Anda di halaman 1dari 32

Basic Operation of the TRACE MS PLUS for Engineers

_____________________________________________________________________________

1. Basic Operation of the TRACE MS


PLUS for Engineers

Introduction
This document assumes the reader already has a working familiarity with
TRACE MS/Voyager/MD800 control via the Xcalibur data system.
TRACE MS Plus requires Xcalibur 1.3.

Configuring TRACE MS Plus


Run the Xcalibur Instrument Configurator.

Select Instrument Configurator from the Windows Start menu.

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 1


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Double-click the TRACE MS Plus button that appears in the instruments


list on the right-hand-side of the dialog.

At time of writing, there were no configurational settings for TRACE MS


Plus. It can be anticipated that there will be some adjustable parameters in a
later release version, to enable such options as 800amu (for ToxLab) and CI
capability.

Launching the Server


The Server can be launched in 2 ways:

1. Running the Tuning application for TRACE MS Plus.

2. Launching Xcalibur.

If Xcalibur is launched, it is assumed that the instrument configuration has


declared the TRACE MS Plus as selected.

2 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

When the Server is running, there will be an icon in the Windows Start
menu, as shown below.

The Icon changes colour depending on the state of the MS.

Yellow indicates Pumped Down and in Standby.

Green Indicates the MS is in Operate.

Red indicates the system is Vented, and therefore in Standby.

Perform a right-click on the icon to open the Server menu.

It is possible to access the Manual Tune dialog from here if it is not already
open.

Other operations that are possible from this menu are:

Automated Tuning and/or Automated Mass Axis Calibration

Manual Mass Calibration, via the Instrument Tune and Calibration…


menu item

Reconfiguration of the MS Inlet (GC, DIP or DCI)

Venting/Pump-down operations

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 3


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Tune Dialog

Opening the Tune dialog


Tune can be launched from its Shortcut on the Windows Desktop, or if not
visible, through the Start Menu, as shown below.

4 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Default Tune dialog

The Tune dialog represents a radical departure from earlier generation user
interfaces. When it first launches the display is seen to be organised into
two regions.

Graphical Display

The top of the dialog is occupied by one or more mass range graphical
displays - dependent on how many have been specified/activated – and each
will have annotation indicating Absolute Intensity, Centroided Mass
Measurement and Display Gain (Zoom) of the individual mass selection.

Tabular Display

The bottom region of the dialog is composed of a table of Scan Functions.


Each entry (row) describes a mass with defined mass span, scan time,

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 5


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

detector voltage setting and peak format and each may be selected in the
graphical region above.

In the example caption, it will be evident that a level of customisation is


possible. Each Scan Function can have an alias name applied to it, for
familiarity. The total number of rows visible on-screen is defined in a
configurational dialog that opens from the View | Options… menu item. It
will be discussed later.

The example shows 7 Scan Functions that have been compiled and together
these define the most routine interactive procedures associated with
maintenance of a MS.

FS1 through FS3 show individual ions from the spectrum of PFTBA
(Heptacosa); these would be used to review sensitivity and mass resolution
for routine analysis.

FS4 has been designed to show a wide mass range and therefore allowing
inspection of the instrument ion background; this would be useful to monitor
contaminants such as column and septum bleeds, hydrocarbon background
from dirty samples and plasticizer ions that could potentially be introduced
during column installations and/or source cleaning.

FS5 through FS7 have been set up to monitor for potential air leaks,
referenced against Helium carrier gas.

Collectively, the table of Scan Functions replaces the earlier technique of


swapping saved parameter display files to accomplish these fundamental
tasks.

The table also permits time windows to be applied to any or all Scan
Functions. Although the default setting is zero, a time range, say 0.00min to
1.00min would enable an acquisition to be performed from the tune dialog,
independent of Xcalibur. This is useful for performing diagnostic manual
mass calibrations very rapidly.

6 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Displaying Tune Parameters


The Windows slider control marked Show Per Method Parameters, at top
right of the dialog will reveal the tune parameters, as shown below.

These parameters are global to all Scan Functions.

Method Editor dialog

The Method Editor dialog for TRACE MS Plus is launched from


Xcalibur’s Instrument Setup dialog

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 7


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Pumping Down and Venting


This is achieved by right-clicking the Server icon at the right hand side of
the Windows taskbar.

Pumping Down

When the instrument is vented, the Server icon will be red (Vented).

To pump down, perform a right-click on the icon to display its menu.

Select the Pump menu item to start the pump down procedure. The Server
icon will flash yellow until high vacuum has been achieved AND
turbomolecular pumps achieve adequate frequency of rotation, where after it
will stop flashing and adopt a continuous yellow appearance.

