Anda di halaman 1dari 176

Introduction to

Capillary Gas Chromatography


Course Number H2615A

Student Manual
Introduction to
Capillary Gas Chromatography
H2615A

Student Manual

Manual Part Number H2615-90000


Printed in the USA Jan, 2001
Notice
The information contained in this document is subject to change without notice.
Agilent Technologies makes no warranty of any kind with regard to this material,
including but not limited to the implied warranties of merchantability and fitness
for a particular purpose.
Agilent Technologies shall not be liable for errors contained herein or for
incidental, or consequential damages in connection with the furnishing,
performance, or use of this material.
No part of this document may be photocopied or reproduced, or translated to
another program language without the prior written consent of Agilent
Technologies, Inc.

Agilent Technologies, Inc


11575 Great Oaks Way
Suite 100, MS 304B
Alpharetta, GA 30319

 2000 by Agilent Technologies, Inc.


All rights reserved
Printed in the United States of America

ii
Table Of Contents
INTRODUCTION TO CAPILLARY GC ....................................................................................1
INTRODUCTION .............................................................................................................................2
WHAT IS CAPILLARY GAS CHROMATOGRAPHY ............................................................................7
SAMPLE PREPARATION CONSIDERATIONS...................................................................................11
COLUMN CONSIDERATIONS ........................................................................................................12
BASIC COLUMN RESOLUTION..............................................................................................22
RESOLUTION EQUATION .............................................................................................................23
FLOW RATE/LINEAR VELOCITY AND RESOLUTION .....................................................................25
EFFICIENCY .................................................................................................................................29
TEMPERATURE AND RESOLUTION ...............................................................................................37
LIQUID PHASE AND RESOLUTION ................................................................................................42
SUMMARY...................................................................................................................................50
SAMPLE INTRODUCTION CONSIDERATIONS..................................................................53
TYPES OF INLET SYSTEMS ...........................................................................................................54
SPLIT/SPLITLESS INLET ...............................................................................................................57
ELECTRONIC PNEUMATIC FLOW CONTROL .................................................................................66
FLOW VS. TEMPERATURE ...........................................................................................................69
SPLIT/SPLITLESS INLET DESIGN ..................................................................................................74
INJECTION MODES ......................................................................................................................77
LINERS ........................................................................................................................................85
DETECTOR PARAMETERS .....................................................................................................87
TYPES OF DETECTORS .................................................................................................................88
DETECTOR SENSITIVITY ..............................................................................................................92
DETECTOR DYNAMIC RANGE......................................................................................................93
FID .............................................................................................................................................96
TCD............................................................................................................................................98
FID AND TCD COMPARISON ....................................................................................................100
ECD..........................................................................................................................................103
NPD..........................................................................................................................................104
EXERCISE 1: DETERMINING EFFICIENCY.....................................................................107
PART I: 30 M X 0.53MM X 0.88UM HP1 COLUMN ....................................................................108
PART II: 30 M X 0.32MM X 0.25UM HP 5 COLUMN ..................................................................111
EXERCISE 2: COMPARING COLUMN DIAMETER AND RESOLUTION ...................113
WORKSHEET .............................................................................................................................114
PACKED COLUMN ANALYSIS: 6' X 1/8" 5 % OV101 ON 80/100 CHROMOSORB.......................115
CAPILLARY COLUMN ANALYSIS: 30 M X 0.53 MM X 0.88 HP1 ...............................................116
CAPILLARY COLUMN ANALYSIS: 30 M X 0.32 MM X 0.25 UM HP5 .........................................117
EXERCISE 3: COMPARING COLUMN LENGTH AND RESOLUTION ........................119
WORKSHEET .............................................................................................................................120
CAPILLARY COLUMN ANALYSIS: 10 M X 0.53 MM X 0.88 UM HP1 .........................................122
CAPILLARY COLUMN ANALYSIS: 20 M X 0.53 MM X 0.88 UM HP1 .........................................123
CAPILLARY COLUMN ANALYSIS: 30 M X 0.53 MM X 0.88 UM HP1 .........................................124
EXERCISE 4: COMPARING COLUMN FILM THICKNESS AND RESOLUTION.......125

iii
WORKSHEET .............................................................................................................................126
CAPILLARY COLUMN ANALYSIS: 10 M X 0.53 MM X 0.88 UM HP1 .........................................128
CAPILLARY COLUMN ANALYSIS: 10 M X 0.53 MM X 2.65 UM HP1 .........................................129
EXERCISE 5: CALCULATING THE SPLIT RATIO ..........................................................131
WORKSHEET .............................................................................................................................132
SPLIT VENT FLOW = 44.8 ML/MIN ............................................................................................133
SPLIT VENT FLOW = 101.7 ML/MIN ..........................................................................................134
SPLIT VENT FLOW = 220 ML/MIN .............................................................................................135
EXERCISE 6: SPLIT AND SPLITLESS INJECTION .........................................................137
WORKSHEET .............................................................................................................................138
APPENDIX..................................................................................................................................141
REFERENCES .............................................................................................................................142
APPLICATION NOTES ................................................................................................................143
BIBLIOGRAPHY .........................................................................................................................154
EXERCISE ANSWERS .................................................................................................................161

iv
Introduction to Capillary GC
Introduction to Capillary GC
Introduction

Introduction

In This Section, You Will Learn:

• Course Description:
– The purpose of this one-day seminar is to introduce the analytical chemist to the
theoretical and practical principles of capillary gas chromatography.
• Course Goals:
– To determine what is needed in order to perform a GC analysis.
– To define resolution and column parameters for capillary columns in order to
achieve good resolution.
– To define the principles for sample introduction techniques (split/splitless).
– To define the principles for detection.
• Student Profile and Prerequisites:
– There are NO prerequisites for this course.
– This course is suitable for anyone desiring an introduction to the theory and
practice of capillary gas chromatography.

Figure 1

2
Introduction to Capillary GC
Introduction

Agenda

• Introduction: • Sample Introduction


– What is Capillary Gas Considerations:
Chromatography? – Split Injection Technique
– Sample Preparation – Electronic Pressure Controlled
Considerations Injection Technique
– Column Considerations – Splitless Injection Technique
• Basic Resolution Equation: • Detector Parameters:
– Column Efficiency Parameters – Definition of Detector Types
• Flowrate and Linear Velocity – Detector Response
• Column Diameter Characteristics
• Column Length – Flame Ionization Detector
• Carrier Gas Molecular Weight
– Capacity Parameters – Thermal Conductivity Detector
• Film Thickness – Electron Capture Detector
• Oven Temperature • Appendix
– Selectivity Parameters – Additional References
• Type of Stationary Phase

Figure 2

3
Introduction to Capillary GC
Introduction

Introduction
What is Capillary Gas Chromatography?

?
?
?? ?

Figure 3

4
Introduction to Capillary GC
Introduction

Typical Gas Chromatograph

MOL-SIEVE FIXED
TRAPS RESTRICTORS

RECORDER/
ELECTROMETER INTEGRATOR
INJECTION
PORT
REGULATORS

FLOW
CONTROLLER

DETECTOR

CARRIER GAS
HYDROGEN

COLUMN
AIR

Figure 4

5
Introduction to Capillary GC
Introduction

Definitions

• Gases:
– Carrier Gas – Pressurized gas used to transport the sample through the
system.
– Detector Gases – Support gases for certain detectors (i.e., FID).
• Sample Introduction:
– Introduces the sample to the carrier gas stream with minimal disruption of
the gas stream.
• Column:
– Achieves separation of the components in the sample.
• Detector:
– Recognizes and responds to sample components as the elute from the
column.
• Data Acquisition:
– Converts the detector signal to a picture chromatograph and provides
manual or automated determination of the identity and amounts of the
sample components.

Figure 5

6
Introduction to Capillary GC
What is Capillary Gas Chromatography

What is Capillary Gas Chromatography

What is a Chromatograph?
Chromatograph and Report generated by the 3365 ChemStation.

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
9
Column Flowrate = 1 mL/min 1 2 6 7 10 11 12
Linear Velocity = 20.5 cm/sec 5
4

Temperature Program:
Isothermal 160 3
Split Mode: 8
Split Ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.442 62321 45133 BV 0.022 0.2127 Pentane
2 2.556 2.9045E+07 1.87002E+07 VV 0.024 99.11066 Solvent
3 2.672 8303 5036 VB 0.026 0.0283+
4 3.100 10129 7143 BB 0.023 0.0346 Undecane
5 3.511 14978 7438 BB 0.032 0.0511 4-Chlorophenol
6 3.816 16871 9417 BB 0.028 0.0576 1-Decylamine
7 4.156 17124 9425 BB 0.029 0.0584 Tridecane
8 4.359 7748 3907 BB 0.031 0.0264 Methyl caprate
9 5.179 33486 14578 BB 0.046 0.1143 Tetradecane
10 6.315 27045 9249 BB 0.046 0.0923 Acenaphthylene
11 6.542 29540 8682 BB 0.053 0.1008 1-Dodecanol
12 6.789 34293 11407 BB 0.047 0.1170 Pentadecane
Total area = 2.9307E +07

Figure 6

7
Introduction to Capillary GC
What is Capillary Gas Chromatography

Typical Chromatograph from the 3396 Integrator

RUN# 54 JAN 3, 1901 00:56:26

SAMPLE NAME: BGC SAMPLE# 1


AUTO TEST

AREA%

RT AREA TYPE WIDTH AREA%


1.178 94134 PB .010 .11476
1.234 4920 BP .011 .00600
1.282 81205504 SPB .016 99.00070
1.361 2650 BP .011 .00323
1.397 24940 PP .019 .03041
1.524 5191 PB .012 .00633
1.952 36455 PB .015 .04444
2.500 50784 PB .020 .06191
2.960 57560 PB .023 .07017
3.530 62750 PB .028 .07650
3.860 29708 PB .031 .03622
5.242 120098 PB .045 .14642
6.715 107378 PB .055 .13091
7.229 99861 PB .062 .12174
8.160 123337 BB .075 .15036

TOTAL AREA=8.2025E+07
MUL FACTOR=1.0000E+00
8

Figure 7

8
Introduction to Capillary GC
What is Capillary Gas Chromatography

What Must Be Done in Order to Perform a GC


Analysis?

?
?
? ?
?

Figure 8

9
Introduction to Capillary GC
What is Capillary Gas Chromatography

What will Be Discussed in This Course

Identification of what must be done; Not how to perform


Operational Tasks.

• Topics Include:
– Sample Preparation Considerations
– Choosing a Column for Best Resolution
– Identifying Choice of Inlets
– Identiying Detectors

• Topics NOT Included:


– How to Install a Column
– How to Set Flows
– How to Calibrate
– How to Use an Auto Injector

10

Figure 9

10
Introduction to Capillary GC
Sample Preparation Considerations

Sample Preparation Considerations

Capillary Columns Require CLEAN SAMPLES!


What are some sample preparation methods?

• ____________ ("Dilute and Shoot")


• ____________ Solution
Centrifugation
Filtration
• ____________
Ion Exchange, Adsorption Columns
Other Solid Phase Extraction
Separatory Funnel
Supercritical Fluid Extraction
Headspace
• _____________
Thermal Desorption
Headspace
• ____________
Purge and Trap
• ____________
Pyrolysis
Derivatization

• Have as ____________ a Sample as Possible.


11

Figure 10

11
Introduction to Capillary GC
Column Considerations

Column Considerations

Columns
What is the primary
function of the column?

?? ??
?

12

Figure 11

12
Introduction to Capillary GC
Column Considerations

Column Selection Techniques

The following are the most common methods used for column
selection:

– Tray a column that is handy.


– Ask a colleague for advice.
– Look for a similar application that has already been published.
– In in doubt, use a non-polar capillary column such asm HP1 or HP5.

These approaches are good starting points, but one must optimize
performance.

13

Figure 12

13
Introduction to Capillary GC
Column Considerations

Example Applications

Type of Column

Carrier Conditions

Oven Conditions

Injection Parameters

Detector Parameters

Sample Information
14

Figure 13

14
Introduction to Capillary GC
Column Considerations

What are the Column Types?


