Child: care, health and development Original Article

doi:10.1111/j.1365-2214.2010.01079.x

Hearing screening for school children: comparison of low-cost, computer-based and conventional audiometry
cch_1079 323..331

B. McPherson, M. M. S. Law and M. S. M. Wong


Division of Speech and Hearing Sciences, Faculty of Education, The University of Hong Kong, Hong Kong, China Accepted for publication 7 January 2010

Abstract
Background There is a need to develop affordable but effective audiometric screening equipment, particularly for use in low-income countries. With advances in computer technology, low-cost computer-based audiometer software has been developed. However, the efcacy of computer-based audiometers in hearing screening and diagnostic assessment requires investigation. The purpose of this study was to evaluate the accuracy of a low-cost, computer-based audiometric system in a school-based hearing screening programme. Methods Eighty children were screened using the computer-based audiometer and with a conventional pure tone screening audiometer. Overall refer rates, as well as frequency and age effects on the accuracy of the computer-based audiometer, were considered. Results There was a signicant relationship between the low-cost, computer-based audiometer and a conventional pure tone audiometer when a 40 dBHL refer criterion was used in school hearing screening and when test results at 500 Hz were excluded from analysis. However, background noise effects and software limitations in the computer-based system had major adverse effects on screening performance. Conclusions The study results and preliminary practical experience with the system suggest that, with further software and hardware improvements, a low-cost, computer-based system may well be feasible for routine school screening in developing countries.

Keywords audiometry, developing countries, hearing loss, school screening Correspondence: Bradley McPherson PhD, Division of Speech and Hearing Sciences, 5/F, Prince Philip Dental Hospital, 34 Hospital Road, Hong Kong, China E-mail: dbmcpher@hkucc.hku.hk

Introduction
The World Health Organization (WHO) has estimated that more than 4% of the worlds population, i.e. 278 million people, has hearing loss. Of this proportion, approximately two-thirds live in developing countries (Smith 2008). The prevalence rate of hearing loss is thought to be high among school-aged children in developing countries. In Swaziland a prevalence rate of 4.1% was noted (Swart et al. 1995). In Kenya, it was reported that 5.6% of school children have mild hearing loss (Hatcher et al. 1995). Studies in South India (Rao et al. 2002) and Nigeria (Olusanya

et al. 2000) have recorded even higher prevalence rates, of 11.9% and 13.9%, respectively. Although the published prevalence rates in school children vary considerably among developing countries because of differences in study methodologies, populations, screening criteria and other factors, hearing loss is generally more prevalent than in developed countries (Smith 2001; Berg et al. 2006). In comparison, the reported prevalence rate for hearing loss in school children in Finland was 2.5% (Marttila 1986) and for Denmark 3.6% (Parving 1999). Hearing loss is a signicant health problem and, if undetected, can have a signicant negative impact on the speech and

2010 Blackwell Publishing Ltd

323

324

B. McPherson et al.

language development, educational attainment and socialemotional development of children. In view of the widespread absence of neonatal hearing screening services in developing countries (Olusanya et al. 2007) hearing screening in schoolaged children in developing countries should be an essential part of public health services. Unfortunately, while routine screening at school entry has been widely adopted in developed countries, children in developing countries are rarely screened for hearing loss (Madriz 2001; McPherson & Olusanya 2008). There are limited nancial resources or trained personnel to provide routine hearing screening for school-aged children. In addition, audiological services are often urban-based and are usually provided in hospitals and private clinics (Gell et al. 1992). People who live in rural areas are often not able to access these services (Jauhiainen 2001; Gomes & Lichtig 2005). In view of the limited resources in developing countries, lowcost but effective hearing screening methods would be of value. The costs of hearing screening may be attributed to the cost of personnel required to carry out the screening, the time required to conduct the screening and the costs of equipment used. Hence to reduce the costs of screening, the method should be rapid, and low-cost equipment that can be easily operated by primary healthcare personnel should be used (Downs 2004). The screening method also must be effective in detecting children with hearing loss. Effectiveness is determined by the sensitivity and specicity of hearing screening (Hind et al. 1999). A screening method is said to be effective only when it achieves both high sensitivity and specicity. School-based hearing screening methods can be classied into two groups: subjective and objective hearing status measures. Subjective approaches include questionnaires and pure tone audiometry. The feasibility of using questionnaires as a hearing screening method has been evaluated in various studies in developing countries (Newton et al. 2001; Gomes & Lichtig 2005). The costs of using questionnaires are considered to be low and the implementation of a questionnaire approach does not require professionals primary health assistants can be employed. However, generally the sensitivity and specicity of questionnaire-based hearing screening have been poor (McPherson & Olusanya 2008). Objective methods such as otoacoustic emission recording and immittance audiometry screening are often used in hearing screening programmes (Roeser & Clark 2004). The equipment costs of these physiologic methods have decreased in recent years and this increases their feasibility as screening methods in developing countries. Nevertheless, the sensitivity and specicity of these methods are also relatively low when compared with the standard subjective screening approach, pure tone audiometry (Nozza 2001). In

