Anda di halaman 1dari 10

Biomatrix/Polymer Composite Material for Heart

Valve Tissue Engineering


CARDIOVASCULAR

Christof Stamm, MD, Amir Khosravi, MD, Niels Grabow, MS, Kathleen Schmohl, PhD,
Nadine Treckmann, BS, Anne Drechsel, BS, Ma Nan, PhD, Klaus-Peter Schmitz, PhD,
Axel Haubold, PhD, and Gustav Steinhoff, MD
Department of Cardiac Surgery, Institute for Biomedical Engineering, Research Center for Cardiac Tissue Replacement, University
of Rostock, Rostock, Germany

Background. Decellularized extracellular matrix has Results. Biocompatibility assays indicated that human
been suggested as a scaffold for heart valve tissue engi- blood vessel cells survive and proliferate on matrix/
neering or direct implantation. However, cell removal polymer hybrid tissue. In vitro activation of cellular and
impairs the physical properties of the valve structure and plasmatic coagulation cascades was lower than with
exposes bare collagen fibers that are highly thrombo- uncoated control tissue. After implantation in the rabbit
genic. Matrix/polymer hybrid valves with improved bio- aorta, matrix/polymer hybrid patches healed well, with
logical and mechanical characteristics may be complete endothelialization, mild leukocyte infiltration,
advantageous. and less calcification than control tissue. Matrix/polymer
Methods. Porcine aortic valves were decellularized hybrid tissue had superior tensile strength and suture
enzymatically and impregnated with biodegradable retention strength, and hybrid valves showed good fluid
poly(hydroxybutyrate) by a stepwise solvent exchange dynamic performance. The two valves in aortic position
process. Biocompatibility was tested in vitro using cell performed well, with complete endothelialization and
proliferation and coagulation assays. Proinflammatory limited inflammatory cell invasion after 12 weeks. Of the
activity was assessed in vivo by implantation of matrix/ two valves in pulmonary position, one failed.
polymer patches in the rabbit aorta. Biomechanic valve Conclusions. Matrix/polymer hybrid tissue valves have
properties and fluid dynamics were tested in a pressure/ good biological and biomechanic characteristics and may
flow-controlled pulse duplicating system. Matrix/ provide superior replacement valves.
polymer hybrid valves were implanted in pulmonary (Ann Thorac Surg 2004;78:2084 –93)
and aortic position in sheep. © 2004 by The Society of Thoracic Surgeons

T he currently available tissue valve prostheses based


on aldehyde-fixed xenogenic tissue are inevitably sub-
ject to calcium phosphate deposition and degeneration. The
when cells are removed and the tertiary structure of fibrous
valve tissue constituents is altered during the decellulariza-
tion process; second, open collagen surfaces are highly
tanning process effectively devitalizes the native cell popu- thrombogenic, because collagen directly induces platelet
lation, denaturizes antigenic protein domains, and changes activation as well as coagulation factor XII. To address these
the scaffold protein architecture rendering in vivo repopu- issues, biomaterial/polymer composite materials based on
lation with recipient cells impossible. Furthermore, none of decellularized vascular matrix scaffolds that are coated with
the currently available heart valve prostheses has the po- biodegradable polymers were developed; the hypothesis
tential for growth, limiting their use in infants and children. that such hybrid tissues exhibit improved biocompatibility
Goal of the current heart valve tissue engineering efforts is in vitro and in vivo was tested.
therefore the development of a valve prosthesis that com-
bines unlimited durability with physiologic blood flow
pattern and biologically inert surface properties [1–3]. Material and Methods
Recently, extracellular heart valve matrix was suggested as
Animals received humane care in compliance with the
a scaffold for tissue engineering, providing the natural valve
architecture and ideal conditions for repopulation with Guide for the Care and Use of Laboratory Animals prepared
recipient cells [4]. However, there are at least two major by the National Academy of Sciences and published by
problems: first, the mechanical tissue properties deteriorate the National Institutes of Health (National Institutes of
Health Publication No 86 to 23, revised 1996). The proto-
Accepted for publication March 25, 2004. col was reviewed and approved by the local animal care
Presented at the Fortieth Annual Meeting of The Society of Thoracic committees.
Surgeons, San Antonio, TX, Jan 26 –28, 2004.
Decellularization
Address reprint requests to Dr Steinhoff, Department of Cardiac Surgery,
University of Rostock, Schillingallee 35, D-18057 Rostock, Germany; Enzymatic removal of cells without altering the biochem-
e-mail: gustav.steinhoff@med.uni-rostock.de. ical characteristics of the extracellular matrix by exposure

