T his handbook presents the basic principles of reversed-phase HPLC for the
analysis and purification of polypeptides. For further details regarding
reversed-phase HPLC separations of polypeptides please refer to the technical
references at the back of the Handbook or contact the Grace Vydac Technical
Support Group.
Table of Contents
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
Mechanism of Interaction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Appendices
Appendix A: Column Characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
Appendix B: The Care and Maintenance of Reversed-Phase Columns . . . . .53
Appendix C: The Effect of Surfactants on Reversed-Phase Separations . . . .56
Appendix D: Ion Exchange Chromatography:
Orthogonal Analytical Techniques . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
Appendix E: The Effect of System Hardware
on Reversed-Phase HPLC Polypeptide Separations . . . . . . . . . . . . . . . . . . . .60
Technical References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
www.gracevydac.com
(株)ケムコ
TEL 06-6375-0395 URL : http://www.chemco.jp/ 1
58
examples where RP-HPLC has been 40 32
28
used to separate similar polypeptides. 15 13 19
9
Insulin-like growth factor with an
84 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99-00
oxidized methionine has been
separated from its non-oxidized Figure 1. RP-HPLC separates rabbit and Figure 2. The number of patents issued by
analogue2 and interleukin-2 muteins human insulin that differ by only a single the United States Patent Office in the years
amino acid. Column: VYDAC® 214TP54 1984–2000 in which VYDAC® Reversed-Phase
Eluent: 27–30% acetonitrile (ACN) in 0.1% HPLC columns are referenced in the
TFA over 25 minutes at 1.5 mL/minute. patented process.
(株)ケムコ
2
TEL 06-6375-0395 URL : http://www.chemco.jp/ 3
A
important in understanding RP-HPLC mobile phase and only a part of the and the hydrophobic phase. 25
polypeptide separations. The separation molecule—the “hydrophobic foot”— Biphenyl
of small molecules involves continuous in contact with the RP surface. Because the number of organic 20
partitioning of the molecules between RP-HPLC separates polypeptides modifier molecules required to
k1 (retention)
15
the mobile phase and the hydrophobic based on subtle differences in desorb a polypeptide—called the ‘Z’
stationary phase. Polypeptides, the “hydrophobic foot” of the number by Geng and Regnier4—is 10
however, are too large to partition polypeptide being separated. very precise, desorption takes place
into the hydrophobic phase; they Differences in the “hydrophobic within a very narrow window of 5
adsorb to the hydrophobic surface foot” result from differences in organic modifier concentration.
0
after entering the column and remain amino acid sequences and differences This results in complete retention 0 10 20 30 40 50 60 70 80 90
adsorbed until the concentration of in conformation. until the critical organic modifier Acetonitrile (%)
organic modifier reaches the critical concentration is reached and sudden
concentration necessary to cause desorption of the polypeptide takes B
desorption (Figure 3). They then place (Figure 4). The sensitivity of 25
Lysozyme Biphenyl
desorb and interact only slightly polypeptide desorption to precise
20
with the surface as they elute down concentrations of organic modifier
k1 (retention)
the column4. accounts for the selectivity of 15
RP-HPLC in the separation of
polypeptides. The sudden desorption 10
Figure 3. Polypeptide enters the column in the mobile phase. The hydrophobic “foot”
of the polypeptide adsorbs to the hydrophobic surface of the reversed-phase material
where it remains until the organic modifier concentration rises to the critical concentration
and desorbs the polypeptide.
(株)ケムコ
4
TEL 06-6375-0395 URL : http://www.chemco.jp/ 5
k' (retention)
the hydrophobic foot but elsewhere changes in the modifier concentration partitioning and adsorption. They 15
in the molecule as well. Kunitani makes isocratic elution difficult desorb more quickly with changes in
10
and Johnson3 found that, due to because the organic modifier organic modifier concentration than
conformational differences, very concentration must be maintained small molecules which partition, 5
similar interleukin-2 muteins could very precisely. Gradient elution is however they desorb more gradually
be separated, including those usually preferred for RP-HPLC than proteins (Figure 6), suggesting 0
0 10 20 30 40 50 60 70 80 90
differing in an oxidized methionine polypeptide separations, even if the a hybrid separation mechanism.