Open Tubular (Capillary)

Wall Coated
Open Tube

Porous Layer 30 Meter X 0.53 mm X 0.88um HP1


Open Tube

Length Diameter Film Type


Thickness of Phase

Packed

Conventional
Packed and Porous
Micro Packed Layer
Bead

6 Feet X 1/8 inch 5% OV101 on 80/100 Chromosorb

Length Diameter Concentration Type of


and Type of Solid Support
Phase

15

Figure 14

15
Introduction to Capillary GC
Column Considerations

Packed Column
Packed Column: 6’ x 1/8” 5% OV101
on 80/100 Chromosorb

Capillary Column Analysis: 1 2


4
Column Flowrate = 30 mL/min 7
Linear Velocity = 12.4 cm/sec
6 9
5
Temperature Program:
Isothermal 130 11 & 12
10
Split Mode: 8
Split Ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 0.246 2879398 4458532 BV 0.015 0.2317 Pentane
2 0.341 1.22622E+09 8.10852E+08 HBS 0.026 98.6695 Solvent
4 1.006 823259 279443 BBT 0.046 0.0662 Undecane
5 1.361 916089 177078 BBT 0.073 0.0737 4-Chlorophenol
6 2.512 1387522 196260 BV 0.106 0.1116 1-Decylamine
8 2.630 749162 97326 VB 0.113 0.0603 Methyl caprate
7 4.12 2650317 232666 BV 0.177 0.2133 Tridecane
9 4.505 2331063 193209 VB 0.187 0.1876 Tetradecane
10 5.678 1879881 106858 BV 0.253 0.1513 Acenaphthylene
11&12 6.830 2807565 146329 VB 0.297 0.2259 1-Dodecanol & Pentadecane
Total area = 1.24275E +09

16

Figure 15

16
Introduction to Capillary GC
Column Considerations

Capillary Column – 0.53mm ID

Capillary Column Analysis:


30 M x 0.53 mm x 0.88 uM HP1 1 2
Column Flowrate = 2.6 mL/min 9
Linear Velocity = 19.7 cm/sec
12
Temperature Program:
Isothermal 180 7&8 10 11
5
Split Mode: 4
6
Split Ratio = 39:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.541 333138 156491 BV 0.033 0.7688 Pentane
2 2.711 4.24499E+07 1.81871E+07 PV 0.036 97.9577 Solvent
4 3.438 27293 11612 BB 0.0630 0.0630 Undecane
5 3.743 44075 13854 BB 0.1017 0.1017 4-Chlorophenol
6 4.206 38866 12775 BB 0.0897 0.0897 1-Decylamine
7&8 4.670 71919 16219 BB 0.1660 0.1660 Tridecane & Methyl caprate
9 5.772 96872 25091 BB 0.2235 0.2235 Tetradecane
10 6.659 79214 16874 BV 0.1828 0.1828 Acenaphthylene
11 6.808 91419 16880 VB 0.2110 0.2110 1-Dodecanol
12 7.420 102251 20686 BB 0.2360 0.2360 Pentadecane
Total area = 4.33349E +07

17

Figure 16

17
Introduction to Capillary GC
Column Considerations

Capillary Column – 0.32mm ID

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
1 2 6 9 11 12
Column Flowrate = 1 mL/min 7 10
Linear Velocity = 20.5 cm/sec 4 5

Temperature Program:
3
Isothermal 160
Split Mode: 8
Split Ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.442 62321 45133 BV 0.022 0.2127 Pentane
2 2.556 2.9045E+07 1.87002E+07 VV 0.024 99.11066 Solvent
3 2.672 8303 5036 VB 0.026 0.0283+ Impurity
4 3.100 10129 7143 BB 0.023 0.0346 Undecane
5 3.511 14978 7438 BB 0.032 0.0511 4-Chlorophenol
6 3.816 16871 9417 BB 0.028 0.0576 1-Decylamine
7 4.156 17124 9425 BB 0.029 0.0584 Tridecane
8 4.359 7748 3907 BB 0.031 0.0264 Methyl caprate
9 5.179 33486 14578 BB 0.046 0.1143 Tetradecane
10 6.315 27045 9249 BB 0.046 0.0923 Acenaphthylene
11 6.542 29540 8682 BB 0.053 0.1008 1-Dodecanol
12 6.789 34293 11407 BB 0.047 0.1170 Pentadecane
Total area = 2.9307E +07

18

Figure 17

18
Introduction to Capillary GC
Column Considerations

What Does the Column Provide?

____________ is a measure of the ability of a column to separate into two


peaks. Separation of compounds , and

Flow
A

Let’s look at how the separation D

occurs.
19

Figure 18

19
Introduction to Capillary GC
Column Considerations

A Trip Down Bourbon Street

Non-Drinker: 1 Bar in 1 minutes


Bourbon Drinker: 1 Bar in 2 minutes
Hurricane Drinker: 1 Bar in 4 minutes

Bourbon Street with 4 Bars


Joe's Place Howard's Two Guys Sam's Spirits

Time for Non-Drinker to Elute at End of Street: ____ minutes.

Joe's Place Howard's Two Guys Sam's Spirits

Time for Bourbon Drinker to Elute at End of Street: ____ minutes.

Joe's Place Howard's Two Guys Sam's Spirits

Time for Hurricane Drinker to Elute at End of Street: ____ minutes.

Non-Drinker Bourbon Drinker Hurricane Drinker


RT = ______ RT = ______ RT = ______
20

Figure 19

20
Introduction to Capillary GC
Column Considerations

21
Basic Column Resolution
Basic Column Resolution
Resolution Equation

Resolution Equation

Basic Resolution Equation

R=
n k
α-1
4
1+k α
Column Retention Selectivity
Efficiency (Capacity)

FACTORS AFFECTING

Flowrate / Linear Velocity Film Thickness Type of Stationary Phase


Column Diameter Oven Temperature
Column Length
Carrier Gas Molecular Wt.

Figure 20

23
Basic Column Resolution
Resolution Equation

Basic Resolution Equation (cont.)

R=
n k
α- 1
4 1+k
α
Column Retention Selectivity
Efficiency (Capacity)

FACTORS AFFECTING

Flowrate / Linear Velocity Film Thickness Type of Stationary Phase


Column Diameter Oven Temperature
Column Length
Carrier Gas Molecular Wt.

Figure 21

24
Basic Column Resolution
Flow Rate/Linear Velocity and Resolution

Flow Rate/Linear Velocity and Resolution

Chromatograms Using Different Flow


Rates/Linear Velocities

Capillary Column Analysis:


30 M x 0.53 mm x 0.88 uM HP1 1 8
2
Column Flow rate = 4.5 mL/min 5 9
Linear Velocity = 34 cm/sec 4 6 11
3 10
Temperature Program:
Isothermal 160
7
Split Mode:
Split ratio = 23:1
FID
1 uL Column Checkout Sample

Pk# Ret. Time Name


1 1.477 Pentane
2 1.609 Solvent
3 2.333 Undecane
4 2.638 4-Chlorophenol
5 3.143 1-Decylamine
6 3.681 Tridecane
7 3.738 Methyl caprate
8 4.970 Tetradecane
9 5.823 Acenaphthylene
10 6.009 1-Dodecanol
11 6.982 Pentadecane

Figure 22

25
Basic Column Resolution
Flow Rate/Linear Velocity and Resolution

Chromatograms Using Different Flow


Rates/Linear Velocities (cont.)

Capillary Column Analysis:


30 M x 0.53 mm x 0.88 uM HP1
Column Flow rate = 10.3 mL/min 12
Linear Velocity = 78 cm/sec 4 8
9
Temperature Program:
5 6&7
Isothermal 160 3 10 11
Split Mode:
Split ratio = 11:1
FID
1 uL Column Checkout Sample

Pk# Ret. Time Name


1 0.640 Pentane
2 0.695 Solvent
3 1.012 Undecane
4 1.145 4-Chlorophenol
5 1.363 1-Decylamine
6 1.597 Tridecane
7 Methyl caprate
8 2.157 Tetradecane
9 2.527 Acenaphthylene
10 2.608 1-Dodecanol
11 3.030 Pentadecane
5

Figure 23

26
Basic Column Resolution
Flow Rate/Linear Velocity and Resolution

Chromatograms Using Different Flow


Rates/Linear Velocities (cont.)

Capillary Column Analysis:


1 2
30 M x 0.53 mm x 0.88 uM HP1
Column Flow rate = 28 mL/min 4 8
9
Linear Velocity = 213 cm/sec 6&7
5 10 11
Temperature Program: 3
Isothermal 160
Split Mode:
Split ratio = 4.5:1
FID
1 uL Column Checkout Sample

Pk# Ret. Time Name


1 0.235 Pentane
2 0.254 Solvent
3 0.372 Undecane
4 0.421 4-Chlorophenol
5 0.501 1-Decylamine
6 0.589 Tridecane
7 Methyl caprate
8 0.793 Tetradecane
9 0.930 Acenaphthylene
10 0.960 1-Dodecanol
11 1.114 Pentadecane

Figure 24

27
Basic Column Resolution
Flow Rate/Linear Velocity and Resolution

Back to Bourbon Street


Column Flowrate or Linear Velocity

Bourbon Street with 4 Bars


Joe's Place Howard's Two Guys Sam's Spirits

Figure 25

28
Basic Column Resolution
Efficiency

Efficiency

Efficiency – Sharp Peaks


Solute
Separation efficiency is a measure of the broadening of
the solute plug as it travels the length of the column.
Start
(Solvent)
Wh = .074
Inert Gas

t m t 'R
(1.477) t R (6.982) Time

We would like to know the actual time the


component spends in the stationary phase.
t R' = t R - tm t R' = 6.982 - 1.477
= 5.51
t R' 2
n = 5.545 n = effective theoretical plates
Wh
n = 30,743
Let's relate n to the length of the column.
n 30743
Plates per meter (N) = = = 1024
or L 30
L (mm)
Height equivalent to a theoretical plate (HETP) = = 0.98 n

Thus, the more efficient the column the bigger the "N" , the smaller the "HETP".

Figure 26

29
Basic Column Resolution
Efficiency

Efficiency and Carrier Gas Linear Velocity

B
HETP = A + +Cu
u

HETP
B{ }
MOLECULAR C RESISTANCE TO MASS TRANSFER

DIFFUSION }
A EDDY DIFFUSION

u opt u

2
Flowrate (ml/min) = r u 60 π

r = Column radius in cm
u = Average linear velocity in cm/sec

col length (cm)


u=
R.T. (sec) of non-retained species

The minimum of the curve represents the smallest


HETP (or largest plates per meter) and thus the
best efficiency. "A" term is not present for
capillary columns.
9

Figure 27

30
Basic Column Resolution
Efficiency

Optimum Flowrate Related to Column Diameter


Generally, the following guidelines may be used for finding the optimum
values for a particular column:

Packed Columns

Optimum Carrier Optimum Carrier


Column Diameter Flowrate Linear Velocity
1/4" 50-60 ml/min 2.6 - 3.2 cm/sec
1/8" 20-30 ml/min 4.2 - 6.3 cm/sec

Capillary Columns

Optimum Carrier Optimum Carrier


Column Diameter Flowrate Linear Velocity
Series 530 mµ (megabore) 3-5 ml/min 22 - 38 cm/sec
320µ m (wide bore) 1-2 ml/min 20 - 41 cm/sec
200 µ m (narrow bore) .4-.6 ml/min 21 - 32 cm/sec
100µ m (high speed) .2-.5 ml/min 42 - 106 cm/sec

10

Figure 28

31
Basic Column Resolution
Efficiency

Column Diameter vs. Efficiency

9000

5000

Efficiency
4000

3000

1800
1200

(n/m)
0.1 0.2 0.25 0.32 0.75 2
0.53 (packed)

Column Internal Diameter (mm)

11

Figure 29

32
Basic Column Resolution
Efficiency

Back to Bourbon Street


Column Diameter vs. Efficiency
Version 1: 4 Bars on Narrow Street
Joe's Place Tw o G uys

H ow ard's Sam's Spirits

Version 2: 4 Bars on Wide Street


Joe's Place Tw o G uys

H ow ard's Sam's Spirits

12

Figure 30

33
Basic Column Resolution
Efficiency

Increased Sensitivity with Capillary Columns

Packed Columns Capillary Columns

Area = 2600 Area = 2600


Height = 5 S/N = 5 Height = 10 S/N = 10
Noise = 1 Noise = 1

Sensitivity = Concentration/Unit Time

13

Figure 31

34
Basic Column Resolution
Efficiency

Basic Resolution Equation (cont.)

n k α -1
R=
4 1+k α
Column Retention Selectivity
Efficiency (Capacity)

FACTORS AFFECTING

Flowrate / Linear Velocity Film Thickness Type of Stationary Phase


Column Diameter Oven Temperature
Column Length
Carrier Gas Molecular Wt.

An "R" value can be calculated:

1.18 (RT - RT2) 1


R=
(W 1+ W 2 )

R = 1.5 Represents Baseline Resolution

14

Figure 32

35
Basic Column Resolution
Efficiency

Back to Bourbon Street


Film Thickness
Version 1: 4 Bars

Joe's Place Tw o G uys

Howard's Sam's Spirits

Version 2: 4 Bars - Each 3 Times Larger

Joe's Place Tw o G uys

Howard's Sam's Spirits

15

Figure 33

36
Basic Column Resolution
Temperature and Resolution

Temperature and Resolution

Oven Temperature
Isothermal 130 degrees

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1 mL/min 1
2
Linear Velocity = 34.2 cm/sec 9
6
Constant Flow On 4 5
7
Temperature Program: 10 11 12
Isothermal 130
Split Mode:
8
Split ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.349 66784 53782 BV 0.020 0.2090 Pentane
2 3.871 3.1668E+07 2.10801E+07 BBA 0.023 99.1278 Solvent
4 3.871 10949 6531 BB 0.027 0.0343 Undecane
5 4.943 16594 6390 BB 0.041 0.0519 4-Chlorophenol
6 5.853 18685 7146 BB 0.041 0.0585 1-Decylamine
7 6.963 18975 6202 BB 0.048 0.0594 Tridecane
8 7.611 8549 2482 BB 0.054 0.0268 Methyl caprate
9 10.304 36848 7892 BB 0.073 0.1153 Tetradecane
10 13.221 32845 4965 BB 0.104 0.1028 Acenaphthylene
11 14.178 30476 4662 BB 0.103 0.0954 1-Dodecanol
12 15.996 37948 5207 BB 0.115 0.1188 Pentadecane
Total area = 3.19466E +07
Efficiency (based on C14 peak) = 2195 plates per meter

16

Figure 34

37
Basic Column Resolution
Temperature and Resolution

Oven Temperature (cont.)


