Modern HPLC
Last 20 years
Nothing really new, but by returning to the basic theory of chromatography, even better columns are on the market: smaller particle sizes which yield faster separations, but require hardware to withstand higher pressures.
SHIMADZU
Solutions for Science Since 1875
What is Chromatography?
Separation of a mixture into individual components. The separation uses a Column (stationary phase) and Solvent (mobile phase). The components are separated from each other based on differences in affinity for the mobile or stationary phase. The goal of the separation is to have the best RESOLUTION possible between components.
SHIMADZU
Solutions for Science Since 1875
SHIMADZU
Solutions for Science Since 1875
Hardware: pumps, injector, detector Column: particle diameter, column size, packing materials, and the dreaded equations Our seminar will focus on the contribution of each factor to perform separations.
SHIMADZU
Solutions for Science Since 1875
SHIMADZU
Solutions for Science Since 1875
Column Types
Reversed-Phase LC
Nonpolar stationary phase: C8, C18 Polar mobile phase: Water, ACN, Methanol The MOST polar compound comes out first
SHIMADZU
Solutions for Science Since 1875
Rugged, general purpose, highly retentive Less retentive, used mostly for peptides & proteins Greater selectivity than alkyl-bonded Moderate retention, normal & rev. phase Weak retention, good for carbohydrates
The cyano column with a high polarity mobile phase (Water/MeOH) will act as a reversed phase column.
SHIMADZU
Solutions for Science Since 1875
Pore size: 100-120 A is typical Surface area: 300-350 m2/g Carbon load: 9-12% for C8, 16-20% for C18
Higher carbon load = better resolution but longer run times Lower carbon load = shorter run times, but may change selectivity vs. higher carbon load SHIMADZU
SHIMADZU
Solutions for Science Since 1875
Capacity Factor,
The relative degree to which an analyte component is delayed as it is eluted through a given system (retentivity).
k = (Vr - V0)/V0 = (tr - t0)/t0
Where Vr = peak retention volume; V0 = column void volume tr = peak retention time; t0 = peak void time
The larger the k, the later the analyte elutes after the void.
SHIMADZU
Solutions for Science Since 1875
0.212
90% 90%
0.472
80% 80%
1.127
2.813 7.666
50% 50%
SHIMADZU
Solutions for Science Since 1875
Temperature Effect on
45C
40C 35C
30C
25C
20C
SHIMADZU
Solutions for Science Since 1875
Dual Piston
Two pistons with equal volume (10 L each) During each stroke, 10 L is delivered Best for low flow rates (< 1 mL/min) Little to NO pulsation, so its ideal for pulse sensitive detectors like RID and CDD
SHIMADZU
Solutions for Science Since 1875
SHIMADZU
Solutions for Science Since 1875
Needle-in-the-flowpath autosampler
Sample loop and needle are a single piece of tubing Loop and needle are cleaned during the run Metering pump draws sample very precisely
Advantages: no sample loss, low carryover Disadvantages: higher cost, more delay volume for gradient
SHIMADZU
Solutions for Science Since 1875
Sample Loading
Sample Injection Everything drawn into the needle goes to the column.
SHIMADZU
Solutions for Science Since 1875
Forced air
Column is in a large chamber with air circulation Better temperature equilibration Room for column switching valves
SHIMADZU
Solutions for Science Since 1875
40 C
35 C 30 C 25 C 20 C
SHIMADZU
Solutions for Science Since 1875
UV-VIS
SHIMADZU
Solutions for Science Since 1875
HPLC Detectors
Refractive Index
Best results when RI of samples is very different from RI of mobile phase Flow cell is temperature controlled with a double insulated heating block.
SHIMADZU
Solutions for Science Since 1875
SHIMADZU
Solutions for Science Since 1875
As the refractive index changes, the image on the photodiode is deflected or unbalanced, and the difference in current to the photodiode is measured.
SHIMADZU
Solutions for Science Since 1875
Isocratic system
Same mobile phase concentration throughout the separation Use 1 pump and pre-mix solvents Use 1 pump and a valve for 4 different solvents Use 2 pumps and vary the amount coming from each pump
SHIMADZU
Solutions for Science Since 1875
Isocratic Separation
SHIMADZU
Solutions for Science Since 1875
Isocratic Separation
To Column
ABCD
Isocratic Separation
To Column
ABCD
To Column
ABCD
REQUIRES degassing
SHIMADZU
Solutions for Science Since 1875
To Column
ABCD
REQUIRES degassing
SHIMADZU
Solutions for Science Since 1875
Next: Troubleshooting and How to Take Care of Your Column and HPLC System
SHIMADZU
Solutions for Science Since 1875
HPLC Troubleshooting
Pressure: too much or too little Leaks: pump, autosampler, detector Reproducibility: pump, autosampler Column Care: Flushing and equilibration
SHIMADZU
Solutions for Science Since 1875
Pump Troubleshooting
SHIMADZU
Solutions for Science Since 1875
Pump Troubleshooting
High Pressure
Outlet frit may be blocked with particles from mobile phase or seal material
Leaks
SHIMADZU
Solutions for Science Since 1875
Pump Troubleshooting
A 1o shift in temperature can result in a 1-2% shift in retention time Avoid drafty locations in the lab Use a column oven when possible
SHIMADZU
Solutions for Science Since 1875
Autosampler Troubleshooting
High Pressure
Particulates from mobile phase, sample, pump may be trapped in the inlet tubing or valve
Leaks
Tighten fittings properly and dont exceed the pressure limit of the autosampler
SHIMADZU
Solutions for Science Since 1875
Autosampler Troubleshooting
Area % Reproducibility
Always degas rinse phase, and use some volume of liquid for rinsing to keep all flow paths in the valves full of liquid Make sure the needle stroke is deep enough to draw sample from the vial Check for leaks on the valve fittings, and the connection to the column inlet
SHIMADZU
Solutions for Science Since 1875
Detector Troubleshooting
Spiky Baseline
Put some restriction on the cell outlet, but not too much! Tubing with 0.005 i.d. is fine.
Leaks
Dont exceed the pressure limit of the cell Use the proper fittings and tighten appropriately
SHIMADZU
Solutions for Science Since 1875
Column Care
Follow MFRs recommendations for solvent compatibility, flow rate, and pressure limits Filter samples when possible
Particulates will build up on the inlet frit over time Connect the outlet to waste, NOT inline with the detector to prevent further contamination
Troubleshooting Summary
Throw away bad parts and columns. Leaks do not fix themselves. If it doesnt pass, you must degas.
SHIMADZU
Solutions for Science Since 1875