addition, objective methods do not aim at measuring hearing per se, but investigate middle and inner ear status. Conventional pure tone audiometry is an accurate method of hearing screening in developing countries (Berg et al. 2006). WHO has suggested that children in developing countries should be screened using an audiometer at school entry (Rao et al. 2002). If this recommendation is followed, a pure tone air conduction audiometer is an essential component of hearing screening. However, the costs associated with it are considerable and the equipment may be unaffordable in many developing countries (McPherson 2008). Low-cost, hand-held screeners have only limited accuracy (McPherson & Knox 1992; Parving et al. 2008). In order to develop affordable equipment for hearing screening in developing countries, the present study investigated the effectiveness of a low-cost, computer-based audiometer in hearing screening for school-aged children. There are several reasons supporting the use of a computer-based audiometer in hearing screening. The application of computer technology in health care has been a continuing trend, including its use in developing countries (Wootton 2001). With advances in technology, digital signals can be stored in a computer memory and the intensity and frequency of an output signal can be accurately adjusted. There are several advantages for computerization of hearing screening programmes. The personal computer is common nowadays in many developing countries, so accessibility to hearing screening equipment, especially in rural areas, may be improved. The cost of purchasing a personal computer is now low when compared with that of purchasing a conventional pure tone audiometer. Pure tone audiometers are specialized devices that are manufactured in small quantities, making them uncommon and relatively costly. Laptop computers that cost approximately $US200 are now being manufactured for low-income nations, further increasing the feasibility of using computer-based audiometry in the developing world (Trucano 2008). A potential additional advantage is that computer-based audiometers can be designed with automatic screening procedures. Therefore, personnel with specialized training may not be required to perform screening. Moreover, the use of computerized methods can make the recording of data and quality assurance monitoring of screening programme activities easier (Hong & Csaszar 2005), as well as facilitate telemedicine applications. As the application of computers in hearing screening is still in its infancy, few researchers have studied the feasibility of using computers in audiometric testing. In one study, an Internetbased, telemedicine audiometric system was compared with conventional audiometry in determining auditory thresholds in

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

Hearing screening using low-cost PC audiometer

325

adults. Telemedicine hearing thresholds varied by less than 2 dBHL from those obtained with conventional audiometry (Givens & Elangovan 2003). Another computerized testing system also reported good threshold reliability with both normal and hearing impaired adults (Henry et al. 2003). In the present research project the feasibility of using a computerbased audiometer in hearing screening rather than in obtaining hearing thresholds was evaluated. Results from a conventional pure tone audiometer were treated as the gold standard in this study as the reliability of conventional pure tone audiometry in hearing screening is usually high (Berg et al. 2006; Sideris & Glattke 2006), and this screening method has been used extensively in past decades. In summary, there is a need to develop more affordable and effective audiometric screening equipment, particularly for use in developing nations. Low-cost, computer-based audiometer software has been developed. However, the efcacy of computer-based audiometers in hearing screening for school children is unknown. The purpose of this study was to evaluate the accuracy of a low-cost, computer-based audiometric system in a school hearing screening programme. In addition, screening personnel were debriefed after the study to identify any perceived problems or inadequacies in the computer-based equipment. Results from this study may help determine whether it is feasible to replace the conventional hearing screening audiometer with more affordable equipment in developing countries.

Screening personnel
Two Cantonese speaking testers who were speech and hearing sciences undergraduate students operated the conventional audiometer and the computer-based audiometer. They had received 12 h of instruction in audiological assessment procedures during their undergraduate training programme.