Isothermal 160 degrees

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5 9 12
Column Flowrate = 1 mL/min 1 2 6 7 10 11
Linear Velocity = 34.2 cm/sec 4 5
Constant Flow On
Temperature Program:
Isothermal 160 8
Split Mode:
Split ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.442 62348 45138 BV 0.022 0.2130 Pentane
2 2.556 2.90175E+07 1.87003E+07 VBA 0.024 99.1321 Solvent
4 3.100 10129 7143 BB 0.023 0.0346 Undecane
5 3.511 15015 7442 BB 0.032 0.0513 4-Chlorophenol
6 3.816 16871 9417 BB 0.028 0.0576 1-Decylamine
7 4.156 17543 9474 BB 0.029 0.0599 Tridecane
8 4.359 7748 3907 BB 0.031 0.0265 Methyl caprate
9 5.179 33486 14578 BB 0.036 0.1144 Tetradecane
10 6.315 27078 9252 BB 0.046 0.0925 Acenaphthylene
11 6.542 29534 8681 BB 0.053 0.1009 1-Dodecanol
12 6.789 34293 11407 BB 0.047 0.1172 Pentadecane
Total area = 2.92716E +07
Efficiency (based on C14 peak) = 1068 plates per meter

17

Figure 35

38
Basic Column Resolution
Temperature and Resolution

Oven Temperature (cont.)


Isothermal 200 degrees

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1 mL/min 9 11
Linear Velocity = 34.2 cm/sec 1 2
Constant Flow On 10
12
Temperature Program:
6 7
Isothermal 200
5
Split Mode: 4
Split ratio = 100:1 8
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.571 90685 59862 BV 0.024 0.3578 Pentane
2 2.633 2.50872E+07 1.50586E+07 VB 0.026 98.9838 Solvent
4 2.841 8765 6242 BB 0.022 0.0346 Undecane
5 2.997 13048 6939 VV 0.030 0.0515 4-Chlorophenol
6 3.082 14917 9536 VV 0.025 0.0589 1-Decylamine
7 3.171 15333 10241 VV 0.024 0.0605 Tridecane
8 3.230 6794 4317 VB 0.025 0.0268 Methyl caprate
9 3.454 29065 18233 BB 0.025 0.1147 Tetradecane
10 3.763 23674 13072 BV 0.029 0.0934 Acenaphthylene
11 3.862 29761 17018 VB 0.028 0.1174 1-Dodecanol
12 4.014 25502 11379 BV 0.035 0.1006 Pentadecane
Total area = 2.53448E +07
Efficiency (based on C14 peak) = 230 plates per meter

18

Figure 36

39
Basic Column Resolution
Temperature and Resolution

Temperature Programming
Rate Time 2
C
2 min
Min
Time 1 10 C Temp 2
Capillary Column Analysis: 0 min min 170
30 M x 0.32 mm x 0.25 uM HP5 Temp 1
130
Column Flowrate = 1 mL/min
Linear Velocity = 34.2 cm/sec
Temperature Program:
Initial Temp 130 1 2
9
12
Initial Time 0 min 10
Rate 10 7
11
6
Final Temp 170
5
Final Time 2 min 4

Split Mode: 8
Split ratio = 100:1
FID
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 1.842 306172 350394 BV 0.019 0.6820 Pentane
2 1.972 4.4062E+07 3.51403E+07 BV 0.033 98.1430 Solvent
4 2.636 269555 27802 BB 0.016 0.0600 Undecane
5 3.071 41444 32610 BB 0.020 0.0923 4-Chlorophenol
6 3.375 44080 38682 BB 0.030 0.0982 1-Decylamine
7 3.687 47924 41829 BB 0.018 0.1067 Tridecane
8 3.852 21588 18339 BB 0.019 0.0481 Methyl caprate
9 4.451 92643 68565 BB 0.021 0.2064 Tetradecane
10 5.179 75694 47476 BV 0.025 0.1686 Acenaphthylene
11 5.276 82445 44760 VB 0.029 0.1836 1-Dodecanol
12 5.462 94778 56319 BB 0.027 0.2111 Pentadecane
Total area = 4.48957E +07
Efficiency (based on C14 peak) = 2599 plates per meter

19

Figure 37

40
Basic Column Resolution
Temperature and Resolution

Basic Resolution Equation (cont.)

R=
n k α -1

α
4 1+k

Column Retention Selectivity


Efficiency (Capacity)

FACTORS AFFECTING

Flowrate / Linear Velocity Film Thickness Type of Stationary Phase


Column Diameter Oven Temperature
Column Length
Carrier Gas Molecular Wt.

α = Relative Retention of a Critical Pair of


Sample Components
1
= t R,j t 1 R,j > t 1 R,i
α 1
t R,i

α > 1 Indicates that Separation can be Achieved

20

Figure 38

41
Basic Column Resolution
Liquid Phase and Resolution

Liquid Phase and Resolution

Liquid Phase Selectivity


Methyl Silicone (HP-1)

Capillary Column Analysis:


10 M x 0.53 mm x 2.65 uM HP1 (Methyl Silicone)
Column Flowrate = 1 mL/min 1 2
Linear Velocity = 27.6 cm/sec
5
Temperature Program: 4 6
Isothermal 130
9 10
Split Mode: 7
Split ratio = 33:1 11
12

FID 8

1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 0.604 133320 227653 BV 0.010 0.3039 Pentane
2 0.881 4.29654E+07 3.00866E+07 BBA 0.024 97.9490 Solvent
4 2.859 37894 11253 BV 0.053 0.0864 Undecane
5 3.774 60737 14524 BB 0.066 0.1385 4-Chlorophenol
6 5.360 70988 11414 BB 0.098 0.1618 1-Decylamine
7 7.334 68144 7811 BV 0.137 0.1553 Tridecane
8 7.544 31576 3719 VB 0.130 0.0720 Methyl caprate
9 12.102 127604 8955 BB 0.223 0.2909 Tetradecane
10 13.386 128482 9041 BB 0.222 0.2929 Acenaphthylene
11 16.068 112143 6044 BB 0.288 0.2557 1-Dodecanol
12 20.157 128763 5433 BB 0.369 0.2935 Pentadecane
Total area = 4.3865E +07

21

Figure 39

42
Basic Column Resolution
Liquid Phase and Resolution

Liquid Phase Selectivity (cont.)


5% Phenyl Methyl Silicone (HP-5)

Capillary Column Analysis:


10 M x 0.53 mm x 2.65 uM HP5 (5% Phenyl Methyl Silicone)
Column Flowrate = 1 mL/min
1
Linear Velocity = 26.9 cm/sec 2
5
Temperature Program: 4
6
Isothermal 110
9
Split Mode: 7 10
Split ratio = 33:1
11 12

FID 8

1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 0.618 154322 233139 BV 0.011 0.2751 Pentane
2 0.923 5.52701E+07 2.60518E+07 BBA 0.037 98.5360 Solvent
4 2.861 33818 9229 BB 0.058 0.0603 Undecane
5 4.445 53379 10021 BB 0.082 0.0952 4-Chlorophenol
6 5.777 62584 8900 BB 0.102 0.1116 1-Decylamine
7 7.420 57894 6349 BB 0.141 0.1032 Tridecane
8 8.375 27113 2747 BB 0.156 0.0483 Methyl caprate
9 12.336 111045 7405 BB 0.230 0.1980 Tetradecane
10 16.623 110612 5914 BB 0.291 0.1972 Acenaphthylene
11 18.037 94467 4603 BB 0.333 0.1738 1-Dodecanol
12 20.709 112933 4515 BB 0.393 0.2013 Pentadecane
Total area = 5.60913E +07

22

Figure 40

43
Basic Column Resolution
Liquid Phase and Resolution

Liquid Phase Selectivity (cont.)


50% Phenyl Methyl Silicone (HP-17)

Capillary Column Analysis:


15 M x 0.53 mm x 2.0 uM HP17 ( 50% Phenyl Methyl Silicone)
Column Flowrate = 2.8 mL/min
Linear Velocity = 26.9 cm/sec
1
2
Temperature Program:
Isothermal 110 4
9
Split Mode: 7
Split ratio = 33:1 6
12
5
FID 11
8 10
1 uL Column Checkout Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 0.549 161267 275185 BV 0.010 0.2074 Pentane
2 0.755 7.63174E+07 6.62005E+07 BBS 0.030 98.5725 Solvent
4 1.301 66872 35187 BB 0.030 0.0860 Undecane
7 3.014 117546 27749 BB 0.066 0.1512 Tridecane
6 3.530 106182 21612 BB 0.075 0.1365 1-Decylamine
9 4.976 223142 33084 BV 0.106 0.2869 Tridecane
5 5.137 97490 17999 VB 0.083 0.1254 4-Chlorophenol
8 5.786 55214 7578 BB 0.115 0.0710 Methyl caprate
12 8.462 227177 20290 BB 0.177 0.2921 Pentadecane
11 12.847 186844 11987 BB 0.245 0.2403 Dodecanol
10 24.611 207428 8105 BB 0.405 0.2667 Acenaphthylene
Total area = 7.77666E +07

23

Figure 41

44
Basic Column Resolution
Liquid Phase and Resolution

Common Stationary Phase Coatings for


Capillary Columns
BONDED STATIONARY PHASES SPECIALTY PHASES

Applications HP Phases Applications HP Phases


Amines, HP-1 or Ultra 1 Purgeable Organics HP-VOC
Hydrocarbons,
Volatile Halogenated HP-624
Pesticides, Chlorinated Pesticides HP-608
PCBs, Phenols,
Sulfur Compounds

Alkaloids, Drugs, HP-5 or Ultra 2


Fatty Acid Methyl, CLASSIC PHASES (NONBONDED)
Esters,
Halogenated
Compunds
Drugs, Glycols, Applications HP Phases
HP 50+ Amino Acid HP-101
Pesticides, Derivatives,
Steroids Essential Oils
Acids, Alcohols, HP-FFAP Drugs, Glycols, HP-17
Aldehydes, Pesticides,
Acrylates, Steroids
Ketones, Nitriles
Alcohols, HP-20M
Alcohols, Free HP-INNO Wax Free Acids,
Acids, Aromatics, Essential Oils,
Essential Oils, Ethers, Glycols,
Solvents Solvents

24

Figure 42

45
Basic Column Resolution
Liquid Phase and Resolution

46
Basic Column Resolution
Liquid Phase and Resolution

Porous Layer Open Tubular Column

Analysis of Low Molecular Weight Substances or Inert Gases at Ambient


Temperature:

Specifications AI 2O 3/KCI AL 2O 3/Na SO


2 4
MolSieve A PoraPLOT Q PoraPLOT S PoraPLOT U CarboPlot P7

Applications C 1-C 10 High Permanent Alcohols, Hydro- Polar He, N 2,


Hydro- Selectivity gases, He, polar and carbons, volatiles, O 2 , CO,
carbons; towards Ne, Ar, O , apolar ketones, nitriles, CO ,
2 2
ppm C -C C N , CH , volatiles esters nitro com- C -C
2
in C -C1 2 4
isomers CO, C H4 , in water; pounds, 1 2

1 4
CD , C2 D6 hydro- alcohols,
hydrogen
4 2 6 carbons aldehydes
isotopes gases, ethane/
solvents ethylene,
ppm water,
'wet'
samples

25

Figure 43

Analysis of low molecular weight substances or inert gases at ambient


temperature.

47
Basic Column Resolution
Liquid Phase and Resolution

Comparison of Plot and Packed Columns


Permanent Gases
1. Hydrogen 5% 2
2. Nitrogen 48.5%
3. Oxygen 13%
4. Argon 0.5%
5. Carbon Monoxide 33% 5

3 Packed
Column
1 4
9 min
Column: 6' x 1/8" SS packed with HayeSep A 80/100 mesh
Oven: 25 C o
Detector: TCD
Flow: He 23 cc/min
Sample: 25 microliters Air plus hydrogen and Carbon Monoxide

Analysis of Laboratory Air


Column: 25 M x 0.32 mm fused silica PLOT
Molsieve 5A (30.0 mm)
(Part No. 19091P-MS2)
Temperature: 30 C o
Carrier Gas: H2, 55kPa (0.55 bar, 8 psi), 56 cm/s
Injector: Splitter, 200 ml/min
T = 150 C o
Detector: TCD o 2 3
T = 180 C
Sample size: 30 ul

1. Argon
2. Oxygen Plot
3. Nitrogen
Column
1

o 8 min

26

Figure 44

48
Basic Column Resolution
Liquid Phase and Resolution

Basic Resolution Equation (cont.)

R=
n k
α -1

4 1+k
α
Column Retention Selectivity
Efficiency (Capacity)

FACTORS AFFECTING

Flowrate / Linear Velocity Film Thickness Type of Stationary Phase


Column Diameter Oven Temperature
Column Length
Carrier Gas Molecular Wt.

27

Figure 45

49
Basic Column Resolution
Summary

Summary

Summary Relationships

1. Using capillary columns requires the use of very ____________ samples.


2. The ____________ the flowrate or linear velocity, the shorter the retention
time.
3. The ____________ the column diameter, the lower the optimum flowrate or
linear velocity.
4. The ____________ the column diameter, the more efficient the column.
5. Increased efficiency results in ____________ peaks and usually better
resolution.
6. Optimum ____________ or _____________ is affected by the type of
carrier gas used.
7. Detector sensitivity is _____________ with more narrow diameter columns
because the signal to noise ratio is increased.
8. Temperature Programming often improves the efficiency and
____________ the time of analysis.

28

Figure 46

50
Basic Column Resolution
Summary

Example Applications

Type of Column

Carrier Conditions

Oven Conditions

Injection Parameters

Detector Parameters

Sample Information
29

Figure 47

51
Basic Column Resolution
Summary

52
Sample Introduction Considerations
Sample Introduction Considerations
Types of Inlet Systems

Types of Inlet Systems

Sample Introduction

Purpose:
To introduce the sample on to the column in the vapor state.
• Syringe Injection:
– Manual injection
– Autosampler injection
• Valve Injection
– Gas sampling valves
– Liquid sampling valves
• Auxiliary Sampling Devices
– Purge and Trap
– Headspace

Figure 48

54
Sample Introduction Considerations
Types of Inlet Systems

Types of Inlet Systems

Purpose:
To allow the insertion of a sample into the gas chromatograph in a repeatable,
reproducible manner. The sample should be representative of the bulk, and
unless specifically desired, should be inserted without chemical change.