Acetonitrile (%)
or in single amino acid substitutions. gradient is very shallow—i.e., a Attempts to correlate peptide retention
Geng and Regnier4 found that the ‘Z’ small change in organic modifier with side chain hydrophobicity have Figure 6. The retention behavior on RP-HPLC
of many peptides is midway between that
number correlates with molecular concentration per unit time. been partially successful, however of proteins and of small molecules.
weight for denatured proteins, tertiary structure in many peptides Pentaphenylalanine, a small peptide, desorbs
more quickly than biphenyl, a small molecule,
however, proteins with intact tertiary limit interactions to only a portion of but more gradually than lysozyme. Small
structure elute earlier than expected the molecule and cause discrepancies peptides appear to chromatograph by a
in the predictions of most models. mixed mechanism.
because only the “hydrophobic foot” Effect of Acetonitrile
is involved in the interaction, while Concentration on Elution It has been shown that the exact
the rest of the protein is in contact location of hydrophobic residues
with the mobile phase. 42% ACN in a helical peptide is important in
predicting peptide retention5. myoglobin having column efficiencies
40% ACN
The adsorption/desorption step only 5–10% of the efficiencies
takes place only once while the Because large polypeptides obtained with small molecules such
polypeptide is on the column. After 39% ACN diffuse slowly, RP-HPLC results as biphenyl. Gradient elution of
desorption, very little interaction in broader peaks than obtained polypeptides, even with shallow
takes place between the polypeptide with small molecules. Peak widths gradients, is preferred, since it results
and the reversed-phase surface and of polypeptides eluted isocratically in much sharper peaks than isocratic
0 Time (min) 20 are a function of molecular weight, elution. Isocratic elution is rarely
subsequent interactions have little
affect on the separation. with large proteins such as used for polypeptide separations.
Figure 5. At 39% ACN, the retention time of
lysozyme is nearly 18 minutes. Increasing
A practical consequence of this the ACN concentration to 40% reduces the
retention time by more than half, to 7.6
mechanism of interaction is that minutes. Increasing the ACN concentration to
polypeptides are very sensitive to 42% reduces the retention time of lysozyme
again by more than half, to 3.1 mintues.
organic modifier concentration. The Column: VYDAC® 214TP54 Eluent: ACN at
sensitivity of polypeptide elution to 39, 40 and 42% in 0.1% aqueous TFA.
(株)ケムコ
6
TEL 06-6375-0395 URL : http://www.chemco.jp/ 7
(株)ケムコ
8
TEL 06-6375-0395 URL : http://www.chemco.jp/ 9
B
adsorbent. Distinct differences in
separation selectivity of the peptides
Monomerically bonded
8
14 is noted, offering yet another option
10
12 in column selection when developing
Figure 9. Low carbon load C18 RP-HPLC Figure 10. Columns: VYDAC® 218TP54 4,5
9
15
16
peptide separations.
7 11
column (B) separated two peptides that were (C18); 214TP54 (C4); Eluent: 0–30 % ACN 1 2 13
3 6
only partially resolved on a standard carbon in 0.1% aqueous TFA over 60 minutes at
load column (A). Columns: A. VYDAC® 1.0 mL/min. Sample: tryptic digest of
218TP52-standard C18, 5 µm, 2.1 x 250 mm β-lactoglobulin A.
0 10 20 30
B. VYDAC® 218LTP52– low carbon load–C18, Minutes
5 µm, 2.1 x 250 mm Eluent: 6 mM TFA/4 mM
HFBA, 11–95% ACN in 75 min at 0.25 mL/min Figure 11. Columns: VYDAC® 218TP54
Sample: Asp–N protein digest. Data courtesy polymeric and 238TP54 monomeric (C18, 5 µm,
of H. Catipon and T. Salati, Genetics Institute, 4.6 x 250 mm) Eluent: 10–40% ACN with
Andover, MA. 0.1% TFA over 30 min. Flowrate: 1.0 mL/min.
(株)ケムコ
10
TEL 06-6375-0395 URL : http://www.chemco.jp/ 11
(株)ケムコ
12
TEL 06-6375-0395 URL : http://www.chemco.jp/ 13
4.81
90% 24.14 Max. 8.7e4 cps.
(See Appendix A, Page 50). Low 3.30 (a)
80%
flow rates can significantly reduce the 70%
5.23
Total-Ion Chromatogram (TIC)
3.06 21.42
53.74
amount of solvent needed for 60% 27.60
polypeptide separations. 50% 5.59 27.96
2.56
40% 32.14
17.02 56.82
32.87
30%
Separation of the Tryptic Digest Separation of the Tryptic Digest 19.10
29.34
Figure 16. Separation of the tryptic digest of bovine serum albumin (BSA) on a 300 µm i.d.
