Cert Densitometer Sn. E311404 - 1
Cert Densitometer Sn. E311404 - 1
mt
@ ratita
SERTIFIKAT KALIBRASI
+{trflIrrri
LK{97IDN
www.pratita.com
tr*ee+iE
':ft4.-gl--*?
H$dfi4&;
i&-{:S.ra{trr
EF5ffiHI
{ALI 8 RAT I O N C E RTI F I CATE
Tanqqal : 19 Desember 2A19 Nomor : 047 ILCIDTM/PPND(III19 F-CP-05-02
Date Number
Pelanqgan .{
: Lokasi Kalibrasi :
Customer Locatian of calibration
PT. Pesat Servis lndustri Lab. Kalibrasi FT. Pratita Prama Nugraha
lkhsan. AMTE
Koordinator Divisi
Division Coordinator
@ratita
Tanscal : 19 Desember2019 Nomor : 04TiLCIDTM/PPND(Ili19 F-CP-05-02
Dafe Number
HASIL KALIBRASI
CALIBRATION RESUTIS
lnstrument
Step lnstrument Measurement
Tablet
readings Deviation Limit Result
readings Uncertainty
Before {oD} (oD)
{oD} Adjusment {oD} {passffail} {oD)
(oD)
Step 3
{0. s1}
0.85 0.93 0.a2 i 0.05 pass
Step 6
{1.80)
1.76 1,80 0.01 r 0.05 pass
U = 0.10
Step 7
i2. 11i
203 2.$ -0.01 * 0.05 pass CL = 95%
litep 10
k=2
(3.03) 2.92 3.00 -0.03 * 0.05 pass
Step 13
f4.s0)
3.80 388 -0 02 r005 pass
Catatan / Nofes.- Pemb*aan alat merupakan nilai rata-rata- diambildari tiga kali pengukuran
Ihe insfrument readings are average value, taken from tbree measurements
- Nilai$ebenarnya adalah nilai pembacaan dikurangi nilai penyimpangan
Deyiafion are reduce fo fire reading for obtaining true reding
- Nilai ketidakpastian didapat dengan tingkat kepercayaan g5% dan fakto;
cakuPan (k) = Z
The uncerlainty is taken d a convidence level 95 % and caverage factar {k) = 2
Halaman 2dari2
. PageZof 2