– Packed column inlet


– Septum-purged packed column inlet
– Split/splitless capillary inlet
– Split-only capillary inlet
– Electronic pressure controlled (EPC) cool on-column
– EPC Purged packed inlet
– EPC Capillary split/splitless inlet

Inlet design must consider the characteristics of capillary columns:


– High efficiency
– Low sample capacity

Figure 49

55
Sample Introduction Considerations
Types of Inlet Systems

Back to Bourbon Street


Four buses of tourists unload at the same time.

Canal Street

Split Vent

Bourbon Street Too many people would overload


Bourbon Street. Therefore, a large
proportion of the tourists go down
Canal Street.

Column
4

Figure 50

56
Sample Introduction Considerations
Split/Splitless Inlet

Split/Splitless Inlet

Split/Splitless Capillary Inlet


SEPTUM

VALVE
FLOW CONTROLLER (SEPTUM)
PURGE VENT

TOTAL FLOW

SPLIT VENT
BACK
(INLET) PRESSURE
PURGE VALVE REGULATOR

COLUMN
HEAD
COLUMN PRESSURE

SEPTUM
TOTAL FLOW PURGE

SPLIT VENT
PURGE VENT

Column flow is set by altering COLUMN HEAD PRESSURE.

SEPTUM PURGE should be set to 2-3 mL/min.


SPLIT VENT flow is TOTAL FLOW minus COLUMN FLOW and
SEPTUM PURGE flow.

Changing TOTAL FLOW should NOT change COLUMN HEAD PRESSURE


or SEPTUM PURGE flow.

Figure 51

57
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Flow Diagram:


Pre-Injection
SEPTUM

(SEPTUM)
2 mL/min PURGE VENT
TOTAL FLOW

50 mL/min

SPLIT VENT

(INLET)
48 mL/min
PURGE VALVE
47.4 mL/min

= CARRIER GAS

LINER

SEAL (Small
grooves in the
seal allow the
0.6 mL/min split to occur.)
COLUMN

Figure 52

58
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Flow Diagram:


Sample Injection

(SEPTUM)
PURGE VENT

TOTAL FLOW

SPLIT VENT

(INLET)
PURGE VALVE

= LIQUID SAMPLE
= CARRIER GAS

COLUMN

Figure 53

59
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Flow Diagram:


Sample Vaporization
(SEPTUM)
PURGE VENT
TOTAL FLOW

SPLIT VENT

(INLET)
PURGE VALVE

Vaporization takes place when sufficient


thermal energy is transferred to the sample.
= CARRIER GAS
= SAMPLE MOLECULES This can occur in the carrier gas, but is
more likely upon contact with a solid
= SOLVENT MOLECULES surface, such as the liner or mixing medium.

COLUMN

Figure 54

60
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Flow Diagram:


Sample/Carrier Mixing
(SEPTUM)
PURGE VENT
TOTAL FLOW

SPLIT VENT

(INLET)
PURGE VALVE

For the concept of SPLIT RATIO to be


valid, the sample (solvent + analyte)
must be mixed with the carrier gas to
= CARRIER GAS give a homogeneous mixture.
= SAMPLE MOLECULES

= SOLVENT MOLECULES At the bottom of the injection port


a small part of this mixture will
transfer to the column, while the
bulk of the mixture will leave the
inlet via the SPLIT VENT.
COLUMN

SPLIT RATIO = COLUMN FLOW + SPLIT VENT FLOW


COLUMN FLOW

Figure 55

61
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Flow Diagram:


Liner Overload
(SEPTUM)
PURGE VENT
TOTAL FLOW

SPLIT VENT

(INLET)
PURGE VALVE

A large injection of a solvent with a


large expansion volume can cause an
overload of the injection port liner.

= CARRIER GAS
This can result in loss of sample out
= SAMPLE MOLECULES the PURGE VENT as well as
= SOLVENT MOLECULES contamination of the in-coming carrier
gas line.

COLUMN

10

Figure 56

62
Sample Introduction Considerations
Split/Splitless Inlet

Split Ration Calculation


Split Ratio = Split Vent Flow + Column Flow
Column Flow

COLUMN
B HEAD
PRESSURE
SEPTUM
TOTAL FLOW PURGE

A
SPLIT VENT PURGE VENT
D C

A B C D
Column Total Column Septum Split Split
I.D. (um) Flow Flow Purge Flow Ratio
200 100 0.5 2.0 97.5 196:1
320 100 1.0 2.0 97 98:1
530 100 3.0 2.0 95 33:1

Split Vent Flow = (Split Ratio)(Column Flow) - Column Flow

11

Figure 57

63
Sample Introduction Considerations
Split/Splitless Inlet

Capillary Liners
Untreated Liners
Split/Splitless 4-mm id, 1000 ul nominal volume
Borosilica glass (untreated) with silanized glass wool plug
PN 19251-60540 Split and splitless modes, especially recommended
for rapid injections with the HP Automatic Sampler
(Univeral Liner)
Splitless 2-mm id, 250 ul nominal volume
PN 18740-80220 quartz glass (untreated), open tube
Splitless mode for slow injections
(manual and HP7671/7672 automatic samplers)
small sample volumes (< 1 ul)
and higher boiling point solvent systems

Deactivated Liners
Packed, tapered 4-mm id, 900 ul nominal volume
PN 5062-3587 Borosilica glass (deactivated), silanized glass wool plug
(deactivated)
Split and splitless modes,
can be self-packed
Unpacked, tapered 4-mm id,900-ul nominal volume
PN 5181-3316 borosilicate glass (deactivated)
Split and splitless modes,
can be self-packed
Dual-tapered 4-mm id, 800 ul nominal volume
PN 5181-3315 borosilicate glass (deactivated)
Split and splitless modes,
but preferred for splitless

TREAT AS ULTRA CLEAN PART!


NOTE: See Supplies Catalog for latest variety of liners.
12

Figure 58

64
Sample Introduction Considerations
Split/Splitless Inlet

Split Injection and Selective Inlets

Split injection may be used with:


– Valves
– Head Space Autosamplers
– Thermal Desorbers
– Purge and Trap Samplers
– Analytical Prolyzers

For additional information on these techniques, see:


GC Inlets – An Introduction, by Mathew Klee, Chapters 9-13.

13

Figure 59

65
Sample Introduction Considerations
Electronic Pneumatic Flow Control

Electronic Pneumatic Flow Control

Standard Split/Splitless Capillary Inlet with


Electronic Pressure Control (EPC)

This inlet has the same features as the


SPLIT PROGRAMMABLE standard inlet plus the following:
PRESSURE

Pressure Control:
Programming: 3 ramps, 0.01-99
psi/min.
TOTAL FLOW Range: 1-100 psi (7-689 kPa) digital
INCR
readout of head pressure.
Flow Range: 0-450 mL/min.
SPLIT/SPLITLESS SEPTUM PURGE
Constant Flow Mode: Inlet pressure is
INLET VENT VENT adjusted real time to maintain constant
INJECTION PORT volumetric flow at column outlet.

14

Figure 60

66
Sample Introduction Considerations
Electronic Pneumatic Flow Control

Example Applications

Type of Column

Carrier Conditions

Oven Conditions

Injection Parameters

Detector Parameters

Sample Information
15

Figure 61

67
Sample Introduction Considerations
Electronic Pneumatic Flow Control

The Analysis Process


Advantages of EPC for GC Inlets

Sample Preparation

Pressure programming at
injection can enhance
Sample Introduction transfer to the column

Pressure/temperature
programming can
Separation optimize separations,
reduce analysis times

Detection Constant flow can


stablize detector
response, improve peak
shapes for quantitation
Data Reduction

16

Figure 62

68
Sample Introduction Considerations
Flow Vs. Temperature

Flow Vs. Temperature

Flow vs. Temperature


30M x 0.32 x 0.25uM HP-5
Constant Flow Mode

FLOW (mL/min)
Constant Pressure Mode

FLOW (mL/min)
1.0
1.0 5 psi .8
.8 .6 0.92 mL/min
.6
.4
.4
.2
.2
50 100 150 200 250 300
50 100 150 200 250 300 TEMPERATURE
TEMPERATURE

PRESSURE (psi)
PRESSURE (psi)

10
10
8
8
6
6
4
4
2
2
50 100 150 200 250 300
50 100 150 200 250 300
TEMPERATURE
TEMPERATURE

1.8
1.6
HETP (mm)

1.4
1.2
1.0
.8
.6
.4
.2

1 2 3 4 5
FLOW

17

Figure 63

1ST GRAPH 2ND GRAPH

Flow Temp Flow Temp


.923 50 .923 50
.725 100 .923 100
.585 150 .923 150
.482 200 .923 200
.404 250 .923 250
.344 300 .923 300

3RD GRAPH 4TH GRAPH

Pressure Temp Pressure Temp


5 50 5 50
5 100 6.16 100
5 150 7.38 150

69
Sample Introduction Considerations
Flow Vs. Temperature

5 200 8.66 200


5 250 9.98 250
5 300 11.3 300

5TH GRAPH

HETP Flowrate
1.001 .24
0.69 .505
0.526 .77
0.456 1.03
0.476 1.53
.583 2.03
1.030 3.43
1.859 4.77

70
Sample Introduction Considerations
Flow Vs. Temperature

EPC Constant Flow Off


Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 0.98 mL/min
Linear Velocity = 20.4 cm/sec 2 10&11
Constant Flow Off 9
12
Initial Pressure = 8 psi
Temperature Program:
Initial Temp 110
Initial Time 3 min 1
7
Rate 10 6
5
Final Temp 180 4 8
Final Time 7 min
FID

1 uL Column Checkout
Sample at split ratio 100:1

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.451 25442 20280 BV 0.020 0.1190 Pentane
2 2.830 2.09577E+07 1.27598E+07 BB 0.026 98.0245 Solvent
4 5.160 19123 10431 BB 0.029 0.0894 Undecane
5 6.504 29426 12560 BB 0.037 0.1376 4-Chlorophenol
6 7.347 30184 13787 BB 0.033 0.1412 1-Decylamine
7 8.114 35244 17470 BB 0.032 0.1648 Tridecane
8 8.489 16991 8203 BB 0.033 0.0795 Methyl caprate
9 9.647 68633 32942 BB 0.033 0.3210 Tetradecane
10 10.774 126729 33598 BB 0.062 0.5927 Acenaphthylen
11 1-Dodecanol
12 11.194 70585 29393 BB 0.038 0.3301 Pentadecane
Total area = 2.13801E +07

18

Figure 64

71
Sample Introduction Considerations
Flow Vs. Temperature

EPC Constant Flow On


Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 0.98 mL/min
Linear Velocity = 20.4 cm/sec
2
Constant Flow On 9 12
Initial Pressure = 8 psi 11
Temperature Program: 1 10
Initial Temp 110
Initial Time 3 min 7
Rate 10 6
Final Temp 180 5
4 8
Final Time 7 min
FID

1 uL Column Checkout
Sample at split ratio 100:1

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 2.455 41934 33092 BV 0.020 0.1861 Pentane
2 2.834 2.20908E+07 1.33684E+07 BV 0.026 98.0502 Solvent
4 5.089 19648 11683 BB 0.027 0.0872 Undecane
5 6.319 30584 14331 BB 0.034 0.1357 4-Chlorophenol
6 7.074 31033 16251 BB 0.029 0.1377 1-Decylamine
7 7.756 35357 20627 BB 0.027 0.1569 Tridecane
8 8.087 17026 9657 BB 0.028 0.0756 Methyl caprate
9 9.103 68292 39287 BB 0.028 0.3031 Tetradecane
10 10.018 67524 32943 BV 0.032 0.2997 Acenaphthylen
11 10.058 58702 33436 VB 0.027 0.2605 1-Dodecanol
12 10.385 69202 38486 BV 0.028 0.3072 Pentadecane
Total area = 2.25301E +07

19

Figure 65

72
Sample Introduction Considerations
Flow Vs. Temperature

Important Considerations in Using EPC

• Bubble meters are ____________ obsolete.


Compare measured and calculated flows during setup and for
troubleshooting.
• EPC pressure/flow calculations apply to flow through
____________ columns.
Calibrate and evaluate conditions for work with packed
columns.
• EPC has significant advantage for the analysis of ____________
mixture and _____________ amounts.

20

Figure 66

73
Sample Introduction Considerations
Split/Splitless Inlet Design

Split/Splitless Inlet Design

Splitless Injection:
Inlet Design Requirements

• The majority of the injected sample is introduced into the


column.
• Higher sensitivity than the split method and therefore, used for
trace analysis.
• Peak broadening must be minimized.