Figure 14. Separation of the tryptic digest of Figure 15. Separation of the tryptic digest of capillary RP-HPLC Sample: 3 pmole. Column: VYDAC® 218MS5.305 5 µm, 300 Å, polymeric-C18
hemoglobin on a microbore RP-HPLC column myoglobin on a capillary RP-HPLC column. reversed-phase (300 µm i.d. x 50 mm L). Flow: 5 µL/min. Mobile phase: A = 0.1% formic acid,
(Reference 26). Column: C18, 1.0 x 250 mm Column: C18 (VYDAC® 218MS), 75 µm i.d. 98% water, 2% ACN. B = 0.1% formic acid, 98% ACN, 2% water. Gradient: Hold 3% B from
(VYDAC® 218TP51). Flow rate: 50 µL/min. capillary.Flow rate: 0.5 µL/min. Eluent: 0 to 5 minutes. Then ramp from 3% B to 50% B at 65 minutes. Final ramp to 75% B at 70 minutes.
Eluent: Gradient from 0 to 40% B over 50 water/TFA/acetonitrile gradient. Detection: MS. (a) Total ion count. (b) Base peak intensity. The base peak is defined as the
minutes, where Solvent A is 0.1% TFA in water single mass peak with maximum amplitude at each time in the chromatogram. The base peak
and Solvent B is 0.1% TFA in acetonitrile. chromatogram emphasizes peaks containing a single predominant molecular species and
deemphasizes heterogeneous peaks and noise. Data courtesy of Applied Biosystems.
(株)ケムコ
14
TEL 06-6375-0395 URL : http://www.chemco.jp/ 15
Interface with mass spectrometry Capillary columns can be interfaced The desorption and elution of ■ It has a low viscosity, minimizing
Direct transfer of the HPLC with electrospray mass spectrometer polypeptides from RP-HPLC column back-pressure;
eluent into the electrospray mass interfaces or even nanoelectrospray columns is accomplished with ■ It has little UV adsorption at low
spectrometer interface is possible interfaces after stream splitting. aqueous solvents containing an wavelengths;
with small bore columns and organic modifier and an ion-pair ■ It has a long history of proven
attomole (10-18) levels of individual An article by Davis and Lee provides reagent or buffer. The organic reliability in RP-HPLC
sample are routinely detected using valuable information for getting the modifier solubilizes and desorbs the polypeptide separations.
sophisticated MS equipment. best performance using microbore polypeptide from the hydrophobic
(See LC/MS, Pages 28–31). and capillary columns44 and is surface while the ion-pair agent or Isopropanol
recommended reading for anyone buffer sets the eluent pH and Isopropanol is often used for large or
Current Trends in embarking on the use of small bore interacts with the polypeptide to very hydrophobic proteins. The major
Small Diameter Columns columns. A number of journal articles enhance the separation. Elution is disadvantage of isopropanol is its
detail the use of mass spectrometers accomplished by gradually raising high viscosity. To reduce the viscosity
with capillary columns41–43 the concentration of organic solvent of isopropanol while retaining its
Narrowbore columns
(Also see Pages 26–29). during the chromatographic run hydrophobic characteristics, we
Narrowbore columns of 2.1 mm i.d.
(solvent gradient). When the solvent recommend using a mixture of 50:50
are run at 100–300 microliters per Examples reaches the precise concentration acetonitrile: isopropanol. Adding
minute. Narrowbore columns are a Microbore. Figure 14 illustrates necessary to cause desorption, the 1–3% isopropanol to acetonitrile has
practical step for most laboratories the separation of a tryptic digest polypeptide is desorbed and elutes been shown to increase protein
to take in reducing solvent usage and of hemoglobin on a microbore from the column. recovery in some cases52.
improving detection sensitivity. Most (1.0 mm i.d.) column.
standard HPLC systems can operate
Organic Modifiers
at these low flow rates with little or Capillary. Figure 15 is an example
no modification. Narrow bore of the separation of a tryptic digest The purpose of the organic solvent
Improved Resolution of Enzyme
columns with flow rates around 200 of myoglobin on a 75 µm i.d. is to desorb polypeptide molecules Subunits Using Low Gradient Slope
microliters/minute interface well with capillary column. from the adsorbent hydrophobic
pneumatically-assisted lectrospray surface. This is typically done by 0.25%/min
mass spectrometer interfaces. Capillary sample load. Figure 16 slowly raising the concentration of
organic solvent (gradient) until the 0.5%/min
illustrates that three picomoles of
Microbore and capillary columns polypeptides of interest desorb
a tryptic digest of BSA can be
Columns of 1.0 mm diameter and and elute.
