2008 Seminar
Max Setup Routines DC 250
• IOT Baseline
dC924 TRC Adjust
dC937 ProCon
• IOT Situational
dC991 Tone Up / Tone Down
dC931 Manual In/Out Setup
2008 Seminar
Max Set Up
• Max Set Up should only be performed when directed
to do so in the Service Manual
• Process Control adjusts Bias voltage, Laser Diode
Power, and Grid/BCR Voltage automatically to maintain
density.
• Process Control maintains Toner Concentration
automatically, to maintain density. It uses the ATC
Sensor and Pixel count to the determine dispense time
needed to attain the target TC.
• Most of the procedures, in Max Set Up, only measure
parameters of process control but do nothing to effect
it. They can be used as a troubleshooting tool but not as
a adjustment to correct Image Quality
10/9/2008 2008 Seminar 3
Entering Diagnostics
Y 0 0 0
M 0 0 0
C 0 0 0
K 0 0 0
• If the machine is functioning correctly, and the customer does not like the way
Process Control sets up density, you can use dC924 to adjust density levels to
satisfy the customer.
• TRC (Tone Reproduction Curve) Adjust is used to adjust image density as desired by
adjusting the center value of tone correction in the range of low/mid/high density for each
Y, M, C and K.
• The Default value (Mid Range) is 0. By adjusting TRC you will affect both prints and
copies.
• This should only be used, as a last resort, if the customer is complaining of color
densities of both prints and copies. These values should always be returned to 0
when baselining the machine (performing max setup and/or calibration)
Y M C K
• The ATC sensor calibration value (sensor gain) is printed on each sensor, and must
be entered whenever a Developer Housing or ATC sensor is replaced This is
accomplished by doing adjustment 9.2 ATC Sensor Set Up (DC 950).
• Locate the ATC Sensor calibration code on the ATC Sensor. This is the 3-digit number in
the 3rd line of text on the label (it will always start with a zero).
• Enter the last 2 digits of this code into the appropriate column of the first row.
• After the routine is completed the Correction Coefficient, and Correction Offset will be
calculated and displayed in the appropriate boxes.
• This procedure must be performed in order for the ATC Sensor to function accurately.
• When Performing this adjustment using the UI make sure you press the Green Start
button on the UI to execute the routine.
Y M C K
Correction Offset 0 0 0 0
• If all the bar code numbers are at 55, they are at default. You must input in the
ATC Sensor Calibration Value (sensor gain) that is printed on each sensor
• This could occur if the CSE did an IOT initialization without doing a save and
restore.
• With this set up the sensors would not function accurately.
• If the original ATC sensors are in the machine use the ATC Sensor Bar Codes
listed on the factory sheets located in tray 1.
2008 Seminar
Procon
• There are no adjustments in ProCon, it is a
measurement tool only
• It should be used to evaluate how the machine thinks it is
performing
• It will also monitor different functions of Process Control
and render judgments if they functioning correctly or not.
• Running this procedure will have the same effect on
Image Quality as the when the machine runs a Mini Set
up.
• Again you have no input it is strictly a measurement tool
• ADC H is used to measure the densities of high area coverage patches (50% for K 75% for CMY).
• ADC Target H is the adjusted target at which the machine should run to maintain correct Image
Density for “High Area” coverage. These Targets are effected by HFSI life and other factors like
Temperature and Humidity.
• Measure ADC H for all colors is the calculated density measurement for High Area patches (2) for
that color. The lower the value the darker the patch the higher the value the lighter the patch.
• The result of ADC H Measure should be within +25 bits of the target value CMY and +50 bits for K
• Things to look for if the ADC H Measure value is not within + 25 Bits CMY or + 50 bits K of the ADC
Target H value
• If Mid/Hi density of copies/prints is acceptable do nothing
• If Mid/Hi density of copies/prints is not acceptable check how ATC is functioning. If ATC is good
go to next step. If ATC is out troubleshoot the Dispense and Development system
• If Mid/Hi density of copies/prints is not acceptable and ATC is good suspect the Xero CRU,
MOB Sensor Assembly, IBT Belt, contacts of the flat cables on the MCU to IOT PWB’s, ROS
LD, MCU PWB, HVPS.
• ADC L is used to measure the densities of Low Area coverage patches (18% CMYK).
