TDS3000 Series Digital Phosphor Oscilloscopes 071-0274-01

This document supports firmware version 2.00 and above.

Copyright Tektronix, Inc. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material. Specifications and price change privileges reserved. Printed in the U.S.A. Tektronix, Inc., P.O. Box 1000, Wilsonville, OR 970701000 TEKTRONIX, TEK, TEKPROBE, and Tek Secure are registered trademarks of Tektronix, Inc. DPX is a trademark of Tektronix, Inc.

OSCILLOSCOPE WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of three (3) years from the date of purchase from an authorized Tektronix distributor. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty. In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product. THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE LISTED PRODUCTS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

PROBE WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and workmanship for a period of one (1) year from the date of purchase from an authorized Tektronix distributor. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Batteries are excluded from this warranty. In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the service center designated by Tektronix, shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product. THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE LISTED PRODUCTS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.

Table of Contents
General Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Contacting Tektronix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv viii ix

Getting Started
Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Functional Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal Path Compensation (SPC) . . . . . . . . . . . . . . . . . . . . Product and Feature Description . . . . . . . . . . . . . . . . . . . . . . . . Acquisition Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Signal Processing Features . . . . . . . . . . . . . . . . . . . . . . . . . Display Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measurement Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Trigger Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Convenience Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Optional Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Transporting the Oscilloscope . . . . . . . . . . . . . . . . . . . . . . . . . . Operating Positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connecting Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using Battery Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating Safely with Battery Power . . . . . . . . . . . . . . . . . Installing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Maximizing Operating Time . . . . . . . . . . . . . . . . . . . . . . . . Charging the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installing an Application Module . . . . . . . . . . . . . . . . . . . . . . . Installing a Communication Module . . . . . . . . . . . . . . . . . . . . . Front-Panel Menus and Controls . . . . . . . . . . . . . . . . . . . . . . . . Using the Menu System . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the Menu Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . Using the Dedicated Controls . . . . . . . . . . . . . . . . . . . . . . . Identifying Items in the Display . . . . . . . . . . . . . . . . . . . . . Using QuickMenus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Front-Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Rear-Panel Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 12 13 14 14 15 15 16 16 16 17 17 18 19 19 110 111 112 113 114 115 117 118 118 121 123 126 128 130 131

TDS3000 Series User Manual

Table of Contents

Application Examples
Taking Simple Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . Using Autoset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Selecting Automatic Measurements . . . . . . . . . . . . . . . . . . Measuring Two Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . Customizing Your Measurements . . . . . . . . . . . . . . . . . . . . Analyzing Signal Detail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Looking at a Noisy Signal . . . . . . . . . . . . . . . . . . . . . . . . . . Separating the Signal from Noise . . . . . . . . . . . . . . . . . . . . Taking Cursor Measurements . . . . . . . . . . . . . . . . . . . . . . . Using Delay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring Jitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Triggering on a Video Signal . . . . . . . . . . . . . . . . . . . . . . . . . . . Capturing a Single-Shot Signal . . . . . . . . . . . . . . . . . . . . . . . . . Optimizing the Acquisition . . . . . . . . . . . . . . . . . . . . . . . . . Using the Horizontal Zoom Function . . . . . . . . . . . . . . . . . Using the Disk Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Saving Screen Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Saving Waveform Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 22 23 24 26 29 210 211 212 213 215 216 220 221 222 223 224 227

Reference
Introduction to Reference . . . . . . . . . . . . . . . . . . . . . . . . . . . . Acquisition Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cursor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Hard Copy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Horizontal Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . QuickMenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Save/Recall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Trigger Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Utility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Vertical Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 32 312 317 321 325 333 339 340 349 360 370

ii

TDS3000 Series User Manual

Table of Contents

Appendices
Appendix A: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix B: Factory Setup . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix C: Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix D: Probe Basics . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TekProbe Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe Guard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ground Leads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . P3010 High-Frequency Compensation . . . . . . . . . . . . . . . . . . . P3010 Replaceable Parts and Accessories . . . . . . . . . . . . . . . . . P6139A Replaceable Parts and Accessories . . . . . . . . . . . . . . . Using Other Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Supported Active Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unsupported Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Appendix E: Performance Verification . . . . . . . . . . . . . . . . . Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Performance Verification Procedures . . . . . . . . . . . . . . . . . . . . Appendix F: General Care and Cleaning . . . . . . . . . . . . . . . General Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A1 B1 C1 D1 D1 D2 D2 D3 D3 D4 D6 D8 D10 D11 D12 E1 E2 E5 F1 F1 F1

Glossary Index

TDS3000 Series User Manual

iii

General Safety Summary


Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it. To avoid potential hazards, use this product only as specified.

To Avoid Fire or Personal Injury


Use Proper Power Cord. Use only the power cord specified for this product and certified for the country of use. Connect and Disconnect Properly. Connect the probe output to the

measurement instrument before connecting the probe to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting the probe from the measurement instrument.
Ground the Product. When operating with AC power, this product is

grounded through the grounding conductor of the power cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before making connections to the input or output terminals of the product, ensure that the product is properly grounded. When operating with battery power, this product must still be grounded. To prevent electric shock, always connect a grounding wire between the ground terminal on the rear panel and earth ground.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all

ratings and markings on the product. Consult the product manual for further ratings information before making connections to the product. Connect the ground lead of the probe to earth ground only.

iv

TDS3000 Series User Manual

General Safety Summary

Replace Batteries Properly. Replace batteries only with the proper type

and rating specified.


Recharge Batteries Properly. Recharge batteries for the recommended

charge cycle only.


Do Not Operate Without Covers. Do not operate this product with

covers or panels removed.


Use Proper Fuse. Use only the fuse type and rating specified for this

product.
Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. Do Not Operate With Suspected Failures. If you suspect there is damage

to this product, have it inspected by qualified service personnel.


Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. Provide Proper Ventilation. Refer to the manuals installation

instructions for details on installing the product so it has proper ventilation.

TDS3000 Series User Manual

General Safety Summary

Safety Terms and Symbols


Terms in This Manual. These terms may appear in this manual:

WARNING. Warning statements identify conditions or practices that could result in injury or loss of life.

CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.

Terms on the Product. These terms may appear on the product:

DANGER indicates an injury hazard immediately accessible as you read the marking. WARNING indicates an injury hazard not immediately accessible as you read the marking. CAUTION indicates a hazard to property including the product.
Symbols on the Product. These symbols may appear on the product:

WARNING High Voltage

Protective Ground (Earth) Terminal

CAUTION Refer to Manual

Battery Information

vi

TDS3000 Series User Manual

General Safety Summary

Preventing Electrostatic Damage


CAUTION. Electrostatic discharge (ESD) can damage components in the oscilloscope and its accessories. To prevent ESD, observe these precautions when directed to do so.

Use a Ground Strap. Wear a grounded antistatic wrist strap to

discharge the static voltage from your body while installing or removing sensitive components.
Use a Safe Work Area. Do not use any devices capable of generating or

holding a static charge in the work area where you install or remove sensitive components. Avoid handling sensitive components in areas that have a floor or benchtop surface capable of generating a static charge.
Handle Components Carefully. Do not slide sensitive components over

any surface. Do not touch exposed connector pins. Handle sensitive components as little as possible.
Transport and Store Carefully. Transport and store sensitive components in a static-protected bag or container.

Battery Recycling
This product may contain an optional Nickel Cadmium (NiCad) battery, which must be recycled or disposed of properly. For the location of a local battery recycler in the U.S. or Canada, please contact: RBRC Rechargeable Battery Recycling Corp. P.O. Box 141870 Gainesville, Florida 32614 (800) BATTERY (800) 227-7379 www.rbrc.com

TDS3000 Series User Manual

vii

Preface
This User Manual describes the capabilities, operation, and applications of the TDS3000 Series Digital Phosphor Oscilloscopes. The following table shows you where to find information in this manual.
If you are looking for: Product overview Installation information Basic operating instructions Details about a product feature Turn to: Product and Feature Description on page 14 Connect Power on page 110 Front Panel Controls on page 118 Reference on page 31 Look up the front-panel button for the feature Application examples Information about selecting a language Information about battery operation Information about making a hard copy Information about probes and probepower limitations Technical specifications Recommended accessories Application Examples on page 21 Configure System on page 361 Using Battery Power on page 110 Hard Copy on page 321 Probe Basics on page D1 Specifications on page A1 Accessories on page C1

viii

TDS3000 Series User Manual

Preface

Contacting Tektronix
Product Support For questions about using Tektronix measurement products, call toll free in North America: 1-800-TEK-WIDE (1-800-835-9433 ext. 2400) 6:00 a.m. 5:00 p.m. Pacific time Or contact us by e-mail: tm_app_supp@tek.com For product support outside of North America, contact your local Tektronix distributor or sales office. Service Support Tektronix offers extended warranty and calibration programs as options on many products. Contact your local Tektronix distributor or sales office. For a listing of worldwide service centers, visit our web site. For other In North America: information 1-800-TEK-WIDE (1-800-835-9433) An operator will direct your call. To write us Tektronix, Inc. P.O. Box 1000 Wilsonville, OR 97070-1000 USA Tektronix.com

Website

TDS3000 Series User Manual

ix

Preface

TDS3000 Series User Manual

Getting Started

Getting Started
In addition to a product and feature description, this chapter covers the following topics: H How to perform a quick functional check, install and compensate passive probes, and compensate the signal path H How to install the power cord, battery pack, and operate the oscilloscope safely with battery power H How to install application modules and communication modules H How to use the menu system H How to identify the oscilloscope controls and connectors

Initial Setup
The following procedures describe how to quickly verify that the oscilloscope is powering up and functioning correctly, compensate passive probes using the built-in compensation signal, and run the signal path compensation (SPC) routine for maximum signal accuracy. H You should perform all initial setup procedures the first time you use the oscilloscope. H You should perform the probe compensation procedure whenever you attach a passive probe for the first time to any input channel. H You should run the signal path compensation routine whenever the ambient temperature changes by 10_ C or more.

TDS3000 Series User Manual

11

Getting Started

Functional Check Perform this quick functional check to verify that your oscilloscope is operating correctly. 1. Connect the oscilloscope power cable (refer to page 19). 2. Turn on the oscilloscope. Wait for the confirmation that all self tests have passed.
PASSED

On/Standby button

3. Connect the oscilloscope probe to channel 1. Attach the probe tip and reference lead to the PROBE COMP connectors.

PROBE COMP

CH 1

4. Push the AUTOSET button. You should see a square wave in the display (approximately 5 V at 1 kHz).

12

TDS3000 Series User Manual

Getting Started

Probe Compensation Perform this adjustment to match your probe to the input channel. This should be done whenever you attach a passive probe for the first time to any input channel. 1. Connect the oscilloscope probe to channel 1. Attach the probe tip and reference lead to the PROBE COMP connectors, then push AUTOSET. If using the probe hook-tip, ensure a proper connection by firmly twisting the tip onto the probe. 2. Check the shape of the displayed waveform.
AUTOSET button

PROBE COMP

CH 1

Overcompensated

Undercompensated

Compensated correctly

3. If necessary, adjust your probe. Repeat as necessary.


L

NOTE. See Appendix D: Probe Basics for more information about the probes provided with your oscilloscope.

TDS3000 Series User Manual

13

Getting Started

Signal Path Compensation (SPC) The signal path compensation (SPC) routine optimizes the oscilloscope signal path for maximum measurement accuracy. You can run the routine anytime but you should always run the routine if the ambient temperature changes by 10_ C or more. To compensate the signal path, do the following steps: 1. Disconnect any probes or cables from the channel input connectors. 2. Push the UTILITY button. 3. Push the System screen button to select Cal. 4. Push the Signal Path screen button. 5. Push OK Compensate Signal Path. This procedure takes several minutes to complete. NOTE. The signal path compensation does not include calibration to the probe tip.

Product and Feature Description


The TDS3000 Series Digital Phosphor Oscilloscope family consists of the models in the table below.
Maximum sample rate 1.25 GS/s 2.5 GS/s 5 GS/s

Model TDS3012 (2 Ch), TDS3014 (4 Ch) TDS3032 (2 Ch), TDS3034 (4 Ch) TDS3052 (2 Ch), TDS3054 (4 Ch)

Bandwidth 100 MHz 300 MHz 500 MHz

14

TDS3000 Series User Manual

Getting Started

Acquisition Features
Separate Digitizers. Ensure accurate timing measurements with

separate digitizers for each channel. Each digitizer can sample at up to the maximum sample rate; acquisition on all channels is always concurrent to provide full single-shot bandwidth on each channel. See page 14.
Normal Acquisition. Acquire 10,000 point waveforms to capture

horizontal detail and then use the zoom detail. See page 311.

function to analyze the

Fast Trigger Acquisition. Acquire up to 3,000 waveforms per second to see rapidly changing signals or intermittent signal irregularities. See page 311. Pretrigger. You can capture signals that occur before the trigger point.

You can position the trigger point at the beginning of the acquisition, at the end, or at any location in between. See page 325.
Delay. You can also delay the acquisition so that it starts after the trigger point. Use delay when you want to acquire the signal at a specific time after the trigger point. See page 327. Peak Detect. See pulses as narrow as 1 ns even at the slower time base

settings. Peak Detect helps you see noise and glitches in your signal. See page 38. Signal Processing Features
Average. Apply averaging to your signal to remove uncorrelated noise

and improve measurement accuracy. See page 39.


Envelope. Use envelope to capture and display the maximum

variation of a signal. See page 39.


Waveform Math. Use waveform math to add, subtract, multiply, or divide waveforms. For example, you can use math to analyze differential signals or calculate a power waveform. See page 374.

TDS3000 Series User Manual

15

Getting Started

Display Features
Color LCD Display. Identify and differentiate waveforms easily with color coding. Waveforms, readouts, and buttons are color matched to increase productivity and reduce operating errors. See page 319. Digital Phosphor. A Digital Phosphor Oscilloscope can clearly display intensity modulation in your signals. The oscilloscope automatically overlays subsequent acquisitions and then decays them to simulate the writing and decay of the phosphor in an analog oscilloscope CRT. This feature results in an intensity-graded waveform display that shows the information contained in the intensity modulation. See page 35. Signal Preview. Use the preview feature to optimize the control

settings when setting up a single-shot acquisition. As you adjust the controls, the adjustments modify the current acquisition to show a preview of how the next acquisition should appear. See page 37. Measurement Features
Cursors. Use cursors to take simple voltage, time, and frequency measurements. See page 312. Automatic Measurements. Choose from a list of 21 automatic

waveform measurements. You can customize the measurements by changing reference levels or by adding measurement gating. See page 333. Trigger Features
Dual Triggers. Use the main (A) trigger system alone or add the B

trigger to capture more complex events. You can use the A and B triggers together to set up a wait-for-time or wait-for-events trigger. See page 350.
Video Trigger. Trigger on video fields or lines to see a stable display of

standard video signals. See page 359.

16

TDS3000 Series User Manual

Getting Started

Convenience Features
Autoset. Use Autoset to quickly set up the vertical, horizontal, and

trigger controls for a usable display. See page 34.


Scope QuickMenu. Use the built-in Scope QuickMenu for simplified

oscilloscope operation. See page 128.


Single Sequence. One button sets the trigger parameters to the correct

settings for a single-shot acquisition (or single-sequence acquisition). See page 33.
Floppy Disk. Use the built-in floppy disk to store and recall wave-

forms and setups, as well as upgrade the oscilloscope firmware and install new features. See page 343.
Probe Support. Use the standard probes or choose an optional probe

for a specific application. See page D1 for information and limitations.


Multilingual User Interface. On-screen menus and messages are in 11 languages. See page 361.

Optional Features
Application Modules. Install Advanced Trigger, Extended Video,

Telecom Mask Test, or FFT application modules to add new test and measurement features. See page C2.
Communication Modules. Install a communication module to add

RS-232, GPIB, VGA, or Ethernet local area network (LAN) ports for remote programmability, sending hardcopy to a LAN printer, or displaying the oscilloscope screen on a monitor. See page 117.
Battery Power. Install a rechargeable battery pack so you can use the

oscilloscope without line power. See page 110.

TDS3000 Series User Manual

17

Getting Started

Transporting the Oscilloscope


When transporting the oscilloscope, wrap the cord around the handle as shown below. Use the supplied cord retainer if the power plug does not have a molded-in retainer. The oscilloscope front cover has a convenient place to store the Reference Manual.

If you are not using a battery, use the accessory tray in the battery compartment to store probes and other accessories. CAUTION. To avoid damage to the disk drive, do not transport the oscilloscope with a floppy disk in the disk drive.

18

TDS3000 Series User Manual

Getting Started

Operating Positions
Use the handle and feet to place the oscilloscope in a convenient operating position.

Connecting Power
To connect a power cord, do these steps: 1. Open the strain relief and slip it over the power cord. 2. Snap the strain relief into its hole in the rear panel of the oscilloscope. 3. Connect the power cord to the power input connector. You can operate the oscilloscope from a grounded mains supply with line voltage between 90 VAC and 250 VAC and frequency between 47 Hz and 440 Hz. The oscilloscope is grounded through the power cord. The line fuse is internal and is not user replaceable.

TDS3000 Series User Manual

19

Getting Started

Using Battery Power You can operate the oscilloscope continuously for approximately two hours from the rechargeable battery pack. A triangle icon in the display ( ) shows when the battery is in use, a power-plug icon ) shows when line power is connected, and a gauge icon ( ( )shows the charge level in the battery. The oscilloscope turns off automatically when the battery runs low; the screen may turn white a few minutes before the automatic shutdown. NiCad batteries appear to lose capacity if not allowed to occasionally discharge completely. Every few months, run your oscilloscope until it shuts down and then recharge the battery completely to keep your battery pack at peak performance. Refer to the General Safety Summary for information about proper nickel-cadmium battery disposal.

110

TDS3000 Series User Manual

Getting Started

Operating Safely with Battery Power WARNING. To avoid electric shock, connect the rear-panel ground terminal to earth ground when operating the instrument from battery power. For safe operation, the oscilloscope chassis should always remain at earth ground potential. Without a connection between the chassis and earth ground, you can receive a shock from exposed metal on the chassis if you connect an input to a hazardous voltage (>30 VRMS, >42 Vpk). Protection against this condition is provided when you attach the Tektronix-supplied grounding wire from the terminal on the rear panel to earth ground. If you use a different grounding wire, it must be at least 18 gauge.

If you choose not to attach the grounding wire, you are not protected against electric shock if you connect the oscilloscope to a hazardous voltage. You can still use the oscilloscope if you do not connect a signal greater than 30 VRMS (42 Vpk) to either the probe tip or common lead. Ensure that all probe common leads are connected to the same voltage. Also, do not connect a grounded device, such as a printer or computer, to the oscilloscope. WARNING. Hazardous voltages may exist in unexpected places due to faulty circuitry in the device under test.

TDS3000 Series User Manual

111

Getting Started

Installing the Battery To install the optional battery pack, do these steps: 1. Open the battery compartment door on the rear panel. 2. Remove the accessory tray.

Battery door (Opened)

3. Slide the battery into the compartment and press it in from both sides until you hear the latches click. 4. Press on both sides of the battery compartment door to snap it closed.

112

TDS3000 Series User Manual

Getting Started

To remove the battery, do these steps: 1. Open the battery compartment door. 2. Raise the handles on each side of the battery and use them to pull the battery out of the oscilloscope.

Maximizing Operating Time To maximize the time that the oscilloscope can operate from a full battery charge, consider doing these things: H Reduce the display backlight intensity (see page 317) H Disconnect unused active probes H Only use passive probes

TDS3000 Series User Manual

113

Getting Started

Charging the Battery The battery charges automatically when the oscilloscope is connected to line power. You can also charge the battery with the optional external charger (TDS3CHG).
Configuration Battery charging in oscilloscope with oscilloscope turned on or off Battery charging with TDS3CHG external charger Typical charging time 18 hours 3 hours

NOTE. Charge the battery before using it for the first time or after prolonged storage.

114

TDS3000 Series User Manual

Getting Started

Installing an Application Module


CAUTION. To avoid damage to the oscilloscope or application module, observe the ESD precautions described on page vii. Optional application packages are available to extend the capability of your oscilloscope. You can install up to four application modules at one time. Application modules can go into the two slots with windows in the upper right corner of the front panel. Two additional slots are hidden behind the two you can see. To install an application module, do these steps: 1. Save any oscilloscope settings and/or reference waveforms to floppy disk before doing these steps. 2. Turn the oscilloscope power off. 3. Open the small door in the upper right corner of the front panel. 4. Use a small screwdriver to remove an existing module if it is necessary to make space for a new application module. 5. Slide the application module into any available slot with the module contacts facing the circuit board. Close the module door. 6. If the application module came with one or more floppy disks, insert the firmware upgrade floppy disk into the disk drive. If there is more than one floppy disk, insert firmware upgrade floppy disk number one into the disk drive. 7. Power on the oscilloscope. The oscilloscope determines whether the firmware update is necessary. If no firmware update is necessary, go to step 9.

TDS3000 Series User Manual

115

Getting Started

8. Push OK Load New Firmware to begin loading the firmware. During the process, the oscilloscope may instruct you to insert a second floppy disk. When the firmware upgrade is complete, the oscilloscope will restart automatically with the new firmware. NOTE. If you power off the oscilloscope, eject a floppy disk, or there is a power outage during the upgrade process, you must power off the oscilloscope and do the firmware upgrade procedure, starting at step 6, before you can use the oscilloscope. 9. Remove the floppy disk.

