&
G. A. Zakaria
Gummersbach Hospital, Academic Teaching Hospital of the University of Cologne, Germany
Objectives
To calculate the perturbation correction factors for a specific chamber To estimate the statistical uncertainty To compare the value calculated by Monte Carlo with theoretical/ experimental value
Absorbed-dose to water dosimetry uses the Spencer-Attix cavity theory to relate the absorbed dose to the gas in the ion chamber, Dgas, to the dose to the surrounding phantom medium, Dmed, by the following expression:
D med
(1)
is the ratio of the spectrum averaged restricted mass collision stopping power for the medium to that of the gas.
_ med L gas
Cont.
In case of radiotherapy dosimetry, water phantom and air filled ionization chamber normally be used. So, the modified cavity theory as follows:
D water
_ L
water air
_ L = D air
water air
(1)
Cont.
Spencer-Attix cavity theory has three necessary assumptions: (1) the cavity does not change the electron spectrum in the medium (2) the dose in the cavity comes from electrons that enter the cavity and not from those that are created within the cavity (3) charged particle equilibrium (CPE) will be exist Unluckily, the ion chambers do not satisfy the above assumptions
Cont.
Presence of chamber stem in the phantom Need correction factor for stem, Pstem Presence of air in ion chamber which is low dense material comparatively with water. Electron spectrum will be changed due to this air cavity in the phantom. Need correction factor for spectrum change, Prepl
Cont.
Finally, Spencer-Attix cavity theory with all over perturbation factor followed by equation (1) is
_ water L = D air Pcel Pstem Pwall Prepl air
D water
( 2)
Cont.
The value of perturbation factors are energy dependent Each chamber should have separate correction factors for energy basis but it is quite impossible, expensive and time consuming. PSDL / SSDL choice a specific energy (60Co) for individual chambers to measure chamber correction factor. Need additional correction factor, KQ
Cont.
Electrode diameter = 1.1 mm Length of electrode = 5 mm Thickness of graphite coat = 0.15 mm Graphite density = 0.82g/cm3 Radius of sensitive volume=2.75 mm Length of sensitive volume = 6.5 mm
Cont.
0.5
Graphite coat
0.69
PMMA
Electrode
0.55
Air cavity Fig.1: Schematic figure of a PTW semiflex 0.125 cm3 ion chamber
Cont.
60Co
spectrum was used for all calculation SSD was = 80 cm Water depth of calculation = 5 cm Field size 10 cm diameter 20cm20cm20cm water phantom was used
Cont.
Photon cutoff energy was 0.001MeV Electron cutoff energy was 0.521MeV 109 particle histories were simulated for each calculation XCOM Photon Cross Sections data are used from NIST (published by Hubbell et al).
Calculation of Pelec
Dose calculation to the chamber effective point of measurement with central electrode, Delec Dose calculation on chamber effective point of measurement without central electrode, Dnoelec
Calculation of Pelec
A
Pelec = DB / DA
Calculation of Pwall
Dose calculation on chamber effective point of measurement without central electrode with air filled, Dwall Dose calculation on chamber effective point of measurement without central electrode and wall, Dnowall The existing wall material replaced by water for Dnowall calculation
Calculation of Pwall
C * Air filled cavity PTW semiflex chamber with only wall D * Air filled cavity PTW semiflex chamber with only wall Effective point of measurement Effective point of measurement
Pwall = DD / DC
Calculation of Prepl
Dose calculation on chamber effective point of measurement without central electrode and wall with water vapor filled, Dsteam Dose calculation on chamber effective point of measurement without chamber in water medium at small voxel, Drepl The existing chamber material replaced by water for Drepl calculation
Calculation of Prepl
E * Effective point of measurement Water vapor
Chamber cavity filled with water vapor Effective point of measurement F * Calculation of dose at small voxel in water
Prepl = DF / DE
Cont.
The calculated value Pcel is found to be 0.995 0.7% This is in good agreement with the published value 0.993 in TRS-398 with PTW 31003 flexible ion chamber.
Cont.
The calculated value of Prepl is found to be 0.992 0.4% The AAPMs TG-51 and TG-21 dosimetry protocols use a value of Prepl = 0.992 for a cylindrical chamber of inner diameter of 6 mm in a 60Co beam. This value is from the work of Cunningham and Sontag who derived Prepl based on analytical calculations and experiments. For the same quantity the IAEAs TRS-398 Codes of Practice use a value of 0.988 which is based on the measured data of Johansson et al.