Easy Quick Manual

Produced by International Clinical Application Team


128BW Quick Manual

Image mode control


- B : Allows you to switch to B mode from scanning mode.
- BB : Allows you to choose either left or right image in the dual B mode.
- M :This button is to turn the M Mode on.
- HAR : Allows you to view the Harmonic Image in B mode. Each probe type decides whether
Harmonic mode is applied or not.
- 3D : Displays images in three dimensions.
- FRE : You can change the probe frequency according to the scanning areas in the probe
where Multi frequency is available.

Menu control
- Report : Reports the outcomes on the screen after measuring all variables.
- Set up : Gets the System set up.
- Application : Allows you to change Application and set parameters probe-by-probe.
Scanning control
- Depth : Adjusts the depth of scanning.
- Zoom : Magnifies scanning areas.
- Focus : Moves the currently indicated focus upward or downward.
- Direction : Shifts the scanning direction of probe to right or left with no probe movement.
- Apex : Shifts the scanning direction of probe to up or down with no probe movement.
- Focus number : Adjusts the number and position of Focusing Points. By pressing the soft
button [1] and [2], you can set the Focusing Point to a desired area or delete it.

Tackball function
- Measure : The menu for complicated measurements and calculations appears depending on
the scanning area at the moment. (set in the Application in Probe menu)
- Indicator :display an arrow-shaped indicator.
- Body Marker :This button is to display the list of Body markers.
- Caliper : Allows you to measure some basic variables such as velocity, distance,
circumference, and area according to the scanning mode at the moment.
- Document : Allows you to attach the explanation about scanning in letters or other symbols
to an image.
Menu control
- Probe : Allows you to shift the kinds of probes.
- SonoView : Executes Sonoview, an image filing software to file the scanned images of patient.
- Image save : Saves the desired image in file and stores it in the memory unit.
- Application : Allows you to change Application and set parameters probe-by-probe.
- ID : display a Setup dialog box to setup the parameters related to the system.
- End exam : Ends the measurement or diagnosis of patients.
TCG control : You can control the TGC values by depth using eight slides.

Freeze :This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.

Print :This button is to print out the image viewed on the current Image Area into a
configured echo printer.
Image Filing : This button is to operate SonoView, which is an Image Filing program.
Store : The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView.
SA PICO Quick Manual



Scanning control
- 2D : Allow you to switch to 2D mode from scanning mode.
- Dual : Displays Multiple 2D modes.
- C : Displays Color Doppler mode.
- PD : Displays Power Doppler mode.
- PW : Displays PW mode.
- M : Displays M mode.
- Gain : You can control Gain 2D,C,PW image mode.

Menu control
- ID : display a Setup dialog box to setup the parameters related to the system.
- Calculation : The menu for complicated measurements and calculations appears depending
on the scanning area at the moment. (Application)
- Measure : Allows you to measure some basic variables such as velocity, distance,
circumference, and area according to the scanning mode at the moment.
- Set up : Gets the System set up.
- Clear : This button is to delete Text, Indicator, Body Mark, and measurement results on the
image.
Probe : Allows you to shift the kinds of probes.
Image Controls
- Focus : Moves the currently indicated focus upward or downward.
- Depth : Adjusts the depth of scanning.
TCG control : You can control the TGC values by depth using eight slides.
Report : Reports the outcomes on the screen after measuring all variables.

Freeze :This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.

Print :This button is to print out the image viewed on the current Image Area into a
configured echo printer.

Store :The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView.
SonoView : This button is to operate SonoView, which is an Image Filing program.
The Shortcut Key functions assigned to the Alpha-numeric keyboard
- A (Application) : Allow you to change Application and set parameters probe-by-probe.
- B (Body mark) : Records the figures describing the scanning areas or posture along with
images. You can change the shapes of Body Mark and Probe, using Change key.
- I (Indicate) : Moves or indicates the arrow while you are explaining the diagnosis regardless
of the measurement.
- M (M-line) : Line appears on 2D image. If changed to M mode or D mode, it appears as
M Line and Sample Volume Line, respectively.
- Q (3d mode) : Displays images in three dimensions.
- T (Text) : Allow you to attach the explanation about scanning in letters or other symbols to an
image.
- X (End exam) : Completing measurement and diagnosis of patients.
- Z (Zoom) : Magnifies scanning areas.
SA8000SE Quick Manual


Image mode & Gain control
- B : This button is to shift the Image Mode to the 2D Mode. Since this system includes basically
2D images except for some special cases, the user cannot turn the 2D off when desired.
- C : This button is to turn the Color Mode on. While the color mode is on, one more press of
this button turns it off.
- D : This button is to turn the PW Spectral Doppler Mode on. While the PW Spectral Doppler
Mode is on, one more press of this button turns it off.
- PD : This button is to turn the Power Doppler Mode on.
- M : This button is to turn the M Mode on.
- Gain :You can control Gain each other image mode.