Application modules

Contacts

NOTE. If you remove an application module, the features provided by the application module become unavailable. You can reinstall the module to restore the features.

116

TDS3000 Series User Manual

Getting Started

Installing a Communication Module


CAUTION. To avoid damage to the oscilloscope or communication module, observe the ESD precautions described on page vii. To install one of the optional communications modules, do these steps: 1. Turn the oscilloscope power off. 2. Press down on the latching tab to remove the blank cover. 3. Slide the communication module into the compartment until the internal connectors are seated and the latching tab locks. 4. Turn power on. The communication module is now ready for your use. To remove a communication module, do these steps: 1. Turn the oscilloscope power off. 2. Press down on the latching tab and then use a small screwdriver to alternately pry out the sides of the communication module. 3. Slide out the communication module and store it in an ESDshielded bag. Install the blank cover if no other communication module is to be installed.
Communication module port GPIB RS-232 Ethernet (LAN) 10BaseT VGA Page A8 in this manual For more information, see Hard Copy on page 321 of this manuall and the TDS3000 Series m t e Programmer Manual

TDS3000 Series User Manual

117

Getting Started

Latching tab

Front-Panel Menus and Controls


The front panel has buttons and controls for the functions you use most often. The front panel has menus to access more specialized functions. Using the Menu System To use the menu system, follow the steps shown on the next two pages.

118

TDS3000 Series User Manual

Getting Started

1. Push a dark-colored front-panel menu button to display the menu you want to use.

MEASURE SAVE/RECALL QUICKMENU

CURSOR

DISPLAY

UTILITY

VERTICAL

HORIZONTAL

TRIGGER

ACQUIRE

2. Push a bottom screen button to select a menu item. If a pop-up menu appears, continue to push the screen button to select an item from the pop-up menu.

TDS3000 Series User Manual

119

Getting Started

3. Push a side screen button to choose a menu item. If the menu item contains more than one choice, push the side screen button again to make the choice.

4. Certain menu choices require you to set a numerical value to complete the setup. Use the general purpose knob to adjust the parameter value. Push the COARSE button to make larger adjustments.

SELECT

COARSE

VERTICAL
POSITION

120

TDS3000 Series User Manual

Getting Started

Using the Menu Buttons You can use the menu buttons to perform many functions in the oscilloscope.

SELECT

MEASURE SAVE/RECALL QUICKMENU

COARSE

CURSOR

DISPLAY

UTILITY

1 2

5 6

1. MEASURE. Performs automated measurements of waveforms. 2. CURSOR. Activates the cursors. 3. SAVE/RECALL. Saves and recalls setups and waveforms to memory or a floppy disk. 4. DISPLAY. Changes the appearance of waveforms and the display screen. 5. QUICKMENU. Activates QuickMenus such as the built-in Scope QuickMenu. 6. UTILITY. Activates the system utility functions, such as selecting a language.

TDS3000 Series User Manual

121

Getting Started

VERTICAL
POSITION

HORIZONTAL
POSITION

TRIGGER
LEVEL

ACQUIRE

SCALE

SCALE

WAVEFORM INTENSITY

7. Vertical MENU. Adjusts the scale, position, and offset of waveforms. Sets the input parameters. 8. Trigger MENU. Adjusts the trigger functions. 9. Acquire MENU. Sets the acquisition modes and horizontal resolution, and resets the delay time.

122

TDS3000 Series User Manual

Getting Started

Using the Dedicated Controls These dedicated buttons and controls generally control waveforms and cursors without the use of menus. 3
SELECT

MEASURE SAVE/RECALL QUICKMENU

2
COARSE CURSOR DISPLAY UTILITY

1
VERTICAL
POSITION

HORIZONTAL
POSITION

TRIGGER
LEVEL

ACQUIRE

7 8 9 10 19
SCALE SCALE

11
WAVEFORM INTENSITY

12 13

18

17 16 15

14

TDS3000 Series User Manual

123

Getting Started

1. COARSE. Causes the general purpose knob and position knobs to make adjustments more quickly. 2. SELECT. Toggles between the two cursors to select the active cursor. 3. General purpose knob. Moves the cursors. Sets numerical parameter values for some menu items. Push COARSE to make adjustments quickly. 4. Vertical POSITION. Adjusts the vertical position of the selected waveform. Push COARSE to make adjustments more quickly. 5. Horizontal POSITION. Adjusts the trigger point location relative to the acquired waveforms. Push COARSE to make adjustments quickly. 6. Trigger LEVEL. Adjusts the trigger level. 7. RUN/STOP. Stops and restarts acquisition. 8. SINGLE SEQ. Sets acquisition, display, and trigger parameters for a single-shot (single-sequence) acquisition. 9. SET TO 50%. Sets the trigger level to the midpoint of the waveform. 10. AUTOSET. Automatically sets the vertical, horizontal, and trigger controls for a usable display. 11. FORCE TRIG. Forces an immediate trigger event. 12. WAVEFORM INTENSITY. Controls waveform intensity. 13. B TRIG. Activates the B trigger. Changes the trigger menu to set the B-trigger parameters. 14. DELAY. Enables delayed acquisition relative to the trigger event. Use horizontal POSITION to set the amount of delay.

124

TDS3000 Series User Manual

Getting Started

15. Horizontal SCALE. Adjusts the horizontal scale factor. 16. Horizontal zoom. Splits the screen and magnifies the current acquisition horizontally. 17. Waveform OFF. Removes selected waveform from the display. 18. Vertical SCALE. Adjusts selected waveform vertical scale factor. 19. CH1, CH2, (CH3, CH4,) MATH. Displays a waveform and chooses the selected waveform. REF shows the reference waveform menu.

20 21

22

23

20. Hard copy. Initiates a hard copy using the port selected in the Utility menu. 21. Power switch. Turns power to on or standby. Power-up time varies from about 15 seconds to 45 seconds, depending on the oscilloscope internal calibration process. 22. Wrist-strap ground. Connect a wrist strap when working with ESD-sensitive circuits. This connector is not a safety ground. 23. MENU OFF. Clears menu from the display.

TDS3000 Series User Manual

125

Getting Started

Identifying Items in the Display The following items may appear in the display; not all items are visible at any given time. Some readouts move outside the graticule area when menus are turned off. 2 3 4 5 6 7

13

9 10 12 11

1. Waveform baseline icons show the zero-volt level of the waveforms (ignoring the effect of offset). The icon colors correspond to the waveform colors. 2. Acquisition readout shows when acquisition is running, stopped, or when acquisition preview is in effect. 3. Trigger position icon shows the trigger location in the waveforms. 4. Expansion point icon shows the point that the horizontal scale expands and compresses around.

126

TDS3000 Series User Manual

Getting Started

5. Waveform record icon shows the trigger location relative to the waveform record. The line color corresponds to the selected waveform color. 6. Trigger status readout show trigger status. 7. Trigger level icon shows the trigger level on the waveform. The icon color corresponds to the trigger source channel color. 8. Cursor and measurement readouts show results and messages. 9. Trigger readouts show the trigger sources, slopes, and levels, and position. 10. Readout shows the delay setting or the trigger location within the record. 11. Horizontal readout shows the main or zoom time/division. 12. Auxiliary waveform readouts show the vertical and horizontal scale factors of the math or reference waveforms. 13. Channel readouts show the channel scale factor, coupling, input resistance, bandwidth limit, and invert status.

14

15

14. Triangle icon with the battery icon indicates a battery is installed and battery power is in use. The battery icon shows the approximate charge level of the battery. See page 111 for important safety information. 15. Power-plug icon with the battery icon indicates a battery is installed but line power is in use. The battery may be charging. The battery icon shows the approximate charge level.

TDS3000 Series User Manual

127

Getting Started

Using QuickMenus The QuickMenu feature simplifies the use of the oscilloscope. When you push the QUICKMENU button, a set of frequently used menu functions show on the display. Then, push the screen buttons around the display to operate the QuickMenu. See page 339 for general instructions to operate QuickMenus.
Using the Scope QuickMenu. Scope is one type of QuickMenu that you

can use to control the basic oscilloscope functions. You can perform many tasks without using the regular menu system. If you need to use a function that is not contained in the Scope QuickMenu, push the button you would normally push to access that function. For example, if you want to add an automatic measurement, push the MEASURE button to set up the measurement. Then, push the QUICKMENU button to return to the Scope QuickMenu with the measurement also in the display. 1 2

7 6

128

TDS3000 Series User Manual

Getting Started

1. Edge Trigger controls. Push these screen buttons to set trigger parameters for edge trigger. 2. Trigger controls if either B trigger or video trigger is selected. 3. Cursor control. Push this screen button to turn on cursors and select the cursor type. Push the SELECT button to toggle between the two cursors to select the active cursor. Use the general purpose knob to move the active cursor. 4. Acquisition controls. Push these screen buttons to set acquisition parameters. 5. Channel vertical controls. Push these screen buttons to set vertical controls for the selected channel. Use the CH1, CH2, CH3, CH4, MATH, and REF buttons to select the channel you want to control. 6. Vertical controls if either the math waveform or a reference waveform is selected. 7. Menu. Push this screen button to select a specific QuickMenu display if more than one is available. NOTE. Items in the Scope QuickMenu not mentioned above are also contained in the regular display. Those items are described on page 126.

Other QuickMenus. Some optional application packages include a

custom QuickMenu display. Those QuickMenus contain specific features that are important for the application.

TDS3000 Series User Manual

129

Getting Started

Front-Panel Connectors
1
1MW, 13pF 150 VRMS CATl 1MW, 13pF 150 VRMS CATl PROBE COMP [ 5V

CH 1 CH 2

CH 3 CH 4

50W 5VRMS

50W 5VRMS

2 1
1MW, 13pF 150 VRMS CATl 1MW, 17pF 150 VRMS CATl PROBE COMP [ 5V

CH 1 CH 2

EXT TRIG

50W 5VRMS

1. PROBE COMP. Square wave signal source to compensate probes. 2. CH 1, CH 2, (CH 3, CH 4). Channel inputs with TekProbe interface. 3. EXT TRIG. External trigger input with TekProbe interface (two-channel models only).

130

TDS3000 Series User Manual

Getting Started

Rear-Panel Connectors
3 1 2
+14.2 V DC 400 mA

1. Parallel printer port. Connect to a printer to make hard copies. 2. CAL switch. For use by authorized service personnel only. 3. DC power output. Provides ~15 V DC accessory power only when the oscilloscope is connected to the AC power line. 4. Ground terminal. Connect to earth ground when using battery power. See page 111 for important safety information. 5. Power input. Attach to an AC power line with integral safety ground.

TDS3000 Series User Manual

131

Getting Started

6 TDS3GM

TDS3VM

TDS3EM

6. GPIB port. Connect to a controller for remote programmability. 7. RS-232 port. Connect to a controller or terminal for remote programmability or printing. 8. VGA port. Connect to a VGA monitor to display the screen image. 9. 10baseT local area network (LAN) Ethernet port. Connect to a 10baseT network for remote printing or programming.

132

TDS3000 Series User Manual

Application Examples

Application Examples
This section presents five common oscilloscope applications: H Taking simple measurements H Analyzing signal detail H Triggering on a video signal H Capturing a single-shot signal H Using the disk drive Each application example highlights different features of the oscilloscope and gives you ideas about using the oscilloscope to solve test problems.

TDS3000 Series User Manual

21

Application Examples

Taking Simple Measurements


You need to see a signal in a circuit, but you do not know the signal amplitude or frequency. Connect the oscilloscope to quickly display the signal and then measure its frequency and peak-to-peak amplitude.

Ch 1

Using Autoset To quickly display a signal, do these steps: 1. Connect the channel 1 probe to the signal. 2. Push the AUTOSET button. The oscilloscope sets vertical, horizontal, and trigger controls automatically. You can manually adjust any of these controls if you need to optimize the display of the waveform. When you are using more than one channel, the autoset function sets the vertical controls for each channel and uses the lowest-numbered active channel to set the horizontal and trigger controls.

22

TDS3000 Series User Manual

Application Examples

Selecting Automatic Measurements The oscilloscope can take automatic measurements of most displayed signals. To measure signal frequency and peak-to-peak amplitude, do these steps: 1. Push the MEASURE button to see the measurement menu. 2. Push the CH 1 button and then push the Select Measurement for Ch1 screen button. 3. Select the Frequency measurement. 4. Push the more screen button until you can select the Pk-Pk measurement. 5. Push the MENU OFF button. The measurements show on the screen and update as the signal changes.

TDS3000 Series User Manual

23

Application Examples

Measuring Two Signals You are testing a piece of equipment and need to measure the gain of its audio amplifier. You have an audio generator that can inject a test signal at the amplifier input. Connect two oscilloscope channels to the amplifier input and output as shown. Measure both signal levels and use these measurements to calculate the gain.

Ch 1 Ch 2

To display the signals connected to channels 1 and 2, do these steps: 1. Push the CH 1 and CH 2 buttons to activate both channels. 2. Push the AUTOSET button.

24

TDS3000 Series User Manual

Application Examples

To select measurements for the two channels, do these steps: 1. Push the MEASURE button to see the measurement menu. 2. Push the CH 1 button and then push the Select Measurement for Ch1 screen button. 3. Select the Amplitude measurement. 4. Push the CH 2 button and then push the Select Measurement for Ch2 screen button. 5. Select the Amplitude measurement.

6. Calculate the amplifier gain using the following equations: output amplitude + 3.155 V + 24.27 130.0 mV input amplitude log(24.27) + 27.7 dB

Gain +

Gain (dB) + 20

TDS3000 Series User Manual

25

Application Examples

Customizing Your Measurements In this example you want to verify that the incoming signal to a piece of digital equipment meets its specifications. Specifically, the transition time from a low logic level (0.8 V) to a high logic level (2.0 V) must be 10 ns or less.

2.0 V 0.8 V 10 ns

To select the rise time measurement, do these steps: 1. Push the MEASURE button to see the measurement menu. 2. Push the CH 1 button and then the Select Measurement for Ch1 screen button. 3. Select the Rise Time measurement. Rise time is typically measured between the 10% and 90% amplitude levels of a signal; these are the default reference levels the oscilloscope uses for rise time measurements. However, in this example you need to measure the time the signal takes to pass between the 0.8 V and 2.0 V levels. You can customize the rise time measurement to measure the signal transition time between any two reference levels. You can set each of those reference levels to a specific percent of the signal amplitude or to a specific level in vertical units (such as volts or amperes).

26

TDS3000 Series User Manual

Application Examples

Setting Reference Levels. To set the reference levels to specific

voltages, do these steps: 1. Push the Reference Levels screen button. 2. Push the Set Levels in screen button to select units. 3. Push the High Ref screen button. 4. Use the general purpose knob to select 2.0 V. 5. Push the Low Ref screen button. 6. Use the general purpose knob to select 800 mV. The measurement verifies that the transition time (3.842 ns) meets the specification ( 10 ns).

Measuring Specific Events. Next you want to see the pulses in the

incoming digital signal, but the pulse widths vary so it is hard to establish a stable trigger. To look at a snapshot of the digital signal, do this step: 1. Push the SINGLE SEQ button to capture a single acquisition.

TDS3000 Series User Manual

27

Application Examples

Now you want to measure the width of each displayed pulse. You can use measurement gating to select a specific pulse to measure. To measure the second pulse, for example, do these steps: 1. Push the MEASURE button. 2. Push the CH 1 button and then push the Select Measurement for Ch1 screen button. 3. Select the Positive Width measurement. 4. Push the Gating screen button. 5. Select Gate With V Bar Cursors to choose measurement gating using cursors. 6. Place one cursor to the left and one cursor to the right of the second pulse. The oscilloscope shows the width measurement (160 ns) for the second pulse.

28

TDS3000 Series User Manual

Application Examples

Analyzing Signal Detail


You have a noisy signal displayed on the oscilloscope and you need to know more about it. You suspect that the signal contains much more detail than you can now see in the display.

TDS3000 Series User Manual

29

Application Examples

Looking at a Noisy Signal The signal appears noisy and you suspect that noise is causing problems in your circuit. To better analyze the noise, do these steps: 1. Push the acquire MENU button. 2. Select the Peak Detect acquisition mode. 3. Increase the WAVEFORM INTENSITY control to see the noise more easily. Peak detect emphasizes noise spikes and glitches in your signal as narrow as 1 ns, even when the time base is set to a slow setting.

Refer to page 38 for more information about peak-detect and the other acquisition modes.

210

TDS3000 Series User Manual

Application Examples

Separating the Signal from Noise Now you want to analyze the signal shape and ignore the noise. To reduce random noise in the oscilloscope display, do these steps: 1. Push the acquire MENU button. 2. Select the Average acquisition mode. Averaging reduces random noise and makes it easier to see detail in a signal. In the example below, a ring shows on the rising and falling edges of the signal when the noise is removed.

TDS3000 Series User Manual

211

Application Examples

Taking Cursor Measurements You can use the cursors to take quick measurements on a waveform. To measure the ring frequency at the rising edge of the signal, do these steps: 1. Push the CURSOR button. 2. Push the Function screen button. 3. Select V Bars cursors. 4. Push the V Bar Units screen button. 5. Select 1/seconds (Hz). 6. Place one cursor on the first peak of the ring using the general purpose knob. 7. Push the SELECT button. 8. Place the other cursor on the next peak of the ring. The cursor D readout shows the measured ring frequency is 227 kHz.

212

TDS3000 Series User Manual

Application Examples

Using Delay You are analyzing a pulse waveform and use the + Width measurement to measure the waveform pulse width. You notice that the measurement is not stable, which implies that there is jitter in the pulse width.

To use delay to see the jitter, do these steps: 1. Push the DELAY button. 2. Adjust the horizontal POSITION control to set the delay close to to the nominal pulse width (210 ms). Push the COARSE button to make delay adjustments more quickly. Push COARSE again to fine tune the delay time. The falling edge of the pulse is now near the center of the screen. When delay is on, the horizontal expansion point separates from the trigger point and remains in the center of the screen.

TDS3000 Series User Manual

213

Application Examples

3. Adjust the horizontal SCALE to a faster time base setting and increase the WAVEFORM INTENSITY to see the jitter in the pulse width.

NOTE. You can toggle the delay function on and off to view signal details at two different areas of interest.

214

TDS3000 Series User Manual

Application Examples

Measuring Jitter To measure the peak-to-peak jitter, do these steps: 1. Push the CURSOR button. 2. Push the Function screen button. 3. Select the V Bars cursors. 4. Push the Bring Both Cursors On Screen screen button to quickly locate the cursors. 5. Place one cursor at the first falling edge and place the other cursor at the last falling edge. 6. Read the peak-to-peak jitter in the D readout (1.40 ms).

You can also measure the minimum and maximum pulse widths. When you select the first cursor, the @ readout displays the minimum pulse width (210 ms). When you select the second cursor, the @ readout displays the maximum pulse width (211 ms).

TDS3000 Series User Manual

215

Application Examples

Triggering on a Video Signal


You are testing the video circuit in a piece of medical equipment and need to display the video output signal. The video output is an NTSC standard signal. Use the video trigger to obtain a stable display.

Ch 1

To trigger on the video fields, do these steps: 1. Push the trigger MENU button. 2. Push the Type screen button to select Video. 3. Push the Standard screen button to select 525/NTSC.

216

TDS3000 Series User Manual

Application Examples

4. Push the Trigger On screen button. 5. Select Odd. 6. Adjust the horizontal SCALE to see a complete field across the screen. 7. Push the acquire MENU button. 8. Push the Resolution screen button. 9. Select Normal acquisition resolution. Normal acquisition resolution is the best choice to acquire a video field signal because the signal contains a great amount of horizontal detail.

If the signal had been noninterlaced, you could choose to trigger on All Fields.

TDS3000 Series User Manual

217

Application Examples

Triggering on Lines. You can also look at the video lines in the field. To trigger on the lines, do these steps:

1. Push the Trigger On screen button. 2. Select All Lines. 3. Adjust the horizontal SCALE to see a complete video line across the screen.

NOTE. The optional Extended Video Application Package adds, among other things, a video QuickMenu, video Autoset, the capability to operate at custom scan rates, and trigger on specific video lines.

218

TDS3000 Series User Manual

Application Examples

Looking at Modulation. A dedicated video waveform monitor clearly

shows the modulation in a video signal. To see a similar modulation display on the oscilloscope screen, do these steps: 1. Start with triggered display of the video lines. 2. Push the acquire MENU button. 3. Push the Resolution screen button. 4. Select Fast Trigger acquisition resolution. 5. Adjust the WAVEFORM INTENSITY control for the amount of modulation you want to see. The oscilloscope now shows signal modulation in shades of intensity and appears similar to the display of a video waveform monitor or an analog oscilloscope. Fast trigger acquisition resolution is the best choice to acquire a video line signal with a rapidly changing shape.

TDS3000 Series User Manual

219

Application Examples

Capturing a Single-Shot Signal


The reliability of a reed relay in a piece of equipment has been poor and you need to investigate the problem. You suspect that the relay contacts arc when the relay opens. The fastest you can open and close the relay is about once per minute so you need to capture the voltage across the relay as a single-shot acquisition. To set up for a single-shot acquisition, do these steps: 1. Adjust the vertical SCALE and horizontal SCALE to appropriate ranges for the signal you expect to see. 2. Push the acquire MENU button. 3. Push the Resolution screen button. 4. Select Normal acquisition resolution. 5. Push the SINGLE SEQ (single sequence) button.

