SERIES
SURGICAL LIGHTS
TEC-B-0002 REV0
1/09
Page 1
TABLE OF CONTENTS
TITLE PAGE
OPTICAL CHARACTERISTICS................................................................................................................ 2
Prescription Labeling
CAUTION: Federal Law (21 CFR 801.109(b)(1)) restricts the sale by or on the order of a licensed physi-
cian.
Distributed by:
SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512 (616) 656-2900
www.skytron.us
Manufactured by:
DKK Dai-Ichi Shomei Co., LTD
Tokyo, Japan
Although current at the time of publication, SKYTRON'S policy of continuous development makes this
manual subject to change without notice.
Page 2
OPTICAL CHARACTERISTICS
CRI 94 94 94 94
Central Illuminance Ec 106,490 lux 90,700 lux 133,000 lux 122,560 lux
Ee Ratio of Irradiance
to Illuminance 3.4 3.7 3.4 3.2
(millwatts/m2 per lux)
Total Irradiance
(watts/m2) 332.2 305.8 413.6 358.6
UV Energy
(watts/m2) 0.0018 0.0016 0.0013 0.0018
LED 5 LED 7
(Shadow Dilution)
Percentage Remaining
Illumination with:
• one mask 45 70
• two masks 49 48
• tube 96 88
• tube & 1 mask 43 53
• tube & 2 masks 44 39
d10 Diameter of a circle around the light field center (point of Illuminance)
where the Illuminance reaches 10% of Ec.
d50 Diameter of a circle around the light field center (point of Illuminance)
where the Illuminance reaches 50% of Ec.
Page 3
AC VOLTAGE
WARNING
This equipment is intended for use by healthcare professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby equipment. It may be necesssary
to take mitigation measures, such as re-orienting or relocating the lighthead or shielding the
location.
LIGHTHEAD LABEL
GRAND RAPIDS, MI · 616.656.2900
LED5 LED7
CAT. NO. LED5 LED7
SAFETY FUSE 2A 2A x 2 2A x 3
TIMELAG TIMELAG TIMELAG
Page 5
SPECIAL USER ATTENTION
Initial use should not begin until after the us- Be sure sterile handle is properly se-
ers have been instructed by the manufacturer's cured before using the lighthead. An
representative. improperly installed handle could fall
out, resulting in possible injury to patient
A routine instructional program must be or surgical staff.
implemented by the facility for proper usage
instructions for all personnel that may operate
this light.
CAUTION
As with the operation of any surgical light, all
hospital personnel should be aware that a cer- Sterilizable handles are subject to nor-
tain amount of care must be exercised to main- mal wear and tear. Always examine the
tain patient safety and to keep your SKYTRON handles for wear or damage to ensure
light fixture performing at peak efficiency. proper and safe operation with the
surgical light.
The following precautions should be reviewed
by all personnel prior to operating the light.
WARNING
NOTE
WARNING
WARNING
LED5
LED5 LED7
LED7 LED5
CEILING
COVER
VERTICAL
110° SUPPORT
VERTICAL TUBE
TRAVEL 360° ROLL
360°
BALANCE
MECHANISM
ROLL
360°
360° PITCH
360°
PITCH
360° 360°
The Aurora series surgical light is indicated for all The lights consist of several modules containing
types of surgical procedures. The Clinical settings LEDs (light emitting diodes) and optical color cor-
include, but are not limited to: The Operating Room, rective reflectors. Each lighthead features indepen-
Labor and Delivery, Emergency Department, Trau- dent focus capability allowing the user to adjust
ma, Intensive Care Unit, Minor Procedure Room, the illumination parameters. The light intensity
etc. The Aurora series surgical light is intended to is adjustable between 10% - 100% and a color
provide local surgical site illumination to any part of temperature of 4000K - 4500K creating the infinite
the patient's body. This is inclusive of all patients capability to adapt to different situations. LEDs of-
requiring surgical intervention. fer low heat radiation and increased illumination
durability.
Page 9
RADIAL ARM
VERTICAL ASSEMBLY
SUPPORT
TUBE
FOCUS
KNOBS
BALANCE
MECHANISM POSITIONING
HANDLES
STERILIZABLE
FOCUS/POSITIONING
HANDLES
Visual Checks
Use the following instructions to operate the light
Prior to start-up, a visual inspection of the following fixture:
points must be undertaken:
1. Position the lightheads as required by grasp-
• Light emission from each lighthead ing the lighthead positioning handles and moving
the lighthead to the desired position. See figure
• Check for cracks, damaged or broken lens. 3.
Avoid use if such damage is evident.
2. Turn the light fixture main power switch "ON"
• Check the operation of the wall intensity at the wall mounted control box and select the
control. desired intensity for each lighthead as required.
See figure 4. The mid-range position will provide
• Check mechanical movements by rotating adequate illumination for most procedures. Full
and articulating each joint. Ensure proper intensity will usually only be required for extreme
operation and emittance of light. deep cavity cases.
• When LED change indicator is illuminated the 3. The Aurora lighting fixture has selectable
LEDs must be evaluated for replacement. color temperatures, 4,000k (soft white) or 4,500k
Refer to Maintenance Manual. (bright white) to accommedate surgeon prefer-
ence. Press the color temperature change button
to select the desired color temperature. The button
is illuminated when 4,500k is selected.
