Slide 1 of 43
April 2007
Pharmaceutical Development
Analytical Method Development
Presenter:
WHO expert
Slide 2 of 43
April 2007
Slide 3 of 43
April 2007
Interchangeability (IC)
INTERCHANGEABILITY (IC) OF MULTISOURCE FPPs = (ESSENTIAL SIMILARITY WITH INNOVATOR FPP) =
IC = PE + BE
|
Slide 5 of 43
April 2007
Pharmaceutical equivalence
FPPs meet same or comparable standards (e.g., marketing authorization, analytical methods)
Same API (chemical and physical equivalence) Same dosage form and route of administration Same strength Comparable labeling
Slide 6 of 43
April 2007
Prequalification requirements
Analytical method validation is required by WHO for the prequalification of product dossiers. Non-compendial ARV APIs and FPPs were/are tested with methods developed by the manufacturer.
Analytical methods should be used within GMP and GLP environments, and must be developed using the protocols and acceptance criteria set out in the ICH guidelines Q2(R1)
Slide 7 of 43
April 2007
Validation of analytical procedures: text and methodology Q2(R1) ICH Harmonized Tripartite Guidelines, (2005)
http://www.ich.org/LOB/media/MEDIA417.pdf
Slide 8 of 43
April 2007
General requirements
Qualified and calibrated instruments Documented methods Reliable reference standards
Qualified analysts
Sample selection and integrity Change control
|
Slide 9 of 43
April 2007
68.26%
95.46%
Slide 10 of 43
April 2007
average = mean
LSL
Slide 11 of 43
April 2007
Capable process
Almost all the measurements of a stable process fall inside the specification limits
USL LSL Cp 6 8 10 12
64 ppm
http://www.itl.nist.gov/div898/handbook/pmc/section1/pmc16.htm
|
Slide 12 of 43
April 2007
Slide 13 of 43
April 2007
Pharmaceutical
To develop a stable and reproducible formulation for the manufacture of bioequivalence, dissolution, stability and pilot-scale validation batches
Methods
To understand the profile of related substances and to study stability and start measuring the impact of key product and manufacturing process parameters on consistent FPP quality
To optimize, scale-up, and transfer a stable and controlled manufacturing process for the prequalification product
To be robust, transferable, accurate, and precise for specification setting, stability assessment, and QC release of prequalified batches
Slide 15 of 43
April 2007
Assay + + + +
+ + +
+
April 2007
+
+ + +
+ +
+ + +
Slide 16 of 43
Trueness
Precision
(Random errors)
Systematic errors
Intra-assay variability
Intra-laboratory variability
Inter-laboratory variability
Repeatability
Intermediate precison
Reproducibility
Source: ISO. 1994. ISO 5725 1-6: Accuracy (Trueness and Precision) of Measurement Methods and Results. ISO, Geneva, Switzerland.
Slide 17 of 43
April 2007
Precise
Accurate
April 2007
Slide 18 of 43
LA, %
50 70 80 100 120 130 Mean
Nevirapine, mg Added 0.499 0.703 0.796 1.001 1.211 1.299 0.918 Recovered 0.495 0.701 0.795 1.005 1.209 1.296 0.917
The data show that the recovery of analyte in spiked samples met the evaluation criterion for accuracy (100 2.0% across 50130% of target concentrations).
|
Slide 19 of 43
April 2007
% Recovery
Red line: LA
|
Slide 20 of 43
April 2007
The precision (VARIABILITY) of an analytical procedure is usually expressed as the standard deviation (S), variance (S2), or coefficient of variation (= relative standard deviation, RSD%.) of a series of measurements. The confidence interval (CI) should be reported for each type of precision investigated.
PRECISION
|
Slide 21 of 43
April 2007
Repeatability expresses the precision (spread of the data, variability) under the same operating conditions over a short interval of time. Repeatability is also termed intra-assay precision.
REPEATABILITY
|
Slide 22 of 43
April 2007
3
4 5 6 Mean
57497
57617 57778 57231 57649
0.299
0.300 0.301 0.298 0.300
The repeatability precision obtained by one analyst in one laboratory was 1.25% RSD for the analyte and, therefore, meets the evaluation criterion of RSD 2%.
STD
RSD 95% CI
|
257
0.4% 270
April 2007
0.0013
0.4% 0.0014
Slide 23 of 43
Intermediate precision expresses within-laboratories variations. #1, #2 and #3: different days, different analysts, different (manufacturing) equipment, etc.
