AGENDA
Company Profile
Disposable & Durable Product
Infusion Pump TE-172
Syringe Pump TE-331
Practice
TERUMO CORPORATION
Multinational Healthcare Company
yang memproduksi alat kesehatan
Histor
y
1921 (Terumo Corporation-Tokyo): Clinical
Thermometers
1960an: Disposable Products: Syringe, Needles,
Blood Collection Tubes
1970an: Oxygenators (USA Factories, Europe)
1980an-Sekarang: Cardiovascular Products,
Global Market
TERUMO FACTORIES
BELGIUM,
EUROPE
HEAD QUARTER
MARYLAND,
USA
TERUMO PRODUCTS
SYRINGE & OTHERS
HOME CARE
BLOOD TRANSFUSION
MEDISAFE
BLOOD BAG
DIGITAL THERMOMETER
TUBE SEALER
TERUGLOVE
SURGICAL GLOVES
URINE BAG
BLOOD PRESSURE
HAND SEALER
KATETER INTRA
VENA
Surflo/ SR*OX
Surflo-W/ SR+DM
Volumetric/ TS*M
1. PT.MBS
Binasukses )
2. PT.APL
Mensa
Product
concept
High accuracy(+/-3-5%
High performance
in compact size
Blood transfusion
Inflow
Inflow
stops
Outflow
1
Outflow
2
Outflow
stops
Pluse
flw adj.
Conventional
Midpress
opens freely
opens not
freely
depresses
completely
depresses
halfway
Stable vol.delivered thanks to
consistent finger stroke & secured
certain volume.
Features
Easy
to use / simple maintenance care in a compact
model
Safety
- Selectable occlusion detection pressure
- Drip sensor (option)
Body weight mode(TE-172)
Computer interface(TE-171C, TE-172C)
Features - Easy to
use
Features - maintenance
Solid aluminum die cast frame
Drip proof construction
History function shock sensor inside
Battery cover
Model
Accuracy
Range
TE-171/172
+/-3%
0.1-1200ml/h
Transfusion
Yes
IV sets Terufusion IV set
type E for pump
Terufusion IV set
for pump
STC-508
+/-5%
1-500ml/h
No
Terufusion IV set
type E for pump
Terufusion IV set
for pump
TE-112
+/-10%
1-300ml/h
No
Terufusion IV sets
for pump
Normal IV sets
Product Line-up
F
TE-172
U
N
C
TE-171
Ward
STC-508
O
N
TE-112
ICU/CCU, ER
Neonates
Flow rate
Volume limit
Input the flow rate and then the pump starts infusion controlling the
motor speed based on the calculation of inner diameter and outer
diameter.
TE-171/172
TPN solution
(Dense)
Drop
sensor
Volume
infused
TE-112
Volume infused is
affected by the solution
viscosity.
--> Needs
compensation
Point to NOTE
sectiona
l view
Solution control
pumps need
Case 1)
Case 1)
with other
Infusion set
sectional
view
Volume to be infused
After 1h operation
Caution
Load
Guide
Air
detection
Finger
section
TS*PA
Type-E set
Caution
Air detector
Occlusion
detection
(Upper)
Occlusion
detection
(Lower)
Occlusion Alarm
When pump stops with occlusion alarm,
Internal pressure in I.V. set is still high !
First, close the clamp of the set and open the door
and then treat necessary actions.
During the solution is kept in I.V. set under
high pressure, the solution may enter the
patient
in a flash when the occlusion condition is
released.
Caution
Broken by shock
The
The actual
actual contained
contained volume
volume in
in
solution
solution bag
bag
Based on Japanese MOH,
500