_____________________________________________________________________________
Introduction
This document assumes the reader already has a working familiarity with
TRACE MS/Voyager/MD800 control via the Xcalibur data system.
TRACE MS Plus requires Xcalibur 1.3.
2. Launching Xcalibur.
When the Server is running, there will be an icon in the Windows Start
menu, as shown below.
It is possible to access the Manual Tune dialog from here if it is not already
open.
Venting/Pump-down operations
Tune Dialog
The Tune dialog represents a radical departure from earlier generation user
interfaces. When it first launches the display is seen to be organised into
two regions.
Graphical Display
The top of the dialog is occupied by one or more mass range graphical
displays - dependent on how many have been specified/activated – and each
will have annotation indicating Absolute Intensity, Centroided Mass
Measurement and Display Gain (Zoom) of the individual mass selection.
Tabular Display
detector voltage setting and peak format and each may be selected in the
graphical region above.
The example shows 7 Scan Functions that have been compiled and together
these define the most routine interactive procedures associated with
maintenance of a MS.
FS1 through FS3 show individual ions from the spectrum of PFTBA
(Heptacosa); these would be used to review sensitivity and mass resolution
for routine analysis.
FS4 has been designed to show a wide mass range and therefore allowing
inspection of the instrument ion background; this would be useful to monitor
contaminants such as column and septum bleeds, hydrocarbon background
from dirty samples and plasticizer ions that could potentially be introduced
during column installations and/or source cleaning.
FS5 through FS7 have been set up to monitor for potential air leaks,
referenced against Helium carrier gas.
The table also permits time windows to be applied to any or all Scan
Functions. Although the default setting is zero, a time range, say 0.00min to
1.00min would enable an acquisition to be performed from the tune dialog,
independent of Xcalibur. This is useful for performing diagnostic manual
mass calibrations very rapidly.
Pumping Down
When the instrument is vented, the Server icon will be red (Vented).
Select the Pump menu item to start the pump down procedure. The Server
icon will flash yellow until high vacuum has been achieved AND
turbomolecular pumps achieve adequate frequency of rotation, where after it
will stop flashing and adopt a continuous yellow appearance.
Venting
When the instrument is pumped down, the Server icon will be yellow
(Operate OFF) or green (Operate ON). The right-click operation will
reveal the following menu:
Select Yes to Vent the MS, if this is required. Venting will be immediate,
with the deactivation of the Vent Valve solenoid attached to the front (or
only) turbomolecular pump.
Autotuning
Select the Instrument Tune and Calibration menu item from the Server icon.
The Tune dialog can be seen to update with the Autotune algorithm.
Ultimately, the log of events can be copied to the clipboard for pasting into a
document. This may be useful for diagnostics, if a problem is suspected, or
as a teaching aid for learning manual tuning.
The Tune dialog will show a full scan across the reference ions of the
internal tuning compound, PFTBA (FC43).
The Reference gas will be turned OFF automatically and Operate will be
set back to its previous state.
This dialog shows the Standard Tune view. There are no Tune parameters
to adjust. It is assumed that Autotune is used to set up the instrument.
When the Tune page is re-opened you will see additional Per Method
Parameters.
Select the Standard Mass Scale Calibration option and follow the
instructions:
There will be an option to print off a report prior to closing the dialog.
The Raw File must contain spectra acquired over an adequate mass range
with satisfactory mass resolution.
Solvent Delays
Scanning acquisition
This is the default mode of acquisition. To add a scan function, literally hit
the Add Full command.
SIM acquisition
There are no constraints on how many ions get acquired in a data file. In
reality, there is an upper limit of 511 individual ions which may be grouped
into any number of SIM groups.
The start and end times can be defined in the table or by moving the
graphical bars with the mouse.
At the time of writing, the grouping control was unfinished for TRACE MS
Plus.
Use the Add Full and Add SIM HTML links to add Scan Functions:
Now use the Preview | All Scans to view all Scan Functions.
They may be edited as needed. There are no constraint on start and end
times, and Scan Functions can overlap.
If the green scan function is clicked-and-dragged, then its position and times
can be varied with the mouse.
Toggle between Scan and SIM functions with the slide controls:
When you click on any graphical scan function, its properties are shown in a
pop-up window: