Instruction Manual Immersion Cooler HAAKE EK45/90

Analyze Detect Measure Control TM

Table of Contents

1.

Key to Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 1.2 Symbols used in this manual . . . . . . . . . . . . . . . . . . Symbols used on the unit . . . . . . . . . . . . . . . . . . . . .

3
3 3

2. 3. 4. 5.

Quality Assurance . . . . . . . . . . . . . . . . . . . . . . . . . . Your Contacts at Thermo Electron (Karlsruhe) GmbH . . . . . . . . . . Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Unit Description . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5.1 5.2 5.3 5.4 5.5 Immersion cooler HAAKE EK45 . . . . . . . . . . . . . . . Immersion cooler HAAKE EK90 . . . . . . . . . . . . . . . Reachable temperature . . . . . . . . . . . . . . . . . . . . . . Temperature control . . . . . . . . . . . . . . . . . . . . . . . . . Cooling circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4 4 5 7
7 7 7 8 8

6.

Unpacking / Setting Up . . . . . . . . . . . . . . . . . . . . . .
6.1 6.2 6.3 6.4 6.5 6.6

Transportation damage? . . . . . . . . . . . . . . . . . . . . . 9 Contents of delivery . . . . . . . . . . . . . . . . . . . . . . . . . 9 Ambient conditions according to EN 61010 . . . . . 9 Resting time after transportation . . . . . . . . . . . . . . . 9 Ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Information concerning the CE sign . . . . . . . . . . . . 10

7.

Functional and Operating Elements . . . . . . . . . . 11


7.1 7.2 7.3 Immersion cooler HAAKE EK45 . . . . . . . . . . . . . . . 11 Immersion cooler HAAKE EK90 . . . . . . . . . . . . . . . 12 Temperature controller . . . . . . . . . . . . . . . . . . . . . . . 13

8.

Mounting Special Accessory (only for EK45) . . 14


8.1 Fitting the evaporator in the optional clamping bracket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

9.

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
9.1 Cooling of liquid baths . . . . . . . . . . . . . . . . . . . . . . . 15

10. Bath Liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 11. Connecting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18


11.1 Connecting to the mains . . . . . . . . . . . . . . . . . . . . . 18

Table of Contents

11.2 Connecting a sensor . . . . . . . . . . . . . . . . . . . . . . . . . 18 11.3 Changing the mains plug . . . . . . . . . . . . . . . . . . . . . 18 11.4 Fuses on the unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18

12. Operating . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
12.1 Displaying the actual temperature . . . . . . . . . . . . . 19 12.2 Setting the desired set temperature . . . . . . . . . . . . 20

13. Faults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
13.1 Messages shown on the display 15 . . . . . . . . . . . . 21 13.2 Faults with the HAAKE EK45 . . . . . . . . . . . . . . . . . 21 13.3 Faults with the HAAKE EK90 . . . . . . . . . . . . . . . . . 22

14. Checking the Parameters . . . . . . . . . . . . . . . . . . . . 23 15. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24


15.1 Cleaning the fins of the liquefier . . . . . . . . . . . . . . . 24 15.2 Discarding the unit . . . . . . . . . . . . . . . . . . . . . . . . . . 24

16. Accessories / Technical Specifications . . . . . . . 25


16.1 Fuse values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

Key to Symbols

1.

Key to Symbols
Warns the user of possible damage to the unit, draws attention to the risk of injury or contains safety notes and warnings. Denotes an important remark.

1.1 Symbols used in this manual

Indicates the next operating step to be carried out and what happens as a result thereof.

1.2 Symbols used on the unit Caution: Read the instruction manual!
set

Activating the input possibilities

up down

Value alteration () higher / () lower

Only for the EK90: Fault display light on = the unit has a fault; off = the unit does not have a fault Reset button (for usage after a fault or interruption).

Contacts at Thermo Electron (Karlsruhe) GmbH

2.

Quality Assurance

Dear customer, Thermo Electron (Karlsruhe) implements a Quality Management System certified according to DIN/EN/ISO 9001:2000. This guarantees the presence of organizational structures which are necessary to ensure that our products are developed, manufactured and managed according to our customers expectations. Internal and external audits are carried out on a regular basis to ensure that our QMS is fully functional. We also check our products during the manufacturing process to certify that they are produced according to the specifications as well as to monitor correct functioning and to confirm that they are safe. This is why we initiate this monitoring process of important characteristics already during manufacturing and record the results for future reference. The Final Test label on the product is a sign that this unit has fulfilled all requirements at the time of final manufacturing. Please inform us if, despite our precautionary measures, you should find any product defects. You can thus help us to avoid such faults in future.