separated on a 300 µm i.d.
less offer significant reductions in
capillary column. Detection was
solvent usage and increases in Acetonitrile (ACN)
by mass spectrometry.
detection sensitivity, however Acetonitrile (ACN) is the most
these may require modifications commonly used organic modifier Figure 17. Column: C18 (VYDAC® 218TP104)
to the HPLC system or the use of because: Flow rate: 1 mL/min. Eluent: Gradient slope
as shown. Gradient from 25–50% ACN in
instruments specifically designed ■ It is volatile and easily removed aqueous TFA. Sample: Subunits of cytochrome
for this purpose. from collected fractions; c oxidase. Data from reference 21.
(株)ケムコ
16
TEL 06-6375-0395 URL : http://www.chemco.jp/ 17
(株)ケムコ
18
TEL 06-6375-0395 URL : http://www.chemco.jp/ 19
(株)ケムコ
20
TEL 06-6375-0395 URL : http://www.chemco.jp/ 21
Figure 21. Elution of five peptides at pH 2.0, 4.4 and 6.5 with phosphate as the buffer.
Column: VYDAC® 218TP54 (C18, 5 µm, 4.6 x 250 mm). Eluent: 15–30% ACN in 30 min at 1.0 Figure 22. Column: VYDAC® 259VHP5415
mL/min; plus A. 20 mM phosphate, pH 2.0 B. 20 mM phosphate, pH 4.4 C. 20 mM phosphate, pH (PS-DVB, 5 µm, 4.6 x 150 mm) Eluent:
6.5 Peptides: 1. bradykinin 2. oxytocin 3. angiotensin II 4. neurotensin 5. angiotensin I. 15–30% ACN over 15 min. with A. 0.1% TFA,
pH 2. B. 15 mM NaOH, pH 9. Flowrate: 1.0
mL/min. Peptides: 1. oxytocin. 2. bradykinin.
3. neurotensin. 4. neurotensin 1-8. 5.
angiotensin III. 6. val-4 angiotensin III.
(株)ケムコ
22
TEL 06-6375-0395 URL : http://www.chemco.jp/ 23
(株)ケムコ
24
TEL 06-6375-0395 URL : http://www.chemco.jp/ 25
T he development of the
electrospray interface to couple
mass spectrometry with HPLC has
The combination of mass spectrometry
with HPLC reduces the need for
chromatographic resolution because
LC/MS Uses Short Columns
for Rapid Analysis
caused a virtual explosion in the of the resolving capacity of the mass The trend in LC/MS toward faster
use of LC/MS in the analysis of spectrometer. Analysis times are analyses with reduced resolution The Use of Low Concentrations
polypeptides. RP-HPLC peptide generally short to best utilize the has led to the use of relatively short of TFA for Peptide Separations
maps are routinely monitored by an sophisticated mass spectrometer. columns with very fast gradients.
on-line mass spectrometer, obtaining Detection sensitivity is often much 1 2 4
The trend toward reduced resolution 0.1% TFA
peptide molecular weights and better with mass spectrometry than 3
causing fragmentation of peptides with UV detection. and faster separations has led to the
to obtain sequence information. use of short columns packed with
smaller particle adsorbents than
normal. The most commonly used
particle size in short columns is three
0.05% TFA
Rapid Separation of Proteins Using Short (50 mm) Column micrometers. Using three micrometer 1 2 4
columns of five to ten centimeter
3
length with fast gradients enables
3 polypeptide separations to be
2 completed in just a few minutes.
1
4
Figure 24 shows the separation of
5 five proteins in less than five 0.02% TFA
1 2 4
minutes using a 50 mm long
column packed with 3 µm particles 3
using a fast gradient.
0 1 2 3 4 5
Minutes
Figure 24. Column: C18, 3 µm 4.6 x 50 mm (VYDAC® 238TP3405). Flow rate: 4.0 ml/min. 0.01% TFA
Eluent: Gradient from 20–45% ACN in aqueous 0.1% TFA in 4 min. Sample: protein standards. 1 2
(1) ribonuclease, (2) insulin, (3) cytochrome c, (4) BSA and (5) myoglobin.