© 2004 by The Society of Thoracic Surgeons 0003-4975/04/$30.00


Published by Elsevier Inc doi:10.1016/j.athoracsur.2004.03.106
Ann Thorac Surg STAMM ET AL 2085
2004;78:2084 –93 BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR

to chemical fixatives was performed as previously de- fibroblasts migrate onto the dish surface, attach, and
scribed [1]. Briefly, hearts were harvested from porcine proliferate. The MTS-tests were performed in decellular-
cadavers (courtesy of Agricultural Research Center, ized matrix treated with various polymer preparations,
Dummerstorf, Germany). The aortic root was prepared, but also with pure polymer samples as well as with
washed in phosphate buffered saline (PBS), and incu- hydrid tissues following several sterilization and storage

CARDIOVASCULAR
bated in 0.05% trypsin solution for 48 hours at 37°C, protocols such as lyophilization, plasmasterilization,
followed by 3 washing steps for 1 hour each. For longer FAD-sterilization, ethylene oxide sterilization, and
storage, the specimens were lyophilized (at ⫺40° C and gamma sterilization (data not shown).
0.05 mbar) and rehydrated before further use, but for
implantation in sheep the valves were processed imme- In Vitro Hemocompatibility
diately. Biocompatibility tests in vitro and in rabbits were Complement and coagulation system activation in re-
performed using human aortic wall tissue, which was sponse to different hybrid tissue preparations were stud-
obtained during routine coronary artery bypass grafting ied in several in vitro assays. Activation of complement
(CABG) operations and was processed in identical factor C3 was assessed by ELISA for C3a-des-Arg, the
fashion. stable metabolite of activated C3 (Progen, Heidelberg,
Germany), following incubation of human plasma with
Polymer Coating hybrid tissue samples for 60 minutes. Representative for
Biodegradable polymers were supplied by Tepha Inc. activation of the plasmatic clotting system, the concen-
(Cambridge, MA). Poly(3-hydroxybutyrate) (P3HB), tration of the prothrombin fragments F1 and F2 was
poly(4-hydroxybutyrate) (P4HB), and poly(3- hydroxybu- measured by ELISA (Behring, Marburg, Germany), again
tyrate-co-4-hydroxybutyrate) (P3/4HB) in powder form after incubation of human plasma with hybrid tissue.
were dissolved in chloroform at 50°C according to the Finally, the response of the cellular clotting system to
chosen concentration. Initially, a dip-coating process was biomatrix/polymer hybrid tissue was studied by measur-
used for fabrication of biomatrix/polymer hybrid tissue: ing platelet factor 4 in human plasma using the Asserach-
Decellularized and lyophilized tissue was repeatedly romPF4 assay system.
immersed in 2% to 6% (w/v) polymer solution followed
by solvent evaporation for two weeks until the chloro- In Vivo Screening Tests
form content was less than 0.2%. Before further use, the Intravascular biocompatibility of various hybrid tissue
tissue was rehydrated in cell culture medium (in vitro preparations was tested in a rabbit model. Adult New
experiments) or phosphate buffered saline (in vivo exper- Zealand White rabbits were anesthetized, heparinized,
iments). When it became evident that this coating may intubated, and ventilated. The abdomen was opened, the
not withstand systemic hemodynamic forces, the proto- abdominal aorta was dissected distal to the renal arteries,
col was switched to a modified polymer impregnation clamped, and incised longitudinally. A patch of bioma-
process: freshly decellularized valve tissue was subjected trix/polymer hybrid tissue measuring approximately 5⫻3
to a wet dehydration process, replacing water with eth- mm was sutured in place using nonresorbable suture.
anol. Then, the specimens were immersed in 1% (w/v) After 1, 3, or 6 months, the animals were sacrificed and
polymer solution for 30 minutes, followed by rehydration the aortic segment containing the patch was explanted
and wet solvent elimination in PBS. and prepared for histology. Sections were stained with
H&E or antibodies for immunohistology, and examined
In Vitro Cell Proliferation by light microscopy. A scoring system was designed to
Whether biomatrix/polymer hybrid valve tissue can be facilitate comparison of histologic findings. The following
repopulated with blood vessel cells, without exhibiting histologic characteristics were studied: endothelialization
cytotoxicity, was tested in series of in vitro experiments. of the luminal patch surface, intima proliferation, inflam-
Tissue samples were prepared as described above, matory infiltration, calcification, cellular migration into
seeded with L929 mouse fibroblasts and incubated under the patch material, thrombus formation, and formation of
standard cell culture conditions for at least 72 hours. Cell a neo-elastica interna (Table 1).
viability was then assessed using the CellTiter96 fluores-
cent cell proliferation assay (MTS test). Because cellular Mechanical Testing
adhesion, proliferation, metabolic activity, and resistance The biomechanical characteristics of various biomatrix/
to toxin are highly dependent on species and cell type, polymer hybrid tissue preparations were extensively
similar tests were performed using a mixed population of evaluated and are described in detail elsewhere [5]. In
human vascular myofibroblasts and endothelial cells, that study, suture retention strength of the aortic conduit
prepared from saphenous vein samples of CABG pa- wall, as well as tensile strength and elastic properties of
tients. The lumen was filled with collagenase A contain- the leaflets were measured. The fluid dynamic properties
ing medium. After incubation for 20 minutes, detached of intact hybrid valves were tested in a pulse-duplicating
endothelial cells were flushed out and cultivated under system according to ISO 5940 standards. High-resolution
standard conditions. The remaining tissue was minced videomorphometric analysis of leaflet motion was
and placed in smooth muscle cell growth medium. Myo- performed.
2086 STAMM ET AL Ann Thorac Surg
BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR 2004;78:2084 –93
CARDIOVASCULAR