The SINGLE SEQ button sets trigger parameters to the correct settings for a single-shot acquisition.

220

TDS3000 Series User Manual

Application Examples

Optimizing the Acquisition The initial acquisition shows the relay contact beginning to open at the trigger point. This is followed by a large spikes that indicate contact bounce and inductance in the circuit. The inductance can cause contact arcing and premature relay failure. Before you take the next acquisition, you can adjust the vertical and horizontal controls to give you a preview of how the next acquisition might appear. As you adjust these controls, the current acquisition is repositioned, expanded, or compressed. This preview is useful to optimize the settings before the next single-shot event is captured. When next acquisition is captured with the new vertical and horizontal settings, you can see more detail about the relay contact opening. You can now see that the contact bounces several times as it opens.

TDS3000 Series User Manual

221

Application Examples

Using the Horizontal Zoom Function If you want to take a close look at a particular spot on the acquired waveform, use the horizontal zoom function. To look closely at the point where the relay contact first begins to open, do these steps: 1. Push the zoom button .

2. Use the horizontal POSITION to place the expansion point close to where the relay contact begins to open. 3. Adjust the horizontal SCALE to magnify the waveform around the expansion point. The ragged waveform and the inductive load in the circuit suggest that the relay contact may be arcing as it opens.

The zoom function works equally well when acquisition is running or is stopped. Horizontal position and scale changes affect only the display, not the next acquisition.

222

TDS3000 Series User Manual

Application Examples

Using the Disk Drive


You need to do some work inside a remote site. You expect to use the oscilloscope to look at signal waveforms and then bring waveform information back to the office to complete a report and perform additional analysis. To do this, take along an IBM-compatible floppy disk.

Remote site

When you need to capture screen images, it may be most convenient to first save them to disk. Once on the disk, you can load the screen images into a PC, print hard copies with a printer attached to the oscilloscope or your PC, or import the screen images into desktop publishing software to produce a report. You can also save waveform data on the disk. From the disk, you can recall waveforms to the oscilloscope display or you can import the data into spreadsheet and Mathcad software to perform additional analysis. If you have oscilloscope setups that you want to use again, you also can store them on the disk. See Save/Recall on page 340 for more information on this capability. For information about remote control of the oscilloscope and network printing, see the AD007 and WSTRO accessories on page C4.

TDS3000 Series User Manual

223

Application Examples

Saving Screen Images While working at the remote site you discover a control signal that you want to capture periodically to show its long-term variation. You want to include these waveforms in a report you are preparing back at the office. Your desktop publishing software can import BMP graphics, so you decide to use this format for the screen images. To setup this configuration, do these steps: 1. Insert a floppy disk into the disk drive. 2. Push the UTILITY button. 3. Push the System screen button to select Hard Copy. 4. Push the Format screen button. 5. Select BMP Windows mono image file format (you may need to push the more screen button several times before you see this choice). 6. Push the Port screen button. 7. Select File to send hard copies to disk. The oscilloscope reads the disk directory and displays its contents.
Naming Files. It is good practice to give disk files descriptive names

so you can recognize them when you get back to the office. You will be saving images of the control signal so you decide CNTRL is a logical target file name to use.

224

TDS3000 Series User Manual

Application Examples

The oscilloscope can append an automatic sequence number to the target file name. This feature is convenient because you want to capture a screen image of the same control signal every five minutes. To set up the target file name and automatic sequence, do these steps: 1. Push the File Utilities screen button. 2. Use the general purpose knob to highlight the file TEK?????.BMP. 3. Select the Rename screen button. 4. Use the screen buttons to clear the existing file name and enter the new file name CNTRL???.BMP. The question marks are placeholders for an automatic number sequence from 000 to 999. 5. Push the OK Accept screen button to setup the target base file name. 6. Push MENU OFF to remove the file list from the display.

TDS3000 Series User Manual

225

Application Examples

Running the Test. To capture the control signal every few minutes, do

these steps: 1. Display the signal, measurements, and menus as you want them to appear in the screen images. 2. Push the hardcopy button .

3. Repeat step 2 every few minutes until you are finished with the test. 4. When you are finished, push UTILITY to see the list of sequential files that have been saved.

The files are labeled with their sequential names as well as the time and date each was created. You can save up to four BMP images or approximately 35 TIFF images on a single 1.44 MB floppy disk. You can also enable file compression (UTILITY > System: Hard Copy > Options) to compress files in zip format so that more files can fit on a floppy disk.

226

TDS3000 Series User Manual

Application Examples

Saving Waveform Data You find another signal that you want to analyze with a spreadsheet program back at the office. To save the waveform data on a disk, do these steps: 1. Display the signal on the oscilloscope screen. 2. Push the SAVE/RECALL button. 3. Push the Save Wfm screen button. 4. Select To File. 5. Select Spreadsheet File Format. The default target file, TEK?????.CSV, is now automatically highlighted. 6. Push the Save To Selected File screen button to save the waveform. 7. Push the File Utilities screen button to see the saved waveform file TEK00000.CSV in the disk directory.

TDS3000 Series User Manual

227

Application Examples

228

TDS3000 Series User Manual

Reference

Introduction to Reference
This chapter contains detailed information about the operation of the oscilloscope. The topics in this chapter are arranged by front-panel button or control group name.
Reference topic Acquire Cursor Display Hard copy Horizontal controls Measure QuickMenu Save/Recall Trigger controls Utility Vertical controls Page 32 312 317 321 325 333 339 340 349 360 370

TDS3000 Series User Manual

31

Acquisition Controls

Acquisition Controls
Run/Stop Button Push the RUN/STOP button to stop and start waveform acquisition. Also push RUN/STOP when you want to resume continuous acquisition after a single-sequence acquisition. Readout in the top left corner of the display shows the status of the acquisition.
Acquisition status readout Run: Roll: Stop: PreVu: Description Acquisition is running. A roll-mode acquisition is running. Acquisition is stopped. Previewing; waiting for trigger.

While acquisition is running or stopped, you can use these controls to examine the waveforms: H Channel buttons to select a channel with horizontal POSITION and SCALE to H Zoom button magnify waveforms (does not affect the actual time base or trigger position settings) H WAVEFORM INTENSITY to adjust gray-scale level H CURSOR button to activate cursors to measure waveforms H MEASURE button to select automatic measurements of waveforms H Hard copy button to print a hard copy

While acquisition is stopped, you can change the vertical and horizontal controls to use in the next acquisition. See page 37 for more information about this function.

32

TDS3000 Series User Manual

Acquisition Controls

Single Sequence Button Push the SINGLE SEQ button to execute a single-shot acquisition. The function of the SINGLE SEQ button depends on the acquisition mode.
Acquisition mode Sample or Peak Detect Envelope N or Average N SINGLE SEQ function One acquisition of each displayed channel is acquired, concurrently N acquisitions of each displayed channel are acquired (N is user adjustable with the general purpose knob)

When you push the SINGLE SEQ button, the oscilloscope does the following things: H The trigger mode is set to Normal H The trigger system arms and the light next to the SINGLE SEQ button turns on After the single-sequence acquisition is complete, acquisition stops and the light next to the SINGLE SEQ button turns off. Push the SINGLE SEQ button again to acquire a new sequence, or push the RUN/STOP button to restart continuous acquisition.

TDS3000 Series User Manual

33

Acquisition Controls

Autoset Button Push the AUTOSET button to automatically adjust the vertical, horizontal, and trigger controls for a usable display. You can manually adjust any of these controls if you need to optimize the display. When you are using more than one channel, the autoset function sets the vertical scale for each channel and positions the channels to prevent overlapping. The autoset function selects the lowest-numbered channel in use and then uses that channel to set the horizontal and trigger controls. The autoset function also changes the following oscilloscope settings: H Acquisition mode is set to Sample H Bandwidth limits are set to Full H Zoom is turned off H Trigger is set to Auto mode and minimum holdoff H Trigger is set to Edge type, DC coupling, and rising slope H B trigger is turned off H XY display format is turned off H Channel 1 is turned on and selected if no active channels are in use If you push the AUTOSET button by accident, you can undo it with these steps: 1. Push the Acquire MENU button. 2. Push the Autoset screen button and then push the Undo Autoset screen button.

34

TDS3000 Series User Manual

Acquisition Controls

Waveform Intensity The WAVEFORM INTENSITY adjusts the intensity of the waveforms in the display. The term Digital Phosphor represents the way this control simulates the intensity control of an analog oscilloscope. At maximum intensity, all waveform points are displayed at full brightness. As you decrease the intensity, you can see intensity grading in the waveforms. The brightest parts of the waveform are the points most frequently acquired; the dimmer parts represent less frequently acquired points. All points decay in intensity over time unless the display persistence is set to infinite. Use a medium intensity setting to get an analog-oscilloscope view of time-varying signals and those containing modulation. Use the maximum intensity setting to view the signals the way most digital oscilloscopes display them. You can turn on display persistence to slow or prevent the decay of the waveform points. With persistence on, you can simulate the function of an analog storage oscilloscope. See page 317 for more information about display persistence. NOTE. The waveform intensity may change when you change oscilloscope acquisition modes or horizontal scale settings. Use the INTENSITY knob to readjust waveform intensity.

TDS3000 Series User Manual

35

Acquisition Controls

Acquire Menu Push the Acquire MENU button to show the acquire menu.
Bottom Mode Side Sample Peak Detect Envelope N Description Use for normal acquisition. Detects glitches and reduces the possibility of aliasing. Captures variations of a signal over a period of time. (Adjust N with the general purpose knob.) Reduces random or uncorrelated noise in the signal display. (Adjust N with the general purpose knob.) Acquires 500 point waveforms at a fast repetition rate. Acquires 10,000 point waveforms with more horizontal detail. Resets the horizontal delay time to zero. Executes the autoset function. (Optional application modules may add choices to execute specialized autoset functions.) Reverts to the settings before the last autoset.

Average N

Horizontal Resolution

Fast Trigger (500 points) Normal (10k points)

Reset Horizontal Delay Autoset

Set to 0 s Normal Autoset

Undo Autoset

36

TDS3000 Series User Manual

Acquisition Controls

Key Points
Vertical and Horizontal Preview. The vertical and horizontal preview

functions let you change the vertical and horizontal controls while acquisition is stopped or while it is waiting for the next trigger. The oscilloscope rescales and repositions the current acquisition in response to the new control settings, and then uses the new settings for the next acquisition. Preview helps you optimize these control settings before the next acquisition; this makes it easier to work with signals that are single-shot or have a low repetition rate. H For more information about vertical preview, see page 373. H For more information about horizontal preview, see page 330. While acquisition is stopped you can make changes to other controls, but those changes will only take affect with the next acquisition. There is no preview for control changes other than the vertical and horizontal controls. The preview functions also do not affect automatic measurements, cursor measurements, or the Math waveform. This data for these functions is always based on the current acquisition. If you rescale or reposition a channel waveform horizontally, it may not appear time-correlated to the automatic measurements, cursor measurements, or the Math waveform.
Acquisition Modes. You can choose one of four acquisition modes: Sample, Peak Detect, Envelope, or Average. The next two pages describe these acquisition modes in detail.

TDS3000 Series User Manual

37

Acquisition Controls

Samples acquired in four acquisition intervals Interval 1 2 3 4

Acquisition mode

Displayed record points (at maximum horizontal magnification) Interval 1 2 3 4

Sample Sample mode acquires one sample in each interval.

Peak Detect

Peak Detect mode uses the lowest and highest amplitudes from two consecutive intervals.

Sample. Use Sample acquisition mode for the fastest acquisition at

any SEC/DIV setting. Sample mode is the default mode.


Peak Detect. Use Peak Detect acquisition mode to limit the possibility of aliasing. Also, use Peak Detect for glitch detection. You can see glitches as narrow as 1 ns.

Peak Detect is only functional for sample rates up to 125 MS/s. For 250 MS/s and faster sample rates, the oscilloscope reverts to Sample acquisition mode, where the narrowest detectable pulse width is 1/(sample rate).

38

TDS3000 Series User Manual

Acquisition Controls

Three acquisitions from one source Acquisition 1 2 3

Acquisition mode

Envelope Envelope uses Peak Detect mode for each individual acquisition. Finds highest and lowest record points over many acquisitions

Average Average uses Sample mode for each individual acquisition. Calculates average value for each record point over many acquisitions

Envelope. Use Envelope acquisition mode to capture the minimum

and maximum extremes of a signal over a specified number of acquisitions (N). The enveloped waveform data acquisition clears and then starts over after each N acquisitions. If you push the SINGLE SEQ button, envelope acquisition stops after N acquisitions. Use the general purpose knob to set the number of acquisitions.
Average. Use Average acquisition mode to reduce random or

uncorrelated noise in the signal you want to display. The averaged waveform is a running average over a specified number of acquisitions (N). If you push the SINGLE SEQ button, average acquisition stops after N acquisitions. Use the general purpose knob to set the number of acquisitions.

TDS3000 Series User Manual

39

Acquisition Controls

If you probe a noisy square wave signal that contains intermittent, narrow glitches, the waveform displayed will vary depending on the acquisition mode you choose.

Sample

Peak Detect

Envelope

Average

310

TDS3000 Series User Manual

Acquisition Controls

Acquisition Resolution. You can choose either Normal or Fast Trigger acquisition resolution. This setting determines the record length that is acquired and affects the other factors shown in the table below. Factor Record length Maximum acquisition rate (approximate) Normal 10,000 points 450 waveforms/s Fast trigger 500 points 3,000 waveforms/s 10X

Maximum horizontal zoom factor 200X

Choose Normal or Fast Trigger acquisition resolution based on the characteristics of the signal you want to acquire.
Signal characteristic Large amount of horizontal detail Shape is stable or changes relatively slowly Single shot High trigger repetition rate Shape changes rapidly Contains modulation Better choice Normal Normal Normal Fast Trigger Fast Trigger Fast Trigger

TDS3000 Series User Manual

311

Cursor

Cursor
Push the CURSOR button to show the cursor menu.
CURSOR

Bottom Function

Side Off H Bars V Bars Bring Selected Cursor to Center Screen Bring Both Cursors On Screen

Desription Turns cursors off. Use to make vertical measurements. Use to make both vertical and horizontal measurements. Moves the active cursor to center screen. Moves any off-screen cursor onto the screen. Sets cursors to move independently. Sets cursors to move together when cursor 1 is selected.

Mode

Independent Tracking

312

TDS3000 Series User Manual

Cursor

Bottom V Bar Units

Side Sec (s) / 1/sec (Hz) Ratio (%) Phase () Use Cursor Positions as %/

Desription Sets horizontal units to seconds or frequency (Hz). Sets V Bar measurement units to percent. Sets V Bar measurement units to degrees. Sets the V Bar measurement scale so that 0% or 0 is the current position of the left V Bar cursor and 100% or 360 is the current position of the right V Bar cursor.

Use 5 divs as %/ Sets the V Bar measurement scale so that 5 screen major divisions is 100% or 360, where 0% or 0 is 2.5 divisions and 100% or 360 is +2.5 divisions from center vertical graticule.

TDS3000 Series User Manual

313

Cursor

Bottom H Bar Units

Side Base

Desription Sets the H Bar units to be the same as the selected waveforms vertical measurement units (volts, IRE, dB, and so on). Sets H Bar units to percent. Sets the H Bar measurement scale so that 0% is the current position of the lowest H Bar cursor and 100% is the current position of the highest H Bar cursor. Sets H Bar measurement scale so that 5 screen major divisions is 100%, where 0% is 2.5 divisions and 100% is +2.5 divisions from the center horizontal graticule.

Ratio (%) Use Cursor Positions as 100%

Use 5 divs as 100%

Key Points
Cursor Movement. Use the general purpose knob to move the active cursor. Push the SELECT button to change which cursor is active. The active cursor is the solid line. Faster Cursor Movement. Push COARSE to set the general purpose

knob to make faster cursor movement.


Finding Cursors. When using zoom, delay, or the fastest time base settings, the cursors can go off screen. If you want to find them, use the Bring Both Cursors On Screen function to move them onto the screen. Fine Cursor Movement. When you magnify the waveform with the

zoom function, you can easily set the cursor to any point on the waveform.

314

TDS3000 Series User Manual

Cursor

D 4.16 V @ 1.78 V

Horizontal Bar cursors

D 6.32 V @ 3.16 V D 5.86 ms @ 1.06 ms

Vertical Bar cursors

D Readout. The D readouts indicate the difference between the cursor

positions.
@ Readout. For H Bar or V Bar cursors, the voltage readout after the

@ symbol indicates the location of the active cursor relative to zero volts. For V Bar cursors, the time readout after the @ symbol indicates the location of the active cursor relative to the trigger point.
Cursor Interaction with Preview. If you change a vertical or horizontal

control setting while acquisition is stopped or it is waiting for trigger, the cursors move with the waveforms and cursor measurements remain valid.
Taking Gray-Scale Measurements. Using cursors is often the best way to

take simple measurements on waveforms that contain significant gray-scale information. The automatic measurements operate only on the latest acquisition, not on the previous acquisitions that are displayed in gray scale. However, you can set the cursors to surround and measure the gray-scale area of the waveform.

TDS3000 Series User Manual

315

Cursor

Cursors at Same Position. If both cursors are at the same position, and

H Bars or V Bars are set to Ratio or Phase, both cursors are set to 0% (or 0). 100%/360 is set to one pixel width away from the cursor position.
V Bars and FFT. When the selected waveform is an FFT waveform, selecting V Bars and Phase sets the measurement to percent. Tracking Mode. In cursor tracking mode, both cursors move together when cursor 1 is selected. Enabling tracking mode automatically selects cursor 1 as the active cursor. If cursor 2 is selected while in tracking mode, only cursor 2 will move.

316

TDS3000 Series User Manual

Display

Display
Push the DISPLAY button to show the display menu.
DISPLAY

Bottom Waveform Display

Side Dots Only Persist Time Set to Min

Description Set to on to see dots only. Set to off to see dots and vectors. Sets persist time. Sets the persistence time to its minimum value.

Clear Persistence Clears any displayed persistence. Backlight Intensity High Medium Low Graticule Full, Grid, Cross Hair, Frame Use for bright ambient conditions. Use for dim ambient conditions. Use to extend battery-operation time. Sets the graticule type.

TDS3000 Series User Manual

317

Display

Bottom XY Display

Side Off (YT) Triggered XY Gated XYZ

Description Turns off an XY display. Turns on triggered XY display. Turns on gated XY display. Displays XY signals when Z-channel signal is above a set level. Only available on 4-channel instruments. Sets Ch2, Ch3, or Ch4 as the Y channel versus Ch1 as X. Sets Ref2, Ref3, or Ref4 as the Y channel versus Ref1 as X. Sets Ch2, Ch3, or Ch4 as the Z-channel gate source, and sets the gating channel threshold level. Selects the color display. Sets all waveforms to high-contrast black and white.

Ch1 (X) Versus Ref1 (X) Versus or Gated By

Color Palette

Normal Monochrome

Key Points
Waveform Dots and Vectors. With Dots Only set to off, the vectors

between samples can be filled; increase the WAVEFORM INTENSITY control to increase the amount of vector fill between samples. The vector fill is most visible on fast signal edges or when horizontal is turned on. zoom Turn Dots Only on if you want to see just the actual samples.

318

TDS3000 Series User Manual

Display

Waveform Persistence. Turn on waveform persistence to slow the

decay of waveform points. You can set the persistence to a specific time or to infinite. Infinite persistence keeps all waveform points in the display until you change a control setting that erases the display.
Display Colors. The channel buttons, waveforms, icons, and readouts

are color-coded to help you identify them easily. The colors are preset and cannot be adjusted. However, you can select the Monochrome Color Palette if you prefer to see all the waveforms in high-contrast black and white.
XY Waveform Triggering. The XY waveform is triggered so you can synchronize periodic input signals to the XY waveform. This feature is useful when only one part of the period contains valid information that you want to see in the XY format. Set the time base and trigger location to acquire just that part of the period.

When you want to see the complete period of the signals regardless of the time base setting, set the trigger source to an unused channel and the trigger mode to Auto.
XY Waveform Scale and Position. If you want to display channel 1 in

the horizontal axis versus channel 3 in the vertical axis, for example, use these controls to scale and position the XY waveform: H Push the CH 1 button and use the vertical SCALE and POSITION controls to set the horizontal scale and position of the XY waveform. H Push the CH 3 button and use the vertical SCALE and POSITION controls to set the vertical scale and position of the XY waveform.

TDS3000 Series User Manual

319

Display

XY Waveform Restrictions. The math waveform, cursor, zoom, and autoset functions do not work in the XY display format. All reference waveforms displayed in XY format must have the same record length (500 or 10,000 points). Gated XYZ. Displays XY signals only when the Z (gating) channel is

true. Gated XYZ is similar to analog oscilloscope modulated XYZ mode except that the displayed XY signal is either on or off; there is no intensity modulation. Gated XYZ is useful for showing constellation diagrams. Push the Gated By screen button to select the Z (gate) source channel. The general purpose knob sets the Z-channel threshold level. Z-channel signals above the set threshold are true and open the XY signal gate; Z-channel signals below the set threshold are false and close the XY signal gate. The gating channel is always a high-true logic; to emulate a low-true gating logic, use the vertical menu to invert the Z-channel signal.