• ADC Target L is the adjusted target at which the machine should run to maintain correct Image
Density for “Low Area” coverage. These targets are effected by HFSI life and other factors like
Temperature and Humidity.
• Measure ADC L for all colors is the calculated density measurement for the Light Area patch for
that color. The lower the value the darker the patch the higher the value the lighter the patch.
• The result of ADC L Measure should be within +25 bits of the target value CMY and +50 bits for K
• Things to look for if the ADC L Measure value is not within + 25 Bits CMY or + 50 bits K of the
ADC Target L value
• If Low Density of copies/prints is acceptable do nothing
• If Low density of copies/prints is not acceptable check how ATC is functioning. If ATC is
good go to next step. If ATC is out troubleshoot Dispense and Development system
• If Low density of copies/prints is not acceptable and ATC is good suspect the Xero CRU,
MOB Sensor Assembly, IBT Belt, contacts of the flat cables on the MCU to IOT PWB’s, ROS
LD, MCU PWB, HVPS.
• These are the three parameters the machine use/adjusts to achieve the ADC H Target. They are
adjusted by process control, there are no manual adjustment procedures. They are used to
control image density.
• Grid Voltage
• Grid Voltage Setting is the Bias setting the machine is using to achieve the ADC H target value
• Grid voltage will change as needed to develop a good ADC H patch.
• For K, Grid Voltage is scorotron voltage, for CMY Grid Voltage is Bias Roll Voltage.
• LD Power
• LD Power is the Laser Diode setting the machine is using to achieve the ADC H target value
• LD will change as needed to develop a good ADC H patch.
• Bias Settings
• Bias Setting is the Developer Bias voltage the machine is using to achieve the ADC H target value
• Bias Setting will change as needed to develop a good ADC H patch.
10/9/2008 2008 Seminar 16
ADC Min 1 ADC Min 2 ProCon On
Y M C K
Measurement of ADC 269 103 106 201
Min 1
Measurement of ADC 328 289 268 326
Min 2
• Used to extrapolate the value for ADC H by measuring two patches Min 1 and
Min 2
• ADC Min 1 must be at least 10 bits lower (darker) then Min 2 (normally lower by
50 bits or more). If the difference is less then 10 bits the setup will fail (ADC
Minisetup Fail)
• ADC H will normally fall between Min 1 and Min 2 on a new machine with new
IBT, Drum, and Developer. If it does not and Mid/Hi density is acceptable there
is no problem.
• Min 1 and Min 2 are also used to calculate LD Power, Grid Voltage, and Bias
voltage (See diagram next slide)
Grid
volt.
Grid BC (large)
Grid BC (small)
Bias
Bias Large volt.
Bias (small)
RADC_Mini2
RADC_Target RADC
RADC_Mini1 LD (small)
LD
• ATC Target is the adjusted target. It is determined by adjusting the standard ATC Target in NVM,
using a sampling of the following: Temperature, Humidity, Developer Life.
• Measured ATC Ave is the actual measured average of several samplings of the ATC sensor. The
lower the value the higher the Toner Concentration. ATC measure should be within + 25 bits of the
ATC Target
• If the difference between the ATC Target and the Measured ATC Ave is greater then + 25 bits run 50
to 100 copies of your standard test pattern. This should stabilize your system and bring the numbers
to within the 25 bit range.
• Things to check if your ATC values continue to drift off and result in CQ problems: Dispenser assy,
Toner Cartridge, Low Toner Sensor, ATC Sensor Set Up not performed, ATC Sensor, Developer
housing
• ADC Patch Fail - The measured ADC Patch density is light (Over 75% of Vclean). Vclean is the bear
IBT belt reading by the ADC Sensor. If this is NG go to the 92-652 RAP in your service
documentation
• ADC Minisetup Fail - The two levels of ADC Patch density created in Mini-Setup: high (LD_OUT+60)
and low (LD_OUT-60) are reversed. If this is NG go to the 92-663 RAP in your service documentation
• ATC Average Fail – The output from the ATC sensor was not detected. If NG go to RAP’s 92-653 (Y),
92-654 (M), 92-655 (C), or 92-656 (K) in your service documentation
• ATC Amplitude Fail – the amplitude value is smaller then 10 bits. If NG go to RAP’s 92-657 (Y), 92-
658 (M), 92-659 (C), or the 92-660 (K) in your service documentation
These values are chosen in a Look Up Table (LUT) based on drum and developer
life.
