 20ppm were not used for they contain more
iron. Higher iron content may cause precipitation, thus no stable reading can be
done. Lower concentrations may also be used for the stability test and a stable
ammonium sulfate stock solution and ammonium thiocyanate solution, but for
Figure 4.1.1 shows the result for stability test which generated the lowest
RSD. The stability test (Figure 4.1.1) for the standard iron solution using 20 ppm
The absorbance reading of the solution that was tested using the
spectrophotometer showed a constant reading in the 120 minute run. Its relative
standard deviation was measured to be 6.56%. The range for acceptable values for
this test should be less than 10% which is the limit set for the method to be
considered stable. The value was within the range and therefore acceptable [28].
Absorbance vs time
0.3
0.2
Abs
0.1
0
0 1000 2000 3000 4000 5000 6000 7000 8000
Time (s)
proposed. This was done with six runs giving an RSD of 8.12%. The value was
0.45
0.4
0.35
Abs
0.3
0.25
0.2
0.15
0 1000 2000 3000 4000
Time (s)
the different iron concentration (Figure 4.3.1). From this result, the changes in the
absorbance were plotted versus the concentration of iron. The curves produced
from the average of the trials have fairly linear characteristics. The figure shows
0.08
0.06
0.04
0.02
0
0 5 10 15 20 25 30 35 40
Concentration
Figure 4.3.1 Calibration Curve for ∆Abs Range: 0.0106 to 0.1235 (n=3)
The square of correlation coefficient and the equation of the curve for
each calibration were calculated using Microsoft Excel. Linearity of the calibration
curve generated is proven by the value of R2 which is 0.9274. It is the best possible
correlation coefficient for linear equations. This means that there is a linear
complex in the different ferric ion standard solutions. A summary of the features
R\% RSD
The calibration curve generated from the three trials performed was
applied with the use of error bars. Error bars in general, indicates estimated errors
Figure 4.3.2 Calibration Curve for ∆Abs Range: 0.0106 to 0.1235 with error bars
Generally, shorter error bars implies that the values are more reliable in
which are increasing as the concentration increases show that there may not be
recommended.
can be reliably detected, but not necessarily quantified, under the stated
conditions of the test. This was calculated using the standard deviation of the
𝑠𝑡𝑑𝑒𝑣(𝑠𝑡𝑎𝑏𝑖𝑙𝑖𝑡𝑦) × 3
𝐿𝑂𝐷 =
𝑚(𝑠𝑙𝑜𝑝𝑒 𝑜𝑓 𝑐𝑎𝑙𝑖𝑏𝑟𝑎𝑡𝑖𝑜𝑛 𝑐𝑢𝑟𝑣𝑒)
0.004959121 × 3
𝐿𝑂𝐷 = = 3.38 𝑀/∆𝑎𝑏𝑠
0.0044
A value of 3.38 was obtained through the calculations. This value means
standard iron solution. Visual observations (Figure 4.3.3) were done. The color
The effectiveness of the sensor was tested by analyzing the iron content of
different milk samples. Figures 4.4.1, 4.4.2, and 4.4.3 shows the absorbance vs
The graph shows the response of the milk samples when ammonium thiocyanate
the expected data. There are several factors that may cause the opposite change
in the trend of the graph for the milk samples as compared to the calibration data.
2.75
Absorbance
2.7
2.65
2.6
0 300 600 900 1200 1500 1800 2100
Time/s
2.7
Absorbance
2.65
2.6
2.55
0 300 600 900 1200 1500 1800 2100
Time/s
2.75
Absorbance
2.7
2.65
2.6
2.55
0 500 1000 1500 2000
Time/s
First the opacity of the milk may be diminished on the formation of the
the casein of milk was tried to lessen the turbidity of milk. The use of a finer
remained. We recommend further studies on this process. Aside from casein there
may be other nutrients existing in the milk sample that reacts with the indicator.
In summary, the iron detection method as a possible sensor was done with
CHAPTER 5
CONCLUSION AND RECOMMENDATIONS
5.1 Conclusion
iron in liquid samples based on its reaction with ammonium thiocyanate was
The iron standards were dissolved in water and were first oxidized to the
Fe (III) form. This was reacted with ammonium thiocyanate resulting in a red
blood coloration which gives significant absorbance readings even in the most
dilute concentrations. The suitable wavelength for reading the absorbance was
The stability test was done at the chosen wavelength using the 20 ppm
iron solution. The calculated RSD value for this test was 6.56%. Repeatability test
using the difference in absorbance between a blank and an iron solution standard
The test for sensitivity as proven by the generated calibration curve has an
The limit of detection, LOD for the calibration curve was calculated to be:
0.004959121 × 3
𝐿𝑂𝐷 = = 3.38 𝑝𝑝𝑚/∆𝐴𝑏𝑠
0.0044
With the advantages presented in this paper, the method proposed could
be suitable for practical uses for the determination of iron present in samples.
After all the necessary tests and experiments, the study has come to
conclude that the method for the iron concentration determination is effective.
