Metode Pengukuran:
Kalibrasi Energi : Sumber Point Source Ba-133, Cs-137 dan Co-60
Kalibrasi Efisiensi : Eu-152
JARAK CUPLIKAN / STANDAR KE DETEKTOR : 25 CM
0.0014000
y = 0.1726x-1.0187
0.0012000
R² = 0.9913
0.0010000
Efisiensi
0.0008000
0.0006000
0.0004000
0.0002000
0.0000000
0 200 400 600 800 1000 1200 1400 1600
Energi (kEV)
SOFTWARE : GENIE 2000
PERHITUNGAN / ANALISIS DATA : EXCELL
METODE : LANGSUNG DARI GRAFIK EFFISIENSI
PERSAMAAN : y = 0.1726x-1.0187, R² = 0.9913
AI-131 = CPSsp/ ε x γ
AI-131 = Aktivitas sampel
CPS sp = cacahan sampel
ε = efisiensi deteksi
γ = Kelimpahan sampel
KOMPONEN : SERTIFIKAT SUMBER STANDAR, UMUR PARO SUMBER
STANDAR, EFISIENSI, CACAHAN SUMBER STANDAR, YIELD SAMPEL,
CACAHAN SAMPEL DAN UMUR PARO SAMPEL
uc A u ( S st )
2
u ( St1 / 2 ) 2
u
2
u u N u ( Sp
u N Sntt
2
2
Spntt
2
t1 / 2 ) 2
Energi Prob. Unc. Prop Area Unc.area CPS Unc CPS Efisiensi Aktifitas
364.5 83.10% 0.5000% 131 12.19% 0.018 0.002 4.24E-04 51.630
KETIDAKPASTIAN GABUNGAN
Komponen Satuan Distribusi U Pembagi Vi ui ci ui ci (ui ci)2 (ui ci )4
Component Unit Distribution (%) Divider (%) (%) (%)
[1] [2] [3] [4] [5] [6] [7] [8] [9] [10] [11]
Sertifikat sumber standar normal 5.000 2 ∞ 2.500 1 2.500 6.250 39.063
Umur paro sumber standar rectangular 0.016 1.7321 ∞ 0.009 1 0.009 0.000 0.000
Efisiensi Deteksi normal 4.990 2.0000 ∞ 2.495 1 2.495 6.225 38.751
Net area sumber standar normal 4.300 2.0000 ∞ 2.150 1 2.150 4.623 21.368
Intensitas sampel rectangular 0.500 1.7321 ∞ 0.289 1 0.289 0.083 0.007
Net area sampel normal 12.190 2.0000 ∞ 6.095 1 6.095 37.149 1380.050
Umur paro sampel rectangular 0.002 1.7321 ∞ 0.001 1 0.001 0.000 0.000
Hasil Pengukuran Sampel I- 131 dengan tingkat keyakinan 2 sigma untuk PAIR 2 :
51.630 ± 7.611 Bq