Fig 1. Decellularization and polymer coating of porcine aortic valves. (A and B) Hematoxylin & eosin stain of the aortic media (A) before and
(B) after enzymatic decellularization demonstrates complete removal of all cellular material. (C–E) Electron microscopy of the luminal valve
surface (C) before decellularization, (D) after decellularization, and (E) after polymer penetration (here: P3/4HB). Note that the modified poly-
mer penetration process leaves the surface porous, facilitating recipient cell adhesion and penetration.

Large Animal Implantation underwent echocardiography and cardiac catheterization


Once the optimal biomatrix/polymer tissue composition and were sacrificed. Morphologic analysis of the hybrid
had been chosen, four porcine hybrid valves were im- valve was carried out by the local pathologist before the
planted in sheep by Experimental Surgical Services valve tissue was divided in three parts and prepared for
(ESS), University of Minnesota. Following approval by further analysis by light microscopy, immunohistology,
the local regulatory bodies, two valves were implanted to and electron microscopy.
replace the native pulmonary root, and three valves were
implanted as freestanding aortic root replacement with Histology
coronary reimplantation. One animal died while under- Hematoxylin & eosin, Mason’s trichrome, and Van Gie-
going aortic valve surgery and was excluded from the son staining of formalin-fixed paraffin-embedded tissue
analysis. Three months after implantation, the animals was performed in standard fashion. For immunohistol-
Ann Thorac Surg STAMM ET AL 2087
2004;78:2084 –93 BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR

CARDIOVASCULAR
Fig 2. Cell proliferation on biomatrix/polymer hybrid tissue in vitro
(MTS test). (A) Cultivated L929 mouse fibroblasts; (B) mixed popu-
lation of human myofibroblasts and endothelial cells. Although
mouse fibroblasts appear to proliferate on P4HB and P3/4HB only,
with virtually no cell growth on P3HB, there is good proliferation of
human cells on all tested polymers. (OD ⫽ optical density; P3HB ⫽
poly[3-hydroxybutyrate]; P4HB ⫽ poly[4-hydroxybutyrate]; P3/4HB
⫽ poly[3-hydroxybutyrate-co-4-hydroxybutyrate].)

ogy, frozen sections were prepared from cryopreserved


tissue and incubated with monoclonal mouse antihuman
CD31 antibody (Dako, clone Nr. JC70A) or monoclonal
mouse antihuman smooth muscle actin antibody (Dako,
Fig 3. Activation of the cellular clotting system, plasmatic clotting sys-
clone Nr. 1A4), followed by detection with peroxidase-
tem, and complement system in human plasma incubated with decellu-
conjugated goat antimouse IgG secondary antibody larized matrix and biomatrix/polymer hybrid tissue assessed by ELISA
(Dako). For analysis of surface morphology and ultra- for (A) platelet factor 4, (B) prothrombin fragments F1 and F2 (pro-
structure, specimens were prepared for scanning elec- thrombin), and (C) C3a-des-Arg. Shown are representative sets of ex-
tron microscopy and viewed using a Philips XL30ESEM periments using the same plasma sample. In the xenogenic setting (por-
(FEI, Hillsboro, OR) electron microscope. cine matrix incubated with human plasma) polymer coating of
decellularized matrix attenuated the activation of both cellular and plas-
matic clotting system, while complement C3 activation remained un-
Results changed. (P3HB ⫽ poly[3-hydroxybutyrate]; P4HB ⫽ poly[4-hydroxy-
butyrate]; P3/4HB ⫽ poly[3-hydroxybutyrate-co-4-hydroxybutyrate].)
The decellularization process removed all cellular com-
ponents of the porcine aortic valve leaflets and the aortic
wall (Figs 1A and 1B). After enzymatic removal of the proteoglycans forms the luminal surface of the valve
endothelial cell layer and cellular components of the scaffold (Figs 1C and 1D). As has been previously re-
media, the fibrous network of collagen and elastin, and ported, the enzymatic digestion process inevitably weak-
2088 STAMM ET AL Ann Thorac Surg
BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR 2004;78:2084 –93
CARDIOVASCULAR

