&
Method Validation
1
Invention of Chromatography
M. Tswett invented chromatography method in the year of 1906 during his
research on plant pigments. He used liquid-adsorption column
chromatography with calcium carbonate as adsorbent (Stationary phase) and
pet ether as eluent (Mobile phase) to separate chlorophylls/pigments
Ether Chromatography
Chlorophyll
Colors
CaCO3
2
Classification of Chromatographic techniques depending
on type of Stationary and Mobile phases
3
HPLC
(High Pressure or High Performance Liquid Chromatography)
(Retention time)
5
Important Parameters for achieving Chromatographic Separation
1. Retention/Capacity factor ( k or k1 )
3. Resolution (R)
4. Efficiency of column
6
Retention (or Capacity) factor, k or k1
t R t0
k
Strength of detector signal
tR t0
tR: Retention time (Peak B)
t0: Non-retention time (Peak A)
t0
(t0 is Retention time of unretained compound
Peak A)
Peak A Peak B
Time
7
Separation Factor (α)
Estimates the extent of separation between adjacent peaks
• Separation factor: Ratio of k’s of two peaks A & B k2
k1 k2 k1
(k 2 k1 )
tR(A) t0
Where, k 1 For Peak A
t0
tR(B) t0
Peak A Peak B Where, k 2 For Peak B
t0
tr2-tr1 tr2-tr1
tR(B) t0
tR(A) t0
Where, K is Retention or Capacity Factor
8
Superior separation Inferior separation
Resolution, RS
t R 2 t R1
RS
1
(W1 W2 )
tR1 tR2 2
Wb1
Wb2
Superior separation
Wb1 Wb2
W1 W2
Inferior separation 9
Efficiency of Column
Theoretical Plate Number (N) & Height Equivalent To Theoretical Plate (HETP)
2
t R 2 .
N 16
tR
5 54
W W1
/2
H Wb1
Wb2
W1/2
H1/2
Superior separation
W
Wb1 Wb2
HETP = L/N
vs Inferior separation 10
Comparison of Column Efficiency
• Column efficiency is dependent on N Value (the theoretical plate number).
• Columns with high plate numbers (More N Value) are more efficient. A column
with a high N will lead to narrower peak at a given retention time than a
column with a lower N.
• Parameters influencing column efficiency:
– Column length (increasing colum length increases efficiency (increases retention of a peak)
– Particle size (decreasing particle size increases efficiency), reduces peak width of a peak.
N 16( t R
2
tR wb )
tR
wb wb
Larger N Smaller N
11
Method development: To Optimize Parameters for achieving Adequate Separation
K1 = Retention or capacity factor t R t0 N = 16 (tR/W)2
N = Theoretical pate number k
α = Separation factor t0
k2
Case Before k1
adjustment ( k 2 k1 )
12
Some common Types of HPLC
1. Ion-exchange chromatography : For Ionic samples (Anion & Cation Exchangers phase)
2. Size exclusion chromatography: Used for Larger molecules (St phase with pores)
3. Normal phase chromatography: Polar stationary phase & non-polar mobile phase
4. Reverse Phase chromatography: Non-Polar stationary phase & polar mobile phase
13
Normal Phase Vs Reverse Phase Chromatography
14
Reverse Phase Synthesis/Preparation
O O O O O O
| | | | | |
OSiOSiOSiOH OSiOSiOSiOR Where R = C8H17 or C18H37
| | | | | | Hydrocarbon chain
O O O O O O (i.e. Octyl C8, or Octadecyl,
| | | | | | C18 derivative)
OSiOSiOSiOH OSiOSiOSiOR
| | | | | |
O O O O O O
• Organic Solvents
– HPLC-grade solvents should be used (Extra pure/costlier than synthesis grade
solvents), e.g. Methanol, Acetonitrile, THF, DMSO
– Aqueous Buffers are used with some % of organic modifiers for acids and bases
(e.g. acidic buffer like Sodium or ammonium dihydrogen phosphate with pH 4 is
used to suppress the ionization of carboxylic acids to avoid peak tailing problem).
