‘Cerificate Number ——_
0184-EPE-00-01-
Page tof 1 a
(Order Number 4501.24.83
Received Date December 27, 2023
Customer PT. Hutama - Adhi - Abipra
Address Jalan Poros KKT Kariangau =
Equipment Name: Timbangan Environment Condition of Calibration
Manufacturer SAYAKI ‘Temperature (304 + 08) | °C
Type 88 -B 6000 Humidity ( 59 + 28) | %RH
Serial Number iwiwet2 - 221100796 Reference EPE.03.02.01.02.01.01
Capacity 6000 g Calibration Date December 27, 2023,
Resolution o1 og Callration Location: Laboratorium
Calibration Report
41. Repeatability of Measurement
‘Standard Indication] __ Repeatabiity Maximum Diference
————
6000.0 0.0) 0.0
2. Correction to Balance Reading
‘Standard Indeation] Unit Under Test ‘Correction ‘Uncertainty, =
CF gs fe Gere eas aes ee tn):
600.0 ‘600.1 “OL 10.
1200.0 1200.4 04 10
1800.0 1800.5 =05, 10
2400.0 2400.5 05 10
"3000.0 3000.6 “06 10.
3600.0 3600.8 “08, 10
4200.0 4200.9, “09. 10 |
4800.0 801.2 “12 10
5400.0 3401.3 is 10.
5000.0 6001.7 “17 io
3. Off Center Loading
cae
Clg Se
00
4, Limit of Performanco 5. Hysteresis
Lm of Performance 27 9 Hysteresis, 00g
‘The Uncertainty is taken at a Confidence Level 95 % For Coverage Factor ( k
Galibrator used
Calirator Name Manufacturer Type Serial Number Traceable to SI Through
Mase Stancerd CROWN FT 4190222 Licoas:1ON
2° Mass Stanara CROWN, Ft 4190223 U048:10N
Calibrated By wit Date of Issued
Typist APS PRE