CARA KALIBRASI LAMPU

DILAKUKAN SETIAP HASIL BERMASALAH DAN SETIAP PERGANTIAN BOLA LAMPU

MENU AWAL --- MAINTENANCE --- MAINTENANCE (SUDUT KANAN ATAS) --- GAIN ADJUST
---STAR ---(PLEASE ASPIRATE) BIARKAN PROSES BERJALAN --- SAVE --- RETURN --- RETURN

SOP

1. HIDUPKAN ALAT DENGAN MENEKAN TOMBOL ON/OFF DIBELAKANG ALAT.


2. CEK TEMPAT PEMBUANGAN LIMBAH
3. SAAT STARTUP AKAN DIMINTA AQUADEST. .IKUTI LANGKAH YANG ADA DI LAYAR.
TUNGGU HINGGA MENU UTAMA MUNCUL
4. UNTUK PEMERIKSAAN KLIK TEST KEMUDIAN PILIH PARAMETER YANG AKAN
DIGUNAKAN
5. KLIK WASH DENGAN MENGGUNAKAN AQUADEST
6. TUNGGU HINGGA LAMPU DALAM KEADAAN READY
7. KLIK WATER BLANK SIAPKAN AQUADEST KEMUDIAN ASPIRATE
8. KLIK REAGEN BLANK SIAPKAN REAGEN BLANK KEMUDIAN ASPIRATE
9. KLIK PADA CALIBRATE SIAPKAN STANDART KEMUDIAN ASPIRATE, KLIK SAVE
10. KLIK SAMPEL SIAPKAN SAMPEL KEMUDIAN ASPIRATE

CATATAN :

 SETIAP PERGANTIAN PARAMETER PEMERIKSAAN LAKUKAN WASH 3X TERLEBIH


DAHULU
 APABILA DALAM PROSES PENGERJAAN SAMPEL PADA MENU MUNCUL
CALIBRATION, LAKUKAN KALIBRASI TERLEBIH DAHULU DENGAN MENEKAN MENU
CALIBRASI DAN MASUKKAN AQUADES. KEMUDIAN LANJUTKAN PEMERIKSAAAN.
 ALAT DALAM KEADAAN STANDBY LAMPU MATI, KETIKA ALAT KITA AKTIFKAN
PERHATIKAN LAMPU TERLEBIH DAHULU YANG TRDAPAT PADA TENGAH LAYAR ATAS.
( KLIK GAMBAR LAMPU SAMPAI BERWARNA DAN TUNGGU SAMPAI LAMPU READY )
ALAT BARU BISA DIPAKAI

CARA MEMATIKAN ALAT

SETELAH SELESAI PEMERIKSAAN KEMBALI KE MENU AWAL – TEKAN SHUTDOWN –


ASPIRATE AQUADES – OK – PLEASE SHUTDOWN – OK – TEKAN TOMBOL POWER DI
BELAKANG ALAT