Fig 4. (A) Explanted abdominal aorta of a rabbit 12 weeks after implantation of a patch consisting of decellularized human aortic wall seen
from the adventitial aspect. The arrow indicates the patch. (B–D) Photomicrographs of explanted specimens 12 weeks after implantation (he-
matoxylin & eosin staining; original magnification 100⫻). The solid arrows indicate patch material; interrupted arrows indicate neoelastica
interna. (B) Decellularized uncoated matrix is clearly distinct from the native aortic tissue with no signs of integration or resorption. There are
no regions of calcification, moderate inflammatory infiltration, and formation of a thick neoelastica interna. (C) Matrix coated with P4HB. The
patch is partially resorbed; the remaining material is heavily calcified. There is formation of a neoelastica interna as well as significant adven-
titial thickening. (D) P3HB-coated hybrid tissue is partially reabsorbed and well integrated in the native aortic tissue. There is little inflamma-
tory infiltration, thin neoelastica interna, and near-normal adventitial tissue.

ens the mechanical characteristics of the valve, so that it strength. Furthermore, the fluid dynamic and morpho-
may not be able to withstand hemodynamic forces in the metric characteristics of polymer-impregnated valves re-
systemic circulation [5]. The polymer impregnation pro- sembled those of native heart valves. As opposed to the
cess, however, led to improved biomechanic properties initially used dip-coating procedure, the modified poly-
in terms of suture retention strength and tensile tissue mer penetration protocol did not alter the surface mor-
phology of the decellularized matrix, preserving the
tissue native texture and microporosity, thus facilitating
Table 1. Morphologic Analysis of Biomatrix/Polymer Hybrid
Material in Vivo (Rabbit Model) adhesion and migration of recipient cells (Fig 1E).

Uncoated P3HB P4HB In Vitro Experiments


Endothelialization 2.5 (2–3) 1.5 (1–2) 2.5 (1–3) In a first series of screening tests, the proliferation
Intima proliferation 2 (1–3) 1.75 (1–2) 1.67 (1–2) capacity of L929 mouse fibroblasts on different biomatrix/
Inflammatory infiltration 2 (2) 1.5 (1–2) 1.67 (1–2) polymer preparations, with and without lyophilization
Calcification 2.75 (2–3) 2.5 (2–3) 1.67 (1–2) and sterilization was assessed. A summary of the cell
Cellular migration 2 (2) 1.75 (1–2) 1.5 (1–2) proliferation data by MTS test in biomatrix coated with
Thrombus formation 3 (3) 3 (3) 2.67 (1,3) P3HB, P4HB, and P3/4HB is shown in Figure 2A. It
Neo-elastica interna 1 (1) 1 (1) 1 (1) appears that P3HB almost completely inhibits cell growth
on the matrix. These experiments were then repeated
Histologic scores of uncoated decellularized valve tissue (uncoated) and
biomaterial/polymer composite tissue (P3HB, P4HB) 3 months after using a mixed population of human endothelial cells and
implantation in the rabbit aorta. Note that the experiments were per- myofibroblasts. It was found that they proliferate very
formed using the polymer dip-coating process. Shown is the arithmetic
mean of the scores in each group and the range of scores assigned in a
well on all matrix/polymer combinations (Fig 2B). Hence,
group (in parenthesis). A scoring system of 1–3 was devised and a specific decellularized biomatrix/polymer hybrid tissue has the
definition assigned to every for each parameter. In general, a score of 3 potential for repopulation even in the xenogenic setting.
represents a favorable result (ie, no calcification, complete endotheliali-
zation), whereas a score of 1 describes an unfavorable result (ie, complete Representative data on hemocompatibility of hybrid tis-
thrombotic occlusion, absence of a neo-elastica interna). sue are shown in Figure 3. Matrix impregnation with
P3HB ⫽ poly(3-hydroxybutyrate); P4HB ⫽ poly(4-hydroxybutyrate). P3HB attenuated the activation of platelet factor 4 in
Ann Thorac Surg STAMM ET AL 2089
2004;78:2084 –93 BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR

Fig 5. Gross morphology of the two por-


cine biomatrix/polymer hybrid valves 12
weeks after implantation in pulmonary
position in sheep. (A) The first valve was
completely degenerated probably due to

CARDIOVASCULAR
bacterial endocarditis. (B) The second
valve was in excellent condition, with
near-normal gross morphology.