Test equipment Computer-based audiometer


A circumaural headphone (Ovann OV880V) and a joystick for subject response (Blazepro USB) were directly connected to a personal computer (IBM ThinkPad laptop PC, model T22). The computer audiometer software Home audiometer softwarev. 1.83 was installed into the PC. The programme can measure thresholds at 125, 250, 500, 750, 1000, 1500, 2000, 3000, 4000, 6000 and 8000 Hz with a nominal range from -10 to 70 dBHL.

Conventional screening audiometer


A portable pure tone screening audiometer (Madsen Micromate) was used for hearing screening. The audiometer was calibrated, according to ISO 389 series standards, prior to screening. The conventional audiometer was tted with circumaural ME-70 enclosures over TDH-39 supra-aural earphones, an arrangement designed to reduce audible background noise in the screening environment.

Method
Participants
Eighty participants, aged 6, 7 and 8 years, were recruited from a normal, urban Hong Kong primary school that had offered to participate in this study. The participant age range was chosen to be comparable to the range of school entry age found in many developing countries (Bommier & Lambert 2000). All participants were Cantonese speaking children with no known language or cognitive impairment. Cantonese speakers were preferred as the instructions were given in Cantonese, and control of this factor reduced the possibility that the participants could fail hearing screening because of an inability to follow instructions. Participants were recruited on a voluntary basis, with consent forms signed by parents and children prior to testing, and all children who volunteered were tested. The research project was approved by the appropriate institutional review board at the University of Hong Kong.

Equipment costs
Costs for the conventional and computer-based equipment are shown in Table 1. Low-cost computer system components were deliberately selected, to reect components typically available and affordable in developing countries. A dedicated, low-cost

Table 1. Comparison of costs between computer-based audiometer installed in an existing computer and conventional audiometers (in US dollars) Computer-based audiometer Software: Home Audiometer USB hub USB joystick Audio headset Laptop computer Total $27 $10 $4 $42 $500 $583 Conventional screening audiometer $1026

$1026

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

326

B. McPherson et al.

laptop computer for the computer-based system may be found in the $US200 to 500 price range the upper price limit is used in Table 1.

Test environment
Hearing screening with both procedures was conducted in a quiet, but not sound-treated room at the primary school over two non-school attendance days. Ambient noise level was measured using a sound level meter (Cesva SC30) on ve occasions, randomly during testing. The noise level was similar on both days and the average noise level was 36.8 dBA. Background noise levels in occupied Hong Kong primary school classrooms are known to be high, an average of 60.7 dBA (Choi & McPherson 2005), and hence times when classrooms were unoccupied were chosen.

used in the computer-based audiometer. The passing criterion was having reliable responses in two out of three trials at each frequency at 40 dBHL, in both ears. Failure to respond at any frequency in either ear was regarded as a screening refer case. Children were instructed to briey press the appropriate response button when they heard a tone, no matter how faint it was. Continuous pure tones were presented for approximately 1 to 1.5 s during screening.

Test procedures: computer-based audiometer


With the computer-based audiometer, hearing thresholds at each frequency (500, 1000, 2000 and 4000 Hz) were automatically determined using an adaptive procedure. Testing commenced in the left ear at 500 Hz, then 1000, 2000 and 4000 Hz. The computer software automatically commenced testing at 500 Hz, although hearing screening conventionally starts with tones at 1000 Hz and 500 Hz is presented last in such a frequency series (Hodgson 1985). The initial intensity level was 40 dBHL. Intensity increased in 3 dBHL steps each time a tone was represented until the subject responded. After each response, the intensity level decreased in 3 dBHL steps. Hearing threshold was determined after several series of responses were made by the subject. Children were instructed in Cantonese to press and hold the appropriate response button while they heard a tone, no matter how soft it was. Continuous pure tones of 1-s duration were presented during screening. Hearing threshold results greater than 40 dBHL at any frequency in either ear were regarded as a screening refer case. The computer-based audiometer was biologically calibrated before testing commenced according to the instructions accompanying the software. Calibration is dependent on access to an individual with known and accurate audiometric thresholds. Calibration was initially validated by comparing thresholds obtained from four individuals with their known audiograms.