3.

Your Contacts at Thermo Electron (Karlsruhe) GmbH

Please get in contact with us or the authorized agent who supplied you with the unit if you have any further questions. Thermo Electron Corporation International / Germany Thermo Electron (Karlsruhe) GmbH Dieselstrasse 4 D-76227 Karlsruhe, Germany Tel. +49(0)721 4094--444 Fax +49(0)721 4094--300 Hotline +49(0)18 05 04 22 53 info.tc.de@thermo.com www.thermo.com The following specifications should be given when product enquiries are made: Unit name printed on the front of the unit and specified on the name plate.
4

Thermo Electron (Karlsruhe) GmbH


TYP V/Hz

Dieselstr. 4 76227 Karlsruhe

Safety Notes

4.

Safety Notes

These notes are intended to draw your attention to risks which only YOU can recognize and avoid or overcome. They are intended to enhance your own safety consciousness. We have set the highest quality standards for ourselves and this unit during development and production. Every unit meets relevant safety regulations. The correct unit usage and proper handling is however solely your responsibility. The following notes must be observed:

This instruction manual must be carefully studied! It contains important information on the connection to the local mains supply, correct unit usage and safe handling. Check for transportation damage during unpacking. Get in contact with supplier and/or carrier for settlement of damage claims. Do not try to start up a damaged unit before the damage has been repaired or you have ascertained the affect of the damage. Ensure that this manual is always at hand for every unit operator. Only use this unit solely for the intended application. Repairs, alterations or modifications must only be carried out by specialist personnel. Consider the manufacturers instruction manuals. Considerable damage can be caused by improper repairs. The Thermo Electron (Karlsruhe) service department is at your disposal for repair work. Do not operate the unit with wet or oily hands. Do not expose the unit to spray water. Do not clean the unit with solvents (fire risk!) - a wet cloth soaked in household detergent is normally sufficient. This device is not designed according to the standard EN 60601-1: 1990 (DIN VDE 0750-1 and IEC 601-1) and should not be operated in rooms used for medical purposes and/or in the vicinity of patients.
5

! ! !

! ! ! !

Safety Notes

The evaporator coil of the unit must not be kinked, bent or twisted. This can lead to the evaporator coil being damaged and to a leakage of refrigerant. The following measures were taken for the protection of the operator:

Protection Class I according to VDE 0106 T1

i.e. protection against electric shocks by grounding all parts within easy reach of the operator. The device must only be connected to mains receptacles with a protective ground.

Protection IP 20 according to EN 60529

i. e. regarding the protection against accidently touching live parts and damage by foreign matter, it has been ensured that foreign bodies with a thickness or diameter of more than 12 mm cannot penetrate. No special precautions were taken against the penetration of water and dust. The device should therefore not be used in a dusty atmosphere or in the neighborhood of spray water. Do not insert wires or tools in any of the openings.

Complete separation from the mains is required when: all dangers caused by this device are to be avoided, Cleaning is carried out, repairs or setting alterations are about to be carried out.

Complete separation means:

Pull out the mains plug!

Unit Description

5.

Unit Description

5.1 Immersion cooler HAAKE EK45 The EK45 is a cooling unit with a one-stage refrigerating circuit with which a maximum low temperature of --45C can be reached. The coolant circuits evaporator coil shaped as spiral-shaped cooling element is connected to the cooling unit via a flexible, specially heat-insulated hose. It can be immersed in vessel with an inner diameter of > 85 mm and a depth of > 200 mm. The cooling unit compressor is air-cooled fully hermetically sealed.

5.2 Immersion cooler HAAKE EK90 Similar to the EK45, however with a two-stage refrigerating circuit with which a maximum low temperature of --90C can be reached. The cooling element is designed as a flexible corrugated coil and can thus be adapted to suit the shape of the vessel available. The inner diameter of the vessel must however not be smaller than 110 mm (2 x the smallest bending radius of 40 mm = 80 mm plus 2 x the diameter of the corrugated coil at 13 mm = 26 mm), whereas the vessel depth can be as small as only 100 mm. The dimensions are the minimum dimensions to which the user can form the cooling element. This should not however be practised often. Frequent bending with very small radii can damage the corrugated coil. If multiple bending is necessary, a minimum diameter of at least 250 mm should be observed (bending radius of 110 mm). 5.3 Reachable temperature The lowest possible temperature in a bath and the shortest cooling down time are dependent on the following:

the volume of liquid to be cooled, the type of liquid and its viscosity at the desired working temperature, the heat conductivity of the liquid to be cooled, the bath vessel and covering insulation, the heat created by any unit immersed in the bath used to recirculate or stir the liquid.
7