Image Controls
- Focus : Up-down switch. The user can make the focusing point go up and down.
- Depth : The user can make the view angle.
- Zoom : Change the size of the Zoom box by controlling the switch. You can select images
by pressing the [Set] button.
- Freq : This button is to change the probe frequency, when you use a probe that supports
Multi Frequency.
Color & Doppler image controls
- Sound : This knob is to control the sound volume.
- Angle : This knob is to control the angle of Sample Volume in Spectral Doppler Mode.
- Power : Turning this knob adjusts the Acoustic Power of the system.
- Base line : This switch is to control the Baseline in Spectral Doppler Mode or Color Doppler
Mode. (including Power Doppler)
- Speed : This switch is to control the Sweep Speed in Spectral Doppler Mode or M Mode.
- Scale : This switch is to control the speed of blood flow or frequency range in Spectral
Doppler Mode.
- Sensitivity : control the Sensitivity values in Color Doppler (Power Doppler) Mode.
- Filter : control the Wall Filter values.
- Steer : This button is to alter the angle of ROI (Region of Interest) in Color Doppler Mode.
This operation is only available for Linear Probe.

TCG control : You can control the TGC values by depth using eight slides.
Measurement and Annotation Controls
- Distance : This button is to measure distance.
- Area : This button is to measure circumference and area.
- Volume : This button is to measure Volume.
- Calc : display some menus that can be possibly measured in the selected Application or Image
Mode.
- Body Marker : This button is to display the list of Body markers.
- Indicator : display an arrow-shaped indicator.
- Text : This button is for a user to input a text on the image.
- Clear : This button is to delete Text, Indicator, Body Mark, and measure results on the image.
Report :This button is to display one of the Report programs, which includes measurement
results of the currently selected Application.
Setup Controls
- Probe : display a dialog box to select or change the probe and Application.
- ID[Patient] : display a Setup dialog box to setup the parameters related to the system.
- End Exam : This button is to finish the diagnosis on the current patient and reset the related
data.

Freeze : This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.

Print : This button is to print out the image viewed on the current Image Area into a
configured echo printer.

Store : The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView.
SonoView : This button is to operate SonoView, which is an Image Filing program.
SA8000 Quick Manual


Image mode & Gain control
- B : This button is to shift the Image Mode to the 2D Mode. Since this system includes basically
2D images except for some special cases, the user cannot turn the 2D off when desired.
- C : This button is to turn the Color Mode on. While the color mode is on, one more press of
this button turns it off.
- D : This button is to turn the PW Spectral Doppler Mode on. While the PW Spectral Doppler
mode is on, one more press of this button turns it off.
- PD : This button is to turn the Power Doppler Mode on.
- M : This button is to turn the M Mode on.
- Gain :You can control Gain each other image mode.

Image Controls
- Focus : Up-down switch. The user can make the focusing point go up and down.
- Depth : The user can make the view angle.
- Zoom : Change the size of the Zoom box by controlling the switch. You can select images
by pressing the [Set] button.
- Freq : This button is to change the probe frequency, when you use a probe that supports
Multi Frequency.
Color & Doppler image controls
- Sound : This knob is to control the sound volume.
- Angle : This knob is to control the angle of Sample Volume in Spectral Doppler Mode.
- Power : Turning this knob adjusts the Acoustic Power of the system.
- Base line : This switch is to control the Baseline in Spectral Doppler Mode or Color Doppler
Mode. (including Power Doppler)
- Speed : This switch is to control the Sweep Speed in Spectral Doppler Mode or M Mode.
- Scale : This switch is to control the speed of blood flow or frequency range in Spectral
Doppler Mode.
- Sensitivity : control the Sensitivity values in Color Doppler (Power Doppler) Mode.
- Filter : control the Wall Filter values.
- Steer : This button is to alter the angle of ROI (Region of Interest) in Color Doppler Mode.
This operation is only available for Linear Probe.

TCG control : You can control the TGC values by depth using eight slides.
Measurement and Annotation Controls
- Distance : This button is to measure distance.
- Area : This button is to measure circumference and area.
- Volume : This button is to measure Volume.
- Calc : display some menus that can be possibly measured in the selected Application or Image
mode.
- Body Marker : This button is to display the list of Body markers.
- Indicator : display an arrow-shaped indicator.
- Text : This button is for a user to input a text on the image.
- Clear : This button is to delete Text, Indicator, Body Mark, and measure results on the image.

Report :This button is to display one of the Report programs, which includes measurement
results of the currently selected Application.
Setup Controls
- Probe : display a dialog box to select or change the probe and Application.
- ID[Patient] : display a Setup dialog box to setup the parameters related to the system.
- End Exam : This button is to finish the diagnosis on the current patient and reset the related
data.

Freeze : This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.

Print : This button is to print out the image viewed on the current Image Area into a
configured echo printer.

Store : The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView.
SonoView : This button is to operate SonoView, which is an Image Filing program.
SA9900 Quick Manual




Image mode & Gain control
- 2D/Gain : This knob-button is to shift the Image Mode to the 2D Mode. Since this system
includes basically 2D images except for some special cases, the user cannot turn the
2D off when desired. You can control 2D Gain by turning the knob-button.
- C/Gain : This button is to turn the Color Mode on. While the color mode is on, one more
press of this button turns it off. You can control Color Gain by turning the knob-
button.
- D/Gain : This button is to turn the PW Spectral Doppler Mode on.
- PD/Gain : This button is to turn the Power Doppler Mode on.
- M/Gain : This button is to turn the M Mode on.