human plasma, indicating less activation of the cellular biocompatibility but is rather stiff and brittle, while
clotting system (Fig 3A). With P3/4HB copolymer there P4HB-coated matrix is soft and pliable, but probably at
was still some attenuation of PF4 release, while P4HB the cost of earlier calcification in vivo. Based on these
alone appeared not to change the matrix-induced PF4 results a copolymer consisting of 82% P3HB and 18%
activation. Activation of the plasmatic clotting cascade as P4HB processed in 1% solution with freshly decellular-
assessed by the concentration of prothrombin fragments ized matrix was chosen for further testing in a large
F1 and F2 was partially suppressed when the matrix was animal model.
coated with biodegradable polymer, irrespective of the The first valve implanted in pulmonary position was
PHB-type used (Fig 3B). Finally, activation of the comple- severely obliterated, with severe diffuse pio-granulo-
ment system in response to hybrid tissue was estimated. matous valvular endocarditis, chronic lympho-histocytic
Again an attenuation of C3a-des-Arg production in the and necrotizing bioprosthetic periarteritis, and intimal
presence of polymer-coated matrix compared to un- fibrous-hyperplasia. Although microbiology at the time
treated decellularized matrix was observed (Fig 3C). of sacrifice was negative, bacterial endocarditis was prob-
ably the cause (Fig 5A). The second valve was in excellent
In Vivo Screening Tests condition, with near-normal gross morphology (Fig 5B).
In the rabbit abdominal aorta patch implantation model, On histology, there was complete endothelial cell lining
various combinations of polymer-matrix composites, but and moderate multifocal granulomatous inflammation
also noncoated xenogenic matrix patches were evaluated limited to the leaflet hinge point (data not shown).
(n ⫽ 3, each). Representative photomicrographs are Particular attention was paid to the two valves im-
shown in Figure 4, and part of the qualitative histologic planted in aortic position. At the time of sacrifice, both
evaluation data are summarized in Table 1. At follow-up, animals were in good condition without clinical signs of
all rabbit aortas were patent and free from blood clot valve dysfunction. Gross morphology and representative
formation, except in one animal with a P4HB-treated photomicrographs are shown in Figure 6. The pathologist
patch. There were no patch aneurysms despite implan- described that there were focal fibrinous deposits on the
tation in the high-pressure system. There was some early leaflets. The leaflets contained homogeneous eosino-
inflammatory cell infiltration (4 weeks) that later resolved philic bundle of collagens, expanded by moderate mul-
in all patches (12 weeks). Recipient blood vessel cells had tifocal granulomatous inflammation (epitheloid macro-
migrated into to patch material in both P3HB and P4HB phages and giant multinucleated cells). As seen in Figure
coated matrix. Importantly, very little or no calcification 6B, there was some intimal thickening particularly of the
occurred in the decellularized and P3HB-coated matrices, luminal surface of the otherwise delicate leaflets. By
in contrast to autologous blood vessel control patches immunohistology, complete endothelial cell lining of the
and some P4HB-coated matrices. There was some degree leaflets and the conduit wall was found. There was
of intimal thickening with formation of a neo-elastica smooth muscle cell migration into the media of the
interna in all hybrid tissue patches, but the vessel lumen leaflets (Fig 7), but very little cellular infiltration of the
was not narrowed in any of the P3HB-coated patches. As conduit wall (Fig 6E). By scanning electron microscopy,
confirmed by immunohistology staining for CD31, there the luminal surface of the leaflets in aortic position
was complete endothelial cell lining of the luminal sur- showed a smooth texture without apparent interruptions
face of matrix/polymer hybrid patches (data not shown). of the intimal integrity (Fig 8).

Large Animal Implantation


Comment
The preliminary biocompatibility and mechanical tests of
various hybrid material preparations indicated that The approach described herein to improve tissue engi-
P3HB-treated xenogenic matrix tends to exhibit better neering heart valve design consisted of the following
2090 STAMM ET AL Ann Thorac Surg
BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR 2004;78:2084 –93
CARDIOVASCULAR

Fig 6. Xenogenic biomatrix/polymer hybrid valve 12 weeks after implantation in aortic position in sheep. (A) Gross morphology reveals the
valve is in good condition, and the distal and proximal suture as well as both coronary orifices are clearly visible. The arterial wall lining is
smooth, the leaflets are delicate and freely mobile; (B) cross section through a leaflet and its hinge point at the conduit wall (Mason’s
trichrome staining). The collagenous valve scaffold is intact. There is some intimal thickening with inflammatory infiltration on the luminal
aspect of the leaflet. Panels C–F reveal the arterial wall of the conduit (hematoxylin & eosin stain): (C) hyperplasia of the tunica intima (ar-
rows), for comparison; (D) shows the native aorta of the same animal; (E) tunica media of the arterial conduit wall in which, at this point,
only few cells appear to have migrated into the media of the conduit wall; (F) for comparison, the native aortic media of the same animal is
illustrated.

steps: (1) extraction of a porcine heart valve and removal prostheses, heart valve tissue engineering strategies have
of all xenogenic cells by enzymatic digestion without attracted considerable attention. The initial approach was
altering the biological properties of the valve matrix based on the fabrication of the entire valve scaffold from
components (no tanning process); (2) penetration of the biodegradable polymers, followed by in vitro seeding
decellularized matrix with biodegradable polymer to with autologous cells and conditioning under simulated
enhance the mechanical characteristics of the porous in vivo conditions before implantation [1–3]. Early results
valve scaffold and to cover thrombogenic matrix compo- were promising, but it soon became clear that the com-
nents. It was shown that coating with poly(hydroxybu- plex three-dimensional structure of the native valve can
tyrate) does indeed improve biocompatibility and me- hardly be achieved with current techniques, and the
chanical properties in vitro, and that such hybrid tissue structural and mechanical properties of the various poly-
heals in well when inserted as a patch in the rabbit aorta. mers are not ideal. Furthermore, in vitro seeding and
Finally, P3/4HB-impregnated xenogenic hybrid valves conditioning with cells of the future recipient is a time-
were implanted in sheep. The two valves implanted in consuming process that would require sophisticated lab-
aortic position functioned well for up to 3 months and oratory equipment in the clinical setting. In addition, it
partially developed the morphologic characteristics of a remains unclear whether the cells actually adhere to the
native aortic valve. scaffold after implantation, and there is evidence that the
Due to the multiple shortcomings of conventional cells found on in vitro-seeded valves months after im-
Ann Thorac Surg STAMM ET AL 2091
2004;78:2084 –93 BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR

Fig 7. Immunohistology staining of a xe-


nogenic biomatrix/polymer hybrid valve
leaflet in aortic position 12 weeks after
implantation in sheep. (A) CD31 staining
demonstrating a confluent layer of endo-

CARDIOVASCULAR
thelial cells covering the leaflet predomi-
nantly on the luminal aspect but also on
the mural surface (original magnification
⫻100). (B) Higher magnification (⫻1000)
illustrating a monolayer of endothelial
cells with multiple CD31⫹ cells migrating
into inner layers of the leaflet. (C) Leaflet
stained for smooth muscle actin (⫻400);
multiple SMA⫹ cells are integrated in the
fibrous scaffold of the hybrid valve tissue.
(D) Stain of the corresponding negative
control.

plantation are in fact cells that have colonized the valve trix scaffold with those of an artificial polymer and
postimplantation, independent from the preseeding pro- demonstrated that polymer-treated valve scaffolds ex-
cess. More recently, natural xenogenic or allogenic heart hibit improved resistance to hemodynamic forces [5].
valve tissue has been propagated as a scaffold for heart These composite heart valves can now be implanted in
valve tissue engineering. In contrast to aldehyde-fixed vivo, and will be repopulated by recipient cells. The
tissue, enzymatically decellularized extracellular matrix polymer coating will then be degraded, while cells mi-
without tanning-induced crosslinks possesses epitopes grate into the extracellular matrix scaffold, restore the
for cellular adhesion receptors, facilitating repopulation natural tissue structure of the native heart valve, and
with tissue-specific celltypes but also inflammatory cells initiate the physiologic turnover of extracellular matrix
[6, 7]. Nonautologous matrix constituents such as colla- components. Once these processes have been completed,
gen, elastin, and proteoglycans have little antigenicity, the neo-valve should have biological and mechanical
given that cellular components are entirely removed [8, characteristics identical to those of the native valve.
9] It has recently been shown that mismatch of HLA-DR
and ABO antigens on endothelial cells in unmodified
valve allografts is associated with accelerated valve fail-
ure [10, 11]. To avoid the immunologic response to graft
cells, we have developed a decellularization process that
removes cells and cellular debris from the heart valve
scaffold by enzymatic digestion, without the use of ma-
trix-altering tanning procedures. Thus, the biological
properties of the extracellular matrix components are
preserved, and the decellularized valve can be repopu-
lated by cells derived from the recipient organism. In an
earlier series of experiments, we reseeded such decellu-
larized valve scaffolds with autologous blood vessel cells
in vitro and implanted those in pulmonary position in
sheep [4]. Three months after implantation, good valve
function and complete histologic restitution of valve
tissue, as well as a confluent endothelial surface were
found. However, aggressive enzymatic decellularization
inevitably weakens the valve tissue, so that the mechan- Fig 8. Surface morphology by scanning electron microscopy (origi-
ical properties do not allow for implantation in the high nal magnification ⫻200) of a xenogenic biomatrix/polymer hybrid
pressure system. Therefore, we sought to combine the valve 12 weeks in aortic position after implantation. Note that the
advantageous properties of the native extracellular ma- luminal surface of the leaflet is perfectly smooth.
2092 STAMM ET AL Ann Thorac Surg
BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR 2004;78:2084 –93

The coating process also serves to attenuate the pro- References


coagulatory activity of bare matrix components. It is well
established that platelet activation occurs when platelets 1. Shinoka T, Breuer CK, Tanel RE, et al. Tissue Engineering
Heart Valves. Valve Leaflet Replacement Study in a Lamb
come in contact with collagen. Platelets directly adhere Model. Ann Thorac Surg 1995;60:513–16.
by binding of platelet collagen receptor to integrins on 2. Sodian R, Hoerstrup SP, Sperling JS, et al. Early In Vivo
CARDIOVASCULAR

collagen, and the intrinsic clotting cascade is initiated Experience with Tissue-Engineered Trileaflet Heart Valves.
when prekallikrein, kininogen, factor XI and factor XII Circulation 2000;102:III22–9.
3. Hoerstrup SP, Sodian R, Daebritz S, et al. Functional Trileaf-
are exposed to collagen. In in vitro tests screening tests let Heart Valves Grown In Vitro. Circulation 2000;102:III44 –9.
we found that activation of both the cellular and the 4. Steinhoff G, Stock U, Karim N, et al. Tissue engineering of
plasmatic coagulation system is indeed attenuated by pulmonary heart valves on allogenic acellular matrix con-
polymer coating of decellularized matrix, while there was duits-in vivo restoration of valve tissue. Circulation 2000;102:
III50 –5.
no apparent change in the extent of complement
5. Grabow N, Schmohl K, Khosravi A, et al. Mechanical and
activation. structural properties of a novel hybrid heart valve scaffold
The use of uncoated decellularized matrix with or for tissue engineering. Artificial Organs 2004 (in press).
without in vitro autologous preseeding for heart valve 6. Courtman DW, Pereira CA, Kashef V, McComb D, Lee JM,
replacement remains controversial. Recently, Leyh and Wilson GJ. Development of a pericardial acellular matrix
biomaterial: biochemical and mechanical effects of cell ex-
colleagues [12] reported that decellularized porcine pul- traction. J Biomed Mater Res 1994;28:655–66.
monary valves implanted in pulmonary position in grow- 7. Bader A, Schilling T, Teebken O, et al. Tissue engineering of
ing sheep functioned well for up to 24 weeks and exhib- heart valves – human endothelial cell seeding of detergent
ited reconstitution of viable valve tissue and only mild acellularized porcine valves. Eur J Cardiotharac Surg 1998;
14:279 –84.
calcification. On the other hand, allogenic sheep valves 8. Yannas IV. Natural materials. In: Ratner BD, Hoffmann AS,
prepared in the same manner calcified rapidly, associ- Schoen FJ, Lemons JE, eds. Biomaterial science. San Diego:
ated with structural and functional deterioration. In a Academic Press, 1996:84 –94.
similar model, the same group observed that in vitro 9. Wilson GJ, Courtman DW, Klement P, Lee JM, Yeger H.
repopulation with autologous cells before implantation Acellular matrix: a biomaterial approach for coronary artery
bypass and heart valve replacement. Ann Thorac Surg
led to severe leaflet degeneration in vivo, while valves 1995;60(Suppl 2):S353.
that were implanted unseeded had clearly superior mor- 10. Baskett RJF, Nanton MA, Warren AE, Ross DB. Human
phologic and functional characteristics [13]. In another leukocyte antigen-DR and ABO mismatch are associated
study, however, they observed progressive degeneration with accelerated homograft valve failure in children: Impli-
cations for therapeutic interventions. J Thorac Cardiovasc
of xenogenic decellularized aortic tissue when implanted Surg 2003;126:232–9.
subcutaneously in rats, and concluded that some recel- 11. Hogan PG, O’Brien MF. Improving the allograft valve: Does
lularization process may be necessary before implanta- the immune response matter? J Thorac Cardiovasc Surg
tion [14]. 2003;126:1251–3.
12. Leyh RG, Wilhelmi M, Rebe P, et al. In vivo repopulation of
Decellularized valves are already commercially avail-
xenogenic and allogenic acellular valve matrix conduits in the
able. Preclinical large animal testing in the xenogenic and pulmonary circulation. Ann Thorac Surg 2003;75:1457–63.
allogenic setting was very promising, with good hemo- 13. Leyh R, Wilhelmi M, Haverich A, Mertsching H. A xenoge-
dynamic function, morphologic reconstitution, and little neic acellularized matrix for heart valve tissue engineering:
calcification for almost 1 year [15–17]. The initial enthu- in vivo study in a sheep model. Z Kardiol 2003;92:938 –46.
14. Walles T, Puschmann C, Haverich A, Mertsching H. Acellu-
siasm has been muted, however, by reports of early lar scaffold implantation–no alternative to tissue engineer-
xenograft failure in humans [18, 19]. Decellularized ho- ing. Int J Artif Organs 2003;26:225–34.
mografts appear to cause less problems in humans, but 15. Goldstein S, Clarke DR, Walsh SP, Black KS, O’Brien MF.
the actual long-term benefit remains to be determined Transpecies Heart Valve Transplant: Advanced Studies of
Bioengineered Xeno-Autograft. Ann Thorac Surg 2000;70:
[20, 21] 1962–9.
In conclusion, we believe that decellularized heart 16. O’Brien MF, Goldstein S, Walsh S, Black KS, Elkins R, Clarke
valve matrix holds great potential for creation of viable, D. The SynerGraft valve: a new acellular (nonglutaralde-
long-lasting replacement valves, and that many of the hyde-fixed) tissue heart valve for autologous recellulariza-
tion. Semin Thorac Cardiovasc Surg 1999;11(suppl):194 –200.
problems in the xenogenic setting, including impaired
17. Elkins RC, Dawson PE, Goldstein S, Walsh SP, Black KS.
biomechanics and residual antigenicity can be overcome Decellularized human valve allografts. Ann Thorac Surg
by pretreatment with biodegradable polymers. However, 2001;71(suppl):S428 –32.
more extensive long-term studies in large animals are 18. Simon P, Kasimir MT, Seebacher G, et al. Early failure of the
clearly needed before clinical use can be considered. tissue engineered porcine heart valve SYNERGRAFT in
pediatric patients. Eur J Cardiothorac Surg 2003;23:1002–6.
19. Sharp MA, Phillips D, Roberts I, Hands L. A cautionary case:
the SynerGraft vascular prosthesis. Eur J Vasc Endovasc
We thank the staff of ESS Experimental Surgical Services, Surg 2004;27:42–4.
University of Minnesota, for their excellence in carrying out the 20. Bechtel JF, Muller-Steinhardt M, Schmidtke C, et al. Evalu-
valve implantations in sheep. The contribution of Dr Heiko ation of the decellularized pulmonary valve homograft (Syn-
Klinge to the early rabbit experiments is acknowledged. We also erGraft). J Heart Valve Dis 2003;12:734 –9.
appreciate the help of the Department of Cardiac Surgery at 21. Sievers HH, Stierle U, Schmidtke C, Bechtel M. Decellular-
University of Luebeck with performing the fluid dynamic ized pulmonary homograft (SynerGraft) for reconstruction
analysis. of the right ventricular outflow tract: first clinical experience.
Z Kardiol 2003;92:53–9.
Ann Thorac Surg STAMM ET AL 2093
2004;78:2084 –93 BIOMATRIX/POLYMER COMPOSITE MATERIALS USED IN HEART VALVE REPAIR