21

Figure 67

74
Sample Introduction Considerations
Split/Splitless Inlet Design

Back to Bourbon Street


One Bus with a Few Tourists

Canal Street

Split Vent

Bourbon Street

Column
22

Figure 68

75
Sample Introduction Considerations
Split/Splitless Inlet Design

Inlet Design and Operation Schematic


Purge Off
Total Septum
Flow Purge
Purge Vent
54 ml/min 53 ml/min 3 ml/min 3 ml/min

1 ml/min
ml/min Inlet Purge Flow 50 ml/min

Split Vent
Purge Control Column
Valve Head
Pressure

To Detector Head Pressure

1m
1 ml/min Gauge

l/m
in
Purge On
Total Septum
Flow Purge
Purge Vent
54 ml/min 3 ml/min Septum Purge Flow 3 ml/min

50 ml/min

Split Vent
Purge Control Column
Valve
Head
Pressure

To Detector Head Pressure


1m

Gauge
l/m
in

23

Figure 69

76
Sample Introduction Considerations
Injection Modes

Injection Modes

Split Mode
1:100 Dilution of the Column Evaluation Sample

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.53 mL/min
1 2
Linear Velocity = 31.7 cm/sec
Constant Flow On
12
Initial Pressure = 12 psi at 110 degrees
9 10 11
Temperature Program:
Initial Temp 110
Initial Time 3 min 7
Rate 10 5 8
4 6
Final Temp 180 3
Final Time 7 min
Split Mode:
Split Vent Flow = 44.8
FID
1 uL Column Checkout
Sample diluted to 1:100
Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 1.574 3.98305E+07 3.85324E+07 BBAS 0.017 97.3335 Pentane
2 1.838 1069149 1209434 BV 0.015 2.6127 Solvent
3 2.074 174 229 BB 0.016 0.0004 Impurity
4 3.583 909 657 BB 0.022 0.0022 Undecane
5 4.761 1302 808 BB 0.025 0.0032 4-Chlorophenol
6 5.567 833 392 BB 0.030 0.0020 1-Decylamine
7 6.211 1683 1105 BB 0.024 0.0041 Tridecane
8 6.553 796 533 BB 0.024 0.0019 Methyl caprate
9 7.585 3318 2189 BB 0.024 0.0081 Tetradecane
10 8.400 3070 1843 BB 0.026 0.0075 Acenaphthylene
11 8.555 2543 1695 BB 0.024 0.0062 1-Dodecanol
12 8.882 3449 2285 BB 0.024 0.0084 Pentadecane
Total area = 4.0916E +07

24

Figure 70

77
Sample Introduction Considerations
Injection Modes

Splitless Mode
1:100 Dilution of the Column Evaluation Sample
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.32 mL/min 1
Linear Velocity = 27.4 cm/sec 2 12
9
Constant Flow On
11
Initial Pressure = 12 psi at 110 degrees 10
Temperature Program:
Initial Temp 40 7
Initial Time 0 min
Rate 10 4 5 8
6
Final Temp 180
Final Time 5 min 3
Splitless Mode:
Purge On Time = 0.77 min.
Split Vent Flow = 44.8
FID

1 uL Column Checkout
Sample diluted to 1:100
Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 1.828 1.20696E+09 7.34754E+07 BBAS 0.199 97.7130 Pentane
2 3.443 2.78214E+07 4639343 BV 0.074 2.2524 Solvent
3 4.460 4628 2001 BV 0.035 0.0004 Impurity
4 7.807 22671 14703 BB 0.024 0.0018 Undecane
5 9.174 26020 14153 BB 0.029 0.0021 4-Chlorophenol
6 9.919 27289 12118 BB 0.032 0.0022 1-Decylamine
7 10.562 39871 28096 BB 0.023 0.0032 Tridecane
8 10.881 19061 13238 BB 0.023 0.0015 Methyl caprate
9 11.842 76661 52722 BB 0.023 0.0062 Tetradecane
10 12.599 68145 43525 BB 0.025 0.0055 Acenaphthylen
11 12.746 62741 43356 BB 0.023 0.0051 1-Dodecanol
12 13.050 80267 52718 BB 0.024 0.0065 Pentadecane
Total area = 1.2352E+09

25

Figure 71

78
Sample Introduction Considerations
Injection Modes

Splitless Injection
Solvent Effect: Initial oven temperature is maintained
below the boiling point of the solvent
causing the solvent to condense at
the head of the column "swelling" the
stationary phase and trapping the
analyte.

S
S S S
S S A S
S
S
S S
S S A S
A A S
S S S S
S A S A
S A
S S S
S S S
A = ANALYTE

S = SOLVENT
S S S
S A S S
S S S S
S A S
S A S
S A S A S S A S S
S S
S S S S S
S A S S S S S A

SOLVENT BOILING POINT INITIAL OVEN TEMP

o
o
( C) ( C)
DICHLOROMETHANE 40 10-30
CHLOROFORM 61 25-50
CARBON DISULFIDE 46 10-35
DIETHYL ETHER 35 10-25
PENTANE 36 10-25
HEXANE 69 40-60
ISO-OCTANE 99 70-90

SOLVENT AND STATIONARY PHASE MUST BE COMPATIBLE


26

Figure 72

79
Sample Introduction Considerations
Injection Modes

No Solvent Effect
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.63 mL/min 12
Linear Velocity = 33.8 cm/sec 1
Constant Flow On
9 11
Initial Pressure = 12 psi at 110 degrees
Temperature Program: 10
Initial Temp 110
Initial Time 3 min
Rate 10 7
Final Temp 180 8
5 6
Final Time 7 min 4
Splitless Mode:
Purge On Time = 0.77 min.
Split Vent Flow = 21 ml/min
Solvent = Pentane
FID
1 uL Column Checkout
Sample diluted to 1:100
Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 1.479 1.25642E+09 6.6195E+07 BB 0.237 99.9661 Pentane
4 3.504 19422 2160 BB 0.108 0.0018 Undecane
5 4.698 22937 2857 BB 0.108 0.0018 4-Chlorophenol
6 5.499 31525 4229 BV 0.101 0.0025 1-Decylamine
7 6.183 40959 8540 VV 0.069 0.0033 Tridecane
8 6.526 19182 4464 VB 0.061 0.0015 Methyl caprate
9 7.571 76488 24623 BB 0.046 0.0061 Tetradecane
10 8.387 67132 21269 BV 0.047 0.0053 Acenaphthylene
11 8.546 63400 25018 VV 0.038 0.0050 1-Dodecanol
12 8.876 80189 34399 VB 0.036 0.0064 Pentadecane
Total area = 1.25685E+07

27

Figure 73

80
Sample Introduction Considerations
Injection Modes

Demonstration of the Purge Function


Purge Off the Entire Run (No Purge On Time)
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.2 mL/min
Linear Velocity = 24.8 cm/sec 1
Constant Flow On
Initial Pressure = 12 psi at 110 degrees 9
4 12
Temperature Program: 7 11
5
Initial Temp 40 6
8
Initial Time 0 min 10
Rate 10
Final Temp 180
Final Time 4 min
Splitless Mode:
Purge On Time = 0.77 min.
Split Vent Flow = 21 ml/min
Solvent = Pentane
FID
1 uL Column Checkout
Sample diluted to 1:100 Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 2.014 8.44044E+08 6.36828E+07 BBAS 0.160 99.9294 Pentane
4 8.083 19245 18112 BBA 0.019 0.0023 Undecane
5 9.449 48844 21638 PV 0.033 0.0058 4-Chlorophenol
6 10.188 42947 20065 PV 0.031 0.0051 1-Decylamine
7 10.820 60389 37891 PV 0.025 0.0071 Tridecane
8 11.136 30027 19510 PV 0.025 0.0036 Methyl caprate
9 12.090 101257 66142 VV 0.024 0.0120 Tetradecane
10 12.863 93363 55617 VV 0.026 0.0111 Acenaphthylene
11 12.989 86959 54461 PV 0.025 0.0103 1-Dodecanol
12 13.291 113651 67001 PV 0.026 0.0135 Pentadecane
Total area = 8.4464E+08

28

Figure 74

81
Sample Introduction Considerations
Injection Modes

Comparing Purge-On Times


Purge On 0.3 minutes
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.6 mL/min
1 8 11
Linear Velocity = 24.3 cm/sec 2
Constant Flow On 10
Initial Pressure = 12 psi at 110 degrees 9
Temperature Program:
6
Initial Temp 40
Initial Time 0 min
Rate 10 3 7
4 5
Final Temp 180
Final Time 4 min
Splitless Mode:
Purge On Time = 0.3 min.
Split Vent Flow = 44 ml/min

FID
1 uL Column Checkout
Sample diluted to 1:100 Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 2.081 7.90249E+08 6.14638E+07 BBAS 0.156 97.3732 Pentane
2 3.759 2.08811E+07 13516 BB 0.061 2.579 Solvent
3 8.100 19300 13516 BB 0.023 0.0024 Undecane
4 9.464 20503 11491 BB 0.028 0.0025 4-Chlorophenol
5 10.209 21299 8985 BB 0.034 0.0026 1-Decylamine
6 10.830 35759 25381 BB 0.022 0.0044 Tridecane
7 11.146 16794 11700 BB 0.023 0.0021 Methyl caprate
8 12.099 70123 48453 BB 0.023 0.0086 Tetradecane
9 12.873 59637 37232 BB 0.025 0.0073 Acenaphthylene
10 12.997 56760 38181 BB 0.023 0.0070 1-Dodecanol
11 13.299 76695 49076 BB 0.025 0.0095 Pentadecane
Total area = 8.11567E+08

29

Figure 75

82
Sample Introduction Considerations
Injection Modes

Comparing Purge-On Times


Purge On 0.7 minutes
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.18 mL/min
Linear Velocity = 24.3 cm/sec 1 2
8 11
Constant Flow On 10
Initial Pressure = 12 psi at 110 degrees 9
Temperature Program:
Initial Temp 40 6
Initial Time 0 min
Rate 10
3 4 7
Final Temp 180 5
Final Time 4 min
Splitless Mode:
Purge On Time = 0.7 min.
Split Vent Flow = 44 ml/min

FID
1 uL Column Checkout
Sample diluted to 1:100 Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 2.038 1.15306E+09 6.32523E+07 BBAS 0.217 97.6632 Pentane
2 3.726 2.71199E+07 4285045 BV 0.079 2.2970 Solvent
3 8.097 23268 14782 BB 0.025 0.0020 Undecane
4 9.465 27011 14779 BB 0.028 0.0023 4-Chlorophenol
5 10.207 23593 10473 BB 0.033 0.0020 1-Decylamine
6 10.831 40884 28411 BB 0.023 0.0035 Tridecane
7 11.147 19385 13693 BB 0.022 0.0016 Methyl caprate
8 12.101 80177 54745 BB 0.023 0.0068 Tetradecane
9 12.874 70910 43647 BV 0.026 0.0060 Acenaphthylene
10 13.000 66676 44021 PB 0.024 0.0056 1-Dodecanol
11 13.301 84329 54693 BB 0.024 0.00715 Pentadecane
Total area = 1.18065E+09

30

Figure 76

83
Sample Introduction Considerations
Injection Modes

Comparing Purge-On Times


Purge On 1.5 minutes
Capillary Column Analysis:
30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.19 mL/min
Linear Velocity = 24.3 cm/sec 1 2
Constant Flow On 8 11
Initial Pressure = 12 psi at 110 degrees 10
Temperature Program: 9
Initial Temp 40 6
Initial Time 0 min
Rate 10 3 7
4 5
Final Temp 180
Final Time 4 min
Splitless Mode:
Purge On Time = 1.5 min.
Split Vent Flow = 44 ml/min

FID
1 uL Column Checkout
Sample diluted to 1:100 Area Percent Report
Pk# Ret. Time Area Height Type Width Area% Name
1 2.025 1.07664E+09 6.34994E+07 BBAS 0.203 97.3248 Pentane
2 3.710 2.91426E+07 4250357 BB 0.087 2.6344 Solvent
3 8.093 23913 14883 BB 0.025 0.0022 Undecane
4 9.459 28871 14475 BB 0.031 0.0026 4-Chlorophenol
5 10.202 27145 11717 BB 0.033 0.0025 1-Decylamine
6 10.829 42931 30353 BB 0.022 0.0039 Tridecane
7 11.144 20543 14155 BB 0.023 0.0019 Methyl caprate
8 12.099 82470 57485 BB 0.023 0.0075 Tetradecane
9 12.872 73083 45772 BB 0.025 0.0066 Acenaphthylene
10 12.997 66749 45916 PB 0.023 0.0066 1-Dodecanol
11 13.299 86018 56588 BB 0.024 0.0078 Pentadecane
Total area = 1.110623E+09

31

Figure 77

84
Sample Introduction Considerations
Liners

Liners

Liners
Untreated Liners
Split/Splitless 4-mm id, 1000 ul nominal volume
Borosilica glass (untreated) with silanized glass wool plug
PN 19251-60540 Split and splitless modes, especially recommended
for rapid injections with the HP Automatic Sampler
(Univeral Liner)

Splitless 2-mm id, 250 ul nominal volume


quartz glass (untreated), open tube
PN 18740-80220 Splitless mode for slow injections
(manual and HP7671/7672 automatic samplers)
small sample volumes (< 1 ul)
and higher boiling point solvent systems

Deactivated Liners
Packed, tapered 4-mm id, 900 ul nominal volume
Borosilica glass (deactivated), silanized glass wool plug
PN 5062-3587 (deactivated)
Split and splitless modes,
can be self-packed

Unpacked, tapered 4-mm id,900-ul nominal volume


PN 5181-3316 borosilicate glass (deactivated)
Split and splitless modes,
can be self-packed

Dual-tapered
4-mm id, 800 ul nominal volume
PN 5181-3315 borosilicate glass (deactivated)
Split and splitless modes,
but preferred for splitless
32

Figure 78

85
Sample Introduction Considerations
Liners

86
Detector Parameters
Detector Parameters
Types of Detectors

Types of Detectors

GC Detector:
A Definition

A GC Detector is a device which senses the presence of a


component different from the carrier gas, and converts that
information to an electrical signal:

– Thermal Conductivity Detector (TCD)


– Flame Ionization Detector (FID)
– Electron Capture Detector (ECD)
– Nitrogen Phosphorus Detector (NPD)
– Flame Photometric Detector
– Electrolyte Conductivity Detector (ELCD, HALL)
– Photoionization Detector (PID)
– Mass Selective Detector (MSD)
– Infrared Detector (IRD)
– Atomic Emission Detector (AED)

Figure 79

88
Detector Parameters
Types of Detectors

Detectors
Types of
Detectors Brief Description
TCD Filament temperature increases as analytes present in the carrier gas pass over it, causing the
resistance to increase.
FID Components burn in a flame producing ions which are collected and converted into a current.

