Calibration Certificate

Instrument Description: Test Equipment Instrument Identification: Type of Instrument: Ax Force Transducer Sx Force Transducer SY Force Transducer MX Force Transducer MY Force Transducer Conditioner Identification: Model Number Ax 2310 Sx 2310 SY 2310 MX 2310 MY 2310 Readout Device Identification: Model Number Ax Labview 6036 card #2 Sx Labview 6036 card #2 Mx Labview 6036 card #2 N1 Labview 6036 card #2 N2 Labview 6036 card #2 Calibration Factors: Full Scale Output: Ax Sx SY MX MY 10 V 10 V 10 V 5 5 V V Amplified Output per Eng. Unit: .1 V/kip 2.00 V/kip 2.00 V/kip 0.05 V/kip in 0.05 V/kip in Serial Number: And PC And PC And PC And PC And PC Channel: 1 2 3 4 5 Serial Number: 072470 072363 72595 088779 088740 Gain: 3380 7700 7080 2460 2520 Excitation: 10 V 10 V 10 V 5 V 5 V Instrument Name: 5D-12-RED-02 5D-12-RED-02 5D-12-RED-02 5D-12-RED-02 5D-12-RED-02 Serial Number: 02 02 02 02 02 Instrument Range: 100 kip 20 kip 20 kip Certificate Number: UB-2006-07-24-06 RED LC Three-Channel Load Cell Location: SEESL

200 in 200 in

Certificate Number: UB-2006-07-24-06

Page 1 of 9

Reference Equipment Reference Instrument Identification: 1. Type of Instrument: Calibration Trace: Caltronix 2. Type of Instrument: Calibration Trace: Caltronix 3. Type of Instrument: Calibration Trace: NIST Traceable NEM-8115: 300k Load Cell 822/2650366-01 Calibration Factors: Full Scale Output: 1. 2. 3. 1. (update) (update) (update) Address: Ketter Hall SEESL University at Buffalo Buffalo, NY 14225 2. Address: Ketter Hall SEESL University at Buffalo Buffalo, NY 14225 3. Address: Northeast Metrology Corp. 2601 Genesee Street Buffalo, NY 14225-2916 Phone/Website: 716-827-3770 www.vantek-nem.com Accreditation: ISO/IEC 17025 Phone/Website: 645-2114 www.nees.buffalo.edu Accreditation: Unamplified Full Scale Output (update) (update) (update) Phone/Website: 645-2114 www.nees.buffalo.edu Accreditation: Amplified Output per Eng. Unit: (update) (update) (update) Ref Force Transducer N2 Fluke 8050A Instrument Name: Scopemeter Certificate Number: 3R394 Instrument Name: Dig Voltmeter Certificate Number: 3R395 Instrument Name: UB#300kip Certificate Number: NMC - 1044276 Serial Number: DM8680589 Cal. Date: 7/6/2005 Serial Number: 4237154 Cal. Date: 07/06/2005 Serial Number: LC300-01 Cal. Date: 6/21/2006 Cal. Exp. Date: 07/06/2006 Instrument Range: 300 kip (compression) Cal. Exp. Date: 6/21/2007 Cal. Exp. Date: 7/6/2006 Instrument Range: Instrument Range: N1 Fluke 133 Scopemeter

Reference Lab Information:

Certificate Number: UB-2006-07-24-06

Page 2 of 9

Calibration Information Calibration Procedure: Basic Description: AXIAL CALIBRATION: The two load cells (Scopemeter, Dig Voltmeter, and RED LC) were placed on top of the UB#300kip reference load cell in the Tinius Olsen machine. The load cells were loaded several cycles to allow the gains of the amplifiers to be adjusted to match the calibrated reference. Scopemeter and Dig Voltmeter were calibrated to be used as normal force references 1 and 2 respectively. SHEAR CALIBRATION: The load cells were set up as shown in Figure 1, then loaded. The gain of the shear conditioner was adjusted such that the shear reading matched that of Dig Voltmeter. MOMENT CALIBRATION: The load cells were set up as shown in Figure 2, then loaded. The gain of the moment conditioner was adjusted such that the moment reading matched that of Dig Voltmeter times the distance from the loading point to moment strain gages. Distance from moment gages to: top = 0.000 in bottom = 0.000 in

Certificate Number: UB-2006-07-24-06

Page 3 of 9

Calibration Data: Certificate Number: Graph: UB-2006-07-24-06 Normal Calibration Data

Normal Calibration

120 100 80 60 Series1 40 20 0 0 -20 20 40 60 80 100 120

Certificate Number: UB-2006-07-24-06

Page 4 of 9

Shear Calibration Data Graph:

Shear Calibration RED LC 2v/kip

SX

12

y = 0.9974x + 0.0176 10

6 Series1 Linear (Series1) 4

0 0 2 4 6 8 10 12

-2

Certificate Number: UB-2006-07-24-06

Page 5 of 9

Moment Calibration Data Graph:

Shear Red Load Cell-02 SY

12

y = 0.968x - 0.0105 10
2 R = 0.9996

Series1 Linear (Series1)

0 0 2 4 6 8 10 12

-2

Certificate Number: UB-2006-07-24-06

Page 6 of 9

Moment Red Load Cell MX

120

y = 5.1483x + 0.2964 100

80

60 Series1 Linear (Series1) 40

20

0 0 5 10 15 20 25

-20

Certificate Number: UB-2006-07-24-06

Page 7 of 9

Moment Red Load Cell MY


20

0 0 5 10 15 20 25

-20

-40 Series1 Linear (Series1) -60

-80

-100 y = -5.4663x - 0.7036 -120

Certificate Number: UB-2006-07-24-06

Page 8 of 9

Calibration Factors: Ax 0.1 V/kip Sx 2V/kip SY 2V/kip MX .05V/kip/in MY .05V/kip/in

Comments:

Personnel Identification: Name: Chris Budden Calibration Period: Cal. Date: 7/21/2006 Cal. Exp. Date: 7/21/2007 Company: UB Signature: Date: 7/21/2006

Certificate Number: UB-2006-07-24-06

Page 9 of 9