Image Controls
- Focus : Up-down switch. The user can make the focusing point go up and down.
- Depth : The user can make the view angle.
- Zoom : Change the size of the Zoom box by controlling the switch. You can select images by
pressing the [Set] button.
- Freq : This button is to change the probe frequency, when you use a probe that supports
multi Frequency.
Color & Doppler image controls
- Sound : This knob is to control the sound volume
- Angle : This knob is to control the angle of Sample Volume in Spectral Doppler Mode
- Power : Turning this knob adjusts the Acoustic Power of the system.
- Base line : This switch is to control the Baseline in Spectral Doppler Mode or Color Doppler
Mode. (including Power Doppler)
- Speed : This switch is to control the Sweep Speed in Spectral Doppler Mode or M Mode
- Scale : This switch is to control the speed of blood flow or frequency range in spectral
Doppler mode.
- Sensitivity : control the Sensitivity values in Color Doppler (Power Doppler) Mode.
- Filter : control the Wall Filter values
- Steer : This button is to alter the angle of ROI (Region of Interest) in Color Doppler Mode.
This operation is only available for linear probe.

TCG control : You can control the TGC values by depth using eight slides.
Measurement and Annotation Controls
- Distance : This button is to measure distance.
- Area : This button is to measure circumference and area.
- Volume : This button is to measure Volume.
- Calc : display some menus that can be possibly measured in the selected Application or Image
Mode.
- Body Marker : This button is to display the list of Body markers.
- Indicator : display an arrow-shaped indicator.
- Text : This button is for a user to input a text on the image.
- Clear : This button is to delete Text, Indicator, Body Mark, and measure results on the image.

Utility Controls
- SonoView : This button is to operate SonoView, which is an Image Filing program.
- Report :This button is to display one of the Report programs, which includes measurement
results of the currently selected Application.
Setup Controls
- Probe : display a dialog box to select or change the probe and Application.
- Patient : display a Setup dialog box to setup the parameters related to the system.
- End Exam : This button is to finish the diagnosis on the current patient and reset the related
data.

- User1-4 : This button is to perform the functions that user has already setup for each button.

Freeze : This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.

Print : This button is to print out the image viewed on the current Image Area into a
configured echo printer.

Store : The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView.
Accuvix XQ Quick Manual


Image Mode & Gain Control
- 2D/Gain : This knob-button is to shift the Image Mode to the 2D Mode. Since this system
includes basically 2D images except for some special cases, the user cannot turn the
2D off when desired. You can control 2D Gain by turning the knob-button.
- C/Gain : This button is to turn the Color Mode on. While the color mode is on, one more press
of this button turns it off. You can control Color Gain by turning the knob-button.
- D/Gain : This button is to turn the PW Spectral Doppler Mode on
- PD/Gain : This button is to turn the Power Doppler Mode on. .
- M/Gain : This button is to turn the M Mode on.
- 3D/4D : If you press the [3D/4D] button the 3D Image Mode is on. If you press the button
again, it is off.

Setup Controls
- Patient : display a Setup dialog box to setup the parameters related to the system.
- End Exam : This button is to finish the diagnosis on the current patient and reset the related
data.
Probe : display a dialog box to select or change the probe and Application.
Image Controls
- Zoom : If you put up this switch, a Zoom box appears. Change the size of the Zoom box by
controlling the switch. You can select images by pressing the [Set] button, or exit the
Zoom mode by pressing the [Exit] button.
- Depth : If this switch is up, the depth of image is shallow. Conversely switch-down results in
a deeper image.
- Focus : The user can make the focusing point go up by putting up the switch and can make
the focusing point go down by putting it down.

Measurement : display some menus that can be possibly measured in the selected
application or image mode. Also you can select [Caliper] menu that you can
carry out basic calculations such as distance and circumference.
Annotation Controls
- Body mark : This button is to display the list of Body markers.
- Indicator : display an arrow-shaped indicator.
- Text : This button is for a user to input a text on the image.
TCG control : You can control the TGC values by depth using eight slides.

Function Controls
- E mail : With this button, you can send the last saved image of the user to the E-Mail address
input at patient data. If there is any problem in setup or connection of Network, a
warning message appears. If E-Mail is not input at patient data, press [E-Mail] button
to move to patient data screen.
- FSI : Press FSI (Full Spectrum Image) button to view the image being scanned more clearly.
The FSI function has 3 steps and appears on the screen as FSI 0.
Freeze :This button is to turn the image scanning on or off. In frozen state, Cine function,
saving, printing out and measuring the image are available.
Print : This button is to print out the image viewed on the current Image Area into a
configured echo printer.

Store : The store button saves the image viewed on the current Image Area in database.
The saved image can be controlled through SonoView ll.
Easy Quick Manual
Medison Co., Ltd.

Produced by International Clinical Application Team

August 2005