8 __________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Venting

When the instrument is pumped down, the Server icon will be yellow
(Operate OFF) or green (Operate ON). The right-click operation will
reveal the following menu:

Selecting the Vent item will display a confirmation prompt, as follows:

Select Yes to Vent the MS, if this is required. Venting will be immediate,
with the deactivation of the Vent Valve solenoid attached to the front (or
only) turbomolecular pump.

_____________Basic Operation of a TRACE MS Plus for Engineers ___________________ 9


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Tuning and Calibrating

Autotuning

Select the Instrument Tune and Calibration menu item from the Server icon.

The following dialog will be displayed:

Select the option for Full System Autotune.

A status dialog will be generated showing a log of events that will


eventually lead to a fully tuned and calibrated instrument.

10 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

The Tune dialog can be seen to update with the Autotune algorithm.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 11


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

12 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Ultimately, the log of events can be copied to the clipboard for pasting into a
document. This may be useful for diagnostics, if a problem is suspected, or
as a teaching aid for learning manual tuning.

Eventually, the system will be tuned and calibration will follow:

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 13


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

The Tune dialog will show a full scan across the reference ions of the
internal tuning compound, PFTBA (FC43).

14 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

And here is the final outcome:

Print the report, if required, otherwise hit the Finish button.

The Reference gas will be turned OFF automatically and Operate will be
set back to its previous state.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 15


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Tuning in Normal Mode


Select the Manual Tune menu item to launch the Tune dialog.

This dialog shows the Standard Tune view. There are no Tune parameters
to adjust. It is assumed that Autotune is used to set up the instrument.

16 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Selecting Ionisation Mode


Different ionisation modes are selectable from the combo bow as shown
below:

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 17


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Tuning in Engineers Mode


To turn ON Engineers Mode, open the Options dialog from the View menu:

The Options dialog will appear as shown:

18 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Select the radio button marked Engineer to reveal all parameters.

When the Tune page is re-opened you will see additional Per Method
Parameters.

It is possible to view many readbacks, power supply rails and perform


various diagnostic tests.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 19


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Performing Mass Calibration


This can be achieved from the menu attached to the Server icon.

Select the Standard Mass Scale Calibration option and follow the
instructions:

20 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

The dialog will show status information as the instrument is calibrated.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 21


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

The Tune dialog will show a scanning acquisition occurring.

Ultimately the process will terminate:

There will be an option to print off a report prior to closing the dialog.

22 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Manual Mass Calibration


Only necessary for diagnostic purposes. Uses an acquired file, and a
selected calibration reference file:

Follow the instructions in the wizard, as shown below:

The Raw File must contain spectra acquired over an adequate mass range
with satisfactory mass resolution.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 23


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Saving Tune Parameters to a named file


Very straightforward, merely open the File | Save As… menu item and
name your tune file.

24 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Setting up MS-only acquisition

Acquiring from Tune Dialog

Need a scan function defining an adequate mass range, as in example below:

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 25


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Specify a file name and location in the options dialog:

Use the File | Start Acquiring menu item to commence:

26 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

You will be able to monitor the elapsed time.

Acquiring from Xcalibur


This is same as with previous systems. Define parameters in Instrument
Method Editor and run from a Sample Sequence in Home Page.

Performing GC/MS acquisitions

Solvent Delays

There is no longer a limit on the positioning or number of solvent delays


within an acquisition.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 27


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Scanning acquisition
This is the default mode of acquisition. To add a scan function, literally hit
the Add Full command.

28 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

SIM acquisition
There are no constraints on how many ions get acquired in a data file. In
reality, there is an upper limit of 511 individual ions which may be grouped
into any number of SIM groups.

The start and end times can be defined in the table or by moving the
graphical bars with the mouse.

At the time of writing, the grouping control was unfinished for TRACE MS
Plus.

Mixed Scan and SIM acquisition


TRACE MS Plus can be used analytically to perform scanning acquisitions
as well as SIM. There is complete freedom of overlap, simultaneous or
discreet groupings.

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 29


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

Use the Add Full and Add SIM HTML links to add Scan Functions:

Now use the Preview | All Scans to view all Scan Functions.

30 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

They may be edited as needed. There are no constraint on start and end
times, and Scan Functions can overlap.

If the green scan function is clicked-and-dragged, then its position and times
can be varied with the mouse.

Toggle between Scan and SIM functions with the slide controls:

_____________Basic Operation of a TRACE MS Plus for Engineers _________________ 31


Basic Operation of the TRACE MS PLUS for Engineers
_____________________________________________________________________________

When you click on any graphical scan function, its properties are shown in a
pop-up window:

32 _________________ Basic Operation of a TRACE MS Plus for Engineers_____________

Anda mungkin juga menyukai