LABKIT MINDRAY BS 120 – ACP

Preparation of working reagent:


Dissolve 1 tablet of R2 with 2 ml. of R1. The working reagent is stable 48 hours at 2-8º C. or 6 hours at
room temperature.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST ACP R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME ACP SAMPLE Vol. (µl) 30
STANDARD No R1 BLANK *
REACTION TYPE KINETIC MIX R BLANK *
PRIM. WAWE 405 LINEARITY RANGE 0.0 U/L. 150.0 U/L.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 16 26 FACTOR 750
INCUBATION TIME PROZONE CHECK *
UNITS U/L. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE *
SENSITIVITY 0
REPLICATES 1
INTERVAL (DAYS)
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 150 IU/L. of Acid phosphatase.

LABKIT
Preparation of working reagent: MINDRAY BS 120 – ALBUMIN
The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST ALB R1 Vol. (µl) 397
No ** R2 Vol. (µl) *
FULL NAME ALBUMIN SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 630 LINEARITY RANGE 0.0 mg/dL. 6.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 0
REPLICATES 1
INTERVAL (DAYS)
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 6 mg/dL. of Albumin.

LABKIT MINDRAY BS 120 – UREA UV LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.
Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST UREA R1 Vol. (µl)

240
No ** R2 Vol. (µl) 60
FULL NAME UREA UV SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE FIXED TIME MIX R BLANK *
PRIM. WAWE 340 LINEARITY RANGE 5.0 mg/dL. 250.0 mg/dL
SEC. WAWE LINEARITY LIMIT *
DIRECTION DECREASE SUBSTRATE LIMIT *
REACTION TIME 2 5 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 250 mg/dL. of Urea.

LABKIT MINDRAY BS 120 – TRIGLYCERIDES LS

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight.
Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST TRIG R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME TRIGLYCERIDES SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 1 mg/dL. 1000 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 18 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS)
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 1000 mg/dL. of Triglycerides.

LABKIT MINDRAY BS 120 – TOTAL PROTEIN

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight and contamination.

Calibration:
Use calibrator included in the kit.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST PT R1 Vol. (µl) 280
No ** R2 Vol. (µl) *
FULL NAME TOTAL PROTEIN SAMPLE Vol. (µl) 7
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 546 LINEARITY RANGE 1.00 g/dL. 15.00 g/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS g/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 15 g/dL. of total proteins.

LABKIT MINDRAY BS 120 – PROT. IN URINE&CFS

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight and contamination.

Calibration:
Use calibrator included in the kit.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST PTUR R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME PROT. URINE SAMPLE Vol. (µl) 10
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 578 LINEARITY RANGE 100 mg/L. 4000 mg/L.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/L. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 4000 mg/L. of Proteins.

LABKIT MINDRAY BS 120 – PHOSPHORUS UV

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight and contamination.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.
PARAMETERS
TEST PHOS R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME PHOSPHORUS SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 340 LINEARITY RANGE 0.3 mg/dL. 15.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 17 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 15 mg/dL. of Phosphorus.

LABKIT MINDRAY BS 120 – LIPASE LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
CALIBRATION WITH STANDARD: Reconstitute the standard included in the kit with 1 ml. of distilled
water. Mix gently until complete solution. Stability: 10 days at 2-8º C. Aliquote the standard solution in
small vials and freeze. Stability: 3 months at -20º C.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST LIP R1 Vol. (µl) 300
No ** R2 Vol. (µl) 60
FULL NAME LIPASE SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE KINETIC MIX R BLANK *
PRIM. WAWE 578 LINEARITY RANGE 1.0 U/L. 250.0 U/L.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 4 11 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS U/L. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 250 IU/L. of Lipase.

LABKIT MINDRAY BS 120 – LDLc DIRECT

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label. Once opened is stable 8
weeks at 2-8º C.

Calibration:
Use serum LDL calibrator included with the kit. Reconstitute the standard included in the kit with 1 ml. of
distilled water. Mix gently until complete solution. Stability: 1 week at 2-8º C or 5 weeks at -20º C.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST LDL R1 Vol. (µl) 225
No ** R2 Vol. (µl) 75
FULL NAME STANDARD No LDL CHOLESTEROL SAMPLE Vol.
(µl) R1 BLANK 3
*
REACTION TYPE FIXED TIME MIX R BLANK *
PRIM. WAWE 578 LINEARITY RANGE 1.0 mg/dL. 1000.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 2 18 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 1000 mg/dL. of LDL-Cholesterol.