Page 11
INTENSITY SAFETY
CONTROL FUSES
BUTTONS
WARNING
WARNING
Be sure sterile handle is properly se-
DO NOT use this fixture in the presence
cured before using the lighthead. An
of FLAMMABLE GASES.
improperly installed handle could fall
out, resulting in possible injury to pa-
tient or surgical staff.
7. When the light is no longer required, return
the lighthead to its full up position. Decrease the
intensity at the wall control, and turn the main
b. Push the handle in, turn it right and left,
power switch "OFF".
and pull the handle out to be certain that it is locked
(PUSH-TWIST-PULL). A distinct click can be heard
NOTE
when the handle is properly engaged.
SKYTRON products are guaranteed
c. To remove the handle, push the release
for proper performance with the use of
button and pull the handle out.
genuine SKYTRON CENTER FOCUS
HANDLES. After market competitive
handles and other disposable handles
CAUTION
will have varying results that could ul-
timately affect the proper performance
Sterilizable handles are subject to nor-
and secure engagement of the center
mal wear and tear. Always examine the
focus handle. Such applications are
handles for wear or damage to ensure
at the discretion of the user to ensure
proper and safe operation with the
patient safety.
surgical light.
Page 12
LIGHTHEAD POSITIONING
General
To obtain the maximum benefit from your The lightheads can be most effectively positioned
SKYTRON surgical lighting system, the following by using the following procedures:
suggestions are offered as a guide for lighthead
positioning. Personnel who are trained in proper 1. Grasp the positioning handles on the light-
lighting techniques can plan and set up the light- head and pull the lighthead down to shoulder
ing arrangements prior to the arrival of the patient. height. Keep the lighthead at approximately a 45°
Factors which should be considered when prepo- angle to easily position the support yoke. See figure
sitioning surgical lights are: 6.
SUPPORT
Surgical Table Placement YOKE
Surgical light positioning requirements change not 2. Using the positioning handles, rotate the
only from procedure to procedure, they also change lighthead around the vertical support until the
from surgeon to surgeon. Final light positioning lighthead is at an approximate 90° angle to the
and adjustment will be directed or done directly by radial arm. See figure 7.
the surgeon. The objective of prepositioning is to
require a minimum of final adjustments after arrival
of the patient. The non-sterile focus control should
be located where it can be reached by non-sterile
personnel and the sterile positioning/focus handle
where they can be reached by the surgeon. U s e
extreme care when prepositioning lightheads.
Bumping lightheads into one another, into walls,
or other equipment may alter bulb alignment which
affects proper focus adjustment.
Page 13
90°
180° 180°
ORTHO NEURO
OB/GYN EENT
DOCTOR SIDE OF TABLE
3. Place the radial arm in the desired position by
pushing or pulling the lighthead by the positioning
5. With the radial arm in proper position, rotate
handles as you walk around the surgical table.
the lighthead to the desired position and install the
sterile positioning/focus handle. Refer to sterile
4. Refer to figure 8 to approximate the desired
handle installation procedure.
radial arm position for locating the lighthead over
the patient.
6. Grasp the positioning handles, place the
lighthead at an angle and move the lighthead to
its full up position.
Page 14
ILLUMINATION TECHNIQUE
Maximum illumination, shadow reduction, and pos- Tilt the lighthead to position the focus control knob
sible obstruction by the surgeon or surgical staff are where it can be easily reached by non-sterile per-
also major concerns for lighthead positioning. The sonnel.
following examples are offered as a basic guide for
lighthead placement for large diameter/satellite, Position the satellite to the left or right according
dual lighthead, or triple lighthead fixtures. to surgeon preference. This allows a second light
source to come from another angle which will help
eliminate obstructions or shadows.
Large Diameter/Satellite Lighthead
Positioning Torso Area
The large diameter lighthead should be pre-posi- For most chest and abdominal procedures, posi-
tioned over the surgical site. The satellite can be tion the large lighthead directly over the surgical
used on either side of the surgeon for augmentation site. See figure 10. Position the radial arm on ap-
and shadow control. proximately a 45° angle from the surgical table
centerline. This position will locate the sterile
The large lighthead should be positioned perpen- focus/positioning handle on the lighthead where it
dicular to the bottom of the surgical cavity. can easily be reached by the surgeon. The focus
control will be where it can easily be reached by
non-sterile personnel. Position the satellite light-
Head head, depending on lighting needs, to augment
the larger lighthead.
To illuminate the head area, position the large
diameter lighthead radial arm parallel to the table
centerline. See figure 9. Position the lighthead be-
hind the surgeon. Tilt the lighthead to the desired
position using pitch axis movement. This will allow
the multiple light sources of the lighthead to pass
around the head and shoulders of the surgeon and
at the same time permit adequate head clearance
for the surgeon.
Figure 10.
Figure 9.
Page 15
Perineum
Figure 11.
Figure 12.
Page 16
30° 30°
120°
Figure 16.
Page 17
When an angled cavity is to be illuminated, at least Close attention to surgical light intensity during the
one of the lightheads should be positioned to be case as well as good quality general illumination
perpendicular to the bottom of the cavity. in the room will help to minimize eye fatigue of
surgical personnel.
For head and perineal work, the lights should be
positioned as they would for a dual system but with Dark surgical drapes will help to reduce reflectance.
a satellite on each side of the surgeon. White drapes should be avoided at all times be-
cause of high reflectance.
Bell-Shaped Cavities
The use of matte and satin finish instruments and
For most surgical procedures the lighthead will be retractors also helps to reduce eye fatigue.
properly focused with all the light beams converged
in one spot at the bottom of the surgical cavity.