ECD As electronegative species pass through the detector, they capture low energy thermal electrons
causing a decrease in cell current.
NPD Nitrogen and phosphorous compounds produce increased currents in a flame enriched with
vaporized alkali metal salt.
FPD Sulphur and phosphorous compounds burn in a flame producing chemiluminescent species which
are monitored at selective wave lengths.
ELCD Halogens, sulphur, or nitrogen compounds are mixed with a reaction gas in a reaction tube. The
productsare mixed with a suitable liquid which produces a conductive solution.
PID Molecules are ionized by excitation with photons from a UV lamp. The charged particles are then
collected, producing a current.
MSD Molecules are bombarded with electrons producing ion fragments which pass into the
spectrometer's mass filter. The ions are filtered based on their mass/charge ratio.
IRD Molecules absorb infrared energy, the frequencies of which are characteristic of the bonds within
that Molecule.
AED Molecules are energized by a plasma source and separated into excited atoms. As electrons return
to their stable state, they emit light, which is element
Specific.
3

Figure 80

89
Detector Parameters
Types of Detectors

Detector Response Characteristics

Sensitivity: The response per amount of sample, that is, the slope of
the response/amount curve. The minimum amount on
the curve is defined as the minimum detectable level
(MDL).
Selectivity: A measure of which categories of compounds will give
a detector response.

Dynamic Range: The range of sample concentrations for which the


detector can provide accurate quantitation.

Figure 81

90
Detector Parameters
Types of Detectors

Comparison of GC Detectors

TCD

FID

ECD
AED
PID
IRD
N-P (N)

N-P (P)
FPD (S)
MSD
(SIM) (SCAN)
ELCD (X)
ELCD (SORN)
10 -15 10 -12 10 -9 10 -6 10 -3
fg pg ng mg mg

pptrillion ppb ppm ppthousand percent


1 ng 1 mg 1ml
1 ppm = = =
1 ml Liter Liter

Figure 82

91
Detector Parameters
Detector Sensitivity

Detector Sensitivity

Capillary Columns and Detector Selectivity

FID ECD NPD (P)

Figure 83

92
Detector Parameters
Detector Dynamic Range

Detector Dynamic Range

Dynamic Range
Dynamic range is a measure of response vs. Quantity
Response is the signal produced by the sample.

Dynamic Range
Response increases reproducibly
Response with increased quantity.

Quantity

Linear Dynamic Range


Response Response increases proportionally
with increased quantity.

Quantity

Non-linear response, as long as it is reproducible, can


be dealt with using non-linear calibration techniques.
7

Figure 84

93
Detector Parameters
Detector Dynamic Range

Single and Multi-Level Calibration

C
C A
C A
C A B
A B
B B

1 0.5 1.5 1.51.0 2.0 2.01.5 3.0 3.02.04.0


LEVEL 1 LEVEL 2 LEVEL 3 LEVEL 4

AREA/HEIGHT

AREA/HEIGHT
Level 4
Level 3
Level 2
Level 1

CONCENTRATION CONCENTRATION
SINGLE POINT FOR MULTI-LEVEL FOR
COMPONENT A COMPONENT C

Figure 85

94
Detector Parameters
Detector Dynamic Range

Operating Conditions for Detectors Commonly


Used in Capillary GC Detection
Flow Rate (ml/min)
Typical Carrier + H2
Type Samples Limits Makeup Air
10-100 pg
FID Hydrocarbons 20-60 30-40 200-500
10 ppb-99%
TCD General 5-100 ng 15-30 n.a. n.a.
10 ppm-100%
ECD Organohalogens 0.05-1 pg 30-60 n.a. n.a.
Chlorinated solvents 50 ppt-1 ppm
& pesticides
Organonitrogen 0.1-10 pg
NPD 20-40 1-5 70-100
& organophosphorus 100 ppt-0.1%
compounds
FPD Sulfur compounds 10-100 pg
(393 nm) 20-40 50-70 60-80
10 ppb-100 ppm
FPD Phosphorus 1-10 pg 20-40 120-170 100-150
(526 nm) compounds 1 ppb-0.1%
PID Compounds ionized 2pg C/sec 30-40 * n.a.
by UV
ELCD 0.5 pg Cl/sec 20-40 80 n.a.
Halogens, N, S 2 pg S/sec
4 pg N/sec
FTIR Molecular vibrations 1000 pg of 3-10 n.a. n.a.
strong absorber
Tunable for any 10 pg to 10 ng
MSD (depending on 0.5-30** n.a. n.a.
species SIM vs. scan)
Tunable for any 0.1 - 20 pg/sec
AED (depending on 60-70 PRESET PRESET
element element

* Refer to Detector Manual


** Dependent upon the type of interface 9

Figure 86

95
Detector Parameters
FID

FID

FID Schematic
Flame ionization detector
designed for use with
capillary columns. FID Detector
Assembly
Air
Inlet

Capillary Column
End-Position
(1-2 mm from Top of Jet) Jet

H2 Inlet
+
Make-Up

Flame ionization
detector gas controls
FID OFF PRESS

AIR FID IGNITOR


OFF OFF

HYDROGEN AUX GAS Exit End of Column

10

Figure 87

96
Detector Parameters
FID

Compounds with Little or No FID Response


Rare gases NH CS 2
3
Nitrogen oxides H2 COS
Silicon halides CO O2
HO CO N
2 2 2
Perhalogenated cpds HCOH HCOOH

Overcoming FID Selectivity

Problem: CO and CO trace analysis


2
(best TCD sensitivity 20-50 ppm)

CO Nickel 350
Good
OR + H2 Catalyst tube CH4
Solution: CO2
Sensitivity as good as 0.1 ppm

11

Figure 88

97
Detector Parameters
TCD

TCD

Single Filament TCD


EXIT
COLUMN flow enters the center
of three ports.

AUXILIARY, or make-up, flow DETECTOR


"sheathes" the column
CELL
minimizing diffusion at
the end of the column.
3.5 mL
FILAMENT VOLUME
REFERENCE flow is
directed to either
one of the outside
ports into the detector
cell. The entry port is
determined by the SWITCHING
SOLENOID, REFERENCE

AUXILIARY
REFERENCE

COLUMN
SWITCHING
SOLENOID

REFERENCE
REFERENCE FLOW = 1.5 x ( COLUMN + AUXILIARY ) FLOW
TCD OFF

AUX GAS
OFF Thermal conductivity
REF
detector gas controls

12

Figure 89

98
Detector Parameters
TCD

Type of Carrier Gas

Thermal Conductivity Relative to Helium


Relative Thermal
Compound Conductivity
Carbon Tetrachloride 0.05
Benzene 0.11
Hexane 0.12
Argon 0.12
Methanol 0.13
Nitrogen 0.17
Helium 1.00
Hydrogen 1.28

13

Figure 90

99
Detector Parameters
FID and TCD Comparison

FID and TCD Comparison

Comparison of FID and TCD Analysis


FID Analysis

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.52 mL/min 12
1
Linear Velocity = 31.4 cm/sec 2 9
Initial Pressure = 12 psi 11
Temperature Program: 10
Initial Temp 110
Initial Time 3 min 6 7
Rate 10 5
Final Temp 180 4
8
Final Time 7 min 3

Split Vent Flow = 44 ml/min

FID
1 uL Column Checkout
Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 1.591 127589 181613 BV 0.012 0.1952 Pentane
2 2.083 6.40112E+07 4.70123E+07 BV 0.023 97.9545 Solvent
3 2.083 20660 23683 BV 0.018 0.0316 Impurity
4 3.599 56809 41836 BB 0.022 0.0869 Undecane
5 4.482 81723 50109 BB 0.026 0.1251 4-Chlorophenol
6 5.526 96577 61596 BB 0.024 0.1478 1-Decylamine
7 6.233 103810 66682 BB 0.025 0.1589 Tridecane
8 6.570 50263 33953 BB 0.024 0.0769 Methyl caprate
9 7.614 203872 122264 BB 0.026 0.3120 Tetradecane
10 8.432 187396 107906 BB 0.027 0.2868 Acenaphthylene
11 8.586 174116 105709 BB 0.025 0.2664 1-Dodecanol
12 8.913 217376 127896 BB 0.025 0.3326 Pentadecane
Total area = 6.53479E+07

14

Figure 91

100
Detector Parameters
FID and TCD Comparison

Comparison of FID and TCD Analysis


TCD Analysis: Low Sensitivity

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.63 mL/min
Linear Velocity = 33.9 cm/sec 1 2
Initial Pressure = 12 psi
Temperature Program:
Initial Temp 110
Initial Time 3 min 8 10 11
Rate 10 9
Final Temp 180 4 5 6
Final Time 7 min 3 7

Split Vent Flow = 44 ml/min

TCD, low sensitivity


1 uL Column Checkout
Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 1.477 122 197 BV 0.010 0.1535 Pentane
2 1.709 78595 64677 BV 0.020 98.7155 Solvent
3 3.359 47 32 BB 0.024 0.0591 Undecane
4 4.532 89 54 BB 0.027 0.1112 4-Chlorophenol
5 5.275 78 52 BB 0.024 0.0985 1-Decylamine
6 5.989 83 49 BB 0.026 0.1037 Tridecane
7 6.327 47 30 BB 0.024 0.0593 Methyl caprate
8 7.381 150 87 BB 0.026 0.1886 Tetradecane
9 8.186 109 59 BB 0.028 0.1370 Acenaphthylene
10 8.360 145 83 BB 0.026 0.1815 1-Dodecanol
11 8.691 153 87 BB 0.028 0.1922 Pentadecane
Total area = 79617

15

Figure 92

101
Detector Parameters
FID and TCD Comparison

Comparison of FID and TCD Analysis


TCD Analysis: High Sensitivity

Capillary Column Analysis:


30 M x 0.32 mm x 0.25 uM HP5
Column Flowrate = 1.63 mL/min
Linear Velocity = 33.9 cm/sec 1 2
Initial Pressure = 12 psi 8 11
10
Temperature Program:
Initial Temp 110
9
Initial Time 3 min 4 5 6
Rate 10
Final Temp 180 3 7
Final Time 7 min

Split Vent Flow = 44 ml/min

TCD, high sensitivity


1 uL Column Checkout
Sample

Area Percent Report


Pk# Ret. Time Area Height Type Width Area% Name
1 1.477 3498 5319 BV 0.011 0.6309 Pentane
2 1.709 526404 292634 BBAS 0.022 94.9492 Solvent
3 3.361 1269 903 BB 0.022 0.2290 Undecane
4 4.535 2458 1457 BB 0.026 0.4433 4-Chlorophenol
5 5.279 2264 1420 BB 0.025 0.4083 1-Decylamine
6 5.991 2135 1349 BB 0.025 0.3850 Tridecane
7 6.330 1281 851 BB 0.024 0.2311 Methyl caprate
8 7.383 4037 2450 BB 0.026 0.7281 Tetradecane
9 8.188 2960 1655 BB 0.028 0.5340 Acenaphthylene
10 8.361 3882 2299 BB 0.027 0.7003 1-Dodecanol
11 8.692 4218 2415 BB 0.027 0.7609 Pentadecane
Total area = 554406

16

Figure 93

102
Detector Parameters
ECD

ECD

Electron Capture Detector Schematic

Anode Purge

Vent
Anode

Plated Ni
Surface
Nickel Plating

Fused Silica
Liner

Makeup Gas Makeup Gas


Adapter
Column
Trace Analysis
Halogenated compounds
Nitro group compounds
Conjugated double bond compound
Common Applications
Environmental (organo chlorine pesticides)
17

Figure 94

103
Detector Parameters
NPD

NPD

Nitrogen Phosphorus Detector Schematic

NPD
Collector

Air
Inlet
Active
Capillary Column Element
End-Position
(1-2 mm from Top of Jet) Jet

H2 Inlet
+
Make-Up

* Active element is a potassium or rubidium alkali metal salt.

Diagram of a nitrogen phosphorous detector designed for


optimal performance with capillary columns.
18

Figure 95

104
Detector Parameters
NPD

Example Applications

Type of Column

Carrier Conditions

Oven Conditions

Injection Parameters

Detector Parameters

Sample Information
19

Figure 96

105
Detector Parameters
NPD

106
Exercise 1: Determining Efficiency
Exercise 1: Determining Efficiency
Part I: 30 M x 0.53mm x 0.88um HP1 Column

Part I: 30 M x 0.53mm x 0.88um HP1 Column

1 2
8
5 9
4 11
3 6
10

Column head pressure: 3 psi


Isothermal 160 degrees FID
1 uL Column Checkout sample
Pk# Ret Time Peak Width Name
1 1.477 0.015 Pentane
2 1.609 0.021 Solvent
3 2.333 0.022 Undecane
4 2.638 0.029 4-Chlorophenol
5 3.143 0.033 1-Decylamine
6 3.681 0.036 Tridecane
7 3.738 0.035 Methyl Caprate
8 4.970 0.051 Tetradecane
9 5.823 0.062 Acenaphthylene
10 6.009 0.063 1-Dodecanol
11 6.982 0.074 Pentadecane

108
Exercise 1: Determining Efficiency
Part I: 30 M x 0.53mm x 0.88um HP1 Column

1) Calculate the Linear Velocity, Flowrate, HETP, and Plates/Meter using peak
8 and use peak 1 as the non-retained species.