Note: In this case, Y, M, C drum life are equal so LUT values are equal. Measured
ADC H is not as close to the ADC H Target when LUT values are used.
• Key Point. If troubleshooting a density problem, Visually inspect output of
Procon “On” and Procon “Off” Print.
If the Procon “On” Print has a density problem, but Procon “Off” Print does not, then
it is best to troubleshoot the process control system (ADC sensor/harnessing, IOT
software, MCU PWBA).
If the Procon “On” Print and the Procon “Off” Print have a density problem, then it is
best to troubleshoot the image path (transfer, developers, ROS, MCU PWBA,
harnessing, DFE)
10/9/2008 2008 Seminar 23
Procon Off
Procon “On” Print Procon “Off” Print
(Y) (M) (C) (K) (Y) (M) (C) (K)
ADC Target H 350 300 310 270 350 300 310 270
Measured ADC H 354 300 299 277 292 266 272 310
Measured ADC Min1 224 205 216 258 227 214 206 260
Measured ADC Min 2 356 339 356 364 364 347 344 363
ADC Target L 957 941 946 888 957 941 946 888
Measured ADC Target L 964 941 942 894 955 933 940 920
Charge Voltage Settings 557 585 590 703 610 610 610 660
Bias Setting 442 463 467 587 482 482 482 555
ATC Target 643 623 623 668 643 623 623 668
Measured ATC Average 650 629 631 686 650 629 634 688
Temperature 28 - - - 28 - - -
Humidity 15 - - - 15 - - -
Temperature Fail OK - - - OK - - -
Humidity Fail OK - - - OK - - -
Y M C K
No. of Cycles 0 0 0 0
• This procedure compares measured toner concentration against a target, and allows manual
adjustment of TC.
• ATC Target is the adjusted target. It is determined by adjusting the standard ATC Target in NVM using
Temperature/Humidity values, and Developer Life.
• Measured ATC Ave is the actual measured average of several samplings of the ATC sensor. The lower
the value the higher the Toner Concentration, the higher the number the lower the toner concentration.
ATC measure should be within + 25 bits of the ATC Target
• For each color that is above or below target, enter a value between -99 and 99 into the top row. This
value is the number of tone up or tone down cycles to be performed. Negative values increase the ATC
Measured Value (less toner concentration); positive values decrease ATC Measured Value (more
Toner Concentration).
• Repeat as required until ATC Measured Value matches ATC Target.
• A more preferred way to stabilize TC would be to run 50 to 100 copies of your standard test pattern
82E13120.
• Things to check if your ATC values continue to drift off and cause CQ problems: Dispenser assy, Toner
Cartridge, Low Toner Sensor, ATC Sensor Set Up not performed, ATC Sensor, Developer housing
Y M C K
Low Density 0 0 0 0
Medium Density 0 0 0 0
High Density 0 0 0 0
• Changes made in the dC919 Adjustment will effect the copy mode only.
• This is a fine adjustment for very hard to please customers.
• You can adjust the center set point + 4 steps with 0 being the default number. If you increase the value
you will cause that color to be darker. If you decrease the value you will cause that color to be lighter.
• The customer has a coarse adjustment for Color Balance on the image quality tab. There are + 3 steps
in the course adjustment.
• dC919 should be used as follows: The customer complains that his output is too light for a certain color.
The customer then selects the Image Quality tab and adjusts color balance for that color. He selects +2
and The color is too light. He then selects +3 and that is too dark. DC919 will allow the CSE to fine
adjust this color by changing the fine adjust center point until the customer is satisfied.
• So, Image Quality Color Balance on the UI is a Coarse Adjustment, CSE Diagnostic Color Balance is a
Fine Adjustment. Again, this would only be used for very hard to please customers.
Reflection Ratio
R G B
Y 46 96 214
C 239 170 99
Degree Marks
Judge 3 3
The purpose of this adjustment is to align the CCD with the Lens. This procedure
should only be performed if the Lens or CCD is replaced, or if the documentation
specifically directs.
Presented by:
Katie Lazarus-Born
Norm Srodawa
Jim Strollo