LIGHT
BEAMS
FOCAL
BELL SHAPED POINT
CAVITY
Figure 17.
To insure proper operation and to extend the life The sterilizable handles are constructed of heat
of your SKYTRON surgical lighting system, the resistant, impact resistant plastic. They should
following preventive maintenance procedures are be cleaned with mild alkaline cleaning products
recommended. WITHOUT active chlorine. Thoroughly rinse off all
cleaners with water. Ensure that the open side of
handles are face down.
NOTE
Recommended sterilization parameters for steriliz-
All repairs should be made using only able handle:
authorized SKYTRON replacement
parts. Steam Sterilization:
134°C (273°F) : 2.3 bars (33.3 PSI)
Daily Maintenance 4 minutes
Daily or between cases, the lighthead exterior
should be wiped down with a mild cleaning agent Always consult current AORN journal recommen-
which will not affect the painted or acrylic parts. dations for proper sterilization procedures.
Do not apply or spray cleaning agents directly on
the lighthead.
Adjustments
Or contact:
SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512
616-656-2900 FAX 616-656-2906
www.skytron.us
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512 • 616.656.2900 • FAX 616.656.2906
PARTS CATALOG
LED SERIES
SURGICAL LIGHT
TEC-B-0009 REV3
3/10
INTRODUCTION
This manual contains the exploded views and replacement parts lists for the serviceable components
of the SKYTRON AURORA Series Surgical Lighting fixtures
Each serviceable part in these exploded views is identified by a reference number. Use this number
to locate necessary part information in the parts list adjacent to the exploded view.
Always use the complete SKYTRON part number and description when ordering replacement
parts.
Abbreviations
As Required........................................................ A/R
Optional ............................................................... opt
Serial Number.................................................... S.N.
Not Shown............................................................ NS
LIGHTHEAD LABEL
GRAND RAPIDS, MI · 616.656.2900
MODEL
SERIAL NO. (CAT. NO.)
WALL CONTROL LABEL
GRAND RAPIDS, MI · 616.656.2900
Although current at time of publication, SKYTRON’s policy of continuous development makes this
manual subject to change without notice.
Page 1
1.Radial Arm Assembly..............................................page 6 2. Balance Mechanism Assembly, LED5..................... page 8
3.Balance Mechanism Assembly, LED7..................page 10 4. Lighthead Yoke Assembly, Model LED5................ page 12
Page 2
5.Lighthead Yoke Assembly, Model LED7................page 14 6. LED Pod Assembly................................................ page 16
29
15. Wall Control Components..................................page 36 16. LED Stand Model Base Assembly...................... page 38
Page 5
4
3
25
2 5
6 7
24
26
23 8
22 9
10
21
20
13
12
11 8
14
15
19
18
17
16
57.031408.01mj
Page 7
1. RADIAL ARM ASSEMBLY
2
10
1 12 11
3 13
4 14
6 13
7
5
8 16
18
15
9 17
19
20
34 19
21
35
36 22
38 37
29
39 43
40
30
41
27
42
5 31
32 24
28
26
6 7 25
33
23
5.010307.04mj
Page 9
2. BALANCE MECHANISM ASSEMBLY MODEL LED 5
2
1
10
12 11
3
13
4 14
6 13
7
5
8
17
15
9 16
18
19
33 18
41 20
34
37 36 35 21
28
40 42
38
29
3 26
30
39
31 23
27
25
24
32
22
7.010307.04mj
Page 11
3. BALANCE MECHANISM ASSEMBLY MODEL LED 7
2
1
3
4
8
6
7
9
15
10
11 18
6
8
13
17
14
16
16
13
10 12
21 23
20 22
11
1
3
24
19
4 5.010307.01mj
Page 13
4. LIGHTHEAD YOKE ASSEMBLY MODEL LED 5
9 4
6
12
7
8
10 14
15
11 22
13
23
21
18
19
32
20 31
19 31
6
11 30
16 17
28 29
27
33
25
12
26
24 7.010307.01mj
Page 15
5. LIGHTHEAD YOKE ASSEMBLY MODEL LED 7
5
4
10 8
9
13
11
15 16
12
15
14
13
57.010307.03mj
Page 17
6. LED POD ASSEMBLY
11
10
9
8
7
11
6 4
23
22 19
21
20
24
4
3 2
18 13
1
14 16
13 15
17
14 13
13 12
57.010307.02mj
Page 19
7. FOCUS MECHANISM ASSEMBLY
1
2
3
4
7
9 8
10
11
12
13 13
14
15
24
16 17 18
32
25 19 20
26
21
27
22
14
29 14 23
28 14
30 14
5.010307.03mj
Page 21
8. LIGHTHEAD COMPONENTS, LED5
2 33
1
3
5 12
13 14
6 15
7 16
8
9 17
18
18
19
10
20
24 25 26
11
27
21 28
22
23
19
30 19
29
31
19
19
7.010307.03mj
Page 23
9. LIGHTHEAD COMPONENTS LED7
3
4
5
13
10
11
12 14
57.031408.06mj
Page 25
10. LIGHTHEAD ASSEMBLY
1
2
3
9 10
4 5
6
11
12 24 23 21
20
7 25
13 18
8
22
14 17
15 19 16
26 31
27 32
15 36
28 35
29 34
30
44
33
45 48
46 37
38
50 39
49
40
47 41
42
51 42 43
42
52
53
42
55 54
47
56 5tv.081308.02
Page 27
11. LIGHTHEAD COMPONENTS, LED5TV
2 1
4 3
31
5 30
29
6 32
7 28 27 25
33 23
8
35 34
35 22 21
36 26
24 18
9
37
20
10
19
17
11 42
12 13 43
41
14 40
39
38
48
44
49
15 45 50 51
16 46
52
47
36
53
54 36
36
55
56
57
58
1
36
59
7tv.081308.01
Page 29
12. LIGHTHEAD COMPONENTS, LED7TV
2 1
4 3
31
5 30
29
6 32
7 28 27 25
33 23
8
35 34
35 22 21
36 26
24 18
9
37
20
10
19
17
11 42
12 13 43
41
14 40
39
38
48
44
49
15 45 50 51
16 46
52
47
36
53
54 36
36
55
56
57
58
1
36
59
7tv.081308.01
Page 31
12. LIGHTHEAD COMPONENTS, LED7TV (continued)
Item Part No. Description Qty.
