Focal Spot Size of -Focus Tubes EN 12543-5:1999: Revision in 2008 possible! Harmonization with ASTM recommended Old: Test object: sphere or wire cross with diameter between 0,9 bis 1,1mm ( 0,01mm) Focus range 5 bis 300 m New: New test object: duplex wire of EN 462-5 or ASTM 2002. Extended standard for and nano focus tubes below 300 m spot Usage of smaller wire diameters than 1 mm. Detectors: Film, CR, DDA, intensifier with computer, Image processing with profile function with line integration, perpendicular to the wire or sphere direction
2
Round wire
simple adjustment!
Result of modeling: Edge transmission increases the unsharpness in the low intensity range
EN 12543-5 EN 12543-4
EN12543-5: Focus size = 2x (90%-50%) = EF +GH EN12543-5: Focus size = 2x (90%-50%) = EF +GH
Background
100 90 80 70 Intensity 60 50 40 30 20 10 0 -30 -20 -10 0 mm 10 20 30 Gauss edge Gauss line FWHH 10-90 %
If the focal spot (line distribution) has a shape of a Gaussian distribution, the 10-90 % edge width corresponds to the full width at half height of the profile.
Background
120
100
Two lines, separated by the full width at half height, show a dip of about 7.3 %.
20
Background
120
100
The Lorentzian distribution is typical for focal spots (line profile) with long range wings of about 50 times the half width at half height. The 10-90 % edge width corresponds to 10 30 % width of the spot line distribution depending on the zero line determination. Typical focal spot line distributions are between Gaussian and Lorentzian.
20
J. Goebbels, Y. Onel AG-CT 2008-05-09 Measurements for focal spot size measurement Unit I Offset Measurement: average over 100 exposures 1 sec Gain Measurement: average over 100 exposures 1 sec Duplex wire exposures: average over 60 exposures 1 sec 225 kV Micro focus tube PE Detector 2048 x 2048 Pixel (0,2 mm)2 SDD: 1147,5 mm SOD: ca. 10-15 mm , Distance of sample as near as possible in front of the tube window! Magnification about 80. 55 kV 100 A, no filter 3227.his Complete Duplex wire 3228.his 3229.his Element 1D Element 1D vertikal Position
9
Software ISee was used inclusive the calibration function. The calibration measure is the 800 m width of the wire 1 of E 2002 IQI
10
Spot width: edge profile as taken corresponding the previous slide presentation.
50 m 30 m 30 m 67 m
Measured Lorentz Gauss 50%
2*90-50% edge
400 350 300 Intensity 250 200 150 100 50 0 0 -50 mm 0,05 0,1 0,15
The first derivative of the edge function is the measured line spread function of the focal spot. This corresponds to a slit camera exposure. 0,2 0,25
12
7000 6000
Imax/2 192,5
5000 4000 Reihe1 3000 2000 1000 0 0 0,05 0,1 pixel 0,15 0,2 0,25
Allocation of values between edge profile and line profile not useful. The asymmetry of the measured profile disturbs the allocation considerably.
grey values
450 400 350 300 Intensity 250 200 150 100 50 0 0 -50 mm 0,05 0,1 0,15 0,2 0,25 Measured Lorentz Gauss 50%
13
Ewert et al. June 2008
The distance between the duplex wires is 10% less than its diameter.
14
Ewert et al. June 2008
Edge method
15
16
Measurement with HD-CR (DRR) immediately after DDA-Measurement (HD-IP) 2*90-50% 33 m Edge profile
Edge profile
20 m pixel size
17
18
2*90-50% = 47 m DDA of PE
19
20
Edge transmission when measured with thin wires Measurement possible Measurement after new calibration!
2*90-50% 18 m
Edge transmission
21
Unit II 100 kV Micro focus tube (Transmission target) CCD Camera + Fiber optic ( 2:1) + GdOS Szintillator Image integration over 20 sec, no corrections SDD: 245,0 mm SOD: ca 3 mm M about 66. 85 kV 90 A, no filters en462.his Element 4D (0,4 mm)
6 m 400 m
22
Conclusion:
Duplex wire EN 462-5 / E 2002 is suitable for the measurement of the focal spot size of micro focus tubes. The limitation to 1 mm wire is not necessary. The standardized method 2*(90%-50%) provides stable values also if edge transmission is observed or focal spots are asymmetrical. The measurement of asymmetric spots with wide wings is considered. The full width at half height of line profiles (calculated as derivative of edge profiles) are almost equal to the edge distribution if calculated by 2*(90%-50%) and close to Gaussian distributions. This applies in many cases for the central part of the spot distribution. Since the method is now applicable with the thin wires of EN 4625 or E 2002 IQI the lower limitation to 5 m is not required anymore.