4
3
0 10 20 30
Minutes
(株)ケムコ
26
TEL 06-6375-0395 URL : http://www.chemco.jp/ 27
with good polypeptide peak shapes the TFA entirely, relying on ion pair 0 5 10 15 20 25 30 35 40 45 50 55
using very low concentrations of TFA. reagents such as acetic acid. 145
MS-MS of peak at 21.78 minutes
y5
140
135 589.3375
K V P Q V S T P T L V E V S R
Tryptic Map Replacing TFA HPLC columns developed for low
130
125
with Acetic Acid (No TFA) TFA use enable the use of a wider
120
115
490.2675
110
selection of ion-pairing reagents to 105
100
95
optimize resolution of peptides.
Intensity, counts
90
85
Peptide separations can now be 80
y3
C4 (214MS54) 75
done with acetic acid or formic 70 361.2241 450.7655
65
acid acid replacing the trifluoroacetic 60
55
y6
I,L 702.4187
50
acid. Mixtures of ion-pair reagents 45
86.0946
R b3
40 325.1907 b7
can also be used to optimize a 35
72.0807 b2 y2
541.6491 y8
b9
30 y7 938.5566
peptide separation. 25 y1 a2 517.3188 803.4648
20 175.1184 b4 b5 552.3654 b10
15 465.5980 813.4511 y9
1051.6245
C18 (218MS54) 10
b6
693.3699 y10 b11
1001.5854
Example of Peptide Isolation 5
0
187.1308
747.4450 1088.6237
y11
0 Minutes 60
and Sequencing 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500
m/z, amu
Reversed-phase HPLC using Figure 27. Reversed-phase separation of tryptic digest peptides of bovine serum albumin (BSA)
Figure 26. Columns: Columns developed for
peptide separations in the absence of TFA. capillary columns with very small followed by MS determination of molecular weights of each peptide followed in turn by MS
Top: C4 (VYDAC® 214MS54); Bottom: C18 fragmentation of each peptide providing data to enable sequencing of the separated peptide.
sample loads coupled with mass Sample: 3 pmole of a tryptic digest of bovine serum albumin. Column: VYDAC® 218MS5.305, 5 µm,
(VYDAC® 218MS54;). Flow rate: 1 ml/min.
Eluent. Gradient from 0–30% ACN in 5 mM spectrometry has become a 300 Å, polymeric-C18 reversed-phase (300 µm i.d. x 50 mm). Flow rate: 5 µL/min. Mobile phase:
HOAc. Sample: Tryptic digest of apotransferrin. powerful tool for the isolation and A = 0.1% formic acid, 98% water, 2% ACN. B = 0.1% formic acid, 98% ACN, 2% water.
Gradient: Hold 3% B from 0 to 5 minutes. Then ramp from 3% B to 50% B at 65 minutes.
Final ramp to 75% B at 70 minutes. Detection: Triple quadrupole MS. Data from Reference 75.
(株)ケムコ
28
TEL 06-6375-0395 URL : http://www.chemco.jp/ 29
Intensity, cps
1e8 30.41 36.79 44.79 50.97
analyzed by Matrix-Assisted Laser MS results are compared to DNA 54.31
55.02
8e7
31.42 35.68 40.23
Desorption Ionization (MALDI) mass or protein databases for identification 36.99 42.97 55.33
6e7 38.21
spectrometry. The results are compared (Figure 28). 29.80 32.94 56.04
4e7 34.06
29.50
to protein and DNA databases for 57.76
2e7 28.49
identification of the isolated proteins.