Reproducibility expresses the precision between laboratories #1, #2 and #3 (collaborative studies, usually applied to standardization of methodology). (Transfer of technology)
Intermediate precision or Reproducibility
|
Slide 24 of 43
April 2007
Assay (mg/5ml) 51.7 52.6 51.9 53.0 52.5 52.3 52.7 52.4 0.49 0.9% 0.51
April 2007
Combined values Mean STD RSD 95% CI 52.5 0.48 0.9% 0.31
Slide 25 of 43
Specificity (selectivity)
Specificity is the ability to assess unequivocally the analyte in the presence of components, which may be expected to be present. Typically these might include impurities, degradants and excipients. An example of specificity criterion for an assay method is that the analyte peak will have baseline chromatographic resolution of at least 2.0 minutes from all other sample components
Stability indicating analytical methods should always be specific.
|
Slide 26 of 43
April 2007
Slide 27 of 43
April 2007
Slide 28 of 43
April 2007
SPECIFICITY degradants
Stress A (%) *
Initial
Acid Peroxide
100.0
99.3 99.8
All others
100.0
Purity angle 0.040 0.105 0.725 0.045 0.120 1.040 0.060 0.110 0.690 NA
Purity threshold 0.280 0.380 1.630 0.280 0.410 1.610 0.270 0.360 1.250 NA
There were no peaks in the placebo chromatogram at the retention times of nevirapine (N), methylparaben (MP) and propylparaben (PP) peaks.
*Sum of N, MP and PP peak areas. The three ingredients can be assessed in the presence of (nonexpected) degradants. The peaks are homogeneous and pure. The method is selective, specific and stability-indicating.
|
Slide 29 of 43
April 2007
Peak B LOQ
Peak A LOD
Baseline noise
Slide 30 of 43
April 2007
3
4 5 6 Mean STD RSD
4196
4303 3932 5238 4242 548 12.9%
7950
8166 7847 8415 7952 402 5.1%
3275
3464 4008 4702 3867 551 14.3%
8292
8050 8368 8284 8113 238 2.9%
Conc. (g/ml)
Conc. (%w/w)
|
0.086
0.017
0.171
0.033
0.107
0.019
0.214
0.039
Slide 31 of 43
April 2007
The limit of quantitation (LOQ) is defined as the lowest concentration of an analyte in a sample that can be determined with acceptable precision and accuracy under the stated operational conditions of the method. The ICH has recommended a signal : noise ratio (SNR) of 10:1.
The LOQ was 0.171 g/ml (Impurity 1) with a signal:noise ratio of 11.3. The RSD for six injections of the LOQ solution was 2%.
|
Slide 32 of 43
April 2007
Linearity
Measured mean Real mean
Precision
Linearity expresses differences in precision at different points of a given range. The linearity of an analytical procedure is its ability (within a given range) to obtain test results, which are directly proportional to the concentration (amount) of analyte in the sample.
Slide 33 of 43
April 2007
Assay mean
Slide 34 of 43
April 2007
Slide 35 of 43
April 2007
Impurity: from the reporting level of an impurity to 120% of the specification. (Unusually potent or toxic impurities, LOD and LOQ should be
commensurate with ICH requirement.)
If assay and purity are performed together as one test and only a 100% standard is used, linearity should cover the range from the reporting level of the impurities to 120% of the assay specification
Slide 36 of 43
April 2007
#1 First measurements
#2, #3, #4, n Series of measurements of the same sample within a relatively short period of time.
Slide 37 of 43
April 2007
3
4 5 10 15 20 25
|
71960
72352 71573 72322 72310 72312 72670
April 2007
0.2%
-0.4% 0.7% -0.3%
An analytical solution prepared from Nevirapine 50mg/5ml Oral Suspension was spiked with Impurity-1 at specification level and stored in a capped volumetric flask on a laboratory bench at uncontrolled room temperature under normal lighting conditions for 25 hours.
Conclusion: the stability of the -0.3% analytical solution of Impurity-1 is -0.3% not a source of variation.
-0.8%
Slide 38 of 43
Slide 39 of 43
April 2007
Robustness
Method parameter STP Flow Wavelength Variation of mobile phase Variation -10% 10% -5nm +5nm -2% +2% -5oC +5oC -0.3 +0.3 tR Impurity 1 0.83 0.83 0.84 0.82 0.83 0.80 0.84 0.82 0.83 0.83 0.83 Impurity 2 1.80 1.81 1.82 1.81 1.81 1.89 1.76 1.80 1.81 1.81 1.80
Column temperature
pH
|
Slide 40 of 43
April 2007
Limits of detection and quantitation Accuracy or recovery from rinsate ( 80%), swabs ( 90%), and process surface ( 70%) Range (lowest level is at least 2x higher than LOQ)
|
Slide 41 of 43
April 2007
Validation should demonstrate that the analytical procedure is suitable for its intented purpose. HPLC systems and method validation deserves special attention during the assessment of dossiers for prequalification.
|
Slide 42 of 43
April 2007
THANK YOU
Slide 43 of 43
April 2007