LABKIT MINDRAY BS 120 – LDH LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST LDH R1 Vol. (µl) 240
No S A RECTION U PRECISION
FULL NAME E W REACTION N
STANDARD No C E TIME I CALIBRATION
REACTION TYPE . D INCUBATION T RULE
PRIM. WAWE W I TIME S SENSITIVITY
REPLICATES ** R2 Vol. (µl) 60
INTERVAL (DAYS) LDH SAMPLE Vol. (µl) 5
DIFFERENCE LIMIT R1 BLANK *
SD KINETIC MIX R BLANK *
BLANK RESPONSE 340 LINEARITY RANGE 5 U/L. 1450 U/L.
ERROR LIMIT LINEARITY LIMIT *
CORREALTION COEF. DECREASE SUBSTRATE LIMIT *
4 14 FACTOR 9690
Linearity: PROZONE CHECK *
U/L. q1 q2
INTEGER q3 q4
PC Abs

*
0
1

0
The method is linear until 1450 IU/L. of LDH.

LABKIT MINDRAY BS 120 – IRON-FZ

Preparation of working reagent:


Working reagent (R.1): Mix one vial R.1 with one vial of R2. Stability: 3 months in the refrigerator (2-8º C).
R3 is ready to be used and must be used as R2 in the instrument.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST PRIM. WAWE Fe 578
No SEC. WAWE **
FULL NAME DIRECTION IRON FZ INCREASE
STANDARD No
REACTION TYPE END POINT
R1 Vol. (µl) MIX R BLANK 425 *
R2 Vol. (µl) LINEARITY RANGE 30 10 µg/dL. 1000 µg/dL.
SAMPLE Vol. (µl) LINEARITY LIMIT 45 *
R1 BLANK SUBSTRATE LIMIT * *
REACTION TIME 1 17 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS µg/dL. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 2
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 1000 µg/dL.. of Iron.

LABKIT MINDRAY BS 120 – HDLc DIRECT

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label. Once opened is stable 8
weeks at 2-8º C.

Calibration:
Use serum HDL calibrator included with the kit. Reconstitute the standard included in the kit with 1 ml. of
distilled water. Mix gently until complete solution. Stability: 1 week at 2-8º C or 5 weeks at -20º C.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST HDL R1 Vol. (µl) 225
No ** R2 Vol. (µl) 75
FULL NAME HDL CHOLESTEROL SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE SEC. WAWE FIXED TIME
PRIM. WAWE DIRECTION 578 INCREASE
MIX R BLANK LINEARITY LIMIT * *
LINEARITY RANGE SUBSTRATE LIMIT 0.0 mg/dL. 250.0 mg/dL.*
REACTION TIME 2 18 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 1
REPLICATES 1
INTERVAL (DAYS) 0
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 250 mg/dL. of HDL-Cholesterol.

LABKIT MINDRAY BS 120 – GPT/ALT LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST GPT R1 Vol. (µl) 240
No U I S PONSE
FULL NAME N S LIBRATION D ERROR LIMIT
STANDARD No I I RULE B CORREALTION
REACTION TYPE T O SENSITIVITY L COEF.
PRIM. WAWE S N REPLICATES A
SEC. WAWE P INTERVAL N Linearity:
C (DAYS) K
DIRECTION R
REACTION TIME E DIFFERENCE R
C A LIMIT E
INCUBATION TIME
S
** R2 Vol. (µl) 60
GPT ONE POINT SAMPLE Vol. (µl) 30
0 R1 BLANK *
KINETIC 1 MIX R BLANK *
340 LINEARITY RANGE 5 U/L. 500 U/L.
LINEARITY LIMIT *
DECREASE SUBSTRATE LIMIT *
3 10 FACTOR 1750
PROZONE CHECK *
U/L. 0 q1 q2
INTEGER q3 q4
PC Abs
The method is linear until 500 IU/L. of GPT/ALT.