320

TDS3000 Series User Manual

Hard Copy

Hard Copy
After you connect a printer and set up the oscilloscope, push the hard-copy button at the left of the display to make a hard copy. You can also store hard-copy images on a floppy disk (in normal or compressed format) and then transfer them later to a PC for printing or use in a report. Connecting a Printer Connect your printer to the parallel printer port on the rear panel of the oscilloscope. If you have one of the optional communication modules installed, you can also connect to the serial RS-232 port or the parallel GPIB port if your printer supports one of these interfaces. You can use the optional TDS3EM Ethernet communication module to connect to network printers. Setting Up to Print Follow these steps to set up the oscilloscope to print a hard copy: 1. Push the UTILITY menu button. 2. Push the System screen button to select Hard Copy. 3. Push the Format screen button and then choose the printer format appropriate for your application. 4. Push the Options screen button to select the image orientation (portrait or landscape) as well as turn on or off hard copy file compression. 5. Push the Ink Saver screen button and select On for most applications. If you want the hard copy colors to be the same as the screen colors, you can select Off (see page 323). 6. Push the Port screen button and select the port that your printer is connected to or select File to save the hard copy on a floppy disk (see page 343 for more information about using the floppy disk). 7. Push the hard copy button .

TDS3000 Series User Manual

321

Hard Copy

Key Points
Printer Formats. The oscilloscope supports the following printers and

file formats.
Format Thinkjet Deskjet mono Deskjet color Laserjet Epson Interleaf TIFF RLE color PCX mono PCX color BMP mono BMP color EPS mono EPS color Bubble Jet DPU-3445 Description HP monochrome ink-jet printer HP monochrome ink-jet printer HP color inkjet printer HP monochrome laser printer Epson 9-pin and 24-pin dot matrix printer *.img Interleaf image object file format *.tif Tag image file format Windows color image file format PC Paintbrush monochrome image file format PC Paintbrush color image file format Windows monochrome image file format Windows color image file format Encapsulated PostScript monochrome image Encapsulated PostScript color image Canon BJC-50, BJC-80 color printers Seiko DPU-3445 thermal printer

322

TDS3000 Series User Manual

Hard Copy

Hard Copy File Compression. When compression is set to on, the

oscilloscope compresses the hard copy data, using the current printer format, into a gnuzip file format, with the extension .gz. Compressing hard copy files lets you store more screen captures on a floppy disk. Centronics output is never compressed. .gz files can be decompressed using PKZIP or WinZip programs.
Color and Gray-Scale Printing. You can print a color hard copy that uses

the display colors. Gray-scale waveform information is printed as shades of color. If you have a Deskjet or Laserjet monochrome printer, gray-scale waveform information is printed as a dithered image.
Ink Saver and Preview. As an alternative to printing the display colors, turn on the Ink Saver function to print a hard copy with a white background. This function saves printer ink while it preserves the color coding of the waveforms and readouts. Ink saver also works with the monochrome print formats.

Push and hold the Preview screen button to show how the colors will appear on the paper.
Clear Spool. You can push the Clear Spool screen button to empty the

print spooler to stop a hard copy operation in progress, if the hard copy port connection is not made due to incompatible settings (such as baud rate), or if you lose the hard copy port connection before the hard copy is complete.

TDS3000 Series User Manual

323

Hard Copy

Date and Time Stamp. To print the current date and time on your hard copies, do these steps:

1. Push the UTILITY button. 2. Push the System screen button to select Config. 3. Push the Set Date & Time screen button. 4. Set the Display Date/Time to On to add the current date and time to the display screen. 5. Push the Menu Off button.
Printer Error Message. To avoid printer errors, always power on the

printer and allow it to complete its initialization process before powering on the oscilloscope. If you see a Hard copy device not responding error message, power the oscilloscope off and then back on, and then try printing again. If the printer still does not work, check that the printer is online, you have selected the correct printer format on the oscilloscope, there are no paper jams, and that the printer cable is firmly connected to the printer and oscilloscope.

324

TDS3000 Series User Manual

Horizontal Controls

Horizontal Controls
Use the horizontal controls to adjust the time base, adjust the trigger location, and to examine waveform details more closely. Horizontal Position Control When delay is off, the horizontal POSITION control moves the trigger point within the acquired waveforms. You can select full pretrigger, full posttrigger, or any point in between. Use a pretrigger setting (trigger position near 100% of the record) to acquire waveforms leading up to a trigger event. For example, if you can trigger on an error condition, the waveforms leading up to the error condition might tell you why the error occurred. Use a posttrigger setting (trigger position near 0% of the record) when you want to acquire waveforms that follow a trigger event. Use a mid-screen setting when you are interested in information both before and after the trigger event.
Trigger point

Pretrigger

Posttrigger

Acquired waveform

Acquired waveform

See Delay Button on page 327 and Zoom Button on page 329 for information about how the horizontal POSITION control operates when these functions are active.

TDS3000 Series User Manual

325

Horizontal Controls

The trigger position is marked with the letter T at the top of the graticule and also in the waveform record icon at the top of the screen.
Trigger marker and horizontal expansion point

Waveform record icon

The small inverted triangle is the horizontal expansion point. When you change the horizontal SCALE setting, the waveforms contract or expand about this point. When delay is off, the horizontal expansion point is the same as the trigger point.

326

TDS3000 Series User Manual

Horizontal Controls

Delay Button Push the DELAY button when you want to delay the acquisition relative to the trigger event. Turn the horizontal POSITION control counterclockwise to increase the delay; the trigger point moves to the left and ultimately outside of the acquired waveform. Then, you can adjust the horizontal SCALE to acquire more detail around the area of interest (center of the screen). When delay is on, the trigger point separates from the horizontal expansion point. The horizontal expansion point stays at the center of the screen. The trigger point can move off the screen; when this happens, the trigger marker turns to point in the direction of the trigger point.
Acquired waveform Trigger point

Delay time Expansion point

Use the delay feature when you want to acquire waveform detail that is separated from the trigger event by a significant interval of time. For example, you can trigger on a sync pulse that occurs once every 10 ms and then look at high-speed signal characteristics that occur 6 ms after the sync pulse.

TDS3000 Series User Manual

327

Horizontal Controls

In the screen example below, the trigger markers show that the trigger point is before the acquired waveform. The delay time, shown in the readout, is the time from the trigger point to the expansion point (center screen).

Delay time readout

The interaction between delay and other functions is summarized in the table below.
Function Trigger point Expansion point Horizontal SCALE Horizontal POSITION Delay off Any point within the acquired waveform Same as the trigger point Sets time base Sets trigger position in the acquired waveform Delay on Can occur prior to the acquired waveform Always center screen Sets time base Sets delay time

328

TDS3000 Series User Manual

Horizontal Controls

Horizontal Scale Control Use the horizontal SCALE control to adjust the time base. When delay is off, the scale expands or contracts around the trigger point. When delay is on, the scale expands or contracts around the center of the screen (see Delay Interactions on page 332 for possible exceptions). When the zoom function is active, use the horizontal SCALE control to adjust the amount of horizontal magnification (the actual time-base setting remains unchanged). The magnified waveforms always expand or contract around the center of the screen. Zoom Button Push the zoom button to magnify the current acquisition along the horizontal axis so you can see more detail. Use the horizontal SCALE control to adjust the amount of magnification. Use the horizontal POSITION control to select the portion of the waveform you want to magnify. When zoom is on, changes you make to these controls do not affect the actual time base or trigger position settings. The split screen shows the entire selected waveform in the upper window to give you a point of reference as you examine the details in the lower window.
Actual time base

Zoom time base

Zoom position

TDS3000 Series User Manual

329

Horizontal Controls

Key Points
Maximum Zoom Magnification Factor. If you are using Normal

acquisition resolution, the maximum horizontal magnification factor is 200X; in Fast-trigger the maximum is 10X.
Horizontal Zoom and Preview. There are two ways you can magnify a

stopped acquisition, horizontal zoom or preview. The differences in how horizontal zoom and preview interact with other functions are shown below.
Function Horizontal SCALE Horizontal POSITION Horizontal zoom Horizontal preview

Sets magnification factor Changes time base for the next acquisition Selects portion of waveform to magnify Turns delay on and off Changes trigger position or delay time for the next acquisition Turns delay on and off

DELAY button Math waveform

Remains valid; magnifies Remains fixed; does not and positions with other track changes to channel waveforms waveforms Remain functional with valid readouts Gray-scale information may be temporarily reduced Remain locked to channel waveforms Gray-scale information is lost

Cursors and automatic measurements Gray-scale

330

TDS3000 Series User Manual

Horizontal Controls

Slow Horizontal Settings. At horizontal SCALE settings of 40 ms/div

or slower, the oscilloscope may display waveforms in roll mode. As the rolling waveform fills the screen, the waveform intensity and vector fill appears to decrease. The oscilloscope automatically reduces the number of displayed points in order to maintain a high acquisition rate: this does not represent a loss of acquired data. When you stop the acquisition, the original waveform intensity and vector fill returns to the display.
Using Zoom and Delay Together. You can use both zoom and delay at the same time to magnify a delayed acquisition. Fast Time-Base Settings. At the fastest time-base settings, only a

portion of the waveform shows in the display. The waveform record icon indicates that portion with brackets. Push the zoom button and then use the horizontal POSITION control to scroll through the entire waveform to see any portion you want. The affected time-base settings are shown below.
Acquisition resolution Normal Fast Trigger Affected time-base settings 100 ns/div to 1 ns/div 4 ns/div to 1 ns/div

At the fastest time-base settings, the maximum zoom magnification factor is also reduced.

TDS3000 Series User Manual

331

Horizontal Controls

Delay Interactions. The maximum delay setting is a function of the

time-base setting and acquisition resolution. If you set a large positive or negative delay, that amount of delay might be automatically reduced if you make the following additional control changes: H Change to a faster time-base setting H Change from Fast Trigger to Normal acquisition resolution If the delay reduction occurs, it may cause the waveform horizontal position to shift.
Negative Delay. You can select up to ten divisions of negative delay.

At the fastest time base settings, you can use negative delay to see more of the waveform that occurs before the trigger point.
Roll Mode Display. To obtain a rolling display similar to a strip-chart

recorder, turn off zoom and delay, select the Auto trigger mode, and set the horizontal SCALE control to 40 ms/div or slower. Subsequent changes to the horizontal SCALE cause the roll-mode display to erase and restart.

332

TDS3000 Series User Manual

Measure

Measure
Push the MEASURE button to show the measure menu.
MEASURE

Bottom Select Measurement Remove Measurement

Side

Description See table on page 337 for a description of the automatic measurements.

Measurement 1 Measurement 2 Measurement 3 Measurement 4

Removes a specific measurement.

All Measurements Removes all measurements. Gating Off Screen Cursors Use to take measurements on the full waveform record. Use to take measurements on the portion of the waveform on screen. Use to take measurements on the portion of the waveform between the V Bar cursors. Moves the active cursor to center screen. Moves any off-screen cursor onto the screen. Automatically uses the best measurement method depending on the signal characteristics. Use to measure pulses. Use to measure other wave shapes.

Bring Selected Cursor to Center Screen Bring Both Cursors On Screen High-Low Setup Auto Select

Histogram Min-Max

TDS3000 Series User Manual

333

Measure

Bottom Reference Levels

Side Set Levels in % or units High Reference Mid Reference Low Reference Set to Defaults

Description Use to choose custom reference levels in relative or absolute units. Sets custom high reference level. Sets custom mid reference level. Sets custom low reference level. Sets reference levels to default values.

Key Points
Choosing Measurements. You can perform up to four automatic

measurements and display them along the right side of the graticule. All four measurements can apply to a single channel or you can spread the measurements over several channels. You can also take measurements on math and reference waveforms. First push a channel, MATH, or REF button to select the waveform you want to measure and then select a measurement. The table beginning on page 337 describes the measurements in detail.
Measurement Interaction with Preview. If you change a vertical or horizontal control setting while acquisition is stopped or is waiting for trigger, measurements track the changes and remain valid. Taking Gray-Scale Measurements. Using cursors is often the best way to

take simple measurements on waveforms that contain significant gray-scale information. The automatic measurements operate only on the latest acquisition, not on the previous acquisitions that are displayed in gray scale. However, you can set the cursors to surround and measure the gray-scale area of the waveform.

334

TDS3000 Series User Manual

Measure

High-Low Setup. The oscilloscope determines the 10%, 50%, or 90%

levels of the waveform and then uses them to calculate the measurements. You can choose the method used to determine these levels: H Histogram sets the values statistically; it finds the most common value either above or below the midpoint (depending on whether it is defining the high or low reference level). Since this statistical approach ignores short-term aberrations (overshoot, ringing, noise), histogram is the best method for measuring digital waveforms and pulses. H Min-max uses the highest and lowest values of the waveform record. This method is best for measuring waveforms that have no large, flat portions at a common value, such as sine waves and triangle waves. H Auto Select chooses one of the above methods automatically depending on the signal characteristics. Auto Select chooses the histogram method if the histogram contains prominent peaks. If not, Auto Select chooses the min-max method.
Measurement Gating. You can use the gating feature to limit measurements to the portion of the waveform that is on the screen or between the cursors.

When you turn Screen gating on, the oscilloscope only uses the waveform points on screen in its measurements. This function is useful at the fastest time base settings or when you want to take is on). measurements on a magnified waveform (when zoom When you turn Cursor gating on, the oscilloscope displays vertical bar cursors. Use the general purpose knob and SELECT button to place the cursors around the area of interest.

TDS3000 Series User Manual

335

Measure

In the example below, the cursors surround the second positive-going pulse so the oscilloscope can measure the width of that pulse.

When gating is off, the oscilloscope takes measurements over the entire waveform record.
Using Measurement Gating with Cursors. If V Bar cursors are already on

when you select Cursor gating, the cursors do both functions at the same time. The cursor readouts are displayed at the same time that the cursors gate the automatic measurements. If H Bar cursors are on when you select cursor gating, the H Bar cursors are turned off.

336

TDS3000 Series User Manual

Measure

Measurement name

Definition

Amplitude Measured over the entire waveform. Amplitude = High (100%) Low (0%) Burst Width Cycle Mean Cycle RMS Fall Time The duration of a burst. Measured over the entire waveform. The arithmetic mean over the first cycle in the waveform. The true Root Mean Square voltage over the first cycle in the waveform. Time that the falling edge of the first pulse in the waveform takes to fall from 90% to 10% of its amplitude. Reciprocal of the period of the first cycle in the waveform. Measured in Hertz (Hz). The value used as 100%. Calculated using either the min/max or the histogram method. Measured over the entire waveform. The value used as 0%. Calculated using either the min/max or the histogram method. Measured over the entire waveform. The maximum amplitude. The most positive peak voltage measured over the entire waveform. The arithmetic mean over the entire waveform. The minimum amplitude. The most negative peak voltage measured over the entire waveform.

Frequency High

Low

Max

Mean Min

TDS3000 Series User Manual

337

Measure

Measurement name Negative Duty Cycle

Definition Measurement of the first cycle in the waNegative Width veform. Negative Duty Cycle + Period

100%

Negative Measured over the entire waveform. Overshoot 100% Negative Overshoot + LowMin Amplitude Negative Width Measurement of the first negative pulse in the waveform. The time between the 50% amplitude points. Measured over the entire waveform. Peak-to-peak = Max Min Period Time it takes for the first complete signal cycle to complete in the waveform. Measured in seconds. Measurement of the first cycle in the waveform. Positive Duty Cycle + Positive Width 100% Period

PkPk

Positive Duty Cycle

Positive Measured over the entire waveform. Overshoot MaxHigh 100% Positive Overshoot + Amplitude Positive Width Measurement of the first positive pulse in the waveform. The time between the 50% amplitude points. Time that the leading edge of the first pulse in the waveform takes to rise from 10% to 90% of its amplitude. The true Root Mean Square voltage over the entire waveform.

Rise Time

RMS

338

TDS3000 Series User Manual

QuickMenu

QuickMenu
QUICKMENU

Push the QUICKMENU button to see a set of frequently used menu functions on the display. QuickMenus simplify oscilloscope operation and can increase your productivity. Scope is a standard QuickMenu for general purpose oscilloscope use. Some optional application modules also include a custom QuickMenu display. See page 128 for an example of the Scope QuickMenu. Key Points
Using the QuickMenus. To use a QuickMenu, push the screen button

that corresponds to the control you need to set. Push the screen button repeatedly to choose one of the settings. The small arrow icon indicates that there are additional settings available that are not shown. You can use most of the front-panel controls at the same time you are using a QuickMenu. For example, if you push a channel button to select a different channel, the QuickMenu changes to show information about that channel.
Using Other Menus. You can still use the regular menus. For example, if you push the MEASURE button, you can set up and take automatic waveform measurements in the usual way. If you return to the QuickMenu, the measurement is still displayed on the screen. Selecting Between QuickMenus. You may have optional application

modules installed that also have QuickMenu displays. To select the QuickMenu you want to use, push the MENU screen button. This menu item is shown only if application modules that contain a QuickMenu are installed.

TDS3000 Series User Manual

339

Save/Recall

Save/Recall
Push the SAVE/RECALL button to show the save/recall menu.
SAVE/RECALL

Bottom Save Current Setup Set

Side To File To Setup 1 ... To Setup 10

Description Saves a setup to disk. Saves a setup to nonvolatile memory.

Recall Saved Set- From File up Recall Setup 1 ... Recall Setup 10 Recall Factory Setup Save Wfm OK Confirm Factory Init To File

Recalls a setup from disk. Recalls a setup from nonvolatile memory. Initializes the setup. Shows file-format choices: Internal, Spreadsheet, or Mathcad (see page 344). Saves the selected waveform to disk. Saves the selected waveform to nonvolatile memory. Recalls a waveform from disk and displays it as a reference waveform.

To Ref1 ... To Ref4 Recall Wfm From File Recall Ref1 ... Recall Ref4

File Utilities Labels

Accesses floppy disk file utilities. Refer to description on page 343. Lets you assign unique labels to reference waveforms and instrument setups stored in nonvolatile memory. Refer to page 345 for a description of how to edit label text.

340

TDS3000 Series User Manual

Save/Recall

Key Points
Saving to Disk. See Saved Waveform Formats on page 344 for

information about saving to and recalling from the disk.


Saving Setups. To save the current setup to nonvolatile memory, push the Save Current Setup screen button and then select one of the ten storage locations. Then, push the OK Overwrite Saved Setup screen button to complete the operation or the MENU OFF button to cancel the operation. Recalling Setups. To recall a setup from nonvolatile memory, push the Recall Setup screen button and then select one of the ten storage locations. Recalling the Factory Setup. Recall the factory setup to initialize the

oscilloscope to a known setup. Appendix B describes the Factory Setup in detail. To recall the factory setup, push the Recall Factory Setup screen button. Then, push the OK Recall Factory Setup screen button to complete the operation.

TDS3000 Series User Manual

341

Save/Recall

Saving a Waveform. To save a waveform to nonvolatile memory, first select the waveform you want to save. Push the Save Wfm screen button and then select one of the four reference waveform locations. See page 376 for an alternate method you can use to save waveforms.

Saved waveforms contain only the most current acquisition; gray-scale information, if any, is not saved.
Displaying a Reference Waveform. To display a waveform stored in

nonvolatile memory, push the REF button and then push the Ref1, Ref2, Ref3, or Ref4 screen button. When a reference waveform is selected, it appears brighter than other reference waveforms. Reference waveforms do not contain gray-scale information.

342

TDS3000 Series User Manual

Save/Recall

Removing a Reference Waveform from the Display. To remove a

reference waveform from the display, push the REF button and then the Ref1, Ref2, Ref3, or Ref4 screen button to select a reference waveform. Push the waveform off button. The reference waveform is still in nonvolatile memory and can be displayed again.
Erasing All Setups and Waveforms. See Tek Secure on page 363 for instructions on how to erase all setups and waveforms saved in nonvolatile memory.

Using the Disk Drive This section describes how to use the disk drive. The File Utilities submenu is shown below.
SAVE/RECALL

Bottom File Utilities

Side Delete Rename Copy Print Create Directory Confirm Delete Overwrite Lock Format

Description Deletes a file. Renames a file. Copies a file to another directory. Prints a file to a printer connected to one of the hard copy ports. Creates a new directory. Turns On or Off a confirmation message before files are deleted. Sets file overwrite protection to On or Off. Formats a disk (erases all files).

TDS3000 Series User Manual

343

Save/Recall

Key Points
Upgrading Firmware. You can use the disk drive to upgrade the

oscilloscope firmware or install new application packages. See the documentation provided with these packages for instructions.
Saved Waveform Formats. Use the Internal waveform format to save a

waveform to disk if you intend to recall it to the display later as a reference waveform. Use the Spreadsheet format to create a comma-separated data file compatible with most spreadsheet programs. Use the Mathcad format if you intend to import the waveform data into Mathcad software.
Navigating the File System. When you insert an IBM-formatted disk and push the File Utilities screen button, the oscilloscope shows a list of directories and files on the disk.