Test procedures: conventional audiometer


Each subject received two hearing screenings, one by conventional audiometer and one by computer-based audiometer. The two screening tests were conducted separately on the same day to minimize any confounding effects of temporal changes in childrens actual auditory status. The order of screening in respect to the two audiometers was random. The two testers were randomly assigned to an audiometer on each day of testing. They were blind to each others screening results. The participants were seated at right angles to the tester to avoid any visual cues during screening and were tted with headphones by the tester. In developing countries, WHO guidelines suggest provision of hearing aids for children with hearing thresholds greater than 30 dBHL in the better ear (WHO 1999). However, in many developing countries children with mild hearing loss (25 to 40 dBHL) are often not screened because priority is given to detecting children with hearing loss that can be most successfully managed through amplication (Arslan & Genovese 1996) or because ambient noise conditions preclude testing at lower sound levels (Jacob et al. 1997). Even isolated village or rainforest locations in developing countries have background noise levels that may at times exceed 40 dBA (Counter 1986). Thus, in this study, the passing criterion was set at a higher intensity level (40 dBHL) across the test frequencies 500, 1000, 2000 and 4000 Hz in order to replicate screening conditions in many developing countries. With the conventional audiometer, the participants were screened with 500 Hz, then 1000, 2000 and 4000 Hz, at 40 dBHL starting from the left ear to replicate the procedure

Follow-up procedures
The parents and the school principal received hearing screening reports of all tested children. Based on the conventional screening audiometer ndings, parents of referral case children were given recommendations regarding follow-up assessments.

Data analysis
To evaluate the accuracy of the computer-based audiometer, the overall referral rates (failed at any one frequency at either ear)

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

Hearing screening using low-cost PC audiometer

327

Table 2. Detail of participants with respect to age, gender and grade No. of participants (%) Age, years 6 7 8 Gender Male Female Grade Primary 1 Primary 2 Primary 3 20 (25) 37 (46) 23 (29) 50 (62.5) 30 (37.5) 26 (32.5) 37 (46.3) 17 (21.2)

Table 3. Comparison of overall referral rates before and after excluding results at 500 Hz Referral rates Before excluding 500 Hz After excluding 500 Hz Computer-based audiometer 56% (45/80) 15% (12/80) Conventional audiometer 13% (10/80) 11% (9/80)

between the computer-based audiometer and conventional pure tone audiometer were compared using c2-test and sensitivity/specicity analysis. The accuracy of the computerbased audiometer as a function of frequency was analysed using a c2-test or Fishers exact test. Agreement between individual test results for each ear was also compared, using kappa values of agreement.

P > 0.05, d.f. = 1) between the results from the two audiometers, when 500 Hz results were excluded. With the conventional hearing screener results as a gold standard, the computer-based screening system had a sensitivity of 1 (95% CI: 0.631), specicity of 0.49 (95% CI: 0.370.61), positive predictive value of 0.56 (95% CI: 0.450.67) and negative predictive value of 0.43 (95% CI: 0.330.55) with the inclusion of 500 Hz results for both audiometers. Excluding 500 Hz results, the computerbased screening system had a sensitivity of 0.78 (95% CI: 0.40 0.96), specicity of 0.92 (95% CI: 0.820.97), positive predictive value of 0.16 (95% CI: 0.090.27) and negative predictive value of 0.83 (95% CI: 0.730.91). In order to determine whether the computer-based and conventional audiometers screen similarly, referral rates at each frequency of 500, 1000, 2000 and 4000 Hz are compared using multiple c2-tests. Table 4 shows that referral rates at the right ear were generally lower than the left ear. Using c2-tests, there were no statistically signicant differences between the two screening procedures in referral rate at any frequencies except at 500 and 1000 Hz for the left ear. Referral rates for the computer-based audiometer with 500 Hz data excluded decreased as age increased with rates of 23% at 6 years, 14% at 7 years and 6% at 8 years of age. However, there was no statistically signicant age effect (c2 = 2.50, P > 0.05, d.f. = 2). The individual child ear test agreement was kappa = 0.20 (SD = 0.06) when results at all frequencies were considered and kappa = 0.62 (SD = 0.13) when 500 Hz results were excluded. These values are considered to reect poor agreement and substantial agreement between test results for individual ears, respectively (Landis & Koch 1977).

