LABORATORIUM KALIBRASI METROLOGI PT.

DIRGANTARA INDONESIA ( IAe )


JL. Pajajaran No : 154 - Bandung 40174 - INDONESIA. Telp : 62-022 - 6054233

CALIBRATION REPORT
Calibration date : 21-11-2019 || Due Date : 21-11-2021
[ Sheet 1 of 2 ] -- No : 1940-001/MET/05/11/2019

Instrument : Steel Rule Resolution : 0.5 mm


Manufacture : Mitutoyo - Japan Nomet : 182-305
Range : 0 - 300 mm User : Laboratorium Uji
Serial Number : 182-305 Location : QA3100
Type : Stainless Tempered Date Received : 20-11-2019
Cal. Location : Lab Dimensi
Room Temp. : 19 °C Humidity : 50.4 %
Barometer : N/A Gravitation : N/A

CALIBRATION RESULT OF Uncertainty at


DEVIATION
POINT CALIBRATION K=2 and CL=95%
mm mm mm mm
0 0.000 0.000 0.29
5 5.000 0.015 0.29
10 10.000 0.016 0.29
15 15.000 0.018 0.29
20 20.000 0.020 0.29
25 25.000 0.020 0.29
30 30.000 0.020 0.29
35 35.013 0.033 0.29
40 40.013 0.033 0.29
45 45.020 0.040 0.29
50 50.020 0.040 0.29
55 55.020 0.040 0.29
60 60.027 0.047 0.29
70 70.033 0.053 0.29
80 80.033 0.053 0.29
90 90.027 0.047 0.29
100 100.033 0.053 0.29
120 120.033 0.053 0.29
150 150.033 0.054 0.29
200 200.033 0.054 0.29
250 250.033 0.054 0.29
300 300.033 0.054 0.29

Instrument Used For Calibration


LF Standard Rule 073159-Traceable to International Standard by PTDI Laboratory-with Calibration Certificate Number :
860-03/MET/05/12/17
Specification Used for Calibration
IK-D029 (Revisi 0) Tolerance for this tool = 0.115 mm

Conclusion

PASS X FAIL
This Equipment is still "In Tolerance"
Before and after Calibration are same

Calibrated & Checked By Checked By


Hamdani Ridwan Darmawan

The report shall not be reproduced except in full without approval of the laboratory can provide assurance that parts of a re port are not taken out of context.

F-DP708.01-01 Error Based


[ Sheet 2 of 2 ] -- No : 1940-001/MET/05/11/2019 | Calibration date : 21-11-2019 || Due Date : 21-11-2021

REKAMAN DATA MENTAH KALIBRASI


Tanggal terima : 20-11-2019 Pembuat : Mitutoyo - Japan
Nama Perusahaan : Laboratorium Uji Type / Model : Stainless Tempered
Alamat : QA3100 Nomor Seri : 182-305
Nama Alat : Steel Rule Kapastas : 0 - 300 mm
Prosedur Kalibrasi : IK-D029 (Revisi 0) Resolusi : 0.5 mm
Temperatur Ruangan : 19 °C Kelembaban : 50.4 %
Nomor Metrologi : 182-305

Penunjukan Alat Penyimpangan Alat ( mm )


No
( mm ) 1 2 3
1 0 0.00 0.00 0.00
2 5 0.00 0.00 0.00
3 10 0.00 0.00 0.00
4 15 0.00 0.00 0.00
5 20 0.00 0.00 0.00
6 25 0.00 0.00 0.00
7 30 0.00 0.00 0.00
8 35 0.02 0.02 0.00
9 40 0.02 0.00 0.02
10 45 0.02 0.02 0.02
11 50 0.02 0.02 0.02
12 55 0.02 0.02 0.02
13 60 0.04 0.02 0.02
14 70 0.02 0.04 0.04
15 80 0.04 0.02 0.04
16 90 0.04 0.02 0.02
17 100 0.04 0.02 0.04
18 120 0.02 0.04 0.04
19 150 0.04 0.04 0.02
20 200 0.04 0.02 0.04
21 250 0.02 0.04 0.04
22 300 0.04 0.04 0.02

Kalibrator Yang Dipergunakan :

LF Standard Rule 073159-Traceable to International Standard by PTDI Laboratory-with Calibration Certificate Number : 860-03/MET/05/12/17

Dikalibrasi Oleh :
Nama Hamdani
Paraf
FR.16-01