Use the general purpose knob to select a directory or file. To change the working directory, select the directory and then push the SELECT button. To move up one directory level, select .. and then push the SELECT button.
Automatic File Numbering. The oscilloscope gives all files created by

the oscilloscope the default name TEK?????, where the question marks are placeholders for an automatic number sequence from 00000 to 99999. You can change the TEK????? file to a new name with up to eight characters. If you use fewer than eight characters and include trailing question marks, the oscilloscope numbers the files sequentially if more than one are saved with the same base name. For example, if you rename the TEK?????.ISF file to TEST??.ISF for a series of saved waveforms, the oscilloscope saves the first one as TEST00.ISF, the second one as TEST01.ISF, up to the last one as TEST99.ISF.

344

TDS3000 Series User Manual

Save/Recall

Editing File, Directory, Reference Waveform, or Instrument Setup Names.

You can edit file names,directory names, reference waveform and instrument setup labels, and Ethernet parameters (TDS3EM only). Use the general purpose knob to select an alphanumeric character. Use the screen buttons described below to edit and enter the new name.

Screen button Enter Char and Back Space Delete Clear and OK Accept MENU OFF

Function Enters the selected character into the field. Moves the cursor to a different character in the field. Deletes the character before the cursor location. Deletes the character at the cursor location. Erases the current field value. Selects the field that you want to edit. Applies all field values. Exits the menu without applying field values.

TDS3000 Series User Manual

345

Save/Recall

Deleting Files. To delete a file, select the file with the general purpose knob, push the Delete screen button, and then push the OK Delete screen button when you see the confirmation screen.

If you do not want to see a confirmation screen each time you delete a file, push the Confirm Delete screen button to set it to Off.
Renaming Files. To rename a file, select the file with the general

purpose knob, push the Rename screen button, and then follow the instructions on page 345. Once a directory is created, you cannot rename it. However, you can delete the directory and create a new one with a new name.
Copying Files and Directories. To copy a file or directory, select the file

or directory with the general purpose knob and push the Copy screen button. Now use the general purpose knob and SELECT button to select a destination directory. Push the copy confirmation screen button to complete the operation.
Printing Files. You can print files through any installed printer port to your printer. This function is especially useful if you want to print hard copy files that you have stored on a disk.

To print a file, select the file with the general purpose knob. Push the Print screen button and then select the port to which your printer is connected. Make sure that the oscilloscope is set to send the correct file format to your printer.
Creating a Directory. To create a directory, use the general purpose knob and SELECT button to select the working directory where you want the new directory to reside. Push the Create Directory screen button and then follow the instructions on page 345.

346

TDS3000 Series User Manual

Save/Recall

Formatting a Disk. The oscilloscope can format 1.44 MB IBM-compatible disks. To format a disk, insert it into the disk drive. Push the Format screen button and then push the OK Confirm Format screen button to confirm the operation. If you decide not to format the disk, push the MENU OFF button to stop the format operation.

CAUTION. To avoid loss of data, do not format disks that already contain important data. When you format a disk, all files and directories will be erased and can never be recovered.

Setting Protections. The oscilloscope provides two protections to help

you prevent accidental loss of data: H Confirm Delete shows a confirmation message each time you try to delete a file. You can turn Confirm Delete off if you do not want to see the message. H Overwrite Lock prevents the oscilloscope from writing over existing files. You can turn Overwrite Lock off if you want to be able to write over existing files.

TDS3000 Series User Manual

347

Save/Recall

File Extensions. Files written by the oscilloscope have the following

extensions. The oscilloscope can only read files with the SET, MSK, and ISF extensions.
File extension *.SET *.ISF *.CSV *.DAT *.TJ *.DJ *.LJ *.IBM *.IMG *.TIF *.RLE *.PCX *.BMP *.EPS *.BJC *.DPU *.GZ *.MSK File type Saved setup file Saved waveform file, Internal format Saved waveform file, Spreadsheet format Saved waveform file, Mathcad format Hard copy file, Thinkjet format Hard copy file, Deskjet format Hard copy file, Laserjet format Hard copy file, Epson format Hard copy file, Interleaf format Hard copy file, TIFF format Hard copy file, RLE format Hard copy file, PCX format Hard copy file, BMP format Hard copy file, EPS format Hard copy file, Bubble Jet format Hard copy file, Seiko DPU-3445 format Gnuzip-compressed hard copy file Mask geometry file (requires TDS3TMT module)

348

TDS3000 Series User Manual

Trigger Controls

Trigger Controls
Push the trigger MENU button to show the trigger menu and then push the Type screen button to select Edge or Video. See pages 355 and 359 for descriptions of Edge and Video trigger. See the manual for the TDS3TRG Advanced Trigger or TDS3VID Extended Video application module for more information if either module is installed. Trigger Level Use the trigger LEVEL control to adjust the trigger level. When you change the trigger level, a horizontal line temporarily appears to show you the level on screen. After the line disappears, the trigger level is marked with a small arrow.

Trigger level marker

Trigger level readout

Set to 50% Push the SET TO 50% button to set the trigger level to the 50% amplitude level of the trigger source waveform.

TDS3000 Series User Manual

349

Trigger Controls

Force Trigger Push the FORCE TRIG button to force an immediate trigger event, even in the absence of a signal. This function is useful in these situations: H If you do not see a waveform on the screen when using Normal trigger mode, push FORCE TRIG to acquire the signal baseline to verify that it is on the screen. H After you push the SINGLE SEQ button to set up for a single shot acquisition, you can push the FORCE TRIG button to do a practice acquisition to verify the control settings. B Trigger To use the B trigger, the A trigger type must be Edge. Push the trigger MENU and B TRIG buttons to show the B-trigger menu and to activate triggering using both the A and B triggers. The light next to the B TRIG button indicates B trigger is active. Push the B TRIG button again to return to the single A trigger.

350

TDS3000 Series User Manual

Trigger Controls

Bottom B Trigger After A

Side B Trigger After A Time

Description Sets the oscilloscope to trigger on the next B-trigger event that occurs after a specified period of time from the A trigger. Use the general purpose knob to set the time value.

Set to Delay Time Sets the B Trigger After A Time value (Bw) then Set to the horizontal Bw value, then (Bw) to 0s sets Bw to zero seconds. Bw is the delay time from the B trigger point to the expansion point (center screen). Set to Min B Events Sets B Trigger After A time to 26.4 ns. Sets the oscilloscope to trigger on the nth B-trigger event after the A trigger. Use the general purpose knob to set the event value. Sets the B events count to 1. Sets Source, coupling, slope, and level for the B trigger. These settings or t e tri er e e etti re i e e e t o imil r etti or are independent of similar settings for the tri er t e A trigger. See page 355 for e or descriptions o t e e me item e cri tio of these menu items.

Set to Min Source Coupling Slope Level

The trigger wait time is the minimum time between the A and B triggers. The trigger wait time is not the same as the horizontal delay time. You can use the horizontal delay function to delay acquisition relative to any trigger event, whether it is from the A trigger alone or from a trigger setup that involves both the A and B triggers.

TDS3000 Series User Manual

351

Trigger Controls

The following figures show wait-for-time and wait-for-events triggering and how they relate to the horizontal delay time.

Pretrigger record

Posttrigger record acquired waveform record

Trigger point A Trigger source A

Trigger point B

delay time

Trigger source B Wait for time trigger wait time

Pretrigger record

Posttrigger record acquired waveform record

Trigger point A Trigger source A Trigger source B Wait for events

Trigger point B

delay time

waiting for nth event (where n=5)

352

TDS3000 Series User Manual

Trigger Controls

After the A trigger event is recognized, the oscilloscope begins counting B trigger events. However, for the first B event to be counted, that event must consist of both an opposite-polarity edge and the edge that is counted. The opposite-polarity edge must occur 5 ns after the A trigger arming event. If this condition is not met, the oscilloscope does not count the first event, which results in a trigger on the n+1st event. Refer to the following figure, where n=3 and A and B trigger slopes are set to rising edge.
Trigger point Arming event A 5 ns Trigger event B (for n=3) Opposite-polarity edge Arming event A Trigger event B (for n=3) 1 2 3

Trigger point

4 (n+1)

TDS3000 Series User Manual

353

Trigger Controls

Trigger Status Readout at the top of the screen shows you the current trigger status. The table below explains the trigger status indicators.
Trigger status Auto Trigd PrTrig Explanation The oscilloscope is acquiring using auto trigger. Valid trigger events, if any, are infrequent. The oscilloscope is acquiring using valid trigger events that are frequent enough to avoid auto triggering. The oscilloscope is acquiring the pretrigger part of the waveforms. This status is shown only at the slowest time/division settings. The oscilloscope has acquired the pretrigger part of the waveforms and is waiting for a valid trigger event. The A trigger event has occurred. The oscilloscope is armed and waiting for a valid B trigger event.

Trig? BTrig?

354

TDS3000 Series User Manual

Trigger Controls

Edge Trigger Use Edge triggering to trigger on the rising or falling edge of the input signal at the trigger threshold.
Bottom Type Edge Source Ch1 ... Ch4 AC Line Sets the trigger source to a specific channel. Sets the trigger source to use the AC line signal (not available when operating with batteries). Selects the external trigger source in two-channel oscilloscopes. Ext/10 rovi e l r er tri er level r e t provides a larger trigger-level range at reduced sensitivity. Sets the trigger source to the lowestnumbered active channel in the display. Selects DC coupling. Rejects frequencies above 30 kHz in the trigger signal. Rejects frequencies below 1 kHz in the trigger signal. DC coupling with low sensitivity to reject noise in the trigger signal. Triggers on the rising edge of a signal. Triggers on the falling edge of a signal. Side Description

Ext Ext/10 Vert

Coupling

DC HF Reject LF Reject Noise Reject

Slope

/ (rising edge) \ (falling edge)

TDS3000 Series User Manual

355

Trigger Controls

Bottom Level

Side Level Set to TTL Set to ECL Set to 50%

Description Use to set the trigger level with the general purpose knob. Sets the trigger level to +1.4 V for TTL logic. Sets the trigger level to 1.3 V for ECL logic (Vee = 5.2 V). Sets the trigger level to the 50% amplitude level of the signal. Enables free-running and roll-mode acquisitions. Triggers only on valid trigger events. Sets holdoff to a specific time. Sets holdoff to a percent of the record duration. Sets holdoff to the minimum value.

Mode & Holdoff

Auto (untriggered roll) Normal Holdoff (time) Holdoff (% of record) Set to Min

Key Points
Displaying the Trigger Source. You do not have to display a channel to

use it as a trigger source.

356

TDS3000 Series User Manual

Trigger Controls

Normal and Auto Mode. Use Normal trigger mode when you want to

trigger only on a valid event. Use Auto trigger mode when you want the acquisition to occur, even when there is not a valid trigger event. Also choose Auto when you want a rolling waveform, with no trigger, at the slower time base settings. See page 332 for more information about roll mode.
External Trigger. Two-channel oscilloscopes have an external trigger input. The trigger level range for the EXT setting is 0.8 V to +0.8 V. The trigger level range for the EXT/10 setting is 8 V to +8 V.

For best external trigger performance, apply a square-wave signal with amplitude greater than the specified minimum level range and with well-defined transitions. Four-channel oscilloscopes do not have a separate external trigger input; however, you can use any of the four channels as your trigger source.

TDS3000 Series User Manual

357

Trigger Controls

Holdoff. You can use holdoff to help stabilize the display of complex

waveforms. After you press the Mode & Holdoff screen button, use the general purpose knob to set the holdoff time as an absolute value or as a percent of the record duration. Holdoff begins when the oscilloscope recognizes a trigger event and disables the trigger system until acquisition is complete. The trigger system remains disabled during the holdoff time.
Acquisition interval Acquisition interval

Trigger level Indicates trigger points

Holdoff

Holdoff

Holdoff

New triggers are not recognized during the holdoff time.

NOTE. For best results, choose Normal trigger mode when using long holdoff settings (10 ms or greater).

358

TDS3000 Series User Manual

Trigger Controls

Video Trigger Choose video triggering to trigger on the odd fields, even fields, or on all the lines of an NTSC, PAL, or SECAM video signal. See the manual for the TDS3VID Extended Video application module for more information if that module is installed.
Bottom Type Video Standard 525/NTSC 625/PAL SECAM Source Trigger On Odd Even All Fields All Lines Mode & Holdoff Triggers on a NTSC signal. Triggers on a PAL signal. Triggers on a SECAM signal. See page 355 for descriptions of these menu items. Triggers on odd or even fields in an iinterlaced signal. terl ce i l Triggers on any field in an interlaced or noninterlaced signal. Triggers on all lines. See page 356 for descriptions of these menu items. Side Description

Key Points
Displaying the Trigger Source. You do not have to display a channel to

use it as a trigger source.


Sync Pulses. When you choose Video, the trigger always occurs on

negative-going sync pulses. If your video signal has positive-going sync pulses, invert the signal using the Vertical menu. See page 371 for information about inverting a signal.

TDS3000 Series User Manual

359

Utility

Utility
The following are examples of what you can do with each of the six branches in the Utility menu: H Use Config to select a language or set the time and date. H Use Apps if an installed application module places items in this menu. See the documentation provided with your application packages for more information. H Use I/O to set up the communication ports. H Use Hard Copy to set up hard copy parameters. Refer to Hard Copy on page 321 for information about setting up and printing a hard copy. H Use Cal to compensate the signal path. H Use Diags to run internal diagnostic routines. Push the UTILITY button to show the utility menu. Then, push the System screen button to choose the branch. The remaining items in the utility menu change depending on the branch you choose.

360

TDS3000 Series User Manual

Utility

Configure System Use the System Config branch to access these functions.
UTILITY

Bottom System Config Language

Side

Description

English Franais Deutsch Italiano Espaol Portugus (Russian) (Japanese) (Korean) (Simplified Chinese) (Traditional Chinese)

Use to choose your native language. Most o cree text appears in the o t on-screen te t e r i t e l language you choose. e o c oo e

Set Date & Time

Display Date/ Time Hour Min Month Day Year OK Enter Date/ Time

Use to turn the date/time display On or Off. Use to set the internal clock with the current hour and minute. Use to set the internal clock with the current month and day. Use to set the internal clock with the current year. Sets the date and time of the internal clock.

TDS3000 Series User Manual

361

Utility

Bottom Battery TimeOuts

Side Power Off TimeOut Backlight TimeOut

Description Use to set the time before an automatic shut down. Use to set the time before the backlight automatically turns off. Erases all nonvolatile waveform and setup memory. Use to see the firmware version.

Tek Secure Erase Memory Version

Key Points
Setting the Date and Time. To set the internal clock with the current

date and time, push the Set Date & Time screen button. Use the general purpose knob to set the values after you push the screen buttons for the year, day and month, hour and minute. Push the OK Enter Date/Time screen button to complete the operation.
Power Off Time-Out. Use this feature to automatically turn the

oscilloscope to standby if it is not being used. Use the general purpose knob to set the power off time-out delay to a fixed time or to (time-out off). Cycle the power switch to turn the oscilloscope back on after an automatic shut down. Power off time-out operates only when you use battery power.

362

TDS3000 Series User Manual

Utility

Backlight Time-Out. Press this button to adjust the backlight time-out delay. This feature automatically turns the backlight off after a period of time if the oscilloscope is not being used. Use the general purpose knob to set the backlight time-out delay to a fixed time or to (time-out off). Push any button to turn the backlight back on after an automatic time-out.

Backlight time-out operates only when you use battery power.


Tek Secure. If you have acquired confidential data on your oscillo-

scope, you may want to execute the Tek Secure function before you return the oscilloscope to general use. The Tek Secure function does the following tasks: H Replaces all waveforms in all reference memories with null sample values H Replaces the current front-panel setup and all stored setups with the factory setup H Calculates the checksums of all waveform memory and setup memory locations to verify successful completion of waveform and setup erasure H Displays a confirmation or warning message if the checksum calculation is successful or unsuccessful After you execute the Tek Secure function, you must turn off the oscilloscope power and then turn the power back on to complete the process.

TDS3000 Series User Manual

363

Utility

I/O System Use the System I/O branch to access these functions.
UTILITY

Bottom System I/O GPIB (TDS3GM only)

Side

Description

Talk/Listen Address Hard Copy (Talk Only) Off Bus Debug

Sets the GPIB address. Sets the GPIB port to talk only for making hard copies. Disables the GPIB port. Enables and disables a message window to help you debug GPIB problems. Sets the baud rate in steps from 1200 to 38400. Use to enable hard flagging (RTS/ CTS) or turn flagging off. Selects end-of-line terminator. Enables and disables a message window to help you debug RS-232 problems. Sets baud rate = 9600, hard flagging = on, and EOL = LF.

RS-232 (all communicatio mo le tions modules)

Baud Rate Flagging EOL Debug

Set RS-232 Parameters to Default Values

364

TDS3000 Series User Manual

Utility

Bottom Ethernet Network Settings (TDS3EM only)

Side Change Instrument Settings

Description Displays a list of fields in which you set oscilloscope Ethernet parameters such as address, instrument name, domain name, and so on. Refer to the TDS3000 Digital Phosphor Oscilloor scope Programmer Manual for procedures to set these fields. Refer to page 345 for an explanation of the field editing menu items. Enables and disables a message window to help you debug Ethernet problems. Adds, renames, or deletes an Ethernet network printer from the oscilloscope printer list. Refer to t e TDS3000 ri ter li t e er the Digital Phosphor Oscilloscope Programmer Manual for procedures to set these iel t e e fields. Refer to page 345 for an e er e or explanation of the field editing menu items. Enables or disables displaying a confirmation message before deleting a printer from the oscilloscope printer list.

DHCP/BOOTP

Debug

Ethernet Printer Settings (TDS3EM only) S o l

Add Printer Rename Printer Delete Printer Confirm Delete

TDS3000 Series User Manual

365

Utility

Key Points
RS-232 Troubleshooting. If you are having difficulties with RS-232 communication, try the following remedies:

H Verify that you are using the correct RS-232 cable and adapters. Most computers require a null-modem connection to the oscilloscope. Most printers require a straight-through connection to the oscilloscope. H Verify that the RS-232 cable is connected to the correct port on your computer or hard copy device. H Reset the RS-232 parameters to defaults and then set the baud rate to match the computer or hard copy device. The default settings (except baud rate) are standard on most computers and hard copy devices. H Enable the debug window to see the RS-232 status, errors, data transmitted, and data received. Refer to the TDS3000-Series Programmer Manual for more information about using the RS-232 port.
GPIB Guidelines. Follow these guidelines when you connect your

oscilloscope to a GPIB network: H Turn off the oscilloscope and all external devices before connecting the oscilloscope to the GPIB network. H Assign a unique device address to the oscilloscope. Two devices cannot share the same device address. H Turn on at least two-thirds of the GPIB devices while using the network. Refer to the TDS3000-Series Programmer Manual for more information about using the GPIB port.
Ethernet Guidelines. Refer to the TDS3000-Series Programmer

Manual for Ethernet setup and troubleshooting guidelines.

366

TDS3000 Series User Manual

Utility

Calibration System Use the System Cal branch to access these functions.
UTILITY

Bottom System Cal Signal Path Factory Cal Cal Due Control

Side

Description

Compensates the signal paths to obtain best measurement accuracy. Used to calibrate the oscilloscope. This is a service function only. Notify After Hours Sets the number of hours of operation of Operation before notifying you that a calibration is due. Notify After Years of Elapsed Time Sets the number of years before notifying you that a calibration is due.

Key Points
Signal Path Compensation. For maximum accuracy at any time, run the signal path compensation routine just before taking critical measurements. To meet accuracy specifications, run the routine if the ambient temperature changes by 10 C or more.

Before running the routine, disconnect any probes or cables from the channel inputs. Then, push the Signal Path and OK Compensate Signal Path screen buttons to confirm that you are ready to proceed. The routine takes several minutes to complete.

TDS3000 Series User Manual

367

Utility

Factory Cal. Service personnel use these functions to calibrate the

oscilloscope internal voltage references using external sources. Refer to your Tektronix field office or representative for assistance with these processes.
Cal Due Control. The calibration due notification occurs only in the power-on screen. Set the controls to if you do not want to be notified when calibration is due.

Diagnostic System Use the System Diag branch to access these functions.
UTILITY

Bottom System Diags Execute Loop

Side

Description

Starts diagnostics. Once Always Until Fail Executes the diagnostic loop once. Executes the diagnostic loop continuously. Executes until a failure occurs. Use to see the previous error log page. Use to see the next error log page.

Error Log

Page Up Page Down

Key Points
Starting Diagnostics. To execute the built-in diagnostic routines,

disconnect all cables or probes from the oscilloscope inputs, and then push the OK Run Test screen button.

368

TDS3000 Series User Manual

Utility

Stopping Diagnostics. Choose how you want the diagnostic routines to

execute: H Loop Once runs all diagnostic routines one time and then stops. H Loop Always runs the diagnostic routines continuously. Push the RUN/STOP and then the MENU OFF buttons to resume normal operation. H Loop Until Fail runs the diagnostic routines until the oscilloscope fails a test or until you cycle the power.
Error Log. The error log contains summary data gathered over the life of the oscilloscope. The error log lists the last 100 errors encountered. The last error in the list is the most recent.

Under normal circumstances, the error log should be empty. Any entry in the error log may indicate a hardware failure or firmware fault. If you can repeatedly cause an entry to be added to the error log, please contact a Tektronix service representative for assistance.