3
4
2
5 12
6
13
14
11
15
16
10
17
7
18
5tv.081308.03
Page 33
13. TV LIGHTHEAD ASSEMBLY
Item Part No. Description Qty.
3
4
5
8
10
57.010307.04
Page 35
14. WALL CONTROL ASSEMBLY
7
14
8
9
10
11
12
13 4
1
2
3
6 57.010307.05
Page 37
15. WALL CONTROL COMPONENTS
3
4
13 11 5
9
14 6
7
12 8 15
10 17
16
21
18
18 19
22 20
18
25 23
24 31
18
30
30
26 30 32
18 33
29
28 32
27
34
091807.01
Page 39
16. LED STAND MODEL BASE ASSEMBLY
13
7 14 27
6
9
11 12
5
3 15 7
4 10 7
1 4
16 28
2 29
18
30
19 7
17
31
32
33
20 37
7
35
21 4
34
30
22
36
23
24
23
25
26
091807.02mj
Page 41
17. STAND MODEL SUPPORT POST ASSEMBLY
SERIES
SURGICAL LIGHTS
7/07
Page 1
TABLE OF CONTENTS
TITLE PAGE
Distributed by:
SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512 (616) 656-2900
www.skytron.us
Manufactured by:
DKK Dai-Ichi Shomei Co., LTD
Tokyo, Japan
Although current at the time of publication, SKYTRON’S policy of continuous development makes this
manual subject to change without notice.
Page 2
Page 3
OPTICAL CHARACTERISTICS
CRI 94 94 94 94
Central Illuminance Ec 106,490 lux 90,700 lux 133,000 lux 122,560 lux
Ee Ratio of Irradiance
to Illuminance
(millwatts/m2 per lux) 3.4 3.7 3.4 3.2
Total Irradiance
(watts/m2) 332.2 305.8 413.6 358.6
UV Energy
(watts/m2) 0.0018 0.0016 0.0013 0.0018
LED 5 LED 7
(Shadow Dilution)
Percentage Remaining
Illumination with:
• one mask 45 70
• two masks 49 48
• tube 96 88
• tube & 1 mask 43 53
• tube & 2 masks 44 39
d10 Diameter of a circle around the light field center (point of Illuminance)
where the Illuminance reaches 10% of Ec.
d50 Diameter of a circle around the light field center (point of Illuminance)
where the Illuminance reaches 50% of Ec.
Page 4
AC VOLTAGE
WARNING
This equipment is intended for use by healthcare professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby equipment. It may be necesssary to
take mitigation measures, such as re-orienting or relocating the lighthead or shielding the
location.
1 PIN PUNCH SET 1/8" - 3/8" 1 METRIC, L-TYPE ALLEN WRENCH SET 1.5-8mm
1 TEFLON TYPE SPRAY LUBRICANT (TRI-FLOW) 1 PENCIL
1 SET OF PHILLIPS SCREWDRIVERS 1 TRUE RMS VOLTMETER
1 SET OF FLAT BLADE SCREWDRIVERS 1 SET OF NON ABRASIVE CLEAN CLOTHS
1 DIGITAL LEVEL 1 MILD CLEANING AGENT, NON ALCOHOL
Page 5
The lighthead Data Labels contain the lighthead
model number, bulb type, fuse type, electrical specifications
and product serial number.
LIGHTHEAD LABEL
GRAND RAPIDS, MI · 616.656.2900
LED5 LED7
CAT. NO. LED5 LED7
SAFETY FUSE 2A 2A x 2 2A x 3
TIMELAG TIMELAG TIMELAG
Page 6
SECTION I MAINTENANCE
1-1. General 1-3. Preventive Maintenance Procedures
To ensure proper operation and extend the life of The following procedures should be done semi-
your SKYTRON surgical lighting fixture, the follow- annually or sooner as needed.
ing preventive maintenance procedures are rec-
ommended. a. Lighthead top cover should be removed and
any accumulation of dust or lint removed.
NOTE
b. Check the focus drive mechanism for proper
All repairs should be made using autho- operation and lubrication. Lubricate the drive
rized SKYTRON replacement parts. gears with lithium grease and all pivot points with a
lightweight oil as necessary.
REMOVE SCREWS
REMOVE TOP
Figure 1-3.
b. Disconnect module connector and loosen the 2. Remove four screws securing Wall Control
1/4 turn screw securing the module. See figure 1- faceplate and remove faceplate.
2.