0
26 28 30 32 34 36 38 40 42 44 46 48 50 52 54 56 58 60
Time, min
Proteomic Analysis of Cellular Proteins by Two-dimensional M+2H
Chromatography and Tandem Mass Spectrometry 2.8e6
(a)
816.2
Precursor ion selected by IDA Max. 2.8e6 cps
2.4e6
Survey scan mass spectrum
Intensity, cps
2.0e6 LC/MS Elongation factor Tu,
1.6e6 Gene PA4265
1.2e6
Band 16
Pseudomonas aeruginosa
8.0e5
Tandem Mass 4.0e5 755.1 843.3
942.0
M+H
597.4 677.1
Cellular Peptide Ion Reversed- spectrometry 766.6 927.4 1034.9 1125.8 1631.3
0
proteins fragments Exchange phase 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700
m/z, amu
213.0 b2 y5
1.20e4 a2 Product ion mass spectrum Max. 1.3e4 cps
(b) 185.0 597.5
1.00e4
LV E T L D S Y I P E P V R
Intensity, cps
y3
8000 371.3
a7 –18 y9
6000 711.5
Mass spectra of y7 y8
1075.5 y12 –18
peptides and MS 4000 342.3 443.3 563.3
874.3 960.5 y10 y11 y12
fragmentation 85.8 175.3 312.0 425.3 538.3 1188.8 1290.5 1419.0
2000 1401.3
Database Identification product ions 157.3
229.0
294.0
479.3 692.3 741.8
855.8
of cellular proteins 0
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500
m/z, amu
(株)ケムコ
30
TEL 06-6375-0395 URL : http://www.chemco.jp/ 31
(株)ケムコ
32
TEL 06-6375-0395 URL : http://www.chemco.jp/ 33
(株)ケムコ
34
TEL 06-6375-0395 URL : http://www.chemco.jp/ 35
Detection of Deamidation Separation of Oxidized Forms of RP-HPLC of Insulin-Like RP-HPLC of Modified Insulin-Like
by RP-HPLC Coagulent Factor from Native Protein Growth Factor Growth Factor
human MetO
growth (298) Oxidized native IGF
hormone MetO MetO ASP
ASP
45
45
S
CYS
CYS
52
52
ASP
ASP
53
53 Met
S
(298) and
46 51
51
46 S
S
CYS
CYS CYS
CYS PHE
PHE ARG
ARG
48 49 50
deamidation
47
MetO
47 48 49 50
S
S S
S desGlyPro
degradation (306) TYR
TYR
44
LEU
LEU
55
S S
S S
CYS
CYS
6
GLY
GLY
7
ALA
ALA
8
8
GLY
GLY
99
(株)ケムコ
36
TEL 06-6375-0395 URL : http://www.chemco.jp/ 37
Hemoglobin variants
A RP-HPLC method using a C4
column has been developed for the
separation of globin chains27. This
method has been used to study
hemoglobin variants in both animals
and humans. RP-HPLC has helped
to detect at least fourteen abnormal
hematological states in humans and
was used to study a silent mutant
involving substitution of threonine
for methionine34.
Figure 39. Eleven subunits of bovine cytochrome c oxidase ranging in MW from 4962 to
56,993 are separated by RP-HPLC. Column: VYDAC® 214TP104 (C4, 10 µm, 4.6 x 250 mm).
Eluent: 25–50% ACN over 50 min, then 50–85% ACN over 17.5 min; all with 0.1% TFA.
Flow rate: 1.0 mL/min. Inset: 35–45% ACN with 0.1% TFA over 40 min. Data from Reference 21
(株)ケムコ
38
TEL 06-6375-0395 URL : http://www.chemco.jp/ 39
(株)ケムコ
40
TEL 06-6375-0395 URL : http://www.chemco.jp/ 41
(株)ケムコ
42
TEL 06-6375-0395 URL : http://www.chemco.jp/ 43
(株)ケムコ
46
TEL 06-6375-0395 URL : http://www.chemco.jp/ 47
(株)ケムコ
48
TEL 06-6375-0395 URL : http://www.chemco.jp/ 49
300
Ethanol Gradient
200
mau
0
0 50 100 150 200 mL
Figure 47. Separation of Xenotropic Murine Leukemia Virus (XMuLV) from target protein during
preparative HPLC. Column: VYDAC® C4, 20-30 mm Elution: Ethanol gradient
Data courtesy of Holly Harker and Marcus Luscher, Amgen, Boulder, Colorado
(株)ケムコ
50
TEL 06-6375-0395 URL : http://www.chemco.jp/ 51
Column
Diameter
Typical
Flow Rate
Sample
Capacity
Maximum Practical
Sample Load
R eversed-phase HPLC columns,
if properly cared for, may give
good performance for over a thousand
Column Storage
RP-HPLC columns can be stored in
(mm) (1) (2) (3) organic solvent and water. For long
sample injections, depending on sample term storage the ion-pairing agent or
Capillary 0.075 0.25 µL/min 0.05 µg preparation and elution conditions. buffer should be rinsed from the
0.15 1 µL/min 0.2 µg Although the following ideas are column and the organic content
0.30 5 µL/min 1 µg specifically applicable to VYDAC® should be at least 50%.