TDS3000 Series User Manual

369

Vertical Controls

Vertical Controls
You can use the vertical controls to select waveforms, adjust the waveform vertical position and scale, and set input parameters. All vertical operations affect the selected waveform. Push a channel button (CH 1, CH 2, CH 3, or CH 4), the MATH button, or the REF button to select a waveform. Vertical Position Control Use the vertical POSITION control to locate the selected waveform in the display. When you change the vertical position, a horizontal line temporarily appears to show you the ground reference level on the screen. After the line disappears, the ground reference level is marked at the left of the graticule. If acquisition is stopped you can still reposition waveforms for analysis. The new position setting is used when acquisition resumes. Waveform Off Push the waveform OFF button to remove the selected waveform from the display. You can still use the channel as a trigger source. Vertical Scale Control Use the vertical SCALE control to set the vertical scale factor of the selected waveform in 1-2-5 increments. If acquisition is stopped you can still rescale waveforms for analysis. The new scale setting is used when acquisition resumes. You can also make fine adjustments to the vertical scale. See Channel Buttons on page 371 for more information.

370

TDS3000 Series User Manual

Vertical Controls

Vertical Menu Push the vertical MENU button to show the vertical menu of the selected waveform. See these pages for more information about the specific vertical menus: H Channel Buttons below H Math Button on page 374 H Ref Button on page 376 Channel Buttons Push a channel button (CH 1, CH 2, CH 3, or CH 4) to select a channel. Each channel button also displays the channel if it is not already displayed. Push the vertical MENU button to show the vertical menu of the selected channel. All vertical operations below affect only the selected waveform.
Bottom Coupling Side DC AC GND Description Sets input coupling to DC. Sets input coupling to AC. Provides a 0 V signal reference. The preamplifier disconnects from the input BNC. Input termination remains connected to the input BNC. Sets input resistance to 50 W or 1 MW, for DC or AC coupling. Use for normal operation. Inverts the polarity of the signal in the preamplifier.

W Invert Invert Off Invert On

TDS3000 Series User Manual

371

Vertical Controls

Bottom Bandwidth

Side Full Bandwidth 150 MHz 20 MHz

Description Sets bandwidth to the full oscilloscope bandwidth. Sets the bandwidth to 150 MHz (not available on some models). Sets the bandwidth to 20 MHz. Enables fine scale adjustment with the general purpose knob. Enables numerical vertical position adjustment. Sets vertical position to center screen. Enables vertical offset adjustment with the general purpose knob. Sets vertical offset to 0 V. Use to set probe gain or attenuation or ro e that o ot ve the for probes t t do not have t e TekProbe II interface. Use to adjust the time skew correction for each probe. Use to set probe skew correction to zero.

Fine Scale Position

Fine Scale Vertical Position Set to 0 divs

Offset

Vertical Offset Set to 0 V

Probe Setup

Voltage Probe Current Probe Deskew Set to 0

Key Points
Using Probes with the TekProbe II Interface. When you attach a probe

with the TekProbe II interface, the oscilloscope sets the channel sensitivity, coupling, and termination resistance automatically to match the probe requirements.

372

TDS3000 Series User Manual

Vertical Controls

Vertical Preview. When you change the vertical POSITION or SCALE

controls while the acquisition is stopped or it is waiting for the next trigger, the oscilloscope rescales and repositions the selected waveform in response to the new vertical control settings. You may see a clipped waveform if the original acquisition went off the screen. The oscilloscope then uses the new settings for the next acquisition. In contrast to horizontal preview, the math waveform, cursors, and automatic measurements remain active and valid when using vertical preview.
The Difference Between Vertical Position and Offset. Vertical position is a display function. Adjust the vertical position to place the waveforms where you want to see them. The waveform baseline locations track adjustments made to their positions.

When you adjust vertical offset you see a similar effect, but it is actually quite different. Vertical offset is applied before the oscilloscope preamplifier and can be used to increase the effective dynamic range of the inputs. For example, you can use vertical offset to look at small variations in a large DC voltage. Set the vertical offset to match the nominal DC voltage and the signal appears in the center of the screen.

0V 1 V 2 V 3 V 4 V 5 V 6 V Offset = 5.2 V

TDS3000 Series User Manual

373

Vertical Controls

50 W Protection. If you select the 50 W termination resistance, the

maximum vertical scale factor is limited to 1 V/div. If you apply excessive input voltage, the oscilloscope automatically switches to 1 MW termination resistance to protect the internal 50 W termination. Math Button Push the MATH button to define the math waveform using the math menu. Also push the MATH button to display or select the math waveform. See the TDS3FFT FFT Application Module User Manual if you have that application module installed.
Bottom Dual Wfm Math Side Set 1st Source to Set Operator to Description Selects the first source waveform. Selects the math operator: +, , or B ,

Set 2nd Source to Selects the second source waveform.

Key Points
Dual-Waveform Math. For the dual-waveform math operations, the two

source waveforms interact with the math operators in the sequence shown below.
Operation + B Expression of math waveform Source 1 + Source 2 Source 1 Source 2 Source 1 Source 2 Source 1 B Source 2

374

TDS3000 Series User Manual

Vertical Controls

Scaling and Positioning the Math Waveform To position or scale a math

waveform, select the math waveform and then adjust it with a vertical POSITION or SCALE control. You can do this whether acquisition is running or is stopped.
Math Interaction with Preview. If you select a channel waveform and then adjust the vertical POSITION or SCALE control while the acquisition is stopped, the math waveform remains fixed. It does not track the changes you see to the channel waveform. The same is true if you adjust the horizontal POSITION or SCALE control under these conditions. Gray-Scale Restriction. Math waveforms are always based on the most

current acquisitions and do not contain any gray-scale information.


Source Waveform Screen Position. When displaying dual-waveform

math waveforms, make sure that the source waveforms do not extend beyond the top or bottom edge of the screen. If part of the source waveform is positioned outside the screen, the math waveform may not display correctly.

TDS3000 Series User Manual

375

Vertical Controls

Ref Button Push the REF button to show the reference menu. Push one of the submenus to display a reference waveform or make it the selected reference waveform.
Bottom Ref 1 Side Description

Save Ch1 to Ref1 Saves channel 1 to Ref 1. Save Ch2 to Ref1 Saves channel 2 to Ref 1. Save Ch3 to Ref1 Saves channel 3 to Ref 1. Save Ch4 to Ref1 Saves channel 4 to Ref 1. Save Math to Ref1 Saves the math waveform to Ref 1.

Ref 2 Ref 3 Ref 4

Identical settings for the reference waveforms Ref 2, Ref 3, and Ref 4.

Key Points
Selecting and Displaying Reference Waveforms. You can display all four

reference waveforms at the same time. Push the submenu button to select a particular reference waveform. The selected waveform appears brighter than other displayed reference waveforms.
Removing Reference Waveforms from the Display. To remove a reference

waveform from the display, select that reference waveform and then push the waveform OFF button .

376

TDS3000 Series User Manual

Vertical Controls

Scaling and Positioning a Reference Waveform You can position and

scale a reference waveform independenty from all other displayed waveforms. Select the reference waveform and then adjust it with a vertical or horizontal POSITION or SCALE control. You can do this whether acquisition is running or is stopped. If a reference waveform is selected, scaling and repositioning of the reference waveform operates the same way whether zoom is turned on or off.
Gray-Scale Restriction. Reference waveforms are always saved from

the most current acquisition and do not contain any gray-scale information.

TDS3000 Series User Manual

377

Vertical Controls

378

TDS3000 Series User Manual

Appendices

Appendix A: Specifications
This appendix contains specifications for the TDS3000 Series oscilloscopes. All specifications are guaranteed unless noted as typical. Typical specifications are provided for your convenience but are not guaranteed. Specifications that are marked with the n symbol are checked in Appendix E: Performance Verification. All specifications apply to all TDS3000 models unless noted otherwise. To meet specifications, two conditions must first be met: H The oscilloscope must have been operating continuously for ten minutes within the operating temperature range specified. H You must perform the Compensate Signal Path operation described on page 14. If the operating temperature changes by more than 10 C, you must perform the Compensate Signal Path operation again. Specifications
Acquisition Acquisition modes Single Sequence Sample (Normal), Peak detect, Envelope, and Average Acquisition mode Sample, Peak Detect Average, Envelope Acquisition stops after One acquisition, all channels simultaneously N acquisitions, all channels simultaneously, N is settable from 2 to 256 (or for Envelope)

TDS3000 Series User Manual

A1

Appendix A: Specifications

Specifications (Cont.)
Inputs Input coupling DC, AC, or GND Channel input remains terminated when using GND coupling. Input impedance, DC coupled Maximum voltage at input BNC (1 MW) i t NC W) 1 MW 1% in parallel with 13 pF 2 pF, TekProbe compatible 50 W 1%; VSWR 1.5:1 from DC to 500 MHz, typical Overvoltage category CAT I Environment (refer to page A12) CAT II Environment (refer to page A12) Maximum voltage 150 VRMS (400 Vpk) 100 VRMS (400 Vpk)

For steady-state sinusoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 Vpk at 3 MHz and above. Maximum voltage at input BNC (50 W) Maximum floating voltage 5 VRMS with peaks 30 V 0 V from chassis (BNC) ground to earth ground, or 30 VRMS (42 Vpk) only under these conditions: no signal voltages >30 VRMS (>42 Vpk), all common leads connected to the same voltage, no grounded peripherals attached Measured on one channel, with test signal applied to another channel, and with the same scale and coupling settings on each channel Frequency range 100 MHz 300 MHz 500 MHz Differential delay, typical TDS3012 TDS3014 100:1 TDS3032 TDS3034 100:1 50:1 TDS3052 TDS3054 100:1 50:1 30:1

Channel-to-channel crosstalk, typical

100 ps between any two channels with the same scale and coupling settings

A2

TDS3000 Series User Manual

Appendix A: Specifications

Specifications (Cont.)
Vertical Number of channels TDS3012, TDS3032, TDS3052 TDS3014, TDS3034, TDS3054 2 plus external trigger input Digitizers SCALE Range (at BNC) t NC Fine SCALE Polarity Position range n Analog bandwidth, 50 W l o typical t W (also t ic l at 1 MW with standard probe) 4

9-bit resolution, separate digitizers for each channel sample simultaneously 1 MW 1 mV/div to 10 V/div Adjustable with 1% resolution Normal and Invert 5 divisions Bandwidth limit set to Full, operating ambient 30 C, derate 1%/C above 30 C Scale range 5 mV/div to 1 V/div 2 mV/div to 4.98 mV/div 1 mV/div to 1.99 mV/div TDS3012 TDS3014 100 MHz 100 MHz 90 MHz 3.5 ns TDS3032 TDS3034 300 MHz 250 MHz 150 MHz 1.2 ns TDS3052 TDS3054 500 MHz 300 MHz 175 MHz 0.7 ns 50 W 1mV/div to 1 V/div

Calculated rise time, typical

Analog bandwidth limit, Selectable between 20 MHz, 150 MHz (not available on typical TDS3012 or TDS3014), or Full Lower frequency limit, AC coupled, typical 7 Hz for 1 MW, reduced by a factor of ten when using a 10X passive probe; 140 kHz for 50 W

TDS3000 Series User Manual

A3

Appendix A: Specifications

Specifications (Cont.)
Vertical Peak detect or Envelope pulse reo e t ic sponse, typicall Minimum width of pulse with amplitude of 2 div to capture 50% or greater amplitude Sample rates 125 MS/s 1 ns DC gain accuracy DC measurement accuracy Sample acquisition mode, typical Sample rates 250 MS/s 1/sample rate

2%, derated at 0.07%/C for temperatures below +18 C and above +28 C, in Sample or Average acquisition mode Measurement type Absolute measurement of any waveform point, and High, Low, Max, and Min measurements. Delta voltage between any two points on a waveform, and all other automatic measurements. DC Accuracy (in volts) [0.021  | reading (offset position) | + offset accuracy + 0.15 div + 0.6 mV] [0.021  | reading | + 0.15 div + 1.2 mV]

n Average acquisition mode (16 averages)

Absolute measurement of any waveform point, and High, Low, Max, and Min measurements. Delta voltage between two points on a waveform, and all other automatic measurements.

[0.021  | reading (offset position) | + offset accuracy + 0.1 div] [0.021  | reading | + 0.05 div]

Offset range

Scale range 1 mV/div to 9.95 mV/div 10 mV/div to 99.5 mV/div 100 mV/div to 995 mV/div 1 V/div to 10 V/div

Offset range 100 mV 1 V 10 V 100 V

Offset accuracy

(0.005  | offset position | + 0.1 div)

0.02 term (gain component) derated at 0.00025/C above 30 C

A4

TDS3000 Series User Manual

Appendix A: Specifications

Specifications (Cont.)
Horizontal Acquisition (horizontal) resolution Record length Acquisition rate, maximum Sample rate range Normal 10,000 points Up to 450 waveforms/s Acquisition resolution Normal Fast trigger Seconds/division range n Sample rate and delay time accuracy Trigger External trigger input, typical External trigger maximum voltage m im m volt e 1 MW in parallel with 17 pF, TekProbe compatible (TDS3012, TDS3032, TDS3052 only) Overvoltage category CAT I Environment (refer to page A12) CAT II Environment (refer to page A12) Maximum voltage 150 VRMS (400 Vpk) 100 VRMS (400 Vpk) TDS3012 TDS3014 100 S/s to 1 GS/s 5 S/s to 1.25 GS/s 4 ns/div to 10 s/div Fast trigger 500 points Up to 3,000 waveforms/s TDS3032 TDS3034 100 S/s to 2.5 GS/s 5 S/s to 2.5 GS/s 2 ns/div to 10 s/div TDS3052 TDS3054 100 S/s to 5 GS/s 5 S/s to 5 GS/s 1 ns/div to 10 s/div

200 ppm over any 1 ms time interval

For steady-state sinusoidal waveforms, derate at 20 dB/decade above 200 kHz to 13 Vpk at 3 MHz and above. External trigger maximum floating voltage 0 V from chassis (BNC) ground to earth ground, or 30 VRMS (42 Vpk) only under these conditions: no signal voltages >30 VRMS (>42 Vpk), all common leads connected to the same voltage, no grounded peripherals attached

TDS3000 Series User Manual

A5

Appendix A: Specifications

Specifications (Cont.)
Trigger n Edge trigger e itivit sensitivity Source Any channel, DC coupled Sensitivity 0.35 div from DC to 50 MHz, increasing to 1 div at oscilloscope bandwidth 100 mV from DC to 50 MHz, increasing to 500 mV at 300 MHz 500 mV from DC to 50 MHz, increasing to 3 V at 300 MHz 3.5 times the DC-coupled limits 1.5 times the DC-coupled limit from DC to 30 kHz, attenuates signals above 30 kHz 1.5 times the DC-coupled limits for frequencies above 80 kHz, attenuates signals below 80 kHz Sensitivity 8 divisions from center of screen, 8 divisions from 0 V if LF REJ trigger coupled 800 mV 8 V Fixed at the midlevel of the AC line

Edge trigger sensitivity, typical

External trigger

External/10 trigger Any channel, NOISE REJ coupled Any channel, HF REJ coupled

Any channel, LF REJ coupled

Trigger level range

Source Any channel

External external External/10 trigger Line SET LEVEL TO 50%, typical

Operates with input signals 45 Hz

A6

TDS3000 Series User Manual

Appendix A: Specifications

Specifications (Cont.)
Trigger Trigger level accuracy, t ic l typical Source Any channel External trigger External/10 trigger Line Trigger holdoff range 250.8 ns to 10 s Sensitivity 0.2 divisions 20 mV 200 mV N/A

Video trigger sensitivity, Triggers on negative sync of NTSC, PAL, or SECAM signal typical t ic l Source Sensitivity Any channel External trigger External/10 trigger B Trigger Range Minimum time between arm and trigger, typical Minimum Pulse Width, typical Maximum Frequency, typical Trigger After Time 13.2 ns to 50 s 5ns from the end of the time period and the B trigger event 0.6 to 2.5 divisions of video sync tip 150 mV to 625 mV of video sync tip 1.5 V to 6.25 V of video sync tip Trigger After B Events 1 event to 9,999,999 events 5 ns between the A trigger event and the first B trigger event B event width, 2 ns B event frequency, 250 MHz

TDS3000 Series User Manual

A7

Appendix A: Specifications

Specifications (Cont.)
Display Display screen Display resolution Backlight intensity, typical Display color External display filter I/O ports Parallel printer port GPIB interface RS-232 interface VGA signal output Probe compensator output, typical Miscellaneous Nonvolatile memory Floppy disk Internal clock Radiated immunity Typical retention time 5 years for front-panel settings, unlimited for saved waveforms and setups 3.5 in, DOS format, 720 KB or 1.44 MB compatible Provides date/time stamp for stored data and the current time and date to the front panel, if enabled. Year-2000 compliant. Per methods of EN50082-1 and EN61000-4-3, the increase in trace noise is not to exceed four major divisions peak-to-peak. Ambient RF fields may induce triggering when trigger threshold is offset less than two major divisions from ground reference. Centronics compatible, DB-25 female connector Available as optional accessory TDS3GM Available as optional accessory TDS3GM or TDS3VM, DB-9 male connector Available as optional accessory TDS3VM, DB-15 female connector, 31.6 kHz sync rate, EIA RS-343A compliant 5.0 V into 1 MW load, frequency = 1 kHz 6.5 in (165 mm) diagonal color liquid crystal 640 horizontal by 480 vertical pixels 200 cd/m2 Up to 16 colors, fixed palette Scratch-resistant tempered glass

A8

TDS3000 Series User Manual

Appendix A: Specifications

Specifications (Cont.)
Power sources AC line power Source voltage Source frequency Power consumption Battery power Operating time, typical Operates the oscilloscope and charges the optional internal battery 90 VRMS to 250 VRMS, continuous range (CAT II) 47 Hz to 440 Hz 75 W maximum Optional accessory TDS3BAT, rechargeable NiCad battery pack 2 hours, depending on operating conditions

Battery charge time, 18 hours in the oscilloscope, 3 hours in the optional external typical charger TDS3CHG Line fuse Environmental Temperature Operating range (no disk installed): +5 C to +50 C Nonoperating range (no disk installed): 20 C to +60 C Typical operating range for floppy disks: +10 C to +50 C Humidity Operating range (no disk installed): 20% to 80% RH below 32 C, derate upper limit to 21% RH at 50 C Nonoperating range (no disk installed): 5% to 90% RH below 41 C, derate upper limit to 30% RH at 60 C Typical operating range for floppy disks: 20% to 80% RH below 32 C, derate upper limit to 21% RH at 50 C Pollution Degree Pollution Degree 2: Typical home or office environment. Internal, not user replaceable

TDS3000 Series User Manual

A9

Appendix A: Specifications

Specifications (Cont.)
Environmental Altitude Operating limit: 3000 m Nonoperating limit: 15,000 m Random vibration Operating: 0.31 gRMS from 5 Hz to 500 Hz, 10 minutes on each axis Nonoperating: 2.46 gRMS from 5 Hz to 500 Hz, 10 minutes on each axis Drop resistance, typical Survives a 152 mm (6 in) drop onto concrete with only cosmetic damage Mechanical Size Height: 176 mm (6.9 in), 229 mm (9.0 in) including handle Width: 375 mm (14.75 in) Depth: 149 mm (5.9 in) Weight Oscilloscope only: 3.2 kg (7.0 lbs) With accessories and carry case: 4.1 kg (9.0 lbs) When packaged for domestic shipment: 5.5 kg (12.0 lbs) Optional battery pack: 2 kg (4.5 lbs)

A10

TDS3000 Series User Manual

Appendix A: Specifications

Specifications (Cont.)
EMC certifications and compliances EMC Compliance: European Union Meets intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities: EN 55011 EN 50082-1 EMC Compliance: Australia/New Zealand Class A Radiated and Conducted Emissions Immunity

Meets the intent of Australian EMC Framework as demonstrated to the following specification: AS/NZS 2064.1/2

EMC Compliance: Russia FCC Compliance: U.S.A.