3. Locate Reset Button as shown in figure 1-
4. Press button and hold for 2 seconds to reset.
CONNECTOR
RESET
Figure 1-2. BUTTON
Figure 1-4.
c. Remove the module from the pad. See figure
1-3. 4. Install Wall Control Faceplate and turn
power back on to fixture.
Page 8
a. Position the lighthead 42" from a white test c. To adjust a misaligned LED module, loosen
surface. Turn main power "ON" and set intensity on the jam nut on the affected focus rod and turn the
low. Adjust focus control to widest pattern diameter adjustment screw as needed to achieve proper
so each light beam can be seen individually. See alignment. See figure 1-6.
figure 1-5.
LIGHTHEAD
INDIVIDUAL
BULB SPOT
42"
FOCUS
ROD
ADJUSTMENT
SCREW
JAM NUT
BULB TEST
CONVERGENCE SURFACE
WARNING
Do not remove lighthead when support arm is in down position. The balance mechanism will be
severely damaged.
Page 9
SECTION II POSITIONING AXES ADJUSTMENT
2-1. General
As a part of normal preventive maintenance, all fixture rotation axes adjustments should be checked
and adjusted as necessary. Refer to figure 2-1 for location of desired check or adjustment procedure.
RADIAL ARM
ASSEMBLY
CEILING COVER
(MOUNTING HUB)
VERTICAL SUPPORT
TUBE (VST)
BALANCE C
MECHANISM E B
(BOM)
A
D
A
D
A-Lighthead Pitch Axis paragraph 2-2 C-Vertical Travel Tension paragraph 2-5
B-Lighthead Roll Axis- D-Lighthead Horizontal
•Model 7 lighthead paragraph 2-3 Rotation paragraph 2-6
•Model 5 lighthead paragraph 2-4 E- Radial Arm Horizontal
Rotation paragraph 2-7
PIN PUNCH
HOLE
SET YOKE
SCREW (2)
b. The lighthead should move freely yet main- d. Rotate the lighthead until a hole is visible and
tain its position without drifting. If an adjustment is insert a pin punch into the hole in the adjustment
required, remove the top cover, refer to figure 2-3 nut.
and proceed as follows:
e. With the nut held captive with the pin punch,
c. Rotate the lighthead until an allen set screw rotate the lighthead clockwise to increase the fric-
is visible through the adjustment hole. Loosen the tion or counterclockwise to decrease the friction.
set screw, rotate the lighthead 180° and loosen the
second set screw. f. Remove pin punch and check adjustment.
Tighten set screws and replace top cover when
adjustment is complete.
Page 11
B
ADJUSTMENT
HOLE
YOKE
Figure 2-4. Model 7 Lighthead Roll Figure 2-5. Model 7 Roll Adjustment
PIN
PUNCH
B
HOLE
SUPPORT
ARM
YOKE
Figure 2-6. Model 5 Lighthead Roll Figure 2-7. Model 5 Roll Adjustment
b. The lighthead should move freely yet main- d. Continue to rotate the yoke until a 5mm hole
tain its position without drifting. If an adjustment is is visible through the hole. Insert a pin punch into
required, refer to figure 2-7 and proceed as follows: the hole in the adjustment nut.
c. Rotate the yoke until an allen set screw is e. With the tension nut held captive with the pin
visible through the adjustment hole and loosen the punch, rotate the yoke clockwise (viewed from the
set screw. Rotate the yoke 180° and loosen the front) to increase the friction and counterclockwise
second set screw. to decrease the friction.
ADJUSTMENT
SLOT
NOTE
BEFORE
BOM
•Recheck the mounting plate to make
sure it is absolutely level. In most
cases, releveling the mounting plate
will solve any drifting problems.
Figure 2-11 . BOM Lubrication h. With the nut held captive by the pin punch,
rotate the lighthead so that the pencil marks are
approximately 1/8" apart. Turn the lighthead clock-
c. If releveling the mounting plate or lubricating wise (viewed from the bottom) to increase the
the BOM needle bearings does not correct the tension and counterclockwise to decrease the ten-
problem, an adjustment may be required. This sion. In some cases it may require considerable
adjustment is made by increasing or decreasing force on the lighthead to make this adjustment.
the bearing preload. Refer to figure 2-12 and
proceed as follows: i. Remove the pin punch and check the lighthead
for proper tension. Repeat adjustment procedure
d. Locate the adjustment hole. This hole is if necessary. When proper adjustment has been
located just beneath the lower VST screws in the achieved, tighten the set screws and install the
BOM and has a slotted head chrome plug in it. chrome plug.
Page 15
a. Check the horizontal rotation axis tension by A regular program of preventive maintenance will
moving the Radial Arms through their full range of increase the life of your equipment and keep it
travel around the center mounting hub. Refer to operating at peak performance.
figure 2-13.
Maintenance must be performed by authorized,
trained maintenance personnel using SKYTRON
authorized replacement parts and service
techniques. Service instructions and parts are avail-
able from SKYTRON.
NOTE
+30V (RED)
+30V (RED)
GND (GREEN)
GND (GREEN)
SIG (WHITE)
SIG (WHITE)
+30V
OWNERS MANUAL
REV 6/08
Page 1
TABLE OF CONTENTS
TITLE PAGE
Although current at time of publication, SKYTRON's policy of continuous development makes this manual
subject to change without notice.