0.50 10 µL/min 2 µg RP-HPLC columns, they also apply
to most other columns.
Chemical Stability
Microbore 1.0 25–50 µL/min 0.05–10 µg Reversed-phase HPLC columns are
Column Protection
stable in all common organic solvents
Column lifetime can be extended by including acetonitrile, ethanol,
Narrowbore 2.1 100–300 µL/min 0.2–50 µg filtering all solvents and samples isopropanol and dichloromethane.
and using an eluent filter and a guard When switching solvents it is
column. We recommend using an important to only use mutually
eluent filter between the solvent miscible solvents in sequence.
Analytical 4.6 0.5–1.5 mL/min 1–200 µg 10 mg delivery system and the injector to Silica-based RP-HPLC columns
trap debris from the solvents, pumps or are stable up to pH 6.5 to 7 and are
mixing chamber. We also recommend not harmed by common protein
Semi-preparative 10 2.5–7.5 mL/min 1,000 µg 50 mg using a guard column between the detergents such as sodium
injector and the column if samples dodecylsulfate (SDS).
contain insoluble components or
Preparative 22 10–30 mL/min 5 mg 200 mg compounds that strongly adsorb to
Pressure and
the material.
Temperature Limits
Column Conditioning RP-HPLC columns are generally
Process 50 50–100 mL/min 25 mg 1,000 mg stable to 60˚C and up to 5,000 psi
100 150–300 mL/min 125 mg 5,000 mg Because of the nature of the
(335 bar) back-pressure. Typical
reversed-phase surface, column
back-pressures for RP-HPLC
performance (resolution, retention)
columns are shown in Figure B-1.
may change slightly during the first few
injections of proteins. A column can
be conditioned by repeated injections
Figure A-1. of the protein until the column
1. Actual flow rates can be a factor of two higher or lower depending on the method.
characteristics remain constant
2. Sample Capacity is the quantity of polypeptide that can be loaded onto the column (requires injection of about 100 µg of
without reducing resolution.
protein) or by injection of 100 µg of
3. Maximum Practical Sample Load is approximately the maximum quantity of sample a commonly available protein, such as
that can be purified with reasonable yield and purity on the column. ribonuclease, followed by running an
acetonitrile/0.1% TFA gradient.
(株)ケムコ
52
TEL 06-6375-0395 URL : http://www.chemco.jp/ 53
(株)ケムコ
54
TEL 06-6375-0395 URL : http://www.chemco.jp/ 55
A. No SDS
B. With 0.5% SDS B. With 0.5% SDS
B. 0.001% SDS
D. 0.1% SDS
Figure C-1. Surfactants affect chromatography Figure C-2. Surfactant affects chromatography
(B) but do not harm column or subsequent (B) but does not harm column or subsequent
separations (C). Column: VYDAC® 218TP54. separations (C). Column: VYDAC® 214TP54 Figure C-3. The presence of even trace amounts of SDS causes a loss in resolution in a peptide
Eluent: 24–95% ACN in 0.1% TFA over Eluent: 24–95% ACN in 0.1% TFA over 30 min map. Column: VYDAC® 218TP52 (Narrowbore). Eluent: 2–80% ACN with 0.06% TFA over 120
30 min at 1.5 mL/min Sample: ribonuclease, at 1.5 mL/min Sample: ribonuclease, insulin, min at 0.25 mL/min. Sample: tryptic digest of carboxymethylated transferrin Data courtesy of K.
insulin, lysozyme, myoglobin and ovalbumin. lysozyme, myoglobin and ovalbumin. Stone and K. Williams. Ref. 12.
(株)ケムコ
56
TEL 06-6375-0395 URL : http://www.chemco.jp/ 57
R eversed-phase chromatography
separates polypeptides on the
basis of hydrophobicity; ion-exchange
reversed-phase and cation exchange
HPLC illustrates the complementary
selectivity of the two techniques
The Benefits of Ion-Exchange
Chromatography
The Benefits of Reversed-Phase
Chromatography
■ Relatively high sample loading ■ A high degree of selectivity based
chromatography separates on the (Figure D-1). On the cation exchange
basis of charge. These complementary column singly-charged oxytocin capacity compared to reversed- on differences in hydrophobicity
separation techniques offer synergistic elutes early, followed by the three phase. or molecular conformation.
capabilities in the analysis and doubly-charged peptides—neurotensin, ■ Resistance to strong reagents such ■ Use of volatile buffers or
purification of proteins and peptides angiotensin II and bradykinin. as 0.1 M NaOH, 0.1 M acid or ion-pairing agents.