2) Using the results from your previous calculations, complete the following
table:

30 M x 0.53 x 0.88 HP1


Column Head Linear
Pressure Velocity Flowrate
(psi) (cm/sec) (mL/min) HETP Plates/Meter
1 10.7 1.4 1.75 571
2 20 2.7 1.2 808
3
4 44 5.9 1.3 768
7 78 10.3 2.0 506
11 121 16 3.53 283
20 213 28 8.4 119

109
Exercise 1: Determining Efficiency
Part I: 30 M x 0.53mm x 0.88um HP1 Column

3) Use the graph below and plot the HETP vs. linear velocity for each of the
columns. Construct on graph paper:
4

3.5

2.5

1.5

0.5

0 20 40 60 80 100 120 140

110
Exercise 1: Determining Efficiency
Part II: 30 M x 0.32mm x 0.25um HP 5 Column

Part II: 30 M x 0.32mm x 0.25um HP 5 Column

1 2 8
5
3 4 6
9 11
10

4) Use the data below to plot HETP vs. LV for this column:
Make the plot on the same graph for comparison:

Column Head
Pressure Linear Flowrate
(psi) Velocity (mL/min) HETP Plates/Meter
2 5.04 0.24 1.001 999
4 10.5 0.505 0.69 1443
6 15.9 0.77 0.526 1900
8 21.3 1.03 0.456 2195
12 31.8 1.53 0.476 2103
16 42 2.03 0.583 1715
28 71.1 3.43 1.030 971
40 99 4.77 1.859 538

111
Exercise 1: Determining Efficiency
Part II: 30 M x 0.32mm x 0.25um HP 5 Column

5) Which column is more efficient?

112
Exercise 2: Comparing Column
Diameter and Resolution
Exercise 2: Comparing Column Diameter and Resolution
Worksheet

Worksheet
Using the following equation for determining resolution quantitatively, compare
three columns by calculating the "R" value for the separation of C-14 and
acenaphthlalene using the data on the next three pages:

RT1 = retention time of the first of the two


R = 1.18 (RT2) – RT1) peaks that one is calculating the
----------------------- resolution of
(W1 + W2)
RT2 = retention time of the second peak
W1 = peak width of first peak
W2 = peak width of second peak

Packed column R = _______________

0.53 column R = _______________

0.32 column R = _______________

Which column diameter yielded the best “R” value?

114
Exercise 2: Comparing Column Diameter and Resolution
Packed Column Analysis: 6' x 1/8" 5 % OV101 on 80/100
Chromosorb

Packed Column Analysis: 6' x 1/8" 5 % OV101 on


80/100 Chromosorb

1 2 4
7
6 9
5
11&12

8 10

Packed Column Analysis: 6' x 1/8" 5 % OV101 on 80/100 Chromosorb


Column Flowrate = 30 mL/min. Linear Velocity = 12.4 cm/sec
Isothermal 130 degrees FID
1 uL HP Column Checkout Sample
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 0.246 2879398 4458532 BV 0.015 0.2317 Pentane
2 0.341 1.22622E+009 8.10852E+008 HB S 0.026 98.6695 Solvent
4 1.006 823259 279443 BB T 0.046 0.0662 Undecane
5 1.361 916089 177078 BB T 0.073 0.0737 4-Chlorophenol
6 2.512 1387522 196260 BV 0.106 0.1116 1-Decylamine
8 2.630 749162 97326 VB 0.113 0.0603 Methyl Caprate
7 4.120 2650317 232666 BV 0.177 0.2133 Tridecane
9 4.505 2331063 193209 VB 0.187 0.1876 Tetradecane
10 5.678 1879881 106858 BV 0.253 0.1513 Acenaphthylene
11 & 6.830 2807565 146329 VB 0.297 0.2259 1-Dodecanol
12 and
Pentadecane

Total area = 1.24275E+009

115
Exercise 2: Comparing Column Diameter and Resolution
Capillary Column Analysis: 30 M x 0.53 mm x 0.88 HP1

Capillary Column Analysis: 30 M x 0.53 mm x


0.88 HP1

1 2
9

12

7&8 11
10
6
4 5

Capillary Column Analysis: 30 M x 0.53 mm x 0.88 HP1


Column Flowrate = 2.6 mL/min Linear Velocity = 19.7 cm/sec
Isothermal 180 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 39.1
Area Percent Report

PK # Ret Time Area Height Type Width Area % Name


1 2.541 333138 156491 BV 0.033 0.7688 Pentane
2 2.711 4.24499E+007 1.81871+007 PV 0.036 97.9577 Solvent
4 3.438 27293 11612 BB 0.0630 0.0630 Undecane
5 3.743 44075 13854 BB 0.1017 0.1017 4-Chlorophenol
6 4.206 38866 12775 BB 0.0897 0.0897 1-Decylamine
7&8 4.670 71919 16254 BB 0.1660 0.1660 Tridecane
9 5.772 96872 25091 BB 0.2235 0.2235 Tetradecane
10 6.659 79214 16874 BV 0.1828 0.1828 Acenaphthylene
11 6.808 91419 16880 VB 0.2110 0.2110 1-Dodecanol
12 7.420 102251 20686 BB 0.2360 0.2360 Pentadecane

Total area = 4.33349E+007

116
Exercise 2: Comparing Column Diameter and Resolution
Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP5

Capillary Column Analysis: 30 M x 0.32 mm x 0.25


uM HP5

1 2 6 7 9
10 11 12
4 5

3
8

Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP5


Column Flowrate = 1 mL/min Linear Velocity = 20.5 cm/sec
Isothermal 160 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 2.442 62321 45133 BV 0.022 0.2127 Pentane
2 2.556 2.90451E+007 1.87002E+007 VV 0.024 99.1066 Solvent
3 2.672 8303 5036 VB 0.026 0.0283
4 3.100 10129 7143 BB 0.023 0.0346 Undecane
5 3.511 14978 7438 BB 0.032 0.0511 4-Chlorophenol
6 3.816 16871 9417 BB 0.028 0.0576 1-Decylamine
7 4.156 17124 9425 BB 0.029 0.0584 Tridecane
8 4.359 7748 3907 BB 0.031 0.0264 Methyl caprate
9 5.179 33486 14578 BB 0.036 0.1143 Tetradecane
10 6.315 27045 9249 BB 0.046 0.0923 Acenaphthylene
11 6.542 29540 8682 BB 0.053 0.1008 1-Dodecanol
12 6.789 34293 11407 BB 0.047 0.1170 Pentadecane

Total area = 2.9307E+007

117
Exercise 2: Comparing Column Diameter and Resolution
Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP5

118
Exercise 3: Comparing Column Length
and Resolution
Exercise 3: Comparing Column Length and Resolution
Worksheet

Worksheet
Using the following equation for determining resolution quantitatively, compare
three columns by calculating the "R" value for the separation of C-14 and
acenaphthlalene using the data from the next three pages:

RT1 = retention time of the first of the two


R = 1.18 (RT2) – RT1) peaks that one is calculating the
----------------------- resolution of
(W1 + W2)
RT2 = retention time of the second peak
W1 = peak width of first peak
W2 = peak width of second peak

10 meter column R = _______________

20 meter column R = _______________

30 meter column R = _______________

Was there an effect on the retention times of the components?

________________________________________________________________

Did doubling the length, double the resolution?

________________________________________________________________

Using the Bourbon Street Analogy, how would we explain the phenomenon of
increasing column lengths?

________________________________________________________________

________________________________________________________________

120
Exercise 3: Comparing Column Length and Resolution
Worksheet

Did the 10 meter column separate all of the peaks?

________________________________________________________________

121
Exercise 3: Comparing Column Length and Resolution
Capillary Analysis: 10 M x 0.53 mm x 0.88 uM HP1

Capillary Analysis: 10 M x 0.53 mm x 0.88 uM HP1

1
2 6
5 9
10
4 7&8 11 12

Column Flowrate = 1 mL/min Linear Velocity = 31.6 cm/sec


Isothermal 150 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 0.527 231973 506649 BV 0.011 0.4622 Pentane
2 0.582 4.91296E+007 4.85842E+007 HBAS 0.018 97.8832 Solvent
4 0.932 39379 45292 BV 0.014 0.0785 Undecane
5 1.080 61167 58218 BB 0.017 0.1219 4-Chlorophenol
6 1.334 78767 59735 BB 0.021 0.1569 1-Decylamine
7&8 1.617 104643 43750 BB 0.034 0.2085 Tridecane
9 2.294 140234 57536 BB 0.038 0.2794 Tetradecane
10 2.652 133393 50311 BB 0.042 0.2658 Acenaphthylene
11 2.844 123994 41236 BB 0.048 0.2470 1-Dodecanol
12 3.376 148918 39432 BB 0.059 0.2967 Pentadecane

Total area = 5.01921E+007

122
Exercise 3: Comparing Column Length and Resolution
Capillary Analysis: 20 M x 0.53 mm x 0.88 uM HP1

Capillary Analysis: 20 M x 0.53 mm x 0.88 uM HP1

1
2

9
5 6 10
4 7
11 12

Column Flowrate = 1 mL/min Linear Velocity = 33.2 cm/sec


Isothermal 150 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 1.005 93377 151911 BV 0.010 0.888 Pentane
2 1.112 1.04351E+008 7.90804E+007 BBAS 0.021 99.2401 Solvent
4 1.768 37032 31069 BV 0.019 00352 Undecane
5 2.056 54979 36934 BB 0.024 0.0523 4-Chlorophenol
6 2.556 15379 7446 BB 0.031 0.0146 1-Decylamine
7 3.063 66376 29476 BV 0.035 0.0631 Tridecane
8 3.126 31186 14012 VB 0.035 0.0297 Methy caprate
9 4.343 129261 38695 BB 0.052 0.1229 Tetradecane
10 5.018 124319 33885 BB 0.057 0.1182 Acenaphthylene
11 5.401 111190 26843 BB 0.065 0.1057 1-Dodecanol
12 6.391 135985 25968 BB 0.082 0.1293 Pentadecane

Total area = 1.0515E+008

123
Exercise 3: Comparing Column Length and Resolution
Capillary Column Analysis: 30 M x 0.53 mm x 0.88 uM HP1

Capillary Column Analysis: 30 M x 0.53 mm x 0.88


uM HP1

1
2

9
5 6
4 7 10
11 12

Column Flowrate = 1 mL/min Linear Velocity = 34.2 cm/sec


Isothermal 150 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 1.460 115928 132851 BBA 0.014 0.1455 Pentane
2 1.620 7.90305E+007 5.3877E+007 HBAS 0.022 99.2190 Solvent
4 2.573 25248 16312 BB 0.024 0.0317 Undecane
5 2.984 37198 17834 BB 0.033 0.0467 4-Chlorophenol
6 3.682 45920 18003 BB 0.039 0.0577 1-Decylamine
7 4.459 44286 15239 BV 0.045 0.0556 Tridecane
8 4.54. 21118 7177 VB 0.046 0.0265 Methy caprate
9 6.325 85934 20143 BB 0.067 0.1079 Tetradecane
10 7.309 85880 17159 BB 0.079 0.1078 Acenaphthylene
11 7.839 71163 13334 BB 0.083 0.0893 1-Dodecanol
12 9.310 89398 13576 BB 0.103 0.1122 Pentadecane

124
Exercise 4: Comparing Column Film
Thickness and Resolution
Exercise 4: Comparing Column Film Thickness and Resolution
Worksheet

Worksheet

Using the following equation for determining resolution quantitatively, compare


two columns by calculating the "R" value for the separation of C-14 and
acenaphthlalene using the data shown on the next two pages:

RT1 = retention time of the first of the two


peaks that one is calculating the
R = 1.18 (RT2) – RT1) resolution of
----------------------- RT2 = retention time of the second peak
(W1 + W2)
W1 = peak width of first peak
W2 = peak width of second peak

0.88 uM film column R = _______________

2.65 uM film column R = _______________

Was there an effect on the retention times of the components?

________________________________________________________________

Did tripling the film thickness triple the resolution?

________________________________________________________________

Using the Bourbon Street Analogy, how would we explain the phenonmenon of
increasing column film thickness?

________________________________________________________________

________________________________________________________________

126
Exercise 4: Comparing Column Film Thickness and Resolution
Worksheet

________________________________________________________________

Did the columns separate all of the peaks?