16
Mixing, Filtration, and Offline Degassing of the Eluent
Decompression
by aspirator
Ultrasonic
Machine
17
HPLC Instrumentation
UV-Visible detector
Solvent Reservoirs
Controller
Binary Pump
Autosampler
Thermostatted
Column Compartment
Detector
18
HPLC Configuration
Manual Injection
Load - Inject
1. Pump: (Transports mobile phase through system)
Isocratic elution
Gradient elution
19
UV-VIS Absorbance Detector Principle (Beer-Lambert law)
C: Concentration
Detection cell
I0 I
A
b
C
A = e·C·b = –log (I / I0)
(A : absorbance, e: absorption coefficient, C = concentration of sample,
b = path length, I = Intensity of light after passing through sample, I0 = Initial
intensity of light)
20
Making HPLC System ready for Analysis
1. Purging Mobile Phase
2. Setting Mobile phase composition/Solvent Bottles A, B, C, D
(Premixed and Online mixing)
3. Setting Column flow rate (typically 0.5 -1.0 mL/min)
4. Column oven temperature (Typically 40 0C)
5. Selecting Detection Wavelength (λmax)
6. Sample injection (Sample Volumes: 5 – 40 μL)
7. Optimizing Elution time or chromatographic run time
Purpose
After an analytical method is developed, it should be validated
to ensure that it is useful for its intended purpose.
1. Specificity/Selectivity
2. Precision (Repeatability of retention times and Peak areas)
3. Accuracy (True value Vs Measured value)
4. Linearity/System calibration
5. Range
6. Limit of detection (LOD)
7. Limit of quantification (LOQ)
8. Method Ruggedness
22
Accuracy & Precision
True value
Inaccurate &
imprecise Accurate and precise
23
Linearity studies /Calibration Curve
y/x = m, Slope i.e. Response factor
y = mx + C
Concentration Peak Area (Zero intercept, c= 0, curve passing through origin)
A1 Calibration curve
C1
A4
A2 A3
Peak area
C2
A2
A3
C3
A1
A4
C4 C1 C2 C3 C4
Concentration 24
Noise, LOD and LOQ
Limit of detection/LOD: An
amount or conc. of a sample
that produces the peak area
S/N = 3 three times the noise value
(S/N = 3)
Noise
Limit of quantification/LOQ: An
amount or conc. of a sample
that produces the peak area
ten times the noise value
(S/N = 10)
25
Maintenance/Troubleshooting of HPLC System
Sr. Care to be taken Consequences
No.
1 Always use MilliQ Water and solvents filtered through 0.2 μm filter for Clogging of column, high pressure, column
preparing mobile phases damage
2 Always degas mobile phase to expel out dissolved air. Causes erratic column flow and poor peak
shapes.
3 Always ensure that there is sufficient solvents available in the reservoir. Entry of air in the system .
4 Always prime pump (Purging) before every analysis and periodically Stagnant mobile phase in reservoir and
even when system is not in use tubing lines - may cause bacterial
growth
5 The detector should be having a good lamp in it (Periodical Checking) Low light intensity interfering absorbance,
hence low sensitivity of detector affecting
LOD and LOQ values
6 The HPLC columns shouldn’t be run at high pressures (A typical column Column bed disturbance/ de-packing/Peak
flow ranges from 0.5 to 1.0 mL/min). shape distortion
7 Don’t overload column with concentrated samples, avoid injecting large Peak distortion problems, Peak tailing or
volume samples, typical injections 5 – 40 μL fronting problems/Low resolution
8 After HPLC run is over, column should be washed with 50:50, Water: Enhances life of column and performance
MeOH or 50:50, Water: Acetonitrile and stored in a solvent as per
storage conditions specified by a column manufacturer (generally
recommended to keep it in 100% Acetonitrile)
26
27
Strong ion exchangers bear functional groups [e.g., quaternary amines or sulfonic acids]
that are always ionized. They are typically used to retain and separate weak ions. These
weak ions may be eluted by displacement with a mobile phase containing ions that are
more strongly attracted to the stationary phase sites. Alternately, weak ions may be
retained on the column, then neutralized by in situ changing the pH of the mobile phase,
causing them to lose their attraction and elute.
Weak ion exchangers [e.g., with secondary-amine or carboxylic-acid functions] may be
neutralized above or below a certain pH value and lose their ability to retain ions by charge.
When charged, they are used to retain and separate strong ions. If these ions cannot be
eluted by displacement, then the stationary phase exchange sites may be neutralized,
shutting off the ionic attraction, and permitting elution of the charged analytes.
28
Some common Types of HPLC
1. Ion-exchange chromatography : For Ionic samples (The stationary bed has an ionically
charged surface of opposite charge to the sample ions. The mobile phase is an aqueous
buffer, where both pH and ionic strength are used to control elution time.
The column is filled with material having precisely controlled pore sizes. Larger
molecules are rapidly moved through the column and the smaller molecules penetrate
inside the porous of the packing particles and elute later.
(Useful for analysis of polymers)
3. Normal phase chromatography: Polar stationary phase & non-polar mobile phase.
4. Reverse Phase chromatography: Non-Polar stationary phase & polar mobile phase