Results
Eighty school children received hearing screening with both computer-based and conventional audiometers. Table 2 shows the demographic characteristics of the participants. Participants with absent responses at 40 dBHL at any frequency at either ear were regarded as refer. Table 3 compares the referral rates before and after excluding results of 500 Hz for both audiometers. The 500 Hz frequency low tone is that most likely to be undetected in noisy environments (McPherson & Olusanya 2008). Before excluding results at 500 Hz, the referral rates for the computerbased audiometer and the conventional audiometer were 56% and 13%, respectively. Results from a c2-test showed a statistically signicant difference in the referral rates (c2 = 33.9, P < 0.05, d.f. = 1). After excluding results at 500 Hz, the referral rates of the computer-based audiometer and the conventional audiometer were 15% and 11%, respectively. A c2-test showed no statistically signicant difference in the referral rates (c2 = 0.49,

Discussion
With the conventional screening results as a gold standard, the specicity of the computer-based system was low and sensitivity high, using the overall screening data. When 500 Hz results were excluded specicity improved and sensitivity declined in value. With a 40 dBHL and 1000, 2000 and 4000 Hz pass/refer criterion the low-cost, computer-based screening audiometer performed comparably to the conventional screening instrument in terms of refer rates, with no signicant difference when results at 500 Hz were excluded. However, the computer-based screening audiometer refer rates were unacceptably high (56% refer) and agreement with conventional screening audiometry for individual ears unacceptably low (kappa = 0.20) when results from all four screen frequencies were used to inform a pass/refer decision. Several

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

328

B. McPherson et al.

Table 4. Comparison of referral rates at each frequency Referral rates (no. of participants) Conventional audiometer Computer-based audiometer 500 Hz Left 6.3% (5) 55% (44) Right* 3.8% (3) 8.8% (7) 1000 Hz Left 3.8% (3) 12.5% (10) Right* 2.5% (2) 2.5% (2) 2000 Hz Left* 3.8% (3) 5% (4) Right* 2.5% (2) 2.5% (2) 4000 Hz Left* 3.8% (3) 3.8% (3) Right* 5% (4) 6.3% (5)

*No signicant difference between two referral rates (P > 0.05).

factors may have contributed to this outcome and these are considered below.

Noise effects in screening


In order to gauge the effect of background noise on hearing screening in this study, the overall referral rates of pure tone audiometry before and after excluding 500 Hz results were compared. The overall referral rate for conventional screening decreased from 13% to 11% when the results at 500 Hz were excluded, comparable to school screening initial refer rates obtained by other researchers (North-Matthiassen & Singh 2007). With exclusion of results at 500 Hz, the referral rate of the computer-based audiometer dropped to a far greater extent (from 56% to 15%), indicating that the computer-based audiometer was much more susceptible to ambient noise. This may have been due to the poorer noise attenuation properties of the non-specialist headphones in the computer-based audiometer. The headphones of the conventional screening audiometer were designed for reliable hearing assessment in non-sound-treated conditions. About 18 dB can be attenuated at 500 Hz using such headphones (Frank et al. 1997). The differences in headphone noise attenuation characteristics may have adversely affected the agreement between the results for the two screening modalities. The headphones in the computer-based system could, in future trials, be replaced by others with better noise attenuation. However, if circumaural noise-excluding headphones are used this would increase the cost of the computer-based audiometer, as such headphones are more expensive to manufacture. Another approach may be to provide low-cost insert earphones that attach to reusable silicone eartips for use with the computer-based system. Insert earphones have superior noise attenuation characteristics to conventional headphones (Wright & Frank 1992) but often are expensive in developing countries and easily occluded with ear wax or discharge (McPherson & Olusanya 2008). Alternatively, testing at 500 Hz could be abandoned. However, 500 Hz is an important frequency for assessing the auditory impact of middle ear disease (Silman et al. 1994) and middle ear disorders are a common cause of hearing loss in many developing countries (Smith et al. 1996).

Participants may have also had difculty in identifying the initially presented 500 Hz tone in the non-sound-treated screening environment. This may have resulted in unreliable responses and could be resolved by revising the software to conform to the ANSI S3.21-1987 guidelines that advise initial testing at 1000 Hz. A 1000-Hz tone will often be easier to immediately detect in a school acoustic environment and 1000 Hz has better testretest reliability (Harris 1945). Screening on days when classrooms are not otherwise occupied can be inconvenient and may present practical problems but it does reduce ambient noise effects and should be considered when planning screening schedules.