LABKIT MINDRAY BS 120 – GOT/AST LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST GOT R1 Vol. (µl) 240
No S I (DAYS) ** 0
FULL NAME E C DIFFERENCE GOT 1
STANDARD No N A LIMIT
REACTION TYPE S T SD KINETIC
PRIM. WAWE I E BLANK 340
SEC. WAWE T S RESPONSE
DIRECTION I I ERROR DECREASE
REACTION TIME V N LIMIT 3 0
INCUBATION TIME I T CORREALTIO
UNITS T E N COEF. U/L.
Y R
PRECISION INTEGER
R V Linearity:
CALIBRATION E A
RULE P L ONE POINT
L
R2 Vol. (µl) LIMIT 60
SAMPLE SUBSTRATE 30
Vol. (µl) LIMIT * Abs
R1 BLANK FACTOR *
MIX R PROZONE 5 U/L. 500 U/L.
BLANK CHECK *
LINEARITY q1 q2 *
10 RANGE q3 q4 1750
LINEARITY PC *
The method is linear until 500 IU/L. of GOT/AST.

LABKIT MINDRAY BS 120 – GLUCOSE LS

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST GLU R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME GLUCOSE SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 0 mg/dL. 500 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE INTERVAL (DAYS)
SENSITIVITY DIFFERENCE LIMIT
REPLICATES SD
BLANK RESPONSE ONE POINT LINEAR
ERROR LIMIT 1
CORREALTION COEF. 1
0
Linearity:

0
The method is linear until 500 mg/dL. of Glucose.

LABKIT MINDRAY BS 120 – γGT LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST GGT R1 Vol. (µl) 240
No B M inearity: **
FULL NAME L I GGT
STANDARD No A T 0
REACTION TYPE N C KINETIC
PRIM. WAWE K O 405
SEC. WAWE R R
DIRECTION E R INCREASE
REACTION TIME S E 3
INCUBATION TIME P A
UNITS O L U/L.
N T
PRECISION INTEGER
S I
CALIBRATION E O
E N
RULE ONE POINT
R C
SENSITIVITY 0
R O
REPLICATES 1
O E
INTERVAL (DAYS)
R F.
DIFFERENCE LIMIT
SD L
I L
R2 Vol. (µl) LIMIT 60
SAMPLE SUBSTRATE 30
Vol. (µl) LIMIT * Abs
R1 BLANK FACTOR *
MIX R PROZONE 2 U/L. 250 U/L.
BLANK CHECK *
LINEARITY q1 q2 *
10 RANGE q3 q4 1190
LINEARITY PC *
The method is linear until 250 IU/L. of Gamma-GT.

LABKIT MINDRAY BS 120 – CHOLESTEROL LS

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST CHOL R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME CHOLESTEROL SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 0 mg/dL. 600 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE INTERVAL (DAYS)
SENSITIVITY DIFFERENCE LIMIT
REPLICATES SD
BLANK RESPONSE ONE POINT LINEAR
ERROR LIMIT 1
CORREALTION COEF. 1

Linearity:

0
The method is linear until 600 mg/dL. of Cholesterol.

LABKIT MINDRAY BS 120 – CHLORIDE

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight and contamination.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST Cl R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME CHLORIDE SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 0 mmol/L. 130 mmol/L.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 17 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS Mmol/L. q1 q2
PRECISION INTEGER q3 q4
PC Abs
CALIBRATION
RULE SD
SENSITIVITY BLANK RESPONSE
REPLICATES ERROR LIMIT
INTERVAL (DAYS) CORREALTION COEF.
DIFFERENCE LIMIT
ONE POINT LINEAR
Linearity: 1
1
0

0
The method is linear until 130 mmol/L. of Chloride.

LABKIT MINDRAY BS 120 – CREATININE-J

Preparation of working reagent:


The reagents are ready to be used and stable at 2-8º C up to the date of expiration specified

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST CREA R1 Vol. (µl) 180
No ** R2 Vol. (µl) 180
FULL NAME CREATININE SAMPLE Vol. (µl) 36
STANDARD No R1 BLANK *
REACTION TYPE FIXED TIME MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 0.20 mg/dL. 15.00 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 2 6 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.01 q3 q4
PC Abs
CALIBRATION
RULE BLANK RESPONSE
SENSITIVITY ERROR LIMIT
REPLICATES CORREALTION COEF.
INTERVAL (DAYS)
DIFFERENCE LIMIT Linearity:
SD
ONE POINT LINEAR
1
1