Page 2
Page 3
OPTICAL CHARACTERISTICS
CRI 94 94 94 94
Central Illuminance Ec 106,490 lux 90,700 lux 133,000 lux 122,560 lux
Ee Ratio of Irradiance
to Illuminance
(millwatts/m2 per lux) 3.4 3.7 3.4 3.2
Total Irradiance
(watts/m2) 332.2 305.8 413.6 358.6
UV Energy
(watts/m2) 0.0018 0.0016 0.0013 0.0018
LED 5 LED 7
(Shadow Dilution)
Percentage Remaining
Illumination with:
• one mask 45 70
• two masks 49 48
• tube 96 88
• tube & 1 mask 43 53
• tube & 2 masks 44 39
AC VOLTAGE
WARNING
This equipment is intended for use by healthcare professionals only. This equipment may cause
radio interference or may disrupt the operation of nearby equipment. It may be necesssary to
take mitigation measures, such as re-orienting or relocating the lighthead or shielding the
location.
LIGHTHEAD LABEL
GRAND RAPIDS, MI · 616.656.2900
LED5 LED7
CAT. NO. LED5 LED7
SAFETY FUSE 2A 2A x 2 2A x 3
TIMELAG TIMELAG TIMELAG
Page 6
SPECIAL USER ATTENTION
NOTE
WARNING
SKYTRON products are guaranteed
Indicates a possibility of personal injury.
for proper performance with the use of
genuine SKYTRON CENTER FOCUS
HANDLES. After market competitive
handles and other disposable handles
CAUTION
will have varying results that could
ultimately affect the proper performance
Indicates a possibility of damage to
and secure engagement of the center
equipment.
focus handle. Such applications are at
the discretion of the user to ensure
patient safety.
NOTE
CAUTION
NOTE
NOTE
CAUTION
NOTE
NOTE
345˚
CEILING VERTICAL
COVER SUPPORT
TUBE
BALANCE
MECHANISM
540˚
540˚ 340˚
340˚
360˚ VERTICAL
SUPPORT
TUBE
ROLL
360˚
110˚
VERTICAL
TRAVEL
360˚
BALANCE
MECHANISM
360˚ ROLL
360˚
PITCH
360˚
The Aurora LFS series combines flatscreen moni- The lighthead models available for use with the
tor mounting with a surgical lighting system from a LFS system include LED5 AND LED7.
single ceiling mount.
The lights consist of several modules containing
The LFS model allows a single flatscreen monitor LEDs (light emitting diodes) and optical color cor-
mount to be combined with up to 3 separate rective reflectors. Each lighthead features inde-
lightheads. The LFSLFS model combines 2 pendent focus capability allowing the user to adjust
flatscreen monitor mounts with up to 2 separate the illumination parameters. The light intensity is
lightheads. adjustable between 10% - 100% and a color tem-
perature of 4000K - 4500K creating the infinite
The LFS radial arm assembly allows up to 90" of capability to adapt to different situations. LEDs
reach for the flatscreen monitor with 345° of rotation offer low heat radiation and increased illumination
capability at the ceiling mount. Vertical travel of up durability.
to 45.5" is provided.
Page 9
The Aurora series surgical lighting system from The adjustable focus mechanism which optimizes
SKYTRON provides high illumination of the operat- the light output by superimposing all the light beams
ing field at varying angles. Combinations of vertical into a single spot can be operated by non-sterile
positioning and multiple rotational capabilities allow personnel using the lighthead mounted focus knob.
the single, dual or triple lighthead models virtually
limitless positioning with a thirteen foot diameter All lightheads also have a removable, sterilizable,
area (triple lighthead models). focus/positioning handle. This allows all final posi-
tioning and focus adjustments or changes to be
The fixtures are single point ceiling mounted with a precisely done by the surgeon. See figure 2.
continuous 360 degree rotation capability at the
ceiling mount end of the radial support arm. See
figure 1. The balance mechanism which is at-
tached to the radial arm by a vertical support tube,
provides the lighthead an additional continuous 360
degree rotation point. The balance mechanism is NON-STERILE
an enclosed spring tension system. This allows SIDE FOCUS KNOB
vertical movement of the lighthead while maintain-
ing the lighthead position without drifting. The yoke
provides additional 360 degree rotation points for
lighthead pitch and roll.
RADIAL ARM
ASSEMBLY
VERTICAL
FOCUS SUPPORT
KNOB TUBE
BALANCE POSITIONING
MECHANISM HANDLE
STERILIZABLE
FOCUS / POSITIONING
HANDLE FOCUS
KNOB
STERILIZABLE
FOCUS / POSITIONING
HANDLE
Visual Checks
Prior to start-up, a visual inspection of the following Use the following instructions to operate the light
points must be undertaken: fixture:
• Light emission from each lighthead 1. Position the lightheads as required by grasp-
ing the lighthead positioning handles and moving
• Check for cracks, damaged or broken lens. the lighthead to the desired position. See figure 3.
Avoid use if such damage is evident.
2. Turn the light fixture main power switch "ON"
• Check the operation of the wall intensity at the wall mounted control box and select the
control. desired intensity for each lighthead as required.
See figure 4. The mid-range position will provide
• Check mechanical movements by rotating adequate illumination for most procedures. Full
and articulating each joint. Ensure proper op intensity will usually only be required for extreme
eration and emittance of light. deep cavity cases.
• When LED change indicator is illuminated the 3. The Aurora lighting fixture has selectable
LEDs must be evaluated for replacement. color temperatures, 4,000k (soft white) or 4,500k
Refer to Maintenance Manual. (bright white) to accommedate surgeon prefer-
ence. Press the color temperature change button to
select the desired color temperature. The button is
illuminated when 4,500k is selected.
