Contents
Terminology Evaluation of grid performance Lead content Grid cut off Potter bucky grid Air gap techniques
Devices used to remove scattered radiation most effectively from a given radiographic field
Invented by DR. GUSTAVE BUCKY 1913
It consists of series of lead foil strips separated by transparent strips Interspaces of grids are filled either with aluminum or some organic compound
Grid function
So most of scattered radiation is absorbed by lead strips &only a small amount passes b\t them.
Terminology
1. 2. 3. 4. Grid ratio Grid pattern Lines per inch Types of grids Linear grid Cross grid Focussed grid Parallel grid
GRID RATIO
Ratio b\t height of lead strips & the distance b\t them .
GRID RATIO r = h\D = 2\0.25 = 8 r = grid ratio h = height of lead strips D = distance b\t lead strips
Grid ratio is usually expressed as two numbers, such as 10:1 First no: - actual ratio Second no: - always 1
LINEAR GRID lead strips are parallel to each other in longitudinal axis. Adv:is to angle the x-ray tube along the length of grid with out loss of primary radiation from grid cutoff .
CROSSED GRID Two super imposed liner grids that have same focusing distance.
Grid ratio = sum of ratios of two linear grids.
Crossed grid cannot be used with oblique technique ie. angulation of tube
FOCUSSED GRID
Made up of lead strips that are slightly angulated so that they focus in space linear\crossed .
Focal distance is the perpendicular distance b\t grid & convergent point\line .
Grids have a focusing range distance within which grid can be used without loss of primary radiation.
Focusing range is fairly wide for a low grid ratio & narrow for a high grid ratio .
PARALLEL GRIDS lead strips are parallel when viewed in cross section. They are focused at infinity&are used in fluoroscopic spot film .
LINES PER INCH No: of lead strips per inch of a grid Lines/inch = 25.4/D+d. D=thickness of interspaces in mm d=thickness of lead strips in mm
The x-ray beam is collimated to a narrow pencil of radiation & the phantom is placed a great distance from grid, so no scatter radiation reaches the grid.
Two measurements are made 1, with grid in place to determine intensity of radiation transmitted through the grid
2, after removal of grid to determine the intensity of radiation directed at the grid Ratio of this gives fractional transmission, multiplied by 100 gives % of transmission.
Tp = Ip\I`p x 100 Tp = primary transmission Ip = intensity with grid Ip` = intensity with out grid .`
Bucky factor .
Ratio of incident radiation falling on grid to the transmitted radiation passing through grid
It measures total quantity of radiation absorbed from x-ray beam by grid It tells us how much patient`s exposure is increased by grid
Higher the bucky factor, greater the exposure factor & radiation dosage to the patient High ratio grid absorb more scatter radiation, so high bucky factor
CONTRAST IMPROVEMENT FACTOR Ratio between contrast with a grid to contrast with out a grid
Contrast improvement factor depends on kVp, field size & phantom thickness.
And . Contrast improvement factor is more closely related to lead content than any other factor.
It is the result of poor geometric relation b\t primary beam & lead foil strips
Cut off is complete & no primary radiation reaches the film when projected image of lead strips are thicker than the width of interspaces. Ie. IN FIGURE if C > B
Lateral decentering
There is uniform of loss of radiation over the entire surface producing a uniformly light radiograph Eg. x-ray tube being positioned lateral to the convergent line
Factors that affect the magnitude of cut off from lateral decentering of film .
1,grid ratio. 2,focal distance . 3,amount of decentering .
The central portion of film is not affected, but periphery is light The cutoff is greater with near than far focus grid distance de-centering
The loss of primary radiation is directly proportional to the grid ratio & distance from convergent line
Equation for calculating the loss of primary radiation. Near focused grid .
L = r c (1\f1 - 1\f0) x100
L = loss of primary radiation F0 = grid focusing distance F1 = target grid distance below the convergence line r = grid ratio c = distance from center of grid
Parallel grids are focused at infinity so they are always used with near focus grid distance decentering.
Combined decentering is easy to recognize It causes an uneven exposure resulting in a film that is light at the farther side & dark at the nearer side
With equal decentering error the amount of cut off is greater with combined decentering below the convergent line than with combined decentering above the convergent line .
MOVING GRIDS
Dr. Hollis E Potter
Potter bucky grid Grids are moved in order to blur out the shadow cast by lead strips seen otherwise as grid lines on the x ray Most moving grids continuously move 1 to 3cm back & forth through out the exposure.
1. Grid must move fast enough to blur its lead lines. If moved slowly either grid lines / random density variations in film are seen 2. Transverse motion of grid should be synchronous with the pulses of x-ray generator, so that the shadow of each strip is superimposed on the shadow of its neighbor
Why does the radiation dose have to increase further in moving grids than non moving types?
1. Due to lateral decentering there will be a 20% reduction in primary radiation 2. Photons are spread out uniformly on the film by a moving grid where as in non moving type, photons concentrate in between lead strips giving 15% more blackening in nonmoving type
Grid selection
8:1 grids will give adequate result below 90 kVp Above 90kVp 12:1 grids are preferable There is little decrease in transmitted radiation beyond 8:1grid & no change between 12:1&16:1, so 12:1is used .
1. The thicker the part, the more advantageous a larger gap 2. An increase in film focus distance to compensate for loss of image sharpness 3. Wider gap means more patient exposure, so increase film focus distance also 4. The first inch of air gap more important than the following inches of air gap
Used in
Chest radiography Magnification radiography
Reference :
Christensen s Basic Physics of Radiology
Thank you.