________________________________________________________________

127
Exercise 4: Comparing Column Film Thickness and Resolution
Capillary Analysis: 10 M x 0.53 mm x 0.88 uM HP1

Capillary Analysis: 10 M x 0.53 mm x 0.88 uM HP1

1
2 6 9
5 10
7&8
4 11 12

Column Flowrate = 1 mL/min Linear Velocity = 48 cm/sec


Isothermal 150 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 0.347 121201 360718 BV 0.006 0.2766 Pentane
2 0.382 4.25931E+007 4.41108E+007 HBAS 0.017 97.2182 Solvent
4 0.615 55319 76048 BV 0.012 0.1263 Undecane
5 0.713 76636 94349 BV 0.014 0.1749 4-Chlorophenol
6 0.882 100878 93840 BV 0.017 0.2303 1-Decylamine
7&8 1.069 145694 78145 BB 0.030 0.3325 Tridecane
9 1.515 190617 100550 BB 0.030 0.4351 Tetradecane
10 1.751 166490 86433 BV 0.030 0.3800 Acenaphthylene
11 1.878 161632 70438 PV 0.036 0.3689 1-Dodecanol
12 2.229 200301 68188 BB 0.046 0.4572 Pentadecane

Total area = 4.38119E+007

128
Exercise 4: Comparing Column Film Thickness and Resolution
Capillary Analysis: 10 M x 0.53 mm x 2.65 uM HP1

Capillary Analysis: 10 M x 0.53 mm x 2.65 uM HP1

1 2
5
6
9 10
4
7&8 11
12

Column Flowrate = 1 mL/min Linear Velocity = 40 cm/sec


Isothermal 150 degrees FID
1 uL HP Column Checkout Sample at Split Ratio = 100:1
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 0.418 164606 376839 BV 0.007 0.3347 Pentane
2 0.542 4.81769E+007 3.22537E+007 BBA 0.026 97.9559 Solvent
4 1.297 41141 23557 BB 0.028 0.0837 Undecane
5 1.617 62624 32634 BB 0.030 0.1273 4-Chlorophenol
6 2.168 79198 27992 BB 0.044 0.1610 1-Decylamine
7&8 2.788 106786 19792 BB 0.077 0.2171 Tridecane
9 4.261 141902 25125 BB 0.089 0.2885 Tetradecane
10 5.036 136041 23624 BB 0.090 0.2766 Acenaphthylene
11 5.456 124564 17972 BB 0.109 0.2566 1-Dodecanol
12 6.618 148478 16669 BB 0.139 0.3019 Pentadecane

Total area = 4.91823E+007

129
Exercise 4: Comparing Column Film Thickness and Resolution
Capillary Analysis: 10 M x 0.53 mm x 2.65 uM HP1

130
Exercise 5: Calculating the Split Ratio
Exercise 5: Calculating the Split Ratio
Worksheet

Worksheet
Using data from the chromatograms on the following pages and the given
formulas,complete the table below:

Column Flowrate (mL/min) = π r2 ū 60


R = Column radius in cm.
ū = Average linear velocity in cm./sec.

col length (cm.)


ū = -----------------------------------------------
R.T. (sec) of non-retained species

Split Vent Flowrate + Column Flowrate


Split Ratio = -------------------------------------------------------
Column Flowrate

Split Vent Column Split Area Count


Flow Flowrate Ratio of C-14

44.8

101.7

220

132
Exercise 5: Calculating the Split Ratio
Split Vent Flow = 44.8 mL/min

Split Vent Flow = 44.8 mL/min


Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP 5

12
1
2 9
11
10

7
6
5
4
8
3

Column Head Pressure: 12 psi


Temperature Program: Init. Temp: 110 Initial Time: 3 min . Rate: 10
Final Temp: 180 Final Temp: 7 min.
FID: 1 uL of Column Checkout Sample
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 1.591 127589 181613 BV 0.012 0.1952 Pentane
2 1.844 6.40112E+007 4.70123E+007 BV 0.023 97.9545 Solvent
3 2.083 20660 23683 BV 0.018 0.0316 Impurity
4 3.599 56809 41836 BB 0.022 0.0869 Undecane
5 4.782 81723 50109 BB 0.026 0.1251 4-Chlorophenol
6 5.526 96577 61596 BB 0.024 0.1478 1-Decylamine
7 6.233 103810 66682 BB 0.025 0.1589 Tridecane
8 6.570 50263 33953 BB 0.024 0.0769 Methyl caprate
9 7.614 203872 122264 BB 0.026 0.3120 Tetradecane
10 8.432 187396 107906 BB 0.027 0.2868 Acenaphthylene
11 8.586 174116 10509 BB 0.025 0.2664 1-Dodecanol
12 8.913 217376 127896 BB 0.025 0.3326 Pentadecane

Total area = 6.53479E+007

133
Exercise 5: Calculating the Split Ratio
Split Vent Flow = 101.7 mL/min

Split Vent Flow = 101.7 mL/min


Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP 5

1
2 9 12
11

10

7
5 6
4
8
3

Column Head Pressure: 12 psi


Temperature Program: Init. Temp: 110 Initial Time: 3 min Rate: 10
Final Temp: 180 Final Time: 7 min
FID: 1 uL of Column Checkout Sample
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 1.593 83755 121735 BV 0.011 0.2454 Pentane
2 1.846 3.35316E+007 3.21424E+007 BV 0.017 98.2293 Solvent
3 2.087 6187 8569 BV 0.016 0.0181 Impurity
4 3.614 25332 18560 BB 0.022 0.0742 Undecane
5 4.800 37070 23577 BB 0.025 0.1086 4-Chlorophenol
6 5.553 40633 24559 BB 0.025 0.1190 1-Decylamine
7 6.254 45626 29838 BB 0.024 0.1337 Tridecane
8 6.593 22138 14686 BB 0.024 0.0649 Methyl caprate
9 7.631 88830 57276 BB 0.024 0.2602 Tetradecane
10 8.448 84758 52205 BB 0.026 0.2483 Acenaphthylene
11 8.601 76148 49580 BB 0.024 0.2231 1-Dodecanol
12 8.928 93985 59704 BB 0.025 0.2753 Pentadecane

Total area = 3.41361E+007

134
Exercise 5: Calculating the Split Ratio
Split Vent Flow = 220 mL/min

Split Vent Flow = 220 mL/min


Capillary Column Analysis: 30 M x 0.32 mm x 0.25 uM HP 5

1
2 11
9

12
10
7
5 6
4 8
3

Column Head Pressure: 12 psi


Temperature Program: Init. Temp: 110 Initial Time: 3 min Rate: 10
Final Temp: 180 Final Time: 7 min
FID: 1 uL of Column Checkout Sample
Area Percent Report
PK # Ret Time Area Height Type Width Area % Name
1 1.599 31703 46779 BV 0.011 0.1763 Pentane
2 1.852 1.77427E+007 1.94882E+007 BV 0.026 0.026 Solvent
3 2.093 2856 3951 BV 0.016 0.0159 Impurity
4 3.619 10121 7515 BB 0.021 0.0563 Undecane
5 4.800 16549 10729 BB 0.024 0.0920 4-Chlorophenol
6 5.560 15419 8910 BB 0.026 0.0858 1-Decylamine
7 6.251 17464 11681 BB 0.024 0.0971 Tridecane
8 6.591 8721 5786 BB 0.024 0.0485 Methyl caprate
9 7.623 33334 21782 BB 0.024 0.1854 Tetradecane
10 8.439 36127 22379 BB 0.026 0.2009 Acenaphthylene
11 8.591 29029 19256 BB 0.024 0.1615 1-Dodecanol
12 8.918 34741 22514 BB 0.024 0.1932 Pentadecane

Total area = 1.79788E+007

135
Exercise 5: Calculating the Split Ratio
Split Vent Flow = 220 mL/min

136
Exercise 6: Split and Splitless Injection
Exercise 6: Split and Splitless Injection
Worksheet

Worksheet
6) How could you improve the sensitivity, assuming that the Integration
parameters have been set for maximum sensitivity?

7) You want to decrease the analysis runtime while maintaining the separation
of components 1 and 2. Comment on the validity of each of the following
approaches.

a. Increase the initial temperature.

b. Shorten the column to 25 m.

138
Exercise 6: Split and Splitless Injection
Worksheet

c. Decrease the “purge on” time.

d. Use a solvent with a higher boiling point.

8) How would you improve the baseline and reduce the solvent tailing for the
following:

139
Exercise 6: Split and Splitless Injection
Worksheet

140
Appendix
Appendix
References

References

References

• High Resolution Gas Chromatography


Third Edition, K.J. Hyver, Editor
Hewlett-Packard 1989 PN 5950-3562
• GC Inlets – An Introduction
Mathew Klee
Hewlett-Packard 1990 PN 5958-9468
• Detectors for Gas Chromatography – A Practical Primer
Rosemary Buffington, Mike Wilson
Hewlett Packard 1987 PN 5958-9433
• Electronic Pressure Control in Gas Chromatography
Sally Stafford
Hewlett-Packard 1993 PN 5182-0842

142
Appendix
Application Notes

Application Notes

Application Notes

Sorted by Industry:
Fuels 1

Chemicals/Polymers 4

Pharmaceutical 5

Toxicology/Forensics 6

Environmental 7

Foods/Favor 8

Bioscience 10

Figure 97

143
Appendix
Application Notes

Application Notes

144
Appendix
Application Notes

Application Notes

145
Appendix
Application Notes

Application Notes

146
Appendix
Application Notes

Application Notes

147
Appendix
Application Notes

Application Notes

148
Appendix
Application Notes

Application Notes

149
Appendix
Application Notes

Application Notes

10

150
Appendix
Application Notes

Application Notes

11

151
Appendix
Application Notes

Application Notes

12

152
Appendix
Application Notes

Application Notes

13

153
Appendix
Bibliography

Bibliography

Bibliography

14

154
Appendix
Bibliography

Bibliography

15

155
Appendix
Bibliography

Bibliography

16

156
Appendix
Bibliography

Bibliography

17

157
Appendix
Bibliography

Bibliography

18

158
Appendix
Bibliography

Bibliography

19

159
Appendix
Bibliography

Bibliography

20

160
Appendix
Exercise Answers

Exercise Answers

Exercise 1 - Determining Efficiency

col length (cm)


A. Linear velocity, u = --------------------------------------------
R.T.(ser) of non-retained species

3000 cm
u = ------------------------------ = 33.8 cm/sec
1.477 min x 60 sec/,in

Flow Rate = π r 2 u 60 d=
0.53m
=
0.053cm
= (3.14) (0.0265 cm) 2 (33.8 cm) (60 sec)
r = 0.0265cm
----------- ---------
sec min
= 4.47 ml/min

L (mm) t R1
HETP = ------------ n = 5.545 ( --------------) 2
n wh

30,000 4.970 - 1.477


= ----------- n = 5.545 ( -------------------) 2
26,011 0.051

= 1.15 n = 26,011

n 26,011
Plates/meter = ------ = -----------
L 30

= 867

B. 33.8 , 4.47 , 1.15 , 867

161
Appendix
Exercise Answers

C.
4


3.5

0.53 Column

2.5


2 Part D

1.5

• 0.32 Column



1


• •
0.5
• •

0 20 40 60 80 100 120 140


E. The 0.32 column is more efficient since it has a lower HETP
minimum.

162
Appendix
Exercise Answers

Exercise 2 - Comparing Column Diameter and


Resolution

1.18 (RT2 - RT1)


R= ---------------------------
(W1 + W2)

1.18 (5.678 - 4.505)


Packed Column: R = -----------------------------
(0.187 + 0.253)

1.18 (1.173)
R = -------------------- = 3.15
0.44

1.18 (6.659 - 5.772)


0.53 Column: R = -----------------------------
(0.2235 + 0.1828)

1.18 (0.887)
R = -------------------- = 2.58
0.4063

1.18 (6.315 - 5.179)


0.32 Column: R = -----------------------------
(0.036 + 0.046)

1.18 (1.136)
R = -------------------- = 16.3
0.082

163
Appendix
Exercise Answers

Exercise 3 - Comparing Column Length and Resolution

1.18 (RT2 - RT1)


R= ---------------------------
(W1 + W2)

1.18 (2.652 - 2.294)


10 Meter Column: R = -----------------------------
(0.038 + 0.042)

1.18 (0.358)
R = -------------------- = 5.28
0.08

1.18 (5.018 - 4.343)


20 Meter Column: R = -----------------------------
(0.052 + 0.057)

1.18 (0.675)
R = -------------------- = 7.31
0.109

1.18 (7.308 - 6.325)


30 Meter Column: R = -----------------------------
(0.036 + 0.046)

1.18 (.983)
R = ----------------------------- = 7.94
0.146

164
Appendix
Exercise Answers

Exercise 4 - Comparing Film Thickness and Resolution

1.18 (RT2 - RT1)


R= ---------------------------
(W1 + W2)

1.18 (1.751 - 1.515)


0.88 uM Film: R = -----------------------------
(0.03 + 0.03)

1.18 (0.236)
R = -------------------- = 4.64
0.06

1.18 (5.036 - 4.261)


2.65 uM Film: R = -----------------------------
(0.089 + 0.090)

1.18 (0.775)
R = -------------------- = 5.11
0.179

165
Appendix
Exercise Answers

Exercise 5 - Calculating the Split Ratio

Split Vent Flow = 44.8

Col Flow = π r2 u 60 (3000) (60)


2
= (3.14) (0.032/2) -------------------
(1.591) (60)
= 1.52 mL/min

44.8 + 1.52
Split Ratio = ------------------ = 30.5
1.52

Area Count = 203,872

Split Vent Flow = 101.7

(3000)
Col Flow = (3.14) (0.016) 2 ------------ = 1.52 mL/min
1.593
= 1.52 mL/min

101.7 + 1.52
Slit Ratio = ------------------ = 67.9
1.52

Area Count = 88,830

Split Vent Flow = 220

(3000)
2
Col Flow = (3.14) (0.016) ------------ = 1.51 mL/min
1.599
= 1.52 mL/min

220 + 1.51
Split Ratio = ------------------ = 146.7
1.51

Area Count = 33,334

166
Appendix
Exercise Answers

Exercise 6 - Split Injection and Splitless Injection

1. Decrease the split ratio. Use a more sensitive detector.

2. A. A good approach if the higher temp will still allow


solvent focusing.
B. May cause the peaks of interest to merge.
C. Decreasing the purge on time may result in loss of
sample and thus, loss of sensitivity.
D. Would allow starting at a higher initial oven temperature.
Probably the best approach.

3. Perhaps hold the purge off longer or increase the purge on


time.
Perhaps use a solvent with a lower boiling point so that a
lower initial oven temp could be used.

167
Appendix
Exercise Answers

168

Anda mungkin juga menyukai