Software effects in screening


Results revealed that referral rates between the two audiometers were signicantly different at the rst two test frequencies (500 and 1000 Hz) in the left ear but not for other frequencies. This cannot be simply explained as an ambient noise effect as noise should affect both ears equally at low frequencies. An alternative explanation, as the test always started with 500 Hz in the left ear, may be that participants were at rst unfamiliar with the computer-based audiometer screening method and this adversely affected their initial performance, despite a short trial run at this frequency before screening commenced. As the computer-based audiometer was an automatic device, the testers were unable to stop the programme and reinstruct children. An improved version of the computer software would allow a complete trial run before actual screening commences. Participants who passed the conventional audiometer screening but failed with the computer-based audiometer may have failed the latter because of fatigue effects, as it took a longer duration to complete. Screening procedures may be challenging for school-aged children, especially younger children with a more limited attention span. The computer-based audiometer frequency and the intensity levels are preset by the software and controlled by participant responses. Children were required to press and hold down a button every time they heard a tone. As it is an automatic system, the testers were not able to stop the

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

Hearing screening using low-cost PC audiometer

329

test and reinstruct children during screening. However, the testers were able to condition or reinstruct children during conventional hearing screening if they found that children were unfamiliar or had difculties with the procedures. Results indicated that the number of referrals with the computer-based audiometer was the greatest for the youngest age group (6 year olds), although this effect was not statistically signicant in the present relatively small study group. The facility to pause the programme and provide reinforcement or reinstruction by the tester during a childs screening assessment would be valuable. A pause function would also allow rest intervals for children at any time during screening. As the hearing thresholds were obtained using 3-dB increments, the duration of the test was much longer (about 15 min) than conventional hearing screening in which only 40 dBHL pure tones were presented (about 5 min). In addition, the screening required more sustained concentration as tones were presented down to individual threshold levels. During the debrieng session with the screening personnel all these issues were highlighted. Screening personnel also suggested that intervals between tone presentations should be irregular in order to minimize the chance of predicting the tone without actual perception. The computer-based programme does not yet have this facility. The software, at least with its default settings, was not well suited to screening in school children.

Key messages
Children in developing countries are rarely screened for hearing, despite high prevalence rates. A barrier to the introduction of screening programmes is the cost of specialist hearing assessment equipment. With the gradual penetration of low-cost personal computers into developing countries it is worthwhile to consider if they could be applied in hearing health services for children. Currently available low-cost software and hardware have major limitations in terms of effectiveness in such screening programmes. With further development, more appropriate computerbased technology may have a role to play in hearing screening for children in developing countries.

Acknowledgements
We thank students and staff of The Endeavourers Leung Lee Sau Yu Memorial Primary School for participating in this study and Tang Kwok Tuen and Lee Hing Yu for their assistance in data collection. We also thank two reviewers for extensive helpful comments. Portions of this work were presented at the 1st International Conference on Prevention and Rehabilitation of Hearing Impairment, Beijing International Convention Center, 2628 April 2007 and at the XXIXth International Congress of Audiology, Hong Kong Convention and Exhibition Centre, 812 June 2008.

Conclusions
Results showed that the low-cost, computer-based and conventional audiometers had comparable refer rates when mid and high frequencies were considered, at the studys 40 dBHL passing criterion. This suggests that low-cost, computer-based audiometers may be a promising alternative for hearing screening of school children in quiet but non-sound-treated environments. With advances in computer technology and decreasing costs, computer-based audiometers integrated into basic personal computers may be a viable alternative for hearing screening in developing countries. However, the high refer rate found at 500 Hz precludes the use of the low-cost computer system, as it is presently congured, in school screening programmes. Further development is necessary to overcome the present constraints imposed by the use of conventional headphones and by the current software design. Successfully overcoming barriers to affordable hearing screening instrumentation may benet large numbers of school children with hearing impairment throughout the developing world.