5.2 Recommendations
Initial steps were done to test the method on real samples such as
powdered milk. Due to time constraints, the actual optical sensor for the purpose
tobacco smoke. Risk assessment and harm reduction is a field where thiocyanate
REFERENCES
[1] Gupta, C.P. (2014). Role of iron (Fe) in body. IOSR Journal of Applied Chemistry,
7(11), 38-46
[2] Vallejos, S., Muñoz, A., Ibeas, S., Serna, F., Garcia, F. C., Garcia, J. M. (2013).
Detecting iron the smart way. Journal of Materials Chemistry A, 48(1), 15435-
[3] Ekici, K.; Agaoglu, S.; Isleyici, O.; Indian Vet. J. 2004, 81, 1284
[4] Mico, C.; Peris, M.; Sanchez, J.; Recatala, L.; Span. J. Agric. Res. 2006, 4, 363.
[6] Niederau, C.; Fischer, R.; Purschel, A.; Stremmel, W.; Haussinger,
[7] Bispo, M. S.; Morte, E. S. B.; Korn, M. G. A.; Teixeira, L. S. G.; Korn, M.; Costa, A.
[9] Ferreira, S. L. C.; Queiroz, A. S.; Melo, A. S. Q.; Assis, J. C. R.; Korn, M. G. A.;
[10] Cooke, K. Why total iron is important. Kentucky Water Watch. Retrieved from
http://www.state.ky.us/nrepc/water/wwhomepg.htm
[12] Pan, J., Cha, T., Chen, H., & Choi, J. (2013). Carbon nanotube-based optical
doi:10.18130/v3hk0t
[14] Willey JD, Kieber RJ, Lancaster RD, Atmos J: Chem. 1996, 25: 149.
[15] Faust BC, Allen JM: Environ Sci Technol. 1993, 27: 1221-
10.1021/es00043a024.
10.1146/annurev.ph.48.030186.003301.
[17] Lieu, P. T., Heiskala, M., Peterson, P. A., & Yang, Y. (2001). The roles of iron in
doi:10.1016/s0098-2997(00)00006-6
3359-3363. doi:10.1182/blood-2002-10-3071
[19] Cadmus, P., Brinkman, S., May, M. (2018).Chronic toxicity of ferric iron for
[20] Rostern NT. (2017). The effects of some metals in acidified water on aquatic
10.19080/OFOAJ.2017.04.555645.
[21] He, Z. L., Yang, X. E., & Stoffella, P. J. (2005). Trace elements in
[22] (PDF) Iron biogeochemistry and its environmental impacts ... (n.d.). Retrieved
from
https://www.researchgate.net/publication/236267914_Iron_biogeochemistry_an
d_its_environmental_impacts_in_freshwater_lakes
[23] Harris, J.E. (1992) Weathering of rock, corrosion of stone and rusting of iron.
[26] Weber, G. (1991). Speciation of iron using HPLC with electrochemical and
doi:10.1007/bf00324473
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 42
http://fisbio.biof.ufrj.br/restrito/bioEstatistica/90_top_especiais/errorbars_stat_si
APPENDICES
APPENDIX A
11 g FeNH4 (SO4 )2
mol 1 mol Fe 55.85 g 1000mg 1
∙ 12H2 O x x x x x
482.25 g 1 mol FeNH4 (SO4 )2 ∙ 12H2 O mol 1g 1L
= 1273.9243 ppm
1.3651 𝑔 1000 𝑚𝑔
𝑥 = 2730.2 𝑝𝑝𝑚
500 𝑚𝐿 𝐻2 𝑂 1𝑔
Volume of 2730.2 ppm NH4SCN solution to prepare 500 ppm solution = 91.57 mL
M 1V 1 = M 2V 2
For 35 ppm:
1273.9243(𝑉) = 35(500)
V = 13.7371 mL
For 30 ppm:
35(𝑉) = 30(500)
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 44
V = 428.5714 mL
For 25 ppm:
30(𝑉) = 25(500)
V = 416.6667 mL
For 20 ppm:
25(𝑉) = 20(500)
V = 400 mL
For 15 ppm:
20(𝑉) = 15(500)
V = 375 mL
For 10 ppm:
15(𝑉) = 10(500)
V = 333.3333 mL
2 Nestogen
8.6 𝑚𝑔 𝐹𝑒 𝑥
= → 𝑥 = 0.8901𝑚𝑔 𝐹𝑒
100 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟 10.3501 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟
3 Bonamil
6.8 𝑚𝑔 𝐹𝑒 𝑥
= → 𝑥 = 0.3037𝑚𝑔 𝐹𝑒
100 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟 4.4658 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟
UNIVERSITY OF SANTO TOMAS CHEMICAL ENGINEERING 45
4 Enfamil
8.4 𝑚𝑔 𝐹𝑒 𝑥
= → 𝑥 = 0.4167𝑚𝑔 𝐹𝑒
100 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟 4.9610 𝑔 𝑃𝑜𝑤𝑑𝑒𝑟
APPENDIX B