Anda di halaman 1dari 4

Bird VS & VSO2 Ventilators

Portable Ventilator Modes o Control o Assist/Control o SIMV o CPAP

Typical Manufacturers Picture

Specifications
Ventilator Parameters Power Modes Control Assist/Control SIMV CPAP Tidal Volume Breath Rate Ranges Tolerance On/Standby

50-2000ml 2-80 bpm

Greater of: 10 ml or 10% Lesser of: 1 breath or 10% of Breath period Greater of: 2 lpm or 10% 0.5 lmp at a setting of 1 lpm 1 lpm at all other settings Greater of: 2cmH2O or 10% Greater of: 2cmH2O or 8% 3% from 21 to 50% 5% from 51 to 100%

Peak flow Sensitivity

10-140 lpm Off, 1-8 lpm

PEEP/CPAP Pressure Support %O2 (VSO2 only)

0-30 cmH2O Off, 1-60 cmH2O 21-100%

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: soma@somatechnology.com Rev 1 6/07

Bias Flow Sigh

10-20 lpm

1 lpm

On/Off 1 sigh every 1 breath period 100 breaths or 7 minutes x1 On/Off 6 seconds max On/Off 3 min max 20-130 cmH2O Ranges 5-120 cmH2O N/A N/A Greater of: 2 cmH2O or 10% +0%, -5% 10 cmH2O Tolerance 5-20 cmH2O 2 cmH2O 21-120 cmH2O 4 cmH2O 2-20 cmH2O 2 cmH2O 21-60 cmH2O 4 cmH2O

Manual Breath Remote Alarm (optional) Inspiratory Hold 100% O2 (VSO2 only) Over Pressure Relief Alarms High pressure Alarm Limit Low Pressure Alarm Limit Low Minute Volume Alarm Apnea Interval Back-up Breath Low O2 (VSO2 only) High O2 (VSO2 only) Vent Inoperative Alarm Silence Alarm Volume Monitors Airway Pressure Manometer

Off, 2-60 cmH2O

0.1 - 99.9 L 10-60 sec 12 bpm or set rate 35 psig 65 psig Red Indicator 60 sec max

Greater of: 15% or 30 ml 0.5 seconds Greater of 1breath or 10% of period 2 psig 2 psig N/A 1 second

50-70 dba at 1 meter 5 dba Ranges -20 120 cmH2O Tolerances Greater of: 2 cmH2O or 5%

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: soma@somatechnology.com Rev 1 6/07

Total Breath Rate (f) I:E Ration (I:E) Exhaled Minute Volume (VE) Peak Inspiratory Pressure (PIP) Mean Airway Pressure (MAP) Positive End Expiratory Pressure (PEEP) Exhaled Tidal Vol. (Vte) Inspiratory time (Ti)

0.250 bpm 99:1 1:99 0 99.9 L

Greater of: 1 bpm or 5% of period Greater of: 50 ms or 5% Greater of: 15% or the measured breath rate Greater of: 2 cmH2O or 5%

0-140 cmH2O

0-99 cmH2O

Greater of: 2 cmH2O or 10%

0-99 cmH2O

Greater of: 2 cmH2O or 10%

0 4,000 ml 0.01 99.99 sec

Greater of: 15% or 15 ml .05 seconds

Indicators Patient Effort Yellow Indicator Power Indicators On: Green indicator; AC: Green indicator Internal Battery Power: Green, Yellow, Red indicator Internal Battery Charge Status: Green, Yellow, Red indicator External Battery Power: Green, Yellow indicator External Battery Charge Status: Green, Yellow, Red indicator Pneumatic Specifications Oxygen Inlet VS Oxygen Bleed In: Max. 80 lpm or 0.5 PSI (.035 Bar) VSO2 Pressurized Oxygen: 40 to 60 PSI (2.8 to 4.2 Bar) Dimensions System VS System: 33.02cm x 27.94cm x 36.83cm (13.0in x 11.0in x 14.5in). VSO2 System: 33.02cm x 27.94cm x 36.83cm (13.0in x 11.0in x 14.5in). (W x H x D)

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: soma@somatechnology.com Rev 1 6/07

Docking Bay

VS System: 30.48cm x 2.54cm x 27.94cm (12.0in x 3.0in x 13.5in). VSO2 System: 30.48cm x 2.54cm x 27.94cm (12.0in x 3.0in x 13.5in) (W x H x D): VS System: 2.54 cm (1.0 in) VSO2 System: 2.54 cm (1.0 in) (W x H x D): VS System: 15 kg (33 lbs.); VSO2 System: 15.5 kg (34 lbs.); Docking Bay: VS System: 1.1 kg (2.5 lbs.); VSO2 System: 1.1 kg (2.5 lbs.). External Battery: VS System: 4.5 kg (10 lbs.); VSO2 System: 4.5 kg (10 lbs.)

External Battery

Weight

Soma Technology, Inc. acknowledges all registered trademarks of manufacturers listed. The technical data given in this publication are for general information and are subject to change without notice.

166 Highland Park Dr. Bloomfield, CT 06002 U.S.A. Tel: (860) 218-2575 Fax: (860) 218-2565 Email: soma@somatechnology.com Rev 1 6/07