This product was certified by the GOST ministry of Russia to be in compliance with all applicable EMC regulations. Emissions comply with FCC Code of Federal Regulations 47, Part 15, Subpart B, Class A Limits

TDS3000 Series User Manual

A11

Appendix A: Specifications

Specifications (Cont.)
Safety certifications and compliances EC Declaration of Conformity Low Voltage (TDS3000 Series and P3010) (P3010) Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities: Low Voltage Directive 73/23/EEC as ammended by 93/68/EEC EN 61010-1/A2:1995 Safety requirements for electrical equipment for measurement, control, and laboratory use EN 61010-2-031:1995 Particular requirements for hand-held probe assemblies for electrical measurement and test equipment HD401 S1 Safety requirements for electronic apparatus UL3111-1 Standard for electrical measuring and test equipment CAN/CSA C22.2 No. 1010.1 Safety requirements for electrical equipment for measurement, control and laboratory use EN 61010-2-031:1995 Particular requirements for hand-held probe assemblies for electrical measurement and test equipment UL1244, Third Edition Electronic measuring and test equipment CAN/CSA C22.2 No. 231.1-M89 Test probes Terminals on this product may have different installation category designations. The installation categories are: CAT III Distribution-level mains (usually permanently connected). Equipment at this level is typically in a fixed industrial location Local-level mains (wall sockets). Equipment at this level includes appliances, portable tools, and similar products. Equipment is usually cord-connected Secondary (signal level) or battery operated circuits of electronic equipment

(P6139A) Approvals (TDS3000 Series and P3010) (P3010) (P6139A) Installation Category Descriptions

CAT II

CAT I

A12

TDS3000 Series User Manual

Appendix B: Factory Setup


The table below lists the state of the oscilloscope after you recall the Factory Setup.
Control Acquire horizontal resolution Acquire mode Acquire number of averages Acquire number of envelopes Acquire run/stop Acquire single sequence Channel selection Coarse Confirm delete Cursor function Cursor H Bar 1 position Cursor H Bar 2 position Cursor H Bar units Cursor V Bar 1 position Cursor V Bar 2 position Cursor V Bar units Cursor tracking Changed by factory setup to Normal (10,000 points) Sample 16 16 Run Off Channel 1 on, all others off No change No change Off 3.2 divisions from the center +3.2 divisions from the center Base 10% of record 90% of record Seconds Independent

TDS3000 Series User Manual

B1

Appendix B: Factory Setup

Control Display graticule type Display backlight Display color palette Dual waveform math function Edge trigger coupling Edge trigger level Edge trigger slope Edge trigger source Horizontal delay Horizontal delay time Horizontal trigger position Horizontal time/div Horizontal zoom Horizontal zoom position Horizontal zoom time/div Measure gating Measure high-low setup Measure high reference Measure low reference Measure mid reference

Changed by factory setup to Full High Normal Ch 1 + Ch 2 DC 0.0 V Rising Channel 1 On 0 ns 10% 400 ms/div Off 50% 400 ms/div Off (Full Record) Auto 90% and 0 V 10% and 0 V 50% and 0 V

B2

TDS3000 Series User Manual

Appendix B: Factory Setup

Control Overwrite lock Reference waveforms Saved setups Trigger holdoff Trigger mode Trigger type Utility language Utility date/time display Utility I/O Utility hard copy Vertical bandwidth Vertical coupling Vertical invert Vertical offset Vertical position Vertical probe setup Vertical volts/div Video trigger standard Video trigger on Waveform file format XY display

Changed by factory setup to No change No change No change 250.8 ns Auto Edge No change On No change No change Full DC 1 MW Off 0V 0 div Voltage, 1 X (unless non-1 X probe attached) 100 mV/div 525/NTSC All lines No change Off

TDS3000 Series User Manual

B3

Appendix B: Factory Setup

B4

TDS3000 Series User Manual

Appendix C: Accessories
Standard Accessories
P3010 10X passive probes (TDS3012 and TDS3014) The P3010 10X passive probes have 100 MHz bandwidth and a CAT II voltage rating of 300 VRMS.

P6139A 10X passive probes (TDS3032, TDS3034, TDS3052 and TDS3054) The P6139A 10X passive probes have 300 MHz or 500 MHz bandwidth and a CAT II voltage rating of 300 VRMS.

Front cover The front cover (200-4416-00) snaps onto the front of the oscilloscope to protect it during transit. The front cover contains a convenient location to store the Reference manual.

Accessory tray The accessory tray (436-0371-00) fits in the battery compartment when you do not have a battery installed. You can use the tray to store probes and other accessories.

Manuals The oscilloscope includes a User Manual, Reference, and Online Tour Disk (020-2263-XX). Insert the disk into the oscilloscope disk drive to see an online tour of the product.

TDS3000 Series User Manual

C1

Appendix C: Accessories

Optional Accessories
TDS3FFT FFT application package (standard with TDS3014, TDS3034, and TDS3054) The FFT application package adds FFT analysis and measurement capability to your oscilloscope. Application packages can be installed by the user. TDS3TRG Advanced Triggering application package (standard with TDS3014, TDS3034, and TDS3054) The advanced triggering application package adds pulse-width, slew-rate, pattern, state, runt, and edge-delay triggering capability to your oscilloscope. Application packages can be installed by the user. TDS3VID Extended Video application package The extended video application package adds video trigger, display, and measurement capabilities to your oscilloscope. Application packages can be installed by the user. TDS3TMT Telecom Mask Test application package The telecom mask test application package adds ITUT G.703, ANSI T1.102 (up to DS3 data rates), and custom mask testing capabilities to your oscilloscope. Application packages can be installed by the user. TDS3GM GPIB/RS-232 communication module This communication module adds GPIB and RS-232 ports to your oscilloscope. You can attach a printer to these ports or use them for remote programmability. Communication modules can be installed by the user and include the TDS3000-Series Programmer Manual (in English only).

C2

TDS3000 Series User Manual

Appendix C: Accessories

Optional Accessories (Cont.)


TDS3VM VGA/RS-232 communication module This communication module adds VGA and RS-232 ports to your oscilloscope. You can attach a printer to the RS-232 port or use it for remote programmability. You can attach a monitor to the VGA port to enhance viewing the screen from a distance. Communication modules can be installed by the user and include the TDS3000-Series Programmer Manual (in English only). TDS3EM Ethernet communication module This communication module adds Ethernet 10baseT and RS-232 ports to your oscilloscope. You can attach the oscilloscope to an Ethernet network for remote printing. Communication modules can be installed by the user and include the TDS3000-Series Programmer Manual (in English only). TDS3CHG external battery charger The battery charger recharges the oscilloscope battery pack in approximately 3 hours.

TDS3BAT rechargeable battery pack An extra rechargeable battery pack provides a high-capacity spare battery for extended portable operation.

TDS3000 Series User Manual

C3

Appendix C: Accessories

Optional Accessories (Cont.)


AC3000 soft case The soft case protects the oscilloscope when not in use. The soft case provides compartments for probes, one spare battery, battery charger, and the User Manual.

RM3000 rack mount kit The rack mount kit contains all the hardware necesary to mount your oscilloscope in a standard rack. The kit requires 7 in of vertical space in the rack.

WSTRO WaveStar Software for Oscilloscopes WaveStar is a Microsoft Office 97-compatible application that connects your oscilloscope to your PC. You can upload and download waveforms and setups. You can drag and drop acquired data into spreadsheets for further analysis, into word processors to integrate with your documentation, or to printers and plotters to make hard copies. Manuals Service Manual (071-0382-XX) provides information about maintenence and module-level repair.

C4

TDS3000 Series User Manual

Appendix D: Probe Basics


This chapter contains basic information about the P3010 or P6139A probes provided with your TDS3000 Series oscilloscope. It also contains information about other probes you can use with your oscilloscope and their limitations.

Probe Descriptions
The P3010 and P6139A are high-impedance passive probes with the following general characteristics.
Characteristic Cable length Compatibility Bandwidth Attenuation Nominal input impedance Maximum working voltage P3010 2m 100 MHz oscilloscope models 100 MHz 10X 10 MW in parallel with 13.3 pF 300 V, CAT II, derated at 20 dB/decade to 50 V above 2.5 MHz P6139A 1.3 m 300 MHz and 500 MHz oscilloscope models 300 MHz or 500 MHz 10X 10 MW in parallel with 8 pF 300 V, CAT II, derated at 20 dB/decade to 50 V above 2.5 MHz

TDS3000 Series User Manual

D1

Appendix D: Probe Basics

Probe Compensation
You should compensate a probe to an oscilloscope input whenever you attach a probe for the first time to any input channel. See page 13 for instructions to compensate probes. When compensating the P3010, only adjust the trimmer marked L.

TekProbe Interface
Probes with the TekProbe interface automatically communicate with the oscilloscope to set the probe type and attenuation factor. If you use a probe without the TekProbe interface, you can set these parameters in the Vertical menu for the channel that the probe is connected to.

D2

TDS3000 Series User Manual

Appendix D: Probe Basics

Probe Guard
A guard around the probe body provides a finger barrier for protection from electric shock.

Guard

WARNING. To avoid electric shock when using the probe, keep fingers behind the guard on the probe body. To avoid electric shock when using the probe, do not touch metallic portions of the probe head while connected to a voltage source.

Ground Leads
Always use a ground lead when you probe a circuit to minimize noise pickup and signal aberrations. Connecting the ground lead to a point near the signal source usually provides the best results. Long ground leads can cause false ringing and aberrations in the acquired waveform that are not in the actual signal. To see the best signal fidelity, use the shortest possible ground lead.

Long Ground Lead ShortGroundLead

TDS3000 Series User Manual

D3

Appendix D: Probe Basics

P3010 High-Frequency Compensation


The P3010 high-frequency compensation should seldom require adjustment; however, your probe may require high-frequency adjustment if either of the following are true: H the probe has high-frequency aberrations H the probe does not perform at the rated bandwidth To perform the high-frequency compensation adjustment you will need a signal source that has all of the following characteristics:

tr<1 ns 1 MHz >1 Vp-p 10% 90%

H square-wave output at 1 MHz H fast rise output with rise time less than 1 ns H output properly terminated

D4

TDS3000 Series User Manual

Appendix D: Probe Basics

Connect the P3010 to the signal source to display a 1 MHz test signal on your oscilloscope. Use the BNC-to-probe tip adapter (013-0277-00) to make the connection. (The display should be similar to that shown below).
10 ns H H

(a) Area Of Waveform Affected By Adjustment

(b) Location of Adjustment

Adjust trimmer H until the waveform is flat on top and has a square leading edge.

TDS3000 Series User Manual

D5

Appendix D: Probe Basics

P3010 Replaceable Parts and Accessories

2 1

Standard Accessories

7 8 9

Optional Accessories

D6

TDS3000 Series User Manual

Appendix D: Probe Basics

P3010 replaceable parts and accessories


Index number 1 2 3 4 5 6 7 8 9 Description Retractable hook tip Probe tip Ground lead, 6 in Marker set (four colors, two each) Adjustment tool BNC-to-probe tip adapter Ground lead, 28 in Ground lead, 12 in IC test tip, package of 10 Part number 013-0107-08 131-4997-01 196-3120-01 016-0633-00 003-1433-01 013-0277-00 196-3120-21 196-3121-01 015-0201-07

TDS3000 Series User Manual

D7

Appendix D: Probe Basics

P6139A Replaceable Parts and Accessories

1 4 3 2 14 12 13 10 Standard Accessories 8 11

5 6

16 15

17

18

19

Optional Accessories

D8

TDS3000 Series User Manual

Appendix D: Probe Basics

P6139A replaceable parts and accessories


Index number 1 2 3 4 5 6 7 8 9 01 11 12 13 14 15 16 17 18 19 Description Compensation box assembly BNC connector Cable cover nipple Cable assembly Ground collar Ground lead, 6 in Ground lead, 2.3 in Marker set (four colors, two each) IC Klipchip grabber, package of 20 Adjustment tool Ground lead with clip, 6 in Probe tip assembly Probe tip cover Retractable hook tip Ground lead, 3 in Probe to connector pin adapter 50 W BNC-to-probe tip termination and adapter Ground contact kit (two each of five lengths) with cover shell Grounding probe tip Part number 206-0440-00 131-3219-00 200-3018-00 174-0978-00 343-1003-01 196-3113-02 195-4240-00 016-0633-00 206-0364-00 003-1433-01 196-3305-00 206-0441-00 204-1049-00 013-0107-06 196-3113-03 015-0325-00 013-0227-00 016-1077-00 013-0085-00

TDS3000 Series User Manual

D9

Appendix D: Probe Basics

Using Other Probes


Optional probes can add capabilities to your oscilloscope that are useful in many applications. Passive Probes You can use the following passive probes without any limitations.
Passive probe P6561A P6562A P6563A P5100 P6015A P6021 P6022 Recommended usage SMT probe, 200 MHz, 10X SMT probe, 350 MHz, 10X SMT probe, 500 MHz, 20X High-voltage probe, 2500 Vpk CAT II, 250 MHz, 100X High-voltage probe, 20 kV DC, 75 MHz, 1000X Current probe, 15 A, 120 Hz to 60 MHz Current probe, 6 A, 935 Hz to 120 MHz

D10

TDS3000 Series User Manual

Appendix D: Probe Basics

Supported Active Probes


The oscilloscope supplies power to active probes. You can use the following active probes as long as the total power required by the probes does not exceed the capacity of the oscilloscope. To determine the total probe load, add the load factors for all probes you want to use. The oscilloscope can supply power to this combination if the sum of the load factors equals 10 or less. All passive probes have a load factor of zero.
Active probe P6205 P6243 P5205 P5210 ADA400A AFTDS AMT75 TCP202 013-0278-00 Recommended usage FET probe, 750 MHz, 10X SMT probe, 1 GHz, 10X High-voltage differential probe, 1300 V, 100 MHz, 50X or 500X High-voltage differential probe, 5600 V, 50 MHz, 100X or 1000X Load factor 0 0 6 6

Differential preamplifier, 10mV sensitiv- 5 ity, DC to 10 kHz Telecom differential 50 W adapter Telecom 75 W adapter Current probe, 15 A, DC to 50 MHz Video display clamp 0 0 4 5

CAUTION. To avoid a measurement error, do not connect active probes with a combined load factor that is greater than 10. The signal distortion caused by such an overload can be subtle (reduced gain, dynamic range, or slew rate).

TDS3000 Series User Manual

D11

Appendix D: Probe Basics

Unsupported Probes
The TDS3000 Series only supports the probes listed in this section of the manual. The oscilloscope may not display a message when you connect an unsupported probe, so make sure that any probe you connect to the TDS3000 is supported.

D12

TDS3000 Series User Manual

Appendix E: Performance Verification


This appendix contains performance verification procedures for the specifications marked with the n symbol. The following equipment, or a suitable equivalent, is required to complete these procedures.
Minimum requirements 3 mV to 4 V, 0.1% accuracy

Description DC Voltage Source

Examples Wavetek 9500 Oscilloscope Calibrator with two 9510 O t t it t o 9 Output Modules

Leveled Sine Wave Gen- 50 kHz to 500 MHz, erator 3% amplitude accuracy Time Mark Generator 50 W feedthrough termination 10 ms period, 50 ppm accuracy BNC connectors

Tektronix part number 011-0099-00

You may need additional cables and adapters, depending on the actual test equipment you use. These procedures cover all TDS3000 oscilloscope models. Please disregard checks that do not apply to the specific model you are testing. Photocopy the test record on the next three pages and use them to record the performance test results for your oscilloscope.

TDS3000 Series User Manual

E1

Appendix E: Performance Verification

Test Record
Serial number Procedure performed by Date

Test Self Test

Passed

Failed

Performance checks Channel 1 1 mV/div DC C me reme t measurement 2 mV/div accuracy cc r c 5 mV/div 50 mV/div 50 mV/div 50 mV delta 90 mV/div 200 mV/div 1 V/div

Low limit 99.25 mV 7.540 mV 101.8 mV 982.4 mV 632.4 mV 340.5 mV 339.3 mV 9.900 V 10.30 V

Test result

High limit 100.8 mV 6.460 mV 98.24 mV 1.018 V 667.6 mV 359.5 mV 290.7 mV 10.10 V 9.698 V

E2

TDS3000 Series User Manual

Appendix E: Performance Verification

Performance checks Channel 2 1 mV/div DC C me reme t measurement 2 mV/div accuracy cc r c 5 mV/div 50 mV/div 50 mV/div 50 mV delta 90 mV/div 200 mV/div 1 V/div Channel 3 1 mV/div DC C me reme t measurement 2 mV/div accuracy cc r c 5 mV/div 50 mV/div 50 mV/div 50 mV delta 90 mV/div 200 mV/div 1 V/div

Low limit 99.25 mV 7.540 mV 101.8 mV 982.4 mV 632.4 mV 340.5 mV 339.3 mV 9.900 V 10.30 V 99.25 mV 7.540 mV 101.8 mV 982.4 mV 632.4 mV 340.5 mV 339.3 mV 9.900 V 10.30 V

Test result

High limit 100.8 mV 6.460 mV 98.24 mV 1.018 V 667.6 mV 359.5 mV 290.7 mV 10.10 V 9.698 V 100.8 mV 6.460 mV 98.24 mV 1.018 V 667.6 mV 359.5 mV 290.7 mV 10.10 V 9.698 V

TDS3000 Series User Manual

E3

Appendix E: Performance Verification

Performance checks Channel 4 1 mV/div DC C me reme t measurement 2 mV/div accuracy cc r c 5 mV/div 50 mV/div 50 mV/div 50 mV delta 90 mV/div 200 mV/div 1 V/div Channel 1 bandwidth Channel 2 bandwidth Channel 3 bandwidth Channel 4 bandwidth Channel 1 trigger tri er sensitivity Channel 2 tri er trigger sensitivity Channel 3 tri er trigger sensitivity Channel 4 tri er trigger sensitivity rising slope falling slope rising slope falling slope rising slope falling slope rising slope falling slope

Low limit 99.25 mV 7.540 mV 101.8 mV 982.4 mV 632.4 mV 340.5 mV 339.3 mV 9.900 V 10.30 V 425 mV 425 mV 425 mV 425 mV stable trigger stable trigger stable trigger stable trigger stable trigger stable trigger stable trigger stable trigger 2 divisions

Test result

High limit 100.8 mV 6.460 mV 98.24 mV 1.018 V 667.6 mV 359.5 mV 290.7 mV 10.10 V 9.698 V +2 divisions

Sample rate and delay time accuracy

E4

TDS3000 Series User Manual

Appendix E: Performance Verification

Performance Verification Procedures


Before beginning these procedures, two conditions must first be met: H The oscilloscope must have been operating continuously for ten minutes in an environment the meets the operating range specifications for temperature and humidity. H You must perform the Compensate Signal Path operation described on page 14. If the operating temperature changes by more than 10 C, you must perform the Compensate Signal Path operation again. The time required to complete the entire procedure is approximately one hour. WARNING. Some procedures use hazardous voltages. To prevent electrical shock, always set voltage source outputs to 0 V before making or changing any interconnections.

TDS3000 Series User Manual

E5

Appendix E: Performance Verification

Self Test This procedure uses internal routines to verify that the oscilloscope functions and passes its internal self tests. No test equipment or hookups are required. Start the self test with these steps: 1. Disconnect all probes and cables from the oscilloscope inputs. 2. Push the UTILITY menu button. 3. Push the System screen button to select Diags. 4. Push the Loop screen button and choose Once. 5. Push the Execute screen button. 6. Push the OK Confirm Run Test screen button. A dialog box displays the result when the self test completes. Push the MENU OFF screen button to continue operation.

E6

TDS3000 Series User Manual

Appendix E: Performance Verification

Check DC Voltage Measurement Accuracy This test checks the DC voltage measurement accuracy in the average acquisition mode. 1. Set the DC voltage source output level to 0 V. 2. Connect the DC voltage source to the oscilloscope channel 1 input as shown below.
DC voltage source

TDS3000 Series Oscilloscope

Ch 1

3. Push the SAVE/RECALL menu button. 4. Push the Recall Factory Setup screen button and then push the OK Confirm Factory Init screen button. 5. Push the acquire MENU button. 6. Push the Mode screen button and then push the Average screen button. 7. Adjust the number of averages to 16 with the general purpose knob. 8. Skip to step 11.

TDS3000 Series User Manual

E7

Appendix E: Performance Verification

9. Move the DC voltage source output cable to the oscilloscope channel you want to check. 10. Push the channel button (CH 1, CH 2, CH 3, or CH 4) for the channel you want to check. 11. Push the MEASURE menu button. 12. Push the Select Measurement screen button. 13. Push the more screen button until you can select the Mean measurement. 14. Push the vertical MENU button. 15. For each row of the table on page E-9, do these steps: a. Set the vertical SCALE control to the setting in the table. b. Set the Invert, Bandwidth Limit, and Offset controls to the settings in the table. c. Set the output of the DC voltage source to the voltage level in the table. d. Verify that the oscilloscope Mean measurement is within the limits listed in the table on page E8. e. For the 50 mV delta measurement, subtract the second 50 mV measurement from the first 50 mV measurement; verify that the difference is within the limits stated in the Delta row of the table. 16. Repeat steps 15a through 15e for each row in the table. 17. Push the waveform off button.

18. Repeat steps 9 through 17 for each channel of the oscilloscope (not including the external trigger input on some models).

E8

TDS3000 Series User Manual

Appendix E: Performance Verification

Vertical SCALE setting 1 mV/div 2 mV/div 5 mV/div 50 mV/div 50 mV/div 50 mV delta1 90 mV/div 2 200 mV/div 1 V/div
1 2

Invert setting Off Off Off Off Off

Bandwidth limit setting 20 MHz 20 MHz 20 MHz Full Full

Offset 96.5 mV 0.0 V 82.5 mV 825 mV 825 mV

Input voltage 100 mV 7 mV 100 mV 1.0 V 650 mV

Low limit 99.25 mV 7.540 mV 101.8 mV 982.4 mV 632.4 mV 340.5 mV

High limit 100.8 mV 6.460 mV 98.24 mV 1.018 V 667.6 mV 359.5 mV 290.7 mV 10.10 V 9.698 V

Off Off On 4

Full 150 MHz 3 150 MHz 3

0.0 V 9.3 V 6.5 V

315 mV 10 V 10 V

339.3 mV 9.900 V 10.30 V

Refer to step 15e on page E8 to calculate 50 mV delta measurement. Push the Vertical MENU button, push the Fine Scale screen button, then use the general purpose knob to adjust setting to 90 mV/div. Use the Full bandwidth setting on the TDS3012 or TDS3014 oscilloscopes. Make sure to turn Invert setting to On for this measurement.

