Supplier Models B Braun Infusomat Space B Braun Outlook 100ES/200ES/ 300ES/400ES Safety Infusion System Good choice for care areas with fewer infusions per patient, as well as facilities that want a bar-code scanner on the pump.

Specifications for Large-Volume Infusion Pumps


Baxter Colleague CXE/CXE 3 CareFusion Alaris System Hospira Plum A+/A+3 Infusion System with Hospira MedNet Software Hospira Symbiq Infusion System Sigma Spectrum

ECRI Institute judgment

Good choice for care areas with many infusions per patient, as well as facilities that want a bar-code scanner on the pump and a removable battery.

Performs adequately in some areas, but DERS functionality is limited, and wireless networking is not provided. In U.S., FDA has ordered that all Colleague pumps be removed from the market and destroyed by July 2012 due to a long-standing failure to correct problems with the pumps. Worldwide, except USA No Yes

Good choice for facilities that want to standardize on one platform for many infusion modes and that want a bar-code scanner on the pump.

Good choice for care areas with Good choice for facilities that want many IV push and piggyback to standardize on one server for deliveries but few bolus doses, as large-volume and PCA pumps. well as facilities that want to standardize on one server for largevolume and PCA pumps.

Good choice for facilities looking for a small pump, a removable battery, and standard administration sets

Where marketed FDA clearance CE Mark (MDD) Configuration

Worldwide Yes Yes

Canada, USA Yes No

North America Yes Yes

Worldwide Yes Yes

Australia, Canada, USA Yes Yes

USA, Puerto Rico Yes No

Number of channels

1 (CXE) 3 (CXE 3)

Up to 4 plus Alaris Auto ID module for bar-code capabilities

2 (A+) 6 (A+3)

1 and 2

Pump capabilities Flow range, mL/hr Syringe delivery MRI conditional Onboard bar-code reader Wireless connection Alarms and indicators Occlusion alarm limits, mm Hg

0.1-1,200 No (Space Syringe module available) No Yes, with Space Station Yes, with Space Station

0.1-999.9 Yes, with syringe adapter set No No (100 ES) Yes (200ES/300ES/400ES) Yes (400 ES two-way, 300 ES oneway outbound) 75, 100, 200, 300, 400, 500, 750

0.1-1,200 Yes No No No

0.1-999 Yes, with syringe adapter set No Yes, with Alaris Auto ID module Yes

0.1-999 Yes No No Yes

0.1-1,000 Yes No No Yes

0.5-999 No No No Yes

225-900

103, 258, 465 (<21 mL/hr) 206, 413, 620 (21-200 mL/hr) 310, 568, 775 (>200 mL/hr) Yes, Guardian

50-525 mm Hg

50-775

50-775 distal 50-250 proximal Yes, MedNet

310, 672, 982 nominal

Dose-error reduction system (smart technology) Pump defaults to DERS on startup Drug library capacity

Yes, DoseGuard

Yes, DoseGuard

Yes, Guardrails Suite MX

Yes, MedNet

Yes

Yes 1,500 drugs, 15 service lines

Yes 300 drugs, 15 service lines

No 500 drugs, 8 service lines

Yes 2,500 unique drug/fluid setups across 30 care areas/profiles

Not specified 1,800 drugs, 15 service lines

Yes 8,000 drugs, 40 service lines

Yes 1,600 drugs, 32 care areas

Drug library organization

Up to 1,500 drug entities (name, dosing unit); one concentration per entity, which may be assigned to any of up to 15 service lines

300 drug entities (name, dosing unit); each entity can have up to 2 concentrations in any of the 15 service lines (total of 9,000 possible entries)

Each of 500 drug entities (name, Each of 2,500 drug entities (name, dosing unit, concentration) can be dosing unit, concentration) can be enabled in up to 8 service lines; 64 enabled in up to 30 service lines of the 500 entities are preconfigured by Baxter

Each of 1,800 drug entities (name, dosing unit, concentration) can be enabled in up to 15 service lines

Each of up to 40 service lines can contain up to 400 drug entities (name, dosing unit, concentration)

Each drug entry includes, at a minimum, the care area, drug name, concentration, dose rate mode, bolus mode, starting dose rate, soft limit, hard limit, volume to be infused, primary or secondary IV container Yes Yes Yes

Bolus dosing limits Clinical advisories Log-analysis software

Yes Yes Standard and custom reports coming 4Q2011

Yes Yes Standard and custom reports available

No No No

Yes Yes, 100 Standard and custom reports available, cloud log analysis for benchmarking

No No Standard and custom reports available

Yes Yes Standard and custom reports available

IDN support Total number of pumps supported on one server

24 pumps per bed and up to 1,000 beds for potential total of 24,000 pumps No No 802.11 a/b/g/n

3,500 pumps

N/A

Alaris PC unit supports 4,000 pumps 2,000 with up to 16,000 channels

2,000

400

Can support >1 hospital on a server Allows for cross-facility log analysis IT considerations 802.11xx protocols supported

Yes Yes a/b/g

NA NA NA

Yes Yes 802.11 a/b/g

Yes Yes 802.11 a/b/g

Yes Yes 802.11 a/b/g

Yes Yes 802.11 b/g

Wireless encryption supported

WPA, WPA2 (Enterprise)

WEP 40, WEP128, WPA-PSK, WPA2- NA PSK

WEP, WPA, WPA2

WEP 64 and 128 (open and shared keys) WPA-TKIP using preshared key WPA2- AES/CCMP using preshared key OPENSSL version 0.98a All message traffic can be SSL encrypted with a 128-bit cipher