Unit Description

The lowest temperatures possible for the EK45 and EK90 specified above were reached in a Dewar vessel with a volume of two liters. Methylcyclohexane was used as the bath liquid and the bath was not stirred. 5.4 Temperature control Both units are equipped with a built-in temperature controller with a red LED display and are supplied with a Pt100 temperature sensor fitted as standard. The sensor should be located in the bath in the direct vicinity of the evaporator coil. A temperature accuracy of approx. 1 to 2C is possible. Any temperature within the working temperature range can be set. A heater is necessary for applications which require temperatures above the respective ambient temperature. The temperature controller features the display of the actual temperature, the set temperature and a control status display. A series of parameters can be entered and the Pt100 sensor is monitored for breakage and short-circuiting. 5.5 Cooling circuits Cooling requires the removal of heat. This is carried out for these units by heat exchangers, which draw off and expel the heat. The cooling capacity for air-cooled units (heat exchanger = liquefier) depends to a great extent on the free flow of cooling air through the two ventilation grids. These grids (one at the front and one at the rear side) must therefore be kept free of all obstacles at all times! The units are run on environmentally-friendly and CFC-free refrigerants.

The cooling capacity is reduced at ambient temperatures of > 30C!

Unpacking / Setting Up

6.

Unpacking / Setting Up

Do not pull, push, twist or similarly mishandle the evaporator coil. It must never be kinked or bent too narrowly. 6.1 Transportation damage?

Notify carrier (forwarding merchant, railroad, post office) Compile a damage report, Inform dealer or manufacturer
(Small problems can often be dealt with on the spot).

6.2 Contents of delivery Pt100 cable length 2500 mm EK45 or EK90 cooler with a Pt100 sensor 6.3 Ambient conditions according to EN 61010
6 60 50

indoors, max. 2000 meters above sea level, ambient temperature 5 ... 40 C, relative humidity max. 80%/31C ( 50%/40C) excess voltage category II, contamination level 2

6.4 Resting time after transportation As we can not guarantee that our coolers are always transported upright, lubrication oil can leak from the compressor into the cooling circuit. If the cooler is started up whilst still in this state, the compressor may be damaged to the lack of oil.

Rest the unit for 24 hours after setting up.

6.5 Ventilation

!
6 2

Keep both ventilation grids 2 and 6 at the front/rear free from obstruction to ensure unhindered air circulation. Blocked ventilation grids lead to increased unit heating which in turn reduces the cooling capacity and thus impairs correct functioning.

Information concerning the CE sign

6.6 Information concerning the CE sign The measuring and control instruments from Thermo Electron (Karlsruhe) carry the CE sign which confirms that they are compatible with the EU guideline 89/336/EEC (electromagnetic compatibility). The tests are carried out according to module H (official sheet L380 of the European Community) as our quality assurance system is certified according to DIN / ISO 9001. It was tested according to the strict EMV test requirements of the EN61326-1/A1 (EMV requirements for electrical equipment for measuring technology, conduction technology and laboratory usage). This means it was tested for interference resistance and interference emission according to public low-voltage mains (household and commercial usage). The following basic standards were applied in detail: Interference resistance: EN61000--4--2 electrostatic discharge EN61000--4--3 electromagnetic fields EN61000--4--4 fast transients EN61000--4--5 surge voltages EN61000--4--6 wire--guided HF--signals EN61000--4--8 magnetic field of mains frequency EN61000--4--11 voltage drop/short--time interruption Interference emission: CISPR16/class B wire--guided interference emission CISPR16/class B radiated interference emission EN 61000--3--2 EN 61000--3--3 voltage variations and flickering over-compensation voltage flows

The application in industrial and commercial (public mains) environments is thus possible. A declaration of conformity is supplied with the ordered unit on request. Our strict standards regarding operating quality and the resulting considerable amount of time and money spent on development and testing reflect our commitment to guarantee the high level of quality of our products even under extreme electromagnetic conditions. Practice however also shows that even units which carry the CE sign such as monitors or analytical instruments can be affected if their manufacturers accept an interference (e.g. the flimmering of a monitor) as the minimum operating quality under electromagnetic compatibility conditions. For this reason we recommend you to observe a minimum distance of approx. 1 m from such units.
10

Functional and Operating Elements

7.