and are often used together because Angiotensin I with four charges 6 M guanidine because of the ■ Freedom from interferences by
of the different separation elutes last. On reversed-phase the polymeric matrix. Relatively crude salt or buffers from ion exchange.
mechanisms. In series they offer peptides elute in the order of oxytocin, solutions can be loaded onto
better purification than can be bradykinin, angiotensin I, neurotensin ion-exchange columns because Ion-exchange chromatography is
achieved with either one alone; in and angiotensin II. The complementary adsorbed matrix components can normally used first, followed by
parallel they offer mutual confirmation selectivities provide two dimensional be removed with strong reagents. reversed-phase chromatography
of analytical results. Comparison of resolving power. ■ Addition of urea, acetonitrile or (Figure D-2). Crude samples can be
the separation of several peptides by non-ionic detergents to break-up loaded onto a polymer-based ion
complexes. exchange column without damaging
■ Optimization of elution selectivity the column; ion-exchange has a high
by adjustment of pH. loading capacity to accomodate
complex samples; and chaotropes
High Performance Reversed-Phase can be added to the sample to break
Comparison of High Performance Reversed-Phase and High Performance
Ion-Exchange Chromatography in the Separation of Peptides and Ion-Exchange Chromatography up protein complexes. The partially
Used in Series to Remove Impurities purified polypeptide, containing salts
in Lysozyme and buffers from the ion exchange
Reversed-Phase (218TP54) Cation Exchange (400VHP575) separation, can then be loaded onto a
5 reversed-phase column. The salts are
5 Cation
Exchange not retained and do not harm the
3 reversed-phase column. Purification
3 lysosome peak
from IEX onto RP based on hydrophobicity or
2
1 2 conformation then takes place
4 impurity on RP
1 4 and the collected sample elutes
in a volatile solution, ready for
impurity on RP Reversed-
Phase
final preparation.
0 25 min 0 40 min
Figure D-2. Strong Cation Exchange Column:
VYDAC® 400VHP575, Cation exchange, 5 µm,
7.5 x 50 mm Eluent: 10 mM phosphate, pH
Figure D-1. Reversed-Phase Column: VYDAC® 218TP54, C18, 5 µm, 4.6 x 250 mm Eluent: 6.5/25% ACN; gradient from 0–0.1 M NaCl in
15–30% ACN in 0.1% TFA over 30 minutes at 1.0 mL/min Strong Cation Exchange Column: 25 min at 1.0 mL/min Reversed-Phase Column:
VYDAC® 400VHP575, Cation exchange, 5 µm, 7.5 x 50 mm Eluent: 10 mM phosphate, VYDAC® 214TP54, C4, 5 µm, 4.6 x 250 mm
pH 2.7/25% ACN; gradient from 0-0.1 M NaCl in 20 min at 1.0 mL/min Sample: 1. oxytocin Eluent: 10–35% ACN in 0.1% TFA, 50 minutes
2. bradykinin 3. angiotensin II 4. neurotensin 5. angiotensin I. at 1.0 mL/min Sample: Lysozyme.
(株)ケムコ
58
TEL 06-6375-0395 URL : http://www.chemco.jp/ 59
(株)ケムコ
62
TEL 06-6375-0395 URL : http://www.chemco.jp/ 63
(株)ケムコ
64
TEL 06-6375-0395 URL : http://www.chemco.jp/ 65
(株)ケムコ
66
TEL 06-6375-0395 URL : http://www.chemco.jp/ 67
(株)ケムコ
68
TEL 06-6375-0395 URL : http://www.chemco.jp/ 69
On-line
experts@vydac.com
www.gracevydac.com
The information contained herein is based on our testing and experience and is offered for the user’s consideration, investigation
and verification. Since operating and use conditions vary and since we do not control such condiitions, we must DISCLAIM
ANY WARRANTY, EXPRESSED OR IMPLIED, with regard to results to be obtained from the use of this product. Test methods are
available on request. © 2002 W.R. Grace & Co.-Conn. All rights reserved. VYDAC is a registered trademark of The Separations
Group, a wholly-owned subsidiary of W. R. Grace & Co.–Conn.
(株)ケムコ
70
TEL 06-6375-0395 URL : http://www.chemco.jp/
FAX 06-6372-0533 E-mail : info@chemco.jp