WEP 64 and 128 (open and shared keys) WPA-TKIP using preshared key WPA2- AES/CCMP using preshared key OPENSSL version 0.98a All message traffic can be SSL encrypted with a 128-bit cipher

Yes

Wireless authentication supported

EAP-TLS (originally certified protocol); EAP-TTLS/MSCHAPv2; PEAPv0/EAP-MSCHAPv2; PEAPv1/EAP-GTC; LEAP; PEAP Ethernet, RS232, USB, master/slave, wireless LAN, PS/2 with Space Station and SpaceCom

PEAPv0-TKIP-MSCHAPv2, PEAPv0AES-MSCHAPv2

NA

PSK, PEAP w/MS-CHAP V2, EAP-TLS, WPA -TKIP with EAP-TLS, EAP-PEAP, EAP-TTLS w/MS-CHAPV2 or PAP EAP-TTLS, EAP-FAST WPA2-AES/CCMP - EAP-TLS, EAPPEAP, EAP-TTLS, EAP-FAST RS232 (RJ45) Ethernet, RS232

WPA -TKIP with EAP-TLS, EAP-PEAP, Yes EAP-TTLS, EAP-FAST WPA2-AES/CCMP - EAP-TLS, EAPPEAP, EAP-TTLS, EAP-FAST Ethernet IrDA

Data port

100ES/200ES/300ES/400ES: Nurse RS232 call interface, Ethernet (configuration and download) 300ES/400ES: Wireless COM ports

Battery Battery type Life, hr @ flow (mL/hr) Recharge time, hr Removable H x W x D, cm (in) NiMH 4 @ 100 6 Yes 6.8 x 21.4 x 12.4 (2.7 x 8.4 x 4.9) NiMH 100ES/200ES: 7 @ 25 300ES/400ES: 5 @ 25 10 No 11.4 x 30.5 x 22.2 (4.5 x 12 x 8.7) Not specified 4 @ 100 (CXE) 2 @ 100 x 3 channels (CXE 3) 12 NiMH (10) 4 @ 25 (4 channels) 4 @ 95% Internal sealed lead-acid rechargeable 6 V Rechargeable lithium-ion Lithium-ion 8 @ 125 (non-wireless) 4 @ 125 (wireless) 12 (non-wireless) 16 (wireless) Yes Standard battery with IV pole clamp: 14.7 x 16.3 x 11.9 (5.8 x 6.4 x 4.7) Wireless battery with IV pole clamp: 16.0 x 16.3 x 11.9 (6.3 x 6.4 x 4.7)

3 @ 125, or 250 mL total @ 126 4 @ 100 6 No A+: 20.3 x 20.3 x 15.2 (8 x 8 x 6) A+3: 48.3 x 38.1 x 35.6 (18 x 15 x 14) 6 No 25.9 x 25.1 x 21.8 (10.2 x 9.9 x 8.6) 1-channel; 27.6 x 27.9 x 25.1 (10.9 x 11 x 8.6) 2-channel

Yes No CXE: 25.9 x 19.7 x 20.3 (10.2 x 7.8 Alaris PC unit: 22.4 x 17.5 x 22.9 x 8) (8.8 x 6.9 x 9); Alaris pump CXE 3: 35.3 x 19.7 x 20.3 (13.9 x module: 22.6 x 8.4 x 14 (8.9 x 3.3 x 5.5) 7.8 x 8)

Weight, kg (lb)

1.4 (3.1)

5 (11)

CXE: 5.5 (12.1) CXE 3: 7.9 (17.5)

Alaris PC unit: 7.2 lb; Alaris pump module: 2.5 lb.

A+: 4.3 (9.5) A+3: 12.7 (28)

4.8 (10.7) 1-channel; 5.4 (12.1) 2channel

Standard battery: 0.72 (1.6) without IV pole clamp, 0.95 (2.1) with pole clamp Wireless battery: 0.75 (1.7) without IV pole clamp, 0.98 (2.2) with pole clamp LVP, 27 dose modes; uses Baxter standard IV sets; field upgradable. 24-hour memory of all setup screens, except for multistep and cyclic modes that are maintained permanently (or until cleared). Separate pump history log and drug event log. Approximately 10,000event capacity.

Other specifications

100ES/200ES/300ES/400ES: IV Drug/dose calculator; battery Colleague Guardian feature allows maintenance mode; drug selection; Safety System; color screen with selection of drug dose and real-time display of dose/rate limits; concentration; 64 preset drug labels rate calculation; standby; battery fully customizable drug/solution check; adjustable bolus rate; data and 436 additional user labels; libraries with loading dose and onlock. Meets requirements of CSA, configurable postocclusion restart; demand bolus capabilities. Meets IEC 601-1, IEC 601-2-24, and UL; delayed start; 10-color display; bar-code scanning via Space Station- requirements of CSA and UL 2601-1; multifunction mounting clamp; based scanner. ISO 9001 certified. enables hard concentration limits and weight limits per patient. 200ES/300ES/400ES: DoseScan bar coding for patient matching, programming, and infusion authorization. 300ES/400ES: DoseCom wireless communication; 400ES accepts wireless drug library upload with Configuration Management Application (CMA) software.

Automated concurrent or secondary Alaris System supports Guardrails delivery through "B" port, which Suite safety software, Alaris SpO2 accepts syringes. module (Masimo SET or Nellcor technology), Alaris EtCO2 module, Alaris PCA module, Alaris Syringe module, Alaris Auto-ID module (bar code).

8.3-in color LCD touchscreen; flashable software for software upgrades; automated set loading; 42 configurable settings; 91 dosing units; bolus, continuous, multistep, and intermittent therapy.