Functional and Operating Elements

7.1 Immersion cooler HAAKE EK45

Front

Rear

12

11

A
!

9 1

10

6 4

A 1 2 3 4 5 6 9 10 11 12

Symbol: Read the instruction manual! Mains switch Ventilation grid front (suction side) Connection hose 1.5 m long Evaporator coil (cooling coil) made from stainless steel 18/8 Mains cable Ventilation grid rear (air exhaust) Name plate Fuses (if this fuse is triggered, see chap. 11.4) Temperature controller with display (see 7.3) Socket for Pt100 sensor

11

Functional and Operating Elements

7.2 Immersion cooler HAAKE EK90

Front
13 14

Rear

11

A
!

12 1 6 10 9 2 4 5
!

A A 1 2 3 4 5 6 9 10 11 12 13 14 Symbol: Read the instruction manual! Mains switch Ventilation grid front (suction side) Connection hose 1.5 m long Evaporator coil (cooling coil) made from stainless steel 18/8 Mains cable Ventilation grid rear (air exhaust) Name plate Fuses (if this fuse is triggered, see chap. 11.4) Temperature controller with display (see 7.3) Socket for Pt100 sensor Fault monitoring lamp Reset key

12

Functional and Operating Elements

7.3 Temperature controller

17

15

16

15 Temperature display (Preset value setting and parameter display) 16 SET key for activating setting functions 17 Up and down key (for increasing and reducing values)

13

Mounting Special Accessory (only for EK45)

8.

Mounting Special Accessory (only for EK45)

8.1 Fitting the evaporator in the optional clamping bracket


1

Slide the metal plates supplied with the bracket into the slot above the evaporator coil. Insert the cooling element through the rectangular opening. Attach the inserted plates using four screws.

Do not kink the evaporator hose!

14

Installation

9.
EK45

Installation

! The

evaporator tube and coil should never be sharply bent nor formed to a small radius! Evaporator hose -- smallest permissible bending radius: 125 mm. Evaporator coil (cooling coil) -- smallest permissible bending radius for the EK90: bending once: 40 mm bending often: 110 mm

EK90

Damage caused by an excessive bending of the coiling coil is not covered by our warranty regulations.

9.1 Cooling of liquid baths


40..110

Immerse total length of evaporator coil 4 into the bath


liquid!

Ensure safe mounting in bath vessel for open-bath circulators with a large opening! (for suitable holders, refer to Thermo Haake circulator accessories) pump jet nozzles, it should be guaranteed that they are adjusted so that a thorough mixing of the liquid will take place in the vicinity of the evaporator coil 4.

In open-bath or immersion circulators with swivelled

15

Bath Liquids

10. Bath Liquids


Immerse the cooling coil (evaporator) of the cooler into a suitable bath vessel. The coil must be completely covered with liquid otherwise ice will form. The selection of the proper bath liquid (heat transfer liquid) influences the capacity of a cooler decisively. The temperature accuracy will decrease the higher viscosity of the heat transfer liquid and the lower its heating capacity is.

-30 to 40C

Water with Antifreeze The amount of antifreeze added should cover a temperature range 10C lower than the operating temperature of the particular application. This will prevent the water from gelling (freezing) in the area of the evaporating coil the surface area of which is much colder than the working temperature. An excess of antifreeze deteriorates the temperature constancy due to its high viscosity.

-70 to 40C

SIL 180 This liquid is very stable in the long term. Due to the high viscosity at low temperatures, the usage of a stirrer is to be recommended.

-50 to 40C -90 to -30C

Synth 60 or Synth 20 Synthetic heat transfer liquids which are however hygroscopic and thus require changing more frequently than usual. Please get in contact with us should you have any questions. We are glad to advise you and can help you to choose a heat transfer liquid suitable for your application Thermo Haake heat transfer liquids are supplied with an EC Safety Data Sheet.

! Important !

The highest working temperature as defined by the EN 61010 (IEC 1010) must be limited to 25C below the flashpoint of the bath liquid.

16

Bath Liquids

! Important !

Please ensure when selecting the heat transfer liquid that no toxic gases can be generated and bear in mind that inflammable gases can build up over the liquid during usage. If water alone or water with a small amount of antifreeze is used as the bath liquid, a thick ice coating can form on the evaporator. This leads to a clearly reduced cooling capacity of the immersion cooler. Remove the ice coating by heating up if necessary. Do not remove the ice coating mechanically!

