What is pacemaker
Device capable of generating artificial
pacing impulses and delivering them to
the heart
Internalimplanted
pacemakers
Permanently
in patients whose
SA nodes fail;
In patients who suffer from permanent
heart block due to hear attack.
The pulse generator must have a power
source lasting for years
External pacemakers
Have a externally worn pulse generator
connected to the electrodes on the
myocardium.
For patients with temporary heart
problems
Also for temporary management of
certain arrhythmias in patients
postoperative time/cardiac surgery
pacemakers
Pacemaker Components
Pulse Generator
Electronic Circuitry
Lead system
Pulse Generator
Lithium-iodine cell is the current standard
battery
Advantages:
Long life 4 to 10 years
Output voltage decreases gradually with time
making sudden battery failure unlikely
Lead Systems
Endocardial leads which are inserted
using a subclavian vein approach
Actively fixed to the endocardium using
screws or tines
Unipolar or bipolar leads
synchronous)
Synchronous-Non
competitive ventricular
programmed
Non
competitive
pacemakers
are
atrial
ventricular
programmed
or
programmed.
Ventricular programmed are connected to
ventricles via electrodes.
There are 2 types: Demand and standby
type.
Both demand and standby type are able
to sense presence or absence of R wave.
R wave inhibited
Output of the R wave inhibited unit (demand)
pacemaker-demand
is suppressed when there are natural R wave.
In case standstill or rate falls below the
preset rate of the pacer, the unit
automatically provides and output to pace
the heart.
Hence pace on demand. demand pacemaker
Demand pacemaker , in the absence of the R
wave, automatically functions in fixed rate
mode.
Demand type..
However if during this period natural beat
occurs in the ventricles, the feedback
circuit, senses the QRS part of the ECG
signal and amplifies it.
This signal is then used to reset the
timing circuit.
Again it awaits for a fixed interval, before
producing the next impulse.
This process repeats.
Atrial programmed
Block
diagram-Refer class notes
pacemakers
Atrial programmed
v1-is the voltage detected by atrial electrodes.
pacemakers
This is amplified and passed through a gate to
monostable multivibrator (MMV) which
produces a pulse v2 of 120ms
(millisecond)duration.
120ms is the approximate delay of the AV
node.
The MMV of pulse 500ms is also triggered by
atrial pulse.
It produces v4, which causes the gate to block
any signals form the atrial electrodes for a
period of 500 ms following the contraction.
The falling edge of the v2 pulse triggers MMV
of 2ms duration.
Atrial programmed
pacemakers
Rate responsive
pacing
The pacemaker is programmed to control