Salicylic acid, the main metabolite of aspirin is integral part of human and animal metabolism. While much
of it is attributable to diet a substantial part is synthesized endogenously.[1]
Aspirin also has an antiplatelet effect by inhibiting the production of thromboxane, which under normal
circumstances binds platelet molecules together to create a patch over damaged walls of blood vessels.
Because the platelet patch can become too large and also block blood flow, locally and downstream,
aspirin is also used long-term, at low doses, to help prevent heart attacks, strokes, and blood
clot formation in people at high risk for developing blood clots.[2] It has also been established that low
doses of aspirin may be given immediately after a heart attack to reduce the risk of another heart attack
or of the death of cardiac tissue.[3][4]
Aspirin was the first discovered member of the class of drugs known as nonsteroidal anti-inflammatory
drugs (NSAIDs), not all of which are salicylates, although they all have similar effects and most have
inhibition of the enzyme cyclooxygenase as their mechanism of action. Today, aspirin is one of the most
widely used medications in the world, with an estimated 40,000 tonnes of it being consumed each year.
[6]
In countries where Aspirin is a registeredtrademark owned by Bayer, the generic term is acetylsalicylic
acid (ASA).[7][8]
You can visualize this process by running the simulation shown below. Click
on the Start button to start the simulation and the Stop button to stop the
simulation.
The light source is set to emit 10 photons per second. Watch the motion of the
photons and observe how some of the photons are absorbed (removed) as the
beam of light passes through the cell containing the sample solution. The
intensity of the light reaching the detector is less than the intensity emitted by
the light source.
Second, the intensity of light (I) passing through the sample solution
is measured. (In practice, instruments measure the power rather than
the intensity of the light. The power is the energy per second, which is
the product of the intensity (photons per second) and the energy per
photon.)
I
T =
I0
A = - log10 T
The transmittance is simply the fraction of light in the original beam that
passes through the sample and reaches the detector. The remainder of the light,
1 - T, is the fraction of the light absorbed by the sample. (Do not confuse the
transmittance with the temperature, which often is given the symbol T.)
In most applications, one wishes to relate the amount of light absorbed to the
concentration of the absorbing molecule. It turns out that the absorbance rather
than the transmittance is most useful for this purpose. If no light is absorbed,
the absorbance is zero (100% transmittance). Each unit in absorbance
corresponds with an order of magnitude in the fraction of light transmitted.
For A = 1, 10% of the light is transmitted (T = 0.10) and 90% is absorbed by
the sample. For A = 2, 1% of the light is transmitted and 99% is absorbed.
For A = 3, 0.1% of the light is transmitted and 99.9% is absorbed.
Note: For each measurement, run the simulation long enough to detect at least
1000 photons. There is substantial random error in the intensity, and the more
photons that are counted, the lower the relative uncertainty in the results.
Iron(III) chloride, also called ferric chloride, is an industrial scale commodity chemical compound, with
the formula FeCl3. The colour of iron(III) chloride crystals depends on the viewing angle: by reflected light
the crystals appear dark green, but by transmitted light they appear purple-red. Anhydrous iron(III)
chloride is deliquescent, forming hydrated hydrogen chloride mists in moist air. It is rarely observed in its
natural form, mineralmolysite, known mainly from some fumaroles.
This page takes a brief look at the Beer-Lambert Law and explains
the use of the terms absorbance and molar absorptivity relating to
UV-visible absorption spectrometry.
Absorbance
The intensity of the light passing through the sample cell is also
measured for that wavelength - given the symbol, I.
If I is less than Io, then obviously the sample has absorbed some of
the light. A simple bit of maths is then done in the computer to
convert this into something called the absorbance of the sample -
given the symbol, A.
Suppose then that you wanted to compare this dye with a different
compound. Unless you took care to make allowance for the
concentration, you couldn't make any sensible comparisons about
which one absorbed the most light.
Suppose this time that you had a very dilute solution of the dye in a
cube-shaped container so that the light travelled 1 cm through it.
The absorbance isn't likely to be very high. On the other hand,
suppose you passed the light through a tube 100 cm long
containing the same solution. More light would be absorbed
because it interacts with more molecules.
Both concentration and solution length are allowed for in the Beer-
Lambert Law.
You will find that various different symbols are given for some of
the terms in the equation - particularly for the concentration and the
solution length. I'm going to use the obvious form where the
concentration of the solution is "c" and the length is "l".
Note: That's obviously "l" for length. The font I'm using
won't distinguish between "l" for length and a capital letter
"I" (for Intensity). That problem disappears in the equation
below - where it is obvious which is which.
That's obviously easier to remember than the first one, but you
would still have to learn the equation for absorbance. It might be
useful to learn it in the form:
Molar absorptivity
That means that you can then make comparisons between one
compound and another without having to worry about the
concentration or solution length.
Values for molar absorptivity can vary hugely. For example, ethanal
has two absorption peaks in its UV-visible spectrum - both in the
ultra-violet. One of these corresponds to an electron being
promoted from a lone pair on the oxygen into a pi anti-bonding
orbital; the other from a pi bonding orbital into a pi anti-bonding
orbital.
Note: These jumps are described in detail on the page
explaining the theory of UV-visible spectrometry.
Depending on your background knowledge, you may have
to read another page first, but there is a link to that on the
theory page.
wavelength of
molar
electron jump maximum absorption
absorptivity
(nm)
lone pair to pi anti-
290 15
bonding orbital
pi bonding to pi
180 10000
anti-bonding orbital
To get around this, you may also come across diagrams in which
the vertical axis is plotted as log10(molar absorptivity).
If you take the logs of the two numbers in the table, 15 becomes
1.18, while 10000 becomes 4. That makes it possible to plot both
values easily, but produces strangely squashed-looking spectra!
Beer-Lambert Law, more commonly known as Beer's Law, states that the optical absorbance of a
chromophore in a transparent solvent varies linearly with both the sample cell pathlength and the
chromophore concentration. Beer's Law is the simple solution to the more general description of
Maxwell's far-field equations describing the interaction of light with matter. In practice, Beer's Law is
accurate enough for a range of chromophores, solvents and concentrations, and is a widely used
relationship in quantitative spectroscopy.
Aλ = ελbc
In an absorbance experiment, light is attenuated not only by the chromophore, but also by reflections from
the interface between air and the sample, the sample and the cuvette, and absorbance by the solvent.
These factors can be quantified separately, but are often removed by defining I0 as the light passing
through a sample "blank" or "baseline" or reference sample (for example, a cuvette filled with solvent but
zero concentration of the chromophore is used as the blank).
Many factors can affect the validity of Beer's Law. It is usual to check for the linearity of Beer's Law for a
chromophore by measuring the absorbance of a series of standards. This "calibration" can also remove
errors in the experiment, the equipment, and the batch of reagents (such as cuvettes of unknown
pathlength).