! Important !

17

Connecting Up

11. Connecting Up
11.1 Connecting to the mains 230 V 115 V Only attach this unit to mains sockets with a grounded earth. Compare the local mains voltage with the specifications written on the name plate. Voltage deviations of +/- 10% are permissible. The socket must be rated as suitable for the total power consumption of the unit. 11.2 Connecting a sensor The temperature sensor supplied should be inserted into socket 12. Any commercially available Pt100 sensor can be used instead of this sensor. A suitable plug is available from Themo Haake under the order no. 087-2432. 11.3 Changing the mains plug (e.g. for Great Britain)

12

Pin configuration Pt100:

This should only be carried out by qualified specialist personnel!

The mains cable wires have the following colors: Brown = Live Blue = Neutral Green/Yellow = Earth 11.4 Fuses on the unit All units are equipped with automatic thermally-triggered fuses. If the fuse has triggered

the fuse does not have to be exchanged -- resetting suf-

fices; a red marking is visible; a certain cooling down time should be allowed (approx. 5 min) before the (dip) switch can be pressed again. ! Do not use tools; do not use force. Both destroy the fuse. ! If the fuse should be triggered again after resetting, the unit probably has a defect. In this case the unit should be sent in for servicing.

18

Operating

12. Operating
1

1
2

Ensure that the evaporator is completely covered with liquid. Switch the unit on via the mains switch 1.

! If the cooler has already been in use, please wait at

least 3 minutes for the EK45 (10 minutes for the EK90) after switching off before switching on again. If the unit is switched on before this time has elapsed, the compressor attempts to start up several times without success. In this case the windings of the compressor motor are in danger of burning through.

The second compressor of the EK90 is always automatically switched on with a delay of 30 seconds. This is carried out to protect the compressor.

! When the unit is started under unfavourable envi-

ronmental conditions, the pressostat safety device may respond after about 1 minute. If this happens, please press the reset button 14. Usually the unit will then start correctly without going into fault status again. However, if fault status appears several times in succession, the unit is defective and must be sent in for servicing.

If the unit shows one of the following messages at the display 15, please see chapter 13: or
15

12.1 Displaying the actual temperature The actual temperature (temperature of the control sensor) is displayed at the red LED display 15 in the normal mode. After any settings have been made, the display automatically switches back to actual temperature display.

19

Operating

12.2 Setting the desired set temperature If no entry is made within 3 seconds the value shown in the display 15 switches automatically back to the actual temperature and the previously selected set value automatically remains stored.
1

15

16 17

Depress the SET key 18 briefly The last stored set value is displayed

Hold pushed the SET key 18 for 2 sec. Select the new set value using the Up 16 and Down 17 keys To memorise the new set point value push the SET key again: The new set value is displayed

18

Normal display (actual value)

Set value level display value

Press SET to store the new value

Change value

20

Faults

13. Faults
13.1 Messages shown on the display 15
15

PFo

No sensor is connected or the connected sensor is broken.

! The unit cools down to the lowest possible


temperature.

PFc

The connected sensor has a short circuit.

! The unit cools down to the lowest possible


temperature. 13.2 Faults with the HAAKE EK45 The compressor is automatically switched off by the safety system in case of it being thermally overloaded. The compressor switches itself back on again automatically after it has cooled down. Possible reasons for the compressor switching itself off:

the ambient temperature is too high the air quantity for cooling is insufficient The liquefier fins are too heavily contaminated The ventilator motor has a defect

Please therefore ensure that,

the permissible maximum ambient temperature of


40_C is not exceeded,

the necessary cooling air quantity can circulate unhindered (do not set up the unit directly against a wall). The cooling air flow rate can be very easily checked: A sheet of paper should remain stuck to the ventilator grid 2 (suction side) when the cooler is in use. If the fault cannot be remedied then

clean the liquefier (see chapter 15)


or

have the unit checked by qualified service personnel


21

Faults

13.3 Faults with the HAAKE EK90: If one of the following faults occurs, the fault control lamp 13 lights up and the unit is switched off:

the ambient temperature is too high the air quantity for cooling is insufficient the liquefier fins are too heavily contaminated the ventilator motor has a defect refrigerant leakage in the 1st stage of the cooling circuit.

