SOP Pembersihan & Steril HOT WIRE FLOW SENSOR

 PEMBERSIHAN
 CELUPKAN HOT WIRE FLOW SENSOR DAN GOYANG KIRI=KANAN SELAMA 2 – 5
MENIT DALAM BEJANA BERISI AQUA DEST/AIR HANGAT SEBANYAK 400 ML
DENGAN SUHU ANTARA 20˚ - 30˚ C
 CELUPKAN KEDALAM LAR VALSURE ENZYMATIC CLEANER 2 ml CAMPURKAN
DENGAN AIR HANGAT SEBANYAK 248 ml DENGAN SUHU ANTARA 20˚ - 30˚ C
,ALTERNATIVE KALAU KESULITAN MENDAPATKAN VALSURE ENZYMATIC CLEANER
BOLEH DIGANTI PAKAI LARUTAN H202 2% 250 ml
 BILAS DENGAN AQUA DEST/AIR HANGAT BARU SEBANYAK 400 Ml ± 1 MENIT
 ULANGI LAGI JIKA MASIH ADA KOTORAN YANG MENEMPEL DI DALAM HOT FLOW
SENSOR

SOP Sterilisasi HW flow sensor PILIH SALAH SATU :
NO 1 .STERILISASI DTT
STERILISASI DTT DENGAN DISINFEKTAN
GLUTARELDEHYDE 2,4% ( LARUTAN CIDEX)
RENDAM SELAMA 25 MENIT
TERAKHIR BILAS DENGAN AQUA BIDEST
NO 2 .STERILISASI AUTOCLAVE
LAKUKAN DENGAN SUHU ANTARA 132˚C - 138˚C
DENGAN WAKTU 15 – 18 MENIT
NO 3. STERILISASI VACUUM STEAM

ATUR TEKANAN ANTARA 10 – 26 PSIG


SUHU DIATUR ANTARA 132˚C - 138˚C
LAKUKAN SELAMA 4 – 8 MENIT
PROSES PENGERINGAN 15 MENIT

Larutan aqua des/cidex


Perhatian :

1. Kalau Ada kotoran /SLEM SEGERA langsung dibersihkan


dengan aqua dest lih cara pembersihan diatas, walaupun
sedang tersambung ke pasien ( lakukan setiap 4 jam
sekali )
2.Segera bersihkan Hot wire flow sensor jika Avea telah
selesai dipakai pasien, jangan dibiarkan / ditunda
pembersihannya seperti SOP diatas
3 . Hot wire Flow sensor jangan sampai terbentur dengan
benda keras, terJatuh , atau dibersihkan dengan
menggunakan Hair dryer atau UT
4. Hot wire Flow sensor dikeringkan dalam drying box
alternative di Angin Anginkan.