References
Arslan, E. & Genovese, E. (1996) Hearing aid systems in undeveloped, developed and industrialized countries. Scandinavian Audiology, 25 (Suppl. 42), 3539. Berg, A. L., Papri, H., Ferdous, S., Khan, N. Z. & Durkin, M. S. (2006) Screening methods for childhood hearing impairment in rural Bangladesh. International Journal of Pediatric Otorhinolaryngology, 70, 107114. Bommier, A. & Lambert, S. (2000) Education demand and age at school enrollment in Tanzania. Journal of Human Resources, 35, 177203. Choi, C. Y. & McPherson, B. (2005) Noise levels in Hong Kong primary schools: implications for classroom listening. International Journal of Disability, Development and Education, 52, 345360. Counter, S. A. (1986) Audiological screening of Amerindians of the Surinam rainforest. Scandinavian Audiology, 15, 5764.

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

330

B. McPherson et al.

Downs, M. P. (2004) Contribution of mild hearing loss to auditory language learning problems. In: Auditory Disorders in School Children. The Law, Identication, Remediation, 4th edn (eds R. J. Roeser & M. R. Downs), pp. 233248. Thieme, New York, NY, USA. Frank, T., Greer, A. C. & Magistro, D. M. (1997) Hearing thresholds, threshold repeatability, and attenuation values for passive noise-reducing earphone enclosures. American Industrial Hygiene Association Journal, 58, 772778. Gell, F. M., White, E. M., Newel, K., Mackenzie, I., Smith, A., Thompson, S. & Hatcher, J. (1992) Practical screening priorities for hearing impairment among children in developing countries. Bulletin of the World Health Organization, 70, 645655. Givens, G. D. & Elangovan, S. (2003) Internet application to tele-Audiology nothin but net. American Journal of Audiology, 12, 5965. Gomes, M. & Lichtig, I. (2005) Evaluation of the use of a questionnaire by non-specialists to detect hearing loss in preschool Brazilian children. International Journal of Rehabilitation Research, 28, 171174. Harris, J. D. (1945) Group audiometry. Journal of the Acoustical Society of America, 17, 7376. Hatcher, J., Smith, A., Mackenzie, I., Thompson, S., Bal, I., Macharia, I., Mugwe, P., Okoth-Olende, C., Oburra, H., Wanjohi, Z., Achola, N., Mirza, N. & Hart, A. (1995) A prevalence study of ear problems in school children in Kiambu district, Kenya, May 1992. International Journal of Pediatric Otorhinolaryngology, 33, 197205. Henry, J. A., Flick, C. L., Gilbert, A., Ellingson, R. M. & Fausti, S. A. (2003) Reliability of computer-automated hearing thresholds in cochlear-impaired listeners using ER-4B canal phone earphones. Journal of Rehabilitation Research and Development, 40, 253264. Hind, S. E., Atkins, R. L., Haggard, M. P., Brady, D. & Grinham, G. (1999) Alternatives in screening at school entry: comparison of the childhood middle ear disease and hearing questionnaire (CMEDHQ) and the pure tone sweep test. British Journal of Audiology, 33, 403414. Hodgson, W. (1985) Basic Audiologic Evaluation. Krieger Publishing Co., Malabar, FL, USA. Hong, O. & Csaszar, P. (2005) Audiometric testing and hearing protection training through multimedia technology. International Journal of Audiology, 44, 522530. Jacob, A., Rupa, V., Job, A. & Joseph, A. (1997) Hearing impairment and otitis media in a rural primary school in South India. International Journal of Pediatric Otorhinolaryngology, 39, 133138. Jauhiainen, T. (2001) Hearing impaired children in developing countries. Scandinavian Audiology, 30 (Suppl. 53), 8384. Landis, J. R. & Koch, G. G. (1977) The measurement of observer agreement for categorical data. Biometrics, 33, 117129. McPherson, B. (2008) Audiology: a developing country context. In: Audiology in Developing Countries (eds B. McPherson & R. Brouillette), pp. 520. Nova Scientic, New York, NY, USA. McPherson, B. & Knox, E. (1992) Test-retest variability using the Liverpool screening audiometer in a eld environment. British Journal of Audiology, 26, 139141.