0
The method is linear until 15 mg/dL. of Creatinin.

LABKIT MINDRAY BS 120 – CK-NAC LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST CK R1 Vol. (µl) 240
No C i**
FULL NAME O tCK-NAC 0
STANDARD No R y
REACTION TYPE R :KINETIC
PRIM. WAWE E 340
SEC. WAWE A
DIRECTION L INCREASE
REACTION TIME T 6 16
INCUBATION TIME I
UNITS O U/L.
N
PRECISION INTEGER
C
CALIBRATION O
E
RULE F. *
SENSITIVITY 0
REPLICATES 1
L
INTERVAL (DAYS)
i
DIFFERENCE LIMIT
n
SD
e
BLANK RESPONSE
a
ERROR LIMIT
r
R2 Vol. (µl) SUBSTRAT q3 q4 60 *
SAMPLE Vol. (µl) E LIMIT PC 5 8095
R1 BLANK FACTOR * *
MIX R BLANK PROZONE *
LINEARITY RANGE CHECK 10 U/L. 2000 U/L.
LINEARITY LIMIT q1 q2 * Abs
The method is linear until 2000 IU/L. of CK-NAC.

LABKIT MINDRAY BS 120 – CK-MB LQ

Preparation of working reagent:


Mix 1 volume of R2 with 4 volumes of R1 (Use this WR as R1 in the instrument). This working reagent is
stable for 15 days at +2+8 C. or 24 hours at room temperature (15-25º C.)

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST E C CK-MB
No R O y: ** 0
FULL NAME R E CK-MB
STANDARD No O F.
REACTION TYPE R KINETIC
PRIM. WAWE L L 340
SEC. WAWE I
DIRECTION M i INCREASE
I
REACTION TIME 18
T n
INCUBATION TIME
C U/L.
UNITS
O e
PRECISION INTEGER
R
CALIBRATION R a
RULE E *
SENSITIVITY A r 0
REPLICATES L 1
INTERVAL (DAYS) T i
DIFFERENCE LIMIT I
SD O t
BLANK RESPONSE N
R1 Vol. (µl) LINEARITY 300 *
R2 Vol. (µl) LIMIT
SAMPLE SUBSTRATE 12
LIMIT * Abs
Vol. (µl)
R1 BLANK FACTOR *
MIX R PROZONE 2 U/L. 1000 U/L.
BLANK CHECK *
LINEARITY q1 q2 *
34 RANGE q3 q4 8255
PC
The method is linear until 1000 IU/L. of CK-MB.

LABKIT MINDRAY BS 120 – CALCIUM A-III

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight and contamination.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST CA R1 Vol. (µl) 300
No ** R2 Vol. (µl) *
FULL NAME CALCIUM SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 630 LINEARITY RANGE 2 mg/dL. 30.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 10 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE INTERVAL (DAYS)
SENSITIVITY DIFFERENCE LIMIT
REPLICATES SD
BLANK RESPONSE ONE POINT LINEAR
ERROR LIMIT 0
CORREALTION COEF. 2

Linearity:

0
The method is linear until 30 mg/dL. of Calcium.

LABKIT MINDRAY BS 120 – BILIRUBIN D - DMSO

Preparation of working reagent:


Ref. 30155, 30156 (Total Bilirubin) and 30150, 30151 (Direct Bilirrubin): Prepare the necessary amount of
working reagent using the following proportion: 30 vol. of R.1 + 1 vol. of R.2. The WR is stable 7 days at
2-8º C. or 24 hours at room temperature.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST BIL D R1 Vol. (µl) 250
No ** R2 Vol. (µl) *
FULL NAME BILIRUBIN SAMPLE Vol. (µl) 3
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 630 LINEARITY RANGE 0.1 mg/dL. 15.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE SD
SENSITIVITY BLANK RESPONSE
REPLICATES ERROR LIMIT
INTERVAL (DAYS) CORREALTION COEF.
DIFFERENCE LIMIT
Linearity: ONE POINT LINEAR
0
1

0
The method is linear until 15 mg/dL. of
Bilirubin.
MINDRAY BS 120 – BILIRUBIN T - DMSO

LABKIT

Preparation of working reagent:


Ref. 30155, 30156 (Total Bilirubin) and 30150, 30151 (Direct Bilirrubin): Prepare the necessary amount of
working reagent using the following proportion: 30 vol. of R.1 + 1 vol. of R.2. The WR is stable 7 days at
2-8º C. or 24 hours at room temperature.
Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST BIL T R1 Vol. (µl) 250
No ** R2 Vol. (µl) *
FULL NAME BILIRUBIN SAMPLE Vol. (µl) 25
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 578 LINEARITY RANGE 0.1 mg/dL. 15.0 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 33 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.1 q3 q4
PC Abs
CALIBRATION
RULE ONE POINT LINEAR
SENSITIVITY 0
REPLICATES 1
INTERVAL (DAYS)
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ERROR LIMIT
CORREALTION COEF. 0

Linearity:
The method is linear until 15 mg/dL. of Bilirubin

LABKIT MINDRAY BS 120 – AMYLASE LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST AMY R1 Vol. (µl) 300
No C r**
FULL NAME O iAMYLASE 0
STANDARD No R t
REACTION TYPE R yKINETIC
PRIM. WAWE E :405
SEC. WAWE A
DIRECTION L INCREASE
REACTION TIME T 3 10
INCUBATION TIME I
UNITS O U/L.
N
PRECISION INTEGER
C
CALIBRATION O
E
RULE F. *
SENSITIVITY 0
REPLICATES 1
L
INTERVAL (DAYS)
i
DIFFERENCE LIMIT
n
SD
e
BLANK RESPONSE
a
ERROR LIMIT
R2 Vol. (µl) SUBSTRAT q3 q4 * *
SAMPLE Vol. (µl) E LIMIT PC 6 3954
R1 BLANK FACTOR * *
MIX R BLANK PROZONE *
LINEARITY RANGE CHECK 5 U/L. 2000 U/L.
LINEARITY LIMIT q1 q2 * Abs
The method is linear until 2000 IU/L. of Amylase.

LABKIT MINDRAY BS 120 – ALP LQ

Preparation of working reagent:


The reagents are ready to use and keep stable until date indicated on the label.

Calibration:
Calibration in front of a factor.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST ALP. R1 Vol. (µl) 240
No E N **
FULL NAME R C ALP 0
STANDARD No R O
REACTION TYPE O E KINETIC
PRIM. WAWE R F. 405
SEC. WAWE L
DIRECTION I LINCREASE
REACTION TIME M i2 10
INCUBATION TIME I n
UNITS T eU/L.
PRECISION C aINTEGER
O r
CALIBRATION R i
RULE R t*
SENSITIVITY E y0
REPLICATES A :1
INTERVAL (DAYS) L
DIFFERENCE LIMIT T
SD I
BLANK RESPONSE O
R2 Vol. (µl) SUBSTRAT q3 q4 60 *
SAMPLE Vol. (µl) E LIMIT PC 5 3300
R1 BLANK FACTOR * *
MIX R BLANK PROZONE *
LINEARITY RANGE CHECK 4 U/L. 825 U/L.
LINEARITY LIMIT q1 q2 * Abs
The method is linear until 825 IU/L. of Alkaline phosphatase.

LABKIT MINDRAY BS 120 – URIC ACID LS

Preparation of working reagent:


The reagent is ready to be used. This solution is stable at 2-8º C up to the date of expiration specified.
Avoid direct sunlight.

Calibration:
Use LABTROL serum calibrator.

Operation conditions:
Introduce in Methods the following data.

PARAMETERS
TEST URI R1 Vol. (µl) 360
No ** R2 Vol. (µl) *
FULL NAME URIC ACID SAMPLE Vol. (µl) 9
STANDARD No R1 BLANK *
REACTION TYPE ENDPOINT MIX R BLANK *
PRIM. WAWE 510 LINEARITY RANGE 0.03 mg/dL. 25.00 mg/dL.
SEC. WAWE LINEARITY LIMIT *
DIRECTION INCREASE SUBSTRATE LIMIT *
REACTION TIME 1 17 FACTOR *
INCUBATION TIME PROZONE CHECK *
UNITS mg/dL. q1 q2
PRECISION 0.01 q3 q4
PC Abs
CALIBRATION
RULE ERROR LIMIT
SENSITIVITY CORREALTION COEF.
REPLICATES
INTERVAL (DAYS) Linearity:
DIFFERENCE LIMIT
SD
BLANK RESPONSE
ONE POINT LINEAR
1
2

0
The method is linear until 25 mg/dL. of Uric acid.