3 4

TDS3000 Series User Manual

E9

Appendix E: Performance Verification

Check Bandwidth This test checks the bandwidth for each channel. 1. Connect the output of the leveled sine wave generator to the oscilloscope channel 1 input as shown below.
Leveled sine wave generator

TDS3000 Series Oscilloscope

Ch 1

2. Push the SAVE/RECALL menu button. 3. Push the Recall Factory Setup screen button and then push the OK Confirm Factory Init screen button. 4. Push the acquire MENU button. 5. Push the Mode screen button and then push the Average screen button. 6. Adjust the number of averages to 16 with the general purpose knob. 7. Push the trigger MENU button. 8. Push the Source screen button and then push the Vert screen button. 9. Push the Coupling screen button and then push the Noise Reject screen button. 10. Skip to step 13.

E10

TDS3000 Series User Manual

Appendix E: Performance Verification

11. Move the output cable of the leveled sine wave generator to the oscilloscope channel you want to check. 12. Push the channel button (CH 1, CH 2, CH 3, or CH4) for the channel you want to check. 13. Set the horizontal SCALE to 10 ms/div. 14. Push the vertical MENU button. 15. Push the Coupling screen button and select 50 W input resistance. 16. Push the MEASURE menu button. 17. Push the Select Measurement screen button. 18. Push the more screen button until you can select the PkPk measurement. 19. Set the vertical SCALE to 100 mV/div. 20. Set the output frequency of the leveled sine wave generator to 50 kHz. 21. Set the output amplitude of the leveled sine wave generator so the peak-to-peak measurement is between 599 mV and 601 mV. 22. Set the horizontal SCALE to 10 ns/div. 23. Set the output frequency of the leveled sine wave generator to the frequency shown in the table below.
Oscilloscope model TDS3012, TDS3014 TDS3032, TDS3034 TDS3052, TDS3054 Frequency 100 MHz 300 MHz 500 MHz

TDS3000 Series User Manual

E11

Appendix E: Performance Verification

24. Verify that the peak-to-peak measurement is 425 mV. 25. Push the waveform off button.

26. Repeat steps 11 through 25 for each channel of the oscilloscope (not including the external trigger input on some models).

E12

TDS3000 Series User Manual

Appendix E: Performance Verification

Check Channel Edge-Trigger Sensitivity This test checks the edge-trigger sensitivity for each channel. 1. Connect the output of the leveled sine wave generator to the oscilloscope channel 1 input as shown below.
Leveled sine wave generator

TDS3000 Series Oscilloscope

Ch 1

2. Push the SAVE/RECALL menu button. 3. Push the Recall Factory Setup screen button and then push the OK Confirm Factory Init screen button. 4. Push the acquire MENU button. 5. Push the Mode screen button and then push the Average screen button. 6. Adjust the number of averages to 16 with the general purpose knob. 7. Push the trigger MENU button. 8. Push the Source screen button and then push the Vert screen button. 9. Set the horizontal SCALE to 10 ns/div. 10. Skip to step 13.

TDS3000 Series User Manual

E13

Appendix E: Performance Verification

11. Move the output cable of the leveled sine wave generator to the oscilloscope channel you want to check. 12. Push the channel button (CH 1, CH 2, CH 3, or CH4) for the channel you want to check. 13. Push the vertical MENU button. 14. Push the Coupling screen button and select 50 W input resistance. 15. Push the MEASURE menu button. 16. Push the Select Measurement screen button. 17. Push the more screen button until you can select the PkPk measurement. 18. Set the vertical SCALE to 500 mV/div. 19. Set the output frequency of the leveled sine wave generator to the frequency shown in the table below.
Oscilloscope model TDS3012, TDS3014 TDS3032, TDS3034 TDS3052, TDS3054 Frequency 100 MHz 300 MHz 500 MHz

E14

TDS3000 Series User Manual

Appendix E: Performance Verification

20. Set the output amplitude of the leveled sine wave generator so the peak-to-peak measurement is approximately 500 mV. 21. Push the SET TO 50% button. Adjust the trigger LEVEL as necessary and then verify that triggering is stable. 22. Push the trigger MENU button. 23. Push the Slope screen button and select the \ (falling) slope. 24. Push the SET TO 50% button. Adjust the trigger LEVEL as necessary and then verify that triggering is stable. 25. Push the Slope screen button and select the / (rising) slope. 26. Push the waveform off button.

27. Repeat steps 11 through 26 for each channel of the oscilloscope (not including the external trigger input on some models).

TDS3000 Series User Manual

E15

Appendix E: Performance Verification

Check Sample Rate and Delay Time Accuracy This test checks the time base accuracy. 1. Connect the output of the leveled sine wave generator to the oscilloscope channel 1 input as shown below.
Time mark generator

TDS3000 Series Oscilloscope

Ch 1

2. Push the SAVE/RECALL menu button. 3. Push the Recall Factory Setup screen button and then push the OK Confirm Factory Init screen button. 4. Set the time mark generator period to 10 ms. Use a time mark waveform with a fast rising edge. 5. If adjustable, set the time mark amplitude to approximately 1 Vp-p. 6. Push the DELAY button to turn delay off. 7. Push the vertical MENU button. 8. Push the Coupling screen button and select 50 W input resistance.

E16

TDS3000 Series User Manual

Appendix E: Performance Verification

9. Set the vertical SCALE to 500 mV/div. 10. Set the horizontal SCALE to 2 ms/div. 11. Adjust the vertical POSITION control to center the time mark signal on the screen. 12. Adjust the trigger LEVEL as necessary to obtain a triggered display. 13. Adjust the horizontal POSITION control to move the trigger location to the center of the screen (50%). 14. Push the DELAY button to turn delay on. 15. Turn the horizontal POSITION control counter-clockwise to set the delay to exactly 10 ms. 16. Set the horizontal scale to 1 ms/div. 17. Check that the rising edge of the marker crosses the center horizontal graticule line within 2 divisions of center graticule. NOTE. One division of displacement from graticule center corresponds to a 100 ppm time base error. This completes the performance verification procedure.

TDS3000 Series User Manual

E17

Appendix E: Performance Verification

E18

TDS3000 Series User Manual

Appendix F: General Care and Cleaning


General Care
Protect the oscillscope from adverse weather conditions. The oscilloscope is not water resistant. Do not store or leave the oscilloscope where the LCD display will be exposed to direct sunlight for long periods of time. CAUTION. To avoid damage to the oscilloscope, do not expose it to sprays, liquids, or solvents.

Cleaning
Inspect the oscillosope as often as operating conditions require. To clean the oscilloscope exterior, perform the following steps: 1. Remove loose dust on the outside of the oscilloscope with a lint-free cloth. Use care to avoid scratching the glass display filter. 2. Use a soft cloth or paper towel dampened with water to clean the oscilloscope. You can use a 75% isopropyl alcohol solution for more efficient cleaning. CAUTION. To avoid damage to the surface of the oscilloscope, do not use any abrasive or chemical cleaning agents.

TDS3000 Series User Manual

F1

Appendix F: General Care and Cleaning

F2

TDS3000 Series User Manual

Glossary

Glossary
AC Coupling A mode that blocks the DC component of a signal but passes the dynamic (AC) component of the signal. Useful for observing an AC signal that is normally riding on a DC signal. Acquisition The process of sampling signals from input channels, digitizing the samples, processing the results into data points, and assembling the data points into a waveform record. The waveform record is stored in memory. Active Cursor The cursor that moves when you adjust the general purpose knob. The @ readout on the display shows the position of the active cursor. Aliasing A false representation of a signal due to insufficient sampling of high frequencies or fast transitions. A condition that occurs when an oscilloscope digitizes at an effective sampling rate that is too slow to reproduce the input signal. The waveform displayed on the oscilloscope may have a lower frequency than the actual input signal. Application Module A small module that you plug in to the front panel to add capabilities to your oscilloscope. You can use up to four application modules at one time. Application Package An optional package that may include an application module, a firmware upgrade, and manuals that add capabilities to your oscilloscope.

TDS3000 Series User Manual

Glossary1

Glossary

Attenuation The degree the amplitude of a signal is reduced when it passes through an attenuating device such as a probe or attenuator (the ratio of the input measure to the output measure). For example, a 10X probe attenuates, or reduces, the input voltage of a signal by a factor of 10. Auto Trigger Mode A trigger mode that causes the oscilloscope to automatically acquire if it does not detect a valid trigger event. Autoset A feature that automatically sets the vertical, horizontal, and trigger controls to provide a usable display. Average Acquisition Mode A mode in which the oscilloscope acquires and displays a waveform that is the averaged result of several acquisitions. The oscilloscope acquires data as in the sample mode and then averages it according to a specified number of averages. Averaging reduces uncorrelated noise in the displayed signal. Backlight The illumination behind the liquid-crystal display. B Trigger A second trigger system that works with the main (A) trigger to capture more complex events. You can use the dual A and B triggers to trigger after a given wait time or number of events. Cursors Paired markers that you can use to make measurements between two waveform locations. The oscilloscope displays the values (expressed in volts, time, or frequency) of the position of the active cursor and the distance between the two cursors.

Glossary2

TDS3000 Series User Manual

Glossary

DC Coupling A mode that passes both AC and DC signal components to the circuit. Available for both the trigger system and the vertical system. Communication Module A optional module that adds I/O ports to the oscilloscope. Delay A means to delay the acquisition to start long after the trigger event has occurred. The trigger point does not have to be within the waveform record when delay is on. Digital Phosphor A term that describes the way the digitizing oscilloscope can simulate the acquisition behavior of an analog oscilloscope. Displayed points vary in intensity depending on the frequency of their acquisition and then decay as if the oscilloscope had an analog CRT (cathode ray tube). Digital Real Time Digitizing A digitizing technique that samples the input signal with a sample frequency of four to five times the oscilloscope bandwidth. Combined with (sinx)/x interpolation, all frequency components of the input up to the bandwidth are accurately displayed. Digitizing The process of converting a continuous analog signal such as a waveform to a set of discrete numbers representing the amplitude of the signal at specific points in time.

TDS3000 Series User Manual

Glossary3

Glossary

Edge Trigger Triggering that occurs when the oscilloscope detects the source passing through a specified voltage level in a specified direction (the trigger slope). Envelope Acquisition Mode A mode in which the oscilloscope acquires and displays a waveform that shows the variation extremes of several acquisitions. External Trigger Triggering that occurs when the oscilloscope detects the external input signal passing through a specified voltage level in a specified direction (the trigger slope). Floating Measurements Voltage measurements where the reference voltage is not earth ground. General Purpose Knob A knob on the front panel that you can use to set parameters, for example, cursor position. The specific parameter assigned to the general purpose knob depends on other selections. Gray Scale The waveform display capable of showing points at different intensities, depending on the frequency of their occurrence. Color gray scale consists of light and dark shades of the color.

Glossary4

TDS3000 Series User Manual

Glossary

Ground (GND) Coupling Coupling option that disconnects the input signal from the vertical system. Ground Lead The reference lead for an oscilloscope probe. Ground Wire A wire that must be connected between the ground terminal and earth ground when using the oscilloscope with battery power. Hard Copy An electronic copy of the display in a format useable by a printer or plotter. Holdoff A specified amount of time that must elapse after a trigger signal before the trigger circuit will accept another trigger signal. Holdoff helps ensure a stable display. Horizontal Bar Cursors The two horizontal bars that you position to measure the voltage parameters of a waveform. The oscilloscope displays the value of the active (moveable) cursor with respect to ground and the voltage value between the bars. Menu A set of labels shown in the display to identify the functions of the bezel buttons. The specific menu contents depend on the menu button you press.

TDS3000 Series User Manual

Glossary5

Glossary

Normal Trigger Mode A mode where the oscilloscope does not acquire a waveform record unless a valid trigger event occurs. It waits for a valid trigger event before acquiring waveform data. Peak Detect An acquisition mode that captures spikes and glitches that may occur between normal sample points. Persistence The decay of waveform points. With persistence turned off, the points decay quickly. With persistence on, the points decay more slowly or not at all, depending on the setting. Pixel A visible point on the display. The display is 640 pixels wide by 480 pixels high. Pretrigger The specified portion of the waveform record that contains data acquired before the trigger event. Preview The oscilloscope capability to show you what the next acquisition might look like if you change control settings while acquisition is stopped or waiting for the next trigger. The new control settings take effect for the next acquisition. Both horizontal and vertical preview are available. QuickMenu An alternate display presentation that lets you control the most commonly used functions with the screen buttons. Optional application packages may have an available QuickMenu display. Record Length The specified number of samples in a waveform. Reference Waveform A saved waveform selected for display. You can save and display up to four reference waveforms.

Glossary6

TDS3000 Series User Manual

Glossary

Roll Mode An acquisition mode useful at slow horizontal scale settings. Roll mode allows you to view the waveform as it is acquired point-by-point. The waveform appears to roll across the display. RS-232 The serial communication port used to connect to a hard-copy device, computer, controller, or terminal. Sample Acquisition Mode A mode in which the oscilloscope creates a record point by saving the first sample during each acquisition interval. This is the default mode of the acquisition system. Sample Interval The time interval between successive samples in a time base. For real-time digitizers, the sample interval is the reciprocal of the sample rate. Sampling The process of capturing an analog input, such as a voltage, at a discrete point in time and holding it constant so that it can be quantized. Scope QuickMenu A built-in QuickMenu that places the most commonly used oscilloscope controls around the screen. With the Scope QuickMenu, you may not need to use the main menu system to operate the oscilloscope.

TDS3000 Series User Manual

Glossary7

Glossary

Screen Buttons The rows of buttons below and to the right of the display that select items in the menus. Selected Waveform The waveform on which all measurements are performed and which is affected by vertical position and scale adjustments. Signal Path Compensation (SPC) The ability of the oscilloscope to minimize the electrical offsets in the vertical, horizontal, and trigger amplifiers caused by ambient temperature changes and component aging. You should run SPC when the ambient temperature varies more than 5_ C from the last SPC or before performing critical measurements. Single Sequence A single shot acquisition if Sample or Peak-detect acquisition mode is selected. A series of N acquisitions if Average or Envelope acquisition mode is selected (N is the number of averages or envelopes). Standby (STBY) The off-like state when the instrument in not in use. Some circuits are active even while the instrument is in the standby state. TekProbe Interface An interface that communicates information between a probe and the oscilloscope, such as the probe type and its attenuation factor. The interface also supplies power to active probes.

Glossary8

TDS3000 Series User Manual

Glossary

Tek Secure A feature that erases all waveform and setup memory locations (setup memories are replaced with the factory setup). Then it checks each location to verify erasure. This feature is useful where the oscilloscope is used to gather security-sensitive data. Time Base The set of parameters that let you define the time and horizontal axis attributes of a waveform record. The time base determines when and how long to acquire record points. Vertical Bar Cursors The two vertical bars you position to measure the time parameter of a waveform record. The oscilloscope displays the value of the active (moveable) cursor with respect to trigger and the time value between the bars. Video Trigger Triggering on the sync pulse of a composite video signal. XY Format A display format that compares the voltage level of two waveform records point by point. It is useful for studying phase relationships between two waveforms. YT Format The conventional oscilloscope display format. It shows the voltage of a waveform record (on the vertical axis) as it varies over time (on the horizontal axis). Zoom The horizontal magnification capability of the oscilloscope. Zoom is a display function; adjusting zoom does not affect the acquired waveform.

TDS3000 Series User Manual

Glossary9

Glossary

Glossary10

TDS3000 Series User Manual

Index

Index
Symbols
.gz file format, 323 AUTOSET button, 34 undo, 34 averaging, 37

A
accessories, C1 acquisition menu, 36 modes, 37 overview, 15 rate, 311 resolution, 311 single shot, 220, 33 status, 32 stopped, 32 waiting for trigger, 32 active cursor, 314 adjusting probes, 13 amplitude measurement, 337 application modules descriptions, C2 installing, 115 applications autoset, 22 averaging, 211 cursors, 212 custom measurements, 26 delay, 213 examples, 21 gray scale, 219 measurements, 23, 25 measuring jitter, 215 peak detect, 210 video, 216 zoom, 222

B
B TRIG button, 350 B trigger, how to use, 350 backlight intensity, 317 timeout, 363 bandwidth selection, 372 battery charging, 114, C3 installation, 112 power, 110 recycling, vii safety, 111 burst width measurement, 337

C
calibration, 14, 367 cleaning, F1 clear spool, 323 COARSE button, 314 color display, 319 printing, 323 communication modules description, C2 installing, 117 compressing hard copy files, 323 constellation diagram, 320

TDS3000 Series User Manual

Index1

Index

cursors active, 312 application example, 212 gating, 28, 335 interactions, 336 locating, 314 measurements, 212 measurements when cursors at same position, 316 menu, 312 readouts, 315 tracking mode, 316 V Bars and FFT measurements, 316 cycle mean measurement, 337 cycle RMS measurement, 337

E
edge trigger, 355 envelope, 37 error log, 369 ESD handling precautions, vii Ethernet, 366 external trigger, 357

F
factory calibration, 368 factory setup, detailed description, B1 fall time measurement, 337 fast trigger, 311 file compression, hard copy, 323 file system application example, 223 extensions, 348 formatting a disk, 347 how to use, 343 protections, 347 firmware upgrade, 344 FORCE TRIG button, 350 frequency measurement, 337 front panel identifying connectors, 130 identifying controls, 121 functional check, 12

D
date/time, how to use, 362 delay example application, 213 how to use, 327 interactions, 328, 331, 332 DELAY button, 327 deskew, probe, 372 diagnostics, 368 digital phosphor, 35 disk, online tour, C1 disk drive application example, 223 how to use, 343 display colors, 319 identifying items in, 126 menu, 317 overview, 16 persistence, 319 roll mode, 332 slow horiz. settings, 331

Index2

TDS3000 Series User Manual

Index

G
gated xyz, 320 GPIB, 366 communication module, C2 gray scale application example, 219 controlling, 35 losing information, 330 measurements, 315 restrictions, 375, 377

L
language, how to select, 361

M
magnify. See zoom math preview, 375 source waveform screen position, 375 math waveform, 374 max measurement, 337 mean measurement, 337 measurements cursor, 212 gating, 28, 335 interactions, 334 menu, 333 reference levels, 27 V Bars and FFT, 316 menus, how to use, 118 min measurement, 337

H
hard copy. See printing hard copy file compression, 323 high level measurement, 337 holdoff, 358 horizontal expansion marker, 326 position, 325 resolution, 311 scale, 329 horizontal preview application example, 221 interactions, 330 horizontal zoom application example, 222 how to use, 329 interactions, 330 maximum, 330

N
negative measurement duty cycle, 338 overshoot, 338 width, 338

I
I/O ports, 364 initial setup, 11 ink saver, 323

O
online tour disk, C1

TDS3000 Series User Manual

Index3

Index

P
peak detect, 37 peak-to-peak measurement, 338 performance verification, E1 period measurement, 338 persistence, 319 positive duty cycle measurement, 338 positive overshoot measurement, 338 positive width measurement, 338 power AC line, 19 battery, 110 probe, D11 power cord, 19 power off timeout, 362 pretrigger, 325 preview application example, 221 horizontal, 330 vertical, 373 printing color, 323 connections, 321 date/time stamp, 324 error message, 324 hard copy file compression, 323 ink saver, 323 preview, 323 printer compatibility, 322 spooler, 323 probe compensation, 13 probe deskew, 372 probes compensating, 13 general information, D1 power limitations, D11 safety information, D3

product description accessories, C1 models, 14 overview, 15 probes, D10

Q
QuickMenu, 128 how to use, 339 identifying menu items, 128

R
readouts, cursors, 315 rear panel, identifying connectors, 131 record length, 311 reference scale and position, 375, 377 waveforms, 376 removing waveforms, 370 rise time measurement, 338 RMS measurement, 338 roll mode, 332 RS-232 communication module, C2 troubleshooting, 366 RUN/STOP button, 32

S
safety, battery operation, 111 sample, 37 save/recall menu, 340 setups, 341 waveforms, 342, 376

Index4

TDS3000 Series User Manual

Index

saved waveforms formats, 344 naming, 344 printing, 346 SELECT button, 314 self test, 368 SET TO 50% button, 349 signal path compensation, 14, 367 signal processing, overview, 15 SINGLE SEQ button, 33 single shot, 33 application example, 220 slow roll mode, 331 specifications, A1 status acquisition, 32 trigger, 354 stopped acquisition, 32

U
undo autoset, 34 utility menus, 360

V
vertical menu, 371, 374, 376 offset, 373 position, 370 preview, 373 scale, 370 video modulation, application example, 219 video trigger, 359 application example, 216 sync pulse, 359

T
Tek Secure, how to use, 363 TekProbe interface, 372, D2 time base controlling, 329 fast settings, 331 time-out, 362 trigger auto, 357 edge, 355 external, 357 holdoff, 358 level, 349 menu, 349 normal, 357 position marker, 326 status, 354 video, 359 XY waveforms, 319

W
WAVEFORM INTENSITY, 35 waveform off, 370 waveform record icon, 127

X
XY waveform controlling, 319 gated xyz, 320 restrictions, 320 triggering, 319

TDS3000 Series User Manual

Index5

Index

Z
zoom application example, 222 how to use, 329 interactions, 330 maximum, 330

Index6

TDS3000 Series User Manual