Page 11
INTENSITY SAFETY
CONTROL FUSES
BUTTONS
WARNING
WARNING
Be sure sterile handle is properly
secured before using the lighthead. An DO NOT use this fixture in the presence
improperly installed handle could fall of FLAMMABLE GASES.
out, resulting in possible injury to patient
or surgical staff.
7. When the light is no longer required, return
the lighthead to its full up position. Decrease the
b. Push the handle in, turn it right and left, intensity at the wall control, and turn the main power
and pull the handle out to be certain that it is locked switch "OFF".
(PUSH-TWIST-PULL). A distinct click can be heard
when the handle is properly engaged. NOTE
c. To remove the handle, push the release SKYTRON products are guaranteed
button and pull the handle out. for proper performance with the use of
genuine SKYTRON CENTER FOCUS
HANDLES. After market competitive
CAUTION handles and other disposable handles
will have varying results that could
Sterilizable handles are subject to ultimately affect the proper performance
normal wear and tear. Always examine and secure engagement of the center
the handles for wear or damage to ensure focus handle. Such applications are at
proper and safe operation with the the discretion of the user to ensure
surgical light. patient safety.
Page 12
LIGHTHEAD POSITIONING
General
To obtain the maximum benefit from your The lightheads can be most effectively positioned
SKYTRON surgical lighting system, the following by using the following procedures:
suggestions are offered as a guide for lighthead
positioning. Personnel who are trained in proper 1. Grasp the positioning handles on the lighthead
lighting techniques can plan and set up the lighting and pull the lighthead down to shoulder height.
arrangements prior to the arrival of the patient. Keep the lighthead at approximately a 45° angle to
Factors which should be considered when easily position the support yoke. See figure 6.
prepositioning surgical lights are:
CAUTION
3. Place the radial arm in the desired position by The upper radial arm of the flatscreen monitor
pushing or pulling the lighthead by the positioning mount should be pre-positioned on the opposite
handles as you walk around the surgical table. side of the table from the surgeon at approximately
90° from the table center line. See figure 9.
4.Refer to figure 8 to approximate the desired
radial arm position for locating the lighthead over
the patient.
ABDOMINAL
180˚ 180˚
ORTHO NEURO
OB/GYN EENT
DOCTOR SIDE OF TABLE
5. With the radial arm in proper position, rotate The lower arm should be positioned under the
the lighthead to the desired position and install the upper arm. In this position the monitor can be easily
sterile positioning/focus handle. Refer to sterile moved up or down the full length of the table without
handle installation procedure. interfering with the lightheads. The monitor can be
pushed up out of the way until it is needed.
6. Grasp the positioning handles, place the
lighthead at an angle and move the lighthead to its Two sterilizable handles are provided for final moni-
full up position. tor positioning or for changes required during the
procedure. Prior to the start of the procedure the
sterilizable handles can be installed. Insert the
sterile handles into the receptacle and turn clock-
wise until tight.
NOTE
Maximum illumination, shadow reduction, and pos- Tilt the lighthead to position the focus control knob
sible obstruction by the surgeon or surgical staff where it can be easily reached by non-sterile
are also major concerns for lighthead positioning. personnel.
The following examples are offered as a basic
guide for lighthead placement for large diameter/ Position the satellite to the left or right according to
satellite, dual lighthead, or triple lighthead fixtures. surgeon preference. This allows a second light
source to come from another angle which will help
eliminate obstructions or shadows.
Large Diameter/Satellite Lighthead
Positioning
Torso Area
The large diameter lighthead should be pre-posi-
tioned over the surgical site. The satellite can be For most chest and abdominal procedures, posi-
used on either side of the surgeon for augmentation tion the large lighthead directly over the surgical
and shadow control. site. See figure 11. Position the radial arm on
approximately a 45° angle from the surgical table
The large lighthead should be positioned perpen- centerline. This position will locate the sterile focus/
dicular to the bottom of the surgical cavity. positioning handle on the lighthead where it can
easily be reached by the surgeon. The focus con-
trol will be where it can easily be reached by non-
Head sterile personnel. Position the satellite lighthead,
depending on lighting needs, to augment the larger
To illuminate the head area, position the large lighthead.
diameter lighthead radial arm parallel to the table
centerline. See figure 10. Position the lighthead
behind the surgeon. Tilt the lighthead to the desired
position using pitch axis movement. This will allow
the multiple light sources of the lighthead to pass
around the head and shoulders of the surgeon and
at the same time permit adequate head clearance
for the surgeon.
Figure 11.
Figure 10.
Page 15
Perineum
In some cases such as cholecystectomies and The large diameter lighthead should be positioned
total abdominal hysterectomies, the surgical cavity at the end of the table for perineal procedures.
may be angled. In cases such as this, the large Locate the radial arm directly in line with the centerline
lighthead should be angled so that the face of the of the table. Once the surgeon has assumed a
lighthead is perpendicular to the bottom of the seated position, the lighthead can be pulled down,
surgical cavity. See figure 12. angled, and adjusted to provide the necessary
illumination over the surgeon’s head and shoul-
ders. See figure 14. The satellite lighthead radial
arm should be positioned approximately 90° from
the other radial arm. Position the satellite lighthead
to the right or left of the large lighthead according to
surgeon preference. In this position, the focus
knobs of both lightheads are located for easy reach
by non-sterile personnel.
Figure 12.
Figure 14.
Figure 13.