Please therefore ensure that,


13 14
PT100

the permissible maximum ambient temperature of


40_C is not exceeded,

the necessary cooling air quantity can circulate unhindered (do not set up the unit directly against a wall). The cooling air flow rate can be very easily checked: A sheet of paper should remain stuck to the ventilator grid 2 (suction side) when the cooler is in use.

If the fault cannot be remedied then

clean the liquefier (see chapter 15)


or

have the unit checked by qualified service personnel


After the fault has been eliminated: After the fault has been eliminated, the RESET key 14 must be pressed in order to start up the unit again.

22

Checking the Parameters

14. Checking the Parameters


The default parameter are adjusted. No alterations are necessary!

! Dont touch the keys on the control panel for 30 seconds after switching it on!
15 16 17

To change the set value see chapter 12.2 Parameter list PR1 Press SET key 18 and arrow key 16 together for 3 sec.: PR1 appears in announcement; this level contains all parameter attainable for the user. Use UP 16 or DOWN 17 to change its value. Press SET to store the new value and move to the following parameter. Calibration Select the code OPb (Probe calibration). The adjustment depends on the measuring range. If you get to the programming level PR2 (by pressing the wrong key) and PAS appears on the display, simply press the SET key 18, then hold down the SET key and arrow key 16 together: the actual temperature appears on the display; the factory setting of the parameters is still effective. How to lock the keybord Keep pressed for more than 3 sec. the arrows 16 and 17. The POF message willbe displayed and the keyboard will be locked. At this point it will be possible only to see the set point or the max. o min. temperature stored. If a key is pressed more than 3 s the POF message will be displayed. To unlock the keyboard Keep pressed together for more than 3 s the arrows 16 and 17, till the PON message will be displayed.
23

18

Maintenance

15. Maintenance
The stainless steel surfaces may after some time show spots and become tarnished. Normal stainless steel cleaners as they are used in the kitchen can be used. Vinegar-based cleaners have proved to be suitable used according to the manufacturers recommendations. Do not use scouring powder! The evaporation coil must be kept clean in order to ensure a long service life. Substances containing acidic or alkaline substances and metal shavings should be removed quickly as they could harm the surfaces causing corrosion. If corrosion (e.g. small rust marks) should occur in spite of this, cleaning with stainless steel caustic agents has proved to be suitable. These substances should be applied according to the manufacturers recommendations. 15.1 Cleaning the fins of the liquefier In order to maintain the cooling capacity of the unit, cleaning has to be done two to four times per year, depending on the grade of soiling. Switch off unit and pull out the plug. For the EK45:
1

Clean the fins with compressed air. For extreme soiling, seek the help of specialist service personnel.

For the EK90:


1

Loosen ventilation grid 2: Rotate the mounting screws 90 in any direction and remove grid. Clean fins with brush or similar tool. Replace grid and push screws back in (do not rotate screws).

2 3

15.2 Discarding the unit: One day the life span of your cooling unit will end. Therefore:

This unit contains ozone-friendly coolant R404a and R23. The unit may however only be discarded by authorized personnel.
24

Accessories / Technical Specifications

16. Accessories / Technical Specifications


Special Accessories Screw attachment for fixing the EK45 to bath walls up to a thickness of 25 mm Set of castors for the EK90 Order No. 333--0452 333--0508

Technical Specs. acc. to DIN58966 Working temp. range C Cooling cap. at 20C W at --10C W at --40C W at --60C W Compressor cooling Refrigerant Cooling circuit Hose length cm Cooling coil dimensions(xL) mm Smallest bending radius mm Evaporator material Dimensions (WxD) mm Unit height (H) mm Weight kg Permissible ambient temperatureC Total wattage max. VA Order No. for 230 V / 50-60 Hz for 115 V / 60 Hz

EK45 --45..40 350 250 air R404A 1 stage 150 81x195 -st. steel 225x465 380 24,5 10..40 1840 328-1451 328-1452

EK90 --90..40 300 280 170 100 air R404A/R23 2 stage 150 13x900 40 st. steel 380x460 490 60 10..40 2760 329-1901 -----

16.1 Fuse values


Unit type EK45 EK90 Mains voltage 230V 115V 230V/50--60Hz 200V/50--60Hz Fuse(s) at the rear panel 2x8 A 1x15 A 2x12 A 2x12 A Fuse(s) in the unit 1xT32 mA 1xT32 mA 1xT630 mA 1xT630 mA

Subject to alterations

Printed in Germany

Part no. 002--8856 A 1.1.040.2--05.04

25