DISCUSSION
DR SCOTT M. BRADLEY (Charleston, SC): In terms of the final DR HENRY L. WALTERS III (Detroit, MI): Did you do any
biopolymer you arrived at—I think it was a combination of two scanning electron micrography on the autopsy specimens of the

CARDIOVASCULAR
polymers— can you give us an idea of how long that takes to valves that you put in circulation to determine the uniformity of
degrade in vivo. endothelialization of the leaflets?

DR STAMM: The lifetime in vivo of P3 and P4HB is different, the DR STAMM: Do you mean like the image that I showed of the
half-lives are different. For P3HB, I believe they are in the leaflet in aortic position?
magnitude of several months, if not years; P4HB degrades much
faster. About the overall total degradation time or half-life of the
polymer combinations, nothing is known. DR WALTERS: That looked like a cross-section light micro-
graph to me. But you had shown some scanning electron
DR ANTONIO CORNO (Lausanne, Switzerland): The compos- micrograph specimens of your combination matrices, and I only
ite valves that you prepared are exposed to different regimen in saw light micrographs of a cross-section of the autopsy speci-
pulmonary artery position regarding the pressure and the oxy- mens after they had been in circulation. I thought the endothe-
gen saturation. Did you detect any difference to the response of lialization was one of the remarkable findings, and I just won-
this valve? dered how uniform it is and if you have done any studies on the
cells to assess their function.
DR STAMM: The valve in pulmonary position that survived the
procedure more or less intact had a complete endothelial lining. DR STAMM: As yet we only have cross-sectional images of both
We also found some smooth muscle cells in the leaflets, but we conventional staining and immunohistology, but the endothelial
were not able to detect a significant difference in terms of lining is complete in all the sections that we’ve looked at. We
histologic appearance between both implantation sites, aortic or have also looked at the entire surface by scanning electron
pulmonary. microscopy and again observed a complete cell lining.

The Society of Thoracic Surgeons: Forty-first Annual


Meeting
Please mark your calendars for the Forty-first Annual www.sts.org. Nonmembers who wish to receive informa-
Meeting of The Society of Thoracic Surgeons, to be held tion on the Annual Meeting may contact the Society’s
in Tampa, Florida, from Jan 24 –26, 2005. The program secretary, Gordon F. Murray.
will provide in-depth coverage of thoracic surgical topics
selected to enhance and broaden the knowledge of car-
diothoracic surgeons. Attendees will benefit from tradi- Gordon F. Murray, MD
tional Abstract Presentations, as well as Surgical Forums, Secretary
Breakfast Sessions, Surgical Motion Pictures, and Town The Society of Thoracic Surgeons
Hall Meetings on specific topics. 633 N. Saint Clair St, Suite 2320
Advance registration forms, hotel reservation forms, Chicago, IL 60611-3658
and details regarding transportation arrangements, as Telephone: (312) 202-5800
well as the complete meeting program, will be mailed to Fax: (312) 202-5801
Society members this fall. Also, complete meeting infor- e-mail: sts@sts.org
mation will be available on the Society’s Web site at website: www.sts.org

© 2004 by The Society of Thoracic Surgeons Ann Thorac Surg 2004;78:2093 • 0003-4975/04/$30.00
Published by Elsevier Inc

Anda mungkin juga menyukai