Page 16
Fixtures containing dual 24" diameter lightheads Tilting the lightheads will give a larger light beam
require some special positioning considerations. angle. See figure 16. Final positioning and focus
The small diameter of the lightheads allows the light adjustments can be done by the surgeon using the
source to be more easily obstructed by the surgical sterile focus/positioning handles. Focus controls
staff. It is very important that the surgical site should be positioned where easily reached by non-
remain illuminated even though the head and hands sterile personnel.
of the surgeon and the surgical staff may be directly
in the central light beam path. In order to minimize
shadowing, the lightheads should be positioned so
that their light beams are angled into the surgical 30˚
cavity. Regardless of the surgical site, these lights
should be positioned to maintain an angle of ap-
proximately 30° about an imaginary line running
perpendicular to the bottom of the surgical site. See
figure 15.
30˚ 30˚
Figure 16.
Figure 15.
Page 17
Triple lighthead systems will either consist of a For most surgical procedures the lighthead will be
large diameter lighthead with two 24" satellites or properly focused with all the light beams converged
three 24" lightheads. There are two basic position- in one spot at the bottom of the surgical cavity.
ing strategies that can be used to obtain the best
illumination possible. The first is to align all three However, in the case of a bell-shaped cavity (for
lightheads to the centerline of the table with the example - total abdominal hysterectomy on an
large lighthead directly over the center of the surgi- obese patient), focusing at the bottom of the cavity
cal site. The second is to cluster the lights in a may cause more shadow problems. The focal
circular arrangement over the surgical site with point for the light beams should be above the
each lighthead about 120° away from each other. bottom of the cavity. See figure 18.
The whole cluster should be positioned to minimize
interference with the head and shoulders of the
surgical staff. See figure 17.
LIGHT
BEAMS
120˚ 120˚
30˚ FOCAL
BELL SHAPED POINT
CAVITY
30˚ 30˚
Figure 18.
120˚
Beyond the focal point, the light beams spread out
Figure 17. like an inverted cone and will more evenly spread
the light throughout the bell-shaped cavity.
General Sterilization
To insure proper operation and to extend the life of The sterilizable handles are constructed of heat
your SKYTRON surgical lighting system, the fol- resistant, impact resistant plastic. They should be
lowing preventive maintenance procedures are cleaned with mild alkaline cleaning products WITH-
recommended. OUT active chlorine. Thoroughly rinse off all clean-
ers with water. Ensure that the open side of handles
NOTE are face down.
All repairs should be made using only Recommended sterilization parameters for
authorized SKYTRON replacement sterilizable handle:
parts.
Steam Sterilization:
134°C (273°F) : 2.3 bars (33.3 PSI)
Daily Maintenance 4 minutes
Daily or between cases, the lighthead exterior Always consult current AORN journal recommen-
should be wiped down with a mild cleaning agent dations for proper sterilization procedures.
which will not affect the painted or acrylic parts. Do
not apply or spray cleaning agents directly on the
lighthead. Adjustments
a. Check the vertical tension adjustment of the c. Check the adjustment for the flatscreen
Height Adjustable Arm for its capacity to support monitor pitch axis. The monitor should move freely
the flatscreen monitor throughout its range of mo- yet maintain its selected position without drifting. If
tion. The monitor should move freely yet maintain an adjustment is necessary, refer to figure 20 and
its selected position without drifting. If an adjust- proceed as follows:
ment is necessary, refer to figure 19 and proceed
as follows:
NOTE
SET SCREW
1/8" PIN PUNCH
COVER PLATE
ADJUSTMENT Figure 20
NUT
Service
Or contact:
SKYTRON
5085 Corporate Exchange Blvd. S.E.
Grand Rapids, MI 49512
1-800-SKYTRON (1-800-759-8766)
Fax. 1-616-656-2906
Page 22 1. HUB AND RADIAL ARM ASSEMBLY
6
23
24
2 22
25 12
1
13
15
3 14 16
17
6
4
18
5
6
18
8
7
19
9
18
18
10
20
11
21
111003.01mj
1. HUB AND RADIAL ARM ASSEMBLY Page 23
3
7
8 4
7 5
9
10
11
12 6
10
13 23
16
23
15 22
17 20
24
19
25
14
25
21 29 26
18
30 27
38
28
43 42
46
36
39 35
44 33
47 40 33
37
41 31
45 34
48
33
32
49
0807.01
2. FLATSCREEN BRACKET ASSEMBLY W/ NYLON STOP PIN Page 25
24
15
14
13 69
68
18
12 23
19 22
11 18 21 67
10
9 8 16 20 31
66
65
2 20
3 64
6 17
5 63
7
1
4
3 2
30 62
28 38
29
28
61
27
60 59
26 36
25 58
37
35 57
48 49 56
5
34
47 50
51
55
42 54
32 46 48
52
53
33
49
4 47
41
45
39
40
43
44
39 0807.02
3. HEIGHT ADJUSTABLE ARM ASSEMBLY Page 27
24
15
14
13 69
68
18
12 23
19 22
11 18 21 67
10
9 8 16 20 31
66
65
2 20
3 64
6 17
5 63
7
1
4
3 2
30 62
28 38
29
28
61
27
60 59
26 36
25 58
37
35 57
48 49 56
5
34
47 50
51
55
42 54
32 46 48
52
53
33
49
4 47
41
45
39
40
43
44
39 0807.02
3. HEIGHT ADJUSTABLE ARM ASSEMBLY (CONTINUED) Page 29