McPherson, B. & Olusanya, B. (2008) Screening for hearing loss in developing countries. In: Audiology in Developing Countries (eds B. McPherson & R. Brouillette), pp. 75105. Nova Scientic, New York, NY, USA. Madriz, J. J. (2001) Audiology in Latin America: hearing impairment, resources and services. Scandinavian Audiology, 30 (Suppl. 53), 8592. Marttila, T. I. (1986) Results of audiometrical screening in Finnish school children. International Journal of Pediatric Otorhinolaryngology, 11, 3946. Newton, V. E., Macharia, I., Mugwe, R., Ototo, B. & Kan, S. W. (2001) Evaluation of the use of a questionnaire to detect hearing loss in Kenyan pre-school children. International Journal of Pediatric Otorhinolaryngology, 57, 229234. North-Matthiassen, C. & Singh, S. A. (2007) The hearing prole among learners in schools in the Western Cape, South Africa. International Journal of Pediatric Otorhinolaryngology, 71, 113118. Nozza, R. J. (2001) Screening with otoacoustic emissions beyond the newborn period. Seminars in Hearing, 22, 415425. Olusanya, B. O., Okolo, A. A. & Ijaduola, G. T. (2000) The hearing prole of Nigerian school children. International Journal of Pediatric Otorhinolaryngology, 55, 173179. Olusanya, B. O., Swanepoel, D. W., Chapchap, M. J., Castillo, S., Habib, H., Mukari, S. Z., Martinez, N. V., Lin, H.-C. & McPherson, B. (2007) Progress towards early detection services for infants with hearing loss in developing countries. BMC Health Services Research, 7, 14. doi:10.1186/1472-6963-7-14. Parving, A. (1999) Hearing screening - aspect of epidemiology and identication of hearing impaired children. International Journal of Pediatric Otorhinolaryngology, 49 (Suppl. 1), S287S292. Parving, A., Srensen, M. S., Christensen, B. & Davis, A. (2008) Evaluation of a hearing screener. Audiological Medicine, 6, 115119. Rao, R. S. P., Subramanyam, M. A., Nair, N. S. & Rajashekhar, B. (2002) Hearing impairment and ear disease among children of school entry age in rural South India. International Journal of Pediatric Otorhinolaryngology, 64, 105110. Roeser, R. J. & Clark, J. L. (2004) Screening for auditory disorders. In: Auditory Disorders in School Children. The Law, Identication, Remediation, 4th edn (eds R. J. Roeser & M. R. Downs), pp. 96123. Thieme, New York, NY, USA. Sideris, I. & Glattke, T. J. (2006) A comparison of two methods of hearing screening in the preschool population. Journal of Communication Disorders, 39, 391401. Silman, S., Silverman, C. A. & Arick, D. S. (1994) Pure-tone assessment and screening of children with middle-ear effusion. Journal of the American Academy of Audiology, 5, 173182. Smith, A. W. (2001) WHO activities for prevention of deafness and hearing impairment in children. Scandinavian Audiology, 30 (Suppl. 53), 93100. Smith, A. W. (2008) Demographics of hearing loss in developing countries. In: Audiology in Developing Countries (eds B. McPherson & R. Brouillette), pp. 2150. Nova Scientic, New York, NY, USA. Smith, A., Hatcher, J., Mackenzie, I. J., Thompson, S., Bal, I., Macharia, I., Mugwe, P., Okoth-Olende, C., Oburra, H. & Wanjohi,

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331

Hearing screening using low-cost PC audiometer

331

Z. (1996) Randomised controlled trial of treatment of chronic suppurative otitis media in Kenyan schoolchildren. Lancet, 348, 11281132. Swart, S. M., Lemmer, R., Parbhoo, J. N. & Prescott, C. A. J. (1995) A survey of ear and hearing disorders amongst a representative sample of Grade 1 schoolchildren in Swaziland. International Journal of Pediatric Otorhinolaryngology, 32, 2334. Trucano, M. (2008) Quick guide to low-cost computing devices and initiatives for the developing world. An infoDev brieng sheet. infoDev / World Bank, Washington, DC.

Wootton, R. (2001) Telemedicine and developing countries-successful implementation will require a shared approach. Journal of Telemedicine and Telecare, 7 (Suppl. 1), 16. World Health Organization (WHO) (1999) WHO/CMB Workshop on Hearing Aids Service-Need and Technology Assessment for Developing Countries, Bensheim, Germany, 2426 November 1998. Authors, Geneva. Wright, D. C. & Frank, T. (1992) Attenuation values for a supra-aural earphone for children and insert earphone for children and adults. Ear and Hearing, 13, 454459.

2010 Blackwell Publishing Ltd, Child: care, health and development, 36, 3, 323331